Found 2 results

510(k) Data Aggregation

    K Number
    K230524
    Date Cleared
    2024-01-19

    (326 days)

    Product Code
    Regulation Number
    890.5650
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Chengdu Cryo-Push Medical Technology Co.,Ltd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cold Compression Wrap Pro is indicated for the temporary relief of minor muscle aches and pains.

    The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneading and stroking of tissues using an inflatable wrap.

    The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.

    Device Description

    Cold Compression Wrap Pro Device consists of a main unit and wraps which simulates kneading and stroking of tissue with the hands by use of inflatable pressure wraps. It is for temporary increase in circulation of the treated areas and temporary relief of minor muscle aches and pains. By inflating the air chambers and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient.

    There are 4 different wraps for Shoulder, elbow/arm, low limbs (Calf and upper leg), ankle/foot. The wrap contains cold pack, and the cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.

    When the device is used for 30 minutes (Default time) at the operature (5°C~ 40°C) specified in the user manual, the temperature of the cold pack is (-18°C ~-4℃) (±2℃).

    AI/ML Overview

    The provided document, K230524, is a 510(k) Premarket Notification for a medical device called "Cold Compression Wrap Pro". This document, issued by the FDA, determines the substantial equivalence of the new device to existing legally marketed predicate devices.

    Based on the provided document, here's a breakdown of the acceptance criteria and the study proving the device meets them:

    Crucially, this is not a clinical study report for an AI/ML medical device. It's a 510(k) submission for a physical medical device (cold compression wrap) with embedded software. Therefore, many of the typical acceptance criteria and study details for AI/ML performance (like MRMC studies, ground truth establishment by experts, and large training/test sets of medical data) are not applicable or present in this document.

    The "study" here refers to the non-clinical performance testing conducted to demonstrate that the device functions as intended and is safe.


    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of "acceptance criteria" with quantitative performance metrics for the device's therapeutic effect (e.g., pain reduction scores, circulation improvement). Instead, it focuses on the device's technical specifications and safety parameters, which serve as its de facto acceptance criteria for this type of device.

    Here's an attempt to extract the closest equivalents to acceptance criteria and reported performance based on the "510(k) Summary" section, particularly the "Substantial Equivalence Comparison" table and "Summary of Non-clinical Performance Testing":

    Acceptance Criteria (Stated or Implied)Reported Device Performance (Subject Device)Comments / Basis for Performance Claims
    Electrical SafetyComplies with IEC 60601-1:2005, AMDI:2012Non-clinical testing
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2:2014Non-clinical testing
    Software ValidationSoftware verification and validation testing conducted; considered "moderate" level of concern.Non-clinical testing; followed FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
    Product Appearance and SizeTested according to internal standardsNon-clinical testing
    Wrap PerformanceTested according to internal standardsNon-clinical testing
    Function TestTested according to internal standardsNon-clinical testing
    Intended Use Equivalence"The Cold Compression Wrap Pro is indicated for the temporary relief of minor muscle aches and pains. The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneading and stroking of tissues using an inflatable wrap. The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable."Claimed "Identical" to predicate devices (K222669 & K133483).
    Operating Temperature of Cold Pack-18℃ ~ -4℃ (±2℃)Maintained consistency with predicate (K222669).
    Main Unit Operating Temperature Range5℃ ~ 40℃ (41°F ~ 104°F)Identical to predicate (K222669).
    Pressure Range0-100 mmHgIdentical to predicate (K222669).
    Pressure Levels20, 40, 60, 80, 100 mmHgIdentical to predicate (K222669).
    Pressure Error Range±15 mmHgIdentical to predicate (K222669).
    Noise Level≤ 55dBIdentical to predicate (K222669).
    BiocompatibilityNot conducted due to use of same patient contact material as predicate device with same nature and duration of contact.Equivalence argument; implied acceptance criteria met by predicate.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document describes non-clinical performance testing (electrical safety, EMC, software validation, product appearance/size, wrap performance, function tests). These tests are typically performed on a small sample of manufactured devices (e.g., a few units, or internal test benches). The specific number is not specified in this summary.
    • Data Provenance: The tests are "non-clinical" and likely conducted in a lab setting by the manufacturer, Chengdu Cryo-Push Medical Technology Co.,Ltd, which is located in China. The data is prospective, in the sense that the company specifically conducted these tests for the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is a physical device, not an AI/ML diagnostic algorithm. Therefore, the concept of "experts establishing ground truth" in the clinical sense (e.g., radiologists interpreting images) is not applicable. The "ground truth" for this device's performance is established by engineering measurements and adherence to recognized standards (e.g., IEC 60601-1 for electrical safety). The qualifications of the testing personnel are not disclosed in this summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as there is no human review or adjudication of data points in the way it would occur for AI/ML diagnostic algorithms (e.g., reviewing medical images). The testing involves objective measurements against predefined engineering specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML diagnostic tools that assist human readers (e.g., radiologists). This product is a physical therapeutic device, not a diagnostic AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable in the context of an AI/ML algorithm. The device has "embedded software" which underwent "software validation". This is a test of the embedded software's functionality and safety (e.g., reliably controlling pressure and time settings), not a standalone diagnostic AI performance evaluation. The device itself operates "as is" when used by a patient, meaning the "algorithm" (software) operates in a standalone manner as part of the device's function, but not as a diagnostic tool.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is based on engineering specifications, recognized international safety standards (e.g., IEC 60601 series), and the functional requirements for a cold compression system. For instance, the ground truth for electrical safety is whether the device actually meets the voltage leakage limits of IEC 60601-1. For the cold pack temperature, it's whether the internal temperature sensor actually registers within the specified range after 30 minutes of operation. This is determined by instrumental measurements and adherence to test protocols, not clinical outcomes or expert consensus on medical data.

    8. The sample size for the training set

    The device is a physical product with embedded software. It is not an AI/ML device that requires a "training set" of data in the common sense (e.g., images for deep learning). The software is likely programmed deterministically or based on engineered control logic, not "trained" on data. Therefore, the concept of a "training set sample size" is not applicable.

    9. How the ground truth for the training set was established

    Since there is no "training set" in the AI/ML sense, this question is not applicable. The "ground truth" for the device's design and software functionality is rooted in engineering requirements, safety standards, and the intended physiological effects of cold compression therapy (which are well-established prior to this device's development).

    Ask a Question

    Ask a specific question about this device

    K Number
    K222669
    Date Cleared
    2022-12-05

    (90 days)

    Product Code
    Regulation Number
    890.5650
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Chengdu Cryo-Push Medical Technology Co., Ltd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A02-P-001:

    The Cryopush Cold Compression Device is indicated for the temporary relief of minor muscle aches and pains. The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneading and stroking of tissues using by an inflatable wrap.

    The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.

    A02-P-002:

    The Cryopush A02-P-002 is indicated for the temporary relief of minor muscle aches and pains.

    The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneading and stroking of tissues using by an inflatable wrap.

    Device Description

    Cryopush Cold Compression Device consists of a main unit and wraps which simulates kneading and stroking of tissue with the hands by use of inflatable pressure wraps. It is for temporary increase in circulation of the treated areas and temporary relief of minor muscle aches and pains. By inflating the air chambers and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The device includes two models A02- P-001 and A02- P-002. The A02- P-001 has one chamber, and A02- P-002 has 2 chambers. The A02- P-001 contains cold pack, and the cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.

    When the A02- P-001 is used for 30 minutes (Default time) at the operating temperature (5°C~ 40°C) specified in the user manual, the temperature of the cold pack is (-18℃~-4℃) (±2℃).

    AI/ML Overview

    The provided document is a 510(k) summary for the "Cryopush Cold Compression Device" (K222669). It details the device's characteristics and its substantial equivalence to a predicate device. However, it does not contain information about a study proving the device meets specific acceptance criteria in the sense of clinical performance or diagnostic accuracy.

    Instead, the document focuses on demonstrating substantial equivalence through non-clinical testing and comparison of technological characteristics. The "acceptance criteria" discussed are related to manufacturing standards, electrical safety, electromagnetic compatibility, and basic functional performance, not clinical efficacy or diagnostic accuracy.

    Therefore, many of the requested points about "acceptance criteria" for performance, sample size, ground truth, experts, and comparative effectiveness studies are not applicable or cannot be extracted from this document, as it outlines a regulatory submission based on substantial equivalence to a predicate device rather than a de novo clinical trial demonstrating novel performance against defined criteria.

    Here's the information that can be extracted based on the provided text, with notes for items not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Internal Standards based on Predicate)Reported Device Performance/Characteristics
    Electrical Safety, EMCDevice complies with IEC 60601-1:2005, AMDI:2012, IEC 60601-1-11:2015, IEC 60601-1-2:2014
    Software ValidationSoftware document determined according to FDA guidance (May 11, 2005) for software in medical devices
    Pressure RangeA02-P-001: 0-100mmHg; A02-P-002: 0-215mmHg (Predicate: 0-240 mmHg)
    Pressure LevelsA02-P-001: 20, 40, 60, 80, 100 mmHg; A02-P-002: 100, 160, 215 mmHg (Predicate: 150, 185, 215 mmHg)
    Pressure Error RangeA02-P-001: ±15mmHg; A02-P-002: ±20mmHg (Predicate: ±25mmHg)
    Cold Pack Temperature (A02-P-001)-18℃ to -4℃ (±2℃) within 30 minutes of operation at 5°C-40°C operating temperature
    Noise Level≤ 55dB (Predicate: ≤ 65dB)
    Working TimeA02-P-001: 10-120 min (default 30 min); A02-P-002: 10-60 min (default 30 min) (Predicate: 20 minutes)
    Keep time (Pressure hold time)10s (Predicate: 1-5s)
    Deflation time20s (Predicate: 1-5s)
    Product Appearance and Size, Wrap Performance, Function Test (Internal Standards)Tested according to internal standards; specific metrics or results are not provided in this summary.
    No adverse impact on safety and effectiveness compared to predicate despite differencesConcluded that differences in cold pack (A02-P-001) and treatment time do not raise adverse impact on safety and effectiveness.
    Overall substantial equivalence to predicate device (K193354)Concluded based on comparison of intended use, technological characteristics, and performances.

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: Not applicable. The document describes non-clinical engineering and performance testing against standards, not a clinical test set with human subjects for performance evaluation.
    • Data provenance: Not applicable. The testing is described as internal testing by the manufacturer ("Chengdu Cryo-Push Medical Technology Co., Ltd") to conform to engineering standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. Ground truth, in the context of expert consensus, is not relevant for the type of engineering and performance testing described. The "ground truth" here is adherence to specified engineering standards and internal performance metrics.

    4. Adjudication method for the test set

    • Not applicable. No expert adjudication process is described for this type of non-clinical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a physical medical device (cold compression device), not an AI-powered diagnostic or assistive technology. Therefore, MRMC studies and "human readers improving with AI" are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is not an algorithm or AI-based device.

    7. The type of ground truth used

    • For engineering and performance tests: The ground truth implicitly refers to the specifications defined by relevant standards (e.g., IEC 60601 series for electrical safety, internal product specifications for pressure, noise, temperature) and the characteristics of the predicate device for comparison of substantial equivalence. There is no "expert consensus," "pathology," or "outcomes data" ground truth established for this type of submission.

    8. The sample size for the training set

    • Not applicable. There is no software or algorithm that would require a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    • Not applicable. No training set is mentioned or implied.