Search Results
Found 3 results
510(k) Data Aggregation
(239 days)
HOTSPUR TECHNOLOGIES, INC.,
The 0.018" Arrow® GPSCath™ Balloon Dilatation Catheter (150 cm) is indicated for Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.
The .018" Arrow GPSCath Balloon Dilatation Catheter (150 cm) is an 0.018" guidewire compatible PTA balloon catheter consisting of a semi-compliant angioplasty balloon fixed at the distal tip and a novel valve technology that allows iniection of fluids through the inflation lumen without removal of the quidewire. Two radiopaque markers identify the balloon working length and two radiopaque markers identify the valve to aid in angiographic placement. The proximal portion of the catheter includes an inflation female luer lock hub on the side of the handle and a quidewire female luer lock hub in the proximal end of the handle.
Balloon inflation function is achieved using an angioplasty balloon mounted to the catheter shaft like a typical PTA balloon catheter. The fluid injection function is achieved through a valve at the proximal end of the balloon that is controlled through the device handle.
This document is a 510(k) summary for the Hotspur Technologies, Inc. 0.018" Arrow GPSCath™ Balloon Dilatation Catheter (150 cm). It describes the device, its intended use, and substantial equivalence to a predicate device based on performance data.
Here's an analysis of the acceptance criteria and study information provided, structured according to your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list numerical acceptance criteria in a table format. Instead, it describes categories of in vitro preclinical tests performed and states that the product passed these tests. The reported device performance is qualitative, indicating compliance with established procedures and attributes.
Test Category | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|
Visual Testing | Product not damaged post-sterilization and packing; handle markings legible. | Performed to ensure product hasn't been damaged and handle markings are legible, and passed. |
Dimensional Testing | Product meets all dimensional performance attributes post-validated EtO sterilization cycle. | Performed to ensure product meets all dimensional performance attributes post-validated EtO sterilization cycle exposure, and passed. |
Performance Testing | Product passes all functional destructive testing post-validated EtO sterilization cycle. | Performed to ensure product passes all functional destructive testing post-validated EtO sterilization cycle exposure, and passed. |
Simulated Use Testing | Product performs and meets functional attributes in an anatomically relevant model. | Performed in an anatomically relevant model to ensure product performs and meets functional attributes, and passed. |
Biocompatibility | Meets requirements of ISO 10993-1. | Testing completed per requirements of ISO 10993-1. |
Sterilization | Achieves a sterility assurance level (SAL) of 1 x 10^-6^ for EtO sterilization. (Validation performed on predicate is applicable). | Sterilization validation performed on the predicate device is applicable due to similarity; demonstrated SAL of 1 x 10^-6^. |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not specify the exact sample sizes (number of units tested) for each type of in vitro preclinical test. It generally states that "tests were performed" or "testing was conducted" without providing quantitative details on the number of samples.
The data provenance is in vitro preclinical tests. There is no mention of human subject data, animal studies, or specific geographic origin of the testing. The tests are described as being "in accordance with the FDA's Guidance for PTCA Catheters."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not applicable as the studies described are in vitro preclinical tests and do not involve human diagnostic or clinical evaluation where expert interpretation of ground truth would be required. The "ground truth" for these tests would be the established engineering specifications, material properties, and functional performance requirements of the device.
4. Adjudication Method for the Test Set
This information is not applicable for the in vitro preclinical tests described. Adjudication methods (like 2+1, 3+1) are typically used for clinical studies involving expert disagreement on a diagnosis or outcome.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no MRMC comparative effectiveness study mentioned. The device is a physical medical device (balloon dilatation catheter), not an AI/software-as-a-medical-device (SaMD) that assists human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable as the device is a physical medical device, not an algorithm or AI.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
For the in vitro preclinical tests, the "ground truth" implicitly used would be:
- Engineering specifications and design requirements: For dimensional testing, visual inspection, and functional performance.
- Established standards: Such as ISO 10993-1 for biocompatibility and validation protocols for sterilization (ensuring SAL).
- Functional performance criteria: For simulated use testing, ensuring the device operates as intended in a model.
There is no mention of expert consensus, pathology, or outcomes data as type of ground truth for these specific tests.
8. The Sample Size for the Training Set
This information is not applicable. There is no "training set" as this is a physical medical device submission based on in vitro testing and substantial equivalence to a predicate, not an AI or machine learning model.
9. How the Ground Truth for the Training Set was Established
This information is not applicable as there is no training set for this device submission.
Ask a specific question about this device
(30 days)
HOTSPUR TECHNOLOGIES, INC.,
The Arrow GPSCath Balloon Dilatation Catheter (150 cm) is indicated for Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.
The Arrow GPSCath Balloon Dilatation Catheter (150cm) is an 0.014" guidewire compatible PTA balloon catheter consisting of a semi-compliant angioplasty balloon fixed at the distal tip and a novel valve technology that allows injection of fluids through the inflation lumen without removal of the guidewire. Two radiopaque markers identify the balloon working length and two radiopaque markers indentify the valve to aid in angiographic placement. The proximal portion of the catheter includes an inflation female luer lock hub on the side of the handle and a guidewire female luer lock hub in the proximal end of the handle.
Balloon inflation function is achieved using an angioplasty balloon mounted to the catheter shaft like a typical PTA balloon catheter. The fluid injection function is achieved through a valve at the proximal end of the balloon that is controlled through the device handle.
This document is a 510(k) summary for the Arrow GPSCath Balloon Dilatation Catheter (150 cm). It is a "Special 510(k)", which means it refers to a previously cleared predicate device (K140351 from Hotspur Technologies). The submission focuses on minor changes and an update to performance data rather than a complete re-evaluation of the device's safety and effectiveness from scratch.
Therefore, the study details you are asking for, particularly related to clinical performance, ground truth, expert adjudication, MRMC studies, or standalone algorithm performance, are not directly applicable or available within this specific 510(k) submission because it relies heavily on the predicate device's existing clearance.
However, I can extract the information that is present concerning the changes and the performance data provided for those changes.
Here's a breakdown based on the provided text, heavily emphasizing that clinical effectiveness studies for a new device are not detailed here because this is a Special 510(k) relying on a predicate:
1. A table of acceptance criteria and the reported device performance
The document mentions "additional testing of key performance data" for an update. It doesn't present a formal table of acceptance criteria and performance results in this summary. Instead, it states that the testing "result[ed] in an update to the Arrow GPSCath Balloon Dilatation Catheter (150cm) Compliance Chart and introducer sheath compatibility."
The "key performance data" tested includes:
- Balloon compliance: (No specific acceptance criteria or reported values are provided in this summary, only that it was tested and updated).
- Balloon rated burst pressure: (No specific acceptance criteria or reported values are provided in this summary, only that it was tested and updated).
- Crossing profile: (No specific acceptance criteria or reported values are provided in this summary, only that it was tested and updated).
Note: For a Class II peripheral transluminal angioplasty catheter, the FDA typically references a set of recognized standards and guidance documents (e.g., ISO, ASTM, FDA guidance for PTA balloon catheters) that would define acceptance criteria for these physical/mechanical properties. This summary implies these tests were done to those standards, but the specific details are not published here.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not specified in this summary. The testing was for "key performance data, including balloon compliance, balloon rated burst pressure, and crossing profile." These are bench/mechanical tests, not clinical studies with human subjects.
- Data Provenance: The testing was conducted by Hotspur Technologies, Inc. (a subsidiary of Teleflex Medical, Inc.) in the USA. These are laboratory/engineering tests not related to patient data.
- Retrospective or Prospective: These are laboratory tests, not clinical studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Number of Experts: Not applicable. The "ground truth" for mechanical properties like compliance, burst pressure, and crossing profile is established through validated test methods and engineering calculations, not through expert consensus in the medical sense.
- Qualifications of Experts: Not applicable for this type of testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: Not applicable. This is not a clinical study involving human assessment or interpretation for which adjudication would be required.
5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: Not applicable. This device is a balloon dilatation catheter, a physical medical device, not an AI or imaging diagnostic tool. Therefore, MRMC studies are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: Not applicable. This device is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Type of Ground Truth: For the mechanical tests (balloon compliance, burst pressure, crossing profile), the "ground truth" would be the scientifically measured and validated physical properties of the device, determined by standardized engineering testing methods and equipment. This is not medical expert consensus, pathology, or outcomes data.
8. The sample size for the training set
- Training Set Sample Size: Not applicable. This device is a physical medical device. There is no "training set" in the context of machine learning or AI.
9. How the ground truth for the training set was established
- Training Set Ground Truth Establishment: Not applicable.
Summary Explanation:
This 510(k) submission (K142300) is a "Special 510(k)" for the Arrow GPSCath Balloon Dilatation Catheter (150 cm). These submissions are used when a modification to an already cleared device does not raise new questions of safety and effectiveness and the modified device remains substantially equivalent to the predicate.
The "study" cited here refers to additional engineering and bench testing conducted to update the device's "Compliance Chart and introducer sheath compatibility" due to a change in the product (presumably a minor manufacturing update or material change that impacted these specific physical properties).
The document explicitly states: "All other performance data applicable to the subject device is the same as for the predicate device (K140351) since there has been no change to the device design, materials, manufacturing processes, or sterilization cycle." This means the original predicate device (K140351) would have had the comprehensive performance data and potentially other studies (e.g., biocompatibility, sterilization validation, aging) to demonstrate its safety and effectiveness. This current submission only addresses the specific changes mentioned.
Therefore, questions related to clinical trials, expert adjudication, AI performance, or large clinical data sets are outside the scope of this specific Special 510(k) summary.
Ask a specific question about this device
(141 days)
HOTSPUR TECHNOLOGIES,INC SUBSIDIARY OF TELEFLEX ME
The Arrow GPSCath™ Balloon Dilatation Catheter (150 cm) is indicated for Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.
The Arrow GPSCath Balloon Dilatation Catheter (150cm) is an 0.014" quidewire compatible PTA balloon catheter consisting of a semi-compliant angioplasty balloon fixed at the distal tip and a novel valve technology that allows injection of fluids through the inflation lumen without removal of the guidewire. Two radiopaque markers identify the balloon working length and two radiopaque markers indentify the valve to aid in angiographic placement. The proximal portion of the catheter includes an inflation female luer lock hub on the side of the handle and a guidewire female luer lock hub in the proximal end of the handle.
Balloon inflation function is achieved using an angioplasty balloon mounted to the catheter shaft like a typical PTA balloon catheter. The fluid injection function is achieved through a valve at the proximal end of the balloon that is controlled through the device handle.
The provided text describes the 510(k) premarket notification for the Arrow GPSCath Balloon Dilatation Catheter (150 cm). This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of its own performance against pre-defined acceptance criteria in a clinical setting. Therefore, many of the requested elements for describing an acceptance criteria study (like sample size for test sets, ground truth methodology, expert qualifications, and MRMC studies) are not applicable or explicitly stated in this type of regulatory submission.
However, based on the provided text, I can infer and extract the following:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't present a table with specific quantitative acceptance criteria or performance metrics directly from a clinical study for the Arrow GPSCath Balloon Dilatation Catheter (150 cm). Instead, it states that the device underwent in vitro preclinical tests to verify safety and effectiveness and assure substantial equivalence to a predicate device.
The "acceptance criteria" for this device, as described, are passing various functional and safety tests, demonstrating biocompatibility, and achieving a certain sterility assurance level. The "reported device performance" is the successful completion of these tests.
Acceptance Criteria Category | Reported Device Performance |
---|---|
Physical and Mechanical Performance (In vitro) | Passed all functional destructive testing post validated EtO sterilization cycle exposure. |
Packaging Integrity | Packaging can withstand the effects of transit and environmental conditions and remain intact and sterile. |
Visual Inspection | Product not damaged through sterilization cycle and packing testing; handle markings legible. |
Dimensional Accuracy | Product meets all dimensional performance attributes post validated EtO sterilization cycle exposure. |
Simulated Use Performance | Performed and met functional attributes in an anatomically relevant model; no corrosion growth. |
Flow Rate & Power Injection Capability | Tested and characterized (specific values not provided, but implies satisfactory performance). |
Biocompatibility | Passed all biocompatibility testing per ISO 10993-1. |
Sterility | Achieved a Sterility Assurance Level (SAL) of 1 x 10^-6 post EtO sterilization. |
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: The document does not specify the exact sample sizes used for each of the in vitro preclinical tests. It mentions "devices" being tested, implying a quantity greater than one, but no specific number.
- Data Provenance: The data provenance is "in vitro preclinical tests" conducted by Hotspur Technologies, Inc., a subsidiary of Teleflex Medical, Inc. The location (country of origin) is not explicitly stated but is implied to be within the US, given the submission to the FDA. The tests are prospective in the sense that they were designed and executed to evaluate the new device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts & Qualifications: Not applicable. The "ground truth" for the in vitro tests would be the established engineering and biological standards (e.g., ISO 10993-1, FDA Guidance for PTCA Catheters) and the objective measurements demonstrating compliance. There were no human experts establishing a "ground truth" in the diagnostic sense for these device performance tests.
4. Adjudication method for the test set:
- Adjudication Method: Not applicable. As these were in vitro engineering and biological tests, the data would have been analyzed against pre-defined quantitative and qualitative specifications. There's no mention of an adjudication process by human reviewers for these types of tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. This document is for a medical device (a catheter), not an AI diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Not applicable. This is a physical medical device, not an algorithm. Its performance is inherent in its design and manufacturing, tested through in vitro methods.
7. The type of ground truth used:
- Type of Ground Truth: The ground truth for these in vitro tests were established engineering specifications, regulatory standards (e.g., ISO 10993-1), and documented procedures based on FDA guidance for PTCA Catheters. For example, for biocompatibility, the ground truth is compliance with ISO 10993-1. For sterility, reaching an SAL of 1 x 10^-6.
8. The sample size for the training set:
- Sample Size for Training Set: Not applicable. This is not an AI/machine learning device. There is no "training set" in this context. The device's design is based on engineering principles and knowledge, not data training.
9. How the ground truth for the training set was established:
- How Ground Truth for Training Set Was Established: Not applicable, as there is no training set mentioned or relevant for this type of device submission.
Ask a specific question about this device
Page 1 of 1