EQ-PRC LC WARFARIN GENOTYPING KIT

K073071 · Trimgen Corporation · ODW · Feb 6, 2009 · Clinical Toxicology

Device Facts

Record IDK073071
Device NameEQ-PRC LC WARFARIN GENOTYPING KIT
ApplicantTrimgen Corporation
Product CodeODW · Clinical Toxicology
Decision DateFeb 6, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3360
Device ClassClass 2

Indications for Use

The eQ-PCR™ LC Warfarin Genotyping kit is an in vitro diagnostic test for the detection and genotyping of two single nucleotide polymorphisms (SNP) in the cytochrome P450 enzyme gene CYP2C9 known as CYP2C9*2 (C430T) and CYP2C9*3 (A1075C), and a SNP in the vitamin K epoxide reductase complex 1 gene VKORC1, known as VKORC1 (–1639G>A) obtained from EDTA – anticoagulated whole blood samples. The eQ-PCR LC Warfarin Genotyping kit is used as an aid in the identification of patients at risk for increased warfarin sensitivity. It is a qualitative assay for use in clinical laboratories upon prescription by the attending physician.

Device Story

In vitro diagnostic kit; detects CYP2C9*2, CYP2C9*3, and VKORC1 (-1639G>A) SNPs. Input: purified DNA from EDTA-anticoagulated whole blood. Operation: real-time PCR amplification followed by melting curve analysis on Roche LightCycler (model 1.2). Principle: FRET-based fluorescence detection using anchor and SNP-specific probes; melting temperature (Tm) discrimination of alleles. Output: qualitative genotype call. Used in clinical laboratories by trained personnel. Results aid physicians in identifying patients at risk for increased warfarin sensitivity, informing clinical decision-making regarding warfarin dosing.

Clinical Evidence

No clinical trials performed. Analytical performance established via bench testing. Accuracy validated against bi-directional sequencing using 159 whole blood samples across three sites. Overall correct call rate was 97.7% (95% CI: 96.18%). Reproducibility study (31 samples, 3 sites, 5 days) showed 98.7% correct call rate. Interference testing confirmed no impact from albumin, bilirubin, triglycerides, heparin, hemoglobin, or common extraction buffers.

Technological Characteristics

Real-time PCR genotyping assay. Components: PCR mix (DNA polymerase, dNTPs), probe mixes (primers/probes for CYP2C9*2, *3, and VKORC1). Detection: FRET-based melting curve analysis. Instrument: Roche LightCycler 1.2. Sample input: 50-200 ng genomic DNA. Storage: -20°C. Software: Automated melting curve-based SNP calling.

Indications for Use

Indicated for patients requiring warfarin therapy to identify risk of increased warfarin sensitivity by genotyping CYP2C9*2, CYP2C9*3, and VKORC1 (-1639G>A) variants from EDTA-anticoagulated whole blood.

Regulatory Classification

Identification

A drug metabolizing enzyme genotyping system is a device intended for use in testing deoxyribonucleic acid (DNA) extracted from clinical samples to identify the presence or absence of human genotypic markers encoding a drug metabolizing enzyme. This device is used as an aid in determining treatment choice and individualizing treatment dose for therapeutics that are metabolized primarily by the specific enzyme about which the system provides genotypic information.

Special Controls

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping System."

*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping Test System.” See § 862.1(d) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k073071 B. Purpose for Submission: New device C. Measurand: Genotype of Cytochrome P450 2C9 (CYP450 2C9) and Vitamin K epoxide reductase complex subunit 1 (VKORC1) D. Type of Test: Genotyping Realtime PCR E. Applicant: TrimGen Corporation F. Proprietary and Established Names: eQ-PCR™ LC Warfarin Genotyping kit G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | Cytochrome P450 2C9 (CYP450 2C9) Drug Metabolizing Enzyme Genotyping System (ODW) | Class II | 21 CFR 862.3360 Drug metabolizing enzyme genotyping system | Toxicology (91) | | Product Code | Classification | Regulation Section | Panel | | INSTRUMENTATION FOR CLINICAL MULTIPLEX TEST SYSTEMS (NSU) | Class II | 21 CFR 862.2570 - INSTRUMENTATION FOR CLINICAL MULTIPLEX TEST SYSTEMS | CLINICAL CHEMISTRY | | VITAMIN K EPOXIDE REDUCTASE COMPLEX SUBUNIT ONE (VKORC1) GENOTYPING SYSTEM (ODV) | Class II | 21 CFR 864.7750 - PROTHROMBIN TIME TEST | HEMATOLOGY | {1} H. Intended Use: 1. Intended use(s): Refer to Indications for use below. 2. Indication(s) for use: The eQ-PCR™ LC Warfarin Genotyping kit is an in vitro diagnostic test for the detection and genotyping of two single nucleotide polymorphisms (SNP) in the cytochrome P450 enzyme gene CYP2C9 known as CYP2C9*2 (C430T) and CYP2C9*3 (A1075C), and a SNP in the vitamin K epoxide reductase complex 1 gene VKORC1, known as VKORC1 (–1639G>A) obtained from EDTA – anticoagulated whole blood samples. The eQ-PCR LC Warfarin Genotyping kit is used as an aid in the identification of patients at risk for increased warfarin sensitivity. It is a qualitative assay for use in clinical laboratories upon prescription by the attending physician. 3. Special conditions for use statement(s): For Prescription use only 4. Special instrument requirements: For use with LightCycler® Real-Time PCR System instrument model 1.2. I. Device Description: eQ-PCR™ LC Warfarin Genotyping Kit reagents The reagents in the eQ-PCR™ LC Warfarin Genotyping Kit are designed to be used specifically with the Roche LightCycler®. They are provided for 32 tests per kit and include: - The PCR Mix is a pre-mixed reagent for PCR amplification including a DNA polymerase, PCR Buffer and dNTPs. - The LC1 is the Probe Mix 1 containing primer and probes designed to detect the 2C9*3 (A1075C) SNP in the CYP2C9 gene and the (–1639 G>A) SNP in the VKORC1 gene. - The LC2 is the Probe Mix 2 containing primers and probes designed to detect the 2C9*2 (C430T) SNP in the CYP2C9 gene. - The NF Water is nuclease-free water. J. Substantial Equivalence Information: 1. Predicate device name(s): Verigene® Warfarin Metabolism Nucleic Acid Test 2. Predicate 510(k) number(s): k070804 3. Comparison with predicate: The following table summarizes the similarities and differences between the new and predicate devices. {2} Similarities: | Characteristics | Nanosphere Warfarin Assay | eQ-PCR™ LC Warfarin Genotyping Kit | | --- | --- | --- | | | Predicate device (K070804) | Proposed device | | Instrument | Verigene system with Verigene cartridge | Roche LightCycler with glass capillary tubes | | Method to determine specificity | Hybridization | PCR Hybridization Melting Curve Analysis | | Assay conditions | A closed system | A closed tube assay | | Procedures | Reaction and SNP analysis require separate operations (transfer cartridge to reader) | PCR and SNP analysis in same tube (no transfer steps) | Differences: | Gene | CYP2C9 and VKORC1 | CYP2C9 and VKORC1 | | --- | --- | --- | | Specimen type | Purified DNA from human whole blood samples | Purified DNA from human whole blood samples | # K. Standard/Guidance Document Referenced (if applicable): Drug Metabolizing Enzyme Genotyping System - Class II Special Controls Guidance Document # L. Test Principle: The eQ-PCR™ LC Warfarin Genotyping Assay uses real time PCR technology to amplify and detect the target DNA. The extracted DNA sample is mixed with a PCR Mix, and an eQ-PCR™ specific Probe Mix reagent containing specific primers and fluorescent labeled probes for the CYP2C9 and/or VKORC1 gene polymorphisms. Amplification and detection are then performed in the Roche Diagnostics LightCycler® Real-Time PCR System instrument model 1.2 using conditions defined in the specific eQ-PCR™ LC Warfarin Genotyping Kit labeling. After the PCR reaction is completed, the LightCycler System automatically proceeds to the melting curve-based detection method. The signal detection is based on fluorescence resonance energy transfer (FRET) technology. Two probes are designed for each SNP {3} and labeled with different fluorophores. The anchor probe selectively hybridizes to the sequence flanking the SNP site, and the SNP detection probe recognizes the SNP and hybridizes to the sequence containing the SNP site. During FRET, the fluorophore on the anchor probe is excited by the light source of the LightCycler and the excitation energy is transferred to the fluorophore on the SNP detection probe. The emitted fluorescence is then measured at the respective wavelength. The SNP detection is performed by melting curve analysis, a method that discriminates different allelic forms of DNA by melting temperature. The SNP detection probe is designed to have a unique melting temperature to the wild type and mutant SNP. With change in temperature, the probe dissociates from the target sequence at a specific melting temperature depending on the allelic composition of SNP. The SNP type is then detected based on the melting curve specific to each SNP. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: To evaluate the within-laboratory, between operator, and between laboratory reproducibility of the eQ-PCR™ LC Warfarin Genotyping Kit on the Roche LightCycler®, three sites (two external and the sponsor's laboratory) and five operators participated. A panel of 31 samples was tested at each of the three sites on each of five days. At the two external sites, two operators per site performed the testing on alternate days. The panel consisted of whole blood (WB) samples collected in EDTA from 20 donors, and 9 genomic DNA samples (cctl below). These samples represented all the genotypes tested by this device. Each site used a different DNA extraction method for isolating DNA from whole blood samples. Each site then tested the resulting extracted nucleic acids using the eQ-PCR™ LC Warfarin Genotyping Kit on the Roche LightCycler®. Genotypes for each sample were determined by bi-directional sequencing and compared with the results from the device. The composition of the reproducibility panel with the call rate for all three sites combined for each sample is described in Table below. There were 6 no calls, but no incorrect calls reported. The entire study produced a correct call rate of $98.7\%$ . Genotypic composition and percent call rate per sample: | Sample ID | Genotype | | | # samples tested | # genotype calls made | # correct calls | # no calls | # incorrect calls | Correct call rate % | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | 2C9*2 | 2C9*3 | VKORC1 | | | | | | | | 1 | wt/wt | wt/wt | GG | 15 | 15 | 15 | 0 | 0 | 100 | | 2 | wt/wt | wt/*3 | GG | 15 | 14 | 14 | 1 | 0 | 93.3 | | 3 | wt/wt | wt/wt | GA | 15 | 15 | 15 | 0 | 0 | 100 | | 4 | wt/wt | wt/wt | GG | 15 | 15 | 15 | 0 | 0 | 100 | | 5 | wt/wt | wt/wt | GG | 15 | 15 | 15 | 0 | 0 | 100 | | 6 | wt/wt | wt/wt | GA | 15 | 14 | 14 | 1 | 0 | 93.3 | | 7 | wt/wt | wt/wt | GA | 15 | 13 | 13 | 2 | 0 | 86.7 | | 8 | wt/*2 | wt/wt | GA | 15 | 14 | 14 | 1 | 0 | 93.3 | {4} | 9 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | 10 | wt/wt | wt/wt | GA | 15 | 15 | 15 | 0 | 0 | 100 | | 11 | wt/wt | wt/wt | GG | 15 | 15 | 15 | 0 | 0 | 93.3 | | 12 | wt/wt | wt/wt | GA | 15 | 15 | 15 | 0 | 0 | 100 | | 13 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | 14 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | 15 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | 16 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | 17 | wt/*2 | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | 18 | wt/*2 | wt/wt | GA | 15 | 15 | 15 | 0 | 0 | 100 | | 19 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | 20 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | 21 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | 22 | wt/*2 | wt/wt | GA | 15 | 14 | 14 | 1 | 0 | 93.3 | | cctl-1 | wt/wt | wt/wt | GG | 15 | 15 | 15 | 0 | 0 | 100 | | cctl-2 | wt/wt | wt/wt | GA | 15 | 15 | 15 | 0 | 0 | 100 | | cctl-3 | wt/wt | wt/wt | AA | 15 | 15 | 15 | 0 | 0 | 100 | | cctl-4 | wt/wt | wt/wt | GG | 15 | 15 | 15 | 0 | 0 | 100 | | cctl-5 | wt/*2 | wt/wt | GG | 15 | 15 | 15 | 0 | 0 | 100 | | cctl-6 | *2/*2 | wt/wt | GA | 15 | 15 | 15 | 0 | 0 | 100 | | cctl-7 | wt/wt | wt/wt | GG | 15 | 15 | 15 | 0 | 0 | 100 | | cctl-8 | wt/wt | wt/*3 | GG | 15 | 15 | 15 | 0 | 0 | 100 | | cctl-9 | wt/wt | *3/*3 | GA | 15 | 15 | 15 | 0 | 0 | 100 | | Total | | | | 465 | 459 | 459 | 6 | 0 | 98.7 | Reproducibility per genotype: | Genotype | # samples tested£ | # correct calls | # incorrect calls | # no calls | Correct call rate % | | --- | --- | --- | --- | --- | --- | | 2C9*2 | 465 | 461 | 0 | 4 | 99.1 | | 2C9*3 | 465 | 464 | 0 | 1 | 99.7 | | VKORC1 | 465 | 464 | 0 | 1 | 99.7 | £ Total tests for each genotype: 31 samples x 3 sites x 5 days repeats = 465 tests {5} Reproducibility per site per Operator: | Site/Operator | Locus | # Samples | # correct calls | # incorrect calls | # no calls | Correct call rate % | | --- | --- | --- | --- | --- | --- | --- | | Site 1/Operator 1 | 2C9*2 | 93 | 91 | 0 | 2 | 97.85 | | | 2C9*3 | 93 | 93 | 0 | 0 | 100 | | | VKORC1 | 93 | 93 | 0 | 0 | 100 | | Site 1/Operator 2 | 2C9*2 | 62 | 62 | 0 | 0 | 100 | | | 2C9*3 | 62 | 62 | 0 | 0 | 100 | | | VKORC1 | 62 | 62 | 0 | 0 | 100 | | Site 2/Operator 3 | 2C9*2 | 93 | 93 | 0 | 0 | 100 | | | 2C9*3 | 93 | 92 | 0 | 1 | 98.9 | | | VKORC1 | 93 | 92 | 0 | 1 | 98.9 | | Site 2/Operator 4 | 2C9*2 | 62 | 60 | 0 | 2 | 96.8 | | | 2C9*3 | 62 | 62 | 0 | 0 | 100 | | | VKORC1 | 62 | 62 | 0 | 0 | 100 | | Site 3/Operator 5 | 2C9*2 | 155 | 155 | 0 | 0 | 100 | | | 2C9*3 | 155 | 155 | 0 | 0 | 100 | | | VKORC1 | 155 | 155 | 0 | 0 | 100 | ## Lot to lot reproducibility results A study was conducted at the sponsor's facility to determine the reproducibility between kit reagent lots of the eQ-PCR™ LC Warfarin Genotyping assay. A panel of 24 whole blood samples was tested using three reagent lots for five days. The results per genotype for each lot are illustrated in the table below. | Loci | Lot | # samples tested | # tests$ | Genotyping calls made | # correct calls | # incorrect calls | # no calls | Correct call rate % | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | 2C9 *2 | Lot 1 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | | | Lot 2 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | | | Lot 3 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | | 2C9 *3 | Lot 1 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | | | Lot 2 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | | | Lot 3 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | | VKORC1 | Lot 1 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | | | Lot 2 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | | | Lot 3 | 24 | 120 | 120 | 120 | 0 | 0 | 100% | $Total tests for each Lot: 24 samples x 5 repeats = 120 tests ## b. Linearity/assay reportable range: Not Applicable {6} c. Traceability, Stability, Expected values (controls, calibrators, or methods): The sponsor conducted stability studies for DNA stored at $-20^{\circ}\mathrm{C}$ and $2 - 8^{\circ}\mathrm{C}$ using 9 commercially available genomic DNA samples with 3 lots of the reagent kit. Stability studies for DNA stored at $-20^{\circ}\mathrm{C}$ were conducted up to one year at 3-month intervals and for DNA stored at $-2 - 8^{\circ}\mathrm{C}$ for 30 days. Results demonstrated consistent accurate results. The sponsor is continuing the stability studies for DNA stored at $-20^{\circ}\mathrm{C}$ studies for 2 yrs. However, the sponsor recommends using DNA from fresh whole blood samples for genotyping assay. Reagents for the Warfarin Genotyping Kit are shipped to the user laboratory on dry ice. The sponsor recommends storing all reagents at $-20^{\circ}\mathrm{C}$ in dark. The kit is stable at $-20^{\circ}\mathrm{C}$ for one year. After first use, all reagents should be stored at $2 - 8^{\circ}\mathrm{C}$ in dark. The sponsor states that under these conditions, reagents are suitable for use up to three months. Controls: The sponsor does not provide positive or negative controls to be used with this assay. However, the sponsor has recommended that a positive (heterozygous and/or homozygous for the three genotypes) sample for each variant, a negative control (a sample that does not contain the mutation of interest, i.e., a wild type sample); and a "Non-Template Control" be included with each test run. All quality control requirements and testing should be performed in conformance with local, state and/or federal regulations. The sponsor conducted preliminary studies to establish the Melting Temperature (Tm) cut-off values to identify the presence of individual SNP positions by the eQ-PCR™ LC Warfarin Genotyping Kit. A total of 64 individual samples were tested to evaluate the melting temperature Cut-Off value for each SNP type. The average melting temperature for each SNP was calculated and the Cut-Off value was determined based on Average value ± 3 SD. The SNP result for each sample was confirmed by bidirectional sequencing. Results are listed in the table below. Tm values obtained for each genotype: | | VKORC1 | | 2C9*2 | | 2C9*3 | | | --- | --- | --- | --- | --- | --- | --- | | | Tm Low (variant) | Tm High (Wt) | Tm Low (variant) | Tm High (Wt) | Tm Low (Wt) | Tm High (variant) | | Min: | 53.31 | 62.36 | 54.52 | 63.47 | 53.26 | 60.32 | | Max: | 55.92 | 64.52 | 58.44 | 66.74 | 56.49 | 63.16 | | Avg. | 54.94 | 63.57 | 57.34 | 65.58 | 55.23 | 61.89 | | SD | 0.80 | 0.65 | 0.90 | 0.59 | 0.90 | 0.86 | The established Cut-Off value for each SNP: | SNP ID | Melting temperature for Wild type | Melting temperature for variant | | --- | --- | --- | | VKORC1 | 63.57 ± 1.95oC | 54.94 ± 2.40oC | | 2C9*2 | 65.58 ± 1.77oC | 57.34 ± 2.70oC | | 2C9*3 | 55.23 ± 2.70oC | 61.89 ± 2.58oC | {7} d. Limit of Detection: A study was conducted at the sponsor's facility to validate the recommended amounts of DNA required for genotyping test using eQ-PCR™ LC Warfarin Genotyping test. Six human genomic DNA samples at 7 different concentrations to obtain 0.1, 1, 10, 50, 100, 200 and 600 ng in the final reaction tube were used and 26 repeats conducted for each test level (total of 3312 tests). These six genomic DNA samples covered all 9 possible 2C9*2, 2C9*3, and VKORC-1 genotypes. Two of the six human genomic DNA samples were tested in two additional concentrations (300 ng & 400 ng) and repeated three times. Based on the results, there were two "no calls" at 0.1 ng DNA. The results of DNA limitation study are given in the tables below. For eQ-PCR™ LC Warfarin Genotyping assay, the sponsor recommends using a total of 50 – 200 ng genomic DNA in the final reaction tube. Call rate based on DNA amount: | Input DNA Amount (ng) | Genotype | # samples | # repeats | # tests | # genotype calls | # correct calls | # incorrect calls | # no calls | Correct call rate (%) | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | 0.1 | 2C9*2 | 6 | 26 | 156 | 468 | 466 | 0 | 2 | 99.6 | | | 2C9*3 | 6 | 26 | 156 | | | | | | | | VKORC1 | 6 | 26 | 156 | | | | | | | 1 | 2C9*2 | 6 | 26 | 156 | 468 | 468 | 0 | 0 | 100 | | | 2C9*3 | 6 | 26 | 156 | | | | | | | | VKORC1 | 6 | 26 | 156 | | | | | | | 10 | 2C9*2 | 6 | 26 | 156 | 468 | 468 | 0 | 0 | 100 | | | 2C9*3 | 6 | 26 | 156 | | | | | | | | VKORC1 | 6 | 26 | 156 | | | | | | | 50^{ε} | 2C9*2 | 6 | 26 | 156 | 468 | 468 | 0 | 0 | 100 | | | 2C9*3 | 6 | 26 | 156 | | | | | | | | VKORC1 | 6 | 26 | 156 | | | | | | | 100^{ε} | 2C9*2 | 6 | 26 | 156 | 468 | 468 | 0 | 0 | 100 | | | 2C9*3 | 6 | 26 | 156 | | | | | | | | VKORC1 | 6 | 26 | 156 | | | | | | | 200^{ε} | 2C9*2 | 6 | 26 | 156 | 468 | 468 | 0 | 0 | 100 | | | 2C9*3 | 6 | 26 | 156 | | | | | | | | VKORC1 | 6 | 26 | 156 | | | | | | | 300 | 2C9*2 | 2 | 3 | 6 | 18 | 18 | 0 | 0 | 100 | | | 2C9*3 | 2 | 3 | 6 | | | | | | | | VKORC1 | 2 | 3 | 6 | | | | | | {8} 9 | 400 | 2C9*2 | 2 | 3 | 6 | 18 | 18 | 0 | 0 | 100 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | 2C9*3 | 2 | 3 | 6 | | | | | | | | VKORC1 | 2 | 3 | 6 | | | | | | | 600 | 2C9*2 | 6 | 26 | 156 | 468 | 468 | 0 | 0 | 100 | | | 2C9*3 | 6 | 26 | 156 | | | | | | | | VKORC1 | 6 | 26 | 156 | | | | | | † The sponsor recommended DNA amount for eQ-PCR Warfarin Genotyping assay Call rate based on genotype: | Allele | Genotype | # samples | # tests | # genotype calls | # correct calls | # incorrect calls | # no calls | Correct call rate (%) | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | 2C9*2 | wt/wt | 4 | 740 | 740 | 739 | 0 | 1 | 99.9 | | | wt/*2 | 1 | 182 | 182 | 182 | 0 | 0 | 100 | | | *2/*2 | 1 | 182 | 182 | 182 | 0 | 0 | 100 | | 2C9*3 | wt/wt | 4 | 728 | 728 | 728 | 0 | 0 | 100 | | | wt/*3 | 1 | 188 | 188 | 188 | 0 | 0 | 100 | | | *3/*3 | 1 | 188 | 188 | 188 | 0 | 0 | 100 | | VKORC1 | G/G | 3 | 552 | 552 | 552 | 0 | 0 | 100 | | | G/A | 2 | 370 | 370 | 370 | 0 | 0 | 100 | | | A/A | 1 | 182 | 182 | 181 | 0 | 1 | 99.5 | | | | Total | 3312 | 3312 | 3310 | 0 | 2 | 99.9 | e. Analytical specificity: The sponsor conducted studies to demonstrate the interference of albumin (5 mg/ml), bilirubin (10 uM), triglycerides (4 mg/ml), heparin (14 U/ml), hemoglobin (80 mg/ml), magnesium chloride (6 mM), and lithium chloride (9 mM) and excess of extraction buffers on eQ-PCR assay. Nine blood samples covering all of the genotypes under evaluation were included. An aliquot of each sample was spiked with the above interfering substances and the results compared against the sample aliquot with no substance. The tests were repeated 6 times for each sample. The test results indicated no interference from above substances at the concentration levels tested and call rates remained unaffected. f. Assay cut-off: Not Applicable. 2. Comparison studies: a. Method comparison with predicate device: The sponsor conducted initial method comparison studies at 3 sites to demonstrate the accuracy of the device in comparison to bi-directional sequencing. This study included whole blood samples collected in EDTA from 150 donors. DNA was extracted at the sponsor's facility and aliquots of 50 unique DNA samples distributed to 3 sites (two external and the sponsor's facility) that participated in the study. The studies on the 150 {9} samples resulted in 3 no-calls and 2 incorrect calls. The sponsor's study using 150 samples did not cover *3/*3 variant genotypes tested by the device. Therefore, the sponsor conducted a second study that included 2 additional *3/*3 samples. Genotyping results confirmed $*3 / *3$ variant genotype for one of the samples, while the other new sample gave an incorrect genotype call of wt// $*2$ , $*3 / *3$ for the 2C9 gene. The sponsor performed a third study that included 7 new samples (4 $*3 / *3$ and 3 $*2 / *2$ samples). There were no incorrect calls or no-calls reported in this third study. Genotype call-rates for three studies combined are given in table below. | Genotype | # of samples | # of correct calls | # of incorrect calls | # of no calls | Correct call rate | | --- | --- | --- | --- | --- | --- | | VKORC1 | 159 | 156 | 0 | 3† | 98.11% | | 2C9*2 | 159 | 155 | 1‡ | 3† | 97.48% | | 2C9*3 | 159 | 155 | 1‡ | 3† | 97.48% | $\ddagger$ The incorrect genotype call reported a genotype of wt/\*2, \*3/\*3. † No-calls resulted from a system failure. The genotype call-rate for each specific genotype based on all three studies combined is given in the table below. | Allele | Genotype | # samples | # genotype calls | # correct calls | # incorrect calls | # no calls | Correct call rate (%) | 95%One-sided Confidence Lower Limit* | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | 2C9*2 | wt/wt | 126 | 126 | 123 | 0 | 3 | 97.6 | 94.40% | | | wt/*2 | 28 | 28 | 27 | 1‡ | 0 | 96.4 | 87.68% | | | *2/*2 | 5 | 5 | 5 | 0 | 0 | 100 | 47.98% | | | Sub-total | 159 | 159 | 155 | 0 | 3 | 97.5 | 94.59% | | 2C9*3 | wt/wt | 140 | 140 | 138 | 0 | 2 | 98.6 | 96.09% | | | wt/*3 | 13 | 13 | 12 | 0 | 1 | 92.3 | 75.32% | | | *3/*3 | 6 | 6 | 5 | 1‡ | 0 | 83.3 | 54.28% | | | Sub-total | 159 | 159 | 155 | 0 | 3 | 97.5 | 96.40% | | VKORC 1 | G/G | 79 | 79 | 77 | 0 | 2 | 97.5 | 93.16% | | | G/A | 63 | 63 | 62 | 0 | 1 | 98.4 | 94.32% | | | A/A | 17 | 17 | 17 | 0 | 0 | 100 | 80.52% | | | Sub-total | 159 | 159 | 156 | 0 | 3 | 98.1 | 95.54% | | Total for Assay | | | 477 | 466 | 2‡ | 9† | 97.7 | 96.18% | $\ddagger$ The incorrect genotype call reported a genotype of wt/\*2, \*3/\*3. † No-calls resulted from a system failure. {10} b. Matrix comparison: The sponsor recommends only fresh whole blood samples only for use with this device. Therefore, Matrix comparison studies were not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not Applicable. b. Clinical specificity: Not Applicable. c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not Applicable. 5. Expected values/Reference range: The following table provides the allele frequency across ethnic groups for the CYP2C9 and VKORC1 variants. Expected Genotypic Frequency of Different Ethnic Groups: | | Caucasian†1 | African†2,3 | Asian†1 | | --- | --- | --- | --- | | 2C9*2 | | | | | wt/wt | 78.7% | 87.0% | 100% | | wt /*2 | 20.4% | 8.7% | 0% | | *2/*2 | 0.9% | 0% | 0% | | 2C9*3 | | | | | wt/wt | 88.0% | N/A | 92.7% | | wt /*3 | 11.6% | 4.3% | 7.3% | | *3/*3 | 0.4% | 0% | 0% | | VKORC1 | | | | | GG | 39.1% | 79.2% | 1.9% | | GA | 46.7% | 20.8% | 18.3% | | AA | 14.2% | 0% | 79.8% | † The data presented are based on the genotypic frequencies in the following 3 publications: 1. Yuan et al., A novel functional VKORC1 promoter polymorphism is associated with inter-individual and inter-ethnic differences in warfarin sensitivity. Human Molecular Genetics 14(13): 1745-1751, 2005. 2. Kirchheimer and Brockmoller. Clinical Consequences of Cytochrome P450 2C9 Polymorphisms. Clinical Pharmacology & Therapeutics 77(1):1-16, 2005. 3. Schelleman et al., Warfarin response and Vitamin K Epoxide Reductase Complex in African Americans and Caucasians. Nature 81(5): 742-747, 2007 {11} N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 12
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...