K Number
K121093
Manufacturer
Date Cleared
2012-11-27

(231 days)

Product Code
Regulation Number
872.6070
Panel
DE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The D1 LED Curing Light & accessories is intended to polymerize resinous dental materials, restorative composite materials, orthodontic brackets bonding and sealing materials that are photo-polymerized in the 430~490nm waveband of visible light.

Device Description

D1 by DXM Co., Ltd. is classified as an Ultraviolet Activator for polymerization (21 CFR 872.6070) because it is a device intended for the photopolymerization light cured dental materials. The D1-Cordless is a universal photo-polymerization light curing source working in cordless conditions and producing visible blue light in the 430 to 490nm waveband of spectrum with a power density of 3,000mW/cm² (11-8mm light-guide). These power densities are sufficient for the product intended uses, namely, Photo-polymerization in the 430-490 nm waveband of visible light cured (VLC) dental materials, restorative composite materials, and orthodontic brackets and orthodontic bonding and sealing materials.

D1 has two operating modes: Plasma Emulation/ Ortho mode: 3000 mW/cm² and High Power mode: 1500 mW/cm2.

Exposure timing intervals for Plasma Emulation mode is 1, 2, 3 seconds, Ortho: approximately 3 seconds on, 1 second off for 16 cycles and High Power mode: 5, 10, 20, 40 seconds.

The D1 Curing light is cordless and is powered by a rechargeable Lithium-lon battery which is charged by a Battery Charger Stand supplied with a certified AC/DC Adapter.

AI/ML Overview

The provided text describes the 510(k) submission for the D1 LED Curing Light, which is a device intended to polymerize resinous dental materials. The submission focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a novel study.

Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

  1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it describes non-clinical bench tests performed to verify specific technical aspects of the device against established standards and expected performance for similar devices.

Acceptance Criterion (Inferred from testing)Reported Device Performance
Verification of Spectral Irradiance OutputConfirmed as part of testing
Verification/Validation of Irradiance IntensityConfirmed as part of testing (specifically, 3,000mW/cm² for Plasma Emulation/Ortho mode and 1,500mW/cm² for High Power mode)
Depth of CureConfirmed as part of testing
Verification of Case Temperature RiseConfirmed as part of testing
Compliance with IEC 60601-1: 1988 +A1 1991, +A2 1995Met through testing
Compliance with ANSI/AAMI/IEC 60601-1-2: 2007Met through testing
Biocompatibility (ISO 10993-5 and ISO 10993-10)Met through testing
  1. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a "test set" in the context of clinical data for performance evaluation in the way one would for a diagnostic or AI algorithm. Instead, it refers to non-clinical bench tests. For these bench tests, sample sizes are not explicitly stated. The data provenance is from non-clinical bench testing conducted in a lab setting, likely by DXM Co., Ltd. or a contracted testing facility. No country of origin for a clinical data set is applicable as no clinical studies were performed.

  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. No ground truth in the context of expert consensus on clinical cases was established, as no clinical studies were performed. The "ground truth" for the non-clinical tests would be the established scientific and engineering principles and the specifications derived from the predicate devices and relevant standards.

  3. Adjudication Method for the Test Set:

    Not applicable, as no clinical test set requiring expert adjudication was used.

  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not applicable to a dental curing light device.

  5. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Not applicable. The D1 LED Curing Light is a physical device, not an algorithm. The "standalone" performance refers to the device's technical specifications and functionality, which were evaluated through bench testing.

  6. Type of Ground Truth Used:

    The "ground truth" for this submission is based on engineering specifications, established international and national standards (e.g., IEC 60601-1, ANSI/AAMI/IEC 60601-1-2, ISO 10993-5, ISO 10993-10), and the performance characteristics of predicate devices. For example, the irradiance intensity of 3,000mW/cm² (Plasma Emulation/Ortho mode) and 1,500mW/cm² (High Power mode) would be a performance characteristic verified against the device's design specifications, which are implicitly derived from the requirements for effectively polymerizing dental materials and the performance of predicate devices.

  7. Sample Size for the Training Set:

    Not applicable. There is no "training set" as this device does not involve machine learning or AI algorithms requiring data for training.

  8. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set.

§ 872.6070 Ultraviolet activator for polymerization.

(a)
Identification. An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.(b)
Classification. Class II.