Search Results
Found 19 results
510(k) Data Aggregation
(193 days)
Andersen Sterilizers, Inc.
Ask a specific question about this device
(156 days)
Andersen Sterilizers, Inc.
The EOGas 4 Endo-SteriTest Rapid Readout Biological Indicator consists of a self-contained biological indicator inoculated with viable Bacillus atrophaeus bacterial spores that is placed in a dedicated biological indicator receptacle mounted on the purge probe in the sterilizer. It monitors the efficacy of the 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.
The EOGas 4 Endo-SteriTest Rapid Readout Biological Indicator (RRBI) consists of a single-use self-contained biological indicator (SCBI) placed in a reusable biological indicator (BI) receptacle. It is designed for monitoring the efficacy of the 6-hour gas exposure at 50°C in an EOCas 4 Ethylene Oxide Gas Sterilizer.
The Bionova BT110 Rapid Readout Biological Indicator (K191021) consists of a plastic vial that serves as the culture tube and a cap including a filter material port to allow ethylene oxide to enter the vial. The plastic vial contains Bacillus atrophaeus spores inoculated onto a paper carrier, and a glass ampoule containing culture medium and a pH indicator. There is a chemical indicator printed on the unit label of the SCBI to indicate EO exposure by changing color from brown/red to green.
Following manufacturer's instructions, the operator inserts the Bionova BT110 Rapid Readout Biological Indicator into the reusable BI receptacle on the dedicated purge probe of the EOGas 4 Ethylene Oxide Gas Sterilizer, and initiates a 6-hour gas exposure at 50°C. After cycle completion, the SCBI is retrieved and activated by crushing the glass ampoule. The chemical indicator on the SCBI changes from brown/red to a green color after ethylene oxide exposure.
The activated SCBI and an unprocessed control are incubated in a Terragene Bionova IC10/20FR, IC10/20FRLCD or MiniBio Auto-Reader Incubator for 4 hours to detect fluorescent activity or 48 hours to detect color change. Evidence of microbial growth by presence of fluorescent activity or color change from blue to yellow must be interpreted as a failure to meet the conditions necessary for sterilization (cycle failed); no fluorescence or no color change indicates conditions for sterilization were achieved (cycle passed).
The provided document describes the EOGas 4 Endo-SteriTest RRBI, a rapid readout biological indicator for monitoring ethylene oxide sterilization cycles. The submission seeks to demonstrate substantial equivalence to a predicate device.
Here's a breakdown of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Functionality | 1) Critical parameters include time, temperature, gas concentration, and relative humidity 35-90% are met; 2) Device is appropriate for monitoring the efficacy of the sterilization process claimed. | Pass |
Shelf Life | Maintains performance specifications (resistance characteristics and correctly indicate pass/fail in cycles) throughout the stated shelf life of 2 years; Stability demonstrates reasonable assurance for effectiveness. | Pass |
Viable Spore Population | $\ge 1.0 \times 10^6$ Bacillus atrophaeus (ATCC 9372) spores. | $\ge 1.0 \times 10^6$ |
Resistance Characteristics | (Implicitly) Must be greater than the same biological indicator placed in the worst-case location of endoscope validation loads. | Met (Implied by "Pass" of full functionality) |
Growth Inhibition | No inhibition of viable spore growth by carrier or primary packaging materials. | Met (Implied by "Pass" of full functionality) |
Holding Time Assessment | (Details not explicitly stated, but expected for BI performance) | Met (Implied by "Pass" of individual studies) |
Reduced Incubation Time Validation | (Details not explicitly stated, but expected for RRBI performance) | Met (Implied by "Pass" of individual studies) |
Recovery Protocols for Recovery Medium | (Details not explicitly stated, but expected for BI performance) | Met (Implied by "Pass" of individual studies) |
Visual Readout Stability | (Details not explicitly stated, but expected for RRBI performance) | Met (Implied by "Pass" of individual studies) |
In-Field Evaluation | (Details not explicitly stated, but expected for BI performance) | Met (Implied by "Pass" of individual studies) |
Notes on Acceptance Criteria:
- The document explicitly lists "Functionality" and "Shelf Life" tests with their acceptance criteria and "Pass" results.
- For the individual studies performed for the Bionova BT110 Rapid Readout Biological Indicators (which is a component of the EOGas 4 Endo-SteriTest RRBI), the document states that "The results of all studies met the established acceptance criteria," without enumerating the specific criteria for each.
2. Sample Sizes Used for the Test Set and Data Provenance
The document does not provide specific sample sizes (e.g., number of BIs or sterilization cycles) for the "test set" studies. It mentions that "The EOGas 4 Endo-SteriTest RRBI has been validated using applicable tests in FDA 2007 'Guidance for Industry and FDA Staff: Biological Indicator (BI) Premarket Notification [510(k)] Submissions', and ANSI/AAMI/ISO 11138-1:2017 'Sterilization of health care products -Biological indicators - Part 1: General requirements'." These standards typically define the requirements for testing, including sample sizes, but the specific numbers used in this submission are not stated.
Data provenance is not explicitly mentioned but is assumed to be prospective bench testing conducted by the manufacturer as part of the validation process for regulatory submission. There is no mention of country of origin for data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. This device is a biological indicator designed to directly measure sterilization efficacy, not to be interpreted by human experts like medical images. The "ground truth" for a biological indicator is the direct observation of microbial growth (or lack thereof) after exposure to a sterilization process, relative to positive and negative controls.
4. Adjudication Method for the Test Set
Not applicable. As noted above, the device's output (microbial growth/no growth, or fluorescent activity/no fluorescent activity) is a direct, objective measurement, not subject to subjective interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This is not an AI-assisted diagnostic or imaging interpretation device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is not an AI algorithm. The device itself (the biological indicator) is a standalone measurement tool.
7. The Type of Ground Truth Used
The ground truth used for biological indicators is the viability of the Bacillus atrophaeus bacterial spores.
- Failed Sterilization: Indicated by observable fluorescent activity or a color change from blue to yellow in the culture medium, signifying spore survival and growth.
- Successful Sterilization: Indicated by no fluorescence and no color change, signifying spore inactivation.
- Positive controls (unprocessed BIs) are expected to show growth, and negative controls (processed BIs from a successful cycle) are expected to show no growth, confirming the validity of the test.
8. The Sample Size for the Training Set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(275 days)
Andersen Sterilizers, Inc.
Sterisheet Sterilization Wraps are single use non-woven sterilization wraps intended to enclose medical devices that are to be sterilized at a healthcare facility. They are used in 3-hour and 6hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.
Sterisheet Sterilization Wraps are single use, non-sterilyzation wraps constructed from cellulose, synthetic fibers (polypropylene), and synthetic binders, with the addition of pigmentation. They are used to enclose medical devices that are to be sterilized by a healthcare provider in 3-hour and 6-hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Devices must be wrapped following manufacturer's instructions. After completion of the sterilization process. Sterisheet Sterilization Wraps maintain sterility of the enclosed medical devices for 6 month.
AN85 EO Indicators, when placed on the outside of the sterilization wraps, may be used to secure the wrapping material on the devices and to indicate ethylene oxide exposure, offering a convenient way to verify processing in the sterilization cycle. The color of the AN85 EO Indicators changes from yellow-green to blue after exposure to ethylene oxide.
The information provided describes the acceptance criteria and a summary of non-clinical testing for the Sterisheet Sterilization Wrap (K200335), which is intended for use in a 6-hour ethylene oxide (EO) gas sterilization cycle.
Here's an analysis of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
Test | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Compliance to ISO 11607-1 | |||
Physical and Chemical Properties | To evaluate the package integrity after EO sterilization | Various standards for various properties (e.g., ISO 536 for Substance, ISO 5636-3 for Permeability, EN 868-2 for Fluorescence and Water repellency) | There was no effect on the physical and chemical properties after EO sterilization. |
Bacterial Filtration Efficiency and Germ Proofness dry and wet challenge test | To demonstrate that Sterisheet provide a microbial barrier property | Bacterial Filtration Efficiency (ASTM F2101); Germ Proofness dry and wet challenge test (DIN 58953-6) | Sterisheet passed all tests and were effective for sterility assurance pre- and post-EO sterilization. |
Material Compatibility | To demonstrate Sterisheet is suitable for use in EO sterilization processes and the associated cycle parameters | The physical, chemical, and microbial barrier properties of Sterisheet are compatible for intended use in EO sterilization per ISO 10993 and USP | The ability of the wraps to act as a microbial barrier was unaffected, and the contents of the packs remained sterile after EO sterilization. |
Biocompatibility | To demonstrate that Sterisheet provides reasonable assurance for safety | Biological evaluation (MEM Elution, Sensitization and Irritation Test) per ISO 10993 and USP | Not direct patient-contacting devices; Biological evaluation Showed that Sterisheet met all accepted performance criteria for non-toxicity. No pigment leaches out of Sterisheet during 6-hour EO exposures. With an additional 6-hour aeration, the residual EO detected in Sterisheet met ISO 10993-7 requirements. |
Shelf Life | To evaluate the physical, chemical, and microbial barrier of the processed Sterisheet at the end of the claimed shelf life | Stability demonstrates reasonable assurance for effectiveness | The physical, chemical, and microbial barrier properties of the wraps were verified at the end of 5 years and met specifications. |
Performance in the 6-hour Gas exposure in the EOGas 4 Ethylene Oxide Gas Sterilizer | |||
Sterilant Penetration | To demonstrate Sterisheet allow sterilization of the wrapped devices in half dose, full dose, simulated-use and in-use testing | Inactivation of 6-Log biological indicators at the worst-case location in the endoscope load wrapped in the Sterisheet for all the testing performed | Data demonstrate that Sterisheet Sterilization Wraps allow penetration of EO and sterilization of the enclosed devices. |
Maintenance of Package Sterility | To demonstrate Sterisheet maintain sterility of the enclosed medical device | The sterilized devices wrapped in Sterisheet remain sterile | Sterility was maintained for 6 months after processing in Sterisheet wrapped endoscopes. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document lists "Performance Testing" and "Summary of bench tests" but does not provide details on specific sample sizes for these tests. It indicates that the tests conform to ISO 11607-1, ASTM F2101, DIN 58953-6, ISO 10993, and USP standards. These are bench-level tests (non-clinical) and do not involve human subjects or geographical data provenance in the typical sense. The testing was conducted to prove the performance of the wrap itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable as the study described is a non-clinical performance evaluation of a sterilization wrap, not a diagnostic device relying on expert interpretation for ground truth. The "ground truth" here is the outcome of standardized physical, chemical, and biological tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the tests are objective laboratory tests rather than subjective assessments requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable as MRMC studies are relevant for diagnostic tools, particularly those involving human interpretation of medical images or data, often with AI assistance. The Sterisheet Sterilization Wrap is a medical device for sterilization, not a diagnostic or AI-powered interpretative tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable for the same reasons as above. The device is not an algorithm. The "standalone" performance here refers to the wrap's inherent ability to function as a microbial barrier and allow sterilant penetration.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the non-clinical tests is established through objective laboratory measurements and validated biological indicators. For example:
- Physical and Chemical Properties: Measured values against established standards (e.g., ISO for substance, permeability).
- Bacterial Filtration Efficiency and Germ Proofness: Efficacy in preventing microbial passage according to ASTM and DIN standards.
- Material Compatibility: Physical, chemical, and microbial barrier properties remaining acceptable after EO exposure.
- Biocompatibility: Absence of toxicity per ISO 10993 and USP (MEM Elution, Sensitization, Irritation).
- Sterilant Penetration: Inactivation of 6-Log biological indicators.
- Maintenance of Package Sterility: Absence of microorganism growth after storage.
8. The sample size for the training set
This information is not applicable. The Sterisheet Sterilization Wrap is not an AI/ML device that requires a "training set." The tests are empirical demonstrations of its physical, chemical, and biological performance.
9. How the ground truth for the training set was established
This information is not applicable as there is no training set for this device.
Ask a specific question about this device
(385 days)
Andersen Sterilizers, Inc.
The EOGas 4 Ethylene Oxide Gas Sterilize is designed to sterilize reusable medical devices that are sensitive to moisture, heat, chemical corrosion, or radiation. The critical process parameters for the two available cycles are summarized below:
Table 1. EOGas 4 Ethylene Oxide Sterilizer cycle parameters
EO Exposure Time | Total Cycle Time | EO Amount | Temperature | Relative Humidity |
---|---|---|---|---|
3 hours | 3.5 hours | 17.6 g ± 5% | 50°C ± 3°C | 35-70% |
6 hours | 7 hours |
The differences between the two options are the length of EO gas exposure and the length of mandatory ventilation after gas exposure; the gas exposure is chosen based on the devices to be sterilized. The appropriate purge probe and process challenge device (PCD) must be used: EOGas 4 SteriTest for a 3-hour gas exposure, EOGas 4 Endo-SteriTest for a 6-hour gas exposure.
The EOGas 4 Ethylene Oxide Gas Sterilizer 3-hour gas exposure is used for surface sterilization of medical devices, including instruments with diffusion-restricted spaces (hinges or mated surfaces), as well as for the sterilization of endoscopes with working length shorter than 1100 mm as specified in the labeling. The EOGas 4 SteriTest PCD is used with the 3-hour gas exposure.
The EOGas 4 Ethylene Oxide Gas Sterilizer 6-hour gas exposure is used for sterilization of duodenoscopes and colonoscopes with working length longer than 1100 mm as specified in the labeling. The EOGas 4 Endo-SteriTest PCD is used with the 6-hour gas exposure.
The EOGas 4 Ethylene Oxide Gas Sterilizer, model AN4000.60 (115V) or AN4000.61 (230V), is intended to sterilize moisture, temperature, chemical corrosion, or radiation-sensitive reusable medical devices in healthcare facilities. The sterilant is a unit dose of 100% ethylene oxide contained in a cartridge, and the sterilization chamber is a gas-impervious flexible sterilization bag.
Here's a summary of the acceptance criteria and study information for the EOGas 4 Ethylene Oxide Gas Sterilizer, based on the provided document:
Acceptance Criteria and Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Minimum Sterilization Parameters | The EO gas exposure time at 50°C in an EOGas 4 sterilizer must result in sterilization of the endoscope loads. | Using a 6-hour EO gas exposure in the EOGas 4 sterilizer, 6-Log biological indicators were consistently inactivated for two duodenoscopes, two colonoscopes, or one colonoscope and one duodenoscope in each load. The minimum parameters for sterilization of two duodenoscopes or colonoscopes are 518 mg/L EO, 46.4°C, and 35% RH. |
Half Dose Validation | For consecutive half dose cycles and full dose cycles: | |
EO concentration is half when half the amount of EO is used. | ||
All 6-Log Bacillus atrophaeus biological indicators, inoculated at the worst-case locations of the tested endoscopes, are inactivated. | Consecutive half dose cycles and full dose cycles were performed. The EO concentration was half when half the amount of EO was used. The cumulative lethality of half dose cycles was half the lethality of full dose cycle. The cycles consistently inactivated all 6-Log Bacillus atrophaeus biological indicators inoculated at the midpoint of the tested channels and at the elevator mechanism of duodenoscopes, as well as the water jet channel of the colonoscopes. | |
Simulated-Use Testing | For consecutive full dose cycles: | |
Biological indicators with 6-Log Bacillus atrophaeus, prepared in an artificial soil and inoculated at the worst-case locations of the tested endoscopes, are inactivated. | Biological indicators with 6-Log Bacillus atrophaeus were prepared in an artificial soil and inoculated at the center of the tested channels and at the elevator mechanism of duodenoscopes as well as the water jet channel of the colonoscopes. Inactivated biological indicators were obtained in all cycles for all duodenoscope and colonoscope loads tested. | |
In-Use Testing | Duodenoscopes and colonoscopes, used on patients, cleaned but not disinfected, are sterilized using the 6-hour cycle at 50°C in an EOGas 4 sterilizer. | Duodenoscopes and colonoscopes, used on patients, were cleaned per hospital protocol but not disinfected, processed using the 6-hour EO exposure at 50°C in the EOGas 4 sterilizer; sterility was tested by a flush method per USP. All test cultures from the processed duodenoscopes and colonoscopes were sterile. |
EO Residuals | After additional aeration following the cycle, EO residuals on duodenoscopes and colonoscopes are evaluated, and the residuals on the endoscopes and accessories meet the requirements of ANSI/AAMI/ISO 10993-7. | Olympus TJF-Q180V, TJF-Q160VF, TJF-Q190V, PJF-160, CF-Q180AL, Pentax ED34-i10T2, ED-3490TK duodenoscopes and EC-3490Li colonoscopes wrapped in Sterisheet must aerate for an additional 6 hours after the cycle. Fujifilm ED-530XT and EC-600HL wrapped in Sterisheet must aerate for an additional 8 hours after the cycle. After 6 hours of additional aeration, EO residuals on the packaging materials met the requirements of ANSI/AAMI/ISO 10993-7. |
Study Information:
-
Sample sizes used for the test set and data provenance:
- Minimum Sterilization Parameters: Performed to define and validate the endoscope loads. The results indicate successful inactivation of 6-Log BIs for "two duodenoscopes, two colonoscopes, or one colonoscope and one duodenoscope in each load." The exact number of each type of load tested isn't specified, but implies multiple runs for different combinations.
- Half Dose Validation: "Consecutive half dose cycles and full dose cycles were performed." No specific number of cycles or endoscopes is provided.
- Simulated-Use Testing: "Inactivated biological indicators were obtained in all cycles for all duodenoscope and colonoscope loads tested." Again, the exact number of cycles or endoscopes is not specified.
- In-Use Testing: "Duodenoscopes and colonoscopes, used on patients... processed using the 6-hour EO exposure." No specific number of patient-used scopes is provided.
- EO Residuals: Tested on "Olympus TJF-Q180V, TJF-Q160VF, TJF-Q190V, PJF-160, CF-Q180AL, Pentax ED34-i10T2, ED-3490TK duodenoscopes and EC-3490Li colonoscopes" and "Fujifilm ED-530XT and EC-600HL." This indicates testing on a variety of endoscope models.
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given the context of a 510(k) submission for a medical device cleared in the U.S., the testing would typically be conducted to U.S. regulatory standards. "In-Use Testing" mentions "used in routine endoscopic procedures in a hospital or clinic setting," suggesting prospective collection of real-world use data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This document describes performance testing of a sterilizer, where the "ground truth" is defined by the inactivation of biological indicators and the adherence to specified chemical and physical parameters. These are objective measures and do not typically involve human "experts" establishing a subjective ground truth, as would be the case for image interpretation or diagnosis. Therefore, no information on experts for ground truth is applicable or provided.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. As noted above, the ground truth is based on objective laboratory measurements (biological indicator cultures, chemical concentrations, temperature, humidity, and residual levels), not a subjective human assessment requiring adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This document pertains to the validation of an ethylene oxide gas sterilizer, not an AI-powered diagnostic or interpretive device. Therefore, no MRMC study or AI assistance evaluation was performed.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a sterilizer, not an algorithm. Its performance is evaluated through its physical and biological sterilization capabilities.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth for the sterilization efficacy tests is primarily established by:
- Biological Inactivation: The complete inactivation of 6-Log Bacillus atrophaeus biological indicators, confirmed by culture methods.
- Physical Parameters: Measurement of critical process parameters (EO concentration, temperature, relative humidity, time) meeting defined specifications.
- Chemical Residuals: Measurement of ethylene oxide residuals meeting ANSI/AAMI/ISO 10993-7 requirements.
- Sterility Testing: For in-use testing, sterility was confirmed by a flush method per USP, indicating no microbial growth.
- The ground truth for the sterilization efficacy tests is primarily established by:
-
The sample size for the training set:
- Not applicable. This device is a physical sterilizer and does not involve AI or machine learning models that require a training set.
-
How the ground truth for the training set was established:
- Not applicable, as no training set is used for this type of device validation.
Ask a specific question about this device
(385 days)
Andersen Sterilizers, Inc.
The EOGas 4 Endo-SteriTest consists of a self-contained biological indicator inoculated with viable Bacillus atrophaeus bacterial spores that is placed in a dedicated biological indicator receptacle mounted on a gold-colored purge probe in the sterilizer. It monitors the efficacy of the 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.
The EOGas 4 Endo-SteriTest consists of a single-use self-contained biological indicator (SCBI) placed in a reusable biological indicator (BI) receptacle. It is designed for monitoring the efficacy of the 6-hour gas exposure at 50°C in an EOGas 4 Ethylene Oxide Gas Sterilizer.
The SCBI, the EZTest-Gas Biological Indicator, consists of a plastic vial that serves as the culture tube, and a cap including a filter material port to allow ethylene oxide to enter the vial. The plastic vial contains Bacillus atrophaeus spores inoculated onto a paper carrier, and a glass ampoule containing modified soybean casein digest broth and phenol red acting as a pH indicator. There is a chemical indicator printed on the unit label of the SCBI to indicate EO exposure.
Following manufacturer's instructions, the operator inserts the EZTest-Gas Biological Indicator into the reusable BI receptacle on the dedicated purge probe of the EOGas 4 Ethylene Oxide Gas Sterilizer, and initiates a 6-hour gas exposure at 50℃. After cycle completion, the EZTest-Gas Biological Indicator is retrieved and activated by crushing the glass ampoule. The chemical indicator on the SCBI changes from blue to a green/brown color depending on the duration of ethylene oxide exposure. The activated SCBI and an unprocessed control are incubated at 30- 35°C for 48 hours and monitored for any color change and/or turbidity. Evidence of microbial growth by color change from red-orange to yellow and/or turbidity must be interpreted as a failure to meet the conditions necessary for sterilization (cycle failed); no color change or turbidity indicates conditions for sterilization were achieved (cycle passed).
The provided text describes the acceptance criteria and the results of studies conducted for the EOGas 4 Endo-SteriTest, a biological indicator for ethylene oxide sterilization.
Here's a breakdown of the requested information based on the document:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
EZTest-Gas Biological Indicator Performance Characteristics | ||
Viable Spore Population Assays | ≥ 10^6 microorganisms per unit; Stable for the stated 2-year shelf life | Meets the specifications for spore population according to 2007 FDA BI guidance and ANSI/AAMI/ISO 11138-1 |
Resistance Characteristics Study (D-value) | D-value ≥ 3 min; Survival/Kill Time: 15 min; D-values remain ±20% of the stated value for the claimed 2-year shelf life | Meets the specifications for resistance characteristics according to 2007 FDA BI guidance and ANSI/AAMI/ISO 11138-1 |
Growth Inhibition | All exposed carriers and packaging components are not bacteriostatic in a worst-case situation | Compliance with 2007 FDA BI guidance and ANSI/AAMI/ISO 11138-1 |
Holding Time Assessment | The resistance characteristics are not altered significantly over a 7-day hold time prior to incubation. | Compliance with 2007 FDA BI guidance |
Recovery Protocols, Medium and Incubation Time | Growth of an inoculum of 10-100 spores of Bacillus atrophaeus after being subjected to the sterilization process. | Compliance with 2007 FDA BI guidance and ANSI/AAMI/ISO 11138-1; Data generated supports an incubation time of 48 hours for EZTest-Gas biological indicators. |
D-Value Comparison for EZTest-Gas Biological Indicator | ||
D-value (in vacuum BIER vs. A-BIER) | D-values are within 20% of the original stated values for all lots of biological indicators tested. | There is no significant difference in the D-values measured in a vacuum BIER vs. an A-BIER. |
Performance Testing for the EOGas 4 Endo-SteriTest | ||
Half Dose Validation | Device is appropriate for monitoring the efficacy of the sterilization process claimed | Partial kill of biological indicators in PCD in half dose cycles, and complete kill in full dose cycles; Biological indicators at the worst-case locations in the endoscope loads are completely inactivated in both half and full dose cycles; EO concentration in the full dose cycles are twice the concentration in the half dose cycles for each endoscope load. |
Pass/Fail Results from the EOGas 4 Sterilizer | Correctly indicate pass/fail in cycles | Data demonstrate the EOGas 4 Endo-SteriTest correctly indicates passed and failed 6-hour gas exposures in the EOGas 4 throughout the stated shelf life. |
Endpoint Color Stability of the Chemical Indicator | Stability to provide reasonable assurance of effectiveness | The endpoint green/brown color was stable for at least 28 days when EZTest-Gas BIs were stored at 20-25°C or at 35-39°C after the EOGas 4 sterilization cycles. |
Shelf Life | Stability of the D-Value and spore population; Correctly indicate pass/fail in cycles throughout the stated shelf life | Maintains performance specifications throughout the stated shelf life of 2 years |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the specific sample sizes (number of biological indicators, number of runs, etc.) for each individual test. It mentions that "lots of biological indicators" were tested for D-value comparison, and refers to "half dose cycles" and "full dose cycles" for the half-dose validation. It does not provide information on the country of origin or whether the data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a biological indicator for sterilization. Its performance is evaluated through microbial growth/inactivation and chemical indicator changes, not by human expert interpretation of images or other data typically requiring expert consensus.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The ground truth for this device is based on objective biological and chemical responses, not subjective expert judgment that would require adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a biological indicator for monitoring sterilization efficacy, not an AI-assisted diagnostic device that involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is a standalone device in the context of it generating its own result (color change, microbial growth/no growth). Its performance is evaluated directly through its biological and chemical responses to sterilization conditions, independent of human intervention for its core function.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth used for this biological indicator is primarily the inactivation or survival of Bacillus atrophaeus bacterial spores when subjected to defined sterilization conditions, and the physical/chemical change of an indicator. This is an objective, measurable biological and chemical response, rather than expert consensus, pathology, or outcomes data.
8. The sample size for the training set
Not applicable. This device is a biological indicator, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable. As this is not a machine learning model, there is no training set or associated ground truth establishment process in that context. The "ground truth" for evaluating the BI's performance is established by microbiological standards for spore inactivation.
Ask a specific question about this device
(275 days)
Andersen Sterilizers, Inc.
AN1036 Dosimeter is a single-use color change chemical indicator. It is calibrated for sterilization temperature and used to verify adequate cumulative ethylene oxide exposure in a 3-hour or 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc.
The AN1036 Dosimeter is a single-use chemical indicator for cumulative ethylene oxide exposure. It is an accessory for the EOGas 4 Ethylene Oxide Gas Sterilizer.
The AN1036 Dosimeter contains a proprietary pH indicator in a glass capillary tube that is sealed on one end and mounted on a plastic tray. It is calibrated for a 50℃ sterilization temperature and responds to ethylene oxide concentration and sterilization time. With exposure to ethylene oxide, the indicator turns from yellow-orange to a dark blue color from the open end toward the closed end. The extent of the color change is proportional to cumulative ethylene oxide exposure. The cycle-specific calibration marks represent adequate cumulative ethylene oxide exposure to inactivate a 6-Log Bacillus atrophaeus biological indicator at the location of the AN1036 Dosimeter after 3-hour gas exposures or at the center of the endoscope lumens after 6-hour gas exposures. The AN1036 Dosimeter is not a replacement for a biological indicator.
Here's a summary of the acceptance criteria and study information for the AN1036 Dosimeter, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Chemical Indicator | Critical parameters with pass/fail criteria; Able to correctly indicate pass/fail results from the EOGas 4 Sterilizer | AN1036 Dosimeters respond to EO concentration, sterilization time, and temperature, at a relative humidity of 35-90%. AN1036 Dosimeters correctly indicate pass/fail in various endoscope loads. A correlation exists between the extent the blue color travels and cumulative EO exposure. |
Biocompatibility | Provides reasonable assurance for safety | AN1036 Dosimeter is not a direct or indirect patient-contacting device. None of the major ingredients is hazardous. No chemicals leach out of the AN1036 Dosimeters during the sterilization cycle. Without any additional aeration, the residual EO detected in Dosimeters met ISO 10993-7 requirements. |
Endpoint Color Stability | Stability demonstrates reasonable assurance for effectiveness | Endpoint blue color is stable for 28 days at 20-25°C or 4 days at 50°C after 6-hour gas exposures. |
Shelf Life | Maintains performance specifications throughout the stated shelf life | Data supports claimed shelf life of 3 years. |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document does not explicitly state a specific numerical sample size for the test set used in the "Chemical Indicator" performance evaluation. It mentions that "AN1036 Dosimeters correctly indicate pass/fail in various endoscope loads."
- Data Provenance: The studies were conducted by Andersen Sterilizers, Inc. as part of their 510(k) submission. The data is retrospective in the sense that it's a report of completed testing, but it's not specified if the data itself was collected retrospectively from past clinical cases or specifically generated for this submission through controlled laboratory experiments. Given the nature of a chemical indicator, it's highly likely it was generated through controlled laboratory testing. The country of origin for the data is implicitly the United States, as Andersen Sterilizers, Inc. is located in North Carolina, and the submission is to the FDA.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The "ground truth" for a chemical indicator's performance would typically be established based on instrumental measurements of ethylene oxide concentration, temperature, and time parameters, and the known biological inactivation levels corresponding to those parameters, rather than expert interpretation of the indicator itself.
4. Adjudication method for the test set:
- This information is not explicitly provided. For a chemical indicator, the "adjudication" would likely involve objective measurements and comparison against established physical/chemical standards, rather than a human consensus process.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human interpretation is involved. The AN1036 Dosimeter is a chemical indicator, not an AI device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- This question is not applicable. The AN1036 Dosimeter is a chemical indicator that works passively based on chemical reactions. It is not an algorithm or an AI system that performs independently.
7. The type of ground truth used:
- The ground truth for the "Chemical Indicator" test is based on the proper functioning of the EOGas 4 Ethylene Oxide Gas Sterilizer under specified critical parameters (Ethylene Oxide concentration, Temperature, Relative Humidity, EO Exposure Time, Total Cycle Time). The indicator's response (extent of color change) is correlated with "adequate cumulative EO exposure to inactivate a 6-Log Bacillus atrophaeus biological indicator." This implies a biological indicator equivalency as the ultimate ground truth, verified under controlled physical and chemical parameters.
8. The sample size for the training set:
- This information is not applicable. The AN1036 Dosimeter is a physical chemical indicator, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established:
- This information is not applicable, as there is no training set for a chemical indicator.
Ask a specific question about this device
(275 days)
Andersen Sterilizers, Inc.
Tyvek Sterilization Pouches with Chevron Seal are intended to enclose medical devices that are to be sterilized in a single pouch configuration at a healthcare facility. They are used in 3-hour and 6-hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.
Tyvek Sterilization Pouches with Chevron Seal are constructed from an uncoated Tyvek backing of fine, continuous, high-density polyethylene fibers, with front material consisting of a clear, laminated polyethylene terephthalate / low density polyethylene (LDPE) or LDPE-ethylene-vinyl acetate copolymer film. The pouches are used to enclose medical devices that are to be sterilized by a healthcare provider in 3-hour and 6-hour gas exposures at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Following manufacturer's instructions, devices are inserted into Tyvek Sterilization Pouches and sealed. The self-seal pouch permits sealing of the pouch without heatsealing equipment, whereas the heat-sealable pouches must be heat sealed prior to the cycle. Andersen Sterilizers' AN85 EO Indicators, when placed on the outside of the sterilization pouches. indicate EO exposure and offer a convenient way to differentiate pouches that have been processed in EO sterilization cycles from unprocessed units. The color of the AN85 EO Indicators changes from yellow-green to blue after exposure to EO. After completion of the sterilization process, the pouches maintain sterility of the enclosed medical devices for 6 months.
The provided document is a 510(k) premarket notification for a medical device (Tyvek Sterilization Pouches with Chevron Seal), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical study with detailed acceptance criteria and performance metrics for a novel AI/ML-driven device.
Therefore, the requested information about acceptance criteria, study design, sample sizes, expert involvement, MRMC studies, standalone performance, and ground truth establishment for an AI/ML device is not applicable to this document.
This submission is for a physical product (sterilization pouches) and details non-clinical performance testing related to its material properties, sterilization efficacy, and shelf life, not an AI/ML algorithm.
Ask a specific question about this device
(261 days)
Andersen Sterilizers, Inc.
AN87 Dosimeter is a single-use color change chemical indicator that is calibrated for sterilization temperature and used to verify adequate cumulative ethylene oxide exposure in the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer manufactured by Andersen Sterilizers, Inc.
The AN87 Dosimeter is a single-use chemical indicator for cumulative ethylene oxide exposure. It is an accessory for the Anprolene AN75 Ethylene Oxide Gas Sterilizer.
The AN87 Dosimeter contains a proprietary pH indicator in a glass capillary tube that is sealed on one end and mounted on a plastic tray. It is calibrated for a 20-29°C sterilization temperature, and responds to ethylene oxide concentration and sterilization time. With exposure to ethylene oxide, the indicator turns from yellow-orange to a dark blue color from the open end toward the closed end. The extent of the color change is proportional to the cumulative ethylene oxide exposure. The calibration mark represents adequate cumulative ethylene oxide exposure to inactivate a 6-Log Bacillus atrophaeus biological indicator at the location of the AN87 Dosimeter. The AN87 Dosimeter is not a replacement for a biological indicator.
The provided text describes the AN87 Dosimeter, a single-use chemical indicator for verifying adequate cumulative ethylene oxide exposure in the Anprolene AN75 Ethylene Oxide Gas sterilizer. The device is claimed to be substantially equivalent to the predicate AN1087 Dosimeter (K150645).
Here's an analysis of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are derived from the performance testing described in section 5.9 of the 510(k) summary, specifically from Table 5-3.
Acceptance Criteria (Test Description) | Reported Device Performance (Result) |
---|---|
Functionality: | Pass: |
1. Critical parameters include temperature, time, and gas concentration at a relative humidity of 35-90%. | Performance characterized in an Andersen Chemical Indicator Evaluator Resistometer and in an Anprolene AN75 Ethylene Oxide Gas Sterilizer using the 12-hour cycle at 20-29°C. The critical parameters to which the AN87 responds include temperature, time, and gas concentration in a relative humidity controlled environment. |
2. Performance parallels that of a biological indicator. | The AN87 indicates adequate cumulative EO exposure in the cycle for all validated loads in the Anprolene AN75 sterilizer. The AN87 performance parallels the performance of a 6-Log Bacillus atrophaeus biological indicator. |
3. A correlation exists between the extent the blue color travels and cumulative EO exposure. | A strong correlation is observed between cumulative EO exposure and the distance the blue color travels in the AN87 Dosimeter. |
Biocompatibility: Not direct or indirect patient-contacting devices; Non-toxic ingredients; Provides reasonable assurance for safety. | Pass: Not direct or indirect patient-contacting devices; Non-toxic ingredients; Provides reasonable assurance for safety. (This is a statement of compliance rather than a test result.) |
Endpoint Color Stability: Stable for at least 28 days at 20-25°C; Stability demonstrates reasonable assurance for effectiveness. | Pass: Using the AN87 Dosimeter at various stages of shelf life, the distance the blue line travels is stable for a minimum of 28 days after the sterilization cycle when AN87 Dosimeters are stored at room temperature (20-25°C). |
Shelf Life: Maintains performance specifications throughout the stated shelf life of 3 years. | Pass: Real-time shelf stability testing with the AN87 Dosimeter supports a shelf life of 3 years. |
2. Sample size used for the test set and the data provenance
The document does not specify the sample size used for the test set for any of the performance tests. It also does not explicitly state the data provenance (e.g., country of origin). The testing was conducted by Andersen Sterilizers, Inc., the manufacturer, implying internal data collection. The tests sound like prospective laboratory experiments rather than retrospective analysis of existing data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not provide information on the number or qualifications of experts used to establish the ground truth for the test set.
4. Adjudication method for the test set
The document does not describe any adjudication method for the test set.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a chemical indicator, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, standalone performance testing was done. The device itself (the AN87 Dosimeter, which is a chemical indicator and not an algorithm in the typical software sense) provides a direct visual indication (color change). The performance tests described characterize the device's ability to accurately reflect cumulative EO exposure without human interpretation beyond observing the color change against a calibration mark. The "algorithm" here is the chemical reaction itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the functionality testing was established by correlating the AN87 Dosimeter's color change with:
- The critical process parameters of the sterilization cycle (temperature, time, gas concentration, humidity).
- The performance of a 6-Log Bacillus atrophaeus biological indicator, which is a widely accepted standard for demonstrating sterilization efficacy. This acts as the direct ground truth for sterilization effectiveness.
8. The sample size for the training set
The concept of a "training set" is not applicable in the context of this chemical indicator. Chemical indicators operate based on predetermined chemical reactions and calibration, not machine learning or AI models that require training data.
9. How the ground truth for the training set was established
As there is no training set for this device, this question is not applicable. The device's calibration and performance are validated against physical and biological standards.
Ask a specific question about this device
(261 days)
Andersen Sterilizers,Inc.
Sterisheet Sterilization Wraps are single use non-woven sterilization wraps intended to enclose medical devices that are to be sterilized at a healthcare facility. They are used in the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer.
Sterisheet Sterilization Wraps are single use, non-sterile sterilization wraps constructed from cellulose, synthetic fibers (polypropylene), and synthetic binders, with the addition of pigmentation. They are used to enclose medical devices that are to be sterilized by a healthcare provider in the 12 hour sterilization cycle at 20-29°C in an Anprolene AN75 Ethylene Oxide Gas Sterilizer. Devices must be wrapped following manufacturer's instructions. After completion of the sterilization process, Sterisheet Sterilization Wraps maintain sterility of the enclosed medical devices for at least 3 months.
The provided document is a 510(k) premarket notification for a medical device called "Sterisheet Sterilization Wrap." It is not a document describing a study of an AI/ML powered device, nor does it contain information about acceptance criteria and performance data for such a device.
The product in question is a physical sterilization wrap, and the performance testing mentioned relates to its physical properties, material compatibility, biocompatibility, shelf life, and its ability to allow sterilant penetration and maintain sterility. The "study that proves the device meets acceptance criteria" refers to bench tests of a physical product, not a study of an AI/ML algorithm's performance.
Therefore, I cannot answer your request as it pertains to an AI/ML device study. The document does not contain any of the information you requested regarding:
- A table of acceptance criteria and reported device performance for an AI/ML algorithm.
- Sample sizes for a test set (in the context of an algorithm).
- Data provenance (country, retrospective/prospective) for an algorithmic test set.
- Number of experts for ground truth establishment or their qualifications.
- Adjudication methods for an algorithmic test set.
- MRMC comparative effectiveness studies.
- Standalone algorithm performance.
- Type of ground truth (expert consensus, pathology, outcomes data).
- Training set sample size or how its ground truth was established.
The document is solely a regulatory submission for a traditional medical device.
Ask a specific question about this device
(260 days)
Andersen Sterilizers,Inc.
The Anprolene SteriTest consists of a self-contained biological indicator inoculated with viable Bacillus atrophaeus bacterial spores that is placed in a dedicated BI receptacle in the sterilizer. It monitors the efficacy of the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized in Table 1.
The Anprolene® SteriTest consists of a single-use self-contained biological indicator (SCBI) placed in a reusable BI receptacle. It is designed for monitoring the efficacy of the 12 hour sterilization cycle at 20-29°C in an Anprolene AN75 Ethylene Oxide Gas Sterilizer.
The SCBI, the EZTest - Gas Biological Indicator, consists of a plastic vial that serves as the culture tube and a cap including a filter material port to allow ethylene oxide to enter the vial. The plastic vial contains Bacillus atrophaeus spores inoculated onto a paper carrier, and a glass ampoule containing modified soybean casein digest broth and phenol red acting as a pH indicator. There is a chemical indicator printed on the unit label of the SCBI to indicate EO exposure.
Following manufacturer's instructions, the operator inserts the SCBI into the reusable BI receptacle on the purge probe of the Anprolene AN75 Ethylene Oxide Gas Sterilizer, and initiates a 12 hour cycle at 20-29°C. After cycle completion, the SCBI is retrieved and activated by crushing the glass ampoule. The chemical indicator on the SCBI changes from blue to a green/brown color depending on the duration of ethylene oxide exposure. The activated SCBI and an unprocessed control are incubated at 30-35℃ for 48 hours, and monitored for any color change and/or turbidity. Evidence of microbial growth by color change from red-orange to yellow and/or turbidity must be interpreted as a failure to meet the conditions necessary for sterilization (failed cycle); no color change or turbidity indicates conditions for sterilization were achieved (passed cycle).
This document describes the Anprolene SteriTest, AN7508.14, a self-contained biological indicator for monitoring ethylene oxide gas sterilization. The information provided focuses on the performance testing to demonstrate substantial equivalence to a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Functionality (Resistance Characteristics) | - Partially positive in each half dose cycle when placed in worst-case loads (maximum fabric, metal, and plastic). |
- Negative in each full dose cycle when placed in worst-case loads (maximum fabric, metal, and plastic).
- Rigorous challenge to the Anprolene AN75 sterilization process, meaning its resistance is greater than the same EZTest-Gas Biological Indicator in worst-case locations of maximum loads. | - Partially positive in each half dose cycle.
- Negative in each full dose cycle.
- Resistance characteristics are greater than the same EZTest-Gas Biological Indicator placed in the worst-case locations of the maximum fabric, metal, and plastic loads. |
| Functionality (Suitability for Monitoring) | Device is appropriate for monitoring the efficacy of the sterilization process claimed (12-hour cycle, 20-29°C, 17.6g ± 5% EO, 35-90% RH). | Passed. |
| Shelf Life | Maintains performance specifications (resistance characteristics and correctly indicate pass/fail in cycles) throughout the stated shelf life of 2 years. Stability demonstrates reasonable assurance for effectiveness. | Passed. |
| Viable Spore Population Assay | (Not explicitly stated in the provided text, but implied as per FDA Guidance and ISO 11138-1 requirements) | Met established acceptance criteria. |
| Carrier and Primary Packaging Materials (Growth Inhibition) Evaluation | (Not explicitly stated in the provided text, but implied as per FDA Guidance and ISO 11138-1 requirements) | Met established acceptance criteria. |
| Holding Time Assessment | (Not explicitly stated in the provided text, but implied as per FDA Guidance and ISO 11138-1 requirements) | Met established acceptance criteria. |
| Reduced Incubation Time Validation | (Not explicitly stated in the provided text, but implied as per FDA Guidance and ISO 11138-1 requirements) | Met established acceptance criteria. |
| Equivalence of Resistance Characteristics (with/without vacuum) | Resistance characteristics are equivalent when measured in a Biological Indicator Resistometer in the presence or absence of a vacuum. | Equivalent. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific number of biological indicators (sample size) used for each individual test (e.g., functionality, shelf life) in numerical terms for the test set. It mentions "each half dose cycle" and "each full dose cycle" and "each worst-case load," implying multiple tests were conducted under these conditions.
The data provenance is from laboratory bench tests conducted by Andersen Sterilizers, Inc. (the manufacturer). The text does not specify the country of origin of the data beyond implicitly being associated with the manufacturer. The studies are prospective in nature, as they are specifically conducted to validate the device's performance against defined criteria.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. This is a biological indicator for sterilization, and the "ground truth" for its performance is established through microbiological testing (e.g., growth or no growth of spores) and physical measurements against established standards, not through human expert interpretation of images or data in the same way a diagnostic AI might require. The "experts" involved would be microbiologists and sterility assurance professionals conducting and interpreting the tests according to recognized standards.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth from multiple human readers in diagnostic studies. For biological indicators, the outcome is determined by the objective results of microbiological assays (e.g., color change indicating growth or no growth) and adherence to established physical/chemical parameters for resistance.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) and assessing how AI assistance impacts reader performance. The Anprolene SteriTest is a biological indicator, a device that provides an objective output (color change/turbidity) to monitor a sterilization process, not a diagnostic tool requiring human interpretation with or without AI.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the performance evaluation of the Anprolene SteriTest is inherently a "standalone" assessment of the device itself. The biological indicator (which can be considered the 'algorithm' in this context, albeit a physical/biological one) is placed in a sterilizer, and its output (color change upon incubation) directly indicates the success or failure of the sterilization cycle. Human involvement is for placement, retrieval, activation, and interpretation of the objective result, but the indicator's performance is intrinsic to its design and biological response, not a human-in-the-loop diagnostic process.
7. The Type of Ground Truth Used
The ground truth used for performance validation is based on microbiological viability (growth or no growth of Bacillus atrophaeus spores) and physical/chemical parameters of the sterilization cycle.
- Microbiological Ground Truth: This is established by observing whether the Bacillus atrophaeus spores on the indicator survive (indicated by growth and color change) or are killed (no growth, no color change) under defined sterilization conditions (full dose, half dose).
- Performance Standards Ground Truth: The tests adhere to established regulatory and consensus standards such as FDA 2007 "Guidance for Industry and FDA Staff: Biological Indicator (BI) Premarket Notification [510(k)] Submissions" and AAMI/ANSI/ISO 11138-1 "Sterilization of health care products – Biological indicators - Part 1: General requirements." These standards define the expected response characteristics (e.g., specific D-value equivalent, partial kills, full kills).
8. The Sample Size for the Training Set
Not applicable. Biological indicators like the Anprolene SteriTest are not "trained" in the way an AI algorithm learns from a training set. Their performance is inherent to their design, the number of spores, and their resistance characteristics. The tests described are for validation, not training.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" for this type of device. The device's performance is based on established scientific principles of microbiology and sterilization, and tested against those principles.
Ask a specific question about this device
Page 1 of 2