(326 days)
The Cold Compression Wrap Pro is indicated for the temporary relief of minor muscle aches and pains.
The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneading and stroking of tissues using an inflatable wrap.
The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
Cold Compression Wrap Pro Device consists of a main unit and wraps which simulates kneading and stroking of tissue with the hands by use of inflatable pressure wraps. It is for temporary increase in circulation of the treated areas and temporary relief of minor muscle aches and pains. By inflating the air chambers and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient.
There are 4 different wraps for Shoulder, elbow/arm, low limbs (Calf and upper leg), ankle/foot. The wrap contains cold pack, and the cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
When the device is used for 30 minutes (Default time) at the operature (5°C~ 40°C) specified in the user manual, the temperature of the cold pack is (-18°C ~-4℃) (±2℃).
The provided document, K230524, is a 510(k) Premarket Notification for a medical device called "Cold Compression Wrap Pro". This document, issued by the FDA, determines the substantial equivalence of the new device to existing legally marketed predicate devices.
Based on the provided document, here's a breakdown of the acceptance criteria and the study proving the device meets them:
Crucially, this is not a clinical study report for an AI/ML medical device. It's a 510(k) submission for a physical medical device (cold compression wrap) with embedded software. Therefore, many of the typical acceptance criteria and study details for AI/ML performance (like MRMC studies, ground truth establishment by experts, and large training/test sets of medical data) are not applicable or present in this document.
The "study" here refers to the non-clinical performance testing conducted to demonstrate that the device functions as intended and is safe.
1. A table of acceptance criteria and the reported device performance
The document does not present a formal table of "acceptance criteria" with quantitative performance metrics for the device's therapeutic effect (e.g., pain reduction scores, circulation improvement). Instead, it focuses on the device's technical specifications and safety parameters, which serve as its de facto acceptance criteria for this type of device.
Here's an attempt to extract the closest equivalents to acceptance criteria and reported performance based on the "510(k) Summary" section, particularly the "Substantial Equivalence Comparison" table and "Summary of Non-clinical Performance Testing":
| Acceptance Criteria (Stated or Implied) | Reported Device Performance (Subject Device) | Comments / Basis for Performance Claims |
|---|---|---|
| Electrical Safety | Complies with IEC 60601-1:2005, AMDI:2012 | Non-clinical testing |
| Electromagnetic Compatibility (EMC) | Complies with IEC 60601-1-2:2014 | Non-clinical testing |
| Software Validation | Software verification and validation testing conducted; considered "moderate" level of concern. | Non-clinical testing; followed FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" |
| Product Appearance and Size | Tested according to internal standards | Non-clinical testing |
| Wrap Performance | Tested according to internal standards | Non-clinical testing |
| Function Test | Tested according to internal standards | Non-clinical testing |
| Intended Use Equivalence | "The Cold Compression Wrap Pro is indicated for the temporary relief of minor muscle aches and pains. The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneading and stroking of tissues using an inflatable wrap. The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable." | Claimed "Identical" to predicate devices (K222669 & K133483). |
| Operating Temperature of Cold Pack | -18℃ ~ -4℃ (±2℃) | Maintained consistency with predicate (K222669). |
| Main Unit Operating Temperature Range | 5℃ ~ 40℃ (41°F ~ 104°F) | Identical to predicate (K222669). |
| Pressure Range | 0-100 mmHg | Identical to predicate (K222669). |
| Pressure Levels | 20, 40, 60, 80, 100 mmHg | Identical to predicate (K222669). |
| Pressure Error Range | ±15 mmHg | Identical to predicate (K222669). |
| Noise Level | ≤ 55dB | Identical to predicate (K222669). |
| Biocompatibility | Not conducted due to use of same patient contact material as predicate device with same nature and duration of contact. | Equivalence argument; implied acceptance criteria met by predicate. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The document describes non-clinical performance testing (electrical safety, EMC, software validation, product appearance/size, wrap performance, function tests). These tests are typically performed on a small sample of manufactured devices (e.g., a few units, or internal test benches). The specific number is not specified in this summary.
- Data Provenance: The tests are "non-clinical" and likely conducted in a lab setting by the manufacturer, Chengdu Cryo-Push Medical Technology Co.,Ltd, which is located in China. The data is prospective, in the sense that the company specifically conducted these tests for the 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is a physical device, not an AI/ML diagnostic algorithm. Therefore, the concept of "experts establishing ground truth" in the clinical sense (e.g., radiologists interpreting images) is not applicable. The "ground truth" for this device's performance is established by engineering measurements and adherence to recognized standards (e.g., IEC 60601-1 for electrical safety). The qualifications of the testing personnel are not disclosed in this summary.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable as there is no human review or adjudication of data points in the way it would occur for AI/ML diagnostic algorithms (e.g., reviewing medical images). The testing involves objective measurements against predefined engineering specifications.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML diagnostic tools that assist human readers (e.g., radiologists). This product is a physical therapeutic device, not a diagnostic AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable in the context of an AI/ML algorithm. The device has "embedded software" which underwent "software validation". This is a test of the embedded software's functionality and safety (e.g., reliably controlling pressure and time settings), not a standalone diagnostic AI performance evaluation. The device itself operates "as is" when used by a patient, meaning the "algorithm" (software) operates in a standalone manner as part of the device's function, but not as a diagnostic tool.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is based on engineering specifications, recognized international safety standards (e.g., IEC 60601 series), and the functional requirements for a cold compression system. For instance, the ground truth for electrical safety is whether the device actually meets the voltage leakage limits of IEC 60601-1. For the cold pack temperature, it's whether the internal temperature sensor actually registers within the specified range after 30 minutes of operation. This is determined by instrumental measurements and adherence to test protocols, not clinical outcomes or expert consensus on medical data.
8. The sample size for the training set
The device is a physical product with embedded software. It is not an AI/ML device that requires a "training set" of data in the common sense (e.g., images for deep learning). The software is likely programmed deterministically or based on engineered control logic, not "trained" on data. Therefore, the concept of a "training set sample size" is not applicable.
9. How the ground truth for the training set was established
Since there is no "training set" in the AI/ML sense, this question is not applicable. The "ground truth" for the device's design and software functionality is rooted in engineering requirements, safety standards, and the intended physiological effects of cold compression therapy (which are well-established prior to this device's development).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 19, 2024
Chengdu Cryo-Push Medical Technology Co.,Ltd % Liz Li Counselor Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District Shenzhen, Guangdong 518000 China
Re: K230524
Trade/Device Name: Cold Compression Wrap Pro Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP, IME Dated: October 31, 2023 Received: November 20, 2023
Dear Liz Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K230524
Device Name Cold Compression Wrap Pro
Indications for Use (Describe)
The Cold Compression Wrap Pro is indicated for the temporary relief of minor muscle aches and pains.
The device is indicated for temporary increase in created areas in people who are in good health, and simulates kneading and stroking of tissues using an inflatable wrap.
The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary K230524
1. Contact Details
1.1 Applicant information
| 510(k) owner's name | Chengdu Cryo-Push Medical Technology Co.,Ltd |
|---|---|
| Address | 102, 105, Zone 20, Huayin Industrial Port, No.618, |
| Kexing Road (West), Wenjiang District, Chengdu 611137 | |
| Sichuan P.R.China | |
| Phone No. | TEL: +86 18086852687 |
| Fax No. | / |
| Name of contact | Zhang Peiyong |
| person | |
| Date Prepared | Nov.18th.2023 |
1.2 Submission Correspondent
| 卓远天成 | Shenzhen Joyantech Consulting Co., Ltd. |
|---|---|
| 1713A, 17th Floor, Block A, Zhongguan Times Square, | |
| Liuxian Avenue, Xili Town, Nanshan District, Shenzhen, | |
| Guangdong Province, China | |
| Phone No. | +86 755-86069197 |
| Contact person | Liz Li |
| liz@cefda.com; grace@cefda.com | |
| Website | http://www.cefda.com |
2. Device information
| Trade name | Cold Compression Wrap Pro |
|---|---|
| Classification | II |
| Classification name | Massager, Powered Inflatable Tube |
| Product code | IRP; IME |
| Regulation No. | 21 CFR 890.5650 |
3. Legally Marketed Predicate Device
| Trade Name | Cryopush Cold Compression Device |
|---|---|
| Classification | II |
| 510(k) Number | K222669 |
| Product Code | IRP; IME |
{5}------------------------------------------------
| Manufacturer | Chengdu Cryo-Push Medical Technology Co., Ltd |
|---|---|
| Trade Name | Portable Therapeutix Squid Active Cold Compressiondevice and Cold Pack for OTC Use |
| 510(k) Number | K133483 |
| Product Code | IRP, IME |
| Manufacturer | MEDIcept, Inc. |
4. Device Description
Cold Compression Wrap Pro Device consists of a main unit and wraps which simulates kneading and stroking of tissue with the hands by use of inflatable pressure wraps. It is for temporary increase in circulation of the treated areas and temporary relief of minor muscle aches and pains. By inflating the air chambers and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient.
There are 4 different wraps for Shoulder, elbow/arm, low limbs (Calf and upper leg), ankle/foot. The wrap contains cold pack, and the cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
When the device is used for 30 minutes (Default time) at the operature (5°C~ 40°C) specified in the user manual, the temperature of the cold pack is (-18°C ~-4℃) (±2℃).
5. Intended use
The Cold Compression Wrap Pro is indicated for the temporary relief of minor muscle aches and pains.
The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneading and stroking of tissues using an inflatable wrap.
The cold wrap is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
{6}------------------------------------------------
6. Substantial Equivalence Comparison
| Item | Subject device | Predicate device K222669 | Predicate device K133483 | Comments |
|---|---|---|---|---|
| Device name | Cold Compression Wrap Pro | Cryopush Cold CompressionDevice | Portable Therapeutix SquidActive Cold Compressiondevice and Cold Pack forOTC Use | / |
| model | A02-C-006 | A02-P-001 | not publicly available | / |
| Regulation number | 21 CFR 890.5650,21 CFR 890.5700 | 21 CFR 890.5650,21 CFR 890.5700 | 21 CFR 890.5650,21 CFR 890.5700 | Identical |
| Classification | II | II | II | Identical |
| Product Code | IRP, IME | IRP, IME | IRP, IME | Identical |
| Intended use | The Cold Compression WrapPro is indicated for thetemporary relief of minormuscle aches and pains.The device is indicated fortemporary increase incirculation of the treatedareas in people who are ingood health, and simulateskneading and stroking oftissues using an inflatablewrap.The cold pack is indicated forlocalized therapy in situationswhere cold temperaturetherapy is necessary ordesirable. | The Cryopush ColdCompression Device isindicated for the temporaryrelief of minor muscle achesand pains.The device is indicated fortemporary increase incirculation of the treatedareas in people who are ingood health, and simulateskneading and stroking oftissues using an inflatablewrap.The cold pack is indicated forlocalized therapy in situationswhere cold temperaturetherapy is necessary ordesirable. | The Squid Active ColdCompression device and ColdPack for OTC use is indicatedfor the temporary relief ofminor muscle aches andpains.The compression device isindicated for temporaryincrease in circulation of thetreated areas in people whoare in good health, andsimulates kneading andstroking of tissues using aninflatable garment.The cold pack is indicated forlocalized therapy in situationswhere cold temperaturetherapy is necessary ordesirable. | Identical |
| Item | Subject device | Predicate device K222669 | Predicate device K133483 | Comments |
| Treatment area/Structure ofSleeves | Shoulder, elbow/arm, lowlimbs (Calf and upper leg),ankle/foot | Low limbs (Calf and upperleg) | Leg, foot, arm, shoulders,lower back, hands | Note |
| OTC or Rx | OTC | OTC | OTC | Identical |
| Environment ofUse | Clinics, hospital, athletetraining environments | Clinics, hospital, athletetraining, and homeenvironments | not publicly available | Note |
| Power source | 100-240V~50/60Hz | 100-240V~50/60Hz | 120V 60 Hz, consumption26W; AC adapter: 120V 60Hz, consumption 36W Lithiumbattery | Identical toK222669 |
| Principle ofoperation/Mechanicalcharacteristics | Intermittent compression 5intensity settings1-20 mmHg2-40 mmHg3-60 mmHg4-80 mmHg5-100 mmHg | Intermittent compression 5intensity settings1-20 mmHg2-40 mmHg3-60 mmHg4-80 mmHg5-100 mmHg | Intermittent compression 4intensity settings1-30 mmHg2-50 mmHg3-70 mmHg4-85 mmHg | Identical toK222669 &Note |
| Working Time | 10min, 20 min, 30 min, 40min, 50 min, 60 min, 70 min,80 min, 90 min, 100 min, 110min, 120 min, default as30min | 10min, 20 min, 30 min, 40min, 50 min, 60 min, 70 min,80 min, 90 min, 100 min, 110min, 120 min, default as30min | 15min | Identical toK222669 |
| Pressure range | 0-100mmHg | 0-100mmHg | 0-85mmHg | Identical toK222669 |
| Pressure levels | 20mmHg,40 mmHg,60mmHg,80 mmHg,100 mmHg. | 20mmHg,40 mmHg,60mmHg,80 mmHg,100 mmHg. | 30mmHg,40 mmHg,70mmHg,85 mmHg | Identical toK222669 |
| Pressure errorrange | ±15mmHg | ±15mmHg | not publicly available | Identical toK222669 |
| Keep time | 5s | 10s | not publicly available | Substantial |
| Deflation time | 6s | 20s | not publicly available | Equivalence |
| Item | Subject device | Predicate device K222669 | Predicate device K133483 | Comments |
| Noise level | ≤ 55dB | ≤ 55dB | not publicly available | Identical toK222669 |
| Wrap Material | Nylon with a PVC laminate | Nylon with a PVC laminate | Nylon with TPU backingwrap-yes | Identical toK222669 |
| Patient contact | Non-conductive attachments | Non-conductive attachments | not publicly available | Identical toK222669 |
| Single Wrapweight | low limbs (Calf and upperleg):480g;Ankle/foot: 338g;Shoulder: 388g;elbow/arm: 308g; | 480g | not publicly available | |
| Wrap Size | low limbs (Calf and upperleg): 330610mm;Ankle/foot: 830420mm;Shoulder: 1080430mm;elbow/arm: 620440mm; | 330*610mm | not publicly available | SubstantialEquivalence |
| Cold pack weight | low limbs (Calf and upperleg):1030g;Ankle/foot: 1282g;Shoulder: 932g;elbow/arm: 632g; | 1030g (±10%) | not publicly available | |
| Cold pack Size | low limbs (Calf and upperleg):570300mm;Ankle/foot: 595260mm;Shoulder: 520360mm;elbow/arm: 440370mm; | 300*570mm | not publicly available | |
| Cold packOperatingtemperature | -18℃~-4°C (±2°C) | -18℃~-4℃ (±2℃) | not publicly available | Identical toK222669 |
| Operatingenvironment | Temperature:5℃~ 40℃ (41°F ~ 104°F)Relative humidity: | Temperature:5℃~ 40℃(41°F ~ 104°F)Relative humidity: | not publicly available | Identical toK222669 |
| Item | Subject device | Predicate device K222669 | Predicate device K133483 | Comments |
| Transportation &Storageenvironment | 10% ~ 90%Atmospheric pressure:700~1060hpa | 10% ~ 90%Atmospheric pressure:700~1060hpa | ||
| Temperature: | Temperature: | |||
| -25℃ | -25℃ | |||
| Relative humidity:10% ~ 90%Atmospheric pressure:700~1060hpa | Relative humidity:10% ~ 90%Atmospheric pressure:700~1060hpa | not publicly available | Identical toK222669 |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Product: Cold Compression Wrap Pro
Chengdu Cryo-Push Medical Technology Co., Ltd
{10}------------------------------------------------
The subject device, Cold Compression Wrap Pro, is substantially equivalent to the predicate device K222669 and K133483. This conclusion is based upon comparison on intended use, principle of operation and technological characteristics, etc.
7. Summary of Non-clinical Performance Testing
The following data were provided in support of the substantial equivalence determination:
- Electrical Safety, Electromagnetic Compatibility 1)
- -IEC 60601-1:2005, AMDI:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- -IEC 60601-1-2:2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests
-
- Software validation
The subject device contains an embedded software system. Software verification and validation testing were conducted as recommended by FDA's Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", and the software for this device was considered as a "moderate" level of concern.
-
- Performance
There are no FDA recognized consensus standards for this device. We tested the following items according to our internal standards,
- Performance
-
-Product Appearance and Size
-
-Wrap Performance
-
-Function Test
-
- Biocompatibility test
The subject device and the predicated device use the same patient contact material, and has the same nature of body contact and contact duration. Therefore, no biocompatibility test was conducted.
8. Clinical testing
N/A
9. Conclusions
The subject device, Cold Compression Wrap Pro, is substantially equivalent to the predicate device K222669 and K133483. This conclusion is based upon comparison on intended use, principle of operation and technological characteristics, etc.
§ 890.5650 Powered inflatable tube massager.
(a)
Identification. A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.(b)
Classification. Class II (performance standards).