Search Filters

Search Results

Found 21 results

510(k) Data Aggregation

    K Number
    K240637
    Date Cleared
    2024-11-04

    (243 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Bionime Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RIGHTEST Blood Glucose Monitoring System Max Tel is intended to the quantitative measurement of glucose (sugar) in fresh capillary whole drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared.

    RIGHTEST Blood Glucose Monitoring System Max Tel is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to montor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    The RIGHTEST Blood Glucose Monitoring System Max Tel is comprised of the RIGHTEST Meter Max Tel and the RIGHTEST Blood Glucose Test Strip Max.

    Device Description

    RIGHTEST Blood glucose monitoring System Max Tel consists of the following devices: Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing Device and Sterile Lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.

    RIGHTEST Blood Glucose Meter Max Tel, when used with the RIGHTEST Blood Glucose Test Strips Max, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of RIGHTEST Blood Glucose Monitoring System Max Tel is verified by the RIGHTEST Control Solution GC700.

    The glucose measurement is achieved by using the amperometric detection meth test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with FAD-glucose dehydrogenase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed.

    AI/ML Overview

    The provided FDA 510(k) summary for the RIGHTEST Blood Glucose Monitoring System Max Tel focuses on demonstrating substantial equivalence to a predicate device, as opposed to providing detailed clinical study results typical of a de novo or PMA submission. Therefore, much of the requested information regarding a comprehensive study proving acceptance criteria for an AI/device for diagnostic purposes (e.g., number of experts, MRMC studies, ground truth establishment for a training set) is not directly present in this document because it is not an AI/Software as a Medical Device (SaMD) submission for a diagnostic algorithm.

    This document describes a glucose monitoring system, which is a medical device rather than an AI-powered diagnostic system that typically involves image analysis or complex algorithmic interpretations of patient data for diagnosis. The "Software Safety Analysis" refers to enabling LTE functionality and adjusting the measurement range, alongside cybersecurity considerations, not the performance of a diagnostic AI.

    However, I can extract the acceptance criteria and performance as described in the document for this specific device:

    Device: RIGHTEST Blood Glucose Monitoring System Max Tel
    Intended Use: Quantitative measurement of glucose (sugar) in fresh capillary whole blood samples for self-testing by people with diabetes at home, as an aid to monitor the effectiveness of diabetes control.


    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the "Discussion of Non-Clinical Tests Performed for Determination of Substantiability" (Section 8) and the "Comparison to Predicate Devices" (Section 7), the acceptance criteria are generally implied by the successful completion and compliance with relevant FDA guidelines for glucose monitoring systems. The performance is reported in terms of demonstrating compliance.

    Acceptance Criteria (Stated/Implied)Reported Device Performance
    Accuracy / Performance Verification:
    Compliance with FDA's accuracy guidelines for Over-the-Counter (OTC) Self-Monitoring Blood Glucose (SMBG) systems. (This is a primary performance metric for glucose meters, though specific numerical targets like ISO 15197 are not detailed in this summary, they are implicit for regulatory acceptance.)The Extreme Glucose Study: "A study conducted on glucose performance using both natural and modified blood samples. The results demonstrated compliance with the FDA's accuracy guidelines for Over-the-Counter (OTC) Self-Monitoring Blood Glucose (SMBG) systems."

    Overall Conclusion: "Results of performance evaluation of RIGHTEST Blood Glucose Monitoring System Max Tel that had no impacts to BGM measurement was conducted to support substantially equivalent to the predicate device..." |
    | Measurement Range: Correct display of "Hi" or "Lo" for out-of-range results. | Hi Lo Display: "The measurement range has been adjusted, and the system displayed a notification indicating 'Hi' or 'Lo'—for results that fall outside the established range." The specific numerical range is 20 - 600 mg/dL (1.1 - 33.3 mmol/L). |
    | Software Functionality and Safety:

    • Successful implementation and validation of LTE functionality.
    • Compliance with FCC testing.
    • Compliance with FDA's cybersecurity guidance. | Software Safety Analysis: "Software adjustments were made to enable LTE functionality and adjusted the measurement range. The LTE function was validated through both FCC compliance testing and laboratory testing. As LTE functionality introduced cybersecurity considerations, we ensured compliance with the FDA's guidance on the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices." |
      | Interference: Performance maintained in the presence of specified interferents. | Interference Data Points: Ascorbic Acid ≥ 3 mg/dL, Conjugated Bilirubin ≥ 30 mg/dL, Uric Acid ≥ 12 mg/dL, Xylose ≥ 8 mg/dL. (Implies performance within specification despite these levels, though the exact outcome of the testing is not described beyond listing the tested interferents) |
      | Other Functional Parameters: Measurement technology, sample type, minimum sample volume, test time, control solution compatibility, operating conditions, storage conditions, shelf life, reagent composition, power saving, coding, monitor, backlight, color, power supply, memory capacity, meter dimension, LCD display area, meter weight, data transmission. | All these parameters are listed as characteristics of the new device, implicitly meeting the predicate device's standards or being deemed acceptable (e.g., LTE network for data transmission is a new feature). |
      | General Acceptance: All laboratory studies met acceptance criteria. | "All laboratory studies that the acceptance criteria were met. Therefore, the performances from these laboratory studies were acceptable." |

    Regarding the other requested points (relevant for AI/SaMD):

    • 2. Sample sized used for the test set and the data provenance: Not specified in the provided document. The reference to "natural and modified blood samples" in "The Extreme Glucose Study" suggests lab-based testing, but no specific sample size or provenance is given.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is not an AI/diagnostic imaging device requiring expert ground truth for interpretation. Ground truth for a glucose meter is typically established by laboratory reference methods (e.g., YSI analyzer).
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This type of adjudication is usually for subjective interpretations by multiple human readers, not for a highly objective measurement device like a glucose meter.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI system assisting human readers.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device itself is a "standalone" measurement device. Its performance is measured directly against laboratory reference standards, but there is no "algorithm only" in the sense of an AI interpreting complex data that a human would usually interpret.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For glucose meters, the ground truth is typically a laboratory reference method (e.g., a YSI analyzer), rather than expert consensus or pathology, as the measurement is quantitative. This is implied by the nature of the device, although not explicitly stated as "YSI" in the document.
    • 8. The sample size for the training set: Not applicable. This device does not use machine learning with a distinct training set in the typical sense of an AI/ML algorithm. Its functionality is based on established electrochemical principles, not pattern recognition learned from a dataset.
    • 9. How the ground truth for the training set was established: Not applicable, for the same reason as point 8.

    In summary, the provided document is a 510(k) summary for a blood glucose monitoring system, emphasizing its substantial equivalence to a predicate device and compliance with general FDA guidelines for such devices. It does not contain the detailed study results and AI-specific ground truth methodologies that would be found in a submission for an AI-powered diagnostic device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231192
    Date Cleared
    2024-01-19

    (267 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Bionime Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RIGHTEST Blood Glucose Monitoring System Max Tel is intended to the quantitative measurement of glucose (sugar) in fresh capillary whole drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared.

    RIGHTEST Blood Glucose Monitoring System Max Tel is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to montor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    The RIGHTEST Blood Glucose Monitoring System Max Tel is comprised of the RIGHTEST Meter Max Tel and the RIGHTEST Blood Glucose Test Strip Max.

    Device Description

    RIGHTEST Blood glucose monitoring System Max Tel consists of the following devices: Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing Device and Sterile Lancets. The RIGHTEST Blood Glucose Test Strip Max is the same as Test Strip Max cleared in K173638.The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.

    RIGHTEST Blood Glucose Meter Max Tel, when used with the RIGHTEST Blood Glucose Test Strips Max, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of RIGHTEST Blood Glucose Monitoring System Max Tel is verified by the RIGHTEST Control Solution GC700.

    The glucose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with FAD-glucose dehydrogenase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed.

    AI/ML Overview

    The provided document pertains to the 510(k) premarket notification for the "RIGHTEST Blood Glucose Monitoring System Max Tel." It primarily focuses on demonstrating substantial equivalence to a predicate device through various non-clinical and clinical tests, particularly outlining the system's accuracy.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The document presents system accuracy results, which serve as the primary performance metrics for the clinical study. The acceptance criteria are implicitly based on the FDA SMBG OTC guidance 2020 (mentioned in the interference section but generally applicable to system accuracy for OTC glucose meters), which typically specifies percentages of results within certain accuracy ranges compared to a reference method.

    Table of Acceptance Criteria and Reported Device Performance:

    Performance Metric (Acceptance Criteria Implicitly from FDA Guidance)Reported Device Performance (Fingertip)Reported Device Performance (Palm)Reported Device Performance (Forearm)
    Accuracy within ±15%98.6% (365 out of 370 tests)97.2% (360 out of 370 tests)97.65% (361 out of 370 tests)
    Accuracy within ±20%100%100%100%
    Accuracy within ±10%333 out of 370 tests323 out of 370 tests316 out of 370 tests
    Accuracy within ±5%231 out of 370 tests211 out of 370 tests226 out of 370 tests

    Note: For glucose meters, the acceptance criteria often involve percentages of results within +/-15 mg/dL for glucose concentrations = 100 mg/dL. The document simplifies this to overall percentages within +/-15%, +/-10%, and +/-5% of a reference bias, and mentions 100% within +/-20%. This implies meeting the standard accuracy requirements for blood glucose meters.

    Study Details: User Performance Study

    1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 370 participants for the User Performance Study.
    • Data Provenance: The document does not explicitly state the country of origin or whether the study was retrospective or prospective. However, user performance studies for regulatory submissions are typically prospective clinical trials. Given the manufacturer (Bionime Corporation) is based in Taiwan and the regulatory consultant is in the US, it's possible the study was conducted in Taiwan or the US, or both.

    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • The document describes a "User Performance Study" where lay users measure glucose. The "System accuracy" section compares these results to a "reference bias" or "comparison method."
    • No information is provided about experts establishing ground truth in the way one might for an AI-powered diagnostic image analysis system requiring expert annotation. For a blood glucose monitoring system, the "ground truth" or reference method is typically established by laboratory-grade glucose analyzers, often using a method like hexokinase or glucose oxidase with a highly accurate spectrophotometer. These are standardized laboratory procedures, not dependent on expert interpretation. The document mentions "reference bias," further suggesting a comparison to a precise laboratory method.

    3. Adjudication Method for the Test Set:

    • Not applicable in the context of a blood glucose monitoring system's accuracy study. Adjudication typically refers to resolving discrepancies among human readers or between human readers and an AI output in diagnostic imaging studies. Here, the comparison is between the device's reading and a precise laboratory reference method.

    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC study was not done. This type of study is relevant for diagnostic imaging systems where multiple human readers assess cases with and without AI assistance to determine the AI's impact on human performance. For a blood glucose meter, the evaluation is direct device performance against a reference standard, not an improvement in human reader performance.

    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This concept isn't directly applicable in the same way it would be for an AI algorithm. The device itself (the meter and strips) is the "standalone" entity that produces a measurement. The "User Performance Study" assesses the "human-in-the-loop" aspect by having lay users operate the device and measure their own blood samples. The system accuracy results directly report the device's performance as used by humans.

    6. The Type of Ground Truth Used:

    • The ground truth (or reference method) for this blood glucose monitoring system study is implied to be highly accurate laboratory-based glucose measurements, against which the device's readings are compared. The term "reference bias" supports this. For blood glucose meters, this reference is typically a carefully calibrated laboratory instrument, not expert consensus or pathology, which are common for AI-based image analysis.

    7. The Sample Size for the Training Set:

    • Not applicable. This document describes the validation of a blood glucose monitoring system, not an AI or machine learning model that requires a "training set." The system's underlying technology is an electrochemical sensor, not a learned algorithm in the AI sense.

    8. How the Ground Truth for the Training Set was established:

    • Not applicable. As no training set for an AI model is mentioned, there's no ground truth establishment for a training set.

    In summary, the document details a traditional validation approach for a medical device (blood glucose monitor) focusing on its accuracy and performance under various conditions, including lay user operation. It does not involve the complex AI-specific testing methodologies (such as MRMC, training sets, or expert adjudication for ground truth) that would be pertinent to AI/ML-driven diagnostic devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221062
    Date Cleared
    2022-09-26

    (168 days)

    Product Code
    Regulation Number
    878.4850
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Bionime Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RIGHTEST Lancing Device GD500 is a reusable lancing device for the single user. It should be used with the proper sterile lancet for the capillary blood sampling. It could automatically inject the lancet into and retract it from the fingertip to obtain a capillary blood sample for glucose monitoring or other test that require one or two drops of blood. A depth adjustable cap allows the best depth of skin penetration for each individual user. Alternative site testing (palm or forearm) can be performed by installing the clear cap on the lancing device.

    GE Lancing Device is a reusable lancing device for the single used with the proper sterile lancet for the capillary blood sampling. It could automatically inject the lancet it from the fingertip to obtain a capillary blood sample for glucose monitoring or other test that require one or two drops of blood. A depth adjustable cap allows the best depth of skin penetration for each individual user. Alternative site testing (palm or forearm) can be performed by installing the clear cap on the lancing device.

    iGlucose Lancing Device is a reusable lancing device for the single user. It should be used with the proper sterile lancet for the capillary blood sampling. It could automatically inject the lancet into and retract it from the fingertip to obtain a capillary blood sample for glucose monitoring or other test that require one or two drops of blood. A depth adjustable cap allows the best depth of skin penetration for each individual user.

    Device Description

    RIGHTEST Lancing Device GD500/ GE Lancing Device/ iGlucose Lancing Device is a mechanical device holding and firing a single-use lancet linearly ahead to prick the skin to collect capillary whole blood from target sites. RIGHTEST Lancing Device GD500/ GE Lancing Device/ iGlucose Lancing Device can be adjusted for 7 levels of depth for a user collecting different amount of capillary blood.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the BIONIME Corporation's RIGHTEST Lancing Device GD500, GE Lancing Device, and iGlucose Lancing Device. It specifically focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study proving a device's performance against specific acceptance criteria for AI or diagnostic accuracy.

    Therefore, many of the requested elements for an AI/diagnostic accuracy study (such as MRMC studies, effect size of human reader improvement with AI, specific ground truth methods, sample sizes for training/test sets related to AI model development, etc.) are not applicable to this document, as it pertains to a lancing device, not an AI-powered diagnostic tool.

    However, I can extract the relevant information regarding the performance data and the basis for substantial equivalence for this medical device.

    Device Under Review: RIGHTEST Lancing Device GD500, GE Lancing Device, iGlucose Lancing Device (reusable lancing devices for capillary blood sampling)

    1. A table of acceptance criteria and the reported device performance:

      Since this is a lancing device and not a diagnostic or AI device, the "acceptance criteria" here relate to safety, functionality, and biocompatibility, rather than diagnostic performance metrics like sensitivity/specificity. The document states that performance data demonstrated safety and efficacy but does not provide quantitative "acceptance criteria" for each test with reported values. Instead, it lists the types of tests performed.

      Acceptance Criteria (Implied by regulatory requirements for lancing devices)Reported Device Performance (Summary from document)
      Biocompatibility (In vitro Cytotoxicity)C. ISO 10993-5 In vitro Cytotoxicity A. was completed.
      Biocompatibility (Skin Irritation)A. ISO 10993-23 Skin Irritation B. was completed.
      Biocompatibility (Skin Sensitization)B. ISO 10993-10 Skin Sensitization (Maximization Test) was completed.
      Functional PerformanceD. Functional tests were validated and completed.
      Cleaning and Disinfection EfficacyE. The cleaning and disinfections on the materials of device were evaluated and tested.

      The document concludes that the device is "substantially equivalent to predicate devices in regard to indications for use, design, and technology, without raising any safety and efficacy risks or concerns," implying that the performance tests met the necessary standards for these types of devices.

    2. Sample size used for the test set and the data provenance:

      This information is not provided in the document as it's not a study on diagnostic accuracy or AI performance. The tests mentioned (biocompatibility, functional tests) would have their own sample sizes based on standard testing protocols for medical devices. The data provenance would be from laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      This is not applicable. Ground truth establishment by experts is relevant for diagnostic accuracy studies, especially for image interpretation. For a lancing device, "ground truth" would relate to measurable physical properties or biological responses (e.g., cytotoxicity, successful blood draw) determined by validated test methods, not expert consensus on images.

    4. Adjudication method for the test set:

      This is not applicable as there is no expert adjudication process for the performance tests of a lancing device in this context.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      This is not applicable. This document is for a lancing device, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      This is not applicable. There is no AI algorithm in this device.

    7. The type of ground truth used:

      For this type of device, the "ground truth" is defined by standardized laboratory test results for biocompatibility (e.g., cell viability assays, skin patch tests) and engineering validation for functional performance (e.g., consistency of lancing depth, successful lancet injection/retraction). It is not pathology or outcomes data in the sense of a diagnostic device.

    8. The sample size for the training set:

      This is not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

      This is not applicable. No training set is used for this type of device.

    Summary of Substantial Equivalence (as reported in the document):

    The submission argues for substantial equivalence based on the following:

    • Same "intended use/indication(s) for use": To obtain capillary blood samples for glucose monitoring or other tests requiring small amounts of blood, for single user, with depth adjustment and alternative site testing capabilities.
    • Similar Technological Characteristics: Both the proposed and predicate devices are mechanical devices that hold and fire a single-use lancet linearly to prick the skin, automatically inject and retract the lancet, have a clear cap for alternative site testing, retract the lancet after use to prevent sharp injury, have mechanical loading and firing functions, and eject the used lancet without touching it.
    • Differences noted: The primary difference is the "depth of penetration." The proposed device offers 7 different depths (0.5 mm to 1.7 mm) compared to the predicate devices' 6 different depths or interchangeable contact tips. The submission states this difference "did not raise other concerns on safety and efficacy."
    • Performance Data: The provided performance data (biocompatibility and functional tests) demonstrated safety and efficacy consistent with the predicate devices and regulatory requirements.

    The document concludes that the RIGHTEST Lancing Device GD500, GE Lancing Device, and iGlucose Lancing Device are "substantially equivalent to the predicate devices On Call Lancing Device and the Genteel Lancing Device (K113332 and K153670)."

    Ask a Question

    Ask a specific question about this device

    K Number
    K190564
    Date Cleared
    2019-06-02

    (89 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Bionime Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rightest Blood Glucose Monitoring System GM700S consists of the Rightest Blood Glucose Monitoring Meter GM700S and the Rightest Blood Glucose Test Strips GS700. The Rightest Blood Glucose Monitoring System GM700S is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700S is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    The Rightest Blood Glucose Monitoring System GM700SB consists of the Rightest Blood Glucose Monitoring Meter GM700SB and the Rightest Blood Glucose Test Strips GS700. The Rightest Blood Glucose Monitoring System GM700SB is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700SB is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    Device Description

    Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System GM700SB, consist of the following devices: Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing device and Sterile lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation. Rightest Meter GM700S and Rightest Meter GM700SB (with Bluetooth), when used with the Rightest Test Strips GM700S, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System GM700SB (with Bluetooth) are verified by the Rightest Control Solution GC700.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Rightest Blood Glucose Monitoring System GM700S and GM700SB, based on the provided document:

    Acceptance Criteria and Device Performance

    The document summarizes the performance tests conducted, but does not explicitly state "acceptance criteria" in a separate section. Based on the provided performance characteristics and the conclusion of "substantial equivalence," the implicit acceptance criteria would be that the device performs comparably to the predicate device and meets established standards for blood glucose monitoring systems.

    Table of Acceptance Criteria (Implicit) and Reported Device Performance

    Feature/Performance MetricImplicit Acceptance Criteria (based on predicate/standards)Reported Device Performance (GM700S/GM700SB)
    Minimum Sample VolumeComparable to predicate (0.75 microliter)0.75 microliter
    Test TimeComparable to predicate (5 seconds)5 seconds
    Hematocrit RangeComparable to predicate (20-65%)20 - 60 %
    Memory CapacitySufficient for user needs (500 results)500 blood glucose test results with date and time
    Power SavingFunctional and user-friendlyTurns off automatically 2 minutes after last user action / Press main button for 3 seconds
    CodingAccurate and user-friendly (Auto coding)Auto coding
    Measurement TechnologyStandard and accurate (Dehydrogenase Electrochemical Sensor)Dehydrogenase Electrochemical Sensor
    Sample TypeSpecified (Fresh capillary whole blood)Fresh capillary whole blood from fingertips, palm, or forearm.
    Measuring RangeClinically relevant (20-600 mg/dL)20 - 600 mg/dL
    Operating TemperatureSuitable for intended use (10 ~ 40°C)10 ~ 40°C (50 ~104 °F)
    Operating Relative HumiditySuitable for intended use (10-90%)10~90%
    MonitorClear display (LCD)LCD display
    Meter Storage ConditionsSpecified (-10 ~ 60°C)-10 ~ 60°C (14~140 °F)
    Test Strip Storage ConditionsSpecified (4 ~ 30℃)4 ~ 30℃ (39~ 86 °F), 10 - 90% relative humidity
    Unit of Measurement DataFixed (mg/dL)Fixed on mg/dL
    Backlight(Predicate has Yes, Device has No, but not a critical difference for substantial equivalence)No
    Strip Shelf Life - Open Vial(4 months)4 months
    Control SolutionSpecified (Rightest Control Solution GC700)Rightest Control Solution GC700
    InterferenceAcceptable levels (specified compounds and concentrations)Ascorbic acid ≥ 3 mg/dL, Uric acid ≥ 10 mg/dL, Xylose ≥ 20 mg/dL
    Reagent CompositionSpecified and comparableFAD-Glucose dehydrogenase 12.4%, Potassium Ferricyanide 49.6%, Non-reactive Ingredients 38.0% (GM700S)
    FAD-Glucose dehydrogenase 12.1%, Potassium ferricyanide 48.5%, Non-reactive ingredients 39.4% (GM700SB)
    Wireless module(GM700S: No, GM700SB: Yes)GM700S: No, GM700SB: No (contradicts table which suggests GM700SB has Bluetooth – this part of the table is inconsistent; it likely implies GM700SB does have a wireless module)
    Power SupplySpecified (CR2032 batteries)One CR2032 battery (GM700S), Two CR2032 batteries (GM700SB)
    Battery LifeSufficient (1000/600 tests)1000 tests (GM700S), About 600 tests (GM700SB)
    User PerformanceLay users obtain accurate resultsStudy shows substantial equivalence

    Study Details

    The document describes one clinical study: a User Performance Study.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: A total of 354 patients participated.
      • Data Provenance: The document does not explicitly state the country of origin. It also does not specify if the study was retrospective or prospective, but "user performance study" typically implies a prospective collection of data from real users.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the user performance study. In blood glucose monitoring studies, ground truth is usually established by a laboratory reference method, not by expert consensus on the device's readings. The study's aim was to show that lay users could obtain accurate results, meaning the comparison would be against a known accurate measurement.
    3. Adjudication Method for the Test Set:

      • The document does not describe an adjudication method. For performance studies of blood glucose meters, the standard is typically comparison to a laboratory reference method, not an adjudication process among experts looking at the device results.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a blood glucose monitoring system, not an AI-assisted diagnostic tool that would involve human readers interpreting images or data with and without AI.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, performance evaluations for the device itself (what could be considered the "algorithm only" in a different context) were conducted. The document mentions "precision, linearity, interference, sample volume, hematocrit, and operated environment" evaluations, which are standalone tests of the device's technical performance characteristics. The user performance study then assessed the human-in-the-loop aspect.
    6. The type of ground truth used:

      • For the user performance study, the ground truth would have been established by a laboratory reference method for blood glucose measurement (e.g., a YSI analyzer), to which the device's readings are compared. The document doesn't explicitly name the reference method, but this is the standard for such studies.
    7. The sample size for the training set:

      • The document does not mention a training set sample size. Blood glucose meters do not typically involve "training sets" in the same way machine learning models do. Their calibration is usually established during manufacturing based on known glucose concentrations.
    8. How the ground truth for the training set was established:

      • N/A, as there is no explicitly mentioned "training set" for an algorithm in the machine learning sense. The device's calibration and performance parameters are established through rigorous engineering and testing processes using samples with known glucose concentrations, typically traced to a reference method, during development and manufacturing.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Bionime Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rightest Blood Glucose Monitoring System Max is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared.

    Rightest Blood Glucose Monitoring System Max is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    The Rightest Blood Glucose Monitoring System Max is comprised of the Rightest Blood Glucose Test Strip Max and the Rightest Meter Max.

    Rightest Blood Glucose Monitoring System Max Plus with Bluetooth function is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared.

    Rightest Blood Glucose Monitoring System Max Plus with Bluetooth function is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    The Rightest Blood Glucose Monitoring System Max Plus is comprised of the Rightest Blood Glucose Test Strip Max and the Rightest Meter Max Plus.

    GE Blood Glucose Monitoring System Max is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared.

    GE Blood Glucose Monitoring System Max is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    The GE Blood Glucose Monitoring System Max is comprised of the GE Blood Glucose Test Strip Max and the GE Meter Max.

    GE Blood Glucose Monitoring System Max Plus with Bluetooth function is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared.

    GE Blood Glucose Monitoring System Max Plus with Bluetooth function is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    The GE Blood Glucose Monitoring System Max Plus is comprised of the GE Blood Glucose Test Strip Max and the GE Meter Max Plus.

    Device Description

    For Over-the-Counter Setting:

    Rightest Blood Glucose Monitoring System Max and Rightest Blood Glucose Monitoring System Max Plus, consist of the following devices:

    Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, lancing device and sterile lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.

    Rightest Meter Max and Rightest Meter Max Plus, when used with the Rightest Test Strips Max, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of Rightest Blood Glucose Monitoring System Max and Rightest Blood Glucose Monitoring System Max Plus are verified by the Rightest Control Solution GC700.

    GE Blood Glucose Monitoring System Max and GE Blood Glucose Monitoring System Max Plus, consist of the following devices:

    Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, lancing device and sterile lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.

    GE Meter Max and GE Meter Max Plus, when used with the GE Test Strips Max, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of GE Blood glucose monitoring System Max and GE Blood glucose monitoring System Max Plus are verified by the Rightest Control Solution GC700.

    AI/ML Overview

    The provided text pertains to a 510(k) premarket notification for a blood glucose monitoring system, not an AI model. Therefore, direct answers to questions regarding acceptance criteria for an AI device, sample sizes for training/test sets, expert adjudication methods, MRMC studies, standalone AI performance, or ground truth establishment for AI are not applicable.

    However, I can extract information related to the performance of the blood glucose monitoring system and present it in a format that mirrors the requested table, as well as address the aspects of the study that were performed.

    Here's a breakdown of the information available and how it relates to your questions:

    Device: Rightest Blood Glucose Monitoring System Max, Rightest Blood Glucose Monitoring System Max Plus, GE Blood Glucose Monitoring System Max, GE Blood Glucose Monitoring System Max Plus. These are glucose test systems (NBW, Class II, 21 CFR 862.1345).

    Study Type: This is a premarket notification (510(k)) seeking substantial equivalence to a predicate device (Rightest Blood Glucose Monitoring System GM720, K140210). The studies described are for the performance of a blood glucose meter, not an AI algorithm.

    Information applicable to the prompt (adapted for a glucose meter):

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a tabulated format for distinct performance metrics but rather presents various technical specifications and study results to demonstrate substantial equivalence. The provided information focuses on the technical specifications and a lay user performance study.

    Acceptance Criteria (Implied/Standard for BGMs)Reported Device Performance (from "Comparison to Predicate Devices" and "Discussion of Non-Clinical Tests")
    Measurement TechnologyDehydrogenase Electrochemical Sensor
    Sample TypeFresh capillary whole blood
    Alternative Sample SiteFingertips, palm, forearm.
    Minimum Sample Volume0.75 microliter
    Test Time5 seconds
    Control Solution3 levels (Level 1, 2, and 4) Rightest Control Solution GC700
    Operating Conditions (Temperature)43 ~111 °F (6 ~ 44°C)
    Operating Conditions (Relative Humidity)10 ~ 90%
    Meter Storage Conditions14140 °F (-1060℃)
    Test Strip Shelf Life (After Opening)4 months
    Hematocrit Range10 - 70 % (New Device) vs. 20 - 65% (Predicate Device)
    Measuring Range10-600 mg/dL (New Device) vs. 20-600 mg/dL (Predicate Device)
    Test Strip Reagent CompositionFAD-Glucose dehydrogenase 12.4 %, Potassium Ferricyanide 49.6 %, Non-reactive Ingredients 38.0 % (New Device) vs. FAD-Glucose dehydrogenase 12.1%, Potassium ferricyanide 48.5%, Non-reactive ingredients 39.4% (Predicate Device)
    InterferenceDopamine HCl > 2.3 mg/dL, Gentisic Acid > 3.0 mg/dL, Glutathione reduced > 35 mg/dL, Hemoglobin > 10,000 mg/dL, Uric Acid > 10 mg/dL, Maltose > 1900 mg/dL, Xylose > 9.0 mg/dL (New Device) vs. Ascorbic acid > 3 mg/dL; Glutathione reduced >60 mg/dL; Uric Acid > 12 mg/dL (Predicate Device)
    Accuracy (Lay User Performance Study)"The study result shows substantial equivalence to comparison method." (Specific numerical performance metrics are not provided in this summary, but would typically be based on ISO 15197 accuracy requirements for blood glucose monitoring systems, e.g., ±15% or ±20% agreement with a reference method for certain glucose concentration ranges).
    Precision, Linearity, Sample Volume, etc.Evaluated to establish performance, functionality, and reliability, with results supporting substantial equivalence. (No specific numerical results for these evaluations are provided in the summary)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: A "User performance study with 357 participants" was performed.
    • Data Provenance: The document does not specify the country of origin. It indicates the manufacturer is Bionime Corporation in Taichung City, Taiwan, and the consultant is in San Juan Capistrano, CA, USA. The study type is for a medical device performance validation, which typically involves prospective data collection. It is implied to be prospective as it's a performance study with participants.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable as the document describes a study for a blood glucose meter, not an AI interpretation model. The "ground truth" for a blood glucose meter is typically established by comparing its readings to a laboratory reference method (e.g., a central lab analyzer's measurement of glucose in the same blood sample), not by expert consensus on image interpretation.

    4. Adjudication Method for the Test Set

    This section is not applicable for a blood glucose meter performance study. Adjudication methods are relevant for subjective interpretations, often in imaging or clinical diagnoses.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    This section is not applicable. An MRMC study is relevant for evaluating the impact of AI systems on human reader performance, typically in diagnostic imaging. This document describes the performance of a blood glucose meter itself.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This section is not applicable as the device is a blood glucose meter. Its performance is inherent to the device's measurement capabilities. The "Lay User Performance Study" evaluates the device's accuracy when used by its intended users (individuals with diabetes at home), which includes a human "user-in-the-loop" aspect, but this is about device usability and accuracy, not AI algorithm standalone performance.

    7. The Type of Ground Truth Used

    The "ground truth" for blood glucose measurements in such studies is typically comparison to a laboratory reference method. The document states that the "study result shows substantial equivalence to comparison method," implying that the device's readings were compared against a highly accurate laboratory standard.

    8. The Sample Size for the Training Set

    This section is not applicable. Blood glucose meters are not "trained" in the way AI models are. Their performance is based on their electrochemical design and calibration.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable as blood glucose meters do not have a "training set" or "ground truth" established for it in the context of machine learning. Their calibration and performance are validated through rigorous lab testing against reference standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173139
    Date Cleared
    2018-06-19

    (263 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Bionime Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rightest Blood Glucose Monitoring System Wiz is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared.

    Rightest Blood Glucose Monitoring System Wiz is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    Rightest Blood Glucose Test Strip Wiz is used with Rightest Blood Glucose Meter Wiz for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm.

    Rightest Blood Glucose Monitoring System Wiz Plus is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. It is intended to be used by a single person and should not be shared.

    Rightest Blood Glucose Monitoring System Wiz Plus is intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

    Rightest Blood Glucose Test Strip Wiz is used with Rightest Blood Glucose Meter Wiz Plus for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm.

    Device Description

    Rightest Blood Glucose Monitoring System Wiz and Wiz Plus are identical with the exception of Bluetooth function for Rightest Blood Glucose Monitoring System Wiz Plus, which allows for wireless information transfer.

    Rightest Blood Glucose Monitoring System Wiz and Wiz Plus consists of the following devices: Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing Device, and Sterile Lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Rightest Blood Glucose Monitoring System Wiz and Wiz Plus. It includes details about the device, its intended use, comparison to a predicate device, and summaries of non-clinical and clinical tests performed.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance (Implicit)

    The document heavily relies on demonstrating "substantial equivalence" to a predicate device (Rightest GM280 Blood Glucose Monitoring System, K170143). While explicit numerical acceptance criteria for accuracy are not provided in the summary sections, the discussion of non-clinical and clinical tests implies that the device performance must be comparable to or within acceptable limits relative to the predicate device and established standards for glucose monitoring systems. Key performance aspects tested, which would imply related acceptance criteria, include:

    Acceptance Criteria (Implied from Tests)Reported Device Performance (Summary Statements)
    Precision (Reproducibility & Repeatability)Verification and validation activities were performed... Evaluations included precision. Specific numerical results are not provided in this summary, but the conclusion states the device demonstrates substantial equivalence.
    Linearity (Accuracy across measurement range)Verification and validation activities were performed... Evaluations included linearity. Specific numerical results are not provided in this summary, but the conclusion states the device demonstrates substantial equivalence.
    Interference (Resistance to interfering substances)Verification and validation activities were performed... Evaluations included interference. Specific numerical results are not provided in this summary, but a table lists substances and their interference thresholds for both new and predicate devices, suggesting equivalence for: Ascorbic acid > 3 mg/dL, Bilirubin-conjugated >10 mg/dL (for new device), Uric Acid >11.8 mg/dL (for new device), Uric Acid > 12 mg/dL (for predicate).
    Sample Volume (Performance with specified minimum volume)Verification and validation activities were performed... Evaluations included sample volume. Device specs state 0.75 microliter minimum sample volume.
    Hematocrit (Performance across a range of hematocrit levels)Verification and validation activities were performed... Evaluations included hematocrit. Device specs state 20-60% hematocrit range.
    User Performance/Usability (Lay user accuracy)A User performance study was performed to demonstrate that lay users could obtain accurate results using the subject device. The study result shows substantial equivalence to predicate device used in finger, palm and forearm position. The study results demonstrate the usability of Rightest Blood Glucose Monitoring System Wiz and Wiz Plus.

    Study Details

    Based solely on the provided text, the information is limited, especially regarding detailed methodologies and numerical study outcomes beyond general statements of equivalence.

    1. A table of acceptance criteria and the reported device performance: See the table above. Explicit numerical acceptance criteria are not provided in this summary. The performance is summarized as demonstrating "substantial equivalence" to the predicate.

    2. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: Not explicitly stated for any of the non-clinical or clinical tests. For the User Performance Study, it mentions "capillary whole blood from fingertip, palm and forearm sample sites" but does not give a number of participants or samples.
      • Data Provenance: Not explicitly stated. Given the submitter's address (Taiwan), the studies were likely conducted in Taiwan, but this is not confirmed. It is a premarket notification, thus the data would be prospective for the purpose of this submission.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned in the provided text. The ground truth method for the "User Performance Study" is implied to be a comparison against the predicate device, but the details of the "true" glucose reference method (e.g., lab analyzer) are not provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not mentioned in the provided text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is for a Blood Glucose Monitoring System, which is an in-vitro diagnostic device, not an imaging AI device that involves human reader interpretation. No human reader involvement or AI assistance is described for this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • For the non-clinical tests (precision, linearity, interference, sample volume, hematocrit), these would inherently be standalone performance evaluations of the device's measurement capabilities.
      • The "User Performance Study" evaluates the accuracy when used by lay users, which is a human-in-the-loop performance, but the 'algorithm' (the device's internal measurement system) itself is performing the measurement.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For the non-clinical tests (precision, linearity, interference, sample volume, hematocrit), the ground truth would typically be established using a traceable, highly accurate laboratory reference method (e.g., a YSI analyzer for glucose). This is not explicitly stated but is standard for such evaluations.
      • For the User Performance Study, the ground truth is implied to be a reference method (likely a lab analyzer) against which both the new device and the predicate device's readings are compared for equivalence.
    8. The sample size for the training set: Not applicable. This is not an AI/ML device that requires a "training set" in the traditional sense of machine learning. It's a glucose monitoring system whose performance is established through empirical testing rather than a learned model.

    9. How the ground truth for the training set was established: Not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Bionime Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rightest Blood Glucose Monitoring System GM280B is intended to the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM280B is intended to be used by a single person and should not be shared.

    The Rightest Blood Glucose Monitoring System GM280B intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

    The Rightest Blood Glucose Test Strip GS280 is used with Rightest meter GM280B for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

    The GE Blood Glucose Monitoring System 182 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE Blood Glucose Monitoring System 182 is intended to be used by a single person and should not be shared.

    The GE Blood Glucose Monitoring System 182 intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE Blood Glucose Monitoring System 182 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

    The GE Blood Glucose Strip 180 is used with GE Blood Glucose Monitoring Meter 182 for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

    The Rightest Blood Glucose Monitoring System GM280 is intended to the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM280 is intended to be used by a single person and should not be shared.

    The Rightest Blood Glucose Monitoring System GM280 intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

    The Rightest Blood Glucose Test Strip GS280 is used with Rightest meter GM280 for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

    The GE Blood Glucose Monitoring System 180 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE Blood Glucose Monitoring System 180 is intended to be used by a single person and should not be shared.

    The GE Blood Glucose Monitoring System 180 intended for self- testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE Blood Glucose Monitoring System 180 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

    The GE Blood Glucose Strip 180 is used with GE Blood Glucose Monitoring Meter 180 for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

    Device Description

    The Rightest and GE Blood Glucose Monitoring System series consists of the following devices: Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, lancing device and sterile lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.

    The Rightest Blood Glucose Meter GM280 and Rightest Blood Glucose Meter GM280B, when used with the Rightest Blood Glucose Test Strips GS280, quantitatively measure glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System GM280 and Rightest Blood Glucose Monitoring System GM280B are verified by the Rightest Control Solution GC550.

    The GE Blood Glucose Meter 180 and GE Blood Glucose Meter 182, when used with the GE Blood Glucose Test Strips 180, quantitatively measure glucose in fresh capillary whole blood. The performance of the GE Blood Glucose Monitoring System 180 and GE Blood Glucose Monitoring System 182 are verified by the Rightest Control Solution GC550.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the GE Blood Glucose Monitoring System series.

    Here's an analysis based on the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The main criterion for a glucose monitoring system's accuracy is typically measured against a laboratory reference method. For this device, the acceptance criteria relates to the accuracy of the blood glucose measurements compared to the YSI 2300 Plus Glucose Analyzer.

    Acceptance Criteria and Reported Device Performance (System Accuracy Study)

    Criterion CategoryAcceptance Criterion (Implicit from success statement)Reported Device Performance (Implicit from success statement)
    System AccuracyThe accuracy of the Rightest Blood Glucose Monitoring System GM280 met the acceptance criteria (specific numerical criteria not detailed in the provided text, but commonly derived from ISO 15197 standards for blood glucose monitoring systems).The study results demonstrate that the accuracy of Rightest Blood Glucose Monitoring System GM280 met the acceptance criteria.
    User Accuracy & Ease of UseThe user accuracy and ease of use met the acceptance criteria (specific numerical criteria not detailed).The study results demonstrate that the user accuracy and ease of use (via participant questionnaire scoring) Rightest Blood Glucose Monitoring System GM280B met the acceptance criteria.

    Note: The exact numerical acceptance criteria for system accuracy (e.g., percentage of readings within a certain deviation from the reference) are not explicitly stated in the provided documents but are implied to have been met. For such devices, commonly, the ISO 15197 standard for blood glucose monitoring systems is referenced, which specifies accuracy requirements (e.g., 95% of results must be within ±15 mg/dL for glucose concentrations

    Ask a Question

    Ask a specific question about this device

    K Number
    K161790
    Date Cleared
    2017-05-02

    (307 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    BIONIME CORPORATION

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iGlucose Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The iGlucose Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The iGlucose Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iGlucose Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use.

    The iGlucose Blood Glucose Test Strips are for use with the iGlucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    Device Description

    The iGlucose Blood Glucose Monitoring System consists of the iGlucose meter, the iGlucose test strips and the Rightest Glucose Control Solutions (GC550 level 2 and 4 (previously cleared under K092052). Bionime is seeking FDA's clearance to market and commercially distribute in the United States the iGlucose Meter and its corresponding iGlucose Test Strips.

    The iGlucose meter is an electronic device that utilizes electrical characteristics technology. When used with its corresponding iGlucose Test Strips, it quantitatively measures glucose in a small drop of fresh capillary whole blood (minimum 0.75 µL). The blood drop is placed on the test strip where it interacts with reagents in the presence of glucose oxidase to produce an electrical current proportional to the amount of glucose in the sample. The oxidase electrochemical sensor of the meter measures the current and using meter software converts it to the corresponding glucose concentration. The glucose result is then displayed on the meter within 5 seconds.

    Other system components include:

    • Rightest lancing device labeled for "Single Patient Use only" as well as its . disposable sterile lancets.
    • Meter charger and cable
    • A clear cap for the lancing device
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the iGlucose Blood Glucose Monitoring System, extracted from the provided FDA 510(k) summary:

    iGlucose Blood Glucose Monitoring System: Acceptance Criteria and Study Details

    1. Acceptance Criteria and Reported Device Performance

    The document states that the iGlucose meter's performance falls within acceptance criteria, and specific metrics are provided from a lay-user trial comparing the iGlucose to a reference method (YSI 2300 Plus).

    Glucose Concentration RangeAcceptance Criteria (relative to YSI)Reported Device Performance
    **
    Ask a Question

    Ask a specific question about this device

    K Number
    K140210
    Date Cleared
    2014-08-22

    (206 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    BIONIME CORPORATION

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rightest Professional Blood Glucose Monitoring System GM720 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm and in venous and arterial whole blood. The Rightest Professional Blood Glucose Monitoring System GM720 is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with auto-disabling, single-use lancing devices. The Rightest Professional Blood Glucose Monitoring System GM720 should not be used for the diagnosis of, or screening for testing neonate blood samples. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

    The Rightest Professional Blood Glucose Test Strips GS720 are for use with the Rightest Professional Blood Glucose Meter GM720 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm and in venous and arterial whole blood samples.

    The Rightest Blood Glucose Monitoring System GM720 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM720 is intended to be used by a single person and should not be shared.

    The Rightest Blood Glucose Monitoring System GM720 is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Rightest Blood Glucose Monitoring System GM720 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

    The Rightest Blood Glucose Test Strips GS720 are for use with the Rightest Blood Glucose Meter GM720 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm.

    Device Description

    The Rightest Blood Glucose Monitoring System GM720 consists of the following devices: Rightest Blood Glucose Meter GM720, Rightest Blood Glucose Test Strip GS720, Rightest Control Solution GC700, lancing device and sterile lancets. The Rightest Blood Glucose Meter GM720, Rightest Blood Glucose Test Strips GS720, and Lancing Device are manufactured by BIONIME Corporation.

    The Rightest Professional Blood Glucose Monitoring System GM720 consists of the following devices: Rightest Professional Blood Glucose Meter GM720, Rightest Professional Blood Glucose Test Strip GS720, and Rightest Control Solution GC700. This system should only be used with auto-disabling, single-use lancing devices and sterile lancets. The Rightest Professional Blood Glucose Meter GM720 and Rightest Professional Blood Glucose Test Strips GS720 are manufactured by BIONIME Corporation.

    Biomine's Rightest™ GM720 system is an electronic device that utilizes the electrical characteristic technology for measuring the glucose level in human blood. A relatively small drop of blood is placed on a disposable test strip coated with FAD-glucose Dehydrogenase (FAD-GDH). Within five seconds, the level of blood glucose will be shown on the digital display screen.

    Rightest™ Blood Glucose Monitoring System GM720, with an auto-coding feature, requires only minimum of 0.75 microliter of blood for the testing, therefore it reduces the time and effort required for testing and improves the compliance of diabetic people to their testing regimens.

    AI/ML Overview

    Device Name: Bionime Rightest™ Blood Glucose Monitoring System GM720, Bionime Rightest™ Professional Blood Glucose Monitoring System GM720

    This response describes the acceptance criteria and the study that proves the device meets those criteria, based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state the numerical acceptance criteria for the system accuracy study in a clear, tabulated format. However, it states that "The accuracy study... demonstrates that the accuracy of Rightest Blood Glucose Monitoring System GM720 met the acceptance criteria." This implies that predefined criteria were used and successfully met.

    Based on the general context of blood glucose monitoring systems, acceptance criteria typically align with standards like ISO 15197 for accuracy. Common criteria are usually expressed as a percentage of measurements falling within a certain deviation from a reference method, particularly at different glucose concentration ranges. Without the specific numerical thresholds, it's impossible to create a precise table.

    However, the reported performance is implicitly that the device did meet these unstated acceptance criteria.

    Note: The document only states that the device "met the acceptance criteria" without providing the specific numerical targets for accuracy (e.g., % within ±15 mg/dL or ±15%). For a comprehensive answer, these numerical criteria would be needed.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • System Accuracy Study (Test Set):

      • Sample Size: 121 patients were recruited.
      • Data Provenance: Not specified in the provided text (e.g., country of origin). The study involved "whole blood (plasma equivalent) glucose values" and "plasma glucose values on a lab instrument," suggesting a clinical setting.
      • Retrospective/Prospective: Not explicitly stated, though the recruitment of "patients" for a study generally implies a prospective collection of data for the purpose of the study.
    • User Performance Study (Test Set):

      • Sample Size: 165 laypersons.
      • Data Provenance: Not specified (e.g., country of origin).
      • Retrospective/Prospective: Not explicitly stated, but recruiting "laypersons" to demonstrate they "could obtain accurate results using the subject device" strongly suggests a prospective study design.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document mentions "plasma glucose values on a lab instrument" as the comparator for the System Accuracy Study.

    • Number of Experts: Not explicitly stated. The ground truth would typically be established by laboratory technicians or medical professionals operating the reference laboratory instrument.
    • Qualifications of Experts: Not explicitly stated. However, personnel operating a "lab instrument" for plasma glucose measurements are expected to be trained and qualified in clinical laboratory procedures.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method for either the System Accuracy Study or the User Performance Study. The System Accuracy Study directly compared the device's readings to a lab instrument's plasma glucose values, implying a direct comparison rather than an adjudication process involving multiple interpretations of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study, comparing human readers with AI assistance versus without AI assistance, was not conducted or reported in the provided text. This device is a blood glucose monitoring system, which directly measures glucose levels and does not involve human interpretation or AI assistance in the reading of results in the way a medical imaging AI would.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone performance study was done. The "System Accuracy Study" evaluated the "Rightest Blood Glucose Monitoring System GM720" by comparing its glucose values directly with a reference lab instrument. This represents the standalone performance of the device itself, without human intervention in the result generation or interpretation beyond the standard operation of the device.

    7. Type of Ground Truth Used

    • For System Accuracy Study: The ground truth was established using plasma glucose values on a lab instrument. This is a highly accurate and standardized method typically used as a reference in clinical laboratories for glucose measurement.
    • For User Performance Study: The ground truth was implicitly the same (laboratory reference method), as the study aimed to demonstrate that lay users could obtain "accurate results" comparable to the performance already established by the system accuracy study.

    8. Sample Size for the Training Set

    The provided document does not mention a training set. This is typical for a medical device like a blood glucose meter, which relies on electrochemical measurement principles rather than machine learning algorithms that require extensive training data. The device's performance is based on its hardware and firmware design, not a learning algorithm developed using a "training set."

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned or applicable for this type of device, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133003
    Date Cleared
    2013-10-24

    (29 days)

    Product Code
    Regulation Number
    862.1660
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    BIONIME CORPORATION

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rightest Control Solutions GC300 are for use with the Rightest Blood Glucose Meter GM300 and the Rightest Blood Glucose Test Strip GS300 to check that the meter and test strips are working together properly and that the test is performing correctly.

    The Rightest Control Solutions GC100 are for use with the Rightest Blood Glucose Meter GM100 and the Rightest Blood Glucose Test Strip GS100 to check that the meter and test strips are working together properly and that the test is performing correctly.

    The Rightest Control Solutions GC110 are for use with the Rightest Blood Glucose Meter GM110 and the Rightest Blood Glucose Test Strip GS300 to check that the meter and test strips are working together properly and that the test is performing correctly.

    The Rightest Control Solutions GC310 are for use with the Rightest Blood Glucose Meter GM310 and the Rightest Blood Glucose Test Strip GS310 to check that the meter and test strips are working together properly and that the test is performing correctly.

    Device Description

    Rightest Control Solution is a single-level, aqueous liquid qlucose control solution containing a measured amount of qlucose that reacts with the test strips to provide a Quality Control Result. This result should fall within the expected range printed on the Test Strip vial, and will verify if the meter and the test strips are working properly. The Rightest Control Solution has a red color to help users see the solution while dispensing onto a test strip.

    Rightest Control Solution is a non-hazardous aqueous glucose control solution containing no human or animal derived materials.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Value Assignment Procedure:Met:
    - Data from repeated measurement evaluated by ANOVA analysis.- "Data from repeated measurement will first be evaluated by ANOVA analysis." (Implied met, as study proceeded to determine mean/CV).
    - Acceptable range of measurement: ±15% of Mean value.- "Only after passing the ANOVA analysis, the data from repeated measurement can proceed to determine mean/CV to establish an acceptable range. The acceptable range of measurement as ±15% of Mean value to correspond to the value assignment ranges currently provided by using Rightest test strips. The test result should fall within the expected range printed on the Test Strip vial, and will verify if the meter and the test strips are working properly." (Implied met, as the acceptable range was established and the conclusion states performance was demonstrated.)
    Closed Bottle Stability (Shelf-Life):Met:
    - 1.5 years (18 months) at 2-30°C.- "Stability characteristics were determined under unopened conditions in real time and accelerated stability studies to demonstrate an closed shelf-life of 1.5 years (18 months) at the recommended storage temperatures, ranging from 2-30°C (36-86°F)."
    Opened Bottle Stability (Open-Vial):Met:
    - 3 months (20 weeks) with less than 10% bias of week 0 values.- "Rightest Control Solution meets stability requirements in real-time stability study as demonstrated by less than 10% bias of week 0 values with an open-vial claim of 3 months (20 weeks)."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Value Assignment: Total of 30 measurements for each test strip type. This was derived from using three lots of control solutions and ten blood glucose meters.
    • Data Provenance: The document does not explicitly state the country of origin for the data for the value assignment study. However, the submitter's identification is "BIONIME CORPORATION NO 694, RENHUA ROAD, DALI DIST., TAICHUNG CITY, TAIWAN 412," suggesting the studies were likely conducted in Taiwan. The studies were prospective as per the description of the testing procedures (e.g., stability studies involving real-time measurements).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This device is a control solution for blood glucose meters, not a diagnostic device that requires human interpretation of images or complex data. Therefore, the concept of "experts establishing ground truth" in the traditional sense (e.g., radiologists) does not apply.

    Instead, the "ground truth" or reference values are established through a statistical process of repeated measurements using the meters and test strips designed to work with the control solution, and then defining an acceptable deviation (±15% of the mean). The expertise involved would be in the design and execution of these analytical studies and the interpretation of statistical results, rather than clinical consensus.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the "ground truth" for this type of device is established through an analytical and statistical process involving repeated measurements, ANOVA analysis, and establishing a mean and acceptable range (±15% of the mean). There is no mention of human adjudication for individual test results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret cases, and the AI's role is to assist or augment that interpretation. This submission is for a control solution, an in-vitro diagnostic (IVD) analytical reagent, which verifies the proper functioning of a blood glucose monitoring system, not for a device that involves human interpretation of medical cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device is a control solution, not an algorithm or AI. Therefore, the concept of a "standalone algorithm performance" study is not applicable. The performance evaluated is the ability of the control solution to produce results within a defined range when used with the specified blood glucose meters and test strips.

    7. The Type of Ground Truth Used

    The ground truth used for establishing the acceptable ranges of the control solution is based on an analytical reference method where:

    • Repeated measurements of the control solution are taken using multiple test strips and meters.
    • Statistical analysis (ANOVA) is applied to these measurements.
    • A mean value is determined, and an acceptable range is established as ±15% of that mean value.

    This is effectively an empirical, statistically derived ground truth based on the expected performance characteristics of the glucose monitoring system components (meters and test strips) when reacting with the control solution.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI, as the device is not an algorithm. However, if we interpret "training set" as the data used to establish the reference values or expected performance of the control solution:

    • For the value assignment procedure: The data used consisted of measurements from three lots of control solutions and ten different blood glucose meters for each test strip type. This provided 30 measurements for each test strip type. This data was used to establish the mean and acceptable ranges.
    • For stability studies: The document states "real time and accelerated stability studies" were conducted, implying ongoing monitoring over time, which would involve a dataset gathered over the stability period. Specific sample sizes for these continuous measurements are not provided beyond the general description.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, for the value assignment, the ground truth was established through a rigorous analytical characterization process:

    1. Repeated measurement: The control solution was repeatedly tested using a defined set of meters and test strips.
    2. ANOVA analysis: Statistical analysis (ANOVA) was performed to evaluate the data from these repeated measurements, likely to ensure consistency across lots and meters.
    3. Mean and acceptable range determination: After passing ANOVA, the data was used to calculate a mean value. This mean then formed the basis for the "ground truth" reference range, with an acceptable deviation defined as ±15% of the mean value. This range corresponds to existing assigned values for the Rightest test strips.

    This process ensures that the control solution's expected readings are consistent with the established performance specifications of the blood glucose monitoring system it is intended to verify.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3