Search Results
Found 2 results
510(k) Data Aggregation
K Number
K060942Device Name
WATER PURIFICATION COMPONENTS FOR HEMODIALYSIS
Manufacturer
Date Cleared
2006-07-07
(92 days)
Product Code
Regulation Number
876.5665Why did this record match?
Applicant Name (Manufacturer) :
**BOB J. JOHNSON **& ASSOCIATES
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
These Water Purification Devices are intended to be used as components of a water purification system to remove organic and inorganic substances from water that is used to dilute dialysis concentrate to form dialysate and to produce purified water for other purposes such as dialyzer reprocessing and equipment rinse and disinfection.
Device Description
The BJJ&A water purification component for hemodialysis will be used with a hemodialysis system. The Water purffication system will remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate, bicarbonate, acetate and sterilant for dialyzer reprocessing.
The purified water will also be used in the equipment disinfection process, equipment rinsing and any other application the clinic/customer deems appropriate. This system will meet all the requirements put forth by the United States Food and Drug Administration (FDA) and the Association for the Advancement of Medical Instrumentation (AAMI) and any other laws that may apply.
The BJJ&A water purfication components, both activated carbon filtration and deionization, utilize no new water purification techniques.
Deionization tanks from both BJJ&A and AmeriWater are utilized to remove dissolved solids from the water. Both utilize mixed bed resin, consisting of anion and cation resin, to remove the charged particles from the water. Both utilize parts and materials which are "NSF" and/or FDA approved.
Activated carbon filtration is utilized by both BJJ&A and AmeriWater to filter out chlorine and chloramines from the water. BJJ&A uses carbon filter made from materials, which are FDA or NSF approved. Both companies use two carbon fitters in a series configuration. A minimum total contact time of 10 empty bed contact time minutes are incorporated into both designs as recommended by the FDA for chlorine and chloramines removal. Both BJJ&A and AmeriWater recommended the chlorine and chloramine levels be checked before each patient shift. Both companies recommend hard plumbed bypass piping of the carbon tanks not be allowed on either component. Both components will utilize an activated carbon with an lodine number of 900 or greater. BJJ&A always recommends using a dual carbon filter in series in every dialysis water system installed, including single patient systems.
Ask a Question
Ask a specific question about this device
K Number
K014169Device Name
WATER PURIFICATION COMPONENTS FOR HEMODIALYSIS
Manufacturer
Date Cleared
2002-02-06
(49 days)
Product Code
Regulation Number
876.5665Why did this record match?
Applicant Name (Manufacturer) :
**BOB J. JOHNSON **& ASSOC
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
Carbon Filters and Water Softeners are intended to be used as components of a water purification system to remove organic and inorganic substances from water that is used to dilute dialysis concentrate to form dialysate and to produce purified water for other purposes such as dialyzer reprocessing and equipment rinse and disinfection
Device Description
The BJJ&A water purification components utilize no new water purification techniques, carbon filtration and water softening. The predicate device system we are claiming SE to, utilizes the exact same water purification principles. The following is a comparison summary of the BJJ&A water treatment components and the predicate device.
BJJ&A and U.S. Filter both utilize a water softener made of materials, which are "NSF" and/or FDA approved. U.S. Filters' softener is installed with a single regeneration tank and electrically interlocked to shut down the RO unit operation during the softener regeneration cycle. BJJ&A softeners are installed in a single regeneration tank configuration such as USF or in a duplex-twin alternating fashion to provide softened water to the system 24 hours a day. While one regeneration tank is in a regeneration cycle the other regeneration tank is on line providing soft water to the system. This will prevent the possibility of softener regeneration or a salt dump into the water purification system during a patient shift. This also eliminates the need to shut down the RO unit for softener regeneration.
Carbon filtration is utilized by both BJJ&A and U.S. Filter to filter out chlorine and chloramines from the water. BJJ&A uses carbon filter made from materials, which are FDA or NSF approved. Both companies use two carbon filters in a series configuration with test ports installed after each filter. A minimum total contact time of 10 empty bed contact time minutes are incorporated into both designs as recommended by the FDA for chlorine and chloramine removal. Both BJJ&A and U.S. Filter recommended the chlorine and chloramine levels be checked before each patient shift. Hard plumbed bypass piping of the carbon tanks is not allowed on either component. Both components will utilize an activated carbon with an Iodine number of 900 or greater. Both BJJ&A and U.S. Filter uses in line carbon regeneration filters. U.S. Filter uses a single carbon filter in its single patient dialysis system. BIJ&A always uses a dual carbon filter in series in every dialysis water system installed, including single patient systems.
BJJ&A and U.S. Filter incorporate emergency equipment bypass lines and valving that are clearly labeled and tagged. Instructions for usage will be included with the water components and staff in-service will be provided to the clinic/customer by BJJ&A. The use of dead legged bypass piping will be avoided. Carbon filters will not have hard plumbed bypass lines. U.S. Filter emergency equipment bypass utilizes a dry bypass technique. This method uses breakable unions and isolation valves at the end of each bypass line so that stagnant water will not be present in the line in the event of an emergency bypass condition. BJJ&A emergency equipment bypass is a wt bypass method. In this method the bypass valve is installed with no more than one half pipe a diameter length in this methou the bypass valve to avoid water stagnation. The use of dead leg bypass piping is unacceptable.
BJJ&A water distribution piping will be configured with no dead legs longer than 6 pipe diameters. U.S. Filter installs sanitization/sanitant introduction valve(s) or introduction port must be incorporated to allow chemical injection of sanitants to allow sanitization of the RO must be incorporated to anow chemical injoined in connections or through the top of the storage system and storage and the the sites for sanitant introduction BJJ&A avoids having any additional deadleg bypass valving installed into the original piping system and this further avoids water stagnation.
Ask a Question