Search Results
Found 3 results
510(k) Data Aggregation
(191 days)
Sterile Syringe with Fixed Safety Needle for Single Use is intended for use in the aspiration of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately to minimize risk of accidental needlesticks.
Sterile Syringe with Fixed Safety Needle for Single Use is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of syringe and fixed needle with a safety mechanism. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
The proposed device is available in a variety combination of syringe volume and needle size.
| Syringe volume | Needle Gauge | Length | Wall type |
|---|---|---|---|
| 1ml | 21G | 5/8", 1" | TW |
| 1ml | 22G | 5/8", 1" | TW |
| 1ml | 23G | 5/8", 1" | TW |
| 1ml | 25G | 5/8", 1" | RW |
This document is a 510(k) Premarket Notification from Gemtier Medical (Shanghai) Inc. for a "Sterile Syringe with Fixed Safety Needle for Single Use." The FDA has determined the device is substantially equivalent to legally marketed predicate devices.
The document does not contain information about an AI/ML-driven medical device. The acceptance criteria and study described are for a physical medical device (syringe and safety needle), focusing on its mechanical, material, and safety performance, rather than an AI algorithm's performance on data.
Therefore, I cannot provide the detailed information requested in the prompt, as it pertains to AI/ML acceptance criteria and studies (e.g., sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types).
The document specifically mentions:
- No clinical study is included in this submission. (Page 10)
- The studies performed are non-clinical tests verifying the physical device's specifications, such as needle performance according to ISO standards, syringe performance, sharp injury protection, simulated clinical study for the safety feature, biocompatibility testing, sterility, shipping, and shelf-life. (Pages 8-9)
Because the provided text is about a physical medical device and not an AI/ML product, the questions related to AI/ML device performance and evaluation methodologies are not applicable.
Ask a specific question about this device
(197 days)
Sterile Syringe with Safety needle for Single Use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
Sterile Syringe for Single Use is a sterile luer lock syringe which is intended to be used with a hypodermic needle for the aspiration and injection of fluids for medical purpose.
Sterile Safety Needle for Single Use is intended to be used with a luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
Sterile Syringe with Safety needle for Single Use is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of syringe (piston, barrel, plunger) and hypodermic needle with a safety mechanism. The proposed device is available in a variety combination of syringe volume and needle size.
The Sterile Syringe for Single Use is intended for manual and single use only, which consists of piston, barrel and plunger. The proposed device is available in a variety syringe volume. The syringe is available in luer lock, which is intended to be connected with a hypodermic needle.
Sterile Safety Needle for Single Use is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of needle tube, retractable cartridge, jointing medium and needle hub. The proposed device is available in variety of needle gauge and needle length. According to the needle length, the safety mechanism is available in 2-part cartridge (outer retractable cartridge and inner retractable cartridge) and 3-part cartridge (outer retractable cartridge, middle retractable cartridge and inner retractable cartridge). The proposed device is compatible for use with a luer lock syringe. After withdrawal of the needle from the body, the attached needle safety mechanism can be manually activated to cover the needle immediately after use to minimize risk of accidental needlesticks.
The provided text describes a submission for K192679 (Sterile Syringe with Safety Needle for Single Use, Sterile Syringe for Single Use, Sterile Safety Needle for Single Use) to the FDA, asserting substantial equivalence to predicate device K170651.
The document discusses acceptance criteria for various tests and confirms that the proposed device met these criteria through non-clinical testing and a simulated clinical study.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
| Test Category | Specific Test / Standard | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Physical, Mechanical, Chemical | Clause 5 of ISO 7886-1:2017 | General requirements | Conforms with requirements of related standards. |
| Clause 6 of ISO 7886-1:2017 | Extraneous matter | Conforms with requirements of related standards. | |
| Clause 7 of ISO 7886-1:2017 | Lubricant | Conforms with requirements of related standards. | |
| Clause 8 of ISO 7886-1:2017 | Tolerance on graduated capacity (Syringe) | Met the requirements of ISO 7886-1. | |
| Clause 9 of ISO 7886-1:2017 | Graduated scale | Conforms with requirements of related standards. | |
| Clause 10 of ISO 7886-1:2017 | Barrel | Conforms with requirements of related standards. | |
| Clause 11 of ISO 7886-1:2017 | Piston/plunger assembly | Conforms with requirements of related standards. | |
| Clause 12 of ISO 7886-1:2017 | Nozzle | Conforms with requirements of related standards. | |
| Clause 13 of ISO 7886-1:2017 | Performance (Syringe) | Met the requirements of ISO 7886-1. | |
| Clause 4.3 of ISO 7864:2016 | Cleanliness (Needle) | Complies with ISO 7864. | |
| Clause 4.4 of ISO 7864:2016 | Limits for acidity or alkalinity (Needle) | Complies with ISO 7864. | |
| Clause 4.5 of ISO 7864:2016 | Limits for extractable metals (Needle) | Complies with ISO 7864. | |
| Clause 4.6 of ISO 7864:2016 | Size designation (Needle) | Complies with ISO 7864. | |
| Clause 4.7 of ISO 7864:2016 | Colour coding (Needle) | Complies with ISO 7864. | |
| Clause 4.8 of ISO 7864:2016 | Needle hub | Complies with ISO 7864. | |
| Clause 4.9 of ISO 7864:2016 | Needle Cap | Complies with ISO 7864. | |
| Clause 4.10 of ISO 7864:2016 | Needle tube | Complies with ISO 7864. | |
| Clause 4.11 of ISO 7864:2016 | Needle point | Complies with ISO 7864. | |
| Clause 4.12 of ISO 7864:2016 | Bond between hub and needle tube | Complies with ISO 7864. | |
| Clause 4.13 of ISO 7864:2016 | Patency of lumen | Complies with ISO 7864. | |
| Clause 5.2 of ISO 9626:2016 | Surface finish and appearance | Complies with ISO 9626. | |
| Clause 5.3 of ISO 9626:2016 | Cleanliness | Complies with ISO 9626. | |
| Clause 5.4 of ISO 9626:2016 | Limits for acidity and alkalinity | Complies with ISO 9626. | |
| Clause 5.5 of ISO 9626:2016 | Size designation | Complies with ISO 9626. | |
| Clause 5.6 of ISO 9626:2016 | Dimensions | Complies with ISO 9626. | |
| Clause 5.8 of ISO 9626:2016 | Stiffness | Complies with ISO 9626. | |
| Clause 5.9 of ISO 9626:2016 | Resistance to breakage | Complies with ISO 9626. | |
| Clause 5.10 of ISO 9626:2016 | Resistance to corrosion | Complies with ISO 9626. | |
| Sterile Barrier Packaging | ASTM F1886/F1886M-16 | Visual inspection | Maintain its integrity. |
| ASTM F88/F88-15 | Seal strength | Maintain its integrity. | |
| ASTM F1929-15 | Dye penetration test | Maintain its integrity. | |
| Sterilization and Shelf Life | ISO 10993-7:2008 | EO residue limit | Did not exceed the limit. |
| ISO 10993-7:2008 | ECH residue limit | Did not exceed the limit. | |
| USP 38-NF 33 <85> | Bacteria Endotoxin Limit (20EU/device) | Did not exceed 20EU/device. | |
| Physical, Mechanical, Chemical, Package Test on aging samples | Maintain its performance during the claimed shelf life. | Maintained its performance during the claimed shelf life. | |
| Biocompatibility | ISO 10993-1 series (Tests for blood path, indirect, <24 hours contact) | No negative impacts from materials. | No negative impacts reported for Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute Systemic Toxicity, Hemolysis, Complement activation, Thromboresistance study, Pyrogen, Particulate testing. |
| Safety Feature | FDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 and ISO 23908:2011 | Meet pre-established criteria (Simulated Clinical Study).Meet acceptance criteria (Safety Feature Test performed on both proposed and predicate). | Met the pre-established criteria. (Proposed device) Met the acceptance criteria. (Proposed device and predicate device) |
| Connect Type (Luer lock) | ISO 80369-7 | Performance of Luer lock. | Met the requirements of ISO 80369-7. |
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the numerical sample sizes for each test set. It mentions tests were performed on the "proposed device" and "aging samples" for shelf-life. In the Safety Feature Test, tests were performed on "both proposed device and predicate device".
The data provenance is from non-clinical tests conducted by Gemtier Medical (Shanghai) Inc. in China (manufacturer location). The nature of these tests (bench testing, simulated clinical study) implies they are prospective in execution, meaning they were designed and performed for the purpose of this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable as the document describes non-clinical and simulated clinical studies for a medical device (syringes and needles), not studies involving human subjects or expert assessment of medical images/data for ground truth. The acceptance criteria are based on compliance with international standards and internal design specifications, not expert consensus on medical conditions.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is not applicable. Adjudication methods are typically used in clinical studies or studies where subjective interpretations need to be resolved. The tests described here are objective performance measurements against established standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a sterile syringe and needle system, not an AI-powered diagnostic or assistive technology that would involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical medical instrument, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the device's performance is established by the international standards (e.g., ISO 7886-1, ISO 7864, ISO 9626, ISO 10993-7, ISO 23908, ISO 80369-7) and internal design specifications against which the device was tested. For biocompatibility, the lack of "negative impacts" is the ground truth. For the safety mechanism, meeting "pre-established criteria" of relevant guidance documents serves as the ground truth.
8. The sample size for the training set
This is not applicable. The document describes the testing and comparison of a physical medical device. There is no training set mentioned, as this is not an AI/ML device.
9. How the ground truth for the training set was established
This is not applicable as there is no training set for this device.
Ask a specific question about this device
(315 days)
The SAFETY BLOOD COLLECTION DEVCIE FOR SINGLE USE is a single-use, sterile, winged venipuncture needle bonded to a flexible tubing with or without a luer adapter and/or tube holder. The SAFETY BLOOD COLLECTION DEVCIE FOR SINGLE USE is used for blood collection and/or the short-term infusion of intravenous fluids (up to 2 hours under direct clinical supervision). The venipuncture needle is designed to be covered with a safety mechanism, which can be activated to cover needle immediately following venipuncture to aid in the protection against accidental needlestick injury.
The Safety Blood Collection Device for Single Use is a single-use, sterile, winged blood collection needle bonded to a flexible tubing with or without a luer adapter and/or tube holder. The Safety Blood Collection/Infusion Set is individually wrapped, sterile, non-pyrogenic and is designed for collection of blood specimens or short-term intravenous administration of fluid (up to 2 hours under direct clinical supervision). When it's equipped with male luer adapter and/ or tube holder, it can be used for taking samples from the patients through vacuum tubes. It is also indicated for short-term intravenous fluids (up to 2 hours under direct clinical supervision) after attaching to an IV set, a transfusion set or to other compatible/appropriate device.
Three models were included in this submission, which are JT-HSQ, JT-HSQ-B, and JT-HSQ-Z.
The provided text is a 510(k) Summary for a medical device called "Safety Blood Collection Device for Single Use." This document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and biocompatibility testing, rather than presenting a clinical study with acceptance criteria and comparative human reader performance as one might find for an AI/ML medical device.
Therefore, many of the requested details, such as sample size for test sets (in the context of clinical studies), data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, and training set details, are not applicable or not provided in this document. The document describes bench testing of physical and chemical properties and biocompatibility, not clinical performance for diagnostic accuracy or human-in-the-loop performance.
Here's an analysis based on the provided text, addressing what can be inferred and highlighting what is absent:
Acceptance Criteria and Device Performance (Based on Non-Clinical Testing):
The acceptance criteria are implicitly defined by the successful adherence to recognized international standards and FDA guidance documents for medical devices of this type. The reported device performance is that it met the requirements of these standards and guidance documents.
| Acceptance Criteria (Implicit from Standards and Guidance) | Reported Device Performance (as stated in document) |
|---|---|
| Biocompatibility (in accordance with ISO 10993 series): | |
| - Cytotoxicity compliance (ISO 10993-5) | Complied with ISO 10993-5 standards |
| - Skin Sensitization compliance (ISO 10993-10) | Complied with ISO 10993-10 standards |
| - Irritation compliance (ISO 10993-10) | Complied with ISO 10993-10 standards |
| - Acute Systemic Toxicity compliance (ISO 10993-11) | Complied with ISO 10993-11 standards |
| - Haemolysis compliance (ISO 10993-4) | Complied with ISO 10993-4 standards |
| - Pyrogenicity compliance (ISO 10993-4) | Complied with ISO 10993-4 standards |
| - Bacterial Endotoxins compliance (ISO 10993-4) | Complied with ISO 10993-4 standards |
| Performance Testing (Adherence to relevant ISO standards and FDA guidance): | |
| - ISO 8536-4:2010 (Infusion equipment): Particulate contamination, leakage, tensile strength, tubing appearance, flow rate, protective caps, chemical requirements, reducing (oxidizable) matter, metal ions, titration acidity or alkalinity, residue on evaporation, uv absorption of extract solution. | Met requirements of ISO 8536-4:2010 |
| - ISO 80369-7:2016 (Small-bore connectors): Dimensional requirement, gauging test, fluid leakage, sub-atmospheric pressure air leakage, stress cracking, resistance to separation from axial load, resistance to unscrewing, resistance to overriding. | Met requirements of ISO 80369-7:2016 |
| - ISO 9626:2016 (Stainless steel needle tubing): Surface finish and visual appearance, cleanliness, limits for acidity and alkalinity, size designation, dimensions, stiffness, resistance to breakage, resistance to corrosion. | Met requirements of ISO 9626:2016 |
| - ISO 7864:2016 (Sterile hypodermic needles): Cleanliness, limits for acidity and alkalinity, limits for extractable metals, size designation, colour coding, needle hub requirement, needle cap requirement, needle tube requirement, needle point requirement, bond between hub and needle tube requirement, patency of lumen, sharps injury protection. | Met requirements of ISO 7864:2016 |
| - ISO 23908:2011 (Sharps injury protection): Simulated clinical performance validation of sharp protection features. | Met requirements of ISO 23908:2011 |
| - FDA Guidance-Simulated Clinical Use Testing of Medical Devices with Sharps Injury Prevention Features: Specific requirements related to simulated clinical use of sharps injury prevention features. | Met requirements of FDA Guidance-Simulated Clinical Use Testing of Medical Devices with Sharps Injury Prevention Features. |
| Comparative Performance (Proposed vs. Predicate Device): | |
| - The proposed device meets the same standards as the predicate device for physical and chemical performance specifications, demonstrating substantial equivalence. | Both proposed and predicate device met the listed standard requirements, and testing results demonstrated substantial equivalences. |
Since this premarket notification is for a conventional medical device (safety blood collection device) and not an AI/ML-based diagnostic or imaging device, the following points are largely not applicable or not detailed in the submission.
-
Sample size used for the test set and the data provenance:
- The document mentions "test devices" for performance testing and "samples" for biocompatibility but does not specify the number of units or lot sizes tested for each standard.
- Data provenance (e.g., country of origin, retrospective/prospective) is not applicable given this is hardware testing against engineering standards, not patient data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for hardware performance and biocompatibility is established by adherence to international engineering and biological safety standards, not expert consensus on clinical data.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for non-clinical bench testing.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI/ML diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For biocompatibility: Ground truth is established by the results of standardized biological tests (e.g., cell cultures, animal irritability tests) that comply with ISO 10993 series.
- For performance: Ground truth is established by meeting the quantified specifications and thresholds defined within various ISO engineering standards (e.g., tensile strength, flow rate, dimensions).
-
The sample size for the training set: Not applicable. This is a physical medical device, not an AI/ML model that requires a training set.
-
How the ground truth for the training set was established: Not applicable, as there is no training set for a physical device.
Ask a specific question about this device
Page 1 of 1