Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K192500
    Date Cleared
    2020-04-06

    (207 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to be used by medical professionals or in the measurement of systolic and diastolic pressure. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.

    Device Description

    Aneroid sphygmomanometer Model HS-20A,HS-20D,HS-201W, HS-201Y,HS-201C1, HS-201Q3/ Aneroid sphygmomanometer with stethoscope, Model HS-50C,HS-50B, HS-50D are a manual non-invasive aneroid sphygmomanometer which respectively uses an inflation cuff wrapped around the upper arm. The cuff is inflated and deflated by a manual inflation bulb. Besides a manometer (Aneroid gauge), the accessories include cuff, inflation bulb, instruction manual and carrying case. Model HS-50C,HS-50D,HS-50B also include a stethoscope. It is conjunction with stethoscope when use.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for an Aneroid Sphygmomanometer. This type of regulatory submission primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing detailed study results of the device meeting specific acceptance criteria in the same way, for example, a novel AI algorithm might. Therefore, some of the requested information, particularly around clinical studies, ground truth establishment, and expert consensus, is not directly applicable or available in this document.

    However, I can extract information related to the device's technical specifications and the standards it aims to meet for substantial equivalence.

    Here's an analysis based on the provided text, addressing the points where information is available:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this medical device, the "acceptance criteria" are primarily defined by the performance requirements of the ISO 81060-1 standard for non-invasive sphygmomanometers, and the "reported device performance" indicates compliance with these standards.

    FeatureAcceptance Criteria (from ISO 81060-1, implicitly)Reported Device Performance (as stated)
    Pressure Accuracy±3mmHg of reading±3mmHg of reading
    Monitor Scale0 to 300mmHg with a minimum interval of 2mmHg0 to 300mmHg with a minimum interval of 2mmHg
    Operating Environment10°C to 40°C (50°F to 104°F) and 15%-85% RH humidity10°C to 40°C (50°F to 104°F) and 15%-85% RH humidity
    BiocompatibilityCompliance with ISO 10993-1, ISO 10993-5, ISO 10993-10Biocompatible as requirement of ISO 10993-1, ISO 10993-5, ISO 10993-10
    Performance StandardCompatible with ISO 81060-1Compatible as requirement of ISO 81060-1

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a "sample size" for a test set in the context of a clinical study or a dataset for an algorithm. This submission references "Non-clinical testing included biocompatibility and bench testing." It does not provide details on the number of units tested or the specific methodology for these bench tests, beyond stating that they were included in the ISO 10993 and ISO 81060-1 standards.

    The data provenance is from non-clinical testing conducted to demonstrate compliance with standards like ISO 10993 and ISO 81060-1. The document does not specify country of origin for this testing data, but the manufacturer is HONSUN (NANTONG) Co., Ltd in China. The testing would be considered bench testing rather than retrospective or prospective clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. For this type of device (a mechanical aneroid sphygmomanometer), ground truth for performance claims is established through a calibrated reference standard (e.g., a manometer calibrated to a known pressure standard) during bench testing, as specified by standards like ISO 81060-1. There would be no "experts" establishing a ground truth in the context of subjective interpretation like with medical imaging.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in studies involving human interpretation of data, often in clinical trials or studies for AI algorithms. For a mechanical device's performance testing (like pressure accuracy), the measurement is objective against a standard, not subject to adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers Improving with AI vs. Without AI Assistance

    No. This device is an aneroid sphygmomanometer, which is a manual non-invasive blood pressure measurement device. It does not incorporate artificial intelligence, nor does it involve "readers" in the context of image interpretation or diagnostic decision-making. Therefore, an MRMC study or an assessment of human reader improvement with AI assistance is not relevant or applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This device is a mechanical instrument and does not contain an algorithm. "Standalone performance" in this context would refer to the device's accuracy and performance specifications as outlined in the table above, which are assessed through bench testing against established standards.

    7. The Type of Ground Truth Used

    For this device, the "ground truth" for pressure accuracy is a calibrated pressure standard as mandated by the ISO 81060-1 standard for non-invasive sphygmomanometers. This standard defines the methods for verifying the accuracy of the device against a known, precise pressure.

    8. The Sample Size for the Training Set

    Not applicable. This device is a mechanical medical instrument. It does not utilize an algorithm, machine learning, or artificial intelligence; therefore, there is no "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this mechanical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170866
    Date Cleared
    2018-02-21

    (335 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Scian Automatic Digital Blood Pressure Monitor LD-576, LD-579 are devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

    Device Description

    Scian Automatic Digital Blood Pressure Monitor, Model LD-575, LD-576 and LD-579

    AI/ML Overview

    This FDA 510(k) clearance letter for the Scian Automatic Digital Blood Pressure Monitor (Models LD-575, LD-576, and LD-579) indicates that the device has been found substantially equivalent to a legally marketed predicate device. However, the provided documents do not contain a detailed study report or specific acceptance criteria and performance data for the device itself.

    The letter confirms the device's regulatory classification, product code, and indications for use, but it does not include the information requested in your prompt regarding acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment.

    To answer your prompt, a separate study report or detailed validation document would be required, which is not present in the provided text. The FDA 510(k) process primarily assesses substantial equivalence, and while performance data is part of that assessment, the specific details you're asking for are usually in separate submissions that aren't typically summarized directly in this type of clearance letter.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1