K Number
K182193

Validate with FDA (Live)

Device Name
S-Line
Date Cleared
2019-01-22

(161 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

S-Line™ orthodontic ceramic bracket is intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position

Device Description

S-Line™ orthodontic ceramic bracket is used to treat malocclusion, the abnormal occlusion of the upper and the lower teeth. It is intended to be applied on the surface of teeth to restore dental esthetics and functionality and it is designed to be used jointly with orthodontic wire.

AI/ML Overview

The applicant, BIO CETEC CO., LTD., submitted a 510(k) premarket notification for their device, S-Line™ Orthodontic Ceramic Bracket, claiming substantial equivalence to the C-Line™ Orthodontic Ceramic Bracket (K163467) as the primary predicate device and DAMON 4Clear (K081415) as a reference device. Transbond™ XT (K073697) was also used as a bonding agent for performance testing.

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state formal "acceptance criteria" with numerical thresholds for performance. Instead, it describes performance tests conducted to demonstrate substantial equivalence by comparing the S-Line™ device to the reference device (DAMON 4Clear) and the primary predicate device (C-Line™). The reported "performance" is that the S-Line™ device's test results demonstrate substantial equivalence.

Performance TestObjectiveComparison AgainstReported Performance of S-Line™
Wire Slot Torque testDemonstrate stability to withstand torque force from wire.Reference device (DAMON 4Clear) (implicit from context)Substantially equivalent
Shear Bonding testDemonstrate bond strength of brackets.Reference device (DAMON 4Clear) (implicit from context)Substantially equivalent
Bracket Removal TestDemonstrate stability of brackets' de-bonding performance from the enamel surface with a plier.Reference device (DAMON 4Clear) (implicit from context)Substantially equivalent
Wire Drag TestMeasure friction between wire and bracket slot.Reference device (DAMON 4Clear) (implicit from context)Substantially equivalent
Door Pull-Out TestMeasure tensile force at the moment of fracture from the orthodontic wire.Reference device (DAMON 4Clear) (implicit from context)Substantially equivalent
Adhesive Strength Bonding TestStudy bonding of an adhesive to tooth structure or a bracket.Not explicitly stated what it was compared against, likely also reference device.Substantially equivalent
Biocompatibility TestsISO 10993-1, 10993-5, 10993-10 standards
CytotoxicityAssess for toxic effects on cells.ISO 10993-5Non-cytotoxic
Mucosal IrritationAssess for irritation to mucosal membranes.ISO 10993-10None Irritation
Skin SensitizationAssess for hypersensitivity reactions.ISO 10993-10Do not show any hypersensitivity

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes used for the non-clinical performance tests. It also does not specify the data provenance (e.g., country of origin, retrospective/prospective). The tests were conducted internally by the manufacturer or a contracted lab to meet ISO standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This information is not applicable as the study is a non-clinical performance study of a device (orthodontic bracket) and does not involve human subjects or expert assessment for ground truth in the context of clinical outcomes or imaging interpretation. Ground truth for these tests would be derived from the physical properties and measurements according to ISO standards.

4. Adjudication Method for the Test Set:

This information is not applicable for non-clinical performance testing. Adjudication methods are typically employed in clinical studies involving human readers/investigators.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

An MRMC comparative effectiveness study was not done. The study described is a non-clinical performance study of a medical device, not a study evaluating human reader performance with or without AI assistance.

6. Standalone Performance:

Standalone performance was done in the context of the device's physical and mechanical properties. The non-clinical performance tests (wire slot torque, shear bond, bracket removal, wire drag, door pull-out, adhesive strength, and biocompatibility) evaluate the device itself without human-in-the-loop interaction.

7. Type of Ground Truth Used:

The ground truth for the non-clinical performance tests was based on physical and mechanical measurements according to specified ISO standards. For example, bond strength would be a measured force, and biocompatibility would be assessed against the criteria of ISO 10993.

8. Sample Size for the Training Set:

This information is not applicable. The S-Line™ Orthodontic Ceramic Bracket is a physical medical device, not an AI/machine learning algorithm, and therefore does not have a "training set."

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Bio Cetec Co., Ltd. % Dave Kim, MBA President MTech Group 8310 Buffalo Speedway Houston, Texas 77025

Re: K182193

Trade/Device Name: S-Line™ Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: October 17, 2018 Received: October 26, 2018

Dear Dave Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

January 22, 2019

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Mary S. Marv S. Runner -S3 Runner -S3 Date: 2019.01.22

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182193

Device Name S-Line™

Indications for Use (Describe)

S-Line™ orthodontic ceramic bracket is intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92

Date: January 22, 2019

1. 510K Applicant / Submitter:

BIO CETEC CO., LTD. RM 1001-1002 New T Castle, 108, Gasan digital 2-ro, Geumcheon-gu, Seoul, Republic of Korea Tel: +82-2-704-2876 Fax: +82-2-704-2877

2. Submission Contact Person

MTech Group. 8310 Buffalo Speedway, Houston, TX 77025 Mr. Dave Kim, MBA Phone: 713-467-2607 Fax: 713-583-8988 Email: davekim@mtech-inc.net

3. Device

Trade / Device Name:S-Line TM
Classification Name:Bracket, Ceramic, Orthodontic
Regulation Number:21CFR 872.5470
Regulation Name:Orthodontic Plastic Bracket
Regulatory Class:II
Product Code:NJM

4. Predicate Device :

Trade / Device Name:C-Line™ Orthodontic Ceramic Bracket
510(k) NumberK163467
Regulation Number:21 CFR 872.5470
Regulation Name:Orthodontic Plastic Bracket
Regulatory Class:II
Product Code:NJM

5. Reference Device 1:

Trade / Device Name:DAMON 4Clear by Ormco Corporation
510(k) Number:K081415
Regulation Number:21 CFR 872.5470
Regulation Name:Orthodontic Plastic Bracket

{4}------------------------------------------------

Regulatory Class:II
Product Code:NJM

Reference Device 2:

Trade/proprietary NameTransbond™ XT
510(k) NumberK073697
Regulation NameAdhesive, bracket and tooth conditioner, resin
Regulation Number21 CFR 872.3750
Regulatory Class:Class II
Product CodeDYH

6. Description:

S-Line™ orthodontic ceramic bracket is used to treat malocclusion, the abnormal occlusion of the upper and the lower teeth. It is intended to be applied on the surface of teeth to restore dental esthetics and functionality and it is designed to be used jointly with orthodontic wire.

7. Indications for Use

S-Line™ orthodontic ceramic bracket is intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

8. Substantial Equivalence Discussion:

[S-Line™ vs. C-Line™(K163467)]

S-Line™ Orthodontic Ceramic bracket is substantially equivalent to C-Line™(K163467). C-Line™(K163467) is also manufactured by BIO CETEC Corporation. The differences of two products include ligating method & bracket structure. The following comparison table is presented to demonstrate substantial equivalence.

Subject DevicePredicate 1SubstantialEquivalenceAnalysis
510(k) NumberK182193K163467-
Device NameS-Line™C-Line™-
Common NameOrthodontic Plastic bracketOrthodontic Ceramic bracket-
ManufacturerBIO CETEC CORPORATIONBIO CETEC CORPORATION-
Indication for UseS-Line™ orthodontic ceramicbracket is intended to bebonded to a tooth to applypressure to a tooth from aflexible orthodontic wire toalter its positionC-Line™ orthodonticceramic bracket is intended tobe bonded to a tooth to applypressure to a tooth from aflexible orthodontic wire toalter its position.SubstantialEquivalence
Material compositionof BracketPolycrystallineAlumina(100%)PolycrystallineAlumina(100%)SubstantialEquivalence
Material compositionof DoorPolycrystallineAlumina(100%)N/ADifferent--
Material compositionof snap ring & pin300 Series Stainless SteelN/ADifferent-
Material compositionof colorants forbracket placementorientationPolyvinylpyrrolidone(25wt%)Sweet whey powder(25wt%)TiO2(25wt%)Food dye(25wt%)Polyvinylpyrrolidone(25wt%)Sweet whey powder(25wt%)TiO2(25wt%)Food dye(25wt%)SubstantialEquivalence
TransparencyHalf-transparencyHalf-transparencySubstantialEquivalence
Bracket designMBT, ROTH designs with andwithout hook, conforming toISO 27020:2010 Dentistry –Brackets and tube for Use inOrthodonticsMBT, ROTH designs withand without hook,conforming to ISO27020:2010 Dentistry –Brackets and tube for Use inOrthodonticsSubstantialEquivalence
Self-ligatingmechanismYesN/ADifferent-
Design partsHook, Slot, Round home, baseDoor and markingHook, Slot, Round home,base and markingSubstantialEquivalence
Bracket In-out(mm)0.65 to 1.080.56 to 1.45SubstantialEquivalence
Bracket Torque(°)-22 to +17-22 to +17SubstantialEquivalence
BracketAngulation(°)0 to 110 to 11SubstantialEquivalence
Available slot sizes0.018 / 0.022 inch0.018 / 0.022 inchSubstantialEquivalence
Orientation markingYesYesSubstantialEquivalence
Sing useYesYesSubstantialEquivalence
Non-SterilePackagingYesYesSubstantialEquivalence
Target PopulationPatients in need of teethalignment correctionPatients in need of teethalignment correctionSubstantialEquivalence
Anatomical SiteTeethTeethSubstantialEquivalence
Location of UseUse only by professionalorthodontistsUse only by professionalorthodontistsSubstantialEquivalence
Bio-compatibilityAll user directly contactingmaterials are compliance withISO10993 requirementsAll user directly contactingmaterials are compliance withISO10993 requirementsSubstantialEquivalence
Candidate DeviceReference DeviceSubstantialEquivalenceAnalysis
510(k) NumberK182193K081415-
Device NameS-Line™DAMON 4Clear-
Common NameOrthodontic Ceramic bracketOrthodontic Ceramic bracket-
ManufacturerBIOCETEC CO., LTD.Ormco Corporation-
Indication forUseS-Line™ orthodontic ceramicbracket is intended to be bonded toa tooth to apply pressure to a toothfrom a flexible orthodontic wire toalter its position.DAMON 4Clear is a ceramicbracket system intended to aid inthe movement of patient teethduring orthodontic treatment.SubstantialEquivalence
Materialcomposition ofBracketPolycrystalline AluminaPolycrystalline AluminaSubstantialEquivalence
Materialcomposition ofDoorPolycrystalline AluminaPolycrystalline AluminaSubstantialEquivalence
Materialcomposition ofsnap ring & pin300 Series Stainless SteelDifferent
Materialcomposition ofcolorants forbracketplacementorientationPolyvinylpyrrolidone(25wt%)Sweet whey powder(25wt%)TiO2(25wt%)Food dye(25wt%)Different
TransparencyHalf-transparencyHalf-transparencySubstantialEquivalence
Bracket designMBT, ROTH designs with andwithout hook, conforming to ISO27020:2010 Dentistry - Bracketsand tube for Use in OrthodonticsMBT, ROTH, High Torque designswith and without hook, conformingto ISO 27020:2010 Dentistry -Brackets and tube for Use inOrthodonticsSubstantialEquivalence
Self-ligatingmechanismYesYesSubstantialEquivalence
Design partsHook, Slot, Round home, base andmarkingHook, Slot, Round home, base andmarkingSubstantialEquivalence
Bracket In-out(mm)0.65 to 1.080.51 to 1.14Different
BracketTorque(°)-22 to +17-11 to +22Different
BracketAngulation(°)0 to 11°2 to 9°Different
Available slotsizes0.018 / 0.022 inch0.018 / 0.022 inchSubstantialEquivalence
OrientationmarkingYesYesSubstantialEquivalence
Single useYesYesSubstantialEquivalence
Non-SterilePackagingYesYesSubstantialEquivalence
TargetPopulationPatients in need of teeth alignmentcorrectionPatients in need of teeth alignmentcorrectionSubstantialEquivalence
Anatomical SiteTeethTeethSubstantialEquivalence
Location of UseUse only by professionalorthodontistsUse only by professionalorthodontistsSubstantialEquivalence
Bio-compatibilityAll user directly contactingmaterials are compliance withISO10993 requirements.All user directly contactingmaterials are compliance withISO10993 requirements.SubstantialEquivalence

{5}------------------------------------------------

[S-Line™ vs. DAMON 4Clear (K081415)]

S-Line™ Orthodontic Ceramic bracket is a self ligating type, same as DAMON 4Clear (K081415), the reference device. The following comparison table is presented to demonstrate substantial equivalence.

{6}------------------------------------------------

{7}------------------------------------------------

S-Line™ Orthodontic Ceramic bracket does not have a new intended use. However, there are differences in some parameters (Material composition of door, Material composition of colorants for bracket placement orientation, Bracket Torque, Bracket Angulation) between S-Line 114 Orthodontic Ceramic bracket and DAMON 4Clear, the reference device.

Material compositions for door and snap ring pin of the subject device are Polycrystalline Alumina(100%) and 300 Series Stainless Steel, respectively. Materials used to S-Line™ are identical to C-Line™ (K163467), the primary predicate device.

Also the colorants for bracket placement orientation are the same as the primary predicate device, C-Line TM (K163467): Polyvinylpyrrolidone(25wt%). Sweet whey powder(25wt%), TiO2(25wt%), Food dye(25wt%).

S-Line ™ is a self ligating mechanism type which is the same as the reference device, DAMON 4Clear.

The bracket in-out, torque and angulation are identical to the primary predicate device whereas there are differences between S-Line™, the subject device and DAMON 4Clear, the reference device. The bracket in-out of the subject device is 0.65 to 1.08 mm whereas the range for the reference device is between 0.51 to 1.14 mm. The bracket torque for the subject device is -22 to +17° wider than the reference device's-11 to +22°. The bracket angulation for the subject device 2 to 11° whereas the reference device's range 2 to 9°.

The differences, however, do not impact the safety and effectiveness as demonstrated in the performance tests.

Transbond™ XT (K073697) is included as a reference device as it is the bonding agent used to demonstrate substantial equivalence in performance testing.

9. Performance Tests (Non-clinical)

Non-clinical performance tests were in accordance with:

{8}------------------------------------------------

  • -. ISO 27020:2010 Dentistry Brackets and tubes for use in Orthodontics
  • -. ISO 11405:2015 Dentistry Testing of adhesion to tooth structure
  • -. ISO 29022:2013 Dentistry Adhesive Notched edge sheer bond strength test.

The following tests for performance comparison between the subject and the reference device have been conducted; wire slot torque strength, shear bond strength, bracket removal test, wire slot drag strength, door pull-out strength and the adhesive strength bonding testing.

Wire Slot Torque test demonstrates stability to withstand the torque force from wire.

Shear Bonding test showed the bond strength of brackets.

The removal test with plier showed stability of brackets de-bonding performance from the enamel surface.

Wire Drag Test measured the friction between wire and bracket slot.

Door Pull-Out Test measured tensile force at the moment of the fracture from the orthodontic wire.

Also the adhesive strength bonding testing was conducted to study bonding of an adhesive to tooth structure or a bracket.

The result of the performance comparison test demonstrates that S-Line™ Orthodontic Ceramic bracket is substantially equivalent to the reference device.

Along with the above tests, biocompatibility testing accordance with ISO 10993-1 has been conducted for S-Line TM orthodontic ceramic bracket.

HumanContactPartTest ItemTest ReportNumberTest StandardTest Result
MucosalmembraneCytotoxicityISO 10993-5Non-cytotoxic
Mucosal IrritationCR587171211ISO 10993-10None Irritation
Skin SensitizationISO 10993-10Do not show anyhypersensitivity

The biocompatibility test results demonstrated that there is no new concern in the cytotoxicity, sensitization, irritation, genotoxicity and no subchronic toxicity.

10. Conclusions:

Based on the information provided in this premarket notification, BIO CETEC CO., LTD. concludes that the S-Line™ Orthodontic Ceramic bracket is substantially equivalent to the primary predicate device and the reference device as described herein.

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.