K Number
K143171Device Name
FilmArray Blood Culture Identification (BCID) Panel for use with the FilmArray 2.0
Manufacturer
Biofire Diagnostics, LLC.
Date Cleared
2015-01-30
(87 days)
Product Code
PEN, OOI, PAM
Regulation Number
866.3365Why did this record match?
Applicant Name (Manufacturer) :
Biofire Diagnostics, LLC.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems. The FilmArray BCID Panel is capable of simultaneous detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants of antimicrobial resistance. The BCID assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system that demonstrate the presence of organisms as determined by Gram stain.
The following gram-positive bacteria, gram-negative bacteria, and yeast are identified using the FilmArray BCID Panel: Enterococci, Listeria monocytogenes, Staphylococci (including specific differentiation of Staphylococcus aureus), Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens), Haemophilus influenzae, Neisseria meningitidis (encapsulated), Pseudomonas aeruginosa, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, and Candida tropicalis.
The FilmArray BCID Panel also contains assays for the detection of genetic determinants of resistance to methicillin (mecA), vancomycin (vanA and vanB), and carbapenems (blakpr) to aid in the identification of potentially antimicrobial resistant organisms in positive blood culture samples. The antimicrobial resistance gene detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, and carbapenems exist.
FilmArray BCID is indicated as an aid in the diagnosis of specific agents of bacteremia and fungemia and results should be used in conjunction with other clinical and laboratory findings. Positive FilmArray results do not rule out co-infection with organisms not included in the FilmArray BCID Panel. FilmArray BCID is not intended to monitor treatment for bacteremia or fungemia.
Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing and epidemiological typing, to identify organisms in the blood culture that are not detected by the FilmArray BCID Panel, and for species determination of some Staphylococci, Enterococci, Streptococci, and Enterobacteriaceae that are not specifically identified by the FilmArray BCID Panel assays.
Device Description
The FilmArray Blood Culture Identification (BCID) Panel is a multiplex nucleic acid test designed to be used with a FilmArray system. The FilmArray BCID pouch contains freeze-dried reagents to perform nucleic acid purification and nested, multiplex PCR with DNA melt analysis. The FilmArray Blood Culture Identification (BCID) Panel simultaneously tests a single positive blood culture sample to provide results for 24 different organisms and organism groups that cause bloodstream infections and three genetic markers that are known to confer antimicrobial resistance (see Table 1).
A test is initiated by loading Hydration Solution and a positive blood culture sample mixed with the provided Sample Buffer into the FilmArray BCID pouch. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and sample Buffer Mix rehydrates the reagents. After the pouch is prepared, the software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.
The instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.
Nucleic acid extraction occurs within the FilmArray pouch using mechanical lysis and standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, a nested multiplex PCR is executed in two stages. During the first stage, a single, large volume, highly multiplexed PCR reaction which includes all primers of the outer primer sets, is performed. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green" Plus, BioFire Defense, LLC). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 21th stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the array captures fluorescent images of the PCR reactions and software interprets the data.
The software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.