Search Results
Found 27 results
510(k) Data Aggregation
(105 days)
Esaote's Model 6450, commercial names MyLabX8 and MyLabX8 eXP, is intended to perform diagnostic general ultrasound studies including: Fetal, Abdominal, Intraoperative (Abdominal), Laparoscopic, Pediatric, Small organ, Neonatal, Neonatal Cephalic, Adult Cephalic, Transvaginal, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Urological, Cardiovascular Pediatric, Transesophageal (cardiac), Peripheral Vessel.
The equipment provides imaging for guidance of biopsy and imaging to assist in the placement of needles and catheters in vascular or other anatomical structures as well as peripheral nerve blocks in Musculosketal applications. The ultrasonic medical diagnostic equipment is intended to be connected to mechanical and electronic ultrasound probes (convex array, linear array and phased array) and Doppler probes.
The Virtual Navigator software option for Esaote 6450 system is intended to support a radiological clinical ultrasound examination (first modality) and follow percutaneous procedures or surgical operations providing additional image information from a second imaging modality (CT, MR, US and PET). The second modality provides additional security in assessing the morphology of the ultrasound image.
Virtual Navigator can be used in the following application: Abdominal, Gynecological, Musculoskeletal, Obstetrics, Pediatric, Urologic, Small Organs, Peripheral Vascular and Transcranial for radiological examinations only.
The second modality image is not intended to be used as a standalone diagnostic image since it represents information of a patient that could not be congruent with the current (actual) patient position and shall therefore always been seen as an additional source of information.
The Virtual Navigator tracking system is contraindicated for patients, personnel and other people who use an electronic life support device (such as a cardiac pacemaker or defibrillator).
Model 6450, commercial names MyLabX8 and MyLabX8 eXP, is a mainframe ultrasound system used to perform diagnostic general ultrasound studies. The primary modes of operation are: B-Mode, Tissue Enhancement Imaging (TEI), M-Mode, Multi View (MView), Doppler (both PW and CW), Color Flow Mapping (CFM), Amplitude Doppler (AD), Tissue Velocity Mapping (TVM), 3D and 4D, Qualitative Elastosonography (ElaXto) and Quantitative Elastosonography (QElaXto).
Model 6450 has the Virtual Navigator software option integrated, designed to support a radiological clinical ultrasound examination (first modality) and follow a percutaneous procedure providing additional image information from a second imaging modality (CT, MR, US and PET). The user is helped in assessing the patient anatomy by displaying the image generated by the 2nd modality.
Model 6450 is equipped with a LCD color display where acquired images and advanced image features are shown. Model 6450 control panel is equipped with a pull-out Qwerty alphanumeric keyboard that allows data entry. The touchscreen has an emulation of the Qwerty alphanumeric keyboard that allows data entry and has additional controls and mode-depending keys, integrated in the touchscreen.
Model 6450 can drive Phased Array (PA), Convex Array (LA), Linear Array (LA), Doppler and Volumetric probes.
Model 6450 is equipped with an internal Hard Disk Drive. Data can also be stored directly to external archiving media (Hard-Disk, PC, server) via a LAN/USB port.
The marketing names for Model 6450 will be MyLabX8 and MyLabX8 eXP.
The provided text does not contain specific acceptance criteria, reported device performance metrics, or detailed study information for the 6450 Ultrasound System. It is a 510(k) summary and clearance letter for a medical device, which primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria.
Therefore, most of the requested information cannot be extracted from this document, as no formal clinical study with performance criteria was conducted for this submission.
Here's a breakdown of what can be inferred or directly stated:
1. Table of Acceptance Criteria and Reported Device Performance
Not applicable. The document explicitly states "No clinical tests were performed." The submission relies on demonstrating substantial equivalence to a predicate device (Esaote 6440 - MyLab9 eXP, K173291), not on meeting specific performance acceptance criteria through a clinical study for this 510(k). The non-clinical tests (acoustic output, biocompatibility, cleaning/disinfection, thermal, electromagnetic, and mechanical safety) are stated to conform to relevant standards but no specific acceptance criteria or performance metrics are detailed in the text.
2. Sample Size Used for the Test Set and Data Provenance
Not applicable. No clinical test set was used as "No clinical tests were performed."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Not applicable. No clinical test set was used, and therefore, no expert-established ground truth for such a set was required or reported.
4. Adjudication Method for the Test Set
Not applicable. No clinical test set was used.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done
No. The document states "No clinical tests were performed," so an MRMC comparative effectiveness study was not conducted.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
Not applicable. The device is an ultrasound system with a software option (Virtual Navigator). While software verification and validation were performed, this is not a standalone algorithm performance study as typically understood for AI/ML devices. The Virtual Navigator software option is intended to support a radiological clinical ultrasound examination, implying a human (radiologist) in the loop.
7. The Type of Ground Truth Used
Not applicable in the context of a clinical performance study. For the non-clinical tests performed (acoustic output, safety, etc.), the "ground truth" is adherence to established engineering and medical device safety standards (e.g., IEC 60601-1, NEMA UD-2).
8. The Sample Size for the Training Set
Not applicable. This submission does not describe an AI/ML algorithm that would undergo training with a dataset in the typical sense. The Virtual Navigator software option is a feature for image guidance using information from a second modality, implying image registration and fusion, but not a trainable algorithm with a distinct training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As no training set is described, no ground truth establishment for a training set is mentioned.
Ask a specific question about this device
(25 days)
MyLabX6 and MyLabX7 are intended to perform diagnostic general ultrasound studies including: Fetal, Abdominal, Intraoperative (Abdominal), Laparoscopic, Pediatric, Small organs, Neonatal Cephalic, Adult Cephalic, Transvaginal, Musculoskeletal (Conventional), Musculosketal (Superficial), Urological, Cardiovascular Adult, Cardiovascular Pediatric, Transoesophageal (cardiac), Peripheral Vessel.
The equipment provides imaging for guidance of biopsy and imaging to assist in the placement of needles and catheters in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.
The ultrasonic medical diagnostic equipment is intended to mechanical and electronic ultrasound probes (convex array, linear array and phased array) and Doppler probes.
The upgraded 6400 systems, (MyLabX6 and MyLabX7), are a mainframe systems equipped with wheels allowing to move the system. The primary modes of operation are for both models: B-Mode, M-Mode, Tissue Enhancement Imaging (TEI), Multi View (MView), Doppler, Color Flow Mapping (CFM), Amplitude Doppler (AD), Tissue Velocity Mapping (TVM), 3D and 4D. Model 6400 manages Qualitative Elastosonography (ElaXto).
Model 6400 is equipped with a LCD color display where acquired images and advanced image features are shown. Model 6400 can drive Phased array, Linear array, Linear array, Doppler probes and Volumetric probes (Bi-Scan probes). The control panel is equipped with a pull-out Owerty alphanumeric keyboard that allows data entry.
Model 6400 is equipped with wireless capability.
Model 6400 is already cleared via K161359.
The marketing name for new devices of Model 6400 will be:
- MyLabX6
- . MyLabX7
MyLabX6 and MyLabX7, defined herein, combine the cleared features of 6400 system with new capabilities, listed below:
-
- Addition of Auto NT (Automatic Nuchal Translucency) option, allows to automatically capture Nuchal Translucency measurement.
-
- MicroV option, only for MyLabX7, recognizes the lowest speeds with ultra-sensitivity for small vessels and slow flow detection.
-
- QPack option (Quantification Curves), only for MyLabX7, provides capabilities to evaluate time/intensity curves of Doppler or CnTI signals within the organ under examination.
-
- 4D STIC option, only for MyLabX7, allows Time/Intensity analysis.
-
- Management of probes L3-11, L4-15, mC 3-11, P 1-5, P 2-9 and P2 5-13. P 1-5, P 2-9 and mc 3-11 are available only for MyLabX7.
-
- Full screen option.
-
- Windows 10 Operative System.
The 6400 new version is manufactured under a quality system compliance with 21CFR 820 requirements and certified according to ISO 9001:2015 and ISO 13485:2016.
This FDA 510(k) summary for the Esaote 6400 Ultrasound System (MyLabX6, MyLabX7) does not contain detailed information about specific acceptance criteria for image quality or clinical performance and the results of a study proving those criteria are met. The document focuses on establishing substantial equivalence to previously cleared predicate devices based on technological characteristics and adherence to safety standards.
Here's a breakdown of the requested information based on the provided text, and what is explicitly stated as not applicable or not provided:
1. Table of Acceptance Criteria and Reported Device Performance
This information is not provided in the document for specific image quality or clinical performance metrics. The submission focuses on substantial equivalence based on overall indications for use and technological similarity to cleared predicate devices, rather than a new performance study with specific acceptance criteria that are numerically defined. The clinical applications listed represent the intended use and are not acceptance criteria for a performance study.
2. Sample size used for the test set and the data provenance
Not applicable/Not provided. No specific clinical test set or data provenance is mentioned as this was not a clinical performance study. The submission relies on the established performance of predicate devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable/Not provided. No specific test set with ground truth established by experts is mentioned.
4. Adjudication method for the test set
Not applicable/Not provided. No specific test set or adjudication method is mentioned.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable/Not provided. The document does not describe any AI component or MRMC study. The device is an ultrasound system, not an AI-powered diagnostic tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable/Not provided. The device is an ultrasound system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable/Not provided. No specific ground truth methodology is mentioned as this was not a clinical performance study for image accuracy or diagnostic efficacy.
8. The sample size for the training set
Not applicable/Not provided. This document describes an ultrasound system, not a device trained on a dataset.
9. How the ground truth for the training set was established
Not applicable/Not provided. This document describes an ultrasound system, not a device trained on a dataset.
Summary of what is provided from the text:
The submission for the Esaote 6400 Ultrasound System (MyLabX6, MyLabX7) is a 510(k) Premarket Notification, which aims to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This type of submission generally relies on demonstrating similar technological characteristics and indications for use, and adherence to recognized performance and safety standards, rather than conducting new, elaborate clinical performance studies with specific acceptance criteria and ground truth validation for novel claims.
The document explicitly states: "No clinical tests were performed." This clarifies that the submission did not include studies with acceptance criteria related to clinical performance, a test set, expert ground truth, or MRMC studies.
The justification for substantial equivalence relies on:
- Predicate Devices: Primary predicate K161359 (6400 – MyLabSeven), and reference predicates K183191 (6420 - MyLabX5) and K173291 (6440 – MyLab9).
- Technological Characteristics: The updated devices (MyLabX6 and MyLabX7) employ the "same fundamental technological characteristics" as the predicate device (Esaote 6400 model cleared via K142008 and K161359). New capabilities (Auto NT, MicroV, QPack, 4D STIC, management of new probes, full screen option, Windows 10 OS) are stated to be "identical" to those already cleared in other Esaote models (6440 and 6420 via K173291 and K183191).
- Non-Clinical Tests: Evaluation for acoustic output, biocompatibility, cleaning and disinfection effectiveness, thermal, electrical, electromagnetic, and mechanical safety, conforming to standards such as IEC 60601-1, IEC 60601-2-37, NEMA UD-2, and NEMA UD-3, among others.
- Indications for Use: The indications for use are consistent with those of previously cleared devices.
Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are interpreted as the regulatory requirement to demonstrate substantial equivalence to legally marketed predicate devices, primarily through engineering and safety testing, and showing that any new features are identical to those already cleared.
Ask a specific question about this device
(15 days)
Esaote's 6420 is intended to perform diagnostic general ultrasound studies including: Fetal, Abdominal, Intraoperative (Abdominal), Laparoscopic, Pediatric, Small organs, Neonatal, Neonatal Cephalic, Adult Cephalic, Transvaginal, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Urological, Cardiovascular Adult, Cardiovascular Pediatric, Transoesophageal (cardiac), Peripheral Vessel.
The equipment provides imaging for guidance of biopsy and imaging to assist in the placement of needles and catheters in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.
The ultrasonic medical diagnostic equipment is intended to be connected to mechanical and electronic ultrasound probes (convex array, linear array and phase array) and Doppler probes.
The upgraded 6420 system (MyLabX5) is compact mainframe system equipped with wheels allowing to move the system. The system sizes and weights allow them to be carried using the handle.
The primary modes of operation are: 2D, M-Mode, Continuous Wave Doppler (CW), Tisue Enhancement Imaging (TEI), Multi View (MView), Doppler, Pulse Wave (PW) Doppler, Color Flow Mapping (CFM), Amplitude Doppler (AD), Tissue Velocity Mapping (TVM), 3D and 4D.
6420 is equipped with a LCD color display where acquired images and advanced image features are shown. 6420 can drive Phased array, Convex array, Linear array, Doppler probes and Volumetric probes (Bi-Scan probes). Model 6420 control panel is equipped with a pull-out Qwerty alphanumeric keyboard that allows data entry.
6420 is equipped with wireless capability.
Model 6420 was previously cleared via K161168.
The upgraded 6420 system, defined herein, combines the cleared features of both the 6440 and 6420 systems with new capabilities, listed below:
-
- Management of Strain 2D.
-
- Addition of Auto NT option.
-
- Addition Auto EF option
-
- Management of probe L 4-15.
-
- Addition of Full screen option.
-
- Operative system Windows 10
The 6420 new version is manufactured under an ISO 9001 and ISO 13485 certified quality system.
The provided document is a 510(k) summary for the Esaote 6420 Ultrasound System. It details an upgrade to an existing device, establishing substantial equivalence to previously cleared predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly present acceptance criteria in terms of quantitative performance metrics, such as sensitivity, specificity, or accuracy, for the device's diagnostic capabilities. Instead, the "acceptance criteria" are implied through establishing substantial equivalence to predicate devices and conformance to recognized safety standards. The reported device performance is therefore based on its similarity to the cleared predicates and its adherence to non-clinical safety requirements.
Acceptance Criteria (Implied by Substantial Equivalence and Standards Conformance)
| Category | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Diagnostic Features | Clinical uses (indications for use) must be substantially equivalent to previously cleared predicate devices (Esaote 6420 via K161168 and Esaote 6440 via K173291), ensuring no change in intended diagnostic capabilities for various clinical applications (Fetal, Abdominal, Pediatric, Small organs, etc.). | The upgraded 6420 system combines cleared features of both the 6440 and 6420 systems. The clinical uses are stated to have not changed from the previously cleared K161168 and K173291 devices. Specific new capabilities added (Strain 2D, Auto NT, Auto EF, L 4-15 probe management, Full screen option, Windows 10 OS) are stated to be identical to those found in the Esaote 6440 (K173291). Many tables list "P" (Previously cleared by FDA) for various clinical applications and modes on the 6420 and associated probes, reinforcing that the device performance for these applications aligns with prior clearances. Examples include various imaging modes (B, M, PWD, Color Doppler, etc.) for applications like Abdominal, Pediatric, Small Organs, Peripheral Vascular, Musculo-skeletal. |
| Technological Characteristics | Fundamental technological characteristics must be the same as predicate devices. New features must be identical to those cleared in predicate devices. | The upgraded 6420 employs the "same fundamental technological characteristics" as its predicate devices. The added features (Strain 2D, Auto NT option, Auto EF option, L 4-15 probe management, Full screen option, Windows 10 OS) are explicitly stated to be "identical to those of Esaote 6440 model, cleared via K173291." |
| Safety and Performance Standards | Conformance to recognized medical device safety standards, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electromagnetic, and mechanical safety. | The upgraded 6420 system was evaluated and found to conform to: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD-3, and NEMA UD-2. The device is manufactured under an ISO 9001 and ISO 13485 certified quality system. |
2. Sample Size Used for the Test Set and Data Provenance:
The document explicitly states: "No clinical tests were performed." Therefore, there is no sample size for a test set and no data provenance from clinical studies. The demonstration of performance relies on substantial equivalence to existing devices and non-clinical evaluations.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Since "No clinical tests were performed," there were no experts used to establish ground truth for a clinical test set.
4. Adjudication Method for the Test Set:
Not applicable, as "No clinical tests were performed."
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable, as "No clinical tests were performed" and this device is an ultrasound system, not an AI-assisted diagnostic tool as described. The document does not mention any AI components or human-in-the-loop performance studies.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable, as this is an ultrasound imaging system, not a standalone algorithm. The performance evaluation is based on the characteristics of the hardware and software features being substantially equivalent to predicates.
7. The type of ground truth used:
Given the statement "No clinical tests were performed," there was no clinical ground truth established through expert consensus, pathology, or outcomes data. The "ground truth" for this 510(k) submission is effectively the established performance and safety profiles of the predicate devices. The submission asserts that the upgraded device's features and overall performance are analogous to these already cleared devices, and non-clinical testing confirms adherence to relevant technical standards.
8. The sample size for the training set:
Not applicable. As "No clinical tests were performed" and the device is not an AI/ML algorithm that requires a training set in the conventional sense, there is no mention of a training set or its size. The upgrades primarily involve hardware and software features already present in other cleared devices.
9. How the ground truth for the training set was established:
Not applicable for the same reasons as #8.
Ask a specific question about this device
(224 days)
G-scan Brio is a Magnetic Resonance (MR) system that produces transversal, sagittal, coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip, imaging the temporomandibular joint and imaging the cervical, the thoracic and the lumbosacral sections as portions of the spinal column.
G-scan Brio images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in use of MR equipment, the images can provide diagnostically useful information.
The changes performed to G-scan Brio, with respect to the cleared version – G-scan Brio K133490 – are due to the improvement of the system performance. These modifications, which do not affect the intended use or alter the fundamental scientific technology of the device, are the following:
- A new Bilateral TMJ Coil
- Introduction of the DPA Lumbar spine coil n° 10 with a new extra-large flexible section
- Introduction of the thoracic spine section examination
- A new software version including the following features:
- o Customization of Image Enhancement
- o Overlay sending to PACS
- o Isotropic 3D acquisition
The provided document is a 510(k) summary for the G-scan Brio, a Magnetic Resonance (MR) system. It describes modifications made to an already cleared device (K133490) to improve system performance.
Based on the content, the document explicitly states that "No clinical tests were performed." This means there is no study described within this document that demonstrates the device meets acceptance criteria through clinical performance.
Therefore, most of the requested information cannot be provided from this particular document.
Here's what can be inferred or stated based on the document:
1. A table of acceptance criteria and the reported device performance
- Acceptance Criteria: Not explicitly stated in terms of clinical performance metrics. The submission focuses on adherence to safety standards and technical characteristic equivalence.
- Reported Device Performance: No clinical performance metrics are reported as no clinical tests were performed. The device's "performance" is demonstrated through non-clinical testing for compliance with safety and performance standards.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable, as no clinical tests were performed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable, as no clinical tests were performed.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable, as no clinical tests were performed.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study was done, as no clinical tests were performed. The device described is an MR system, not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No standalone performance study was done, as no clinical tests were performed. The device is a diagnostic imaging system, designed to be interpreted by medical experts.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable, as no clinical tests were performed.
8. The sample size for the training set
- Not applicable, as no clinical tests were performed.
9. How the ground truth for the training set was established
- Not applicable, as no clinical tests were performed.
Summary of Non-Clinical Tests (as provided in the document):
The document details that the G-scan Brio was evaluated through non-clinical tests to demonstrate substantial equivalence to its predicate device. These tests focused on:
- Medical electrical equipment safety
- Risk management
- Software verification
- Image quality
The device was found to conform to the following medical device safety standards:
- IEC 60601-1
- IEC 60601-1-2
- IEC 60601-1-6
- IEC 60601-2-33
- ISO 14971
- ISO 62304
- IEC 62366
- NEMA MS-1
- NEMA MS-3
The conclusion states that "The non-clinical testing demonstrates that the G-scan Brio is as safe, as effective, and performs as well as or better than the predicate. G-scan Brio is substantially equivalent to the legally marketed devices to applicable medical device safety and performance standards."
In essence, the device's acceptance criteria in this submission are met by demonstrating compliance with recognized safety and performance standards through non-clinical testing, rather than through a clinical performance study.
Ask a specific question about this device
(43 days)
Esaote's Model 7410 is a compact ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonalic, Adult Cephalic, Small organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transrectal, Pediatric and Other: Urologic. The 7410 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.
Model 7410 is a portable system equipped with a handle. The system size and weight allow it to be carried using its handle. The primary modes of operations are: B-Mode, M-Mode, Tissue Enhancement Imaging (TEI), Multi View (MView), Doppler, Color Flow Mapping (CFM), Amplitude Doppler (AD), Tissue Velocity Mapping (TVM), 3D and 4D. 7410 is equipped with a LCD color display where acquired images and advanced image features are shown. 7410 system can drive Phased, Convex, Linear array, Doppler probes and Volumetric probes (Bi-Scan probes). On 7410 system the touchscreen has an emulation of the Qwerty alphanumeric keyboard that allows data entry. 7410 system is equipped with wireless capability. Model 7410 has been designed to be powered by battery.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Esaote 7410 Ultrasound System, presented in the requested format:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document is a 510(k) summary for an ultrasound system, which primarily demonstrates substantial equivalence to predicate devices rather than establishing novel performance acceptance criteria or conducting a dedicated performance study against new criteria. Therefore, the "acceptance criteria" here are implicitly the performance specifications and safety standards met by the predicate devices. The "reported device performance" is the assertion that the new device (7410 System) meets these same technological characteristics and standards.
| Acceptance Criteria (Implicit from Predicate Devices & Standards) | Reported Device Performance (Esaote 7410 Ultrasound System) |
|---|---|
| Technological Characteristics- Equivalent fundamental technological characteristics to predicate devices (Esaote 7400, Esaote Europe 6420)- Equivalent clinical uses to predicate devices- Compliance with IEC60601-1- Compliance with IEC60601-1 and IEC60601-2-37 safety requirements- Acoustic Output Display feature per AIUM / NEMA standards (equivalent Ispta and MI maximal values)- Similar measurements and analysis package to predicate devices- Digital storage capabilities, including network connectivity- Wireless capability- Battery powering capability | - The 7410 system employs the same fundamental technological characteristics as its predicate devices (Esaote 7400 via K111032, K132331, K132466; and Esaote Europe 6420 via K141486).- Clinical uses for the 7410 system are equivalent to those of the Esaote 7400 model and Esaote Europe 6420.- Designed to meet IEC60601-1.- Designed to meet IEC60601-1 and IEC60601-2-37 safety requirements.- Provides an Acoustic Output Display feature per AIUM / NEMA standards, with equivalent Ispta and MI maximal values.- Provides similar measurements and analysis package.- Has digital storage capabilities, including network connectivity.- Equipped with wireless capability, substantially equivalent to 7400 cleared for wireless capability.- Designed to be powered by battery, substantially equivalent to 7400 cleared for battery powering. |
| Safety Standards / Non-Clinical Tests- Acoustic output evaluation- Biocompatibility evaluation- Cleaning and disinfection effectiveness evaluation- Thermal, electromagnetic, and mechanical safety evaluation- Conformity to IEC 60601-1- Conformity to IEC 60601-1-2- Conformity to IEC 60601-2-37- Conformity to NEMA UD-3- Conformity to NEMA UD-2 | - The device was evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, as well as thermal, electromagnetic, and mechanical safety.- Found to conform to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD-3, and NEMA UD-2.- The system is manufactured under an ISO 9001:2000 and ISO 13485 certified quality system. |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "No clinical tests were performed." This is a 510(k) submission based on substantial equivalence to predicate devices and non-clinical engineering/safety testing, not a performance study involving a test set of patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
Not applicable, as no clinical tests were performed and thus no ground truth was established by experts for a test set.
4. Adjudication Method for the Test Set
Not applicable, as no clinical tests were performed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/CADe/CADx device and no MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is an ultrasound imaging system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable, as no clinical performance study was conducted requiring ground truth. The basis for clearance is substantial equivalence to legally marketed predicate devices and conformity to recognized safety standards.
8. The sample size for the training set
Not applicable, as no algorithm requiring a training set was developed or evaluated.
9. How the ground truth for the training set was established
Not applicable, as no algorithm requiring a training set was developed or evaluated.
Summary of the Document's Nature:
This document is a 510(k) Pre-market Notification for the Esaote 7410 Ultrasound System. Its primary purpose is to demonstrate substantial equivalence to already legally marketed predicate ultrasound devices (Esaote 7400 and Esaote Europe 6420). This means the new device shares the same intended use and fundamental technological characteristics, and any differences do not raise new questions of safety or effectiveness. The "studies" performed are non-clinical engineering and safety tests to ensure the device meets recognized industry standards for medical electrical equipment and diagnostic ultrasound. No new clinical performance studies were conducted to establish specific accuracy, sensitivity, or specificity metrics for diagnostic image interpretation by the device itself.
Ask a specific question about this device
(168 days)
S-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip, imaging of the temporomandibular joint and imaging the cervical spine and the lumbar spine sections as portions of the spinal column.
S-scan images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
The changes performed on the modified S-scan device, with respect to the cleared version - Sscan K121650 - are due to the improvement of the system performance. These modifications, which do not affect the intended use or alter the fundamental scientific technology of the device, are the following:
-
- Upgrading of the electronics
-
- Patient bed unlocking function
-
- A new software version
This document is a 510(k) summary for the S-scan Magnetic Resonance (MR) system from Esaote S.p.A. It describes a modified version of the S-scan device compared to its cleared predicate (K121650).
Here's an analysis of the provided text regarding acceptance criteria and the study:
Important Note: The provided 510(k) summary is for a modified device, and the changes are described as "improvement of the system performance" due to "Upgrading of the electronics," "Patient bed unlocking function," and "A new software version." Crucially, it states that these modifications "do not affect the intended use or alter the fundamental scientific technology of the device." As a result, this 510(k) relies on the substantial equivalence to the predicate device (K121650) and does not describe a new study with explicit acceptance criteria for a novel AI/software feature with reported device performance results.
Therefore, I cannot directly extract "acceptance criteria and the study that proves the device meets the acceptance criteria" in the typical sense of a new AI algorithm's performance study from this particular document. The document essentially states that because the fundamental scientific technology and intended use are unchanged, a new major performance study against specific acceptance criteria for the new components is not required to demonstrate substantial equivalence for the overall device.
However, I can still extract information related to the device and its intended function, which would implicitly encompass its performance requirements.
1. Table of acceptance criteria and the reported device performance
Based on the provided text, there are no explicit acceptance criteria or reported device performance metrics stated for the modified S-scan's new software/electronics within this 510(k) summary. This is because the submission is a "Special 510(k)" relying on the fundamental equivalence to the predicate, K121650, where performance was presumably established. The modifications are described as improvements that do not alter the fundamental scientific technology or intended use.
The intended function of the device is to produce diagnostic MR images. Therefore, the implicit acceptance criterion is that the modified device must continue to produce diagnostically useful images of equivalent or improved quality compared to the predicate device, as determined by a medical expert.
| Acceptance Criteria (Implicit for MRI Systems) | Reported Device Performance (Not explicitly stated in this 510(k) for the modifications) |
|---|---|
| Produce transversal, sagittal, coronal, and oblique cross-section images of specified body parts (limbs, joints, spinal column). | The device continues to perform its intended function of producing these image types. |
| Images correspond to spatial distribution of protons, determining MR properties (T1, T2, nuclei density, flow velocity, chemical shift). | The device's fundamental scientific technology in producing these images remains unchanged. |
| Images provide diagnostically useful information when interpreted by a medical expert trained in MR. | Implied: The diagnostic utility is maintained or improved due to performance upgrades, but no specific metrics are provided in this summary. |
| Maintain or improve image quality characteristics (e.g., signal-to-noise ratio, spatial resolution, contrast) previously established for the predicate device. | Implied: The "improvement of the system performance" via electronics and software upgrades aims to enhance these aspects, but no quantitative results are given. |
| Device functions safely and effectively. | Implied: The modifications are cleared as substantially equivalent, suggesting safety and effectiveness are maintained. |
Regarding the study proving the device meets acceptance criteria:
The document doesn't describe a specific clinical study for the modified S-scan to prove it meets new or modified acceptance criteria as if it were a novel AI algorithm. Instead, the "proof" is based on the argument of substantial equivalence to the predicate device (K121650). The modifications are considered "improvements" that do not alter the fundamental scientific technology. This type of submission (a Special 510(k)) typically relies on verification and validation of the engineering changes to ensure the device continues to meet its established performance specifications, rather than a full-scale clinical trial to re-establish diagnostic accuracy.
Therefore, for aspects 2-9, the answer is largely that this information is not applicable to this specific 510(k) summary because it doesn't describe a new performance study to establish primary diagnostic performance of a new algorithm but rather an upgrade to an existing MR imaging system.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable. This 510(k) does not describe a new performance study with a test set of images for diagnostic evaluation by a new algorithm. The "test set" for the system itself would be part of the predicate device's original clearance (K121650) and internal engineering verification and validation testing for the upgrades.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. See point 2.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. See point 2.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This 510(k) is for a conventional MR imaging system, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This 510(k) is for an MR imaging system; it does not involve a standalone algorithm for diagnostic interpretation. The device's output (images) requires interpretation by a medical expert. ("When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.")
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Not applicable for this 510(k) submission. For the original predicate device (K121650), the ground truth for validating MR image quality would typically involve various phantoms, physical measurements, and potentially clinical correlation with other imaging modalities or pathologies if clinical validation was performed.
8. The sample size for the training set
- Not applicable. This device is an MR imaging system; it is not described as involving a machine learning algorithm with a "training set" in this context. The "new software version" likely refers to operational software for the MRI, not an AI diagnostic algorithm.
9. How the ground truth for the training set was established
- Not applicable. See point 8.
Ask a specific question about this device
(132 days)
Esaote's Model 6400 is a mainframe ultrasound system diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric and Other: Urologic. The 6400 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.
Esaote's Model 7400 is a compact ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric and Other: Urologic. The 7400 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.
Model 6400 is a mainframe system equipped with wheels allowing them to move the system. Model 7400 is a portable system equipped with a handle. The system size and weight allow it to be carried using its handle. The primary modes of operation are for both models: B-Mode, M-Mode, Tissue Enhancement Imaging (TEI), XView, Multi View (MView), Trapezoidal View (TPView), Doppler, Color Flow Mapping (CFM), Amplitude Doppler (AD) and Tissue Velocity Mapping (TVM). Both 6400 and 7400 are equipped with a LCD color display where acquired images and advanced image features are shown. Both 6400 and 7400 can drive Phased, Convex, Linear array and Doppler probes. 6400 control panel is equipped with a pull-out qwerty alphanumeric keyboard that allows data entry. On 7400 model the touchscreen has an emulation of the qwerty alphanumeric keyboard that allows data entry. Both 6400 and 7400 models are equipped with wireless capability. Model 7400 was designed to be powered by battery.
Both 6400 and 7400 have been cleared via K111302.
6400 and 7400 Upgrade, defined herein, combine the cleared features of both 6400 and 7400 systems with other new software capabilities, listed below:
-
- Management of motorized probes (Bi-Scan probes) that allow volumetric acquisition.
-
- Management of 3D/4D mode of operation.
-
- Management of the acquisition of consecutive ultrasound images on extended surface. The final image is composed of the consecutive frames placed side (Panoramic View or VPan).
The 6400 and 7400 Upgrades are manufactured under an ISO 9001:2000 and ISO 13485 certified quality system.
This 510(k) summary for the Esaote 6400 and 7400 Ultrasound Systems (K132231) indicates that no clinical tests were performed to establish acceptance criteria or demonstrate device performance.
Instead, the submission relies on demonstrating substantial equivalence to previously cleared predicate devices and conformity to applicable medical device safety and performance standards.
Therefore, most of the requested information regarding acceptance criteria and performance studies, if they were based on clinical data, cannot be extracted from this document as such studies were explicitly stated as not performed.
However, I can extract information related to the device's technological characteristics and the non-clinical tests performed.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance:
Since no clinical studies were performed, there are no specific performance metrics (e.g., sensitivity, specificity, accuracy) or quantitative acceptance criteria for diagnostic capability that can be reported from this document.
The acceptance criteria are implicitly based on:
- Substantial Equivalence: The new features and devices (6400 and 7400 Upgrades) are deemed substantially equivalent to previously cleared predicate devices (Esaote 6200, 7340, 6400, 7400 models). This means their performance is expected to be no worse than the predicate devices, which were already deemed safe and effective.
- Conformity to Safety Standards: The devices meet various international and national safety standards.
| Acceptance Criteria Category | Specific Criteria/Standard | Reported Device Performance |
|---|---|---|
| Technological Equivalence | Clinical Uses | Equivalent to predicate devices (K111302, K100931, K081794, K091009, K110688) |
| Bi-Scan probe management | Equivalent to predicate devices | |
| 3D/4D mode of operation | Equivalent to predicate devices | |
| Panoramic View (VPan) | Equivalent to predicate devices | |
| Safety Standards | IEC 60601-1 | Conforms to standard |
| IEC 60601-1-2 | Conforms to standard | |
| IEC 60601-2-37 | Conforms to standard | |
| NEMA UD-3 (Acoustic Output Display) | Conforms to standard, equivalent Ispta and MI maximal values to predicate | |
| NEMA UD-2 (Acoustic Output Measurement) | Conforms to standard | |
| Non-Clinical Tests | Acoustic Output | Evaluated and conforms |
| Biocompatibility | Evaluated and conforms | |
| Cleaning & Disinfection Effectiveness | Evaluated and conforms | |
| Thermal Safety | Evaluated and conforms | |
| Electromagnetic Safety | Evaluated and conforms | |
| Mechanical Safety | Evaluated and conforms |
2. Sample Size Used for the Test Set and Data Provenance:
- Not Applicable. No clinical test set data was used as "No clinical tests were performed." The submission relies on comparative data with predicate devices and non-clinical engineering tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Not Applicable. No clinical tests were performed, so no expert ground truth for a clinical test set was established as part of this submission.
4. Adjudication Method for the Test Set:
- Not Applicable. No clinical tests were performed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This device is an ultrasound imaging system, not an AI-assisted diagnostic tool in the sense of software interpreting images for human readers. No MRMC comparative effectiveness study was done for AI assistance. The submission focuses on hardware and software feature upgrades to existing ultrasound systems.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is an ultrasound imaging system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not Applicable. For the purpose of establishing clinical performance, no ground truth data from clinical trials was used, as no clinical trials were performed. The "ground truth" for substantial equivalence is based on the cleared status and established safety and effectiveness of the predicate devices and conformity to engineering standards.
8. The sample size for the training set:
- Not Applicable. No clinical data was used for training purposes, as no clinical tests or AI/machine learning components requiring such training are described.
9. How the ground truth for the training set was established:
- Not Applicable. No clinical training set was used.
Ask a specific question about this device
(232 days)
OrthoCAD is an option that provides the morphometry of the lumbo-sacral section of the spine, by means of semi-automatic segmentation of MR images, the generation of the relative 3D model and calculation of the significant geometrical properties of the vertebral bodies and spinal canal. When this data is interpreted by a trained physician, it can yield information that may assist diagnosis.
The OrthoCAD software option is a software package intended to be used with Esaote Gscan system cleared via K111803. OrthoCAD provides the morphometry of the lumbosacral section of the spine, by means of semi-automatic segmentation of MR images, the generation of the relative 3D model and calculation of the significant geometrical properties of the vertebral bodies and spinal canal. When this data is interpreted by a trained physician, it can yield information that may assist diagnosis.
G-scan is a Magnetic Resonance (MR) system, which produces images of the internal structures of the patient's limbs and joints.
The OrthoCAD system allows you to visualize, analyse and compare Magnetic Resonance images. The system is connected to a database that enables the physician to keep track of all the patients examined over time.
The images are acquired by running FSE T2 Rel sequences on the G-scan of the lumbosacral tract of the vertebral spine, in the sagittal plane, and are transferred to the OrthoCAD database following acquisition.
When the MR images are stored on the OrthoCAD database, the user can proceed with a manual or semi-automatic (wizard) segmentation of the vertebral bodies (from L1 up to S1) and of the spinal canal. During the segmentation of the anatomical elements, 3D models are constructed based on the segmented structures.
When the segmentation procedure has been terminated, the user can proceed with the evaluation of the following significant clinical parameters:
- . Vertebral wedging
- Listhesis index .
- Intervertebral translation index .
- Intervertebral angles .
- Vertebral collapse index .
- . Spinal curvature
- . Spinal canal thickness
- . Spinal canal section
- Foramen area . .
Following this process, the endoscopic virtual navigation within the segmented spinal canal is enabled.
Finally, if the user has worked on MR images acquired both in the clinostatic and orthostatic mode, the measures calculated and the virtual navigation of the two the positions can be compared, and a report containing all the information is produced.
OrthoCAD is made up of six environments:
- . Patient Management: contains the functions required for the display and management of patients stored in the database associated with the system.
- Home: keeps track of the procedures executed overtime for the selected patient . (analyses present, status of examinations associated with the various analyses, etc.).
- . Segmentation: carries out the functions used for the segmentation and those related to the construction of 3D models of anatomic elements
- Measurements: includes all tools required to measure the clinical parameters . used for the analysis of the currently selected exam.
- Navigation: enables endoscopic virtual navigation within the segmented . anatomical structures by means of the definition of anatomic points, in order to construct one (or more) navigation routes.
- . Comparison: enables the comparison of two different examinations within the same analysis or within different analyses provided they are the same type. This environment enables:
- . The simultaneous display, or superimposed display when required, of anatomical elements which belong to the two volumes being compared.
- 미 The simultaneous display of the different measurements, with an indication of the main differences between these values.
Here's an analysis of the provided 510(k) summary for the OrthoCAD software option, detailing the acceptance criteria and study information:
1. Acceptance Criteria and Reported Device Performance
The provided document does not explicitly state quantitative acceptance criteria (e.g., a specific percentage accuracy or precision threshold). Instead, it describes a series of comparative tests that were performed to demonstrate the device's performance against manual methods. The reported performance indicates that these tests were conducted, and the summary's conclusion is that the device "met performance requirements and is as safe and effective as the predicate devices."
Based on the "Tests performed" section, the implicit acceptance criteria are related to the consistency, variability, repeatability, reproducibility, and correctness of measurements when compared to manual segmentation and measurement.
| Acceptance Criteria (Inferred from "Tests performed") | Reported Device Performance |
|---|---|
| Comparison between manual and semi-automated segmentation on lumbar and first sacral vertebrae. | Tests performed; device "met performance requirements." |
| Comparison between manual and semi-automated segmentation on spinal canal. | Tests performed; device "met performance requirements." |
| Comparison of manual and semi-automated measurements and evaluation of variability, repeatability, and reproducibility. | Tests performed; device "met performance requirements." |
| Validation of new software OrthoCAD in its correctness in measuring MRI images of the Lumbar spine. | Tests performed; device "met performance requirements." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size used for the test set (number of images, cases, or segmented structures).
The data provenance is implicitly from MRI images of the lumbosacral tract of the vertebral spine acquired using an Esaote G-scan system. The country of origin of the data is not explicitly stated but can be inferred to be associated with Esaote S.p.A., an Italian company. The document does not specify if the data was retrospective or prospective.
3. Number of Experts Used to Establish Ground Truth and Qualifications
The document does not explicitly state the number of experts used to establish the ground truth for the test set, nor does it detail their specific qualifications (e.g., radiologist with X years of experience). It refers to the "manual segmentation" as the comparative method, implying that experts performed this manual segmentation, but the specifics are absent.
4. Adjudication Method for the Test Set
The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the test set. The comparison is described as being between "manual and semi-automated segmentation" and "manual and semi-automated measurements," suggesting that the manual method served as the reference, but how disagreements or variations in the manual process were handled is not detailed.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not mention or describe a multi-reader multi-case (MRMC) comparative effectiveness study. The focus is on comparing the semi-automated software's performance against manual methods, rather than assessing the improvement of human readers with AI assistance.
6. Standalone Performance Study
Yes, a standalone performance study was implicitly done. The "Tests performed" section describes comparisons between the semi-automated segmentation and measurements (algorithm only) and manual methods. This indicates that the algorithm's performance was evaluated independently from a human-in-the-loop scenario, by comparing its outputs directly to the manually established ground truth.
7. Type of Ground Truth Used
The type of ground truth used was expert consensus / manual segmentation and measurement. The tests involved a "comparison between the manual and semi-automated segmentation" and "comparison of the manual and semi-automated measurements," indicating that the gold standard was derived from manual interpretation and manipulation by human experts.
8. Sample Size for the Training Set
The document does not specify the sample size used for the training set.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set was established. In fact, it doesn't mention a distinct training set at all, which is common for older 510(k) submissions, especially for software that is more rules-based or semi-automatic rather than deep learning-based. It's possible the "manual segmentation" and "measurements" used for testing were also part of the development and refinement process, but a separate, explicitly defined training set with its ground truth establishment is not detailed.
Ask a specific question about this device
(28 days)
G-scan Brio is a Magnetic Resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip, imaging the temporomandibular joint and imaging the cervical spine and the lumbar spine sections as portions of the spinal column.
G-scan Brio images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
The changes performed on the modified G-scan device (G-scan Brio), with respect to the cleared version - G-scan K111803 - are due to the improvement of the system performance. These modifications, which do not affect the intended use or alter the fundamental scientific technology of the device, are the following:
-
- A new patient table that can be moved manually in the inner/outer direction of the gantry.
-
- A new footboard and seat for vertical examinations.
-
- A patient table extension.
-
- A back support.
-
- A new step for patient positioning.
-
- A new control panel on the front of the magnet.
-
- Upgrading of the hydraulic circuit.
-
- Upgrading of the electronics.
-
- 4-channels Lumbar spine coils N.17 and 18 introduction, cleared via K110802.
-
- A new software release.
The G-scan Brio is a Magnetic Resonance (MR) system.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance | Study Type |
|---|---|---|
| Safety and Performance Standards Adherence | Met all listed industry performance and safety standards, and demonstrated to be as safe and effective as predicate devices. | Non-clinical testing |
| IEC 60601-1:1988 (Medical Electrical Equipment - General Safety) | Met | Non-clinical testing |
| IEC 60601-1-1:2000 (Medical Electrical Equip - Safety for Medical Electrical Systems) | Met | Non-clinical testing |
| IEC 60601-1-2:2001 (General safety - Electromagnetic compatibility) | Met | Non-clinical testing |
| IEC 60601-1-4:2000 (Programmable electrical medical systems) | Met | Non-clinical testing |
| IEC 60601-2-33 (2008) (Safety of magnetic resonance equipment) | Met | Non-clinical testing |
| IEC 62304:2006 (Medical device software) | Met | Non-clinical testing |
| ISO 10993-1:2003 (Biological evaluation of medical devices) | Met | Non-clinical testing |
| ISO 14971:2007 (Application of risk management to medical devices) | Met | Non-clinical testing |
| NEMA MS-1-2008 (SNR in Diagnostic MRI) | Met | Non-clinical testing |
| NEMA MS 2-2008 (Geometric Distortion in Diagnostic MRI) | Met | Non-clinical testing |
| NEMA MS 3-2008 (Image Uniformity in Diagnostic MRI) | Met | Non-clinical testing |
| NEMA MS 5-2010 (Slice Thickness in Diagnostic MRI) | Met | Non-clinical testing |
| Maintenance of Intended Use | Demonstrated that modifications did not affect the intended use or alter the fundamental scientific technology. | Non-clinical testing |
| Image Quality (Implied) | The device is intended to produce diagnostically useful images when interpreted by a medical expert. Conformance to NEMA MS standards (SNR, Geometric Distortion, Image Uniformity, Slice Thickness) directly relates to image quality. | Non-clinical testing |
2. Sample size used for the test set and the data provenance:
The provided document describes non-clinical testing focused on engineering and performance standards, not a clinical study involving human subjects or patient data. Therefore, there is no "test set" in the sense of a dataset of patient images with associated ground truth for evaluating diagnostic accuracy or a similar clinical performance endpoint. The testing involved the device in a laboratory setting to verify its adherence to technical specifications and safety standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. As explained above, this was a non-clinical engineering and performance study. There was no "ground truth" related to medical diagnoses established by experts for a test set of images. The "ground truth" was the technical specifications and standards (e.g., specific SNR values, distortion limits).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. No human interpretation of images for diagnosis or "adjudication" was performed as part of this non-clinical performance data. The device's performance was measured against established technical standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, an MRMC comparative effectiveness study was not done. The G-scan Brio is an MRI hardware system, and the filing focuses on technical and safety performance, not the diagnostic performance of a read by a human expert with or without AI assistance. The document predates widespread AI integration in MR system hardware.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
No, a standalone algorithm performance study was not done. The G-scan Brio is an MRI system that requires interpretation by a human expert. The "performance data" refers to the system's compliance with technical and safety standards, not the performance of an independent algorithm for diagnostic interpretation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this performance data was primarily technical specifications and established industry standards (e.g., NEMA standards for SNR, geometric distortion, image uniformity, slice thickness). The verification involved confirming the physical and technical parameters of the device met these engineering and safety benchmarks.
8. The sample size for the training set:
Not applicable. This document describes the performance of an MR imaging system (hardware and associated software), not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. As this is not an AI algorithm, there was no training set or associated ground truth establishment process for a training set.
Ask a specific question about this device
(34 days)
S-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf. knee. thigh and hip, imaging of the temporomandibular joint and imaging the cervical spine and the lumbar spine sections as portions of the spinal column.
S-scan images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
The changes performed on the modified S-scan device, with respect to the cleared version – S-scan K110802 - are due to the improvement of the system performance. These modifications, which do not affect the intended use or alter the fundamental scientific technology of the device, are the following:
- A new step for patient positioning.
- A kit for patient positioning during hip examination using 4-channels lumbar spine coil 17 and 18.
- Geometric distortion correction.
- FSE parameters optimization
- FSE acceleration and improving.
- FSE sampling frequency management.
- Streaming acquisition performance increasing.
- Configurable send and export of XBONE echoes.
- Environment for the management of the virtual film.
- Data matching for patient coming from a worklist server.
- SpeedUp technique introduction, cleared via K120748.
- A new software version.
This 510(k) submission (K121650) for the S-Scan MR system describes modifications to an already cleared device (K110802), rather than a new device that requires extensive clinical validation to demonstrate its performance against acceptance criteria. Therefore, the information typically associated with acceptance criteria and a detailed study proving performance (e.g., specific metrics, ground truth establishment, sample sizes for training/testing datasets, expert qualifications, adjudication methods, or MRMC studies) is not explicitly provided in the document.
The submission focuses on demonstrating that the modifications to the S-Scan device do not affect its intended use or alter its fundamental scientific technology, and thus the device remains substantially equivalent to its predicate. The "acceptance criteria" here are implied to be that the improved system performance from the modifications does not negatively impact image quality or diagnostic utility against the baseline established by the predicate device.
However, based on the provided text, we can infer some aspects related to the device performance implicitly.
Here's an analysis based on the document, addressing as many of your points as possible:
1. A table of acceptance criteria and the reported device performance
Since this is a Special 510(k) for device modifications, explicit "acceptance criteria" for a new device's performance are not detailed in the provided text. The implicit acceptance criterion is that the modified device's performance is at least equivalent to, or improved compared to, the predicate device (K110802), specifically in areas where modifications were made, without compromising the overall diagnostic utility.
| Acceptance Criterion (Implicit) | Reported Device Performance (Implicit from modifications) |
|---|---|
| No degradation of image quality for diagnostic interpretation. | System performance improvements via: - Geometric distortion correction. - FSE parameters optimization. - FSE acceleration and improving. - FSE sampling frequency management. - Streaming acquisition performance increasing. - SpeedUp technique introduction. These modifications aim to enhance image quality and acquisition efficiency. |
| Maintenance of intended use and fundamental scientific technology. | Stated: "These modifications, which do not affect the intended use or alter the fundamental scientific technology of the device." The images "can provide diagnostically useful information" when interpreted by a medical expert. |
| Compliance with general controls and relevant regulations. | Reviewed by FDA for substantial equivalence and allowed to market. |
2. Sample size used for the test set and the data provenance
The document does not specify a separate "test set" sample size or data provenance for a performance study. Since the modifications are for an already cleared device, it's likely that internal verification and validation activities were performed by the manufacturer, rather than a new clinical study with a distinct test set described in the 510(k). No information is provided regarding the country of origin or whether data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided. Given the nature of a Special 510(k) for modifications to an established MR system, it's improbable that a new, extensive ground truth establishment process for a clinical test set was required and detailed in this submission. The "ground truth" for the diagnostic utility of MR images, in general, is implicitly established by the experience and training of "a medical expert trained in the use of MR equipment."
4. Adjudication method for the test set
Not specified in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study is not mentioned. The device described is an MR system, not an AI-assisted diagnostic tool, so the concept of "human readers improve with AI vs without AI assistance" does not apply to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The S-Scan is an MR imaging system, not a standalone algorithm. Its output, images, requires interpretation by a human expert.
7. The type of ground truth used
The ground truth is implied to be expert interpretation of MR images. The document states: "When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information." There is no mention of pathology or outcomes data as direct ground truth for this submission, as the submission focuses on system modifications.
8. The sample size for the training set
Not applicable. This document describes modifications to an MR imaging system. It is not an AI/algorithm-based device that would typically have a "training set" in the context of machine learning.
9. How the ground truth for the training set was established
Not applicable, as there is no mention of a training set for an AI/algorithm.
Ask a specific question about this device
Page 1 of 3