Search Results
Found 8 results
510(k) Data Aggregation
(113 days)
ADVANCED MEDICAL OPTICS, INC.
The Vitrectomy Cutter is a device intended to remove vitreous matter from the eye and the Irrigation Sleeve provides irrigation during the vitrectomy procedure. These devices are for use with the Whitestar Signature™ System.
The Vitrectomy Cutter is a device intended to remove vitreous matter from the eye and the Irrigation Sleeve provides irrigation during the vitrectomy procedure. The vittectomy mode is used to cut and remove vitreous matter from the anterior segment of the eye during secondary intraocular lens implantation following vitreous loss associated with trauma or during primary cataract surgery. It can also be used to provide vitreous cutting in the posterior segment by enabling surgeons to remove vitreous matter for the treatment of various posterior segment conditions.
This is a 510(k) premarket notification for a medical device cleared by the FDA. As such, it does not contain a study that proves the device meets acceptance criteria in the manner requested (e.g., with detailed performance metrics, sample sizes, expert involvement, and statistical analyses typically associated with clinical trials or AI performance evaluations).
The document is a regulatory submission demonstrating substantial equivalence to a previously cleared predicate device, rather than a de novo clinical study proving novel performance criteria. Clinical studies as described in your request are not typically required for 510(k) clearances if substantial equivalence can be established through other means, such as performance testing detailed below.
Therefore, many of the requested fields cannot be directly extracted or are not applicable in their usual sense for this type of submission.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document describes the device and states that "The physical and performance characteristics of the vitrectomy cutter and the irrigation sleeve are substantially equivalent to those of the predicate devices." It also states "The products will be manufactured in compliance with both FDA (QSR) and ISO standards and have demonstrated to be safe and effective for the indicated use."
Specific, quantitative acceptance criteria and their corresponding reported device performance values are not explicitly detailed in the provided text. The assessment for this type of device typically involves engineering specifications and functional performance tests (e.g., cutting rate, aspiration flow, material compatibility, sterilization efficacy, etc.) to demonstrate equivalence to the predicate device. These specific test results are usually found in the full 510(k) submission, which is more extensive than this summary.
Acceptance Criteria (Implied) | Reported Device Performance (Implied) |
---|---|
Physical Characteristics equivalent to predicate | Substantially equivalent to predicate devices (K950287) |
Performance Characteristics equivalent to predicate | Substantially equivalent to predicate devices (K950287) |
Manufactured to FDA (QSR) standards | In compliance with FDA (QSR) standards |
Manufactured to ISO standards | In compliance with ISO standards |
Safe for indicated use | Demonstrated to be safe for indicated use |
Effective for indicated use | Demonstrated to be effective for indicated use |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified in the provided text. Performance testing for medical devices like this typically involves a defined number of units, but this information is not in the summary.
- Data Provenance: Not specified, but likely obtained from internal laboratory testing conducted by Advanced Medical Optics, Inc. in the USA. Given it's a 510(k) for an existing type of device, it's almost certainly retrospective in the sense that it relies on established engineering principles and performance testing benchmarks rather than prospective clinical trials to prove efficacy.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable (N/A): This type of medical device submission does not typically involve "ground truth" derived from expert consensus in a clinical diagnostic sense, as it is a surgical tool. Its performance is assessed through engineering and functional testing rather than diagnostic accuracy.
4. Adjudication method for the test set:
- Not Applicable (N/A): Adjudication methods like 2+1 or 3+1 are used in clinical studies for diagnostic outcomes. For this device, performance is evaluated against engineering specifications and predicate device equivalence, not expert adjudication of clinical findings.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No: This is a surgical instrument, not a diagnostic imaging device or an AI-powered system that assists human readers. Therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant and was not conducted.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No: This device is a manual surgical instrument and does not involve an algorithm or AI operating in a standalone capacity.
7. The type of ground truth used:
- Engineering Specifications / Predicate Device Performance: The "ground truth" for this submission is implicitly the established safety and performance profile of the predicate device (K950287 UnuVit and 4422CE) and adherence to recognized industry standards (FDA QSR, ISO). Performance is measured against these established benchmarks.
8. The sample size for the training set:
- Not Applicable (N/A): This device does not use machine learning or AI, so there is no "training set."
9. How the ground truth for the training set was established:
- Not Applicable (N/A): As there is no training set for an AI/ML algorithm, this question is not relevant.
Ask a specific question about this device
(122 days)
ADVANCED MEDICAL OPTICS, INC.
Used to fold and assist in inserting TECNIS® 1-Piece intraocular lenses into the eye.
The One Series™ Ultra cartridge is used to fold and assist in the insertion of a TECNIS® 1-Piece soft acrylic intraocular lens (IOL) into the eye following cataract extraction. The cartridge, Model 1VIPR30 is a single-use component composed of polypropylene which is injection molded and coated with a lubricious coating. The cartridge is provided sterile. The IOL is loaded into the proximal portion of the cartridge using forceps. The cartridge is then placed in a reusable titanium handpiece, which advances the IOL through the tube section of the cartridge and delivers it into the eye.
The provided text is related to a 510(k) premarket notification for a medical device called the "ONE SERIES™ Ultra Cartridges" (later renamed ONE SERIESTM Ultra Cartridges). This document primarily serves as an FDA clearance letter and a 510(k) summary, establishing substantial equivalence to predicate devices rather than detailing a specific clinical study with granular acceptance criteria and performance metrics.
Therefore, much of the requested information regarding acceptance criteria, specific study design, sample sizes, expert qualifications, and ground truth establishment, as typically found in detailed study reports, is not present in these documents. The document focuses on demonstrating equivalence through bench testing and biocompatibility.
Here's an analysis based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria or specific performance metrics in a tabular format as would be found in a detailed study report. It broadly states that the device was found equivalent to predicate devices based on "material and performance characteristics."
Acceptance Criteria (Inferred/General) | Reported Device Performance (General) |
---|---|
Material equivalence to predicate devices | Found equivalent |
Sterilization method equivalence | Found equivalent |
Intended use equivalence | Found equivalent |
Mode of operation equivalence | Found equivalent |
Performance: Folding of IOL | Verified to be equivalent to predicate devices |
Performance: Delivery of IOL | Verified to be equivalent to predicate devices |
Performance: Retention of IOL | Verified to be equivalent to predicate devices |
Performance: Cosmetic properties of IOL | Verified to be equivalent to predicate devices (implying no negative impact on cosmetic properties) |
Performance: Dimensional properties of IOL | Verified to be equivalent to predicate devices (implying no negative impact on dimensional properties) |
Performance: Optical properties of IOL | Verified to be equivalent to predicate devices (implying no negative impact on optical properties) |
Biocompatibility | Conducted and verified |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified. The document mentions "Bench testing and biocompatibility testing," but does not provide details on the number of cartridges or IOLs tested.
- Data Provenance: Not specified. It's likely the testing was conducted in a laboratory setting by the manufacturer (Advanced Medical Optics, Inc.) in the USA, as indicated by their address. The testing is non-clinical.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- This is not applicable for this type of submission. The performance testing described (folding, delivery, retention, cosmetic, dimensional, optical properties) would likely be assessed against engineering specifications and predicate device performance, not by expert human interpretation in the way an AI diagnostic device would be.
4. Adjudication Method for the Test Set
- Not applicable as the testing involves objective measurements against specifications rather than expert consensus on subjective interpretations.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- No. This is not an AI-assisted diagnostic device, but rather a physical medical device (intraocular lens cartridge). Therefore, MRMC studies and AI-related effectiveness metrics are not relevant or present in this document.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
- No. This device is an intraocular lens cartridge, not an algorithm.
7. The Type of Ground Truth Used
- The "ground truth" for the device's performance would be engineering specifications, material standards, and benchmark performance of the predicate devices. For example, successful folding and delivery without damage to the IOL would be objectively verifiable against defined criteria. Biocompatibility would be assessed against established standards for medical devices.
8. The Sample Size for the Training Set
- Not applicable. This device is not an AI algorithm and therefore does not have a "training set" in the machine learning sense. The "training" for such a device involves product development, engineering, and manufacturing processes to meet design specifications, not data-driven model training.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. See point 8.
Ask a specific question about this device
(53 days)
ADVANCED MEDICAL OPTICS, INC.
The AMO Ophthalmic Surgical System is a modular ophthalmic microsurgical system that facilitates both anterior segment (i.e., cataract) and posterior segment (i.e., vitreoretinal) ophthalmic surgery. The posterior segment surgery module is optional. The modular design allows the users to configure the system to meet their surgical requirements.
The AMO Ophthalmic Surgical System is a device designed for use in performing anterior segment ophthalmic surgery (cataract) and posterior segment ophthalmic surgery. The device is intended for use in performing phacoemulsification, phacofragmentation, diathermy, irrigation/aspiration, vitrectomy, extrusion and silicone oil injection and extraction.
The provided document describes the AMO Ophthalmic Surgical System, a device for anterior and posterior segment ophthalmic surgery. However, it does not include information about acceptance criteria, a study proving device performance against such criteria, sample sizes for test or training sets, ground truth establishment, or any details about AI/ML algorithms or their evaluation.
This 510(k) submission is for a traditional medical device (phacofragmentation system) and focuses on demonstrating substantial equivalence to predicate devices through technological characteristics and compliance with general safety and performance standards. It does not involve AI/ML components or associated performance studies typical for AI-powered devices.
Therefore, I cannot fulfill your request for information regarding acceptance criteria, study details, sample sizes, ground truth, or MRMC studies, as these elements are not present in the provided text regarding the AMO Ophthalmic Surgical System.
Ask a specific question about this device
(311 days)
ADVANCED MEDICAL OPTICS, INC.
Use LENS PLUS® Rewetting Drops while wearing your lenses to moisten and rehydrate them. Also, use LENS PLUS® Rewetting Drops to relieve minor irritation, discomfort and/or blurring which may occur while wearing your lenses.
Lens Plus® Rewetting Drops is a sterile, preservative free, buffered, isotonic, aqueous solution containing sodium chloride and boric acid. The product is packaged in a plastic sterile reclosable vial which allows remaining drops can be used throughout the day (12 hours).
The provided text describes a 510(k) summary for "Lens Plus® Rewetting Drops" and its substantial equivalence to a predicate device. This document focuses on the safety and performance of a medical device (rewetting drops) and not on the performance of an AI/ML algorithm.
Therefore, the requested information categories concerning AI/ML algorithm performance (such as sample sizes for test sets, ground truth establishment, expert qualifications, MRMC studies, standalone performance, and training set details) are not applicable to this document. The study described is a non-clinical bacteriostasis study.
Here's the relevant information that can be extracted from the provided text, structured to address the spirit of your request:
Acceptance Criteria and Device Performance Study for Lens Plus® Rewetting Drops (K042562)
The primary "acceptance criterion" for this device, as detailed in the 510(k) summary, revolves around demonstrating the ability of bacteria to not significantly grow and survive in the reclosable vials containing the rewetting drops over a designed period, specifically to support a 12-hour discard statement.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|
No significant growth and survival of specified bacteria over time | Staphylococcus aureus, Escherichia coli, Pseudomonas aeruginosa, Candida albicans, and Aspergillus niger showed no significant growth and survive over the designed period. |
Support for 12-hour discard statement | "The data support the desired discard statement on label of up to 12 hours." |
2. Sample size used for the test set and the data provenance
- Sample Size (Test Set): Not explicitly stated in terms of number of vials or bacterial cultures beyond the types of bacteria tested. The study mentions testing against five specific microorganisms.
- Data Provenance: The study was conducted "in accordance with Micro Appendix C of the Premarket Notification (510(K)) Guidance Document for contact Lens Care Products." This is a non-clinical, controlled laboratory study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This was a laboratory study evaluating bacteriostasis, not requiring human expert interpretation in the way AI/ML ground truth is established.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This was a laboratory study.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML device.
7. The type of ground truth used
The "ground truth" here is the objective microbial growth/survival data obtained from the bacteriostasis study, evaluated against established microbiological standards for such products.
8. The sample size for the training set
Not applicable. This is not an AI/ML device, and no training set was used.
9. How the ground truth for the training set was established
Not applicable. No training set was used.
Ask a specific question about this device
(22 days)
ADVANCED MEDICAL OPTICS, INC.
The SOVEREIGN® High Vacuum Pack is indicated for use with the SOVEREIGN® phacoemulsification machine. The SOVEREIGN® High Vacuum pack includes a filter protected flow restriction orifice used during cataract surgery to minimize vacuum surges in the aspiration line. The SOVEREIGN® Compact High Vacuum Pack is indicated for use with the SOVEREIGN® Compact phacoemulsification machine. The SOVEREIGN® Compact High Vacuum pack includes a filter protected flow restriction orifice used during cataract surgery to minimize vacuum surges in the aspiration line.
The SOVEREIGN® High Vacuum Pack and the SOVEREIGN® Compact High Vacuum Pack are accessory kits indicated for use with the SOVEREIGN® phacoemulsification machine and the SOVEREIGN® Compact phacoemulsification machine, respectively, to perform an irrigation/aspiration (I/A) procedure or a phacoemulsification procedure during cataract surgery. The tubing assembly component of the kit incorporates a flow restriction feature for preventing/neutralizing post occlusion surge following an occlusion break at the distal end of a phaco tip. The materials, basic scientific concepts, physical properties and intended use of the device kits are similar to those of the identified predicate devices.
The provided text describes a 510(k) summary for the SOVEREIGN® High Vacuum Pack and SOVEREIGN® Compact High Vacuum Pack, which are accessory kits for phacoemulsification machines used in cataract surgery. The summary focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed study with acceptance criteria and performance metrics in the typical sense of a diagnostic medical device.
However, based on the information provided, we can infer some "acceptance criteria" and the nature of the "study" conducted for this type of submission. The primary goal was to prove substantial equivalence to existing legally marketed devices.
Here's an interpretation of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Inferred from Substantial Equivalence Claim) | Reported Device Performance |
---|---|
Equivalence in Intended Use: Device performs irrigation/aspiration or phacoemulsification during cataract surgery, minimizes vacuum surges. | The SOVEREIGN® High Vacuum Pack and SOVEREIGN® Compact High Vacuum Pack are indicated for use with the respective SOVEREIGN® phacoemulsification machines to perform irrigation/aspiration or phacoemulsification during cataract surgery. They include a filter protected flow restriction orifice to minimize vacuum surges in the aspiration line, similar to predicate devices. |
Equivalence in Technological Characteristics: Materials, method of sterilization, mode of operation. | "The technological characteristics of the SOVEREIGN® High Vacuum Pack were compared to those of the predicate devices and were found to be equivalent with respect to the materials, method of sterilization, intended use, and mode of operation." |
Equivalence in Performance Characteristics (Nonclinical): Flow resistance and chamber maintenance. | "Bench testing was conducted which verified that the performance characteristics (flow resistance and chamber maintenance) associated with the SOVEREIGN® High Vacuum Pack were equivalent to that of the predicate devices." |
Safety and Effectiveness: No new questions of safety or effectiveness. | "AMO® has demonstrated through its evaluation of the SOVEREIGN® High Vacuum Pack that the device is equivalent to the predicate devices with respect to intended use, technological characteristics, and safety and effectiveness." |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not specify a "sample size" in the context of clinical patient data. The evaluation was primarily bench testing. For bench testing, the sample size would refer to the number of units tested. This information is not provided in the document.
- Data Provenance: The data provenance is nonclinical bench testing. The country of origin for the testing is not explicitly stated. It is considered retrospective in the sense that it compares performance to established predicate devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: Not applicable. This was bench testing, not an expert-driven clinical evaluation requiring ground truth establishment in that manner.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. Bench tests typically involve direct measurement and comparison to predefined engineering specifications or predicate device performance, not expert adjudication in the clinical sense.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- MRMC Study: No, an MRMC study was not done. This device is a surgical accessory, not an AI-powered diagnostic tool.
- Effect Size of AI Improvement: Not applicable.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Standalone Performance: Not applicable. This device is a physical accessory used with a phacoemulsification machine, hence inherently "human-in-the-loop" in its application during surgery. No standalone algorithm performance was assessed.
7. The Type of Ground Truth Used
- Type of Ground Truth: The "ground truth" for the bench testing was derived from established performance characteristics of the predicate devices (e.g., their known flow resistance and chamber maintenance capabilities). This is a form of engineering specification/benchmark comparison, rather than expert consensus, pathology, or outcomes data in a clinical trial.
8. The Sample Size for the Training Set
- Training Set Sample Size: Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The device design and verification were based on engineering principles and comparisons to existing predicate designs.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set Establishment: Not applicable, as there was no training set in the context of machine learning.
Ask a specific question about this device
(202 days)
ADVANCED MEDICAL OPTICS, INC.
COMPLETE® BLINK-N-CLEAN® Lens Drops is indicated to lubricate and rewet soft (hydrophilic) contact lenses, disposable lenses and extended wear lenses, as well as rigid gas permeable lenses before application and during wear.
COMPLETE® BLINK-N-CLEAN® Lens Drops is a sterile, isotonic, buffered, solution containing lubricants, preservatives, buffers, surfactants, ancillary ingredients, and purified water. The product is a clear, colorless solution packaged in plastic bottles with controlled dropper tips.
The provided text describes a 510(k) summary for COMPLETE® BLINK-N-CLEAN® Lens Drops, which is a medical device. However, the information available in the document is primarily related to its regulatory submission and non-clinical testing for compatibility and protein removal, not advanced AI/ML device studies as implied by the detailed questions about acceptance criteria, ground truth, expert adjudication, and MRMC studies.
Therefore, many of the requested details cannot be extracted or are not applicable to the type of device and study described.
Here's an attempt to answer based on the provided text, highlighting where information is not available:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Compatibility with Lenses: | |
Average changes in diameter (silicone acrylate and fluorosilicone acrylate RGP) | Within established acceptance criteria, comparable to controls. |
Average changes in power (silicone acrylate and fluorosilicone acrylate RGP) | Within established acceptance criteria, comparable to controls. |
Average changes in base curve (silicone acrylate and fluorosilicone acrylate RGP) | Within established acceptance criteria, comparable to controls. |
Absence of surface deposits, discoloration, and/or deformities. | No evidence observed. |
Protein Removal: | |
Significant protein removal compared to predicate devices. | Significant amount of protein removed compared to predicate devices. |
Preclinical Safety and Efficacy: | Established in P910075/S7. |
Clinical Safety and Acceptability: | Established in 910075/S7. |
2. Sample size used for the test set and the data provenance:
- Sample Size for Compatibility Test: Thirty (30) regimen cycles of silicone acrylate and fluorosilicone acrylate rigid gas permeable (RGP) lenses.
- Data Provenance: Not specified in terms of country of origin or whether it was retrospective/prospective for the compatibility and protein removal studies. It appears to be non-clinical laboratory testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable as the reported studies are non-clinical (compatibility and protein removal) and do not involve expert interpretation or ground truth establishment in the context of clinical expert consensus. The "ground truth" here is objective physical measurements and chemical analysis.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable to the non-clinical studies described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a lens care solution, not an AI/ML diagnostic or assistive device that would involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a lens care solution, not an algorithm. The reported studies evaluate the physical and chemical performance of the solution itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For Compatibility: Objective physical measurements (diameter, power, base curve) and visual observations (surface deposits, discoloration, deformities) of contact lenses after exposure to the product.
- For Protein Removal: Quantitative analytical methods measuring protein accumulation and removal from human-worn contact lenses in a simulated in-eye use.
8. The sample size for the training set:
- Not applicable. This is not an AI/ML device; therefore, there is no "training set."
9. How the ground truth for the training set was established:
- Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(86 days)
ADVANCED MEDICAL OPTICS, INC.
Use blink™ CL Lubricant Eye Drops to lubricate and rewet soft and rigid gas permeable (RGP) contact lenses; to help relieve dryness, irritation and discomfort that may be associated with lens wear; and to cushion lenses by placing a drop on the lens prior to application on the eye.
blink™ CL Lubricant Eye Drops is a sterile, isotonic, buffered solution containing a lubricant, a preservative, buffers, tonicity agents, and purified water. The product is a clear, colorless solution packaged in plastic bottles with controlled dropper tips.
1. Table of Acceptance Criteria and Reported Device Performance
Test Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Nonclinical Studies | ||
Lens Compatibility | Compatible with both soft (hydrophilic) and RGP contact lenses to be considered substantially equivalent to the control. | The product is compatible with soft (hydrophilic) and RGP contact lenses and substantially equivalent to the control. |
Solution Compatibility | Compatible with leading contact lens care products on the market. | The product is compatible with leading contact lens care products. |
Preservative Uptake and Release | Very little, if any, uptake of the preservative into or onto soft or RGP lenses, with any absorbed amounts quickly released, indicating compatibility and acceptability for use. | There is very little, if any, uptake of the preservative in or onto soft or RGP lenses. Any amounts taken up are quickly released. The product is compatible and acceptable for use with soft (hydrophilic) and RGP contact lenses. |
Contact Lens Wetting Angle | Substantially equivalent in wetting properties to predicate lubricating and rewetting products. | blink™ CL Lubricant Eye Drops is substantially equivalent in wetting properties to the predicate devices. |
Microbiological Studies (Preservative Efficacy) | Meets acceptance criteria for Preservative Effectiveness Testing as outlined in ISO 14730:2000(E), "Ophthalmic optics – Contact lens care products - Antimicrobial preservative efficacy testing and guidance on determining discard date." | The product meets the acceptance criteria for Preservative Effectiveness Testing as outlined in ISO 14730:2000(E). |
Microbiological Studies (Sterility) | Meets USP Sterility test requirements. | The product meets USP Sterility test requirements. |
Stability | Remains stable for the labeled shelf-life based on accelerated testing. | Accelerated testing predicts that the product will remain stable for the labeled shelf-life. |
Toxicology | Non-cytotoxic, non-irritating, and well-tolerated. | blink™ CL Lubricant Eye Drops is non-cytotoxic, non-irritating and well-tolerated. |
Clinical Studies | ||
Safety and Acceptability | Safe, acceptable, and substantially equivalent to the control. | The investigational formulation is safe, acceptable, and substantially equivalent to the control. |
2. Sample Size and Data Provenance
- Test Set Sample Size: Not explicitly stated for each non-clinical test. The clinical study was a "multi-center" and "parallel-group" evaluation, but specific patient numbers are not provided.
- Data Provenance: The document does not specify the country of origin for the data. The clinical study was conducted by AMO (Advanced Medical Optics). No information is provided on whether it was retrospective or prospective, though a "multi-center, double-masked, randomized, parallel-group, one-month evaluation" suggests a prospective design.
3. Number of Experts and Qualifications
- No information provided regarding the number or qualifications of experts used to establish ground truth for the test set in either non-clinical or clinical studies. The clinical study was "double-masked," implying that both participants and evaluators were unaware of treatment assignments, but it does not specify who established the clinical endpoints or evaluated outcomes for ground truth.
4. Adjudication Method
- No information provided on adjudication methods for establishing ground truth in either non-clinical or clinical studies.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. The studies described are primarily aimed at demonstrating the equivalence of the device to predicate devices, not at evaluating the improvement of human readers with or without AI assistance. The device is a lubricant eye drop, not an AI-powered diagnostic or assistive tool.
6. Standalone (Algorithm Only) Performance
- Not applicable. The device is a lubricant eye drop and does not involve an algorithm or AI component.
7. Type of Ground Truth Used
- Nonclinical Studies: Ground truth was established through in-vitro laboratory testing and adherence to recognized standards (e.g., ISO 14730:2000(E) for preservative efficacy, USP Sterility test requirements). The comparison was often against "control" or "predicate" devices, meaning the performance of an established, similar product served as the reference.
- Clinical Studies: Ground truth for safety and acceptability would have been based on clinical evaluations and observations, likely including patient reported outcomes and objective clinical assessments, compared against a "control" in a "double-masked" study.
8. Sample Size for the Training Set
- Not applicable. This device is a pharmaceutical product (lubricant eye drops), not an AI/ML model that requires training data.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(84 days)
ADVANCED MEDICAL OPTICS, INC.
COMPLETE® brand Multi-Purpose Solution is indicated for the care of soft (hydrophilic) contact lenses. Use this product, as recommended by your eye care practitioner, to:
- Chemically (NOT HEAT) Disinfect .
- Clean ●
- Rinse .
- Store
- Remove Protein .
- Condition .
COMPLETE® brand MULTI-PURPOSE Solution is a sterile, isotonic, buffered, solution containing hydroxypropyl methylcellulose as a lubricant, preserved with polyhexamethylene biguanide 0.0001%, a phosphate buffer, Poloxamer 237 as a surfactant, edetate disodium as a chelating agent, sodium chloride, potassium chloride, and purified water. COMPLETE® brand MULTI-PURPOSE Solution contains no chlorhexidine or thimerosal.
The product is a clear, colorless solution packaged in plastic bottles with controlled dropper tips.
The provided text is a 510(k) summary for COMPLETE® brand Multi-Purpose Solution (K030092). This document focuses on demonstrating substantial equivalence to a predicate device for a contact lens care solution, rather than providing detailed acceptance criteria and a study design for evaluating a novel medical device's performance against specific metrics. Therefore, many of the requested categories are not directly applicable or explicitly stated in the document.
However, based on the information provided, here's an attempt to address your request:
Acceptance Criteria and Device Performance for COMPLETE® brand Multi-Purpose Solution (K030092)
Note: The document describes in-vitro and clinical studies to support substantial equivalence for a contact lens solution, not a medical imaging or diagnostic device. As such, the typical metrics for device performance (e.g., sensitivity, specificity, accuracy) are not relevant here. Instead, the focus is on physical, chemical, and subjective user experience.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria (Inferred from Study Goals) | Reported Device Performance (COMPLETE® brand Multi-Purpose Solution) |
---|---|---|
In-Vitro - Physical Properties | Viscosity Equivalence: Viscosity of COMPLETE® Solution with HPMC should be comparable to original COMPLETE® without HPMC and other multipurpose solutions. | "In Vitro Studies were conducted to compare the physical and chemical properties of COMPLETE® Solution with other multi-purpose solutions in order to evaluate the effects of adding hydroxypropyl methylcellulose (HPMC)." (General statement; no specific numeric values are provided in this summary.) |
In-Vitro - Lens Interaction | HPMC Uptake & Release: Demonstrate uptake of HPMC onto the lens surface during soak and subsequent release in saline over time. | "The studies show uptake of HPMC onto the lens surface during the soak period, and its subsequent release, over time, in saline. This layer of adsorbed HPMC appears to enhance the lens' ability to retain moisture and remain more wettable." (Positive finding, indicating the desired mechanism of action.) |
In-Vitro - Surface Wetting | Surface Fluid Retention: Comparable effect on surface fluid retention compared to B&L's ReNu® MultiPlus® and Alcon's OPTI-FREE® Express. | "Compare the effect of three COMPLETE® Solution formulations to B&L's ReNu® MultiPlus® and Alcon's OPTI-FREE® Express on surface fluid retention." (Study was conducted, results are implied to be satisfactory for substantial equivalence, but specific comparative data is not provided.) |
In-Vitro - Wetting Angle | Receding Contact Angle: Comparable effect on receding contact angle (as a measure of wetting) to various multipurpose solutions. | "Examine the effect of various multipurpose solutions on receding contact angle as a measure of wetting." (Study was conducted, results are implied to be satisfactory for substantial equivalence, but specific comparative data is not provided.) |
Clinical - Subjective Comfort | Comfort & Acceptability Equivalence: Subjective comfort and acceptability of COMPLETE® with HPMC should be substantially equivalent to original COMPLETE® without HPMC and competitor regimens. | "This subjective data shows that COMPLETE® Solution, used with a rub regimen, has comfort substantially equivalent to other marketed contact lens care multipurpose solutions." (Positive finding, demonstrating equivalence in user experience.) |
Clinical - Rub vs. No-Rub Efficacy | Functional Equivalence (Rub vs. No-Rub): COMPLETE® Solution used in a "no-rub" regimen should demonstrate substantial equivalence to its use in a "rub" regimen. | "Objective data from an additional clinical study demonstrate the substantial equivalence of COMPLETE® Solution used in a 'no-rub' regimen, with COMPLETE® Solution used in a 'rub' regimen." (Positive finding, supporting the flexibility of use with equivalent performance.) |
2. Sample Size and Data Provenance for Test Set
- Clinical Studies: "Study subjects in three clinical studies" were used for subjective comfort and acceptability. An "additional clinical study" provided objective data for rub vs. no-rub regimens.
- Sample Size: Not explicitly stated. The number of participants in each study is not provided.
- Data Provenance: Not explicitly stated. However, given the context of a US regulatory submission (510(k)), it is highly probable the data originated from studies conducted in the United States and were prospective in nature, as they were designed to evaluate the new formulation.
3. Number of Experts and Qualifications for Ground Truth
- Not applicable directly. This is a contact lens solution, not an interpretive device. The "ground truth" for subjective comfort comes from the study subjects themselves via questionnaires. For objective data (rub vs. no-rub), unspecified "objective data" was collected, likely related to lens cleanliness or disinfection efficacy, which would be measured by laboratory techniques rather than expert consensus on interpretation.
4. Adjudication Method for Test Set
- Not applicable. There is no "adjudication" in the sense of multiple experts independently reviewing and then resolving discrepancies for medical image interpretation. Clinical subjective data was gathered via questionnaires, and objective data via laboratory methods.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No. This type of study is relevant for diagnostic imaging systems where human readers interpret cases. The provided document concerns a contact lens solution.
6. Standalone (Algorithm Only) Performance
- Not applicable. The device is a chemical solution, not an algorithm.
7. Type of Ground Truth Used
- Clinical Studies:
- Subjective Data: User-reported comfort and acceptability via questionnaires.
- Objective Data: Unspecified objective measures related to contact lens care performance (e.g., disinfection, cleaning, protein removal efficacy) for the rub vs. no-rub regimen comparison. This would typically be laboratory-based measurements.
- In Vitro Studies: Laboratory measurements of physical (viscosity, surface fluid retention, receding contact angle) and chemical (HPMC uptake/release) properties.
8. Sample Size for Training Set
- Not applicable. This isn't a machine learning model, so there isn't a "training set" in that sense. The studies described are for validation of the product's performance and equivalence.
9. How Ground Truth for Training Set was Established
- Not applicable. See point 8.
Ask a specific question about this device
Page 1 of 1