K Number
K170437
Date Cleared
2017-11-01

(261 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sterisheet Sterilization Wraps are single use non-woven sterilization wraps intended to enclose medical devices that are to be sterilized at a healthcare facility. They are used in the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer.

Device Description

Sterisheet Sterilization Wraps are single use, non-sterile sterilization wraps constructed from cellulose, synthetic fibers (polypropylene), and synthetic binders, with the addition of pigmentation. They are used to enclose medical devices that are to be sterilized by a healthcare provider in the 12 hour sterilization cycle at 20-29°C in an Anprolene AN75 Ethylene Oxide Gas Sterilizer. Devices must be wrapped following manufacturer's instructions. After completion of the sterilization process, Sterisheet Sterilization Wraps maintain sterility of the enclosed medical devices for at least 3 months.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called "Sterisheet Sterilization Wrap." It is not a document describing a study of an AI/ML powered device, nor does it contain information about acceptance criteria and performance data for such a device.

The product in question is a physical sterilization wrap, and the performance testing mentioned relates to its physical properties, material compatibility, biocompatibility, shelf life, and its ability to allow sterilant penetration and maintain sterility. The "study that proves the device meets acceptance criteria" refers to bench tests of a physical product, not a study of an AI/ML algorithm's performance.

Therefore, I cannot answer your request as it pertains to an AI/ML device study. The document does not contain any of the information you requested regarding:

  • A table of acceptance criteria and reported device performance for an AI/ML algorithm.
  • Sample sizes for a test set (in the context of an algorithm).
  • Data provenance (country, retrospective/prospective) for an algorithmic test set.
  • Number of experts for ground truth establishment or their qualifications.
  • Adjudication methods for an algorithmic test set.
  • MRMC comparative effectiveness studies.
  • Standalone algorithm performance.
  • Type of ground truth (expert consensus, pathology, outcomes data).
  • Training set sample size or how its ground truth was established.

The document is solely a regulatory submission for a traditional medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 1, 2017

Andersen Sterilizers,inc. William Andersen President 3154 Caroline Drive Haw River, North Carolina 27258

Re: K170437

Trade/Device Name: Sterisheet Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: September 28, 2017 Received: October 6, 2017

Dear William Andersen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K170437

Device Name Sterisheet® Sterilization Wrap

Indications for Use (Describe)

Sterisheet Sterilization Wraps are single use non-woven sterilization wraps intended to enclose mat are to be sterlized at a healthcare facility. They are used in the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized below in Table 1.

Table 1. Critical sterilization cycle parameters in the Anprolene AN75 Ethylene Oxide Gas Sterilizer
EO AmountTemperatureRelative HumidityEO Exposure TimeTotal Cycle Time
17.6 g ± 5%20-29°C35-90%12 hours14 hours

The product code for Sterisheet Sterilization Wraps is listed in Table 2.

Table 2. Product code for Sterisheet Sterilization Wraps
Product DesignationSterisheet S88 Blue
Product Code0129
Sizes18"x18", 20"x20", 24"x24", 30"x30", 36"x36", 40"x40", 45"x45", 48"x48", 54"x54", and 54"x72"

The 12 hour cycle in the Anprolene Oxide Gas Sterilizer has been validated to sterilize a load of up to 24 bs of metal surgical instruments, 3 lbs of fabric, or 3.5 lbs of plastic devices (combined weight of wrapped devices and tray). Sterility was maintained for at least 3 months after processing in an Anprolene sterilization cycle.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

510(k) Summary K170437

5.1 Applicant's Name and AddressAndersen Sterilizers, Inc.Establishment Registration Number 30046347103154 Caroline DriveHaw River, NC 27258
5.2 Contact PersonWilliam K. Andersen, BE, MD, FAAOSPresidentPhone: 336-376-8622, Fax: 336-376-5428
5.3 Date of PreparationOctober 20, 2017
5.4 Device
Proprietary NameSterisheet® Sterilization Wrap
Common NameSterilization Wrap
ClassificationClass II (21 CFR 880.6850)
Product CodeFRG
5.5 Predicate Device
Device NameSterisheet® Sterilization Wrap
510(k) numberK152291
ManufacturerAriowiggins Medical Inc

The predicate Sterisheet Sterilization Wrap (K152291) is approved for use in the EOGas 4 Ethylene Oxide Gas Sterilization system manufactured by Andersen Sterilizers, Inc.

This 510(k) submission modifies the indications for use of the predicate device in order to include it as a component in the Anprolene® AN75 Ethylene Oxide Gas Sterilization system manufactured by Andersen Sterilizers, Inc. No modifications were made to the manufacturing method, technology, or intended use.

5.6 Device Description

Sterisheet Sterilization Wraps are single use, non-sterile sterilization wraps constructed from cellulose, synthetic fibers (polypropylene), and synthetic binders, with the addition of pigmentation. They are used to enclose medical devices that are to be sterilized by a healthcare provider in the 12 hour sterilization cycle at 20-29°C in an Anprolene AN75 Ethylene Oxide Gas Sterilizer. Devices must be wrapped following manufacturer's instructions. After completion of

{4}------------------------------------------------

the sterilization process, Sterisheet Sterilization Wraps maintain sterility of the enclosed medical devices for at least 3 months.

AN85/AN86 EO Indicators, when placed on the outside of the sterilization wraps, may be used to secure the wrapping material on the devices and to indicate ethylene oxide exposure, offering a convenient way to verify processing in the sterilization cycle. The color of the AN85/AN86 EO Indicators changes from vellow-green to blue after exposure to ethylene oxide.

5.7 Indications for Use

Sterisheet Sterilization Wraps are single use non-woven sterilization wraps intended to enclose medical devices that are to be sterilized at a healthcare facility. They are used in the 12 hour sterilization cycle at 20-29°C in the Anprolene AN75 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized in Table 5-1.

Table 5-1. Critical sterilization cycle parameters in the Anprolene AN75 Ethylene Oxide Gas Sterilizer

Ethylene OxideTemperatureRelativeHumidityEthylene OxideExposure TimeTotal Cycle Time
17.6 g ± 5%20-29°C35-90%12 hours14 hours

The product code for Sterisheet Sterilization Wraps is listed in Table 5-2.

Product DesignationSterisheet S88 Blue
Product Code0129
Sizes18"x18", 20"x20", 24"x24", 30"x30", 36"x36",40"x40", 45"x45", 48"x48", 54"x54", and 54"x72"

Table 5-2. Product code for Sterisheet Sterilization Wraps

The 12 hour cycle in the Anprolene AN75 Ethylene Oxide Gas Sterilizer has been validated to sterilize a load of up to 24 lbs of metal surgical instruments. 3 lbs of fabric, or 3.5 lbs of plastic devices (combined weight of wrapped devices and tray). Sterility was maintained for at least 3 months after processing in an Anprolene AN75 sterilization cycle.

5.8 Device Comparison

The technological characteristics of the subject Sterilization Wraps are identical to the predicate device (K152291) - both are intended for the same use, use the same technology, and are designed in the same way.

The only difference between the subject Sterisheet Sterilization Wrap and the predicate device is the sterilization cycle for which the subject Sterilization Wrap is indicated. The

{5}------------------------------------------------

difference raises no issues related to safety or effectiveness of the subject device in the sterilization cycle. A comparison between the devices is listed in Table 5-3.

ElementsPredicate Sterisheet SterilizationWraps (K152291)Subject Sterisheet SterilizationWraps
ManufacturerArjowiggins HealthcareArjowiggins Healthcare
Intended UseTo enclose medical devices, allowsterilization of the enclosed devices,and maintain sterility of the encloseddevicesIdentical
Indications forUseSterisheet Sterilization Wraps are singleuse non-woven sterilization wrapsintended to enclose medical devices thatare to be sterilized at a healthcarefacility. They are used in the 3 hoursterilization cycle at 50°C in the EOGas4 Ethylene Oxide Gas Sterilizer.Sterisheet Sterilization Wraps are singleuse non-woven sterilization wrapsintended to enclose medical devices thatare to be sterilized at a healthcarefacility. They are used in the 12 hoursterilization cycle at 20-29°C in theAnprolene AN75 Ethylene Oxide GasSterilizer.
MaterialsCellulose, synthetic fibers(polypropylene), and synthetic bindersIdentical
DesignCellulose allows EO to pass through thewrap but prevents microorganisms fromcrossing through the wrap, providing amicrobial barrier for the wrappeddevices after sterilization;Synthetic fibers increase mechanicalresistance;Synthetic binders enhance drapeability,strength, softness, and fluid repellencyIdentical
Wrap ShapeSquare or rectangularIdentical
Configuration inLoadDouble sequential envelope wrap isrecommendedIdentical
Shelf Life5 years from date of manufactureIdentical
Aeration Time≥ 6 hours≥ 6 hours

Table 5-3. Device Comparison

{6}------------------------------------------------

5.9 Performance Testing

Sterisheet Sterilization Wraps conform to all applicable requirements for packaging for terminally sterilized medical devices for EO sterilization, based on ISO 11607-1. Performance testing was conducted to show that Sterisheet Sterilization Wraps perform as intended to allow sterilization and maintain sterility of the enclosed medical device. Sterilization efficacy testing demonstrated a sterility assurance level of 10-6 using the half dose validation method under worst-case conditions. Shelf life studies demonstrated after completion of the Anprolene AN75 sterilization process, sterility is maintained for at least 3 months. The performance of Sterisheet Sterilization Wraps is summarized in Table 5-4.

Table 5-4. Summary of bench tests performed to demonstrate safety and effectiveness of
Sterisheet Sterilization Wraps
TestDescriptionResults
Compliance to ISO 11607-1
Package IntegrityPorous material providing a microbial barrier;Physical and chemical properties are maintainedMeets requirements
Material CompatibilityCompatibility with respect to forming and sealingprocess-suitable folding and drapeability;Suitable for use in EO sterilization processesMeets requirements
BiocompatibilityNot direct patient-contacting devices;Materials are non-toxic and meet ISO 11607-1requirements;Biological evaluation meets acceptable criteria;Bio-burden control;Provides reasonable assurance for safetyPass
Shelf LifePhysical properties and microbial barrier of theprocessed Sterisheet Sterilization Wraps wereverified at the end of the claimed shelf life of 5years:Stability demonstrates reasonable assurance foreffectivenessPass
Performance in the Anprolene AN75 Cycle
Sterilant PenetrationEO penetrated the wraps under worst-case half-dose conditions or an extreme biologicalchallenge scenario, and inactivated 6-Logbiological indicatorsAllow a sterilityassurance level of 10-6for the sterilization cycle
Maintenance ofPackage SterilitySterility was maintained for at least 3 months afterprocessing in an Anprolene AN75 sterilizer.Pass

5.10 Conclusion

Based on the intended use, technological characteristics, and nonclinical tests performed, the subject device Sterisheet Sterilization Wrap (K170437) is as safe and effective as the legally marketed predicate device, Sterilization Wrap (K152291) intended for used in the EOGas4 sterilization system.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).