Search Results
Found 1 results
510(k) Data Aggregation
(699 days)
Foshan Wenjian Medical Instrument Co., Ltd.
Dental High-speed Turbine Handpiece (WJ-114, WJ-122, WJ-124, WJ-134, WJ-132, WJ-144, WJ-142, WJ -154, WJ -162, WJ -162) is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Dental Low-speed Turbine Handpiece (WJ-414, WJ-424, WJ-422) is intended for removing carious material, excess filling material, cavity and crown preparations and restorations and restorations, root canal preparations and polishing teeth.
The Dental High-speed Turbine Handpiece is the dental clinic, hospital treatment for patients with tooth disease tools, which is an instrument for drilling, grinding, repairing. It composed of handpiece and a connector, including models WJ-114, WJ-112, WJ-124, WJ-134, WJ-132, WJ-144, WJ -142, WJ -154, WJ -152, WJ -164, WJ -162.
The Dental Low-speed Handpiece is the dental clinic, hospital treatment for patients with tooth disease tools, which is an instrument for drilling, grinding, repairing. It composed of handpiece and a connector, including low speed air motor, straight handpiece and geared angle handpiece, for the model WJ-414, WJ-412, WJ-424, WJ-422
This document is a 510(k) summary for a dental handpiece, not a study report. It aims to demonstrate substantial equivalence to predicate devices, not to prove the device meets specific acceptance criteria through a clinical study. Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment is not present in this type of regulatory submission.
However, based on the provided text, I can extract information about performance tests conducted to support substantial equivalence, which implicitly serve as acceptance criteria for functionality and safety.
Here's a breakdown of the available information and why other requested details are not present:
1. A table of acceptance criteria and the reported device performance
The document doesn't present explicit "acceptance criteria" in a numerical or pass/fail table format, as one might find in a clinical study report for a diagnostic device. Instead, it relies on performance tests aligning with recognized standards to demonstrate substantial equivalence to predicate devices. The "Elements of Comparison" table (pages 6-7) serves a similar purpose by comparing the subject device's characteristics against predicate devices, implying that if the subject device meets or is comparable to the predicate's performance/specifications, it is acceptable.
Below is an attempt to create a table based on the provided "Test Summary" and "Elements of Comparison", reinterpreting compliance with standards and equivalence to predicate devices as meeting "acceptance criteria".
Acceptance Criteria Category (Implicit) | Specific Standard/Characteristic | Subject Device Performance/Status |
---|---|---|
Biocompatibility | Cytotoxicity, Sensitization, Irritation | Complied with ISO 10993-5 and ISO 10993-10 standards |
Performance (Functional) | Dental Handpiece Performance | Complied with ISO 14457:2012 standards |
Electrical Safety | Medical Electrical Equipment | Complied with IEC 60601-2-60 standards (and IEC 60601-1 for fiberoptic models) |
Sterilization & Cleaning Validation | Reprocessing Validation | Complied with ISO 17665-1, ISO 17665-2, ASTM ST79, and FDA Guidance Document "Reprocessing Medical Devices..." |
Bur Extraction Force | Minimum Force | 28N (Compared to predicate's 28N and 30N) |
Maximum Air Pressure (High-speed) | Operating Range | 200kPa ~ 250kPa (29.01 psi~ 36.25 psi) (Comparable to predicate's 177kPa ~ 301kPa) |
Maximum Air Pressure (Low-speed) | Operating Range | 245 ~ 392 kPa (35.53 psi~ 56.85 psi) (Comparable to predicate's 245 ~ 392 kPa and 36-43psi) |
Speed in RPMs (High-speed) | Operating Range | 300,000rpm ~ 400,000rpm (Comparable to predicate's 300,000-400,000rpm and 320,000-400,000rpm) |
Speed in RPMs (Low-speed) | Operating Range | 18,000 ~ 20,000 rpm (Comparable to predicate's 18,000-22,000rpm and up to 20,000rpm) |
Conformity with Shanks | Type of Shanks | Type 3 per ISO 1797-1 |
Coupling Dimensions | Standard Compliance | Complied with ISO 3964 |
Hose Connections | Type of Connections | Type 1 for 2-hole model / Type 2 for 4-hole model |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the 510(k) summary. Performance tests typically involve laboratory testing of a small number of device samples rather than human subject test sets. The document states "Performance testing was provided," but doesn't detail the number of units tested for each standard. Data provenance is also not mentioned, but given the manufacturer's location (Foshan, China), it's highly likely tests were conducted in China or by certified labs on behalf of the manufacturer. These are not "test sets" in the clinical study sense.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable and not provided. This document describes premarket notification for a Class I dental handpiece, which typically involves demonstrating substantial equivalence through engineering, performance, and biocompatibility testing against recognized standards and predicate devices, not clinical trials requiring expert-established ground truth. Expert review, if any, would be for the regulatory submission itself by the FDA, not for establishing ground truth for device performance data (which comes from lab tests).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable and not provided. Adjudication methods are relevant for clinical studies where multiple experts evaluate ambiguous outcomes. This document details laboratory and performance testing, which does not involve this kind of adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable and not provided. MRMC studies are used for diagnostic devices, particularly those involving image interpretation by human readers, often with AI assistance. This device is a dental handpiece (a surgical/treatment tool), not a diagnostic or AI-assisted interpretation system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable and not provided. This refers to AI algorithm performance. The device is a mechanical dental instrument.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
This is not applicable in the context of this device and report. For the performance tests conducted, the "ground truth" would be the specification limits defined by the relevant ISO and IEC standards (e.g., a specific bur retention force, a defined RPM range). For biocompatibility, it's compliance with established biological safety requirements.
8. The sample size for the training set
This is not applicable and not provided. "Training set" refers to data used to train machine learning algorithms. This device is a mechanical tool.
9. How the ground truth for the training set was established
This is not applicable and not provided. See point 8.
Ask a specific question about this device
Page 1 of 1