Search Results
Found 2 results
510(k) Data Aggregation
(74 days)
ADVANCED BIO-SURFACES, INC.
The OrthoGlide Lateral Knee Implant is intended for use in the osteoarthritic knee, where substantial amounts of cartilage have been lost as a result of the disease. The device is indicated for uncemented use in the treatment of moderate degeneration of the lateral compartment of the knee (grade II-IV chondromalacia) with no more than minimal degeneration (grade I-II chondromalacia, no loss of joint space) in the medial and patellofemoral compartments in patients with osteoarthritis.
The OrthoGlide Lateral Knee Implant is placed in the lateral compartment of the knee between the tibial plateau and femoral condyle by means of minimally invasive surgery. The instruments are intended to aid in the surgical preparation of the implant site and implant placement. The OrthoGlide is made of a Cobalt-Chrome-Molybdenum Alloy. Device geometry and ligament tension combine to keep the implant in place. The implant covers the lateral tibial plateau. The device is designed to improve the alignment of the knee, returning the joint to a less valgus position. Realignment of the knee distributes the weight-bearing forces across the joint and helps restore the normal relationships of the articular surfaces and the surrounding capsular, ligamentous and muscular structures. The device is designed to help relieve pain by providing an articulating surface with a low coefficient of friction and high durability. Device geometry improves knee alignment and joint spacing. The device surface is smooth and when wet, is intended to mimic the lubricious surface previously provided by the articular cartilage.
The advanced Bio-Surfaces, Inc. OrthoGlide® Lateral Knee Implant did not undergo clinical studies with acceptance criteria and measured performance in the traditional sense as this was a 510(k) premarket notification for a Class II medical device. The device was found substantially equivalent to predicate devices based on non-clinical testing.
Here's a breakdown of the information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Since this was a 510(k) submission based on substantial equivalence to predicate devices, there isn't a direct "acceptance criteria" table with measured performance like in a clinical trial. Instead, the "acceptance criteria" revolved around demonstrating that the device performed similarly to the predicates and met established standards through bench and cadaver testing.
Acceptance Criteria Category | Reported Device Performance (Summary from provided text) |
---|---|
Physical Attributes & Durability (Bench Testing) | Demonstrated through tensile elongation, load deflection, cyclic fatigue resistance, material consistency and stability, and processing control. |
Surgical Technique & Instrument Verification (Cadaver Testing) | Verified the surgical technique and instruments. |
Physiological Effects (Cadaver Testing) | Verified desired physiological effects of stability and range of motion preservation. |
Biocompatibility | Meets internationally recognized standards for biocompatibility (ISO 10993). |
Sterility | Meets internationally recognized standards for sterility (EN550). |
Material Specifications | Conforms to material specifications. |
Substantial Equivalence to Predicate Devices | Substantially equivalent to FDA approved predicate devices (K003269, K033242, K053094) regarding indications for use, materials, technological characteristics, and surgical techniques, offering no additional risks or safety concerns. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set:
- Bench Testing: Not specified in terms of number of samples, but implied to be sufficient to demonstrate the listed physical attributes and durability.
- Cadaver Testing: Not specified in terms of the number of cadavers used.
- Data Provenance: The data is from non-clinical testing (bench and cadaver studies) conducted by Advanced Bio-Surfaces, Inc. The document does not specify the country of origin for the cadavers or the exact location of the bench testing facilities, but the manufacturer is based in Minnetonka, MN, USA. The studies are by nature prospective for the device's design verification and validation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This type of information is generally not applicable to non-clinical bench and cadaver testing for medical devices, particularly for a 510(k) submission not involving clinical data. The "ground truth" for bench testing would be defined by the specifications and standards (e.g., ISO, EN) the device is tested against. For cadaver testing, the "ground truth" would be the successful execution of the surgical technique and observation of desired physiological effects by medical professionals/surgeons involved in the cadaver labs.
The document does not mention the number or qualifications of experts involved in the cadaver evaluations.
4. Adjudication Method for the Test Set
Not applicable, as this was not a clinical study requiring adjudication of expert opinions on patient outcomes. The evaluation of bench and cadaver tests would be based on objective measurements and observations against pre-defined criteria.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical testing was not used to determine substantial equivalence." and "The OrthoGlide as well as the predicates are submitted without clinical information."
- Effect Size: Not applicable, as no MRMC study was conducted.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. The OrthoGlide® Lateral Knee Implant is a physical implant, not an AI algorithm. Therefore, "standalone" performance in the context of AI is not relevant. The performance of the implant itself is assessed through the bench and cadaver testing described.
7. Type of Ground Truth Used
- Bench Testing: Engineering specifications, material standards (e.g., tensile strength, fatigue limits), and international voluntary standards (e.g., ISO 10993 for biocompatibility, EN550 for sterility).
- Cadaver Testing: Direct observation and measurement of surgical technique execution, instrument fit, knee stability, and range of motion on cadaveric specimens, likely by surgical experts or engineers.
8. Sample Size for the Training Set
Not applicable. As a physical implant, there is no "training set" in the context of machine learning or AI. The design and manufacturing processes are refined through engineering development and testing.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of medical device.
Ask a specific question about this device
(95 days)
ADVANCED BIO-SURFACES, INC.
The Orthoglide Medial Knee Implant is intended for use in the osteoarthritic knee, where substantial amounts of cartilage have been lost as a result of the disease. The device is indicated for uncemented use in the treatment of moderate degeneration of the medial compartment of the knee (grade II-IV chondromalacia) with no more than minimal degencration (grade I-II chondromalacia, no loss of joint space) in the lateral and patellofemoral compartments in patients with ostcoarthritis.
The Orthoglide Medial Knee Implant is placed in the medial compartment of the knee between the tibial plateau and femoral condyle by means of minimally invasive surgery. The instruments are intended to aid in the surgical preparation of the implant site and implant placement. The Orthoglide is made of a Cobalt-Chrome-Molybdenum Alloy. Device geometry and ligament tension combine to keep the implant in place. The implant covers the tibial plateau. The device is designed to improve the alignment of the knee, returning the joint to a more valgus position. Realignment of the knee distributes the weight-bearing forces across the joint and helps restore the normal relationships of the articular surfaces and the surrounding capsular, ligamentous and muscular structures.
The device is designed to help relieve pain by providing an articulating surface with a low coefficient of friction and high durability. Device geometry improves knee alignment and joint spacing. The device surface is smooth and when wet, is intended to mimic the lubricious surface previously provided by the articular cartilage.
This 510(k) submission for the Orthoglide® Medial Knee Implant does not contain a study demonstrating that the device meets specific acceptance criteria in the manner typically seen for AI/ML-based medical devices or diagnostics. Instead, it demonstrates substantial equivalence to predicate devices through a combination of engineering bench testing and non-clinical evaluations.
The document states: "The Orthoglide as well as the predicates are submitted without clinical information." This explicitly indicates that no clinical studies, which would typically involve human subjects and outcome measures to establish specific acceptance criteria and performance against them, were conducted for this submission.
Therefore, many of the requested elements for AI/ML device studies are not applicable in this context. I will address the parts that are relevant based on the provided text.
Description of Acceptance Criteria and Study for Orthoglide® Medial Knee Implant
The Orthoglide® Medial Knee Implant's acceptance criteria are framed around demonstrating substantial equivalence to legally marketed predicate devices, rather than achieving specific performance metrics in a clinical study. The "study" in this context refers to a series of bench tests and cadaver evaluations designed to ensure the device's physical attributes, durability, material consistency, and surgical compatibility, mirroring what would be expected of the predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a substantial equivalence submission based on non-clinical data, the acceptance criteria are not in the form of diagnostic performance metrics (e.g., sensitivity, specificity). Instead, they relate to engineering and material properties, and the "reported performance" is that the device meets these criteria.
Acceptance Criteria Category | Specific Criteria (Implicit or Explicit in Document) | Reported Device Performance (Summary) |
---|---|---|
Mechanical/Physical | Tensile elongation | Demonstrated adherence to specifications |
Load deflection | Demonstrated adherence to specifications | |
Cyclic fatigue resistance | Demonstrated adherence to specifications | |
Material consistency and stability | Demonstrated adherence to specifications | |
Processing control | Demonstrated adherence to specifications | |
Biocompatibility | Conformance to ISO 10993 | Meets ISO 10993 standards |
Sterility | Conformance to EN550 | Meets EN550 standards |
Material Specifications | Conformance to material specifications | Meets material specifications (Cobalt-Chrome-Molybdenum Alloy) |
Surgical Technique | Verification of surgical technique (cadaver) | Verified surgical technique, instruments |
Physiological Effects (Cadaver) | Stability, angular correction, range of motion preservation | Verified desired physiological effects |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not applicable. The "test set" here refers to the samples used in bench testing and cadaver evaluations, not a clinical patient cohort. The document does not specify the number of individual components or cadavers used for these tests.
- Data Provenance: The tests were conducted internally by Advanced Bio-Surfaces, Inc. as part of their design verification and validation process. The origin of the raw materials (Cobalt-Chrome-Molybdenum Alloy) would typically be from suppliers, but specific "country of origin" for the test data itself is not stated beyond the manufacturer's location in Minnetonka, MN, USA. All data appears to be prospective in the sense that it was generated specifically for this device's development and submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This concept is not applicable in the context of this 510(k) submission. "Ground truth" for an AI/ML device typically involves expert interpretation of medical images or clinical data. Here, the "truth" is established by adherence to engineering standards, material specifications, and observed physical behaviors in controlled lab and cadaveric settings. There's no mention of a panel of experts establishing ground truth for diagnostic or prognostic interpretations.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods (e.g., 2+1, 3+1) are used for resolving discrepancies in expert interpretations to establish ground truth in clinical/diagnostic studies (typically AI/ML). The "truth" in this submission relies on objective measurements from bench testing and observations during cadaveric procedures, not on expert consensus that requires adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. Such studies are relevant for evaluating the impact of an AI algorithm on human reader performance, which is not pertinent to this implant device's submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done
Yes, in a sense. The "standalone" performance here refers to the inherent physical and mechanical properties of the device itself, as evaluated through bench testing and cadaver studies. This is the primary mechanism used to demonstrate substantial equivalence, rather than a clinical study with human patients. The device's performance (e.g., fatigue resistance, biocompatibility) was assessed independently of human interaction beyond the necessary execution of the tests themselves.
7. The Type of Ground Truth Used
The "ground truth" for this submission consists of:
- Engineering Standards and Specifications: Adherence to established and voluntary standards (e.g., ISO 10993 for biocompatibility, EN550 for sterility) and the manufacturer's own design specifications for mechanical properties, materials, and processing.
- Observed Physical Properties: Measurements from tensile strength tests, load deflection, and cyclic fatigue, compared against predetermined engineering requirements.
- Cadaveric Observations: Direct observation and measurement during cadaver testing to verify surgical technique, instrument function, and physiological effects like stability, angular correction, and range of motion preservation.
8. The Sample Size for the Training Set
Not applicable. This device is a mechanical implant, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1