510(k) Data Aggregation
(308 days)
A patient interface, face mask, for use with CPAP and bi-level systems used in the treatment of adult (>30 kg) OSA and / or ventilatory support. Two styles (with exhalation port and without exhalation port). Single use only
The proposed patient interface face mask incorporates a number of features, which are designed to maximize seal and comfort, and maintain the mask in the correct position throughout use. Adhesive foam to seal to the patient face. One size - medium. Models with and without Integral fixed leak (exhalation) port.
Here's a breakdown of the acceptance criteria and the study information based on the provided text, categorized as requested:
1. Table of Acceptance Criteria and Reported Device Performance
Feature/Metric | Acceptance Criterion (Predicate Device Performance) | Reported Device Performance (ARCimed CPAP mask) |
---|---|---|
Exhaust Flow Range | Pressure / Flow (lpm) | |
3 cm H2O / 17.8 lpm | ||
10 cm H2O / 31.5 lpm | ||
20 cm H2O / 42.6 lpm | ||
30 cm H2O / 51.7 lpm | ||
40 cm H2O / 60.1 lpm | ||
(Based on ResMed Quattro K063122) | Pressure / Flow (lpm) | |
3 cm H2O / 19.9 lpm | ||
10 cm H2O / 33.3 lpm | ||
20 cm H2O / 47.1 lpm | ||
30 cm H2O / 58.0 lpm | ||
40 cm H2O / 67.8 lpm | ||
Pass / fail +/- 15% (Implied acceptance within this range of predicate) | ||
Pressure Drop (Resistance to Flow) | ResMed – K063122 | |
0.59 cm H2O @ 50 lpm | ||
1.2 cm H2O @ 100 lpm | 0.35 cm H2O @ 50 lpm | |
0.81 cm H2O @ 100 lpm | ||
CO2 Rebreathing | ResMed – K063122 | |
0.25% EtCO2 (Measured change from baseline) | 0.06% EtCO2 (Measured change from baseline) | |
Adhesive Seal | Advanced Warming – K950771 | |
Offers a tight seal | ||
No performance requirements other than it will remain attached to patient's face | We are claiming no performance requirements other than the mask will seal and remain attached to the patient's face. | |
Vented/Non-vented models (Anti-Asphyxia Valve compatibility) | Yes – ResMed – K063122 | |
Offer a mask with optional anti-asphyxia valve | Vented and Non-vented models requires attachment to a circuit with anti-asphyxia valve incorporated demonstrated to activate AAV valve equivalent to predicates. Should be used with AAV with minimum opening pressures of |