K Number
K190564

Validate with FDA (Live)

Date Cleared
2019-06-02

(89 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rightest Blood Glucose Monitoring System GM700S consists of the Rightest Blood Glucose Monitoring Meter GM700S and the Rightest Blood Glucose Test Strips GS700. The Rightest Blood Glucose Monitoring System GM700S is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700S is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

The Rightest Blood Glucose Monitoring System GM700SB consists of the Rightest Blood Glucose Monitoring Meter GM700SB and the Rightest Blood Glucose Test Strips GS700. The Rightest Blood Glucose Monitoring System GM700SB is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700SB is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

Device Description

Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System GM700SB, consist of the following devices: Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing device and Sterile lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation. Rightest Meter GM700S and Rightest Meter GM700SB (with Bluetooth), when used with the Rightest Test Strips GM700S, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System GM700SB (with Bluetooth) are verified by the Rightest Control Solution GC700.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Rightest Blood Glucose Monitoring System GM700S and GM700SB, based on the provided document:

Acceptance Criteria and Device Performance

The document summarizes the performance tests conducted, but does not explicitly state "acceptance criteria" in a separate section. Based on the provided performance characteristics and the conclusion of "substantial equivalence," the implicit acceptance criteria would be that the device performs comparably to the predicate device and meets established standards for blood glucose monitoring systems.

Table of Acceptance Criteria (Implicit) and Reported Device Performance

Feature/Performance MetricImplicit Acceptance Criteria (based on predicate/standards)Reported Device Performance (GM700S/GM700SB)
Minimum Sample VolumeComparable to predicate (0.75 microliter)0.75 microliter
Test TimeComparable to predicate (5 seconds)5 seconds
Hematocrit RangeComparable to predicate (20-65%)20 - 60 %
Memory CapacitySufficient for user needs (500 results)500 blood glucose test results with date and time
Power SavingFunctional and user-friendlyTurns off automatically 2 minutes after last user action / Press main button for 3 seconds
CodingAccurate and user-friendly (Auto coding)Auto coding
Measurement TechnologyStandard and accurate (Dehydrogenase Electrochemical Sensor)Dehydrogenase Electrochemical Sensor
Sample TypeSpecified (Fresh capillary whole blood)Fresh capillary whole blood from fingertips, palm, or forearm.
Measuring RangeClinically relevant (20-600 mg/dL)20 - 600 mg/dL
Operating TemperatureSuitable for intended use (10 ~ 40°C)10 ~ 40°C (50 ~104 °F)
Operating Relative HumiditySuitable for intended use (10-90%)10~90%
MonitorClear display (LCD)LCD display
Meter Storage ConditionsSpecified (-10 ~ 60°C)-10 ~ 60°C (14~140 °F)
Test Strip Storage ConditionsSpecified (4 ~ 30℃)4 ~ 30℃ (39~ 86 °F), 10 - 90% relative humidity
Unit of Measurement DataFixed (mg/dL)Fixed on mg/dL
Backlight(Predicate has Yes, Device has No, but not a critical difference for substantial equivalence)No
Strip Shelf Life - Open Vial(4 months)4 months
Control SolutionSpecified (Rightest Control Solution GC700)Rightest Control Solution GC700
InterferenceAcceptable levels (specified compounds and concentrations)Ascorbic acid ≥ 3 mg/dL, Uric acid ≥ 10 mg/dL, Xylose ≥ 20 mg/dL
Reagent CompositionSpecified and comparableFAD-Glucose dehydrogenase 12.4%, Potassium Ferricyanide 49.6%, Non-reactive Ingredients 38.0% (GM700S)FAD-Glucose dehydrogenase 12.1%, Potassium ferricyanide 48.5%, Non-reactive ingredients 39.4% (GM700SB)
Wireless module(GM700S: No, GM700SB: Yes)GM700S: No, GM700SB: No (contradicts table which suggests GM700SB has Bluetooth – this part of the table is inconsistent; it likely implies GM700SB does have a wireless module)
Power SupplySpecified (CR2032 batteries)One CR2032 battery (GM700S), Two CR2032 batteries (GM700SB)
Battery LifeSufficient (1000/600 tests)1000 tests (GM700S), About 600 tests (GM700SB)
User PerformanceLay users obtain accurate resultsStudy shows substantial equivalence

Study Details

The document describes one clinical study: a User Performance Study.

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: A total of 354 patients participated.
    • Data Provenance: The document does not explicitly state the country of origin. It also does not specify if the study was retrospective or prospective, but "user performance study" typically implies a prospective collection of data from real users.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the user performance study. In blood glucose monitoring studies, ground truth is usually established by a laboratory reference method, not by expert consensus on the device's readings. The study's aim was to show that lay users could obtain accurate results, meaning the comparison would be against a known accurate measurement.
  3. Adjudication Method for the Test Set:

    • The document does not describe an adjudication method. For performance studies of blood glucose meters, the standard is typically comparison to a laboratory reference method, not an adjudication process among experts looking at the device results.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a blood glucose monitoring system, not an AI-assisted diagnostic tool that would involve human readers interpreting images or data with and without AI.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, performance evaluations for the device itself (what could be considered the "algorithm only" in a different context) were conducted. The document mentions "precision, linearity, interference, sample volume, hematocrit, and operated environment" evaluations, which are standalone tests of the device's technical performance characteristics. The user performance study then assessed the human-in-the-loop aspect.
  6. The type of ground truth used:

    • For the user performance study, the ground truth would have been established by a laboratory reference method for blood glucose measurement (e.g., a YSI analyzer), to which the device's readings are compared. The document doesn't explicitly name the reference method, but this is the standard for such studies.
  7. The sample size for the training set:

    • The document does not mention a training set sample size. Blood glucose meters do not typically involve "training sets" in the same way machine learning models do. Their calibration is usually established during manufacturing based on known glucose concentrations.
  8. How the ground truth for the training set was established:

    • N/A, as there is no explicitly mentioned "training set" for an algorithm in the machine learning sense. The device's calibration and performance parameters are established through rigorous engineering and testing processes using samples with known glucose concentrations, typically traced to a reference method, during development and manufacturing.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the words "Convert to PDF" in a simple, sans-serif font. The text is horizontally aligned and appears to be a title or heading. The background is a plain, light color, providing contrast for the dark text.

Image /page/0/Picture/2 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 2, 2019

Bionime Corporation c/o Feng-Yu Lee Dynamic Biotech Inc. dba IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212 San Juan Capistrano, CA 92675

Re: K190564

Trade/Device Name: Rightest Blood Glucose Monitoring System GM700S Rightest Blood Glucose Monitoring System GM700SB Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: February 22, 2019 Received: March 5, 2019

Dear Feng-Yu Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190564

Device Name

Rightest Blood Glucose Monitoring System GM700S

Indications for Use (Describe)

The Rightest Blood Glucose Monitoring System GM700S consists of the Rightest Blood Glucose Monitoring Meter GM700S and the Rightest Blood Glucose Test Strips GS700.

The Rightest Blood Glucose Monitoring System GM700S is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700S is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared.

The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K190564

Device Name

Rightest Blood Glucose Monitoring System GM700SB

Indications for Use (Describe)

The Rightest Blood Glucose Monitoring System GM700SB consists of the Rightest Blood Glucose Monitoring Meter GM700SB and the Rightest Blood Glucose Test Strips GS700.

The Rightest Blood Glucose Monitoring System GM700SB is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700SB is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared.

The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Bionime Corporation

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K190564

    1. Submitter's Identification:
      BIONIME CORPORRATION NO 100, Sec. 2, Daging St., South Dist., 40242 Taichung City, Taiwan Phone Number: 886-4-23692388 FAX Number: 886-4-22617568

c/o IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212 San Juan Capistrano, CA 92675 Contact Person: Feng-Yu Lee Phone Number: 1-949-218-0929 Fax Number: 1-949-218-0928

Date Summary Prepared: May 2nd, 2019

    1. Name of the Device: Rightest Blood Glucose Monitoring System GM700S Rightest Blood Glucose Monitoring System GM700SB
    1. Common or Usual Name: Blood Glucose Monitoring System
Product CodeClassificationRegulation SectionPanel
NBW; System, Test, BloodGlucose, Over-the-CounterClass II21 CFR 862.1345Clinical Chemistry75

4. Device Description:

Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System GM700SB, consist of the following devices:

Blood Glucose Meter, Blood Glucose Test Strip, Control Solution, Lancing device and Sterile lancets. The Blood Glucose Meter, Blood Glucose Test Strips, and Lancing Device are manufactured by BIONIME Corporation.

Rightest Meter GM700S and Rightest Meter GM700SB (with Bluetooth), when used with the Rightest Test Strips GM700S, quantitatively measure glucose in fresh whole blood samples from capillary. The performance of Rightest Blood Glucose Monitoring System GM700S and Rightest Blood Glucose Monitoring System

{5}------------------------------------------------

GM700SB (with Bluetooth) are verified by the Rightest Control Solution GC700.

Intended Use: న.

Rightest Blood Glucose Monitoring System 700S (I)

The Rightest Blood Glucose Monitoring System GM700S consists of the Rightest Blood Glucose Monitoring Meter GM700S and the Rightest Blood Glucose Test Strips GS700. The Rightest Blood Glucose Monitoring System GM700S is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700S is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared.

The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

(II) Rightest Blood Glucose Monitoring System 700SB

The Rightest Blood Glucose Monitoring System GM700SB consists of the Rightest Blood Glucose Monitoring Meter 700SB and the Rightest Blood Glucose Test Strips GS700. The Rightest Blood Glucose Monitoring System GM700SB is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The Rightest Blood Glucose Monitoring System GM700SB is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The systems should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

6. Predicate Device Information:

The subject devices are substantially equivalent to the brand of Rightest Blood Glucose Monitoring System noted below.

Name:Rightest Blood Glucose Monitoring System GM720
Device Company:Bionime Corporation
510(K) Number:K140210
    1. Comparison to Predicate Devices:

{6}------------------------------------------------

Bionime Corporation

Subject DevicesPredicate Device
SpecificationRightest GM700SRightest GM700SB(GM700S w/ Bluetooth)Rightest GM720
Indications for UseSameQuantitative measurement of glucosein fresh capillary whole blood samplesdrawn from the fingertip, palm andforearm as an aid to monitor theeffectiveness of diabetes control
Minimum SampleVolume0.75 microliter
Test Time5 seconds
Hematocrit20 - 60 %20 - 65 %
Memory Capacity500 blood glucose test results with date and time
Power SavingTurn off automatically 2 minutes after last user action / Press the main button for 3 seconds.
CodingAuto coding
MeasurementTechnologyDehydrogenase Electrochemical Sensor
SampleFresh capillary whole blood from fingertips, palm, or forearm.
Measuring Range20 - 600 mg/dL
OperatingTemperature10 ~ 40°C (50 ~104 °F)
Operating RelativeHumidity10~90%
MonitorLCD display
Meter StorageConditions-10 ~ 60°C (14~140 °F)
Test Strip StorageConditions4 ~ 30℃ (39~ 86 °F), 10 - 90% relative humidity
Unit of MeasurementDataFixed on mg/dL
BacklightNoYes
Strip Shelf Life -Open Vial4 months
Control SolutionRightest Control Solution GC700
InterferenceAscorbic acid ≥ 3 mg/dLUric acid ≥ 10 mg/dLXylose ≥ 20 mg/dLAscorbic acid (≥ 3 mg/dL),Dopamine (≥1.25 mg/dL),L-Dopa (≥2 mg/dL),Xylose (≥10 mg/dL)Uric Acid (≥16 mg/dL)
Reagent1. FAD-Glucose dehydrogenase 12.4 %2. Potassium Ferricyanide 49.6 %3. Non-reactive Ingredients 38.0 %1. FAD-Glucose dehydrogenase 12.1%2. Potassium ferricyanide 48.5%3. Non-reactive ingredients 39.4%
Wireless moduleNoNo
Power SupplyOne CR2032 batteryTwo CR2032 batteries
Battery Life1000 testsAbout 600 tests
Meter Dimension82mm45mm15.5mm71.5 mm *39.5 mm *14.0 mm
LCD display area34 mm*27.5mm25.02 mm *32.7 mm
Meter Weight$43.0\pm 5$ g with batteries$50 \pm 5$ g with batteries
ColorWhite/GrayBlack

{7}------------------------------------------------

Bionime Corporation

8. Technology Characteristics:

The Rightest Blood Glucose Monitoring System GM700S, GMS700SB are electronic device that uses electrochemical methodologies. The system quantitatively measures blood glucose levels using an amperometric method. The system employs flavin adenine dinucleotide-glucose dehydrogenase (GDH-FAD) enzyme chemistry. The electrons generated during this reaction are transferred from the blood to the electrodes. The magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter.

9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:

Verification and validation of test results were evaluated to establish the performance, functionality and reliability of the subject devices. The evaluation included precision, linearity, interference, sample volume, hematocrit, and operated environment.

10. Discussion of Clinical Tests Performed:

User Performance Study:

A User performance study was performed to demonstrate that lay users could obtain accurate results using the subject device. The study was performed using capillary whole blood from fingertip, palm and forearm sample sites. A total of 354 patients participated. The study result shows substantial equivalence to predicate device used in finger, palm and forearm position.

    1. Conclusions:
      Results of performance evaluation of the Rightest Blood Glucose Monitoring System subject devices demonstrate that the candidate devises are substantial equivalence to the predicate device, Rightest Blood Glucose Monitoring System GM720.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.