Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K020735
    Date Cleared
    2002-07-12

    (128 days)

    Product Code
    Regulation Number
    882.1320
    Why did this record match?
    Applicant Name (Manufacturer) :

    CATHAY INTERNATIONAL, LTD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SOF-PACH™ Reusable Cutaneous Neurostimulation Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. SOF-PACH™ Reusable Electrodes are designed and intended to be used with marketed. Electrical Stimulators i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) of PGF (Pulsed Galvanic Stimulation).

    These electrodes will include the precaution statement: Federal Law restricts the device to sale by or on the order of a licensed practitioner or therapist.

    Device Description

    SOF-PACH™ Reusable Neurostimulation Electrodes are non-sterile laminated, flexible structures composed of materials commonly used in this application:

    First layer - Various cloths, tapes, etc. including Tricot/polyester fabric, polyethylene foam or a polypropylene substrate, coated with biocompatible adhesive.

    Second layer Electrically conductive carbon coated with or without Ag/AgCl, and Activated Carbon Mesh material.

    Third layer Biocompatible conductive hydrogel coupling media (CATHAY's Gel) which has passed the required skin sensitivity testing criteria as specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact. These tests include Cytotoxicity, Sensitization and Primary Skin Irritation Tests.

    The electrodes are designed for single-patient/multiple application use. They have a low profile construction designed for comfort under clothing and utilize tan-tone fabric for reduced visibility. Because of the adhesive nature of the biocompatible hydrogel, no securing materials are required to secure the device to the patient's skin. The electrodes have two different types of contact points that can be used to link the stimulation device to the electrodes. These contact points are compatible with standard marketed Neurostimulation devices.

    1. Lead wire assembly 6" wire with .080 inch diameter recessed female socket connected to one side of the wire. The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR Part 898
    2. Male snap assembly Two-part sensor and stud mated together.
    AI/ML Overview

    This document is a 510(k) summary for the SOF-PACH™ Reusable Neurostimulation Electrodes (K020735). It describes the device, its technological characteristics, and how its safety and effectiveness were established by comparison to predicate devices.

    Here's an analysis of the provided text in relation to your questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Safety: Skin Irritation/SensitivityCATHAY's Gel passed Cytotoxicity, Sensitization, and Primary Skin Irritation Tests according to Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1.
    Effectiveness: Impedance Levels (point-to-point @ 1 K-Hz)SOF-PACH™ electrode impedance values (125-250 ohms) were comparable to other marketed electrodes tested.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "CATHAY's Gel has passed the required skin sensitivity testing criteria" and "point-to-point impedance testing for all manufacturers' electrodes (including SOF-PACH™) ranged from 125-250 ohms". However, the specific sample sizes for these tests (both for the gel biocompatibility and the impedance testing) are not explicitly stated in the provided text.

    Data Provenance: The document does not specify countries of origin for the test data. The biocompatibility tests are likely conducted in a lab setting, and impedance tests on the devices themselves. There is no indication of retrospective or prospective patient data, as the tests are on the device itself and its components, not involving human subjects in a clinical trial for effectiveness.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this 510(k) submission. The "ground truth" for this device's acceptance criteria relies on standardized material testing (biocompatibility) and direct physical measurement (electrical impedance), not on expert clinical interpretation of data.

    4. Adjudication Method for the Test Set

    This is not applicable. As mentioned above, the acceptance criteria are based on standardized tests and direct measurements, not on subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, an MRMC comparative effectiveness study was not done. This device is a neurostimulation electrode, and its safety and effectiveness are established through testing its physical and material properties, and comparison to predicate devices' stated characteristics, not through analyzing human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This device is a physical medical device, not an AI or software algorithm.

    7. The Type of Ground Truth Used

    The ground truth for the safety criteria (skin irritation/sensitivity) was established by passing standardized biocompatibility tests (Cytotoxicity, Sensitization, and Primary Skin Irritation Tests) as specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1.

    The ground truth for the effectiveness criteria (impedance) was established by direct electrical measurement using a standard voltmeter (point-to-point impedance at 1 K-Hz). The acceptance was based on comparability to predicate devices.

    8. The Sample Size for the Training Set

    This is not applicable. This submission describes a physical medical device, not an algorithm that requires a training set. The device itself is manufactured, and its components and final product are tested.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1