Search Results
Found 4 results
510(k) Data Aggregation
(88 days)
Aspen Glove Sdn. Bhd.
Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.
The subject device in 510(K) notification is a blue nitrile examination gloves powder free tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I). The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)
The provided document is a 510(k) Premarket Notification for "Blue Nitrile Examination Gloves Powder Free Tested for Use with Chemotherapy Drugs." This document clears a medical device (examination gloves) for market and is not an AI/ML device study. Therefore, most of the requested information about acceptance criteria and study design for AI/ML performance is not applicable within this context.
However, I can extract information related to the performance of these physical medical gloves as detailed in the provided document, specifically related to their physical properties and chemical resistance.
Here's the relevant information based on the document:
Study Purpose: To demonstrate that the "Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy Drugs" are as safe, effective, and perform as well as the legally marketed predicate device (Harbour Health Powder Free Nitrile Examination Glove, Blue (Tested for use with Chemotherapy Drugs) K210944).
Lack of AI/ML Study: This document does not describe a study involving an AI/ML device. It pertains to a physical medical device (examination gloves). Therefore, specific criteria and study details related to AI/ML device performance (such as sample size for test/training sets for AI, data provenance for AI, number of experts for AI ground truth, human reader improvement with AI assistance, standalone AI performance) are not present in this document and cannot be provided.
Acceptance Criteria and Device Performance (for Physical Device - Gloves):
The performance of the gloves was evaluated against ASTM standards for various physical properties and for resistance to chemotherapy drugs. The acceptance criteria and reported device performance are presented in the tables below.
Table 1: Physical Properties Acceptance Criteria and Device Performance
Test Method (Standard) | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D6319-2019 | To determine the length of the gloves | Min 220 mm for Size Small & Min 230 mm for all other sizes | Small: 245 mm; Medium: 246 mm; Large: 248 mm; X-Large: 248 mm |
ASTM D6319-2019 | To determine the width of the gloves | Small: 80+/-10 mm; Medium: 95+/-10 mm; Large: 110+/-10 mm; X-Large: 120+/-10 mm | Small: 80 mm; Medium: 92 mm; Large: 105 mm; X-Large: 115 mm |
ASTM D6319-2019 | To determine the thickness of the gloves | Palm 0.05 mm min; Finger 0.05 mm min for all sizes | Palm: 0.09 mm (all sizes); Finger: 0.15mm (all sizes) |
ASTM D6319-2019 | To Determine the physical properties - Tensile strength | Before Ageing: Tensile Strength 14Mpa Min for all sizes; After Ageing: Tensile Strength 14Mpa Min for all sizes | Before Ageing: >18.52 Mpa (all sizes); After Ageing: >17.54 Mpa (all sizes) |
ASTM D6319-2019 | To Determine the physical properties - Ultimate Elongation | Before Ageing: Ultimate Elongation 500% Min for all sizes; After Ageing: Ultimate Elongation 400% Min for all sizes | Before Ageing: >690% (all sizes); After Ageing: >665% (all sizes) |
ASTM D5151-2019 | To determine the holes in the gloves | AQL 2.5 | Gloves Passes AQL 1.5 |
ASTM D6124-06 (Reapproved 2017) | To determine the residual powder in the gloves | 2 Mg/Glove Max | (Material Mediated Pyrogenicity) |
Table 3: Chemotherapy Drug Permeation Acceptance Criteria and Device Performance
(Note: The "acceptance criteria" here is implied by the predicate device's performance for comparison, and the requirement to report the minimum breakthrough detection time. A longer time generally indicates better performance.)
Tested Chemotherapy Drug Name & Concentration | Predicate Device Minimum Breakthrough Detection Time | Proposed Device Minimum Breakthrough Detection Time |
---|---|---|
Carmustine (BCNU) (3.3 mg/ml) | 14.5 Minutes | 34.3 Minutes |
Carboplatin (10 mg/ml) | >240 Minutes | >240 Minutes |
Cisplatin (1 mg/ml) | >240 Minutes | >240 Minutes |
Cyclophosphamide (Cytoxan) (20 mg/ml) | >240 Minutes | >240 Minutes |
Dacarbazine (10.0 mg/ml) | >240 Minutes | >240 Minutes |
Doxorubicin HCl (2 mg/ml) | >240 Minutes | >240 Minutes |
Etoposide (20 mg/ml) | >240 Minutes | >240 Minutes |
Fluorouracil (50 mg/ml) | >240 Minutes | >240 Minutes |
Ifosfamide (50 mg/ml) | >240 Minutes | >240 Minutes |
Methotrexate (25 mg/ml) | >240 Minutes | >240 Minutes |
Mitomycin C (0.5 mg/ml) | >240 Minutes | >240 Minutes |
Mitoxantrone (2 mg/ml) | >240 Minutes | >240 Minutes |
Paclitaxel (6 mg/ml) | >240 Minutes | >240 Minutes |
Thiotepa (10 mg/ml) | 47.4 Minutes | 87.3 Minutes |
Vincristine Sulfate (1 mg/ml) | >240 Minutes | >240 Minutes |
Drugs tested by Predicate but NOT Proposed: | Busulfan (6mg/ml) (>240 min), Cytarabine (100 mg/ ml) (>240 min), Docetaxel (10 mg/ml) (>240 min), Mechlorethamine HCL (1 mg/ml) (>240 min) | Not Tested |
Note on warnings: | Warning: Do not use with Carmustine (BCNU) & Thiotepa |
2. Sample Size Used for the Test Set and Data Provenance:
- The document primarily refers to testing done in accordance with ASTM standards. These standards specify sample sizes for testing (e.g., for AQL testing for pinholes, or for physical properties). The exact number of samples for each specific test (e.g., how many gloves for tensile strength or how many for chemical permeation) is not explicitly stated in the summary, but it would be prescribed by the referenced ASTM standards (e.g., ASTM D6319-2019, ASTM D5151-2019, ASTM D6124-06, ASTM D6978-05).
- Data Provenance: The applicant for the device is Aspen Glove Sdn. Bhd., located in Malaysia. The testing would presumably have been conducted there or by a certified lab, but specific country of origin for the data (i.e., where the tests were physically performed) is not explicitly stated beyond the applicant's location. The testing would be prospective in nature as it evaluates the manufactured product.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
- This is not applicable as this is a physical medical device and not an AI/ML device requiring establishment of "ground truth" by clinical experts for image interpretation or similar tasks. The "ground truth" for the physical properties and chemical resistance is established by validated laboratory testing methods conforming to recognized international standards (ASTM, ISO).
4. Adjudication Method for the Test Set:
- Not applicable. This is not a study requiring adjudication of expert interpretations but rather laboratory testing against established physical and chemical standards.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:
- Not applicable. This is a physical medical device, not an AI/ML device that assists human readers.
6. Standalone (Algorithm Only) Performance:
- Not applicable. This is a physical medical device, not an algorithm.
7. Type of Ground Truth Used:
- For physical and biocompatibility properties: Ground truth is established by validated laboratory testing methods against recognized industry standards (ASTM, ISO) with numerical or pass/fail criteria.
- For chemotherapy drug permeation: Ground truth is established by laboratory testing per ASTM D6978-05 (2019), measuring the minimum breakthrough detection time.
8. Sample Size for the Training Set:
- Not applicable. This is not an AI/ML device requiring a training set.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable. This is not an AI/ML device.
Ask a specific question about this device
(131 days)
Aspen Glove Sdn. Bhd.
Latex Examination Powder Free Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-05 (Reapproved 2015), Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are natural in color (No color is added) and are powder free.
The document provided is a 510(k) premarket notification for "Latex Examination Powder Free Gloves." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, primarily through non-clinical performance data against established standards for medical gloves, rather than clinical studies using AI.
Therefore, the requested information regarding AI study details (sample sizes for test and training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth for AI, how ground truth for training was established) is not applicable to this submission, as it does not involve an AI/ML powered device.
However, I can extract the acceptance criteria and device performance based on the non-clinical tests described.
1. Table of Acceptance Criteria and Reported Device Performance
Test Method & Purpose | Acceptance Criteria (from Standards) | Reported Device Performance |
---|---|---|
Dimensions | ||
ASTM D3578-05 (Reapproved 2015) - Length | Min 230 mm for all sizes | X-Small: 236 mm, Small: 236 mm, Medium: 242 mm, Large: 248 mm, X-Large: 252 mm |
ASTM D3578-05 (Reapproved 2015) - Width | X-Small: 70+/-10 mm, Small: 80+/-10mm, Medium: 95+/-10 mm, Large: 111+/-10 mm, X-Large: 115+/-10 mm | X-Small: 76 mm, Small: 85 mm, Medium: 98 mm, Large: 106 mm, X-Large: 116 mm |
Physical Properties | ||
ASTM D3578-05 (Reapproved 2015) - Thickness | Palm: 0.08 mm min, Finger: 0.08 mm min for all sizes | All sizes (X-Small to X-Large): Palm 0.12 mm, Finger 0.14 mm |
ASTM D3578-05 (Reapproved 2015) - Tensile Strength (Before Ageing) | 18 Mpa Min for all sizes | X-Small: 22.08 Mpa, Small: 22.16 Mpa, Medium: 22.22 Mpa, Large: 22.30 Mpa, X-Large: 22.32 Mpa |
ASTM D3578-05 (Reapproved 2015) - Tensile Strength (After Ageing) | 14 Mpa Min for all sizes | X-Small: 18.50 Mpa, Small: 18.56 Mpa, Medium: 18.65 Mpa, Large: 18.74 Mpa, X-Large: 18.76 Mpa |
ASTM D3578-05 (Reapproved 2015) - Ultimate Elongation (Before Ageing) | 650% Min for all sizes | X-Small: 856%, Small: 868%, Medium: 874%, Large: 880%, X-Large: 882% |
ASTM D3578-05 (Reapproved 2015) - Ultimate Elongation (After Ageing) | 500% Min for all sizes | X-Small: 842%, Small: 850%, Medium: 855%, Large: 860%, X-Large: 862% |
ASTM D3578-05 (Reapproved 2015) - Stress at 500% Elongation (Before Ageing) | 5.5 Mpa Max for all sizes | X-Small: 5.1 Mpa, Small: 5.1 Mpa, Medium: 5.2 Mpa, Large: 5.2 Mpa, X-Large: 5.2 Mpa |
Barrier Integrity | ||
ASTM D5151-06 (Reapproved 2015) - Detection of Holes (Water Tight) | AQL 2.5 | Gloves Pass AQL 1.5 |
Other Properties | ||
ASTM D6124-06 (Reapproved 2017) - Residual Powder on Medical Gloves | 2 Mg/Glove Max | X-Small: 0.21 mg/glove, Small: 0.21 mg/glove, Medium: 0.22 mg/glove, Large: 0.22 mg/glove, X-Large: 0.22 mg/glove |
ASTM D5712-95 (Reapproved 2010) - Analysis of Protein in Natural Rubber | 200 µg/dm² Max for all sizes | All sizes (X-Small to X-Large): 32 µg/dm² |
ISO 10993-10:2010(E) - Primary Skin Irritation | Not an irritant | Under the condition of study, not an irritant |
ISO 10993-10:2010(E) - Dermal Sensitization | Not a sensitizer | Under the conditions of the study, not a sensitizer |
ISO 10993-5:2009(E) - In vitro cytotoxicity | Non-cytotoxic | Under the conditions of the study, non-cytotoxic |
ISO 10993-11:2017(E) / USP - Material mediated Pyrogenicity | Non-pyrogenic | Under the conditions of the study, non-pyrogenic |
USP 42 - Bacterial Endotoxin Test |
Ask a specific question about this device
(127 days)
Aspen Glove Sdn. Bhd.
A Sterile Latex Surgical Gloves Powder Free is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
The proposed device, Sterile Latex Surgical Gloves Powder Free is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, as per standard ASTM D3577 -09(2015), Standard Specification for Rubber Surgical Gloves. The classification is: Type I - gloves compounded primarily from natural rubber latex. The proposed device is Sterile Latex Surgical Gloves Powder Free, and variants of different sizes. All variants share the same natural color (No color is added). The proposed device is sterilized using Gamma method to achieve the Sterility Assurance Level (SAL) of 10 and place in a sterility maintained package to ensure a shelf life of 3 years.
The provided document is a 510(k) Premarket Notification for a medical device: "Sterile Latex Surgical Gloves powder free." This document focuses on demonstrating substantial equivalence to an existing predicate device, rather than proving the device meets clinical acceptance criteria for an AI/ML medical device. Therefore, many of the requested points related to AI/ML study design (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth for training set) are not applicable to this type of submission.
However, the document does contain information about performance testing against established standards, which serve as the acceptance criteria for this specific device.
Here's an analysis of the provided information, addressing the relevant points and noting those that are not applicable:
Device Name: Sterile Latex Surgical Gloves powder free
Submission Type: 510(k) Premarket Notification (for substantial equivalence)
1. A table of acceptance criteria and the reported device performance
The document provides extensive tables comparing the subject device's performance to the predicate device and against the specifications outlined in relevant ASTM standards. These standards define the "acceptance criteria" for the physical and chemical properties of surgical gloves.
Summary Table of Acceptance Criteria and Reported Performance (Selected Key Parameters)
Criteria Category | Performance Criterion Source (Standard/Section) | Acceptance Criteria | Reported Device Performance (K211478) | Proof of Meeting Criteria |
---|---|---|---|---|
Physical Dimensions | ASTM D3577-09 (Reapproved 2015) | Min 265mm (Length, all sizes) | 385mm - 396mm (All sizes) | Met |
ASTM D3577-09 (Reapproved 2015) | Various +/- 6mm (Width, per size) | 74mm - 116mm (All sizes) | Met | |
ASTM D3577-09 (Reapproved 2015) | Min 0.10mm (Finger Thickness, all sizes) | 0.20mm | Met | |
ASTM D3577-09 (Reapproved 2015) | Min 0.10mm (Palm Thickness, all sizes) | 0.15mm | Met | |
ASTM D3577-09 (Reapproved 2015) | Min 0.10mm (Cuff Thickness, all sizes) | 0.12mm | Met | |
Physical Properties | ASTM D3577-09 (Reapproved 2015) | 24 Mpa minimum (Tensile Strength, Before Aging) | 27.2 Mpa - 29.7 Mpa | Met |
ASTM D3577-09 (Reapproved 2015) | 18 Mpa minimum (Tensile Strength, After Aging) | 22.1 Mpa - 24.5 Mpa | Met | |
ASTM D3577-09 (Reapproved 2015) | 750% minimum (Ultimate Elongation, Before Aging) | 864% - 874% | Met | |
ASTM D3577-09 (Reapproved 2015) | 560% minimum (Ultimate Elongation, After Aging) | 727% - 734% | Met | |
ASTM D3577-09 (Reapproved 2015) | 5.5 MPa Max (Stress at 500% before ageing) | 5.1 Mpa (Average Value) | Met | |
Barrier Integrity | ASTM D3577-09 & ASTM D5151-06 | Max 1.5 AQL (Pinhole AQL, Before & After Aging) | 1.0 AQL (Gloves Passes) | Met |
Chemical Residues | ASTM D6124-06 | $\le$ 2 mg/Glove (Powder Residue) | 0.26 mg/glove - 0.62 mg/glove (All sizes) | Met |
ASTM D3577 & ASTM D5712-15 | $\le$ 50µg/dm² (Protein Content) | 38 µg/dm² | Met | |
Biocompatibility | ISO 10993-10 | Non-irritant | Non-irritant | Met |
ISO 10993-10 | Non-sensitizer | Non-sensitizer | Met | |
ISO 10993-5:2009 | Non-cytotoxic | Cytotoxic | NOT Met (Qualified) | |
ISO 10993-11:2017 | No systemic toxicity concern | No systemic toxicity concern | Met | |
ISO 10993-11:2017 / USP 41 | No material mediated pyrogenicity response | No material mediated pyrogenicity response | Met | |
USP 42 | NMT 20 EU/pair of gloves (Bacterial Endotoxin) |
Ask a specific question about this device
(127 days)
Aspen Glove Sdn. Bhd.
Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.
This document is a 510(k) summary for a medical device (Blue Nitrile Examination Glove Powder Free), demonstrating its substantial equivalence to a predicate device. This type of submission generally relies on non-clinical performance data rather than clinical studies for novel AI/ML devices. Therefore, a direct application of the typical "acceptance criteria" and "study that proves the device meets the acceptance criteria" for AI/ML performance (e.g., accuracy, sensitivity, specificity, MRMC studies) is not present here.
However, we can frame the information provided within the context of the requested questions by translating the device's technical specifications and performance data into a pseudo-"acceptance criteria" and "proof of meeting criteria."
Here's a breakdown based on the provided document:
1. A table of acceptance criteria and the reported device performance
For a physical device like a medical glove, "acceptance criteria" relate to its physical and biocompatibility properties, and "performance" refers to the test results demonstrating compliance with these standards.
Characteristic | Standard/Test Method | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Dimensions | ASTM D6319-10 (Reapproved 2015) | Length Min 230 mm; Width depends on size (e.g., Medium: 95+/-10mm) | Meets/Exceeds (e.g., Length Min 230mm, Medium Width 92mm) |
Physical Properties | ASTM D6319-10 (Reapproved 2015) | Before Aging: Tensile Strength min 14 Mpa, Ultimate Elongation Min 500% | |
After Aging: Tensile Strength min 14 Mpa, Ultimate Elongation Min 400% | Meets/Exceeds (e.g., Before Aging: 18.62Mpa, 694%; After Aging: 17.68Mpa, 668% for Medium) | ||
Thickness | ASTM D6319-10 (Reapproved 2015) | Palm min 0.05 mm, Finger min 0.05 mm | Meets/Exceeds (e.g., Palm 0.09mm, Finger 0.15mm for all sizes) |
Powder Residue | ASTM D6319-10 / ASTM D6124-06 | ≤2 mg/glove | Meets (e.g., 0.22 mg/glove for Medium) |
Water Tightness (Holes) | ASTM D5151-06 (Reapproved 2015) | Passes AQL-1.5 | Passes AQL 1.5 |
Biocompatibility: | |||
Primary Skin Irritation | ISO 10993-10:2010(E) | Not an irritant | Not an irritant |
Dermal Sensitization | ISO 10993-10:2010(E) | Not a sensitizer | Not a sensitizer |
In vitro cytotoxicity | ISO 10993-5:2009(E) | Non-cytotoxic | Cytotoxic (This is a discrepancy or a nuance to be explored within biocompatibility, as the acceptance criteria states "non cytotoxic" but result states "cytotoxic." Often, a cytotoxic result in vitro requires further in vivo investigation, but for gloves, the overall safety profile is considered.) |
Acute Systemic Toxicity | ISO 10993-11:2017(E) | Extracts do not pose a systemic toxicity concern | Extracts do not pose a systemic toxicity concern |
Material Mediated Pyrogenicity | ISO 10993-11:2017(E) / USP 41 | Non-pyrogenic | Non-pyrogenic |
Bacterial Endotoxin | USP 42 | NMT 20 EU/pair of gloves |
Ask a specific question about this device
Page 1 of 1