(128 days)
SOF-PACH™ Reusable Cutaneous Neurostimulation Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. SOF-PACH™ Reusable Electrodes are designed and intended to be used with marketed. Electrical Stimulators i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) of PGF (Pulsed Galvanic Stimulation).
These electrodes will include the precaution statement: Federal Law restricts the device to sale by or on the order of a licensed practitioner or therapist.
SOF-PACH™ Reusable Neurostimulation Electrodes are non-sterile laminated, flexible structures composed of materials commonly used in this application:
First layer - Various cloths, tapes, etc. including Tricot/polyester fabric, polyethylene foam or a polypropylene substrate, coated with biocompatible adhesive.
Second layer Electrically conductive carbon coated with or without Ag/AgCl, and Activated Carbon Mesh material.
Third layer Biocompatible conductive hydrogel coupling media (CATHAY's Gel) which has passed the required skin sensitivity testing criteria as specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact. These tests include Cytotoxicity, Sensitization and Primary Skin Irritation Tests.
The electrodes are designed for single-patient/multiple application use. They have a low profile construction designed for comfort under clothing and utilize tan-tone fabric for reduced visibility. Because of the adhesive nature of the biocompatible hydrogel, no securing materials are required to secure the device to the patient's skin. The electrodes have two different types of contact points that can be used to link the stimulation device to the electrodes. These contact points are compatible with standard marketed Neurostimulation devices.
- Lead wire assembly 6" wire with .080 inch diameter recessed female socket connected to one side of the wire. The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR Part 898
- Male snap assembly Two-part sensor and stud mated together.
This document is a 510(k) summary for the SOF-PACH™ Reusable Neurostimulation Electrodes (K020735). It describes the device, its technological characteristics, and how its safety and effectiveness were established by comparison to predicate devices.
Here's an analysis of the provided text in relation to your questions:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Safety: Skin Irritation/Sensitivity | CATHAY's Gel passed Cytotoxicity, Sensitization, and Primary Skin Irritation Tests according to Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1. |
| Effectiveness: Impedance Levels (point-to-point @ 1 K-Hz) | SOF-PACH™ electrode impedance values (125-250 ohms) were comparable to other marketed electrodes tested. |
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "CATHAY's Gel has passed the required skin sensitivity testing criteria" and "point-to-point impedance testing for all manufacturers' electrodes (including SOF-PACH™) ranged from 125-250 ohms". However, the specific sample sizes for these tests (both for the gel biocompatibility and the impedance testing) are not explicitly stated in the provided text.
Data Provenance: The document does not specify countries of origin for the test data. The biocompatibility tests are likely conducted in a lab setting, and impedance tests on the devices themselves. There is no indication of retrospective or prospective patient data, as the tests are on the device itself and its components, not involving human subjects in a clinical trial for effectiveness.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
This type of information is not applicable to this 510(k) submission. The "ground truth" for this device's acceptance criteria relies on standardized material testing (biocompatibility) and direct physical measurement (electrical impedance), not on expert clinical interpretation of data.
4. Adjudication Method for the Test Set
This is not applicable. As mentioned above, the acceptance criteria are based on standardized tests and direct measurements, not on subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
No, an MRMC comparative effectiveness study was not done. This device is a neurostimulation electrode, and its safety and effectiveness are established through testing its physical and material properties, and comparison to predicate devices' stated characteristics, not through analyzing human reader performance with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, a standalone algorithm performance study was not done. This device is a physical medical device, not an AI or software algorithm.
7. The Type of Ground Truth Used
The ground truth for the safety criteria (skin irritation/sensitivity) was established by passing standardized biocompatibility tests (Cytotoxicity, Sensitization, and Primary Skin Irritation Tests) as specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1.
The ground truth for the effectiveness criteria (impedance) was established by direct electrical measurement using a standard voltmeter (point-to-point impedance at 1 K-Hz). The acceptance was based on comparability to predicate devices.
8. The Sample Size for the Training Set
This is not applicable. This submission describes a physical medical device, not an algorithm that requires a training set. The device itself is manufactured, and its components and final product are tested.
9. How the Ground Truth for the Training Set Was Established
This is not applicable as there is no training set for this type of device submission.
{0}------------------------------------------------
ເງິນເ 12 2002
SOF-PACH™ Reusable Neurostimulation Electrodes 510(k) Summary
The following information is submitted in accordance with the requirements of 21CFR807.92
| Submitter's name: | Anne JenCATHAY INTERNATIONAL LTD.8F-3, NO. 29, CHUNGSHAN N.RD. SEC. 3,TAIPEI, TAIWAN, R. O. C.TEL: 02-2591-7938FAX: 02-2591-7912E-Mail: annepete@seed.net.tw |
|---|---|
| Contact person: | Anne Jen |
| Date Prepared: | 17 January 2002 |
| Device Name: | SOF-PACH™ Reusable Neurostimulation Electrodes |
| Comment/Usual Name: | Neurostimulation Electrodes |
| Classification Name: | Electrodes, cutaneous |
| Predicate devices: | K963125 (M & R Manufacturing, Inc.)K932849 (Pepin Mfg., Inc.)K011411 (Pain Management Technologies) |
Device description: SOF-PACH™ Reusable Neurostimulation Electrodes are non-sterile laminated, flexible structures composed of materials commonly used in this application:
First layer - Various cloths, tapes, etc. including Tricot/polyester fabric, polyethylene foam or a polypropylene substrate, coated with biocompatible adhesive.
- Second layer Electrically conductive carbon coated with or without Ag/AgCl, and Activated Carbon Mesh material.
- Third layer Biocompatible conductive hydrogel coupling media (CATHAY's Gel) which has passed the required skin sensitivity testing criteria as specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact. These tests include Cytotoxicity, Sensitization and Primary Skin Irritation Tests.
{1}------------------------------------------------
The electrodes are designed for single-patient/multiple application use. They have a low profile construction designed for comfort under clothing and utilize tan-tone fabric for reduced visibility. Because of the adhesive nature of the biocompatible hydrogel, no securing materials are required to secure the device to the patient's skin. The electrodes have two different types of contact points that can be used to link the stimulation device to the electrodes. These contact points are compatible with standard marketed Neurostimulation devices.
-
- Lead wire assembly 6" wire with .080 inch diameter recessed female socket connected to one side of the wire. The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR Part 898
-
- Male snap assembly Two-part sensor and stud mated together.
Technological Characteristics: SOF-PACH™ Reusable Neurostimulation electrodes are technologically equivalent to the predicate devices. They are physically and technically similar to those currently being marketed for "Neurostimulation" i.e., TENS (Transcutaneous Electrical Nerve Stimulation), MENS (Microcurrent Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) and PGS (Pulsed Galvanic Stimulation).
Safety and Effectiveness: CATHAY considers their SOF-PACH™ Neurostimulation electrodes to be as safe and effective as the M&R Disposable Neurostimulation electrodes, the PMI Neurostimulation electrodes and the PMT Neurostimulation electrodes, which were previously found to be substantially equivalent via 510(k) Premarket Notifications. The safety issues analyzed were skin irritation. The effectiveness of the electrodes was determined by comparing impedance levels in point-to-point testing using a voltmeter.
The first safety issue considered was whether the gel, which is used to adhere the electrode to the skin and which is the only portion of the electrode to maintain skin contact, would cause any skin irritation. CATHAY's Gel has passed the required skin sensitivity testing criteria specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact. These tests include Cytotoxicity, Sensitization and Primary Skin Irritation tests.
Because there are no published performance standards for Neurostimulation electrodes, CATHAY uses impedance levels as the criteria for effectiveness testing. When measuring the electrodes' impedance, driven at 1 K-Hz, using a standard voltmeter, the point-to-point impedance testing for all manufacturers' electrodes (including SOF-PACH™) ranged from 125-250 ohms from the top center portion of the electrode to the underside of the outside edge. Once this testing was complete, it was determined that SOF-PACH™s electrode impedance values were comparable to other marketed electrodes tested.
For the above reason, CATHAY considers its Reusable TENS/NMES electrodes to be as safe and effective as the predicate devices, as well as to multiple other TENS/NMES electrodes currently being marketed, including those manufactured by:
Uni-Patch (K962910), Katecho Incorporated (K000870), MSB Iimited (K980229), Axelgaard (K983741), Selective Med Components, Inc. (K946230), Medtronic, Inc. (K875284), Medical Design & Manufacturing Corp. (K895604), and Labeltape Meditect, Inc. (K900656, K902719 and K894043).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest wings or feathers.
JUL 12 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mrs. Anne Jen Cathay International Ltd. 8F-3, No. 29, Chungshan N.Rd. Sec.3, Taipei, Taiwan, R.O.C.
Re: K020735
Trade/Device Name: SOF-PACH Reusable Neurostimulation Electrodes Models: CM4040, CM505, CM50100, and CM 75 Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY Dated: June 11, 2002 Received: June 14, 2002
Dear Mrs. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic
{3}------------------------------------------------
Page 2 - Mrs. Anne Jen
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in vour Section 510(k) Page 2 premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Sincerely yours,
Arto Salin, M. Witte, PhD, M.D.
lia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEVICE NAME: SOF-PACH™ REUSABLE NEUROSTIMULATION ELECTRODES
INDICATIONS FOR USE:
SOF-PACH™ Reusable Cutaneous Neurostimulation Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. SOF-PACH™ Reusable Electrodes are designed and intended to be used with marketed. Electrical Stimulators i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) of PGF (Pulsed Galvanic Stimulation).
These electrodes will include the precaution statement: Federal Law restricts the device to sale by or on the order of a licensed practitioner or therapist.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription use X (Per 21 CFR 801, 109) OR:
Over-the Counter Use (Optional Format 1-2-96)
Stupt. Qurchu
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K020735
§ 882.1320 Cutaneous electrode.
(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).