K Number
K233933
Device Name
CIRRUS™ HD-OCT Model 6000
Date Cleared
2024-05-17

(155 days)

Product Code
Regulation Number
886.1570
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device. It is indicated for in-vivo viewing, axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures, including cornea, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head. The CIRRUS™ HD-OCT Reference Database is a quantitative tool used for the comparison of retinal nerve fiber layer thickness, macular thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of healthy subjects. CIRRUS™ HD-OCT AngioPlex angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid. The CIRRUS™ HD-OCT is indicated for use as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.
Device Description
The subject device is a computerized instrument that acquires and analyses cross-sectional tomograms of anterior ocular structures (including comea, retinal nerve fiber layer, macula, and optic disc). It employs non-invasive, non-contact, low-coherence interferometry to obtain these high-resolution images. CIRRUS 6000 has a 100kHz scan rate for all structural and angiography scans. The subject device uses the same optical system, and principle of operation as the previously cleared CIRRUS 6000 (K222200) except for the reference database functionality. The subject device contains a newly acquired reference database which was collected on K222200. This study data compares macular thickness, ganglion cell thickness, optic disc and RNFL measurements to a reference range of healthy eyes as guided by the age of the patient and /or optic disc size. Reference database outputs are available on Macular Cube 200x200, and Optic Disc Cube 200x20 scan patterns. All other technical specifications have remained the same as the predicate K222200.
More Information

CIRRUS HD-OCT Model 500, 5000 (K181534)

No
The document describes a reference database and statistical analysis (regression) for comparison, not AI/ML algorithms for image analysis or diagnosis.

No

The device is described as an imaging and diagnostic device to aid in the detection and management of ocular diseases, not for therapeutic purposes.

Yes
The "Intended Use / Indications for Use" section explicitly states: "The CIRRUS™ HD-OCT is indicated for use as a diagnostic device to aid in the detection and management of ocular diseases...".

No

The device description explicitly states it is a "computerized instrument that acquires and analyses cross-sectional tomograms" and employs "non-invasive, non-contact, low-coherence interferometry to obtain these high-resolution images." This indicates the device includes hardware components for image acquisition, not just software.

Based on the provided text, this device is not an In Vitro Diagnostic (IVD).

Here's why:

  • IVDs are used to examine specimens derived from the human body. The text explicitly states the CIRRUS™ HD-OCT is a "non-contact, high resolution tomographic and biomicroscopic imaging device" indicated for "in-vivo viewing" and imaging of ocular structures. This means it is used directly on a living patient, not on a sample taken from a patient.
  • The device description focuses on imaging technology and analysis of in-vivo images. It describes the use of low-coherence interferometry to obtain images of ocular structures.
  • The intended use and indications for use describe imaging and measurement of ocular structures within the eye. This is consistent with an in-vivo imaging device.
  • The reference database is used for comparison of measurements taken from the in-vivo images to a database of healthy subjects. This is a diagnostic aid based on imaging data, not on the analysis of a biological specimen.

Therefore, the CIRRUS™ HD-OCT is an in-vivo imaging device, not an In Vitro Diagnostic.

N/A

Intended Use / Indications for Use

The CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device. It is indicated for in-vivo viewing, axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures, including cornea, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head.

The CIRRUS™ HD-OCT Reference Database is a quantitative tool used for the comparison of retinal nerve fiber layer thickness, macular thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of healthy subjects.

CIRRUS™ HD-OCT AngioPlex angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid.

The CIRRUS™ HD-OCT is indicated for use as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.

Product codes

OBO

Device Description

The subject device is a computerized instrument that acquires and analyses cross-sectional tomograms of anterior ocular structures (including cornea, retinal nerve fiber layer, macula, and optic disc). It employs non-invasive, non-contact, low-coherence interferometry to obtain these high-resolution images. CIRRUS 6000 has a 100kHz scan rate for all structural and angiography scans.

The subject device uses the same optical system, and principle of operation as the previously cleared CIRRUS 6000 (K222200) except for the reference database functionality.

The subject device contains a newly acquired reference database which was collected on K222200. This study data compares macular thickness, ganglion cell thickness, optic disc and RNFL measurements to a reference range of healthy eyes as guided by the age of the patient and /or optic disc size. Reference database outputs are available on Macular Cube 200x200, and Optic Disc Cube 200x20 scan patterns. All other technical specifications have remained the same as the predicate K222200.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Optical Coherence Tomography (OCT)

Anatomical Site

cornea, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, optic nerve head

Indicated Patient Age Range

18 years of age and older

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

A prospective, multi-site reference database (RDB) study was conducted at eight (8) clinical sites across the USA to develop the CIRRUS 6000 RDB for macular thickness and optic nerve head scan values.

The Subject Device's RDB was created to help clinicians assess and effectively compare a patient's measurements to that of healthy subjects, representative of the general population.

The eligibility criteria included subjects 18 years of age and older, best corrected visual acuity (BCVA) of 20/40 or better in either eye, IOP

§ 886.1570 Ophthalmoscope.

(a)
Identification. An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.(b)
Classification. Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

0

May 17, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Carl Zeiss Meditec Inc Tanesha Bland Senior Regulatory Affairs Specialist- US SSC 5300 Central Parkway Dublin, California 94568

Re: K233933

Trade/Device Name: CIRRUS™ HD-OCT Model 6000 Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO Dated: April 3, 2024 Received: April 8, 2024

Dear Tanesha Bland:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

1

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Elvin Y. Ng -S

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices

2

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

Submission Number (if known)

K233933

Device Name

CIRRUS™ HD-OCT Model 6000

Indications for Use (Describe)

The CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device. It is indicated for in-vivo viewing, axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures, including cornea, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head.

The CIRRUS™ HD-OCT Reference Database is a quantitative tool used for the comparison of retinal nerve fiber layer thickness, macular thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of healthy subjects.

CIRRUS™ HD-OCT AngioPlex angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid.

The CIRRUS™ HD-OCT is indicated for use as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

Image /page/4/Picture/0 description: The image features the word "ZEISS" in bold, white letters against a solid, dark blue background. The font is sans-serif and appears to be a simple, blocky design. The overall impression is clean and professional, likely representing the logo of the company Zeiss.

Page 1 of 28

In accordance with 21 CFR 807.92 the 510(k) Summary for the CIRRUS Model 6000 RDB is provided below.

SUBMITTER 1.

| Applicant: | Carl Zeiss Meditec Inc
5300 Central Parkway
Dublin, CA 94568
USA |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Contact Name | Vidita Desai
Senior Regulatory Affairs Specialist
Carl Zeiss Meditec, Inc.
5300 Central Parkway Dublin, CA 94568
Mobile:(925) 413-1644
E-mail: vidita.desai@zeiss.com (preferred) |
| Primary Correspondent | Tanesha Bland
Senior Regulatory Affairs Specialist USA
Carl Zeiss Meditec, Inc.
5300 Central Parkway Dublin, CA 94568
Mobile:(925) 216-7963
E-mail: Tanesha.bland@zeiss.com (preferred) |
| Secondary Correspondent | Maria Golovina, MS MBA
Head of Regulatory Affairs – USA
Carl Zeiss Meditec USA Inc.
5300 Central Parkway, Dublin CA USA 94568
Mobile: 925.216.1078
Email:Maria.Golovina@zeiss.com (preferred) |

Date Prepared:

May 3, 2024

5

Page 2 of 28

SUBJECT DEVICE 2.

Device Trade Name:CIRRUS HD-OCT Model 6000
Classification:21CFR886.1570 Optical Coherence Tomography (OCT)
Regulatory Class:II
Product Code:OBO

3. PREDICATE DEVICE

Predicate Device:CIRRUS HD-OCT Model 6000 (K222200)
Classification:21CFR886.1570 Optical Coherence Tomography (OCT)
Regulatory Class:II
Product Code:OBO
Reference Device:CIRRUS HD-OCT Model 500, 5000 (K181534)
Classification:21CFR886.1570 Optical Coherence Tomography (OCT)
Regulatory Class:II
Product Code:OBO

4. DEVICE DESCRIPTION

The subject device is a computerized instrument that acquires and analyses cross-sectional tomograms of anterior ocular structures (including comea, retinal nerve fiber layer, macula, and optic disc). It employs non-invasive, non-contact, low-coherence interferometry to obtain these high-resolution images. CIRRUS 6000 has a 100kHz scan rate for all structural and angiography scans.

The subject device uses the same optical system, and principle of operation as the previously cleared CIRRUS 6000 (K222200) except for the reference database functionality.

The subject device contains a newly acquired reference database which was collected on K222200. This study data compares macular thickness, ganglion cell thickness, optic disc and RNFL measurements to a reference range of healthy eyes as guided by the age of the patient and /or optic disc size. Reference database outputs are available on Macular Cube 200x200, and Optic Disc Cube 200x20 scan patterns. All other technical specifications have remained the same as the predicate K222200.

6

K233933 - 510(k) Summary INDICATIONS FOR USE 5.

Page 3 of 28

The CIRRUS™ HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device. It in-vivo viewing, axial cross-sectional, and three-diment of anterior and posterior ocular structures, including cornea, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, and optic nerve head.

The CIRRUS™ HD-OCT Reference Database is a quantitative tool used for the comparison of retimal nerve fiber layer thickness, macular thickness, ganglion cell plus inner plexiform layer thickness, and optic nerve head measurements to a database of healthy subjects.

CIRRUS™ HD-OCT AngioPlex angiography is indicated as an aid in the visualization of vascular structures of the retima and chorod.

The CIRRUS™ HD-OCT is indicated for use as a diagnostic device to aid in the detection and management of ocular diseases including, but not limited to, macular holes, cystoid macular edema, diabetic retinopathy, age-related macular degeneration, and glaucoma.

SUBSTANTIAL EQUIVALENCE TO PRIMARY PREDICATE 6.

Table 1. Subject to Predicate Device Comparison Table - Indications for Use

| Device | Subject Device -CIRRUS 60000
with RDB | Predicate Device – CIRRUS 6000 | Reference Device - CIRRUS
5000 (K181534) | Equivalency
Analysis |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Indications for
Use | CIRRUS™ HD-OCT is a non-contact,
high resolution tomographic and
biomicroscopic imaging device. It is
indicated for in-vivo viewing, axial
cross-sectional, and three-dimensional
imaging and measurement of anterior
and posterior ocular structures,
including cornea, retina, retinal nerve
fiber layer, ganglion cell plus inner
plexiform layer, macula, and optic
nerve head.
CIRRUS™ HD-OCT Reference
Database is a quantitative tool used
for the comparison of retinal nerve
fiber layer thickness, macular
thickness, ganglion cell plus inner
plexiform layer thickness, and optic
nerve head measurements to a
database of healthy subjects. | CIRRUS HD-OCT is a non-contact,
high resolution tomographic and
biomicroscopic imaging device
intended for in-vivo viewing, axial
cross-sectional, and three-dimensional
imaging of anterior and posterior ocular
structures. The device is indicated for
visualizing and measuring anterior and
posterior ocular structures, including
cornea, corneal epithelium, retina,
retinal nerve fiber layer, ganglion cell
plus inner plexiform layer, macula, and
optic nerve head.
CIRRUS AngioPlex OCT Angiography
with is indicated as an aid in the
visualization of vascular structures of
the retina and choroid. | CIRRUS HD-OCT is a non-
contact, high resolution
tomographic and biomicroscopic
imaging device intended for in-
vivo viewing, axial cross-
sectional, and three-dimensional
imaging of anterior and posterior
ocular structures. The device is
indicated for visualizing and
measuring anterior and posterior
ocular structures, including
cornea, corneal epithelium, retina,
retinal nerve fiber layer, ganglion
cell plus inner plexiform layer,
macula, and optic nerve head. The
CIRRUS normative databases are
quantitative tools indicated for the
comparison of retinal nerve fiber
layer thickness, macular thickness,
ganglion cell plus inner plexiform | Identical to the
reference device. |

7

K233933 - 510(k) SummaryPage 4 of 28
DeviceSubject Device -CIRRUS 60000
with RDBPredicate Device – CIRRUS 6000Reference Device – CIRRUS 5000 (K181534)Equivalency Analysis
CIRRUS™ HD-OCT AngioPlex
angiography is indicated as an aid in
the visualization of vascular structures
of the retina and choroid.CIRRUS HD-OCT is indicated as a
diagnostic device to aid in the detection
and management of ocular diseases
including, but not limited to, macular
holes, cystoid macular edema, diabetic
retinopathy, age-related macular
degeneration, and glaucoma.layer thickness, and optic nerve
head measurements to a database
of normal subjects.

CIRRUS AngioPlex OCT
Angiography with is indicated as
an aid in the visualization of
vascular structures of the retina
and choroid.

CIRRUS HD-OCT is indicated as
a diagnostic device to aid in the
detection and management of
ocular diseases including, but not
limited to, macular holes, cystoid
macular edema, diabetic
retinopathy, age-related macular
degeneration, and glaucoma. | | |
| | CIRRUS™ HD-OCT is indicated for
use as a diagnostic device to aid in the
detection and management of ocular
diseases including, but not limited to,
macular holes, cystoid macular
edema, diabetic retinopathy, age-
related macular degeneration, and
glaucoma. | | | | |

Table 2. Subject to Predicate Device Comparison Table – Technical Characteristics

| Device | CIRRUS™ HD-OCT 6000 v.
11.7
Subject Device | CIRRUS™ HD-OCT 6000
Predicate Device (K222200) | Reference Device –
CIRRUS 5000 (K181534) | Analysis |
|--------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-----------|
| Device
Classification Name | Tomography, Optical
Coherence | Tomography, Optical
Coherence | Tomography, Optical
Coherence | Identical |
| Generic/
Common Name | Optical Coherence Tomography
(OCT) | Optical Coherence Tomography
(OCT) | Optical Coherence
Tomography (OCT) | Identical |
| Classification
Product Code | | | | Identical |
| Class | II | II | II | Identical |
| Technology | Spectral Domain (Spatially
encoded Frequency Domain
and Fourier Domain Principle) | Spectral Domain (Spatially
encoded Frequency Domain
and Fourier Domain Principle) | Spectral Domain (Spatially
encoded Frequency Domain
and Fourier Domain
Principle) OCT | Identical |

8

K233933 - 510(k) Summary Page 5 of 28
OCTOCT
Illumination
Sources used in
InstrumentLight Emitting Diode 700 nm –
Iris Viewer
LSO Super Luminescent Diode
750 nm
OCT Super Luminescent Diode
840 nmLight Emitting Diode 700 nm –
Iris Viewer
LSO Super Luminescent Diode
750 nm
OCT Super Luminescent Diode
840 nmLight Emitting Diode
700 nm – Iris Viewer
LSO Super Luminescent
Diode 750 nm
OCT Super Luminescent
Diode 840 nmIdentical
Models600060005000Identical to the predicate
device.
OCT IMAGING
MethodologySpectral domain OCT (SD-
OCT)Spectral domain OCT (SD-
OCT)Spectral domain OCT (SD-
OCT)Identical
OCT Optical
SourceOCT Super Luminescent
Diode, 840 nm wavelengthOCT Super Luminescent Diode,
840 nm wavelengthOCT Super Luminescent
Diode, 840 nm
wavelengthIdentical
Optical Power1200 μW at the cornea +/- 300
μW (0.9 - 1.5 mW) at the
cornea.1200 μW at the cornea +/- 300
μW (0.9 - 1.5 mW) at the
cornea.only for OCT AngiographyIdentical to the predicate
device.
Axial Scan Depth
(Max)- Retina• 2.9 mm for 12 mm 1
Line 100x HD Raster,
HD Angio 8x8, Angio
8x8, and Angiography
12x12
• 2.0 mm for all other
scans• 2.9 mm for 12 mm 1
Line 100x HD Raster,
HD Angio 8x8, Angio
8x8, and Angiography
12x12
• 2.0 mm for all other
scans2.0 mmIdentical to the predicate
device.
Axial Scan Depth
(Max)- Retina2.0 mm (in tissue), 1024 pixels
per A-scans for all other scan
except for the ones listed above2.0 mm (in tissue), 1024 pixels
per A-scans for all other scan
except for the ones listed above2.0 mm (in tissue), 1024
pixels per A-scans.Identical
K233933 - 510(k) SummaryPage 6 of 28
Axial Scan Depth
(Max) –Anterior
Segment2.0 mm (Anterior 5-Line
Raster) – 1024 points2.0 mm
(Anterior Segment Cube) –
1024 points
5.8 mm (Anterior Chamber) –
2048 points
2.9 mm (Wide Angle-to-Angle) –
1024 points
2.9 mm (HD Angle) – 1024
points
2.0 mm (HD Cornea) – 1024
points
2.0 mm (Pachymetry) – 1024
points2.0 mm (Anterior 5-Line
Raster) – 1024 points2.0 mm
(Anterior Segment Cube) –
1024 points
5.8 mm (Anterior Chamber) –
2048 points
2.9 mm (Wide Angle-to-Angle) –
1024 points
2.9 mm (HD Angle) – 1024
points
2.0 mm (HD Cornea) – 1024
points
2.0 mm (Pachymetry) – 1024
points2.0 mm (Anterior 5-Line
Raster) – 1024 points2.0 mm
(Anterior Segment Cube) –
1024 points
5.8 mm (Anterior Chamber) –
2048 points
2.9 mm (Wide Angle-to-
Angle) – 1024 points
2.9 mm (HD Angle) – 1024
points
2.0 mm (HD Cornea) – 1024
points
2.0 mm (Pachymetry) –
1024 pointsIdentical
External Anterior
Segment LensAnterior Chamber Lens
Cornea LensAnterior Chamber Lens
Cornea LensAnterior Chamber Lens
Cornea LensIdentical
Transverse Scan
Range
(Lateral range in
degrees)
Retina10° x 0° on retina (Minimum)
42° x 42° on retina (Maximum)10° x 0° on retina (Minimum)
42° x 42° on retina (Maximum)10° x 0° on retina
(Minimum)
31° x 31° on retina
(Maximum)Identical to the predicate
device.
Transverse Scan
Range –Anterior
Segment3 mm (Minimum)
15 mm (Maximum)3 mm (Minimum)
15 mm (Maximum)3 mm (Minimum)
15 mm (Maximum)Identical
Axial Resolution5 μm (in tissue)5 μm (in tissue)5 μm (in tissue)Identical
Transverse
Resolution – Retina≤ 15 μm (in tissue)≤ 15 μm (in tissue)≤ 15 μm (in tissue)Identical
Transverse
Resolution –
Anterior Segment20 μm (in tissue)
20 μm @ 6mm FOV
25 μm @ 9mm FOV20 μm (in tissue)
20 μm @ 6mm FOV
25 μm @ 9mm FOV20 μm (in tissue)
20 μm @ 6mm FOV
25 μm @ 9mm FOVIdentical
K233933 - 510(k) Summary
Page 7 of 2845 μm @ 15mm FOV45 μm @ 15mm FOV25 μm @ 9mm FOV
45 μm @ 15mm FOV
Scan patterns
Non-angiographyLine, circle, crosshair, raster (a series of closely spaced lines, aka cube scan), radial scans and combinations of the above.Line, circle, crosshair, raster (a series of closely spaced lines, aka cube scan), radial scans and combinations of the above.Line, circle, crosshair, raster (a series of closely spaced lines, aka cube scan), radial scans and combinations of the above.Identical
Scan types
Non-angiographyMacular Cube 512 x 128 = 6 mm x 6 mmMacular Cube 512 x 128 = 6 mm x 6 mmMacular Cube 512 x 128 = 6 mm x 6 mmIdentical
Macular Cube 200 x 200 = 6 mm x 6 mmMacular Cube 200 x 200 = 6 mm x 6 mmMacular Cube 200 x 200 = 6 mm x 6 mmIdentical
Optic Disc Cube 200 x 200 = 6 mm x 6 mmOptic Disc Cube 200 x 200 = 6 mm x 6 mmOptic Disc Cube 200 x 200 = 6 mm x 6 mmIdentical
Not availableNot available5-Line RasterIdentical to the predicate device.
HD (high-definition) Raster
  • HD 1 Line 100x (2.9 mm depth and up to 12 mm in length)
  • HD 21 Line
  • HD Radial
  • HD Cross
  • HD 5-Line Raster | HD (high-definition) Raster
  • HD 1 Line 100x (2.9 mm depth and up to 12 mm in length)
  • HD 21 Line
  • HD Radial
  • HD Cross
  • HD 5-Line Raster | HD (high-definition) Raster
  • HD 1 Line 100x (2.0 mm depth and up to 9 mm in length)
  • HD 21 Line
  • HD Radial
  • HD Cross
  • HD 5-Line Raster | Identical to the predicate device. |
    | | | Anterior Segment Cube 512 x 128 | Anterior Segment Cube 512 x 128 | Anterior Segment Cube 512 x 128 | Identical |
    | K233933 - 510(k) Summary | | Page 8 of 28 | | | |
    | | Anterior Segment 5-Line Raster | Anterior Segment 5-Line Raster | Anterior Segment 5-Line Raster | Identical | |
    | | HD Angle | HD Angle | HD Angle | Identical | |
    | | Anterior Chamber | Anterior Chamber | Anterior Chamber | Identical | |
    | | Wide Angle-to-Angle | Wide Angle-to-Angle | Wide Angle-to-Angle | Identical | |
    | | HD Cornea | HD Cornea | HD Cornea | Identical | |
    | | Pachymetry | Pachymetry | Pachymetry | Identical | |
    | OCT Angiography
    scans | Scans: | Scans: | Scans: | Scans: | |
    | | AngioPlex 3x3 mm scan | AngioPlex 3x3 mm scan | AngioPlex 3x3 mm scan | Identical | |
    | | AngioPlex 6x6 mm scan | AngioPlex 6x6 mm scan | AngioPlex 6x6 mm scan | Identical | |
    | | AngioPlex 8x8 mm scan,
    (2.9mm scan depth) | AngioPlex 8x8 mm scan,
    (2.9mm scan depth) | AngioPlex 8x8 mm scan,
    (2.0mm scan depth) | Identical to the predicate
    device. | |
    | | AngioPlex 12x12 mm scan
    (2.9mm scan depth) | AngioPlex 12x12 mm scan
    (2.9mm scan depth) | Not Available | Identical to the predicate
    device. | |
    | | AngioPlex HD 6x6 mm scan | AngioPlex HD 6x6 mm scan | Not Available | Identical to the predicate
    device. | |
    | | AngioPlex HD 8x8 mm scan
    (2.9mm scan depth) | AngioPlex HD 8x8 mm scan
    (2.9mm scan depth) | Not Available | Identical to the predicate
    device. | |
    | | 6 x 6 mm Montage AngioPlex | 6 x 6 mm Montage AngioPlex | Not Available | Identical to the predicate
    device. | |
    | | 8 x 8 mm Montage AngioPlex | 8 x 8 mm Montage AngioPlex | Not Available | Identical to the predicate
    device. | |
    | K233933 - 510(k) Summary | | Page 9 of 28 | | | |
    | Retina Tracking | FastTrac Retinal Tracking
    Track-to-prior to Scan
    Acquisition tracking. | FastTrac Retinal Tracking
    Track-to-prior to Scan
    Acquisition tracking. | FastTrac Retinal Tracking
    Track-to-prior to Scan
    Acquisition tracking. | Identical | |
    | OCT Angiography
    Algorithms
    (OCTA) | - En face Algorithm
  • Segmentation Algorithm
  • Z-Motion Correction
    Algorithm
  • Flow Contrast Algorithm
    (intensity based + phase=
    complex-based OCT
    Angiography) | - En face Algorithm
  • Segmentation Algorithm
  • Z-Motion Correction
    Algorithm
  • Flow Contrast Algorithm
    (intensity based + phase=
    complex-based OCT
    Angiography) | - En face Algorithm
  • Segmentation
    Algorithm
  • Z-Motion Correction
    Algorithm
  • Flow Contrast
    Algorithm (intensity
    based + phase=
    complex-based OCT
    Angiography) | Identical | |
    | RNFL, Macula,
    Anterior Segment | ANALYSIS AND REPORTS
    Analyses for the Optic Disc:
    • ONH/RNFL OU
    Analysis
    • Guided Progression
    Analysis
    • Advanced Visualization
    • En face
    • 3D Visualization
    • Panomap | Analyses for the Optic Disc:
    • ONH/RNFL OU
    Analysis
    • Guided Progression
    Analysis
    • Advanced Visualization
    • En face
    • 3D Visualization
    • Panomap | Analyses for the Optic
    Disc:
    • ONH/RNFL OU
    Analysis
    • Guided
    Progression
    Analysis
    • Advanced
    Visualization
    • En face
    • 3D Visualization | Identical | |

9

10

11

12

13

K233933 - 510(k) SummaryPage 10 of 28
• Single Eye Summary• Panomap
Analyses for the Macula: Macular thickness Macular change Advanced RPE PanoMap Advanced visualization High-definition image En face Analysis 3D Visualization Ganglion Cell Guided ProgressionSingle Eye Summary Analyses for the Macula: Macular thickness Macular change Advanced RPE PanoMap Advanced visualization High-definition image En face Analysis 3D Visualization Ganglion Cell Guided Progression
Analyses for the Anterior Segment: Anterior Segment Analysis High-Definition Image Analysis Anterior Chamber Scan Analysis HD Angle Analysis HD Cornea AnalysisAnalyses for the Anterior Segment: Anterior Segment Analysis High-Definition Image Analysis
Anterior Chamber Scan Analysis HD Angle Analysis HD Cornea Analysis

14

K233933 - 510(k) SummaryPage 11 of 28
Pachymetry Analysis with Epithelial Thickness Wide Angle-to-Angle Analysis 3D visualizationPachymetry Analysis with Epithelial Thickness Wide Angle-to-Angle Analysis 3D visualizationAnterior Chamber Scan Analysis HD Angle Analysis HD Cornea Analysis Pachymetry Analysis with Epithelial Thickness Wide Angle-to-Angle Analysis 3D visualization
RNFL Thickness AnalysisOCT Fundus image with RNFL Calculation circle OCT Image Extracted from Calculation Circle with RNFL Segmentation RNFL Thickness Map Deviation from Normal Map Average ThicknessOCT Fundus image with RNFL Calculation circle OCT Image Extracted from Calculation Circle with RNFL Segmentation RNFL Thickness Map Average Thickness Quadrant Thicknesses Clock Hour ThicknessesOCT Fundus image with RNFL Calculation circle OCT Image Extracted from Calculation Circle with RNFL Segmentation RNFL Thickness Map Deviation from Normal MapIdentical to the reference device.
K233933 - 510(k) SummaryPage 12 of 28
Quadrant Thicknesses Clock Hour Thicknesses Symmetry value TSNIT (or NSTIN format) RNFL thickness graph with color coding as shown in the Reference DatabaseSymmetry value TSNIT(or NSTIN format) RNFL thickness graph.Average Thickness Quadrant Thicknesses Clock Hour Thicknesses Symmetry value Normative database TSNIT RNFL thickness graph with color coding
Optic Nerve Head (ONH) analysisRim Area / Disc Area Average C/D Ratio Vertical C/D Ratio: Cup Volume Neuro-retinal Rim Thickness B-scan cross section of the ONH Automatic outline of disc margin Automatic outline of cup edge ONH B-scan slices and segmentation Reference ONH database color codingRim Area / Disc Area Average C/D Ratio Vertical C/D Ratio: Cup Volume Neuro-retinal Rim Thickness B-scan cross section of the ONH Automatic outline of disc margin Automatic outline of cup edge ONH B-scan slices and segmentation Disc sizeRim Area / Disc Area Average C/D Ratio Vertical C/D Ratio: Cup Volume Neuro-retinal Rim Thickness B-scan cross section of the ONH Automatic outline of disc margin Automatic outline of cup edge ONH B-scan slices and segmentation Normative ONH database color codingIdentical to the reference device.
K233933 - 510(k) Summary
Page 13 of 28
Disc size
Guided Progression
Analysis (GPA) for
RNFLMultiple RNFL Thickness Maps OCT Fundus image with Calculation circle Overall Average RNFL Thickness Graph Superior RNFL Thickness Graph Inferior RNFL Thickness Graph RNFL Thickness Profile RNFL Summary: RNFL Thickness Map Progression RNFL Thickness Profiles Progression Average RNFL Thickness Progression Manual selection option of scans to include in change analysisMultiple RNFL Thickness Maps OCT Fundus image with Calculation circle Overall Average RNFL Thickness Graph Superior RNFL Thickness Graph Inferior RNFL Thickness Graph RNFL Thickness Profile RNFL Summary: RNFL Thickness Map Progression RNFL Thickness Profiles Progression Average RNFL Thickness Progression Manual selection option of scans to include in change analysisMultiple RNFL Thickness Maps Multiple Deviation from Normal Maps OCT Fundus image with Calculation circle Overall Average RNFL Thickness Graph Superior RNFL Thickness Graph Inferior RNFL Thickness Graph RNFL Thickness Profile RNFL Summary: RNFL Thickness Map Progression RNFL Thickness Profiles Progression Average RNFL Thickness Progression Manual selection option of scans to include in change analysisIdentical to the predicate
device.
K233933 - 510(k) SummaryPage 14 of 28
Guided Progression
Analysis (GPA) for
Ganglion Cell/IPL- Sequential display of
Macular images in Guided
Progression Analysis
(GPA) for Ganglion Cell
Layer/Inner Plexiform
Layer (GCL/IPL)
  • Multiple GCL/IPL
    Thickness Maps
  • Overall Average GCL/IPL
    Thickness Graph
  • Superior GCL/IPL
    Thickness Graph
  • Inferior GCL/IPL
    Thickness Graph
  • GCL/IPL Summary:
  • GCL Thickness Map
    Progression
  • GCL Thickness
    Progression | - Sequential display of
    Macular images in Guided
    Progression Analysis
    (GPA) for Ganglion Cell
    Layer/Inner Plexiform
    Layer (GCL/IPL)
  • Multiple GCL/IPL
    Thickness Maps
  • Overall Average GCL/IPL
    Thickness Graph
  • Superior GCL/IPL
    Thickness Graph
  • Inferior GCL/IPL
    Thickness Graph
  • GCL/IPL Summary:
  • GCL Thickness Map
    Progression
  • GCL Thickness
    Progression | - Sequential display of
    Macular images in
    Guided Progression
    Analysis (GPA) for
    Ganglion Cell
    Layer/Inner
    Plexiform Layer
    (GCL/IPL)
  • Multiple GCL/IPL
    Thickness Maps
  • Overall Average
    GCL/IPL Thickness
    Graph
  • Superior GCL/IPL
    Thickness Graph
  • Inferior GCL/IPL
    Thickness Graph
  • GCL/IPL Summary:
  • GCL Thickness Map
    Progression
  • GCL Thickness
    Progression | Identical |
    | Macular Thickness
    Analysis and
    Display | - Fundus Image with scan
    cube overlay
  • Average thickness and
    volume table
  • ETDRS grid map with
    reference database color
    coding
  • Slice navigators | - Fundus Image with scan
    cube overlay
  • Average thickness and
    volume table
  • ETDRS grid map
  • Slice navigators
  • Horizontal B-scan (X-
    image) | - Fundus Image with
    scan cube overlay
  • Average thickness
    and volume table
  • ETDRS grid map
    with normative
    database color coding
  • Slice navigators | Identical to the reference
    device |

15

16

17

18

K233933 - 510(k) SummaryPage 15 of 28
Horizontal B-scan (X-image)
Vertical slice - A-scan (Y-image)
3-D color ILM-RPE thickness map
3-D surface map of ILM
3-D surface map of RPE
Segmentation line toggle
Measurement calipers
Automatic fovea finding and coordinates
Assorted controls for aligning ETDRS grid
High-Resolution Images option
Edit Layers (enhanced interface)
Zoom controls
Save Image options
Movie optionVertical slice - A-scan (Y-image)
3-D color ILM-RPE thickness map
3-D surface map of ILM
3-D surface map of RPE
Segmentation line toggle
Measurement calipers
Automatic fovea finding and coordinates
Assorted controls for aligning ETDRS grid
High-Resolution Images option
Edit Layers (enhanced interface)
Zoom controls
Save Image options
Movie optionHorizontal B-scan (X-image)
Vertical slice - A-scan (Y-image)
3-D color ILM-RPE thickness map
3-D surface map of ILM
3-D surface map of RPE
Segmentation line toggle
Measurement calipers
Automatic fovea finding and coordinates
Assorted controls for aligning ETDRS grid
High-Resolution Images option
Edit Layers (enhanced interface)
Zoom controls
Save Image options
Movie option

19

K233933 - 510(k) SummaryPage 16 of 28
High Definition
Display- Fundus image with OCT b-
scan reference line (slice
navigator) overlays
  • Enlarged B-scan image of
    selected line
  • Up to 5 thumbnail images | - Fundus image with OCT b-
    scan reference line (slice
    navigator) overlays
  • Enlarged B-scan image of
    selected line
  • Up to 5 thumbnail images | - Fundus image with
    OCT b-scan reference
    line (slice navigator)
    overlays
  • Enlarged B-scan
    image of selected line
  • Up to 5 thumbnail
    images | Identical |
    | Ganglion Cell OU
    Analysis | - GCA Thickness Map
  • GCA Deviation Map
  • Six sectors of average
    thickness – color coded per
    RDB limits
  • Horizontal Macular B-scan
  • Vertical Macular B-scan
  • Average GCL+IPL
    thickness (chart)
  • Data available on screen &
    on printout | - GCA Thickness Map
  • Six sectors of average
    thickness
  • Horizontal Macular B-scan
  • Vertical Macular B-scan
  • Average GCL+IPL
    thickness (chart)
  • Minimum GCL+IPL
    thickness (chart)
  • Data available on screen &
    on printout | - GCA Thickness Map
  • GCA Deviation Map
  • Six sectors of average
    thickness - color
    coded per NDB limits
  • Horizontal Macular
    B-scan
  • Vertical Macular B-
    scan
  • Average GCL+IPL
    thickness (chart)
  • Minimum GCL+IPL
    thickness (chart)
  • Data available on
    screen & on printout | Identical to the reference
    device |
    | K233933 - 510(k) Summary | | Page 17 of 28 | | |
    | Single Eye
    Summary
    (Data from macula
    scan and optic disc
    scan are shown in
    one report) | Macula Analysis:
  • en face image of macular
    cube
  • Macular Thickness map
  • ETDRS grid sector
    thicknesses with
    comparison to reference
    data
  • Macular B-scan
    ONH/RNFL Analysis:
  • en face image of Optic
    Disc Cube
  • RNFL Thickness map
  • Data Table with RNFL and
    ONH data
  • RNFL TSNIT Thickness
    graph
  • ONH B-scan
  • Data available on screen &
    on printout | Macula Analysis:
  • en face image of macular
    cube
  • Macular Thickness map
  • ETDRS grid sector
    thicknesses
  • Macular B-scan
    ONH/RNFL Analysis:
  • en face image of Optic
    Disc Cube
  • RNFL Thickness map
  • Data Table with RNFL and
    ONH data
  • RNFL TSNIT Thickness
    graph
  • ONH B-scan
  • Data available on screen &
    on printout | Macula Analysis:
  • en face image of
    macular cube
  • Macular Thickness
    map
  • ETDRS grid sector
    thicknesses with
    comparison to
    normative data
  • Macular B-scan
    ONH/RNFL Analysis:
  • en face image of
    Optic Disc Cube
  • RNFL Thickness map
  • Data Table with
    RNFL and ONH data
  • RNFL TSNIT
    Thickness graph
  • ONH B-scan
  • Data available on
    screen & on printout | Identical to the reference
    device |
    | Advanced RPE
    Analysis | - Advanced RPE Analysis | - Advanced RPE Analysis | - Advanced RPE
    Analysis | Identical |
    | PanoMap Analysis | - Macular Cube 512x128 or
  • Macular Cube 200x200
    and
  • Optic Disc Cube 200x200
    scan for the same eye. | - Macular Cube 512x128 or
  • Macular Cube 200x200 and
  • Optic Disc Cube 200x200
    scan for the same eye.
  • Montage of Macular Cube | - Macular Cube
    512x128 or
  • Macular Cube
    200x200 and
  • Optic Disc Cube | Identical |
    | | K233933 - 510(k) Summary
    Page 18 of 28 | | | |
    | Wellness Exam
    Report | • Montage of Macular Cube
    and Optic Disc Cube OCT
    en face images

This is an OU wide field
analysis that combines
information from the macular
thickness analysis. PanoMap
analysis, RNFL and ONH
analysis and ganglion cell OU
analysis into one report. | • and Optic Disc Cube OCT
en face images

This is an OU wide field
analysis that combines
information from the
macular thickness analysis.
PanoMap analysis, RNFL
and ONH analysis and
ganglion cell OU analysis
into one report. | 200x200 scan for the
same eye.
• Montage of
Macular Cube and
Optic Disc Cube OCT
en face images
• Not Available | Identical to the predicate
device. |
| OCT Angiography
Report / Analysis | • OCT Angiography Report
• OCT Angiography Change
Analysis | • OCT Angiography Report
• OCT Angiography Change
Analysis | • OCT Angiography
Report
• OCT Angiography
Change Analysis | Identical |
| ORCC & RPE-to-
RPE Fit Slabs (pre-
set) in OCTA
Analysis | • ORCC (Outer Retina to
ChorioCapillaris), Sub-
RPE, RPE-to-RPEFit slabs
in OCTA Analysis | • ORCC (Outer Retina to
ChorioCapillaris), Sub-
RPE, RPE-to-RPEFit slabs
in OCTA Analysis | Not Available | Identical to the predicate
device. |

20

21

22

K233933 - 510(k) SummaryPage 19 of 28
Anterior Segment -
Analysis with
Quantitative
Measurement ToolsAnterior 5-Line Raster Thickness calipers Anterior Segment Cube 3D Visualization Thickness calipers Pachymetry Scan Corneal thickness Epithelial thicknessAnterior 5-Line Raster Thickness calipers Anterior Segment Cube 3D Visualization Thickness calipers Pachymetry Scan Corneal thickness Epithelial thicknessAnterior 5-Line Raster Thickness calipers Anterior Segment Cube 3D Visualization Thickness calipers Pachymetry Scan Corneal thickness Epithelial thicknessIdentical
ONH Angiography
Two-Visit
Comparisonvisual side by side comparison of two angiography visits on the analysis screen and corresponding report print out.visual side by side comparison of two angiography visits on the analysis screen and corresponding report print out.visual side by side comparison of two angiography visits on the analysis screen and corresponding report print out.Identical
Reference Database,
DiverseRetinal Nerve Fiber Layer (RNFL) Thickness Macular Thickness Optic Nerve Head (ONH) Parameters Ganglion Cell/IPL ThicknessNot ApplicableRetinal Nerve Fiber Layer (RNFL) Thickness Macular Thickness Optic Nerve Head (ONH) Parameters Ganglion Cell/IPL ThicknessEquivalent to the reference device.
CIRRUS 6000 RDB provides additional colors when the measurements are within overlapping confidence intervals of the quantile limits.
FUNDUS IMAGING
MethodologyLine Scanning OphthalmoscopeLine Scanning OphthalmoscopeLine Scanning OphthalmoscopeIdentical

23

| | K233933 - 510(k) Summary
Page 20 of 28 | | | |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Optical Source | Super Luminescent Diode
(SLD)
750 nm | Super Luminescent Diode
(SLD)
750 nm | Super Luminescent Diode
(SLD)
750 nm | Identical |
| Optical Power | 20 Hz | > 20 Hz | > 20 Hz | Identical |
| Transverse
Resolution | 25 µm (in tissue) | 25 µm (in tissue) | 25 µm (in tissue) | Identical |
| | IRIS IMAGING | | | |
| Methodology | CMOS camera and LED
illumination | CMOS camera and LED
illumination | CMOS camera and LED
illumination | Identical |
| Optical source | 1280 x 1024 | 1280 x 1024 | 1280 x 1024 | Identical |
| Resolution | During alignment | During alignment | During alignment | Identical |
| Live iris image | Light emitting diode (LED),
700 nm | Light emitting diode (LED),
700 nm | Light emitting diode (LED),
700 nm | Identical |
| | FIXATION | | | |
| Internal fixation
source | LED Array, 21 positions | LED Array, 21 positions | LED Array, 9 positions | Identical to the predicate
device. |
| Internal fixation
focus adjustment | -20D to +20D (diopters) | -20D to +20D (diopters) | -20D to +20D (diopters) | Identical |
| External fixation
source | Mechanically adjustable arm
with LED at the tip | Mechanically adjustable arm
with LED at the tip | Mechanically adjustable arm
with LED at the tip | Identical |
| | ELECTRICAL, PHYSICAL, ENVIRONMENTAL | | | |
| Configuration | Patient module, computer,
media and power supply
integrated into single compact
module | Patient module, computer,
media and power supply
integrated into single compact
module | Patient module, computer,
media and power supply
integrated into single
compact module | Identical |
| Computer | High performance multi-core
processor CORE i7 @ 3.6 GHz,
32GB RAM, 2TB Hard Disk,
Windows 10, SSD | High performance multi-core
processor CORE i7 @ 3.6 GHz,
32GB RAM, 2TB Hard Disk,
Windows 10, SSD | High performance multi-
core processor CORE i7 @
3.1GHz, 16GB RAM, 2TB
Hard Disk, Windows 10 | Identical to the predicate
device. |
| K233933 - 510(k) Summary | | Page 21 of 28 | | |
| Input Devices | Computer mouse/keyboard | Computer mouse/keyboard | Computer mouse/keyboard | Identical |
| Display | 22" Widescreen HD
(Resolution 1920 x 1080) | 22" Widescreen HD
(Resolution 1920 x 1080) | 19" Color Flat Panel Display
(Resolution 1280 x 1024) | Identical to the predicate
device. |
| Network | Network and additional USB
connectors under rear cover. | Network and additional USB
connectors under rear cover. | Network and additional USB
connectors under rear cover. | Identical |
| Optical media
formats supported | 6 USB Media ports | 6 USB Media ports | 6 USB Media ports | Identical |
| Weight | 35 kg (77 lbs) (without
monitor) | 35 kg (77 lbs) (without
monitor) | 36 kg (80 lbs) | Identical to the predicate
device |
| Dimensions | 62.2L x 42.5W x 49.4H (cm) | 62.2L x 42.5W x 49.4H (cm) | 62.2L x 42.5W x 49.4H (cm) | Identical |
| Electrical rating
(115V) | 100-120 V~ 50-60 Hz 6.3A | 100-120 V~ 50-60 Hz 6.3A | 100-120 V~ 50-60 Hz 6.3A | Identical |
| Power rating (115V) | 500W | 500W | 350W | Identical to the predicate
device |
| Fuse rating (115V) | T 6.3 A 250V | T 6.3 A 250V | T 5A 250V | Identical to the predicate
device |
| Max amperage and
voltage rating | T 6.3A 250V | T 6.3 A 250V | T 5A 250V | Identical to the predicate
device |
| Environmental
Conditions:
Transport and
Storage | Temp. -40° to +70° C
30% to 75% (excluding
condensation)
Atmospheric Pressure
500 to 1060 hPa | Temp. -40° to +70° C
30% to 75% (excluding
condensation)
Atmospheric Pressure
500 to 1060 hPa | Temp. -40° to +70° C
30% to 75% (excluding
condensation)
Atmospheric Pressure
500 to 1060 hPa | Identical |
| Environmental
Conditions:
Operation | Temperature:
+10 to +35° С
Relative Humidity:
30% to 75% (excluding
condensation)
Atmospheric Pressure:
700 to 1060 hPa | Temperature:
+10 to +35° С
Relative Humidity:
30% to 75% (excluding
condensation)
Atmospheric Pressure:
700 to 1060 hPa | Temperature:
+10 to +35° C
Relative Humidity:
30% to 75% (excluding
condensation)
Atmospheric Pressure:
700 to 1060 hPa | Identical |

24

25

K233933 - 510(k) Summary Page 22 of 28
Enclosure-
Flammability
RatingsUL 94V-0UL 94V-0UL 94V-0Identical
UX
Protocol buttons- Workflow buttons filter scan patterns for acquisition- Workflow buttons filter scan patterns for acquisitionNot AvailableIdentical to the predicate device
Scan Protocols- user-selectable scan protocols on the patient management screen that define a specific set of scans to be acquired.- User-selectable scan protocols on the patient management screen that define a specific set of scans to be acquired.Not AvailableIdentical to the predicate device
Preferred Analysis- allows the user to preconfigure the preferred analysis to auto load. This is a UI update that allows the user to see first 4 preferred analyses every time- Allows the user to preconfigure the preferred analysis to auto load. This is a UI update that allows the user to see first 4 preferred analyses every timeNot AvailableIdentical to the predicate device
Anterior Segment
Caliper Tool- This tool is able to snap to the surfaces identified by the software, including anterior cornea and posterior cornea.This tool is able to snap to the surfaces identified by the software, including anterior cornea and posterior cornea.Not AvailableIdentical to the predicate device
Analysis switch eye- allows the user to switch to the 'other' eye for the same analysesallows the user to switch to the 'other' eye for the same analysesNot AvailableIdentical to the predicate device
New Circle Tool for
Angiography- a manually placed circle tool for annotation purposesa manually placed circle tool for annotation purposesNot AvailableIdentical to the predicate device
K233933 - 510(k) SummaryPage 23 of 28
New Freehand Tool
for AngiographyA scalable circle and
freeform measurement tools
for the OCT Angiography
and ONH Angiography
enface image which
displays estimated values
for area and
perimeter/diameterA scalable circle and freeform
measurement tools for the OCT
Angiography and ONH
Angiography enface image
which displays estimated values
for area and perimeter/diameterNot AvailableIdentical to the predicate
device

26

27

Non-Clinical Performance Testing

Non-clinical system testing was provided on the predicate K222200 and no new non-clinical information is provided in this submission since sterlity, shelf-life, biocompatibility, and animal testing was not performed as there were no modifications made to the device in this regard.

Other verification and validation testing (such as software) was conducted, and documentation was provided as recommended by the FDA's Guidance for Industry and Staff "Content of Premarket Submissions for Device Software Functions (June 2023)". All testing passed.

Clinical Performance Testing

Repeatability and Reproducibility testing was leveraged from the K22200; and thus, was not provided in this submission.

The addition of the reference database required new testing which is summarized below.

A prospective, multi-site reference database (RDB) study was conducted at eight (8) clinical sites across the USA to develop the CIRRUS 6000 RDB for macular thickness and optic nerve head scan values.

The Subject Device's RDB was created to help clinicians assess and effectively compare a patient's measurements to that of healthy subjects, representative of the general population.

The eligibility criteria included subjects 18 years of age and older, best corrected visual acuity (BCVA) of 20/40 or better in either eye, IOP