Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K072727
    Device Name
    INRATIO 2 PT MONITORING SYSTEM
    Manufacturer
    HEMOSENSE, INC.
    Date Cleared
    2007-10-26

    (30 days)

    Product Code
    GJS
    Regulation Number
    864.7750
    Why did this record match?
    Applicant Name (Manufacturer) :

    HEMOSENSE, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The INRatio 2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio 2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio 2 PT Monitoring System is not intended to be used for screening purposes.
    Device Description
    Like the predicate device, the INRatio 2 PT Monitoring System performs a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor. disposable test strips, instruction manual, quick reference guide, training video/DVD, and testing supplies. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The Test Strips have not been modified from the design described in K020679.
    Ask a Question
    K Number
    K021923
    Device Name
    INRATIO SELF-TEST
    Manufacturer
    HEMOSENSE, INC.
    Date Cleared
    2002-10-24

    (135 days)

    Product Code
    GJS
    Regulation Number
    864.7750
    Why did this record match?
    Applicant Name (Manufacturer) :

    HEMOSENSE, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The HemoSense INRatio is an in vitro diagnostic system that provides a quantitative prothrombin time results, expressed in seconds and as an International Normalized Ratio (INR). It uses fresh capillary whole blood and is intended for is intended for quantitative prothrombin time testing of fresh, capillary whole blood for monitoring of oral anticoagulation therapy by trained patients or their caregivers, on the prescription or other order of a treating physician.
    Device Description
    The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage prothrombin time test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control tests are required for the INRatio system. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set range, it indicates a problem with the test and an error message is reported on the display instead of a result. The INRatio is the same as the INRatio, except that the labeling has been modified for readability by the lay user.
    Ask a Question
    K Number
    K020679
    Device Name
    INRATIO
    Manufacturer
    HEMOSENSE, INC.
    Date Cleared
    2002-05-06

    (63 days)

    Product Code
    GJS
    Regulation Number
    864.7750
    Why did this record match?
    Applicant Name (Manufacturer) :

    HEMOSENSE, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The HemoSense INRatio is an in vitro diagnostic system that provides a quantitative prothrombin time results, expressed in seconds and as an International Normalized Ratio (INR). It uses fresh capillary whole blood and is intended for use by health care professionals in monitoring patients who are on warfarin-type (coumarin) anticoagulation therapy. This device is not intended to be used for screening purposes.
    Device Description
    The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage prothrombin time test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control tests are required for the INRatio System. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set range, it indicates a problem with the test and an error message is reported on the display instead of a result.
    Ask a Question

    Page 1 of 1