Search Results
Found 2 results
510(k) Data Aggregation
(289 days)
Bio Cetec Co., Ltd.
The Bio-TackS Orthodontic Mini Implant is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.
Fixed on jawbone the Bio-TackS Orthodontic Mini Implant is a screw-shaped product used as a fixed point for orthodontic treatment and it is applied in order to secure sufficient physical and mechanical fixing source for treatment.
The provided text describes the Bio-TackS Orthodontic Mini Implant and its substantial equivalence to a predicate device, focusing on non-clinical performance and biocompatibility. It explicitly states that clinical testing was not required (Section 9, page 8). Therefore, the information requested about acceptance criteria and studies proving device performance with human data (e.g., sample size for test set, experts for ground truth, MRMC study, standalone performance, ground truth type for test/training sets) is not available within this document.
The document primarily focuses on demonstrating substantial equivalence through bench testing (non-clinical performance tests) and biocompatibility testing.
Here's the available information relevant to the acceptance criteria and the study that proves the device meets them, based on non-clinical data:
1. A table of acceptance criteria and the reported device performance:
The document doesn't explicitly state quantitative acceptance criteria for the "performance tests" in a table format. Instead, it describes performing tests according to a standard (ASTM F543) and comparing the subject device's results to the predicate device, concluding "similar performance."
Reported Device Performance (from Section 8, page 6-7):
Test | Acceptance Criteria (Implicit: Similar to Predicate) | Reported Device Performance |
---|---|---|
Sterilization and Shelf Life | Sterilization validation per ISO 11137-1, 11137-2, 11137-3 (VDmax25 method). Shelf life demonstrated by accelerated aging with evaluations for seal peeling (ASTM F88), dye penetration (ASTM F1929), and sterility testing. | Sterilization validation performed (ISO 11137-1, -2, -3, VDmax25 method). Shelf-life demonstrated by accelerated aging with seal peeling, dye penetration, and sterility testing. |
Insertion-Removal Torque Test | Mechanical stability similar to predicate device (evaluated at 1440 degrees at 1 rpm). | Both samples (BioTackS BSH1810S and Osstem OSSH1810) demonstrated "almost equivalent torque values" when inserted and removed. |
Rotational Fracture Torque Strength Test | Material yield strength similar to predicate device. | "The tapered implant fracture occurred in the same tread section." Both implants made from the same raw material (titanium alloy-ELI grade 23), indicating "material yield strengths of both products are the same." All screw-head types tested, results similar. |
Axial Pull-Out Strength Testing | Effectiveness of mechanical design and elimination similar to predicate device. | "The test data demonstrated similar results due to the similar pitch tread thickness of the samples" (BioTackS BSH1406S and Osstem OSSH1406). |
Biocompatibility - Cytotoxicity | Non-cytotoxic (per ISO 10993-5) | Non-cytotoxic |
Biocompatibility - Acute Systemic Toxicity | Non-acute systemic toxicity (per ISO 10993-11) | Non-acute systemic toxicity |
Biocompatibility - Intracutaneous (Intradermal) Reactivity | No irritation (per ISO 10993-10) | None Irritation |
Biocompatibility - Local Lymph Node Assay (LLNA) | No hypersensitivity (per ISO 10993-10) | Do not show any hypersensitivity |
2. Sample size used for the test set and the data provenance:
- Bench Tests (Performance Tests): For each specific test (Insertion-Removal torque, Rotational fracture torque, Axial pull-out), the document mentions testing specific models of the Bio-TackS and Osstem implants (e.g., BSH1810S, OSSH1810 for insertion-removal torque). It does not specify the exact number of samples tested for each condition but implies at least one sample of the "worst-case" or specific diameter/length combinations.
- Biocompatibility Tests: Sample sizes are not explicitly stated for these tests (e.g., number of cells for cytotoxicity, number of animals for toxicity/reactivity tests).
- Data Provenance: The tests are explicitly non-clinical bench and laboratory tests. No country of origin for the data is specified, but the applicant (BIO CETEC CO., LTD.) is from the Republic of Korea.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable as no clinical test set requiring human expert ground truth was conducted. The "ground truth" for the non-clinical tests is based on objective measurements against established standards (e.g., ASTM F543, ISO 11137, ISO 10993 series).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable, as no clinical test set requiring adjudication was conducted.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable, as no clinical study (MRMC or otherwise) involving human readers or AI assistance was conducted.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable, as this device (an orthodontic mini implant) is a physical medical device, not an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the non-clinical performance tests, the "ground truth" is measured physical and mechanical properties against established engineering and material science standards (ASTM F543), and for biocompatibility, it is biological response as assessed by ISO 10993 standards and laboratory assays.
8. The sample size for the training set:
Not applicable. This is a physical device, and the evaluation did not involve machine learning or AI models with training sets.
9. How the ground truth for the training set was established:
Not applicable (as above).
Ask a specific question about this device
(161 days)
BIO CETEC CO., LTD.
S-Line™ orthodontic ceramic bracket is intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position
S-Line™ orthodontic ceramic bracket is used to treat malocclusion, the abnormal occlusion of the upper and the lower teeth. It is intended to be applied on the surface of teeth to restore dental esthetics and functionality and it is designed to be used jointly with orthodontic wire.
The applicant, BIO CETEC CO., LTD., submitted a 510(k) premarket notification for their device, S-Line™ Orthodontic Ceramic Bracket, claiming substantial equivalence to the C-Line™ Orthodontic Ceramic Bracket (K163467) as the primary predicate device and DAMON 4Clear (K081415) as a reference device. Transbond™ XT (K073697) was also used as a bonding agent for performance testing.
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state formal "acceptance criteria" with numerical thresholds for performance. Instead, it describes performance tests conducted to demonstrate substantial equivalence by comparing the S-Line™ device to the reference device (DAMON 4Clear) and the primary predicate device (C-Line™). The reported "performance" is that the S-Line™ device's test results demonstrate substantial equivalence.
Performance Test | Objective | Comparison Against | Reported Performance of S-Line™ |
---|---|---|---|
Wire Slot Torque test | Demonstrate stability to withstand torque force from wire. | Reference device (DAMON 4Clear) (implicit from context) | Substantially equivalent |
Shear Bonding test | Demonstrate bond strength of brackets. | Reference device (DAMON 4Clear) (implicit from context) | Substantially equivalent |
Bracket Removal Test | Demonstrate stability of brackets' de-bonding performance from the enamel surface with a plier. | Reference device (DAMON 4Clear) (implicit from context) | Substantially equivalent |
Wire Drag Test | Measure friction between wire and bracket slot. | Reference device (DAMON 4Clear) (implicit from context) | Substantially equivalent |
Door Pull-Out Test | Measure tensile force at the moment of fracture from the orthodontic wire. | Reference device (DAMON 4Clear) (implicit from context) | Substantially equivalent |
Adhesive Strength Bonding Test | Study bonding of an adhesive to tooth structure or a bracket. | Not explicitly stated what it was compared against, likely also reference device. | Substantially equivalent |
Biocompatibility Tests | ISO 10993-1, 10993-5, 10993-10 standards | ||
Cytotoxicity | Assess for toxic effects on cells. | ISO 10993-5 | Non-cytotoxic |
Mucosal Irritation | Assess for irritation to mucosal membranes. | ISO 10993-10 | None Irritation |
Skin Sensitization | Assess for hypersensitivity reactions. | ISO 10993-10 | Do not show any hypersensitivity |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample sizes used for the non-clinical performance tests. It also does not specify the data provenance (e.g., country of origin, retrospective/prospective). The tests were conducted internally by the manufacturer or a contracted lab to meet ISO standards.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
This information is not applicable as the study is a non-clinical performance study of a device (orthodontic bracket) and does not involve human subjects or expert assessment for ground truth in the context of clinical outcomes or imaging interpretation. Ground truth for these tests would be derived from the physical properties and measurements according to ISO standards.
4. Adjudication Method for the Test Set:
This information is not applicable for non-clinical performance testing. Adjudication methods are typically employed in clinical studies involving human readers/investigators.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
An MRMC comparative effectiveness study was not done. The study described is a non-clinical performance study of a medical device, not a study evaluating human reader performance with or without AI assistance.
6. Standalone Performance:
Standalone performance was done in the context of the device's physical and mechanical properties. The non-clinical performance tests (wire slot torque, shear bond, bracket removal, wire drag, door pull-out, adhesive strength, and biocompatibility) evaluate the device itself without human-in-the-loop interaction.
7. Type of Ground Truth Used:
The ground truth for the non-clinical performance tests was based on physical and mechanical measurements according to specified ISO standards. For example, bond strength would be a measured force, and biocompatibility would be assessed against the criteria of ISO 10993.
8. Sample Size for the Training Set:
This information is not applicable. The S-Line™ Orthodontic Ceramic Bracket is a physical medical device, not an AI/machine learning algorithm, and therefore does not have a "training set."
9. How the Ground Truth for the Training Set Was Established:
This information is not applicable for the same reason as above.
Ask a specific question about this device
Page 1 of 1