Search Results
Found 6 results
510(k) Data Aggregation
(248 days)
DXM Co., Ltd
The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 400-480 nm waveband of visible light.
Cybird is a LED curing light intended for polymerization of light-cured materials by dental professionals. This product is effective on various light-cured materials. Cybird's body is made from industrial-grade aluminum which is a durable material intended to ensure heat dissipation. The Cybird features multiple curing modes to ensure functionality.
The provided text describes the 510(k) summary for the Cybird LED Curing Light (Model: Cybird WPT) and its comparison to a predicate device. It explicitly states that clinical testing was not required to demonstrate substantial equivalence, meaning a study proving the device meets specific acceptance criteria based on clinical outcomes or human reader performance was not performed or provided.
Instead, the submission relies on non-clinical tests and a comparison of technical specifications to a previously cleared predicate device. Therefore, it's not possible to provide all the requested information, specifically points 2 through 9, as they pertain to clinical or human-reader studies.
Here's a breakdown of what information can be extracted based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not present a formal "acceptance criteria" table with specific thresholds that the device had to meet to be deemed effective in a clinical context. Instead, it compares the performance attributes of the Cybird WPT to its predicate device, the Cybird LED Curing Light (Model: Gold & XD, K173876). The "acceptance criteria" here implicitly relates to demonstrating that the new device performs "similar or better" than the predicate in non-clinical tests and that the differences do not "represent significant issues to clinical evaluation."
Performance Attribute | Predicate Device (Cybird Gold/XD - K173876) | Subject Device (Cybird WPT) | Comparison / "Acceptance" Status |
---|---|---|---|
Intended Use | Polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 400-480 nm waveband of visible light. | Same as predicate. | Same |
Device Design | 4.2VDC Lithium Ion battery charger | Inductive Wireless Charger | Different (Same user convenience; spectral irradiance identical) |
Operational Modes (Plasma/Ortho) | 2, 3, 4 sec (Plasma); 55 sec (Ortho) | 3, 6, 9 sec (Plasma); 55 sec (Ortho) | Different (Longer curing time to achieve equivalent polymerization with black light guides) |
Light Source | LED light, blue and violet wavelengths; 8 light head diameter | Same as predicate. | Same |
Accessories | Not specified. | Not specified. | Not specified. |
Composition of Patient-Contacting Materials | Aluminum, anodized white; Barrier Sleeves; Light Shield | Same as predicate. | Same |
Light Guide | Fiber optic glass Clear (11x8, 11x11) | Fiber optic glass Black (11x8, 11x11) | Different (Less light penetration through black guides) |
Light Intensity (mW/Cm²) | Plasma E. Mode/Ortho Mode: 11x8 Clear Shor (2581); 11x11 Clear Short (1800); High Power Mode: 11x8 Clear Shor (1521); 11x11 Clear Short (1043) | Plasma E. Mode/Ortho Mode: 11x8 Black Long (2034); 11x11 Black Long (1294); High Power Mode: 11x8 Black Long (1104); 11x11 Black Long (675) | Different (WPT's irradiance is ~20-30% lower with black light guide) |
Peak Wavelength | Dual peak: Approximately 405nm, 460nm (GOLD), 460nm (XD) | Dual peak: Approximately 405nm, 460nm | Same |
Depth of Cure (mm) - Spident Escom 100 | Plasma mode (2,3,4 sec): 11x8 Clear: 3.21 | Plasma mode (3,6,9 sec): 11x8 Black: 3.08 | Similar (Performances are similar using Spident ESCOM – K110428) |
Parameters of Disinfection | Chemical disinfection with approved cleaning/sanitizing agents (Cavicide, Isopropyl alcohol (75%), Ethyl alcohol based, Lysol disinfectant (alcohol-based only)) | Same as predicate. | Same |
Usability/Ergonomics | 2 buttons – 1 power, 1 mode select | Same as predicate. | Same |
Power Source | 3.7V Li-ion Battery | 3.7V Li-ion Battery with inductive wireless charging | Same (Battery type) |
Curing Hardness | Tested | Tested | Similar or better (demonstrated that the Cybird WPT LED Curing light performed similar or better compared with the predicate device. ) |
Spectral Irradiance Output | Tested | Tested | Similar or better (demonstrated that the Cybird WPT LED Curing light performed similar or better compared with the predicate device. ) |
Summary of Acceptance: The Cybird WPT LED Curing Light was deemed substantially equivalent based on:
- Identical intended use.
- Demonstration through non-clinical performance and material testing that differences in device design (wireless charger, adjusted curing times for black light guides) and resulting light intensity do not negatively impact the device's ability to achieve polymerization of dental materials, and that the depth of cure is similar to the predicate.
- Compliance with voluntary standards for biocompatibility, electrical safety, EMC, and software verification/validation.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable. No clinical test set data was used. Non-clinical tests were performed on the device itself and material samples. The text does not specify the sample sizes for these non-clinical tests (e.g., how many devices were tested for irradiance intensity or depth of cure) nor the country of origin of this testing data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. No clinical "ground truth" established by experts was used, as clinical testing was not required for this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. No clinical test set requiring adjudication was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is an LED curing light, not an AI-powered diagnostic or assistive tool. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is hardware for polymerization; it does not involve an algorithm with standalone performance in the context of diagnostic or interpretive tasks. Software verification and validation were done for its internal operation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the non-clinical tests, the "ground truth" or reference for performance would be established by:
- Validated laboratory methods and standards: Such as those for measuring irradiance intensity, spectral output, and depth of cure (e.g., ADA Specification No. 48).
- Predicate device performance: The performance of the predicate device (Cybird Gold / XD) served as the benchmark for comparison for most attributes.
- Material specifications: For biocompatibility, materials were tested against ISO standards (ISO 10993-5, 10993-10).
8. The sample size for the training set
- Not applicable. This device does not use machine learning or AI models that require a training set.
9. How the ground truth for the training set was established
- Not applicable. No training set was used.
Ask a specific question about this device
(34 days)
DXM Co., Ltd
The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 400-480 nm waveband of visible light.
Cybird is a LED curing light intended for rapid polymerization of light-cured materials by dental professionals. This product effectively reduces polymerization time on various lightcured materials and provides excellent treatment results every time. Cybird's body is made from industrial-grade aluminum which ensures its durability and excellent heat dissipation. The Cybird features multiple curing modes for maximum functionality.
The Cybird curing light produces visible blue light in the 405 nm (Gold) & 465 nm (Gold / XD) peak waveband of spectrum with a power density of 1,500mW/cm² (High Power Mode), 2,500 mW/cm2 (Ortho Mode, Gold), 2,700mW/cm2 (Ortho Mode, XD). These power densities are sufficient for the product intended uses.
405 nm (Gold) & 465 nm (Gold / XD) peak waveband of visible light cured (VLC) orthodontic brackets and orthodontic bonding and sealing materials.
Exposure times can be set for 1.5, 2, 2,5, 3 seconds for XD in Ortho Mode; 2, 3, 4 seconds for Gold in Ortho Mode; and 5,10,15, or 20 seconds in High Power Mode.
The provided text describes the 510(k) premarket notification for the "Cybird LED Curing Light" and its substantial equivalence to a predicate device. However, it does not explicitly detail acceptance criteria and a study proving the device meets those criteria in the way a clinical trial or algorithm performance study would.
Instead, the document focuses on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical performance testing and compliance with various standards. The "acceptance criteria" here are implied to be the standards and measurements taken to show the device performs similarly or equivalently to the predicate and adheres to regulatory requirements.
Here's an attempt to structure the information based on your request, acknowledging the limitations of the provided text:
Acceptance Criteria and Device Performance
The "acceptance criteria" for this device are primarily demonstrated through non-clinical performance tests and compliance with established standards, aiming to show substantial equivalence to a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Test/Standard | Acceptance Goal (Implied) | Reported Device Performance (Cybird LED Curing Light) |
---|---|---|---|
Curing Performance | Curing hardness | Sufficient polymerization for intended use | Performed; results not explicitly quantified as pass/fail against specific target, but implies successful curing. |
Curing Performance | Depth of cure (per ADA Specification No. 48) | Sufficient depth of cure for intended use (e.g., 2.5mm of predicate) | Plasma mode (2 sec): 2.035mm - 2.380mm (for various materials); High Power mode (10 sec): 2.370mm - 3.795mm (for various materials). |
Light Intensity | Maximum light intensity measurements at 2mm | Within acceptable range for intended use; comparable to predicate | 1,500mW/cm² (High Power Mode), 2,500 mW/cm2 (Ortho Mode, Gold), 2,700mW/cm2 (Ortho Mode, XD). |
Light Spectrum | Spectral irradiation plots of all power modes | Appropriate waveband for polymerization (e.g., 400-480 nm) | Dual peak: 405nm, 465nm (GOLD); Single peak: 465nm (XD) - within 400-480 nm range. |
Biocompatibility | In vitro cytotoxicity (AAMI / ANSI / ISO 10993-5) | No cytotoxicity | Conformity with AAMI / ANSI / ISO 10993-5:2009(R) 2014. |
Biocompatibility | Irritation and skin sensitization (AAMI / ANSI / ISO 10993-10) | No irritation or sensitization | Conformity with AAMI / ANSI / ISO 10993-10:2010. |
Electrical Safety | Essential performance testing (AAMI ANSI ES 60601-1) | Compliance with safety standards | Conducted in accordance with AAMI ANSI ES 60601-1:2005/(R) 2012 and A1:2012. |
EMC Testing | Electromagnetic Compatibility (IEC 60601-1-2) | Compliance with EMC standards | Conducted in accordance with IEC 60601-1-2 Ed.3: 2007-03. |
Software | Verification and Validation (IEC 62304) | Compliance with software lifecycle processes | Conducted as recommended in IEC62304:2006 and FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." |
Substantial Equivalence | Comparison to predicate device | No new questions of safety or effectiveness compared to predicate | All differences (LED light source, light guard diameter, power mode, exposure time) found not to raise new questions of safety and effectiveness. |
2. Sample Size and Data Provenance
- Test Set Sample Size: The document does not specify a "test set" in the context of discrete data points for a predictive model. The testing involved various materials (e.g., Handae Chemical, Hanfil; Spident Escom 100; Dentkist Charmfil Plus; Mediclus Anycom) for depth of cure and hardness. The number of samples for each material or measurement is not provided.
- Data Provenance: The tests are non-clinical (laboratory-based performance tests) and were conducted by DXM Co., Ltd. (Republic of Korea). The data is prospective in the sense that it was generated specifically for this 510(k) submission.
3. Number of Experts and Qualifications for Ground Truth
- This information is not applicable as the document describes non-clinical performance testing of a physical device, not an AI/ML algorithm requiring expert-established ground truth for image interpretation or diagnosis. The "ground truth" for these tests would be the physical measurements and observations made by trained technicians and engineers in a laboratory setting.
4. Adjudication Method for the Test Set
- This information is not applicable for the reasons stated above. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical data reviewed by multiple experts, not for physical device performance measurements.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for evaluating the impact of AI on human reader performance for tasks like diagnostic interpretations. The "Cybird LED Curing Light" is a physical dental device, not a diagnostic AI tool.
6. Standalone (Algorithm Only) Performance Study
- No, a standalone (algorithm only) performance study was not done. This is also not applicable as the device is not an algorithm but a medical device.
7. Type of Ground Truth Used
- The "ground truth" for the tests described is based on physical measurements and adherence to recognized standards. For example, Depth of Cure was measured per ADA Specification No. 48, biocompatibility per ISO 10993 standards, and electrical safety per IEC 60601 standards. These are objective, empirically verifiable metrics, not expert consensus or pathology in a clinical sense.
8. Sample Size for the Training Set
- This information is not applicable. There is no "training set" as this is not an AI/ML device.
9. How Ground Truth for the Training Set Was Established
- This information is not applicable for the reasons stated above.
Ask a specific question about this device
(231 days)
DXM CO., LTD.
The D1 LED Curing Light & accessories is intended to polymerize resinous dental materials, restorative composite materials, orthodontic brackets bonding and sealing materials that are photo-polymerized in the 430~490nm waveband of visible light.
D1 by DXM Co., Ltd. is classified as an Ultraviolet Activator for polymerization (21 CFR 872.6070) because it is a device intended for the photopolymerization light cured dental materials. The D1-Cordless is a universal photo-polymerization light curing source working in cordless conditions and producing visible blue light in the 430 to 490nm waveband of spectrum with a power density of 3,000mW/cm² (11-8mm light-guide). These power densities are sufficient for the product intended uses, namely, Photo-polymerization in the 430-490 nm waveband of visible light cured (VLC) dental materials, restorative composite materials, and orthodontic brackets and orthodontic bonding and sealing materials.
D1 has two operating modes: Plasma Emulation/ Ortho mode: 3000 mW/cm² and High Power mode: 1500 mW/cm2.
Exposure timing intervals for Plasma Emulation mode is 1, 2, 3 seconds, Ortho: approximately 3 seconds on, 1 second off for 16 cycles and High Power mode: 5, 10, 20, 40 seconds.
The D1 Curing light is cordless and is powered by a rechargeable Lithium-lon battery which is charged by a Battery Charger Stand supplied with a certified AC/DC Adapter.
The provided text describes the 510(k) submission for the D1 LED Curing Light, which is a device intended to polymerize resinous dental materials. The submission focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a novel study.
Here's an analysis of the provided information concerning acceptance criteria and supporting studies:
-
Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it describes non-clinical bench tests performed to verify specific technical aspects of the device against established standards and expected performance for similar devices.
Acceptance Criterion (Inferred from testing) | Reported Device Performance |
---|---|
Verification of Spectral Irradiance Output | Confirmed as part of testing |
Verification/Validation of Irradiance Intensity | Confirmed as part of testing (specifically, 3,000mW/cm² for Plasma Emulation/Ortho mode and 1,500mW/cm² for High Power mode) |
Depth of Cure | Confirmed as part of testing |
Verification of Case Temperature Rise | Confirmed as part of testing |
Compliance with IEC 60601-1: 1988 +A1 1991, +A2 1995 | Met through testing |
Compliance with ANSI/AAMI/IEC 60601-1-2: 2007 | Met through testing |
Biocompatibility (ISO 10993-5 and ISO 10993-10) | Met through testing |
-
Sample Size Used for the Test Set and Data Provenance:
The document does not specify a "test set" in the context of clinical data for performance evaluation in the way one would for a diagnostic or AI algorithm. Instead, it refers to non-clinical bench tests. For these bench tests, sample sizes are not explicitly stated. The data provenance is from non-clinical bench testing conducted in a lab setting, likely by DXM Co., Ltd. or a contracted testing facility. No country of origin for a clinical data set is applicable as no clinical studies were performed.
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. No ground truth in the context of expert consensus on clinical cases was established, as no clinical studies were performed. The "ground truth" for the non-clinical tests would be the established scientific and engineering principles and the specifications derived from the predicate devices and relevant standards.
-
Adjudication Method for the Test Set:
Not applicable, as no clinical test set requiring expert adjudication was used.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC comparative effectiveness study was done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not applicable to a dental curing light device.
-
Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
Not applicable. The D1 LED Curing Light is a physical device, not an algorithm. The "standalone" performance refers to the device's technical specifications and functionality, which were evaluated through bench testing.
-
Type of Ground Truth Used:
The "ground truth" for this submission is based on engineering specifications, established international and national standards (e.g., IEC 60601-1, ANSI/AAMI/IEC 60601-1-2, ISO 10993-5, ISO 10993-10), and the performance characteristics of predicate devices. For example, the irradiance intensity of 3,000mW/cm² (Plasma Emulation/Ortho mode) and 1,500mW/cm² (High Power mode) would be a performance characteristic verified against the device's design specifications, which are implicitly derived from the requirements for effectively polymerizing dental materials and the performance of predicate devices.
-
Sample Size for the Training Set:
Not applicable. There is no "training set" as this device does not involve machine learning or AI algorithms requiring data for training.
-
How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set.
Ask a specific question about this device
(87 days)
DXM CO., LTD.
Indicated for the use of local anesthetics for infiltration and nerve block anesthesia administered prior to, or in conjunction with dental procedures.
Denjector is a gun type electronic dental anesthetic cartridge syringe. This device is used with injectable local anesthetics packaged in cartridge form. The device consists of: Injector gun, battery, battery charger stand, DC adapter for battery charger stand, Cartridge holder and Cartridge pushing/pulling pin. Standard needles may be used with the device.
This document is a 510(k) summary for the Denjector, an electronic dental anesthetic cartridge syringe. It indicates that no clinical tests were performed to establish its safety and effectiveness. Instead, the submission relies on non-clinical tests and a comparison to a predicate device to claim substantial equivalence. Therefore, the information requested regarding acceptance criteria and a study proving those criteria cannot be extracted from this document, as such a study was not conducted.
Here is a summary of the information that can be extracted, along with explanations for the missing data:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria | Reported Device Performance |
---|---|
Not applicable | Not applicable |
(No specific acceptance criteria for performance were defined or tested clinically in this submission.) | (No device performance metrics from a clinical study are reported as no clinical study was performed.) |
Explanation: The document explicitly states, "Clinical testing was not conducted." Therefore, there are no acceptance criteria derived from clinical performance and no reported device performance against such criteria. The substantial equivalence claim is based on non-clinical standards and comparison to a predicate.
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not applicable (No clinical test set was used).
- Data Provenance: Not applicable (No clinical data was used).
Explanation: As no clinical testing was performed, there is no test set or associated data provenance.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Number of Experts: Not applicable (No clinical test set requiring ground truth established by experts).
- Qualifications: Not applicable.
Explanation: Without clinical testing, there was no need for experts to establish ground truth for a test set.
4. Adjudication Method for the Test Set:
- Adjudication Method: Not applicable.
Explanation: No clinical test set means no adjudication method was necessary.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- MRMC Study Done: No.
- Effect Size: Not applicable.
Explanation: An MRMC comparative effectiveness study involves human readers and clinical cases. Since no clinical testing was done, no such study was conducted.
6. Standalone (Algorithm Only) Performance Study:
- Standalone Study Done: No.
Explanation: The "Denjector" is a hardware device (an electronic dental anesthetic cartridge syringe), not a software algorithm. Therefore, a standalone algorithm-only performance study is not relevant. The testing mentioned (EN 60601-1, EN 61000-3-2, and EN 61000-3-3) refers to electrical safety and electromagnetic compatibility standards for medical devices, which are not performance studies in the clinical sense.
7. Type of Ground Truth Used:
- Type of Ground Truth: Not applicable.
Explanation: There are no clinical outcomes or diagnostic assessments to establish ground truth for, as no clinical study was performed.
8. Sample Size for the Training Set:
- Sample Size: Not applicable.
Explanation: The Denjector is a hardware device, not an AI/ML algorithm that requires a training set.
9. How Ground Truth for the Training Set Was Established:
- Method: Not applicable.
Explanation: As no training set was used, no ground truth needed to be established for it.
In summary, the provided 510(k) submission for the Denjector explicitly states that no clinical trials were conducted. The device's safety and effectiveness were established through non-clinical testing against recognized standards (EN 60601-1, EN 61000-3-2, EN 61000-3-3) and a comparison to a predicate device (CC:3; computer controlled syringe by DENTSPLY Midwest Professional Ltd., K983105) to demonstrate substantial equivalence.
Ask a specific question about this device
(12 days)
DXM CO., LTD.
The Cybird LED Curing Light is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photo-polymerized in the 430-490 nm waveband of visible light.
Cybird by DXM Co., LTd is classified as an Ultraviolet Activator for polymerization(21 C.f.r.872.6070) because it is a device intended for the photo-polymerization light cured dental materials. The Cybird is a universal photo-polymerization light curing source working in cordless conditions and producing visible blue light in the 440 to 490nm waveband of spectrum with a power density comprised 1,000W/cm², 11-8mm light-guide).These power densities are sufficient for the product intended uses ,namely, Photopolymerization in the 440-490 nm waveband of visible light cured (VLC) dental materials , Photo – polymerization in the 440-490nm waveband of visible light cured(VLC) restorative composite materials, and Photo-polymerization in the 440-490 nm waveband of visible light cured (VLC) orthodontic brackets and orthodontic bonding and sealing materials. Exposure times can be set for 5,10,15, or 20 seconds.
The DXM Co., Ltd. Cybird LED Curing Light is a device intended for the photo-polymerization of light-cured dental materials. The submission states that no clinical testing was conducted to establish its safety and effectiveness. Instead, the device's substantial equivalence to a predicate device, the Satelec Mini L.E.D (K032465), was demonstrated through non-clinical testing.
Therefore, the following information cannot be extracted from the provided text: A table of acceptance criteria and the reported device performance, sample size used for the test set and data provenance, number of experts for ground truth and qualifications, adjudication method, MRMC comparative effectiveness study results, standalone performance, type of ground truth used, sample size for training set, and how ground truth for the training set was established.
However, based on the provided text, the following details regarding the acceptance criteria and the study that "proves" the device meets them can be summarized:
Non-Clinical Acceptance Criteria and Reported Performance (via equivalence to a predicate device):
The acceptance criteria for the Cybird LED Curing Light were not explicitly stated as quantitative thresholds for performance metrics. Instead, the device's acceptability was based on demonstrating substantial equivalence to a legally marketed predicate device (Satelec Mini L.E.D, K032465) through non-clinical testing. This implies that the device's performance, as measured by these non-clinical tests, should be comparable to or within acceptable limits relative to the predicate device and established safety standards.
Acceptance Criteria Category | Reported Device Performance (via Non-Clinical Testing) |
---|---|
Intended Use | Demonstrated substantial similarity to the predicate device in its intended use, which is the photo-polymerization of resinous dental materials, restorative composite materials, and orthodontic brackets, bonding, and sealing materials that are photo-polymerized in the 430-490 nm waveband of visible light. |
Operational Characteristics | Produces visible blue light in the 440 to 490nm waveband of spectrum with a power density comprised of 1,000W/cm² (11-8mm light-guide). Exposure times can be set for 5, 10, 15, or 20 seconds. These power densities are sufficient for the intended uses. |
Safety - Electrical Safety | Conforms to EN 60601-1: 1990 +A,+B. No design characteristics violated the requirements of this standard or resulted in safety hazards. |
Safety - Electromagnetic Compatibility (EMC) | Conforms to EN 60601-1-2: 2001, EN 61000-3-2:2000, EN61000-3-3: 1998/A1:2001, and EN 55011: 1998/A1:1999, Class B. No design characteristics violated the requirements of these standards or resulted in safety hazards. |
Function | Demonstrated substantial similarity to the predicate device without any design characteristics violating standard requirements or resulting in safety hazards. |
Study Information (Non-Clinical Equivalence Study):
-
Sample size used for the test set and the data provenance: Not applicable in the traditional sense of a clinical trial. The "test set" consisted of the Cybird LED Curing Light itself, undergoing various engineering and electrical safety tests. The "data provenance" is from these specific non-clinical tests performed on the device by DXM Co., Ltd. There is no mention of country of origin for the data or whether it was retrospective or prospective in the context of human subjects.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in this context, was established by compliance with international and national standards for electrical safety and electromagnetic compatibility (e.g., EN 60601-1, EN 61000 series, EN 55011). These standards themselves are established by expert consensus in their respective fields, but no specific number or qualification of experts for this particular device's testing is mentioned beyond the presumption that the company performing the tests had qualified personnel.
-
Adjudication method: Not applicable. The process was one of compliance testing against defined standards. Either the device met the standard (passed) or it did not (failed). There was no independent adjudication of test outcomes by multiple parties mentioned.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an LED curing light, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a standalone hardware product for polymerization, not an algorithm.
-
The type of ground truth used: The ground truth for proving the device's safety and effectiveness was based on compliance with established international and national standards for electrical safety (EN 60601-1), electromagnetic compatibility (EN 60601-1-2, EN 61000 series, EN 55011), and demonstration of functional equivalence (power output, wavelength) to the predicate device for its intended use.
-
The sample size for the training set: Not applicable. There was no machine learning model or algorithm to train.
-
How the ground truth for the training set was established: Not applicable, as there was no training set.
Ask a specific question about this device
(91 days)
DXM CO., LTD.
The Hawk Intraoral Camera and Accessories are indicated for use to provide a view of mouth to assist the dentist in describing dental procedures and to show the patient the mouth before and after dental procedures.
Hawk Intraoral Camera composed of light (25g), small (185 x10mm) hand piece along with docking station. The hand piece consists of a focusing mechanism to assist the doctor in taking video of the patient's mouth. The hand piece connects to docking station via cable. The docking station attaches directly to monitor, TV, or computer via standard composite connection. It uses high definition imaging (480 lines) to capture images at 78 degrees field of view. Hawk Intraoral Camera includes also footswitch.
The provided 510(k) summary for the "Hawk Intraoral Camera and Accessories" does not describe any acceptance criteria or a study proving the device meets them. This type of submission, especially for a Class I device seeking substantial equivalence, generally relies on demonstrating equivalence to a predicate device through non-clinical testing of safety and performance against recognized standards, rather than clinical efficacy studies with specific performance metrics.
Here's why the requested information cannot be extracted from this document:
- No Acceptance Criteria: The document does not define specific metrics (e.g., sensitivity, specificity, accuracy, image resolution thresholds, field of view clarity) that the device must achieve to be considered effective.
- No Performance Study: There is no study described that measures the device's performance against predefined acceptance criteria. The document explicitly states: "Clinical testing was not conducted."
- Reliance on Substantial Equivalence: The approval is based on demonstrating substantial equivalence to a predicate device (Veracan Intraoral camera and accessories, K021083) in terms of intended use, operation, safety, and function, supported by non-clinical testing against electrical and medical device safety standards (EN 60601-1, EN 60601-2, EN 61000-3-2 and EN 61000-3-3).
Since the specific information requested about acceptance criteria and a performance study is not present in the provided text, the table and other details cannot be filled as requested. The available information primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing.
Explanation of Missing Elements:
- Table of acceptance criteria and reported device performance: Not present. The submission focuses on meeting safety standards and equivalence to a predicate, not specific performance metrics.
- Sample size, data provenance, number of experts, adjudication, MRMC study, standalone performance, type of ground truth, training set sample size, how ground truth for training was established: All these elements are relevant to performance studies, especially those involving AI or diagnostic devices, which were not conducted for this camera. The device is a viewing tool, not a diagnostic one requiring such detailed performance evaluation.
Summary of available relevant information:
- Intended Use: To provide a view of the mouth to assist the dentist in describing dental procedures and to show the patient the mouth before and after dental procedures.
- Predicate Device: Veracan Intraoral camera and accessories (K021083).
- Non-Clinical Testing Standards: EN 60601-1, EN 60601-2, EN 61000-3-2, and EN 61000-3-3. These are safety and electromagnetic compatibility standards, not performance criteria for diagnostic accuracy or image quality exceeding basic functionality.
- Clinical Testing: Not conducted.
- Conclusion: Substantially equivalent to the predicate and safe and effective for its intended use based on non-clinical testing against standards.
Ask a specific question about this device
Page 1 of 1