    Device Name
    ARROW GPSCATH BALLOON DILATATION CATHETER
    Manufacturer
    HOTSPUR TECHNOLOGIES,INC SUBSIDIARY OF TELEFLEX ME
    Date Cleared
    2014-07-02

    (141 days)

    Product Code
    LIT
    Regulation Number
    870.1250
    Why did this record match?
    Applicant Name (Manufacturer) :

    HOTSPUR TECHNOLOGIES,INC SUBSIDIARY OF TELEFLEX ME

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Arrow GPSCath™ Balloon Dilatation Catheter (150 cm) is indicated for Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.
    Device Description
    The Arrow GPSCath Balloon Dilatation Catheter (150cm) is an 0.014" quidewire compatible PTA balloon catheter consisting of a semi-compliant angioplasty balloon fixed at the distal tip and a novel valve technology that allows injection of fluids through the inflation lumen without removal of the guidewire. Two radiopaque markers identify the balloon working length and two radiopaque markers indentify the valve to aid in angiographic placement. The proximal portion of the catheter includes an inflation female luer lock hub on the side of the handle and a guidewire female luer lock hub in the proximal end of the handle. Balloon inflation function is achieved using an angioplasty balloon mounted to the catheter shaft like a typical PTA balloon catheter. The fluid injection function is achieved through a valve at the proximal end of the balloon that is controlled through the device handle.
    Ask a Question

    Page 1 of 1