Search Results
Found 6 results
510(k) Data Aggregation
(26 days)
GMP Medicare Sdn. Bhd.
An examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs with Gastric Acid, Xylazine and Fentanyl Citrate in accordance with ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. chemotherapy drugs; Gastric Acid; Xylazine; Fentanyl
Nitrile Powder Free Examination Glove, Blue Tested for Use with 60 Chemotherapy Drugs, Gastric Acid, Xylazine and Fentanyl
This document is a 510(k) clearance letter for a medical device: "Nitrile Powder Free Examination Glove, Blue Tested for Use with 60 Chemotherapy Drugs, Gastric Acid, Xylazine and Fentanyl". This type of device is a physical barrier and does not involve AI or imaging analysis, which are the typical contexts for the detailed acceptance criteria and study designs you've requested.
Therefore, many of the specific questions you've asked, such as those related to AI algorithm performance (e.g., sample size for training set, number of experts for ground truth, MRMC studies) are not applicable to this device.
However, I can extract the closest analogous information from the provided text about the device's performance against its stated "Indications for Use," which serves as its functional acceptance criteria.
Understanding the Device and its "Acceptance Criteria":
The device is a medical glove. Its primary "acceptance criteria" are related to its ability to prevent permeation by various substances, specifically:
- 60 chemotherapy drugs
- Gastric Acid
- Xylazine
- Fentanyl Citrate
The reported "device performance" is measured by the "Minimum Breakthrough Detection Time (Min.)" when tested according to ASTM D6978-05. The acceptance criterion implied for most tested substances is "no breakthrough detected up to 240 minutes" (i.e., >240 minutes). For two specific drugs, Carmustine and Thiotepa, the breakthrough times are explicitly lower (11 minutes and 39 minutes, respectively), and the label includes a warning that these should not be used with these drugs.
Here's the information parsed from the document, addressing your questions where applicable, and stating "Not Applicable" (N/A) for those that don't fit the context of a physical barrier device or an AI/imaging study:
1. A table of acceptance criteria and the reported device performance
For most substances, the implied acceptance criterion is "no breakthrough detected up to 240 minutes". The reported performance generally meets this, with specific exceptions noted.
Substance/Chemotherapy Drug | Concentration (mg/ml or other) | Acceptance Criterion (Based on typical performance) | Reported Device Performance (Minimum Breakthrough Detection Time - Min.) |
---|---|---|---|
Most Chemotherapy Drugs (58 out of 60 listed) | Varies (e.g., Amethopterin 25.0, Bleomycin Sulfate 15.0) | >240 minutes breakthrough time | >240 |
Carmustine | 3.3 | N/A (specific warning) | 11.0 |
Thiotepa | 10.0 | N/A (specific warning) | 39.0 |
Gastric Acid | Not specified (Simulated) | >240 minutes breakthrough time | >240 (no breakthrough detected up to 240 minutes) |
Xylazine | 100 mg/ml | >240 minutes breakthrough time | >240 (no breakthrough detected up to 240 minutes) |
Fentanyl Citrate | 100mcg/2ml | >240 minutes breakthrough time | >240 (no breakthrough detected up to 240 minutes) |
Note: The "acceptance criterion" column for Carmustine and Thiotepa is marked N/A because the device performance is clearly below the general threshold of >240 minutes, and instead, the clearance document acknowledges this lower performance and specifies a warning against their use. This means the device is cleared despite not meeting the >240 min criterion for these two specific chemicals, but with a critical warning to users.
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified in the provided text. The testing was conducted according to ASTM D6978-05, which would stipulate the required sample sizes for such tests, but the specific number used for this device is not reported.
- Data Provenance: The manufacturer is "GMP Medicare Sdn. Bhd." located in Malaysia. The testing methodology (ASTM D6978-05) is an international standard. Whether the testing was performed in Malaysia or another country is not specified, nor is whether the data is retrospective or prospective, though performance testing for device clearance is typically prospective by nature.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- N/A. This device is a physical barrier. "Ground truth" in this context is established by objective laboratory testing (permeation time using analytical methods), not by human expert interpretation (like radiologists for medical images).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- N/A. Adjudication methods like 2+1 or 3+1 are used for establishing consensus among human annotators/experts, typically in medical imaging or clinical trials. The "test set" here refers to physical samples of gloves subjected to a standardized chemical permeation test. The results are quantitative and objectively measured.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- N/A. This is a physical device, not an AI-assisted diagnostic tool. No human "readers" or AI assistance are involved in its primary function or testing for this clearance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- N/A. This is a physical device, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Objective laboratory measurement/Chemical Permeation Testing: The "ground truth" for this device's performance is the direct measurement of breakthrough time using the standardized ASTM D6978-05 method. This involves analytical detection of the permeating substance on the non-contact side of the glove.
8. The sample size for the training set
- N/A. This is a physical device. There is no "training set" in the context of machine learning. Design and development of the glove material (e.g., nitrile composition) would involve material science and manufacturing processes, but not a machine learning training set.
9. How the ground truth for the training set was established
- N/A. As above, no training set for an algorithm is involved.
Ask a specific question about this device
(219 days)
GMP Medicare SDN BHD
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
This glove was tested for use with chemotherapy drugs and Fentanyl Citrate as pet ASTM D6978-05 (Reapproved 2013), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Nitrile Powder Free Examination Glove, Blue Tested for Use with 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim
This document is a 510(k) premarket notification for a medical device: "Nitrile Powder Free Examination Glove, Blue Tested for Use with 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim." The core of the submission revolves around demonstrating the glove's resistance to permeation by chemotherapy drugs and Fentanyl.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Nitrile Powder Free Examination Glove's resistance to chemotherapy drugs and Fentanyl permeation are based on the ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The reported device performance is the "Minimum breakthrough Detection Time (min)" for each chemical.
No | Test Chemotherapy Drug | Concentration (mg/ml) | Acceptance Criteria (Implicit from context: generally, higher is better, and values over 240 mins are good, though specific pass/fail thresholds are not explicitly stated as 'acceptance criteria' in the common sense, rather the demonstration of resistance) | Reported Device Performance (Minimum breakthrough Detection Time (min)) |
---|---|---|---|---|
1 | Amethopterin | 25 | Resistance to permeation | >240 |
2 | Bleomycin Sulfate | 15 | Resistance to permeation | >240 |
3 | Busulfan | 6 | Resistance to permeation | >240 |
4 | Carboplatin | 10 | Resistance to permeation | >240 |
5 | Carmustine | 3.3 | Resistance to permeation (Note: Explicit warning given due to low permeation time) | 11.0 |
6 | Cisplatin | 1 | Resistance to permeation | >240 |
7 | Cyclophosphamide | 20 | Resistance to permeation | >240 |
8 | Cytarabine | 100 | Resistance to permeation | >240 |
9 | Dacarbazine | 10 | Resistance to permeation | >240 |
10 | Docetaxel | 10 | Resistance to permeation | >240 |
11 | Doxorubicin HCl | 2 | Resistance to permeation | >240 |
12 | Ellence | 2 | Resistance to permeation | >240 |
13 | Etoposide | 20 | Resistance to permeation | >240 |
14 | Fludarabine | 25 | Resistance to permeation | >240 |
15 | Fluorouracil | 50 | Resistance to permeation | >240 |
16 | Gemcitabine | 38 | Resistance to permeation | >240 |
17 | Idarubicin | 1 | Resistance to permeation | >240 |
18 | Ifosfamide | 50 | Resistance to permeation | >240 |
19 | Irinotecan | 20 | Resistance to permeation | >240 |
20 | Mechlorethamine | 1 | Resistance to permeation | >240 |
21 | Melphalan | 5 | Resistance to permeation | >240 |
22 | Methotrexate | 25 | Resistance to permeation | >240 |
23 | Mitomycin C | 0.5 | Resistance to permeation | >240 |
24 | Mitoxantrone | 2 | Resistance to permeation | >240 |
25 | Oxaliplatin | 5 | Resistance to permeation | >240 |
26 | Paclitaxel | 6 | Resistance to permeation | >240 |
27 | Rituximab | 10 | Resistance to permeation | >240 |
28 | Thiotepa | 10 | Resistance to permeation (Note: Explicit caution given due to low permeation time) | 39 |
29 | Topotecan | 1 | Resistance to permeation | >240 |
30 | Trisenox | 1 | Resistance to permeation | >240 |
31 | Vincristine Sulfate | 1 | Resistance to permeation | >240 |
32 | Vinorelbine | 10 | Resistance to permeation | >240 |
33 | Fentanyl Citrate | 100 mcg/2ml | Resistance to permeation | >240 |
The Study that Proves the Device Meets the Criteria:
The study referenced is the testing performed "as per ASTM D6978-05 (Reapproved 2013), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs." This standard is a recognized methodology for evaluating the integrity of medical gloves when exposed to hazardous chemicals. The table provided directly presents the results of this testing, showing the breakthrough detection times for each specified drug and Fentanyl Citrate.
Remaining Information Not Present in the Document:
The provided document (an FDA 510(k) clearance letter and Indications for Use statement) does not contain the detailed technical information required to answer the following questions:
- Sample size used for the test set and the data provenance: Not mentioned. The document only presents chemical permeation test results.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is a physical performance test for a glove, not an AI or diagnostic device requiring expert interpretation of results for "ground truth." The "ground truth" here is the direct measurement of chemical breakthrough time.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This refers to consensus building among human readers for diagnostic interpretation, which is not relevant for a glove permeation test.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for AI-assisted diagnostic devices, not protective equipment like gloves.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is for AI algorithms.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): The ground truth is established by the direct measurement of chemical permeation through the glove material according to the ASTM D6978-05 standard. It is a direct physical measurement.
- The sample size for the training set: Not applicable. This refers to the training of machine learning models, which is irrelevant for a glove permeation test.
- How the ground truth for the training set was established: Not applicable. This refers to the training of machine learning models.
Ask a specific question about this device
(363 days)
GMP Medicare SDN BHD (F25)
A patient examination glove is disposable glove intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner
Not Found
This document is an FDA 510(k) clearance letter for Nitrile Examination Powder Free Gloves. It does not contain information about an AI/ML medical device, its acceptance criteria, or a study proving that such a device meets acceptance criteria.
The information provided in the prompt, such as "A table of acceptance criteria and the reported device performance," "Sample sized used for the test set," "Number of experts used to establish the ground truth," and "If a multi reader multi case (MRMC) comparative effectiveness study was done," are all relevant to the evaluation of AI/ML-driven medical devices. However, this document is for a physical medical device (gloves) and therefore does not include any of these details.
To answer your request, I would need a document related to an AI/ML medical device, not a physical one.
Ask a specific question about this device
(232 days)
GMP MEDICARE SDN BHD (F25)
A patient examination glove is a disposable glove intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Nitrile Examination Powder Free Glove, Black; Nitrile Powder Free Glove, Orange
The provided text is a 510(k) premarket notification decision letter from the FDA for Nitrile Examination Powder Free Gloves. This document does not contain any information related to acceptance criteria, study details, or device performance for an AI/ML powered medical device.
Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as the input material does not pertain to such a device or study.
Ask a specific question about this device
(145 days)
GMP MEDICARE SDN BHD
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Nitrile Blue Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs
The provided document is a 510(k) premarket notification decision letter from the FDA regarding Nitrile Blue Powder Free Examination Gloves. This document describes the regulatory classification and general controls applicable to the device, and importantly, includes the "Indications for Use" statement and a table of chemotherapy drug permeation test results.
Therefore, the acceptance criteria and study proving the device meets the acceptance criteria are related to the performance of the gloves against permeation by chemotherapy drugs, not an AI/ML algorithm or a diagnostic device, which is what the prompt's questions are geared towards.
Given the nature of the device (examination gloves) and the content of the document, most of the questions in the prompt are not applicable. The information provided is for a physical medical device, not a software or AI-based diagnostic tool.
However, I can extract information relevant to the performance claims and their supporting study as presented in the document, reinterpreting the prompt's questions for this context.
Here's an attempt to answer the questions based only on the provided document, acknowledging that most questions are designed for AI/ML or diagnostic devices and do not fit:
Device Under Evaluation: Nitrile Blue Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs
1. A table of acceptance criteria and the reported device performance
For the specific claim "Tested for Use with Chemotherapy Drugs," the acceptance criterion is implied by the standard referenced (ASTM D6978-05) and the reported "Breakthrough Detection Time." While an explicit "acceptance criterion" value (e.g., must be > X minutes) is not stated in this document, the demonstration of high breakthrough times for most drugs indicates meeting a performance expectation set by the standard for protective gloves.
Chemotherapy Drugs | Concentration | Reported Breakthrough Detection Time in Minutes | Implicit Acceptance Criterion (as demonstrated performance) |
---|---|---|---|
Carmustine (BCNU) | 3.3mg/ml | 11.1 | Permeation Resistance Demonstrated |
Cisplatin | 1.0mg/ml | > 240 | High Permeation Resistance |
Cyclophosphamide (Cytoxan) | 20.0mg/ml | > 240 | High Permeation Resistance |
Dacarbazine (DTIC) | 10.0mg/ml | > 240 | High Permeation Resistance |
Doxorubicin Hydrochloride | 2.0mg/ml | > 240 | High Permeation Resistance |
Etoposide (Toposar) | 20.0mg/ml | > 240 | High Permeation Resistance |
Fluorouracil | 50.0mg/ml | > 240 | High Permeation Resistance |
Paclitaxel (Taxol) | 6.0mg/ml | > 240 | High Permeation Resistance |
Thiotepa | 10.0mg/ml | 40.1 | Permeation Resistance Demonstrated |
Amethopterin Hydrate | 25.0mg/ml | > 240 | High Permeation Resistance |
Methotrexate | 25.0mg/ml | > 240 | High Permeation Resistance |
Mitomycin C | 0.5mg/ml | > 240 | High Permeation Resistance |
Vincristine Sulfate | 1.0mg/ml | > 240 | High Permeation Resistance |
2. Sample size used for the test set and the data provenance
The document references "ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs." This standard would specify the sample size requirements (e.g., number of gloves tested per drug). This specific document does not state the sample size used for the permeation tests.
- Data Provenance: The tests were conducted according to a recognized ASTM standard, implying a controlled laboratory setting. The country of origin of the data is not specified, but the applicant (GMP MEDICARE SDN.BHD.) is based in Malaysia ("MY"). The testing would typically be done in a lab, which could be internal or external.
- Retrospective or Prospective: Not applicable in the context of physical product testing. These are laboratory characterization tests on manufactured samples.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to this type of device and study. The "ground truth" for chemical permeation is established by the standardized test method (ASTM D6978-05) itself, which involves specialized laboratory equipment and procedures, not human expert interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are used to resolve disagreements in expert interpretations (e.g., in medical image reading). This study involves objective chemical permeation measurements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document is about a physical product (gloves) and their chemical resistance, not an AI or diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This refers to the standalone performance of an algorithm. The "device" here is a pair of gloves. The performance (breakthrough time) is inherently "standalone" in the sense that it's a direct measurement of the glove material properties, not an assisted human process.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the performance claims would be the objective measurements obtained according to ASTM D6978-05. This standard defines how permeation is measured and what constitutes a "breakthrough detection time." It's a direct physical measurement/test outcome, not an interpretation from experts, pathology, or outcomes data.
8. The sample size for the training set
Not applicable. There is no "training set" in the context of physical medical device performance testing like chemical permeation. This concept applies to machine learning models.
9. How the ground truth for the training set was established
Not applicable, for the same reason as point 8.
Ask a specific question about this device
(258 days)
GMP MEDICARE SDN BHD (F25)
Glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Chemotherapy Drug Permeation
- The following chemicals have been tested with these gloves
Chemotherapy Drug | Concentration | Breakthrough Detection Time in Minutes |
---|---|---|
Carmustine (BCNU) | 3.3mg/ml | *10.1 |
Cisplatin | 1.0mg/ml | >240 |
Cyclophosphamide (Cytoxan) | 20.0mg/ml | >240 |
Dacarbazine (DTIC) | 10.0mg/ml | >240 |
Doxorubicin Hydrochloride | 2.0mg/ml | >240 |
Etoposide (Toposar) | 20.0mg/ml | >240 |
Fluorouracil | 50.0mg/ml | >240 |
Paclitaxel (Taxol) | 6.0mg/ml | >240 |
Thiotepa | 10.0mg/ml | *20.2 |
Amethopterin Hydrate | 25.0mg/ml | >240 |
Methotrexate | 25.0mg/ml | >240 |
Mitomycin C | 0.5mg/ml | >240 |
Vincristine Sulfate | 1.0mg/ml | >240 |
- Please note that the following drugs have extremely low permeation times:
Carmustine (BCNU): 10.1 minutes and Thiotepa: 20.2 minutes
Nitrile Blue Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs
Glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
This document describes the FDA's substantial equivalence determination for Nitrile Blue Powder Free Examination Gloves. It focuses on their safety and effectiveness, particularly when exposed to chemotherapy drugs, based on performance criteria.
Here's an analysis of the provided information, framed by your request for acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for permeation by chemotherapy drugs are not explicitly stated as numerical targets in the document. However, they are implied by the testing methodology and the presentation of "Breakthrough Detection Time in Minutes." The general acceptance would be to demonstrate a reasonable barrier against these drugs, with the specific times providing the performance metrics.
Chemotherapy Drug | Concentration | Implied Acceptance Criteria (e.g., minimum breakthrough time) | Reported Device Performance (Breakthrough Detection Time in Minutes) |
---|---|---|---|
Carmustine (BCNU) | 3.3mg/ml | (Not explicitly stated, but lower is worse) | *10.1 |
Cisplatin | 1.0mg/ml | (Not explicitly stated) | >240 |
Cyclophosphamide (Cytoxan) | 20.0mg/ml | (Not explicitly stated) | >240 |
Dacarbazine (DTIC) | 10.0mg/ml | (Not explicitly stated) | >240 |
Doxorubicin Hydrochloride | 2.0mg/ml | (Not explicitly stated) | >240 |
Etoposide (Toposar) | 20.0mg/ml | (Not explicitly stated) | >240 |
Fluorouracil | 50.0mg/ml | (Not explicitly stated) | >240 |
Paclitaxel (Taxol) | 6.0mg/ml | (Not explicitly stated) | >240 |
Thiotepa | 10.0mg/ml | (Not explicitly stated, but lower is worse) | *20.2 |
Amethopterin Hydrate | 25.0mg/ml | (Not explicitly stated) | >240 |
Methotrexate | 25.0mg/ml | (Not explicitly stated) | >240 |
Mitomycin C | 0.5mg/ml | (Not explicitly stated) | >240 |
Vincristine Sulfate | 1.0mg/ml | (Not explicitly stated) | >240 |
Note: The document highlights that Carmustine and Thiotepa had "extremely low permeation times" compared to others, indicating these might be areas of concern if specific minimum times were regulatory requirements.
2. Sample Size for the Test Set and Data Provenance
- Sample Size: The document does not explicitly state the sample size (number of gloves or tests) used for the chemotherapy drug permeation study. The table presents drug-specific results, but not the number of replicates or individual tests contributing to those results.
- Data Provenance: The study was conducted "as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs." This standard is a highly detailed, globally recognized protocol for testing these products, ensuring reproducibility and reliability of the data. The country of origin of the data is not specified beyond the Malaysian company (GMP Medicare SDN. BHD.) submitting the device. It is a prospective study as tests were conducted to specific ASTM standard.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not applicable and therefore not provided in the document. The "ground truth" for this device's performance is established by the results of physical and chemical testing (specifically, permeation testing against a standardized protocol), not by expert consensus or interpretation of images/data. The ASTM D6978-05 standard itself defines the methodology for determining this ground truth (breakthrough time).
4. Adjudication Method for the Test Set
This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical or diagnostic studies where expert opinions or interpretations are being harmonized. For material performance testing, the results are derived directly from the test procedure according to the ASTM standard, not from human adjudication of results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This device is a medical glove, and its performance for chemotherapy drug permeation is assessed through laboratory material testing, not through human reader interpretation of cases or clinical scenarios. Therefore, an MRMC study and the concept of "improving human readers with AI" are irrelevant to this submission.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. This is not an AI/algorithm-based device. Its performance is purely based on the physical barrier properties of the glove material when exposed to specific chemicals.
7. Type of Ground Truth Used
The ground truth used is based on physical and chemical testing results derived from a standardized protocol: ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. This standard uses instrumental analysis to detect when a chemotherapy drug permeates through the glove material.
8. Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device; therefore, there is no "training set" in the context of developing an algorithm. The testing described is for performance validation, not for training a model.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for an AI/ML algorithm. The "truth" for the product's performance is established through direct, standardized laboratory testing as outlined in point 7.
Ask a specific question about this device
Page 1 of 1