Search Results
Found 78 results
510(k) Data Aggregation
K Number
K242801Device Name
MICOR 700 System (N/A); MICOR 700 drive (FG-50631); MICOR 700 extractor (FG-50621); MICOR 700 vitrector (FG-51185)
Manufacturer
Carl Zeiss Meditec Cataract Technology Inc.
Date Cleared
2025-06-11
(267 days)
Product Code
HQC
Regulation Number
886.4670Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec Cataract Technology Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The intended use of the MICOR® 700 Lens Removal System is fragmentation and removal of cataracts, as well as associated procedures such as irrigation and aspiration, as well as ancillary functions such as vitreous aspiration and cutting.
Device Description
MICOR 700 System (N/A); MICOR 700 drive (FG-50631); MICOR 700 extractor (FG-50621); MICOR 700 vitrector (FG-51185)
Ask a Question
K Number
K243878Device Name
CLARUS (700)
Manufacturer
Carl Zeiss Meditec, Inc.
Date Cleared
2025-04-17
(120 days)
Product Code
QER
Regulation Number
886.1120Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The CLARUS 700 ophthalmic camera is indicated to capture, display, annotate and store images to aid in the diagnosis and monitoring of diseases and disorders occurring in the retina, ocular surface and visible adnexa. It provides true color and autofluorescence imaging modes for stereo, widefield, ultra-widefield, and montage fields of view.
The CLARUS 700 angiography is indicated as an aid in the visualization of vascular structures of the retina and the choroid.
Device Description
The CLARUS 700 is an active, software controlled, high resolution ophthalmic imaging device for In-vivo imaging of the human eye. Imaging modes include True color, Fundus Auto-fluorescence with green excitation, Fundus Auto-fluorescence with blue excitation, Fluorescein Angiography, Stereo External eye and Fluorescein Angiography- Indocyanine green angiography (FA-ICGA). All true color images can be separated into red, green and blue channel images to help enhance visual contrast of details in certain layers of the retina.
The CLARUS 700 angiography imaging aids in the visualization of the vascular structures of the retina and the choroid. With a single capture, CLARUS 700 produces a 90º high definition widefield image. Widefield images are automatically merged to achieve a 135º ultra-widefield of view. The CLARUS 700 makes use of a deep learning algorithm for Optic Nerve Head (ONH) detection. The ultra-widefield montage on CLARUS 700 is no longer dependent just on the patient accurately fixating their gaze on the internal fixation. With the ONH detection, the software will find the optic nerve and determine based on the image(s) captured where the patient was gazing at the point of capture. The CLARUS 700 device allows clinicians to easily review and compare high-quality images captured during a single exam while providing annotation and caliper measurement tools that allow in-depth analysis of eye health. CLARUS 700 is designed to optimize each patient's experience by providing a simple head and chin rest that allows the patient to maintain a stable, neutral position while the operator brings the optics to the patient, facilitating a more comfortable imaging experience. The ability to swivel the device between the right and left eye helps technicians capture an image without realigning the patient. Live IR Preview allows the technician to confirm image quality and screen for lid and lash obstructions, prior to imaging, ensuring fewer image recaptures.
The CLARUS 700 device's principle of operation is Slit Scanning Ophthalmic Camera also referred to as Broad Line Fundus Imaging (BLFI). During image capture, a line of illumination passes through the slit and scans across the retina. A 2D monochromatic camera captures the returned light to image the retina. A single sweep of the illumination is used to illuminate the retina for image capture. Repeated sweeps of near infrared light are used for a live retina view for alignment. Red, green and blue LEDs sequentially illuminate to generate true color images. Blue and green LED illumination enables Fundus Autofluorescence (FAF) imaging. Fluorescein Angiography images are captured with green LED illumination at a wavelength that stimulates fluorescence of the injected sodium fluorescein dye. The principle of operation of CLARUS 700 has not changed since the previous clearance, K191194.
The CLARUS 700 system is mainly comprised of an acquisition device, all-in-one PC, keyboard, mouse, instrument lift table and external power supply.
The device hardware is based on the predicate CLARUS 700 (K191194) hardware. The new ICGA imaging mode on the device required the following hardware changes as stated in the summary above:
- Lightbox for Infrared (IR) Laser
- Modified Slit filter – FA/ICG Slit Excitation Filter – new coating, no change to FA
- Modified Turret Filter 1- FA/ICG Dual Band barrier filter – new coating, no change to FA.
- Added Turret Filter 2 – Added second filter. Same coating as Turret Filter 1 to eliminate cornea reflex band in ICG images with a different shape.
- Added Large Alignment Tool (LAT)
- Added ICG Power Meter Tool
The CLARUS software provides the user the capability to align, capture, review and annotate images. The software has two installation configurations: Software installed on the Instrument (Acquisition & Review) as well as Software installed on a separate 'Review Station' (Laptop or Computer) (only Review).
The DEVICE software version 1.2 is based on the predicate CLARUS 700 software version 1.0 (K191194).
Added capability for DEVICE software version 1.2 include:
- **Simultaneous capture of Fluorescein Angiography (FA) + Indocyanine Green Angiography (ICGA)**
- **Angiography Movie:** Capture of multiple pictures in sequence, after a single press of a button. Available for FA, ICGA and Simultaneous FA+ICGA.
- **Early Treatment Diabetic Retinopathy Study (ETDRS)** – Manual placement of ETDRS grids (7 field ETDRS and Macula ETDRS) over the pictures:
- The ETDRS 7-fields grid in CLARUS is a display of the standard 7-fields in Color Fundus Photography used to determine an ETDRS (Early Treatment Diabetic Retinopathy Study) level for patients with Diabetic Retinopathy. These 7-fields in and around the macular region are displayed in one single widefield image according to definitions followed by the gold-standard 7-field images using narrow-field fundus cameras.
- The Macular ETDRS grids display assists in the identification of an ETDRS level in nine subfields centered around the fovea.
- **ICGA Boost Mode:** user-selectable option for ICGA capture that increase used light to obtain better picture at later phase.
- **8 up view:** addition of the possibility to view eight pictures side by side (currently it is only possible to see 1, 2, 4, 16)
The CLARUS 700 device meets the requirements of ISO 10940:2009 standard. The device technical specifications are identical to the predicate device.
Ask a Question
K Number
K241174Device Name
INTRABEAM (700); And accessories (INTRABEAM SMART Stand, INTRABEAM SMART Spherical Applicator, INTRABEAM Spherical Sizer Set, INTRABEAM Needle Applicator)
Manufacturer
Carl Zeiss Meditec AG
Date Cleared
2025-01-10
(259 days)
Product Code
JAD
Regulation Number
892.5900Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec AG
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The INTRABEAM is intended for use in radiotherapy treatments.
The INTRABEAM SMART Spherical Applicator is used with the INTRABEAM to deliver a prescribed dose of intraoperative radiation to the treatment margin or tumor bed during intracavity and intraoperative radiotherapy treatments.
The safety and effectiveness of the INTRABEAM as a replacement for whole breast irradiation in the treast cancer has not been established.
The INTRABEM Needle Applicator (comprising the Needle Applicator and guide shafts) is intended for use in combination with the INTRABEAM to intraoperatively administer radiation to tissue including irradiation of intracranial tumors.
The INTRABEAM SMART Stand is designed as an instrument support and positioning unit for the INTRABEAM.
The INTRABEAM Spherical Sizer Set shall support the doctor (surgeon and/or radiation oncologist) in assessing which spherical-shaped applicator shall be used for the radiation therapy procedure, involving INTRABEAM.
Device Description
The INTRABEAM 700 is a radiation therapy device intended for targeted treatments of selected lessions for minimally invasive, intraoperative, interstital, intracavity and contact radiation therapy of tumors or tumor beds within the body of cancer patients. By applying the radiation source in conjunction with various applicators, a prescribed dose of low energy radiation can be delivered to the target volume. The delivery of the radiation dose is controlled via the integrated control unit and software.
The INTRABEAM 700 is provided as a mobile workstation. Like the previously cleared versions of the INTRABEAM system, the INTRABEAM 700 provides several tools for Quality Assurance of radiation delivery, which are intended to verify the proper functioning of the radiotherapy treatment system.
The main components of the INTRABEAM 700 system are:
- INTRABEAM Workplace - mobile cart containing the following:
- Control Console 700 (CC700)
- Computer with Software Version 5.0
- Touchscreen monitor and mouse
- UNIDOS Romeo Electrometer
- V-guide
- XRS 4 X-ray Source
- Quality Assurance Tools: PAICH, PDA, and Ionization Chamber with Ionization Chamber Holder
- radiance Third party treatment planning simulation software
The INTRABEAM SMART Stand is connected to the INTRABEAM 700 and is used to mount the X-ray generator (XRS 4) and the necessary applicator, in order to deliver the prescribed radiation dose at the target site.
The INTRABEAM Spherical Applicator is a sterile disposable product that shall be placed in contact with the tumor mass and/or tumor resection cavity to deliver a prescribed dose of intraoperative radiation.
The INTRABEAM Spherical Sizer Set is a sterile disposable product that shall be placed in contact with body part and/or tumor mass to help support the doctor (surgeon and/or radiation oncologist) in assessing which spherical-shaped applicator shall be used for the radiation therapy procedure, involving INTRABEAM.
The INTRABEAM Needle Applicator has not been updated since the last clearance, K162568.
Ask a Question
K Number
K233911Device Name
VISULAS combi
Manufacturer
Carl Zeiss Meditec AG
Date Cleared
2024-09-06
(269 days)
Product Code
HQF
Regulation Number
886.4390Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec AG
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
VISULAS combi is intended for use in photocoagulating and photodisrupting ocular tissues in the treatment of diseases of the eye, including:
• Photocoagulation of the retina
- Trabeculoplasty
- · Iridotomy
- · Posterior capsulotomy
· Posterior membranectomy
This device is Prescription Use (Rx) only.
Device Description
VISULAS combi is an ophthalmic laser used for standard photocoagulation treatments of ocular tissue with 532 nm wavelength and standard photodisruption treatments of ocular tissues at a wavelength of 1064 mm. VISULAS combi is operated in the following treatment modes:
- single-spot mode (software license VERTE) -
- multi-spot mode (software license VITE)
- YAG disruption mode (software license YAG). -
VISULAS combi consists of a laser console, touch control panel, optional laser light applicators such as laser slit lamp or laser indirect ophthalmoscope, foot switch and instrument table. The device can also be used with various accessories, such as SL Imaging Solution, contact lenses or Applanation Tonometer.
Ask a Question
K Number
K240215Device Name
BLUE 400; BLUE 400 S
Manufacturer
Carl Zeiss Meditec AG
Date Cleared
2024-06-28
(154 days)
Product Code
QFX
Regulation Number
882.4950Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec AG
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
BLUE 400 and BLUE 400 S are accessories to the surgical microscope and allow the fluorescence observation of fluorophores with an excitation peak between 400 nm and the fluorescence emission observation comprising the spectrum in a spectral band of 620 - 710 nm.
The ZEISS BLUE 400 and BLUE 400 S are surgical microscope accessories used in fluorescent visualization of suspected grade III and IV gliomas during neurosurgery.
Device Description
The BLUE 400 and BLUE 400 S are fluorescence accessories to qualified surgical microscopes, intended to allow intraoperative viewing of malignant glioma tissue grade III and IV under fluorescence. The overall system is comprised of excitation (illumination) and emission (observation) filters to detect fluorescence and are optimized in conjunction with the drug to pass light between 620 – 710 nanometers. The BLUE 400 S filters allow the surgical microscope to produce excitation light in a wavelength range covering at least 400 - 410 nanometers that excites an approved optical imaging agent and enables the surgeon to observe the emitted fluorescent signal in the oculars or on a display. Fluorescence of marked brain tissue helps visualization of tissue associated with Grade III & IV glioma during neurosurgeries.
Compared to the blue visualization of the surrounding non-fluorescent tissue in the BLUE 400 image, BLUE 400 S is designed to visualize the surrounding nonfluorescent tissue more similar to white light impression, while tumor visualization of grade III and IV glioma remains consistent. With the visualization of non-fluorescent anatomy in an almost white light impression, BLUE 400 S is expected to allow PplX visualization with less frequent switching between fluorescence and white light imaging modes.
BLUE 400 and BLUE 400 S can be installed only into qualified ZEISS surgical microscopes. For these accessories to be used with a qualified ZEISS surgical microscope, the critical components of the surgical microscope need to fulfill the clinically relevant parameters for the Indications for Use of BLUE 400 and BLUE 400 S.
Ask a Question
K Number
K233933Device Name
CIRRUS HD-OCT Model 6000
Manufacturer
Carl Zeiss Meditec Inc
Date Cleared
2024-05-17
(155 days)
Product Code
OBO
Regulation Number
886.1570Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device. It is indicated for in-vivo viewing, axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures, including cornea, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head.
The CIRRUS™ HD-OCT Reference Database is a quantitative tool used for the comparison of retinal nerve fiber layer thickness, macular thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of healthy subjects.
CIRRUS™ HD-OCT AngioPlex angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid.
The CIRRUS™ HD-OCT is indicated for use as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.
Device Description
The subject device is a computerized instrument that acquires and analyses cross-sectional tomograms of anterior ocular structures (including comea, retinal nerve fiber layer, macula, and optic disc). It employs non-invasive, non-contact, low-coherence interferometry to obtain these high-resolution images. CIRRUS 6000 has a 100kHz scan rate for all structural and angiography scans.
The subject device uses the same optical system, and principle of operation as the previously cleared CIRRUS 6000 (K222200) except for the reference database functionality.
The subject device contains a newly acquired reference database which was collected on K222200. This study data compares macular thickness, ganglion cell thickness, optic disc and RNFL measurements to a reference range of healthy eyes as guided by the age of the patient and /or optic disc size. Reference database outputs are available on Macular Cube 200x200, and Optic Disc Cube 200x20 scan patterns. All other technical specifications have remained the same as the predicate K222200.
Ask a Question
K Number
K233421Device Name
RESCAN 700
Manufacturer
Carl Zeiss Meditec AG
Date Cleared
2024-03-08
(150 days)
Product Code
OBO
Regulation Number
886.1570Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec AG
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
RESCAN 700 provides non-contact, high resolution optical coherence tomographic (OCT) and biomicroscopic imaging of the anterior and posterior segment of the eve via an ophthalmic surgical microscope. The RESCAN 700 is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging of posterior ocular structures, including retina, macula, and optic disc, as well as imaging of anterior ocular structures, including the cornea, lens and anterior chamber angle.
RESCAN 700 uses the assistance system (CALLISTO eye) that provides non-diagnostic video documentation and image capture for ophthalmic surgeries. The assistance system allows the remote control of RESCAN 700.
Device Description
RESCAN 700 brings Spectral Domain OCT technology to the ZEISS ophthalmic surgical microscopes (e.g. ARTEVO 800). Used in conjunction with the assistance system, CALLISTO eye, OCT images taken intra-operatively are presented on the monitor and may also be seen within the surgeon's oculars using the surgical microscopes integrated data injection system (IDIS). OCT images may be stored for subsequent retrieval using CALLISTO eye's data management system. RESCAN 700 can be controlled via the touch panel of the assistance system or via the foot control panel of an ophthalmic surgical microscope
Ask a Question
K Number
K232944Device Name
CALLISTO eye
Manufacturer
Carl Zeiss Meditec AG
Date Cleared
2023-12-21
(92 days)
Product Code
NFJ
Regulation Number
892.2050Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec AG
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
CALLISTO eye Software is a software device intended for remote control of ophthalmic surgical microscopes of ARTEVO 750/850 and RESCAN 700, and display images of the anterior and posterior segment of the eye.
CALLISTO eye Software is indicated as graphical guidance aid to insert, align, position, and register an intraocular lens (IOL) including toric IOLs, limbal relaxing incisions, and capsulorhexis during anterior segment surgical procedures.
Device Description
CALLISTO eye software version 5.0 is a new release sporting a new user interface but carries the clinical feature set of software version 3.7.2: it supports the digital visualization technology and connectivity of ARTEVO 750 / ARTEVO 850 and provides connectivity to the QUATERA700. CALLISTO eye enables the video visualization of the anterior segments of the eye and allows the connection and remote control of a surgical microscope with and without OCT Camera. It is designed for high patient throughput and can be used for teaching purposes.
CALLISTO eye is an assistance system that processes real-time video images that can be displayed on the CALLISTO eye Panel PC for viewing by the surgeon and the surgical staff in the operating room. The same video images can be viewed by the surgeon through the eyepiece of the connected surgical microscope. CALLISTO eye provides Assistant Functions displaying treatment templates as screen overlays and Cockpits displaying patient and device information as screen overlays. Both functions assist the surgeon during procedures such as limbal relaxing incisions, capsulorhexis, and alignment of toric intraocular lenses (TIOL). All treatment templates are based on preoperative clinical data of a particular patient and shall be defined by the surgeon prior surgery. These templates can be displayed on the CALLISTO eye Panel PC, through the eyepiece of the surgical microscope equipped with a data injection system (IDIS (WITH VERSION 5.0 RELABELED AS ADVISION)) of the ARTEVO 750 or on a 3D monitor connected to the ARTEVO 850. While using "ASSISTANCE markerless" configuration, CALLISTO eye can utilize the preoperative diagnostic data from the Zeiss IOLMaster and may provide the reference and target axis as required to align a toric intraocular lens without the otherwise required ink marks.
Transmission of the diagnostic data from the IOLMaster to CALLISTO eye takes place via USB stick or via a data network connected to a DICOM compatible MIMPS server such as FORUM. The DICOM functionality allows the indirect communication with other DICOM compatible diagnostic devices and patient information systems to exchange patient data (e.g. medical devices work lists).
Ask a Question
K Number
K232051Device Name
VISULAS green
Manufacturer
Carl Zeiss Meditec
Date Cleared
2023-10-24
(106 days)
Product Code
HQF
Regulation Number
886.4390Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
VISULAS green is intended for use in photocoagulating ocular tissues of the treatment of diseases of the eve, including
- · Photocoagulation of the retina
- Trabeculoplasty
- · Iridotomy
Device Description
VISULAS green is an ophthalmic laser used for standard photocoagulation treatments of ocular tissue with 532 nm wavelength. VISULAS green is operated in single-spot mode or in multi-spot mode.
VISULAS green consists of a laser console, touch control panel, optional laser light applicators such as laser slit lamp or laser indirect ophthalmoscope, foot switch and instrument table. The device can also be used with various accessories, such as SL Imaging Solution, contact lenses or Applanation Tonometer.
Ask a Question
K Number
K230350Device Name
VISULAS yag
Manufacturer
Carl Zeiss Meditec Inc
Date Cleared
2023-09-22
(225 days)
Product Code
HQF
Regulation Number
886.4390Why did this record match?
Applicant Name (Manufacturer) :
Carl Zeiss Meditec Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
VISULAS yag is intended for use in photodisrupting ocular tissue in the treatment of diseases of the eye, including Posterior capsulotomy, Iridotomy and Posterior Membranectomy.
This device is for Prescription Use (Rx) only.
Device Description
VISULAS yag uses a Q-switched, flashlamp-pumped solid-state laser for photodisruption treatments of diseases of the eye, including posterior capsulotomy, iridotomy and membranectomy. Laser radiation is generated by means of a neodymium-doped yttrium aluminum garnet (Nd:YAG) gain medium inside the laser source. The emitted laser radiation with a near-infrared wavelength of = 1064 nm has a pulse duration of < 4 ns (full-width half-maximum; FWHM) and a focal diameter of 6.5 um ± 20%. The maximum energy output per pulse is 9 to 13 mJ in single-burst mode.
Ask a Question
Page 1 of 8