Search Results
Found 5 results
510(k) Data Aggregation
(169 days)
Ask a specific question about this device
(227 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner
It is a powder-free chloroprene patient examination gloves made by on-line polymer-coating and mild on-line chlorination process. The process modifies the surface characteristics without the use of any dusting or donning powder. The gloves are non-sterile, disposable and for single use only.
Characteristics:
- Ambidextrous with beaded cuff
- Blended synthetic latex (Chloroprene and Nitrile Latex)
- Blue or green Coloured ●
- Five sizes: X-Small, Small, Medium, Large and X-Large. ●
The provided document is a 510(k) Premarket Notification for Chloroprene Examination Gloves Powder Free (Blue, Green). This document focuses on demonstrating substantial equivalence to a predicate device, which involves comparing material, design, and performance characteristics for a medical device (examination gloves), not an AI/ML-driven device.
Therefore, the requested information regarding AI/ML device acceptance criteria and study details (such as sample size for test/training sets, data provenance, expert adjudication, MRMC studies, standalone performance, and ground truth establishment) is not applicable to this document.
What the document does provide is a detailed comparison of the physical and biological characteristics of the proposed gloves against established ASTM standards and the predicate device.
Here's a summary of the acceptance criteria and reported performance for the Chloroprene Examination Gloves Powder Free (Blue, Green), as extracted from the provided text:
1. A table of acceptance criteria and the reported device performance:
| Characteristic | Standard | Acceptance Criteria | Reported Device Performance (Subject Device K213993) |
|---|---|---|---|
| Product Code | - | LZA (Same as predicate) | LZA |
| Intended Use | - | Intended for medical purposes, worn on examiner's hand/finger to prevent contamination between patient and examiner (Same as predicate) | Intended for medical purposes, worn on examiner's hand/finger to prevent contamination between patient and examiner |
| Design | - | Powder Free, Non-Sterile, Ambidextrous, Beaded Cuff (Same as predicate) | Powder Free, Non-Sterile, Ambidextrous, Beaded Cuff |
| Indications for Use | - | A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner (Same as predicate) | A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner |
| Construction | - | Ambidextrous, Polymer Coated or Chlorinated, Powder Free Nitrile (Similar to predicate's Ambidextrous, Polyacrylic Polymer Inner Coating, Powder Free Nitrile) | Ambidextrous, Polymer Coated or Chlorinated, Powder Free Nitrile |
| Single Use | - | Yes (Same as predicate) | Yes |
| Packaging | - | Packed in Dispenser Boxes (Same as predicate) | Packed in Dispenser Boxes |
| Dimension (Length) | ASTM D6977-19 | XSmall, Small: 220 mm minMedium, Large, XLarge: 230 mm min | Meet 220mm min (XSmall, Small)Meet 230mm min (Medium, Large, XLarge) |
| Dimension (Thickness) | ASTM D6977-19 | Palm: 0.05 mm minFinger: 0.05 mm min | Meet 0.05mm min (Palm)Meet 0.05mm min (Finger) |
| Dimension (Width) | ASTM D6977-19 | XSmall: 70 ± 10 mmSmall: 80 ± 10 mmMedium: 95 ± 10 mmLarge: 110 ± 10 mmXLarge: 120 ± 10 mm | Meet 70 ± 10 mm (XSmall)Meet 80 ± 10 mm (Small)Meet 95 ± 10 mm (Medium)Meet 110 ± 10 mm (Large)Meet 120 ± 10 mm (XLarge) |
| Physical Properties (Before Aging) | ASTM D6977-19 | Tensile Strength: Min. 14 MPaUltimate Elongation: Min. 500% | Meet 14MPa minMeet 500% min |
| Physical Properties (After Aging) | ASTM D6977-19 | Tensile Strength: Min. 14 MPaUltimate Elongation: Min. 400% | Meets 14MPa minMeet 400% min. |
| Water Leak Test (Before Aging) | ASTM D6977-19 | G-I, AQL 2.5 (FDA GII, AQL 2.5) | Meet AQL 1.5 |
| Water Leak Test (After Aging) | ASTM D5151-19 | (Assumed same as Before Aging in this context - AQL 2.5 for general examination gloves) | Meet AQL 2.5 |
| Powder Free Residue | ASTM D6977-19ASTM D6124-06 | Max. 2 mg/glove | Meet 2mg/glove max. |
| Biocompatibility: Primary Skin Irritation Test | ISO 10993-10 | No Animal Irritation | Passes. Conclusion: Did not cause an irritant response. |
| Biocompatibility: Skin Sensitization Test | ISO 10993-10 | No Animal Irritation (No sensitization effect) | Passes. Conclusion: Did not produce a skin sensitization effect. |
| Biocompatibility: In Vitro Cytotoxicity | ISO 10993-5 | No cytotoxicity reactivity at 100.0% extract concentrations under the conditions of this test | Exhibited cytotoxicity response at 100% of extract concentration. (Remarked as "Different" compared to predicate which had "No Information") – It still passed the overall safety conclusion. |
| Biocompatibility: Acute Systemic Toxicity | ISO 10993-11 | No adverse biological reaction | Passed. Conclusion: Showed no adverse biological reaction. (Remarked as "Different" compared to predicate which had "No Information") |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated as a single "test set" sample size for a clinical study. The data presented are results of various laboratory and biocompatibility tests on the gloves. These tests typically involve a specific number of gloves or samples per test as per the ASTM and ISO standards referenced (e.g., AQL for water leak tests, specific batches for physical property tests). The exact number of gloves or samples for each test is not detailed in this summary, but the results confirm they "Meet" or "Pass" the specified standards.
- Data Provenance: The manufacturer is Careglove Global SDN BHD, located in Malaysia. The testing was conducted to comply with international standards (ASTM and ISO). The document does not specify whether the data is retrospective or prospective in the context of a clinical trial, as this is a premarket notification for a Class I general medical device based on engineering and biocompatibility testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This is a physical medical device (examination gloves) where "ground truth" is established through adherence to standardized laboratory test methods (ASTM, ISO) for physical properties and biocompatibility, not by expert interpretation of medical images or patient outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. As per point 3, this is a physical device tested against objective standards, not an AI/ML system requiring human expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for this device's performance is based on objective measurements defined by international standards (ASTM D6977-19, ASTM D5151-19, ASTM D6124-06, ISO 10993-10, ISO 10993-5, ISO 10993-11) for physical properties (e.g., length, thickness, tensile strength, elongation, freedom from holes, powder residue) and biocompatibility (e.g., skin irritation, sensitization, cytotoxicity, systemic toxicity).
8. The sample size for the training set
- Not Applicable. This is not an AI/ML device; there is no "training set."
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI/ML device; there is no "training set" or establishment of ground truth in that context.
Ask a specific question about this device
(327 days)
A sterile latex surgical glove, powder free is a disposable device intended for medical purposes that is worn by operating room personnel to protect a surgical wound from contamination.
The subject device is a powder-free variation of the class I latex surgical gloves made by on-line polymercoating process. This process modifies the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder.
The provided document is a 510(k) summary for a medical device: "Sterile Latex Surgical Gloves, Powder Free." It details the device's characteristics, comparison to a predicate device, and the results of non-clinical testing. It explicitly states that clinical testing was "Not applicable." Therefore, the document does not contain information about a study proving the device meets acceptance criteria related to a diagnostic or AI-based medical device performance study.
The questions in the prompt relate to the evaluation of an AI or diagnostic device, asking about sample sizes, expert ground truth, MRMC studies, and training sets. This information is not relevant to a 510(k) submission for surgical gloves, which are assessed based on physical properties, sterility, and biocompatibility.
However, I can extract the acceptance criteria and reported device performance for the surgical gloves based on the provided tables.
Here's the information derived from the document, tailored to the context of surgical gloves:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally based on meeting the specifications of ASTM D3577-15 and other relevant standards (ISO 11737-2, ISO 10993-10, ISO 10993-5, ISO 10993-11, USP General Chapter 40). The reported device performance indicates whether the subject device "Meets" or "Passes" these specifications.
| Characteristic | Standard | Specification | Reported Device Performance (Subject Device K190241) |
|---|---|---|---|
| Dimension (Length) | ASTM D3577-15 | ||
| Length (size: 5.5), mm | 245 min | Meet 245mm min | |
| Length (size: 6.0), mm | 265 min | Meet 265mm min | |
| Length (size: 6.5), mm | 265 min | Meet 265mm min | |
| Length (size: 7.0), mm | 265 min | Meet 265mm min | |
| Length (size: 7.5), mm | 265 min | Meet 265mm min | |
| Length (size: 8.0), mm | 265 min | Meet 265mm min | |
| Length (size: 8.5), mm | 265 min | Meet 265mm min | |
| Length (size: 9.0), mm | 265 min | Meet 265mm min | |
| Dimension (Thickness) | ASTM D3577-15 | ||
| Thickness (cuff), mm | 0.10 min | Meet 0.10mm min | |
| Thickness (palm), mm | 0.10 min | Meet 0.10mm min | |
| Thickness (finger), mm | 0.10 min | Meet 0.10mm min | |
| Dimension (Width) | ASTM D3577-15 | ||
| Width (size: 5.5), mm | 70 ± 6 | Meet 70 ± 6 mm | |
| Width (size: 6.0), mm | 76 ± 6 | Meet 76 ± 6 mm | |
| Width (size: 6.5), mm | 83 ± 6 | Meet 83 ± 6 mm | |
| Width (size: 7.0), mm | 89 ± 6 | Meet 89 ± 6 mm | |
| Width (size: 7.5), mm | 95 ± 6 | Meet 95 ± 6 mm | |
| Width (size: 8.0), mm | 102 ± 6 | Meet 102 ± 6 mm | |
| Width (size: 8.5), mm | 108 ± 6 | Meet 108 ± 6 mm | |
| Width (size: 9.0), mm | 114 ± 6 | Meet 114 ± 6 mm | |
| Water Leak Test | ASTM D3577-15, ASTM D5151-06 | ||
| Before Aging, AQL | G-I, AQL 1.5 (FDA GII, AQL 1.5) | Meet AQL 1.5 | |
| After Aging, AQL | Meet AQL 1.5 | ||
| Physical Properties (Before Aging) | ASTM D3577-15 | ||
| Tensile Strength (MPa) | Min. 24 | Meet 24MPa min. | |
| Ultimate Elongation (%) | Min. 750 | Meet 750% min | |
| Stress at 500% Elongation (MPa) | Max. 5.5 | Meet 5.5 MPa max | |
| Physical Properties (After Aging) | ASTM D3577-15 | ||
| Tensile Strength (MPa) | Min. 18 | Meets 18MPa min | |
| Ultimate Elongation (%) | Min. 560 | Meet 560% min. | |
| Residual Powder Content | ASTM D3577-15, ASTM D6124-06 | Max. 2mg/glove | Meet 2mg/glove max. |
| Extractable Protein Content | ASTM D5712-10 | 50 µg/dm² and below | Meet Specification |
| Biocompatibility Tests | |||
| Primary Skin Irritation Test | ISO 10993-10 | No Animal Irritation | Passes (did not cause an irritant response) |
| Skin Sensitization Test | ISO 10993-10 | No Animal Sensitization | Passes (did not produce a skin sensitization effect) |
| Cytotoxicity Test | ISO 10995-5 | Non-cytotoxicity | moderate-cytotoxicity at 6 cm²/mL extraction |
| Acute Toxicity Oral | ISO 10993-11 | No toxic | No adverse biological reaction |
| Pyrogenic Test | USP General Chapter 40 | Non pyrogenic | Non pyrogenic |
Regarding the specific questions about an AI/diagnostic device evaluation:
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
This information is not applicable. The device is a surgical glove, not a diagnostic or AI device. The testing involved physical and chemical properties and biocompatibility, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
This information is not applicable. Ground truth as understood in AI/diagnostic studies (e.g., disease presence) is not relevant for surgical gloves. The "ground truth" for the gloves are the established international standards for their physical, chemical, and biological properties.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not applicable. Adjudication methods are used in interpreting ambiguous diagnostic findings, which is not relevant for surgical gloves.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. MRMC studies are used to evaluate human reader performance with and without AI assistance in diagnostic tasks. This device is a surgical glove.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable. Standalone algorithm performance is relevant for AI devices. This device is a physical product.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For surgical gloves, the "ground truth" is defined by adherence to recognized international standards (e.g., ASTM, ISO, USP) for physical properties, sterility, and biocompatibility. This involves laboratory testing using defined methodologies, not clinical outcomes data or expert diagnostic consensus.
8. The sample size for the training set:
This information is not applicable. Training sets are used for machine learning. This document describes testing of manufactured physical products, not an AI model.
9. How the ground truth for the training set was established:
This information is not applicable. There is no training set for a surgical glove.
Ask a specific question about this device
(85 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
It is the powder-free variation of the class I latex patient examination gloves made by on-line polymer-coating and mild on-line chlorination process modifies the surface characteristics and causes it to remain tackfree without the use of any dusting or donning powder.
Based on the provided document, the device in question is Powder Free Nitrile Examination Gloves, Blue (colored). This document is a 510(k) Summary for a medical device which typically focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a clinical study proving device performance against specific acceptance criteria in the context of an AI-powered diagnostic device.
Therefore, many of the requested elements for an AI device (like human reader improvement with AI assistance, standalone performance, training set details, expert qualifications for ground truth) are not applicable to this type of device and submission.
However, I can extract the acceptance criteria and performance data for the physical and biological characteristics of the gloves as presented in the "Summary of the Technological Characteristic" section.
Here's the breakdown:
1. A table of acceptance criteria and the reported device performance
| Characteristic | Acceptance Criteria (Predicate / Subject Device) | Reported Device Performance (Subject Device) |
|---|---|---|
| Dimension | ||
| Length (size: XSmall) | Meet 220mm min | Meet 220mm min |
| Length (size: Small) | Meet 220mm min | Meet 220mm min |
| Length (size: Medium) | Meet 230mm min | Meet 230mm min |
| Length (size: Large) | Meet 230mm min | Meet 230mm min |
| Length (size: XLarge) | Meet 230mm min | Meet 230mm min |
| Thickness (palm) | Meet 0.05mm min | Meet 0.05mm min |
| Thickness (finger) | Meet 0.05mm min | Meet 0.05mm min |
| Width (size: XSmall) | Meet 70 ± 10 mm | Meet 70 ± 10 mm |
| Width (size: Small) | Meet 80 ± 10 mm | Meet 80 ± 10 mm |
| Width (size: Medium) | Meet 95 ± 10 mm | Meet 95 ± 10 mm |
| Width (size: Large) | Meet 110 ± 10 mm (Subject); 111 ± 10 mm (Predicate) | Meet 110 ± 10 mm |
| Width (size: XLarge) | Meet 120 ± 10 mm | Meet 120 ± 10 mm |
| Physical Properties | ||
| (Before Ageing) Tensile Strength | Meet 14MPa min. | Meet 14MPa min. |
| (Before Ageing) Elongation | Meet 500% min | Meet 500% min |
| (After Aging) Tensile Strength | Meets 14MPa min | Meets 14MPa min |
| (After Aging) Elongation | Meet 400% min. | Meet 400% min. |
| Water Leak Test (Before Aging) | Meet AQL 1.5 | Meet AQL 1.5 |
| Water Leak Test (After Aging) | Meet AQL 2.5 | Meet AQL 2.5 |
| Residual Powder Content | Meet 2mg/glove max. | Meet 2mg/glove max. |
| Biocompatibility Test | ||
| Primary Skin Irritation Test | Passes; No irritant response | Passes; No irritant response |
| Skin Sensitization Test | Passes; No skin sensitization effect | Passes; No skin sensitization effect |
2. Sample sized used for the test set and the data provenance:
The document does not specify the exact sample sizes used for each test (e.g., how many gloves were tested for length, tensile strength, or water leaks). It simply states that the subject device "Meet[s]" or "Passes" the criteria. The provenance of the data is not explicitly stated beyond being presented in a submission from Careglove Global Sdn Bhd, Malaysia, implying the testing was conducted to support their device. It's standard for these types of tests to be prospective, laboratory-based evaluations.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This device is a physical product (examination gloves), not an AI-powered diagnostic tool requiring expert interpretation for "ground truth". The performance criteria are physical and chemical measurements or biological response assessments, not based on expert consensus.
4. Adjudication method for the test set:
Not applicable for the same reasons as #3. Quality control of physical properties does not typically involve expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an AI device. The "standalone performance" is the inherent physical and chemical properties of the glove itself.
7. The type of ground truth used:
The "ground truth" for the physical and chemical properties of the gloves is established by standardized laboratory testing methods (e.g., likely conforming to ASTM standards like D6310, as mentioned in the document). For biocompatibility, the ground truth is established by the biological response observed in appropriate in vivo or in vitro models as per standard biocompatibility testing protocols.
8. The sample size for the training set:
Not applicable. This is not an AI device that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. This is not an AI device.
Ask a specific question about this device
(322 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
It is the powder-free variation of the class I latex patient examination gloves made by on-line polymercoating and mild on-line chlorination process. The process modifies the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder.
The document is a 510(k) Premarket Notification for "Latex Examination Gloves Powder Free". It does not describe an AI/ML medical device, therefore, the requested information for acceptance criteria and study details for such a device is not applicable.
The document focuses on demonstrating substantial equivalence to a predicate device (K152593) for patient examination gloves.
However, I can extract the acceptance criteria and performance of the device based on the non-clinical tests mentioned, as they serve a similar purpose of demonstrating safety and effectiveness for this medical device.
1. Table of acceptance criteria and the reported device performance
| Parameter | Acceptance Criteria (ASTM Specifications) | Reported Device Performance (K161833) | Status |
|---|---|---|---|
| Dimension (Length) | |||
| Size XSmall | 220 mm min. | 224mm-236mm | Meet Specification |
| Size Small | 220 mm min. | 225mm-240mm | Meet Specification |
| Size Medium | 230 mm min. | 235mm-246mm | Meet Specification |
| Size Large | 230 mm min. | 235mm-245mm | Meet Specification |
| Size XLarge | N/A | N/A | N/A |
| Thickness | |||
| Palm | 0.08 mm min. | 0.12mm-0.14mm | Meet Specification |
| Finger | 0.08 mm min. | 0.15mm-0.17mm | Meet Specification |
| Width | |||
| Size XSmall | 70 ± 10 mm | 73mm-75mm | Meet Specification |
| Size Small | 80 ± 10 mm | 82mm-87mm | Meet Specification |
| Size Medium | 95 ± 10 mm | 94mm-97mm | Meet Specification |
| Size Large | 111 ± 10 mm | 102mm-107mm | Meet Specification |
| Size XLarge | N/A | N/A | N/A |
| Physical Properties | |||
| Tensile Strength, Before Aging | 18 MPa min. | 21.18MPa-26.17MPa | Meet Specification |
| Stress at 500% Elongation | 5.5 MPa max. | 3.0MPa-4.2MPa | Meet Specification |
| Ultimate Elongation, Before Aging | 650% min. | 750.20%-820.20% | Meet Specification |
| Tensile Strength, After Aging | 14 MPa min. | 18.28MPa-23.88MPa | Meet Specification |
| Ultimate Elongation, After Aging | 500% min. | 550.40%-700.50% | Meet Specification |
| Water Leak Test, 1000 ml (AQL) | |||
| Before Aging | 2.5 | 1.5 and below | Meet Specification |
| After Aging | 2.5 | 2.5 and below | Meet Specification |
| Extractable Protein | |||
| Water Extractable Protein | 50 µg/dm² | 50 µg/dm² and below | Meet Specification |
| Residual Powder Content | |||
| Residual Powder Content | 2 mg/glove max. | 2 mg/glove and below | Meet Specification |
2. Sample size used for the test set and the data provenance
The document does not explicitly state the sample sizes for each specific test (e.g., how many gloves were tested for length, thickness, or tensile strength). However, the tests were performed "as per ASTM D3578-05, Standard Specification for Rubber Examination Gloves and 21 CFR 800.20." These standards typically define the sampling plans. The data provenance is Malaysia, as the applicant, Careglove Global SDN BHD, is located in Seremban, Negeri Sembilan, Malaysia. The tests are non-clinical and would be considered prospective in the sense that they were conducted specifically for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not an AI/ML device that requires expert ground truth for image or data interpretation. The ground truth for the glove's performance is established by objective measurements against established industry standards (ASTM D3578-05).
4. Adjudication method for the test set
Not applicable, as it's not an AI/ML device requiring human adjudication of performance. The performance is determined by direct physical and chemical measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML device.
7. The type of ground truth used
The ground truth is based on established industry standards and regulatory requirements, specifically:
- ASTM D3578-05 (Standard Specification for Rubber Examination Gloves)
- 21 CFR 800.20 (dealing with devices that prevent contamination)
- ISO 10993-10 (Biological evaluation of medical devices - Tests for irritation and skin sensitization)
These standards define the acceptable performance limits for the characteristics tested.
8. The sample size for the training set
Not applicable. This is not an AI/ML device, so there is no "training set."
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device, so there is no "training set" or ground truth for it. The ground truth for the device's performance is established by the aforementioned ASTM and ISO standards for physical characteristics and biocompatibility.
Ask a specific question about this device
Page 1 of 1