Search Results
Found 10 results
510(k) Data Aggregation
(270 days)
The Pocket Fetal Doppler (Models CONTEC 10D, CONTEC 10E and CONTEC10F) is used to detect the fetal heart rate. The device should be used by health care professionals including nurses, midwives, and specialized technicians in the hospital, clinic, community and home. The device is intended for use at or after 12 weeks gestation.
The Pocket Fetal Doppler is a hand-held fetal Doppler used for non-invasive measurement and numerical display of the fetal heart rate (FHR) utilizing pulsed-wave Doppler ultrasound. The Pocket Fetal Doppler includes three models: CONTEC10E and CONTEC10E and CONTEC10F. All models have four components: upper shell, display and probe. The material for probe cap and shell is ABS. The device contains an ultrasonic signal transmitter and receiver, analog signals processing unit, FHR calculating unit, and LCD display control unit. The device is used to detect the fetal heart rate. The device should be used by health care professionals including nurses, midwives, and specialized technicians in hospital, clinic, community and home. This device can detect a fetal heart rate value at or after twelve weeks gestation.
The CONTEC10D is powered by a 3.7V lithium battery, the ultrasonic signal is continuously transmitted at a frequency of 2.5MHz.
The CONTEC10E and CONTEC10F is powered by two 1.5V batteries (AA LR6), the ultrasonic signal is continuously transmitted at a frequency of 2.5MHz.
The FDA 510(k) summary for the Pocket Fetal Doppler (CONTEC 10D, 10E, 10F) does not contain the level of detail typically found in a clinical study report for device acceptance criteria and performance, especially for AI/ML-based devices. This is a traditional medical device (fetal Doppler) and not an AI/ML device, hence some of the requested information (like number of experts for ground truth, adjudication methods, MRMC studies, or training set details) is not applicable or provided in this document.
However, based on the provided text, here's an attempt to extract the relevant information regarding performance and acceptance, and clarify what is not applicable or available:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are primarily based on its ability to accurately measure fetal heart rate (FHR) within specified ranges and its compliance with relevant electrical safety, EMC, and performance standards.
1. Table of Acceptance Criteria and Reported Device Performance:
| Feature/Test | Acceptance Criteria (from testing standards/specifications) | Reported Device Performance |
|---|---|---|
| FHR Measuring Range | 50 BPM ~ 240 BPM | 50 BPM ~ 240 BPM |
| FHR Resolution | 1 BPM | 1 BPM |
| FHR Accuracy | ±2 BPM | ±2 BPM |
| Acoustic Output (ISATA) | < 20 mW/cm² (FDA guidance) | CONTEC 10D: < 10 mW/cm²CONTEC 10E/F: < 20 mW/cm² |
| Biocompatibility | Meets FDA Guidance "Use of International Standard ISO 10993-1..." recommendations (Attachment G) | Device considered biocompatible (passed testing per ISO 10993 standards) |
| Electrical Safety | Complies with IEC 60601-1 and IEC 60601-1-11 | Complies with IEC 60601-1 and IEC 60601-1-11 |
| EMC | Complies with IEC 60601-1-2 | Complies with IEC 60601-1-2 |
| Performance (Ultrasonic) | Complies with IEC 60601-2-37 | Complies with IEC 60601-2-37 |
| Battery Safety | Complies with EN 62133 | Complies with EN 62133 |
| Use Life | Met design specifications | Passed Use Life testing |
| Battery Life | Met design specifications | Passed Battery life testing |
| Battery Indicator Performance | Met design specifications | Passed Battery indicator testing |
| Software Verification & Validation | Consistent with Basic level per 2023 FDA guidance "Content of Premarket Submissions for Device Software Functions" | Documentation consistent with Basic level |
2. Sample Size Used for the Test Set and Data Provenance:
The document describes non-clinical testing to verify design specifications and compliance with standards. It does not mention a "test set" in the context of patient data or clinical trials with a specific sample size. The performance metrics (FHR range, resolution, accuracy) are likely derived from bench testing using simulators or controlled inputs, and not from a dataset of patient recordings.
- Sample Size: Not specified as it's non-clinical, likely involves a series of tests on manufactured units.
- Data Provenance: Not applicable in the context of patient data. The tests are performed on the device itself.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
Not applicable. This is a hardware device measuring a physiological parameter. The "ground truth" for its performance is established by reference to calibrated measurement equipment and standardized test protocols, not by expert human interpretation of data.
4. Adjudication Method for the Test Set:
Not applicable. There is no human interpretation or "test set" for adjudication, as it's a non-clinical device performance study.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
No, an MRMC study was not done. This type of study is relevant for AI/ML diagnostic aids where the impact on human reader performance is evaluated. This device is a direct measurement tool for FHR, not an interpretive aid.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:
Yes, the performance characteristics (FHR measuring range, resolution, accuracy, acoustic output) are essentially standalone performance metrics of the device as an instrument, independent of human interpretation during a diagnostic process. The stated performance data (e.g., FHR accuracy ±2 BPM) represents the device's inherent capability.
7. The Type of Ground Truth Used:
The "ground truth" for the non-clinical performance tests is based on technical specifications and adherence to international and national standards. For example:
- FHR accuracy would be verified against a precisely known or simulated fetal heart rate signal.
- Acoustic output would be measured using specialized acoustic measurement equipment.
- Biocompatibility is determined by laboratory testing against ISO 10993 standards.
8. The Sample Size for the Training Set:
Not applicable. This device does not employ machine learning or AI that would require a "training set" of data.
9. How the Ground Truth for the Training Set was Established:
Not applicable, as there is no training set for this traditional medical device.
Ask a specific question about this device
(228 days)
B-Ultrasound Diagnostic System in a general-purpose, digital ultrasound diagnostic system for abdomen, gynecology, obstetric, urology, and small-parts application.
The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs (Thyroid, Galactophore, Testis), neonatal cephalic, peripheral vasculo-skeletal (both conventional and superficial), The device is intended to adult, pregnant woman, pediatric and neonate.
The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional.
This device is not intended for home use.
Modes of operation include B, M, PWD(PW), B+M, 2B, 4B.
B-Ultrasound Diagnostic System in a general-purpose, digital ultrasound diagnostic system for abdomen, gynecology, obstetric, urology, and small-parts application.
The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, peripheral vascular, and musculo-skeletal (both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate.
The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional.
This device is not intended for home use.
Modes of operation include B, M, PWD(PW), B+M, 2B, 4B.
Please refer to the acoustic output declaration for each transducer as following pages.
B-Ultrasound Diagnostic System is a Track 3, diagnostic ultrasound system, which has a main unit and two probes (Broadband convex array probe and Broadband convex array probe), powered by a 14.4V lithium battery, the ultrasonic signal is continuously transmitted at a frequency of 2-12MHz.
The B-Ultrasound Diagnostic System of six functional modules: Power module, Ultrasound front-end module, Keyboard module, Industrial control board, LCD screen, and Interface board. When electrical pulses are applied to the piezoelectric wafer (transducer), it will produce ultrasound waves of a certain frequency, which will enter the human body. Due to the ultrasound waves traveling through different tissues in the human body with different acoustic impedances (caused by differences in density and ultrasound transmission speed), the reflected echoes generated by the surfaces of different organs in the human body will be different. The reflected echoes of different sizes will be received by the piezoelectric wafer, and then converted into electrical pulse signals. After digital beamforming and other processing, those signals are formed into standard video signals that can be displayed on the monitor screen as cross-sectional images of the organ.
The provided text is a 510(k) summary for the "B-Ultrasound Diagnostic System" (K232895). It primarily focuses on demonstrating substantial equivalence to a predicate device (K170856) based on non-clinical testing. Crucially, it explicitly states there was "No clinical study included in this submission," and therefore, it does not contain the detailed information about acceptance criteria and clinical study results that you requested.
Therefore, I cannot extract the information required to populate the table or answer questions 2-9 from the provided text. The document indicates that the device's substantial equivalence and performance were evaluated through non-clinical tests and compliance with various international standards related to electrical safety, EMC, performance, and biocompatibility.
Here's an explanation of why I cannot fulfill your request based on the provided document:
- No Clinical Study Data: The document explicitly states "No clinical study is included in this submission." This means there is no data from human studies or specific performance metrics from such studies.
- Focus on Substantial Equivalence: The primary objective of this 510(k) summary is to demonstrate that the new device is "substantially equivalent" to a legally marketed predicate device, not necessarily to prove its absolute performance against specific clinical acceptance criteria in a human study.
- Non-Clinical Testing: The "Non-Clinical Test Conclusion" section lists various standards (IEC, ISO) that the device complies with, but these are general safety and performance standards for medical electrical equipment and ultrasound devices, not acceptance criteria for a specific clinical task or diagnostic accuracy.
- Lack of Performance Metrics for Specific Clinical Use: While the indications for use are listed (e.g., abdomen, gynecology, pediatric), the document doesn't provide performance metrics (like sensitivity, specificity, accuracy, dice score etc.) related to these applications, which would typically be found in a clinical study report.
In summary, the provided document does not contain the necessary information to describe the acceptance criteria and the study that proves the device meets those criteria in a clinical context.
Ask a specific question about this device
(226 days)
The Color Doppler Ultrasound Diagnostic System is a general-purpose ultrasound system. It is intended for use by, or under the direction of a qualified and trained physician for ultrasound imaging, measurement, display and analysis of the human body and fluid. The device is intended for use in a hospital environment.
The systems support the following clinical applications:
Ultrasound examination and diagnosis of abdomen, obstetrics, gynecology, small organ (Thyroid, Galactophore, Testis) and cardiac. Modes of operation include: 3D Imaging mode, B, M, PW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Combined modes: B/M, B/Color M, B/PWD , B/Color/PWD ,B/Power/PWD.
The apparatus is high resolution ultrasound diagnostic system, which adopts digital beam forming(DBF), real-time dynamic aperture(RDA), real-time dynamic beam apodization (DRA), real-time dynamic receive focusing (DRF), digital-control dynamic frequency scanning (DFS), frame correlation etc. technologies. The image is clear, stable, and high resolution.
It adopts embedded operating system, which greatly improve the product security and compatibility, and enhance the data processing ability.
The apparatus is taken more conveniently for its humane operation interface and management system, flexible expansibility and compatibility, pop-up menu and keyboard design.
Users can chose different kinds of display modes,such as:B, 2B, 4B, BM, M. Image real-time, freeze, memory, calling up and large capability cine loop can be realized; the device possesses multilevel scanning depth, dynamic range, sound output power and focus number, focus space, focus position adjusting etc. functions.
With controllable frame correlation, gamma correction, image reversing, histogram, depth advance, zoom magnification, no distortion real-time part magnification etc. abundant image processing functions. It is convenient for choice of optimal diagnostic image with perfect image pigeonhole, browse, management functions. Measure, calculation, and report are even complete and convenient with powerful measure software package and perfect measurement.
The device supports linear probe and convex probe. It has extensive application that is suitable for clinic diagnosis, such as abdomen, obstetrics, gynecology, small organs, cardiology etc. .
The provided text is a 510(k) Summary for a Color Doppler Ultrasound Diagnostic System. It aims to demonstrate substantial equivalence to a predicate device and includes information on non-clinical testing and a comparison table. However, it explicitly states:
"No clinical study is included in this submission."
This means there is no information provided in the document regarding acceptance criteria, device performance from a clinical study, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance. The document focuses on demonstrating substantial equivalence through non-clinical testing and comparison of technical specifications and intended use, not clinical efficacy or performance.
Therefore, I cannot fulfill your request for the specific details of acceptance criteria and the study that proves the device meets them because this information is not present in the provided text.
The closest information provided is:
- Non-Clinical Test Conclusion: "Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device."
- Applied Standards: A list of standards indicating safety and performance compliance (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359, ISO 10993-5, ISO 10993-10, IEC 62133-2, IEC TR 60601-4-2). These standards typically involve technical performance and safety assessments rather than clinical performance acceptance criteria.
- Acoustic Output within FDA guidelines: $Ispta.3 \le 720 mW/cm^2$, MI < 1.9, TI < 6.0. These are safety limits, not clinical performance metrics.
Without a clinical study, the acceptance criteria and performance data requested (e.g., related to diagnostic accuracy, sensitivity, specificity, etc.) are not available in this document.
Ask a specific question about this device
(243 days)
The Pocket Fetal Doppler (Models CONTEC10C and CONTEC10CL) is used to detect the fetal heart rate. The device should be used by health care professionals including nurses, midwives, and specialized technicians in the hospital, clinic, community and home. The device is intended for use at or after 12 weeks gestation.
The Pocket Fetal Doppler is a hand-held FHR detection device, it is used for non-invasive measurement and display of the fetal heart rate (FHR) utilizing Doppler ultrasound. The Pocket Fetal Doppler consists of two models (CONTEC10C, CONTEC10CL). Both models have two hand-held components, a main unit and a probe. The device contains components of ultrasonic signal transmitter and receiver, analog signals processing unit, FHR calculating unit, and LCD display control unit.
The provided text is a 510(k) Premarket Notification for a Pocket Fetal Doppler. It details the device's characteristics, indications for use, comparison to a predicate device, and results of non-clinical testing. However, it does not include information about acceptance criteria for performance metrics (beyond resolution and accuracy values), NOR does it describe a study involving human readers or experts to prove the device meets these criteria.
Therefore, many of the requested points cannot be extracted from the provided text.
Based on the information available:
1. A table of acceptance criteria and the reported device performance
The document provides performance specifications for the device and a predicate, but doesn't explicitly state "acceptance criteria" in a pass/fail sense in this section. It lists the device's performance characteristics.
| Performance Parameter | Acceptance Criteria (Not explicitly stated as such, but implied by device specification) | Reported Device Performance (Subject Device) | Reported Device Performance (Predicate Device) |
|---|---|---|---|
| FHR Measuring Range | (Not stated as separate criteria, but within acceptable range for fetal heart rate) | 50 BPM ~ 240 BPM | 50-210 bpm |
| Resolution | 1 BPM | 1 BPM | 1 bpm |
| Accuracy | ±2 BPM | ±2 BPM | ±2 bpm |
| Acoustic Output (ISATA) | < 20 mW/cm² (FDA recommendation) | < 20 mW/cm² | 2.0MHz – 17.24 mW/cm²2.5MHz – 18.57 mW/cm²3.0MHz – 11.496 mW/cm² |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
Not provided in the document. The testing described is non-clinical performance and safety testing, not human use studies with a test set of data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable. No human experts were used for a test set in the context of device performance claims based on this document. The testing focused on device specifications and compliance with standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. No test set involving human interpretation/adjudication was described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a direct measurement tool (Fetal Doppler) and does not involve AI assistance or human interpretation/improvement studies as described in the document.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This largely describes the type of testing performed. The device's performance (FHR range, resolution, accuracy) was tested in a standalone capacity against specifications, not as part of an algorithm's performance in image interpretation. The testing validated the device's ability to measure FHR.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the device performance metrics (FHR range, resolution, accuracy), the "ground truth" would be established by calibrated measurement equipment and reference standards typically used in medical device testing, not biological ground truth like pathology or outcomes.
8. The sample size for the training set
Not applicable. This document does not describe the development or testing of an AI algorithm with training data.
9. How the ground truth for the training set was established
Not applicable. No training set for an AI algorithm was described.
Ask a specific question about this device
(219 days)
The Electronic Sphygmomanometer and Automatic Blood Pressure Monitor are intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.
The proposed devices, ABPM50 Automaitc Blood Pressure Monitor, CONTEC08A and CONTEC08C Electronic Sphygmomanometers are battery driven automatic non-invasive Blood Pressure Monitor. They can automatically complete the inflation and BP measurement, which can measure systolic, and diastolic blood pressure as well as the pulse rate at upper arm within its claimed range and accuracy via the oscillometric technique. User can select the unit of the measurement: mmHg or kPa. ABPM50 and CONTEC08A and CONTEC08C can be only used on adult individuals.
The devices have the data storage function for data review including measurement time, systolic blood pressure, diastolic blood pressure and pulse rate.
ABPM50 and CONTEC08A have physiological over-limit prompt function which can be turned on or off by users. When the measurement results exceed the over-limit prompt limit, the physiological over-limit prompt function will be triggered. The over-limit can be set by users, and the low limit must be lower than the corresponding high limit.
In addition, all of the three proposed devices have technical over-limit prompt function, which will be triggered when the battery voltage is low, and this technical over-limit prompt function can not be cancelled unless being closed or the power replaced.
The provided FDA 510(k) summary (K202757) for the Electronic Sphygmomanometer, Automatic Blood Pressure Monitor CONTEC08A/CONTEC08C/ABPM50 details the non-clinical and clinical tests performed to demonstrate substantial equivalence to a predicate device.
Here's an breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this medical device, being a non-invasive blood pressure measurement system, are primarily based on established international standards for accuracy and safety. The reported device performance is presented as compliance with these standards and specific measurement ranges.
| Category | Acceptance Criteria (Standard/Range) | Reported Device Performance/Compliance |
|---|---|---|
| Accuracy | ISO 81060-2 (Clinical Investigation of Intermittent Automated Measurement Type) | "Total 85 subjects are included in each clinical study, the results of proposed device meet the performance parameters claimed in user manual, and the proposed device complies with ISO 81060-2." (Page 6) "The clinical trial results verify that the clinical accuracy of the proposed device is not inferior to that of predicate device." (Page 6) |
| Static Pressure | ±3 mmHg | ±3 mmHg (Page 8, Table 7-1) |
| Pulse Accuracy | ±5% | ±5% (Page 8, Table 7-1) |
| Electrical Safety | IEC 60601-1, IEC 60601-1-11 | Complies with IEC 60601-1, IEC 60601-1-11 (Page 7, Table 7-1) |
| Electromagnetic Compatibility (EMC) | IEC 60601-1-2:2014 | Complies with IEC 60601-1-2:2014 (Page 6, Table 7-1) |
| Performance Test | IEC 80601-2-30 Edition 1.1 2013-07 (Particular Requirements for Basic Safety and Essential Performance of Automated Non-Invasive Sphygmomanometers) | Complies with IEC 80601-2-30 Edition 1.1 2013-07 (Page 7, Table 7-1) "the proposed device has passed the IEC8601-2-30 and ISO81069-2 test, the performance of the proposed device can be guaranted..." (Page 9, Analysis 2) |
| Biocompatibility | ISO 10993-5 (In Vitro Cytotoxicity), ISO 10993-10 (Irritation and Skin Sensitization) | Complies with ISO 10993-5 and ISO 10993-10. (Page 7, Table 7-1) "has passed the Biocompatibility test (including Cytotoxicity, Skin Sensitization, Skin Irritation)..." (Page 9, Analysis 3) |
| Measurement Range (BP) | Varies (Predicate: 30-255 mmHg) | 10-270 mmHg (Proposed device, Page 8, Table 7-1) |
| Measurement Range (Pulse) | Varies (Predicate: 40-199 bpm) | 40-240 BPM (Proposed device, Page 8, Table 7-1) |
| Operating Environment | +5 °C | "has passed the IEC 60601-1-1 test, the safety and performance of the proposed device can be guaranteed." (Page 9, Analysis 1) |
| Storage Environment | -20 °C~+55 °C, ≤95% RH, 700 hPa~1060 hPa (No corrosive gas and drafty) | "has passed the IEC 60601-1-1 test, the safety and performance of the proposed device can be guaranteed." (Page 9, Analysis 1) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: "Total 85 subjects are included in each clinical study." (Page 6)
- Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. However, given that the manufacturer and correspondent are based in China, it is highly probable the clinical study was conducted in China. The study appears to be prospective as it describes a clinical investigation for the purpose of regulatory submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
The document states that the clinical study was conducted "in accordance with ISO 81060-2". ISO 81060-2 (Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type) outlines specific requirements for the observers and the comparison method to establish ground truth.
While the exact number and qualifications of experts (e.g., clinicians, trained observers) are not explicitly stated in this FDA submission summary, ISO 81060-2 typically requires:
- At least two trained observers to measure blood pressure using a reference sphygmomanometer simultaneously with the device under test.
- These observers must be trained and certified to perform auscultatory blood pressure measurements (e.g., using a mercury column or validated alternative) with high accuracy and reproducibility. The standard often specifies requirements for their training and regular performance validation to ensure inter-observer agreement.
4. Adjudication Method for the Test Set
The document does not explicitly describe an "adjudication" method in the context of expert consensus, as this is typically seen in studies where multiple readers interpret images or data and discrepancies need resolving. For blood pressure measurement, ISO 81060-2 mandates a specific comparison method where the device's measurements are compared against simultaneous auscultatory measurements obtained by trained observers. The standard defines how these measurements are to be collected and statistically analyzed (e.g., mean difference and standard deviation between the device and the reference measurements) to determine accuracy, rather than an "adjudication" of differing expert opinions on a specific diagnosis.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Improvement with AI vs. Without AI Assistance
This device (Electronic Sphygmomanometer/Automatic Blood Pressure Monitor) is a medical measurement device, not an AI-based diagnostic tool that assists human readers in interpreting complex data like medical images. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, which is common for AI tools in radiology or pathology, was not performed or applicable here. The study focused on demonstrating the device's accuracy and safety against established standards and a predicate device through direct measurement comparison.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the primary evaluation of this device is a standalone performance assessment. The device is designed to automatically measure and display blood pressure and pulse rate without continuous human intervention during the measurement process. The clinical study evaluated the device's direct measurement performance against a gold standard (auscultatory measurements) as per ISO 81060-2.
7. The Type of Ground Truth Used
The ground truth used for the clinical study was expert consensus/reference measurement through a standardized protocol. Specifically, it involved:
- Simultaneous blood pressure measurements taken by trained observers using a reference method (typically auscultation with a mercury sphygmomanometer or validated equivalent) alongside the device under test.
- This is the standard approach defined in ISO 81060-2 for validating automated non-invasive sphygmomanometers.
8. The Sample Size for the Training Set
The document does not mention a training set because this medical device approval is for a physical measurement device, not a machine learning or AI algorithm that requires a separate training set. The "study" described is a clinical validation study to confirm the accuracy and safety of already developed hardware and software, rather than a development and training phase for a learning algorithm.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set" in the context of an AI/ML algorithm for this type of device, this question is not applicable. The device's underlying measurement algorithms are based on established oscillometric principles and do not "learn" from a dataset in the way a deep learning model would.
Ask a specific question about this device
(155 days)
The TP500 Infrared Thermometer is intended to measure human body temperature by measuring from 1-3cm distance to the forehead. The device can be used on people of all ages. It is recommended that the device be operated by an adult. The device is intended to be used in the hospital and other clinical environments.
The Infrared thermometer (model TP500) is a hand-held, battery powered, electronic thermometer that converts a subject's forehead temperature, using the infrared energy emitted in the area around the subject's forehead to an oral equivalent temperature when measured from 1-3cm to the center of the subject's forehead with no contact. The device can be used on people of all ages and is intended to be used in the hospital and other clinical environments.
Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided FDA 510(k) summary for the Contec Medical Systems Co., Ltd. Infrared Thermometer (Model TP500).
This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed report of all raw study data. Therefore, some information, especially regarding the specifics of how the ground truth was established for the training set, might not be fully elaborated.
1. Table of Acceptance Criteria and Reported Device Performance
The core performance of an infrared thermometer relates to its accuracy and repeatability. The primary standard for this, as cited, is ASTM E1965-98. The acceptance criteria are implicit in meeting this standard and are explicitly stated in the "Accuracy" row of the comparison table.
| Acceptance Criterion (Standard/Requirement) | Reference | Reported Device Performance (TP500) |
|---|---|---|
| Accuracy | ASTM E1965-98 / User Manual Claim | ±0.3°C (0.5°F) |
| Clinical Bias | ASTM E1965-98 | Evaluated and met performance parameters claimed |
| Clinical Uncertainty | ASTM E1965-98 | Evaluated and met performance parameters claimed |
| Clinical Repeatability | ASTM E1965-98 | Evaluated and met performance parameters claimed |
| Electrical Safety | IEC 60601-1 | Complies |
| Electromagnetic Compatibility (EMC) | IEC 60601-1-2 | Complies |
| Performance (General) | ISO 80601-2-56 | Complies |
| Biocompatibility | ISO 10993-5:2009 & ISO 10993-10:2010 | Complies (materials tested and shown not to affect effectiveness/safety) |
| Software Verification & Validation | FDA Guidance (May 11, 2005) | Met requirements |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: A "controlled human clinical study" was conducted with 108 subjects. These subjects included three age groups:
- Age 0-1 (35 subjects)
- Age 1-5 (33 subjects)
- Age above 5 (40 subjects)
- The study included both febrile and afebrile persons.
- Data Provenance: Not explicitly stated, but given the manufacturer is China-based, it is highly probable the data was collected in China. The study is described as a "controlled human clinical study," implying a prospective design for gathering the clinical accuracy data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This detail is not provided in the summary. For a thermometer, the "ground truth" would typically be established by a highly accurate reference thermometer (e.g., a rectal thermometer or an oral thermometer operated by a trained healthcare professional, depending on the body site), rather than "experts" in the sense of multiple radiologists reviewing images. The summary states that "clinical bias, clinical uncertainty and clinical repeatability have been evaluated per clinical validation for infrared thermometer." This implies a comparison to a gold standard measurement, but the specifics of how many personnel were involved in taking these gold standard measurements or their specific qualifications are not detailed.
4. Adjudication Method for the Test Set
Adjudication methods (like 2+1, 3+1) are typically used for subjective assessments, such as image interpretation, to resolve discrepancies among experts. For objective measurements like temperature, an adjudication method is generally not applicable. Instead, the ground truth is established by a primary, highly accurate measurement method. The summary does not mention any adjudication process.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. MRMC studies are typically for medical imaging systems where human readers interpret images with and without AI assistance. This device is an infrared thermometer, not an imaging system designed for human interpretation with AI assistance. The study conducted was a clinical accuracy study comparing the device to a gold standard measurement, and implicitly, to the predicate device's performance per the equivalence claim.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
The device itself is a standalone thermometer. Its "performance" is inherently "algorithm only" in the sense that it provides a temperature reading directly. The clinical study evaluated the direct performance of the device's sensor and internal processing (algorithm) against a gold standard. So, yes, a standalone performance evaluation of the device's measurement capabilities was performed.
7. The Type of Ground Truth Used
The type of ground truth used was comparative measurements against a reference standard. The clinical study states that "clinical bias, clinical uncertainty and clinical repeatability have been evaluated per clinical validation for infrared thermometer." This means the device's readings were compared against highly accurate, established methods of temperature measurement to determine its accuracy and precision. While not explicitly stated as "rectal temperature" or "oral temperature," for a clinical thermometer, these are the typical gold standards depending on the population and clinical context. The study was conducted in accordance with ASTM E1965-98, which outlines the methodology for such clinical validations including the use of reference thermometers.
8. The Sample Size for the Training Set
Not applicable/Not explicitly stated for an AI training set. This device is a traditional medical device (infrared thermometer), not an AI/ML-based device that learns from a large training dataset in the same way, for example, a diagnostic imaging AI would. Its "algorithm" is based on physics (infrared radiation) and calibrated against known standards, not trained on patient data in an iterative machine learning process. Therefore, there is no "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As explained in point 8, this is not an AI/ML device that requires a training set and associated ground truth in that context. The device's internal calibration and performance are established through engineering design, component selection, and adherence to performance standards like ISO 80601-2-56 and ASTM E1965-98, which involve physical calibrations, not data-driven machine learning training.
Ask a specific question about this device
(258 days)
CONTEC™ Electrocardiograph, ECG90A, is intended to acquire ECG signals from adult or children patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. The proposed devices also have measurements and diagnostic interpretation functions, which are offered to clinician on an advisory basis. Digital Electrocardiographs shall be used in healthcare facilities by doctors and or trained healtheare professionals.
This product is an electrocardiograph collecting 12-lead ECG signal simultaneously and printing ECG waveform with thermal printing system, which features in, recording and displaying ECG waveform with manual or auto mode, measuring and diagnosing ECG waveform parameters automatically, prompting for "Lead off" and "Lack of paper", optional interface languages(Chinese/ English), case database management.
This product is used to assist doctors to analyze and print the ECG waveform of children and adults, which can be applied to ECG room, ward, hospital, etc.
The ECG machine provides medical department human ECG wave group for pattern and rhythm analysis that applied in clinical diagnosis and research.
Essential Performance:Sync collection and display of 12-lead ECG.Display of HR.Print ECG wave with different format,Filter settings,and Auto-measurement of ECG data.
ECG90A is divided into button system, printing system, power supply system, LCD display system, signal collection system, control system
Button system: to collect the user's key operation input, send the control system to process,
Printing system: print the print data sent by the control system, heat the printer head to record the waveform on the thermal printing paper, and feedback whether the lack of paper;
Power system: provide power supply for each system module;
The proposed device adopts two kinds of power supply modes: build-in lithium battery DC power supply and AC-DC adapter power supply.
AC-DC adapter power supply: after the external power adapter connecting with the proposed device, the AC voltage is reduced to 12V via the AC-DC power adapter. The 12V voltage is used to charge the lithium battery via the charging management chip;meanwhile the 12V voltage is reduced to 8.6V via a voltage converter, which is used to power the following circuit and printer. Then the 8.6V voltage is reduced again to 5.0V voltage via the voltage converter, and the 5.0V voltage pass through power isolation and is used to power the signal collection and processing module; meanwhile the 5V voltage will be reduces to 3.3 Vvoltage, which is used to power the control module. Build-in lithium battery DC power supply: when the proposed device is connected with AC-DC adapter power supply, the build-in lithium battery will be charged automatically. The charged battery will power the whole system as DC power supply when the proposed device is not connected with the external AC-DC adapter power supply
LCD display system: display various of settings of waveform information to the user for viewing :
Signal collection system: collect the waveform data,submit the final data to the control system to process; the signal acquisition and process module adopts floating and optical isolation to reduce the external disturbance to signal. After filtering and amplification, the A/C conversion of acquired ECG signal is completed on MCU (Micro Controller Unit); and then the data is transmitted to control system.
after processing the data will be displayed on the LCD screen and the waveform will be printed on the thermal paper. MCU can identify and response the key operation through connecting line, and MCU can connect and communicate with the SD card and USB through the signal line.
Control system: the MCU is used to control the whole system operation. The device parameters setting, the processing of acquired signal, the printing, screen display and charging are all controlled by this module.
The provided document is a 510(k) summary for the CONTEC™ Electrocardiograph (ECG90A). It does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria for AI-related functions. The document explicitly states: "No clinical study is included in this submission."
Therefore, I cannot provide the requested information for acceptance criteria, device performance, sample sizes, ground truth establishment, or any details related to AI or MRMC studies, as these are not present in the provided text.
The information primarily focuses on establishing substantial equivalence to a predicate device (CONTEC™ Electrocardiograph, ECG300G) based on non-clinical tests and technical specifications.
Ask a specific question about this device
(240 days)
CMS600P2 B-Ultrasound Diagnostic System in a general-purpose, digital ultrasound diagnostic system for abdomen,gynecology, obstetric, urology,and small-parts application.
The system is intended to use for the following type of studies: fetal organ, abdominal, pediatric, small organs, neonatal cephalic, transvaginal, peripheral vascular,and musculo-skeletal (both conventional and superficial). The device is intended to adult, pregnant woman, pediatric and neonate.
The system is a prescription device intended to be used by or on the order of a physician or similar qualified health care professional.
This device is not intended for home use.
The apparatus is high resolution ultrasound diagnostic system, which adopts digital beam forming(DBF), real-time dynamic aperture(RDA), real-time dynamic beam apodization (DRA), real-time dynamic receive focusing (DRF), digital-control dynamic frequency scanning (DFS), frame correlation etc. technologies. The image is clear, stable, and high resolution.
It adopts embedded operating system, which greatly improve the product security and compatibility, and enhance the data processing ability.
The apparatus is taken more conveniently for its humane operation interface and management system, flexible expansibility and compatibility, pop-up menu and keyboard design.
Users can chose different kinds of display modes,such as:B, 2B, 4B, BM, M. Image real-time, freeze, memory, calling up and large capability cine loop can be realized; the device possesses multilevel scanning depth, dynamic range, sound output power and focus number, focus space, focus position adjusting etc. functions.
With controllable frame correlation, gamma correction, image reversing, histogram, depth advance, zoom magnification, no distortion real-time part magnification etc. image processing functions. It is convenient for choice of optimal diagnostic image with image pigeonhole, browse, management functions. Measure, calculation, and report features are provided to facilitate image measurement.
The device supports linear probe and convex probe. The frequency of the device is 2.0 MHz-10.0 MHz. It has extensive application that is suitable for clinic diagnosis, such as abdomen, obstetrics, gynecology, small organs, cardiology and urology etc. .
The device is consist of three parts: mainframe, transducer(probe), adapter.
The provided text describes the 510(k) submission for the CMS600P2 B-Ultrasound Diagnostic System, asserting its substantial equivalence to a predicate device (EMP-2100 Full Digital Ultrasound Diagnostic Device).
Important Note: This document focuses on demonstrating substantial equivalence to a predicate device, as opposed to proving novel performance against specific clinical acceptance criteria through a clinical trial with a human-in-the-loop AI system. Therefore, the information provided will pertain to demonstrating equivalence to an existing medical device's performance, rather than an AI's performance against specific clinical metrics. Many of the requested details, such as "effect size of human readers improving with AI vs. without AI assistance," "standalone (algorithm only) performance," or "adjudication method for the test set," are typically associated with AI/ML-driven devices and are not applicable to this traditional ultrasound system's 510(k) submission.
Here's an analysis based on the provided document:
Acceptance Criteria and Device Performance (Based on Substantial Equivalence)
The acceptance criteria for the CMS600P2 B-Ultrasound Diagnostic System are primarily based on demonstrating substantial equivalence (SE) to a legally marketed predicate device (EMP-2100 Full Digital Ultrasound Diagnostic Device) in terms of:
- Intended Use and Indications for Use
- Technological Characteristics (e.g., imaging modes, frequency range, resolution, safety standards)
- Safety and Effectiveness (demonstrated through non-clinical testing and comparison to predicate)
The performance of the proposed device is assessed against the performance characteristics of the predicate device.
Table of Acceptance Criteria and Reported Device Performance (Comparison to Predicate Device)
| Acceptance Criterion (Category / Characteristic) | Predicate Device Performance (EMP-2100) | Proposed Device Performance (CMS600P2) | Substantial Equivalence (SE) / Remark |
|---|---|---|---|
| General | |||
| Device Name | EMP-2100 Full Digital Ultrasound Diagnostic Device | CMS600P2 B-Ultrasound Diagnostic System | / (Different names, but comparable) |
| Classification Name | System, Imaging, Pulsed Echo, Ultrasonic; Transducer, Ultrasonic, Diagnostic | System, Imaging, Pulsed Echo, Ultrasonic; Transducer, Ultrasonic, Diagnostic | SE |
| Product Code | IYO, ITX | IYO, ITX | SE |
| Regulation Number | 892.1560, 892.1570 | 892.1560, 892.1570 | SE |
| Intended Use | General-purpose, digital ultrasound diagnostic system for various applications (abdomen, gynecology, obstetric, urology, small-parts, cardiology, etc.) | General-purpose, digital ultrasound diagnostic system for various applications (abdomen, gynecology, obstetric, urology, small-parts, cardiology, etc.) | SE (Similar intended use) |
| Technical Specifications (Probe specific examples) | |||
| C3.5-80R60-A16A (Convex Probe) | Comparable Predicate Data Listed | ||
| Frequency | 3.5MHz (for C128-50 Convex) | 3.5MHz | Analyse 1 (Differences in performance specs, but tested for accuracy and effectiveness) |
| Detect Depth | ≥200 mm (C128-50 Convex) | ≥160 mm | Analyse 1 |
| Resolution (Lateral) | ≤3(depth≤80) ≤4(80 ≤130) (for C128-50 Convex) | ≤3(Depth≤80) ≤4(80≤130) | Analyse 1 |
| Resolution (Axial) | ≤1 (depth≤50) ≤1 (depth≤40) (for C128-50 Convex) | ≤2(Depth≤80) ≤3(80 ≤130) | Analyse 1 |
| L7.5-80L40-A16A (Linear Probe) | Comparable Predicate Data Listed | ||
| Frequency | 6.5MHz (L096-42C Linear) | 7.5MHz | Analyse 1 |
| Detect Depth | ≥90 mm (L096-42C Linear) | ≥50 mm | Analyse 1 |
| Resolution (Lateral) | ≤2(depth≤60) (L096-42C Linear) | ≤2(Depth≤40) | Analyse 1 |
| Resolution (Axial) | ≤0.5(depth≤60) (L096-42C Linear) | ≤1 (Depth≤50) | Analyse 1 |
| E6.5-80R13-A16A (Endo-vaginal Probe) | Comparable Predicate Data Listed | ||
| Frequency | 6.5MHz (C080-13G Micro Convex) | 6.5MHz | Analyse 1 |
| Detect Depth | ≥90 mm (C080-13G Micro Convex) | ≥40 mm | Analyse 1 |
| Resolution (Lateral) | ≤2(depth≤60) (C080-13G Micro Convex) | ≤2(Depth≤30) | Analyse 1 |
| Resolution (Axial) | ≤0.5(depth≤40) (C080-13G Micro Convex) | ≤1 (Depth≤40) | Analyse 1 |
| System Features | |||
| Monitor Size | 10 inch | 10.1 inch | SE |
| Display Mode | B, B+B, B+M, M, 4B | B, 2B, BM, M, 4B | SE (Similar) |
| Image gray scale | 256 level | 256 level | SE |
| Image Storage | 16 | 2048 frame | Analyse 1 (Internal memory difference, but tested functionality) |
| Cine Loop | 256 frame | 600 frame | Analyse 1 (Internal memory difference, but tested functionality) |
| Image Flip | Up/down, left/right | Up/down, left/right, black/white | Analyse 1 (Added capability but main function similar) |
| Image Process | Various | Various | Analyse 1 (Different specific functions, but overall image processing similar) |
| Measurement | Distance, circumference, area, volume, histogram, profile, time velocity heart rate | Distance, circumference, area, volume, heart, pregnant age, fetal weight, expected date | Analyse 1 (Different specific measurements, but overall measurement capability similar) |
| Notation | Date, Time, Name, Sex, Age, Doctor, Hospital name, probe frequency, etc. | Date, time, name, No., sex, age, doctor, hospital name, probe frequency, etc. | SE |
| Net weight | 10kg | 2.3 kg (include probe) | SE (Weight difference not considered to impact safety/efficacy if performance is SE) |
| Power supply | AC 100-240V, 47-60Hz | DC15V; Adapter:100 V~240 V, 50 Hz/60 Hz | Analyse 2 (Difference in power supply, but both passed IEC60601-1 safety test) |
| Dimensions (mm) | 365 (H) X 292 (W) X 380 (D) | 292 (L) × 232 (W) × 45 (H) | SE (Differences in size, but doesn't affect SE of core function) |
| Configuration | mainframe, transducer(probe) | mainframe, transducer(probe), adapter | SE (Both include core components) |
| Safety & Standards Compliance | |||
| Biocompatibility | ISO10993-5 & ISO10993-10 | ISO10993-5 & ISO10993-10 | SE |
| Electrical Safety | IEC60601-1 | IEC60601-1 | SE |
| EMC | IEC60601-1-2 | IEC60601-1-2 | SE |
| Performance Standards | UD2 | IEC 60601-2-37; IEC61157 | Analyse 1 (Different specific standards, but both demonstrate performance) |
| Acoustic Output | Meet the requirements of Track 1 | Meet the requirements of Track 3 | Analyse 1 (Different tracks, but both demonstrate compliance) |
| Skin Contacted Material | Probe Cover (ABS), Acoustic Lens (Silicone elastomer) | Probe Cover (ABS), Acoustic Lens (Silicone elastomer) | SE |
| Operation Environment | Temperature 5°C -40°C, RH up to 90%RH, Air pressure 70-106Kpa | Temperature: +10°C~ +40°C, RH: 30%~ 75%, Atmosphere pressure: 700 hPa ~ 1060 hPa | SE (Within acceptable ranges) |
| Storage environment | Temperature -5-40°C, RH up to 90%RH (no water drop), Air pressure 70-106Kpa | Temperature: -10°C~+55°C, RH: ≤93%, no condensation, Atmosphere pressure: 700 hPa ~ 1060 hPa | SE (Within acceptable ranges) |
| Sterile | No | No | SE |
| Single Use | No | No | SE |
Study Proving Acceptance Criteria Met: Non-Clinical Testing and Comparison
The study proving the device meets the acceptance criteria is primarily a non-clinical testing and comparison study against a predicate device, as explicitly stated in the document.
1. Sample Size Used for the Test Set and Data Provenance:
- Test Set: No separate "test set" in the sense of a distinct dataset for an algorithm's performance evaluation is mentioned, as this is a physical diagnostic device, not an AI/ML product. The "test set" would implicitly refer to the device itself undergoing various performance and safety tests.
- Data Provenance: Not applicable in the context of imaging data. The provenances are the device specifications and direct measurements from non-clinical tests conducted on the proposed device and compared to the predicate's known specifications and performance.
- Retrospective/Prospective: Not specified for the "data" as it pertains to device testing against engineering specifications and regulatory standards. The non-clinical tests were conducted on the proposed device.
2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable. This submission is for a traditional ultrasound diagnostic system, not an AI/ML-driven device requiring expert consensus for image interpretation or "ground truth" establishment beyond the standard physical and electrical performance measurements. The "ground truth" here is the adherence to engineering specifications and safety standards, as well as the established performance of the predicate device.
3. Adjudication Method for the Test Set:
- Not applicable. No human interpretation or adjudication of diagnostic outputs from the device is described for the purpose of establishing clinical ground truth for a test set. Adjudication methods are typically relevant for studies evaluating human reader performance, especially with AI assistance.
4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- No. An MRMC study was not conducted. The submission explicitly states: "No clinical study is included in this submission." This type of study is common for AI-assisted diagnostic tools to show improvement in human reader performance, which is not the scope of this 510(k) for a basic ultrasound system.
5. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
- Not applicable. This is not an AI/ML algorithm requiring standalone performance evaluation. It is an ultrasound diagnostic system where the user (physician) interprets the images.
6. The Type of Ground Truth Used:
- For the non-clinical tests (e.g., electrical safety, EMC, performance parameters like resolution and depth): The "ground truth" is established by engineering specifications, international consensus standards (IEC, ISO), and the documented performance characteristics of the legally marketed predicate device.
- For biocompatibility: The "ground truth" is established by ISO 10993 series standards for in vitro cytotoxicity, irritation, and delayed-type hypersensitivity.
7. The Sample Size for the Training Set:
- Not applicable. This is not an AI/ML device that requires a "training set" of data.
8. How the Ground Truth for the Training Set Was Established:
- Not applicable. As no training set for an AI/ML algorithm is involved, this question is irrelevant.
Ask a specific question about this device
(335 days)
The Pulse Oximeter CMS50EW is a non-invasive device intended for spot-check or continuous monitoring of non-invasive oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate through the finger of adult patients in home and hospital environments (including clinical use internist/surgery, anesthesia, and intensive care settings). The device is reusable and not intended for out-of-hospital transport use.
The proposed device, Pulse Oximeter CMS50EW is a battery powered fingertip device, which can display % SpO2, pulse rate value and pulse strength. It is based on digital blood oxygen technology.
The proposed device is composed of power module, signal acquisition module, control and signal processing module. The power source of it is a built-in rechargeable lithium battery. And it has alarm function, to raise the user's attention for exceeding of physiological limit and device error via visual and audio alarming indicator. And the pulse oximeter can communicate with computer by USB data wire and Bluetooth.
The proposed device, Pulse Oximeter CMS50EW, is a modification to the predicate device, CMS50E, cleared under K090671; the modification is to add the blue tooth functionality, however, the oximetry technology including the sensor was unchanged.
The provided text focuses on the 510(k) summary for the Contec Medical Systems Co., Ltd.'s Pulse Oximeter CMS50EW, which is seeking substantial equivalence to its predicate device, the Pulse Oximeter CMS50E (K090671). The key modification is the addition of Bluetooth functionality, while the oximetry technology and sensor remain unchanged.
Here's an analysis of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria provided are primarily in the context of performance specifications for SpO2 and Pulse Rate (PR) measurement, which are identical to the predicate device. The "reported device performance" is implicitly that the device meets these existing specifications due to the unchanged oximetry technology.
| Metric | Acceptance Criteria (Predicate) | Reported Device Performance (Proposed) |
|---|---|---|
| SpO2 Accuracy | 70% | 70% |
| SpO2 Range | 0%-100% (Same for predicate) | 0%-100% (Identical to predicate) |
| PR Measurement Range | 30bpm~250bpm (Same for predicate) | 30bpm~250bpm (Identical to predicate) |
| PR Accuracy | ±2bpm or ±2% (select the larger) (Same for predicate) | ±2bpm or ±2% (select the larger) (Identical to predicate) |
| Electrical Safety | Comply with IEC 60601-1 (Same for predicate) | Complies with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2 (Same for predicate) | Complies with IEC 60601-1-2 |
| Wireless Function | Complies with FDA guidance for Radio Frequency Wireless Technology | Wireless transmission performance, data integrity, wireless coexistence, and security of wireless data complied with acceptance criteria. |
2. Sample size used for the test set and the data provenance
The document explicitly states: "Clinical testing was not necessary in this submission to support the device modification because the oximetry technology including the sensor is unchanged from the predicate device."
Therefore, there is no information provided regarding a specific clinical test set, sample size, or data provenance (country of origin, retrospective/prospective) for this device modification. The equivalence is based on the predicate device's established performance and non-clinical testing for the new Bluetooth functionality.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
As no clinical testing was deemed necessary for the oximetry performance, there is no mention of experts establishing ground truth for a test set related to SpO2 or PR accuracy for the CMS50EW. Any such data would pertain to the original clearance of the predicate device (K090671) and is not detailed here.
4. Adjudication method for the test set
Not applicable, as no clinical test set for oximetry performance was performed for this submission.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a pulse oximeter, not an AI-assisted diagnostic tool that would involve human readers or MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to the performance of the device's core functionality (SpO2 and PR measurement). The device's performance is standalone in the sense that it provides direct measurements. The non-clinical tests conducted (IEC standards, wireless transmission tests) assess this standalone performance. No detailed "algorithm only" performance separate from the device's integrated operation is described in a way that suggests a disaggregated study.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the SpO2 and PR measurements, the ground truth would typically be established by established measurement techniques (e.g., co-oximetry for SpO2, ECG for heart rate) during the original clinical validation of the predicate device. However, this specific submission does not detail the ground truth methodology for this device because it relies on the predicate's established oximetry technology. For the new wireless functionality, the "ground truth" would be the expected technical specifications and performance indicators for wireless transmission.
8. The sample size for the training set
Not applicable, as this is not an AI/machine learning device requiring a training set in the conventional sense. The device's oximetry technology is formula-based, using the Lambert-Beer Law.
9. How the ground truth for the training set was established
Not applicable for the same reason as point 8. The device's operation is based on established biophysical principles and calibrated with known values, not trained on a dataset.
Ask a specific question about this device
(53 days)
CONTEC08C Electronic Sphygmomanometer is intended to measure the systolic, diastolic and mean blood pressure as well as pulse rate via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm. It can be used on adult individuals.
The proposed device, CONTEC08C Electronic Sphygmomanometer is battery driven automatic non-invasive Blood Pressure Monitor. The device can automatically complete the inflation, deflation and BP measurement, which can measure systolic, diastolic and mean blood pressure as well as the pulse rate at upper arm within its claimed range and accuracy via the oscillometric technique. User can select the unit of the measurement: mmHg or kPa. It can be only used on adult individuals.
The device has the data storage function for data review including measurement time, systolic blood pressure, diastolic blood pressure, mean blood pressure and pulse rate. These records can be uploaded to PC via USB and processed with the PC software.
The device has technical alarm function which will be triggered when the battery voltage is lower than 3.7V, and this alarm can not be cancelled unless being closed or the power replaced.
Here's an analysis of the provided text regarding the acceptance criteria and study for the CONTEC08C Electronic Sphygmomanometer:
The provided document describes the clearance of a non-invasive blood pressure monitor (sphygmomanometer) through the 510(k) pathway, which focuses on demonstrating substantial equivalence to a legally marketed predicate device. This type of submission generally relies on performance testing against recognized standards rather than extensive clinical efficacy studies typically seen for novel devices.
Acceptance Criteria and Reported Device Performance
The device's performance is primarily assessed against the recognized standard ANSI/AAMI SP10: 2002 & A1: 2003 & A2: 2006. This standard outlines requirements for automatic sphygmomanometers, including accuracy specifications.
| Acceptance Criteria (from ANSI/AAMI SP10) | Reported Device Performance |
|---|---|
| Blood Pressure Accuracy | |
| - Mean difference: ≤ ±5 mmHg | Met standard requirements |
| - Standard deviation: ≤ 8 mmHg | Met standard requirements |
| Pulse Rate Accuracy | |
| (Not explicitly stated in provided text, but implied to be met per the standard.) | Met standard requirements |
| Other safety and performance standards | |
| - IEC 60601-1: 1988 +A1: 1991 +A2: 1995 (General safety) | Complies with standard |
| - IEC 60601-1-2: 2007 (EMC) | Complies with standard |
Note: The provided text states that "Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: ... ANSI/AAMI SP10: 2002 & A1: 2003 & A2: 2006." While it confirms compliance, specific numerical performance results (e.g., the exact mean difference and standard deviation achieved) are not detailed in this summary. However, meeting ANSI/AAMI SP10 implicitly means meeting its accuracy requirements for blood pressure (mean difference ≤ ±5 mmHg and standard deviation ≤ 8 mmHg).
Study Details
Given the nature of a 510(k) for a sphygmomanometer tested against ANSI/AAMI SP10, the "study" referred to is a specific blood pressure accuracy validation protocol as defined by that standard.
-
Sample size used for the test set and the data provenance:
- Sample Size: The document does not explicitly state the number of subjects (the sample size) used for the blood pressure accuracy testing according to ANSI/AAMI SP10. The standard itself specifies minimum subject requirements for various test phases (e.g., 85 subjects for the clinical validation of accuracy).
- Data Provenance: The document states that "Bench tests were conducted." The sponsor and submission correspondent are based in China (CONTEC MEDICAL SYSTEMS CO., LTD and Mid-Link Consulting Co., Ltd.), which suggests the testing was likely conducted in China. The data is prospective, meaning it was collected specifically for the purpose of validating the device.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- For blood pressure validation, the "ground truth" is typically established by trained observers (auscultators) using a mercury sphygmomanometer or validated reference device according to a standardized protocol. ANSI/AAMI SP10 requires at least two trained observers to simultaneously obtain reference blood pressure measurements.
- Qualifications of experts: These observers must be specifically trained and certified as competent in auscultatory blood pressure measurement. While the document doesn't detail their specific qualifications, compliance with ANSI/AAMI SP10 implies that appropriately qualified personnel were used.
-
Adjudication method for the test set:
- ANSI/AAMI SP10 specifies an adjudication method for observer differences. Typically, if the two observers' readings differ by more than a predefined amount (e.g., 4 mmHg), a third observer or a re-measurement may be required, or the data point may be discarded. The exact protocol would follow the standard.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is an automatic electronic sphygmomanometer and does not involve "human readers" interpreting images or data with or without AI assistance in the way an MRMC study would apply. Its function is to provide direct measurements. Therefore, an MRMC comparative effectiveness study was not performed and is not applicable here.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the primary testing described is a standalone performance test of the device. The device's algorithm (oscillometric technique) measures blood pressure automatically without human input during the measurement phase. The "bench tests" and compliance with ANSI/AAMI SP10 are evaluations of this standalone performance.
-
The type of ground truth used:
- For blood pressure devices, the ground truth is expert auscultatory measurements (or another validated reference method) conducted concurrently with the device measurements on human subjects, as dictated by the ANSI/AAMI SP10 standard.
-
The sample size for the training set:
- The document does not refer to a distinct "training set" for an algorithm to be learned or optimized in the sense of machine learning. The device utilizes an oscillometric algorithm which is a well-established technique for blood pressure measurement. Any internal calibration or algorithm refinement by the manufacturer would have been part of the device's design and development, not typically described as a "training set" in a 510(k) summary for such a device. The testing described is validation of the final device.
-
How the ground truth for the training set was established:
- As there is no "training set" explicitly mentioned or applicable in the context of an AI/machine learning model as typically discussed, this question is not relevant to the provided information. The device's underlying algorithm is based on established oscillometric principles rather than a data-driven "training" process from a ground truth dataset in the AI sense.
Ask a specific question about this device
Page 1 of 1