Search Results
Found 1 results
510(k) Data Aggregation
K Number
K955164Device Name
S.A.R.S. II (SEMI-AUTOMATIC-RECONSTITUTING-SYSTEM)
Manufacturer
Date Cleared
1996-03-20
(128 days)
Product Code
Regulation Number
880.5860Why did this record match?
Applicant Name (Manufacturer) :
HABLEY MEDICAL TECHNOLOGY CORP.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Habley Medical Technology Corporation S.A.R.S. II™ (Semi-Automatic-Reconstituting-System) is intended for the mixing of an approved two component drug and for delivering the mixed drug (according to the drug labeling) to a receptacle such as an IV Bag for subsequent patient administration.
Device Description
Habley Medical Technology Corporation S.A.R.S. II™ (Semi-Automatic-The Reconstituting-System) is made up of three components: 1) a syringe adapter assembly, 2) a vial adapter assembly, and 3) a syringe.
The syringe adapter assembly attaches to the piston syringe via a Luer lock connector. The syringe adapter assembly consists of a central cavity which houses an elastomeric valve. The needle cannula is situated within the elastomeric valve and is held in place by a retaining plate. A cannula carrier and coil compression springs are also contained in the syringe adapter assembly and are engaged when the three components are coupled together. When coupled the cannula passes through and is fixed to the cannula carrier. Movement of the cannula and cannula carrier against the coil compression spring causes extension into the interior and opens end of the barrel of the conventional syringe.
The piston syringe that attaches to the syringe adapter assembly is composed of a conventional piston syringe (21 CFR 880.5860). The syringe to be used with the S.A.R.S. II™ will be purchased from a manufacturer of legally marketed syringes. The S.A.R.S. II™ is compatible with any size piston syringe with a standard Luer lock fitting.
The vial adapter assembly attaches to the syringe adapter assembly via an interlocking This consists of inwardly extending lugs positioned in the vial adapter mechanism. assembly that are aligned with outer slots of a retainer ring on the syringe adapter assembly. The interlock of the two components cause an automatic alignment of the activation pin.
The vial adapter assembly includes the threaded vial adapter, safety cup, spring arms to secure the vial, needle cannula, and elastomeric seal. The spring arms activate the piercing mechanism after the components are threaded together. The safety cup size and shape is conformable to conventional drug vials. The syringe adapter assembly includes the elastomeric valves, needle cannula, and latches for attachment to vial adapter assembly or IV Bag. The needle cannula provides a stainless steel fluid pathway between the piston syringe and the vial of an approved drug component or drug.
The Habley Medical Technology Corporation S.A.R.S. II™ (Semi-Automatic-Reconstituting-System) will be commercialized in the following configurations:
- . S.A.R.S. II™ including the vial and syringe adapter assemblies. This device will be packaged in a blister pack and provided sterile.
- . S.A.R.S. II™ including the vial assembly and syringe adapter assemblies and a piston syringe having a volume of 5, 10, 20 or 30 ml.
Ask a Question
Ask a specific question about this device