K Number
K152612

Validate with FDA (Live)

Date Cleared
2016-05-26

(255 days)

Product Code
Regulation Number
862.3360
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The eSensor® Warfarin Sensitivity Saliva Test is an in vitro diagnostic for the detection and genotyping of the *2 and *3 alleles of the cytochrome P450 (CYP450) 2C9 gene locus and the Vitamin K epoxide reductase C1 (VKORC1) gene promoter polymorphism (-1639G>A) from genomic DNA extracted from human saliva samples collected using the Oragene® Dx and ORAcollect® Dx devices, as an aid in the identification of patients at risk for increased warfarin sensitivity.

The eSensor® XT-8 instrument is an in vitro diagnostic device intended for genotyping multiple mutations or polymorphisms in an amplified DNA sample utilizing electrochemical detection technology.

Device Description

The kit consists of the eSensor® Warfarin Sensitivity Saliva Test cartridge, the eSensor® Warfarin Sensitivity Saliva Test amplification reagents (including PCR mix and DNA polymerase), the eSensor® Warfarin Sensitivity Saliva Test detection reagents (including exonuclease, probes and hybridization buffer ingredients) and the eSensor® XT-8 System. One eSensor® Warfarin Sensitivity Saliva Test Kit has sufficient materials for 24 tests.

AI/ML Overview

The provided document, a 510(k) summary for the eSensor® Warfarin Sensitivity Saliva Test, details performance data primarily focused on a new specimen collection kit (ORAcollect®·Dx Device) and its impact on the existing device's performance. The study aims to demonstrate that incorporating this new collection kit does not compromise the accuracy or reliability of the Warfarin Sensitivity Saliva Test.

Here's an analysis of the acceptance criteria and study findings:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state 'acceptance criteria' in terms of specific numerical thresholds for accuracy, reproducibility, or call rates. However, the studies consistently aim for 100% agreement with the ground truth (bi-directional sequencing), and 100% correct calls with 0% incorrect calls. The "Method Comparison" section indicates an aspiration for high concordance and call rate.

Metric (Implicit Acceptance Criterion)Reported Device Performance
Reproducibility
Sample-to-Sample, Lot-to-Lot, Day-to-Day, Operator-to-Operator Agreement with Sequencing (CYP2C92, CYP2C93, VKOR)100% Agreement (60/60 correct calls across 3 operators for each SNP, totaling 180 successful calls out of 180 attempts for each SNP, or 120/120 for each donor when considering total calls).
Multi-center Reproducibility Agreement with Sequencing100% Agreement (30/30 for Site 1, 30/30 for Site 2, 29/29 for Site 3, after final pass, excluding one sample that did not meet input criteria).
Method Comparison
Overall Concordance with DNA Sequencing99.4% concordance for all polymorphisms (after retests).
First-pass Call Rate98.1%
Final Pass Call Rate99.4%
% Agreement per Genotype (post-retest)2C9*2: 99.0% (wt/wt), 100.0% (wt/2), 100.0% (2/2) **2C93: 100.0% (wt/wt), 95.0% (wt/*3), 100.0% (*3/*3) VKORC1: 98.4% (G/G), 100.0% (G/A), 100.0% (A/A) (Note: Lower 95% LCBs for less common genotypes due to smaller sample sizes).
Interference Studies
Endogenous Interfering Substances Agreement with Sequencing100% Agreement (for Amylase, Hemoglobin, IgA, Total Protein - all 14 samples tested per substance).
Exogenous Interfering Substances Agreement with Sequencing100% Agreement (for Eating, Drinking, Chewing Gum, Smoking, Mouthwash, Brushing Teeth at immediate and 30-minute time points - sample sizes ranging from 5 to 9 per group).

2. Sample Sizes and Data Provenance

  • Reproducibility (Sample-to-Sample, Lot-to-Lot, Day-to-Day, Operator-to-Operator):

    • Test set sample size: 10 donors, with 3 samples each (total 30 samples collected). These were processed by 3 operators, with genotyping data evaluated for each operator for each of the 3 SNPs. This implies 20 tests per SNP per operator (based on the "20" under "Samples Tested" for each SNP for each operator), totaling 60 tests per SNP across all operators, or 120 calls per SNP when considering all samples and operators for the "Summary of Results by Sample and Genotype" table.
    • Data Provenance: Not explicitly stated, but likely prospective and from a controlled lab environment.
  • Multi-center Reproducibility:

    • Test set sample size: 30 donors. Multiple saliva samples collected from each.
    • Data Provenance: Saliva samples were collected from 3 sites. Two sites were described as "professional setting" with "supervised collections," and the third site had "unsupervised collections." One sample from each donor was transported to three independent sites for extraction. All eSensor Warfarin Sensitivity Saliva testing was conducted at Site 1. This suggests a prospective data collection design, with samples from multiple (unspecified) locations. No country of origin is mentioned.
  • Method Comparison:

    • Test set sample size: 156 saliva samples.
    • Data Provenance: Not explicitly stated, but implies a prospective or retrospectively collected set of human saliva samples. No country of origin is mentioned.
  • Interfering Substances (Endogenous):

    • Test set sample size: 14 donors, each providing 4 saliva samples.
    • Data Provenance: Not explicitly stated, but likely prospective, controlled lab study.
  • Interfering Substances (Exogenous):

    • Test set sample size: Varied per activity, ranging from 5 to 9 donors per activity group. Each donor provided samples at two time-points (immediate and 30 minutes post-activity).
    • Data Provenance: Not explicitly stated, but likely prospective, controlled lab study.

3. Number of Experts and Qualifications for Ground Truth

The ground truth for all studies was established by bi-directional DNA sequencing. This is a laboratory-based method, not typically requiring "experts" in the sense of clinical specialists. The interpretation of sequencing results is a standard molecular biology technique. No specific number of experts or their qualifications (e.g., geneticists, molecular biologists) are mentioned for establishing the ground truth, as it's assumed to be a direct, objective measurement.

4. Adjudication Method for the Test Set

The studies used bi-directional DNA sequencing as the gold standard, not a human expert adjudication process. The agreement was calculated directly between the device's genotypes and the sequencing results. If there were discrepancies in the initial device results (e.g., initial miscalls or no-calls in the method comparison), retests were performed. For instance, in the method comparison, two miscalls were attributed to "operator error (sample mix-up) occurring during the first XT8 testing" and one remaining no-call to "an operator error at the purification step." This suggests an internal investigation and re-testing process rather than an external adjudication panel.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This device is a diagnostic test for genotyping, not an imaging device or a system requiring human interpretation with or without AI assistance. Therefore, the concept of "readers improving with AI vs. without AI" is not applicable here. The device itself performs the genotyping.

6. Standalone Performance Study

Yes, a standalone performance study was done. All the performance data described (reproducibility across operators/sites, method comparison with sequencing, and interfering substances studies) directly assesses the performance of the eSensor® Warfarin Sensitivity Saliva Test (algorithm/device only, without human-in-the-loop performance influencing the genotyping call itself, though human operators are involved in sample preparation and running the test). The percentage agreement and call rates reflect the direct output of the device compared to the ground truth.

7. Type of Ground Truth Used

The primary ground truth used for all studies was bi-directional DNA sequencing. This is a highly accurate molecular method for determining genetic sequences and polymorphisms (genotypes).

8. Sample Size for the Training Set

The document is a 510(k) summary for a labeling modification (addition of a new specimen collection kit) to an already cleared device. It primarily presents validation data for the continued performance of the device with the new collection kit. It does not mention a "training set" for the eSensor® Warfarin Sensitivity Saliva Test itself, as this device likely relies on established molecular biology principles and probes designed against known genetic targets, rather than machine learning algorithms that require extensive training data. It's a deterministic diagnostic test, not an AI model that learns from data in the way a computer vision algorithm would.

9. How Ground Truth for the Training Set Was Established

As noted above, the concept of a "training set" doesn't directly apply in the context of this device. The device's design is based on known genetic sequences and electrochemical detection rather than a machine learning model. The validation studies verify its performance against the established ground truth of bi-directional DNA sequencing.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 26, 2016

GENMARK DIAGNOSTICS, INCORPORATED ALAN MADERAZO VP, QUALITY, REGULATORY, & CLINICAL AFFAIRS 5964 LA PLACE COURT CARLSBAD CA 92008

Re: K152612

Trade/Device Name: eSensor Warfarin Sensitivity Saliva Test Regulation Number: 21 CFR §862.3360 Regulation Name: Drug Metabolism Enzyme Genotyping Test Regulatory Class: II Product Code: ODW, ODV, NSU Dated: April 21, 2016 Received: April 22, 2016

Dear Mr. Maderazo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152612

Device Name eSensor® Warfarin Sensitivity Saliva Test

Indications for Use (Describe)

The eSensor® Warfarin Sensitivity Saliva Test is an in vitro diagnostic for the detection and genotyping of the *2 and *3 alleles of the cytochrome P450 (CYP450) 2C9 gene locus and the Vitamin K epoxide reductase C1 (VKORC1) gene promoter polymorphism (-1639G>A) from genomic DNA extracted from human saliva samples collected using the Oragene® Dx and ORAcollect® Dx devices, as an aid in the identification of patients at risk for increased warfarin sensitivity.

The eSensor® XT-8 instrument is an in vitro diagnostic device intended for genotyping multiple mutations or polymorphisms in an amplified DNA sample utilizing electrochemical detection technology.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

eSensor® Warfarin Sensitivity Saliva Test

Attached is a 510(k) summary as described in 21 CFR 807.92

Sponsor Information

Submitted By:
Name:GenMark Diagnostics, Incorporated
Address:5964 La Place CourtCarlsbad, CA 92008(760) 448-4300
Company Contact:
Contact:Alan Maderazo, Ph.D., RAC
VP, Quality Assurance, Regulatory and Clinical Affairs
Phone:760-448-4308
Fax:760-683-6961
Email:al.maderazo@genmarkdx.com

Date Prepared: May 23, 2016

General Information

Trade Name: eSensor® Warfarin Sensitivity Saliva Test

eSensor® Warfarin Sensitivity Saliva Test

Device DescriptionDrug metabolizing enzyme genotyping testProthrombin time test,
Medical SpecialtyHematology
Product CodeODW, ODV
Device Class2
Regulation number862.3360, 864.7750

Instrument (XT-8)

Device DescriptionInstrumentation for clinical multiplex test systems
Medical SpecialtyClinical Chemistry
Product CodeNSU
Device Class2
Regulation number862.2570

{4}------------------------------------------------

Predicate device: eSensor® Warfarin Sensitivity Saliva Test, K110786

This premarket application is a labeling modification to a previously cleared device: addition of a new specimen collection kit.

Device Description

The kit consists of the eSensor® Warfarin Sensitivity Saliva Test cartridge, the eSensor® Warfarin Sensitivity Saliva Test amplification reagents (including PCR mix and DNA polymerase), the eSensor® Warfarin Sensitivity Saliva Test detection reagents (including exonuclease, probes and hybridization buffer ingredients) and the eSensor® XT-8 System. One eSensor® Warfarin Sensitivity Saliva Test Kit has sufficient materials for 24 tests.

The polymorphisms genotyped by the eSensor® Warfarin Sensitivity Saliva Test are shown in Table 1.

Table 1. Polymorphisms in the eSensor® Warfarin SensitivitySaliva Test Panel
PolymorphismAllele†
CYP450 2C9 430 C>T*2
CYP450 2C9 1075A>C*3
VKORC1 -1639G>AGG, GA, or AA
† CYP450 2C9 allele designations as established by the HumanCytochrome P450 (CYP) Allele Nomenclature Committee(http://www.cypalleles.ki.se/cyp2c9.htm). The major alleles ofthese polymorphisms are designated as *1.

{5}------------------------------------------------

Intended Use

The eSensor® Warfarin Sensitivity Saliva Test is an in vitro diagnostic for the detection and genotyping of the *2 and *3 alleles of the cytochrome P450 (CYP450) 2C9 gene locus and the Vitamin K epoxide reductase C1 (VKORC1) gene promoter polymorphism (-1639G>A) from genomic DNA extracted from human saliva samples collected using the Oragene®Dx and ORAcollect®•Dx devices, as an aid in the identification of patients at risk for increased warfarin sensitivity.

The eSensor® XT-8 instrument is an in vitro diagnostic device intended for genotyping multiple mutations or polymorphisms in an amplified DNA sample utilizing electrochemical detection technology.

{6}------------------------------------------------

Comparison to Predicate

This premarket application is a labeling modification to a previously cleared device: addition of a new specimen collection kit (the ORAcollect®·Dx Device manufactured by DNA Gentek). The data supporting the use of this specimen collection kit are provided in K152464.

Similarities
ItemProposed DevicePredicate Device (K110786)
Intended UseFor the detection and genotyping of the *2 and*3 alleles of the cytochrome P450 (CYP450) 2C9 gene locus and the Vitamin K epoxide reductase C1 (VKORC1) gene promoter polymorphism (-1639G>A)Same
Indications for Useas an aid in the identification of patients at risk for increased warfarin sensitivitySame
Device ComponentsTest cartridge, amplification reagents (including PCR mix and DNA polymerase), detection reagents (including exonuclease, probes and hybridization buffer ingredients) and the eSensor® XT-8 System.Same

{7}------------------------------------------------

Performance Data

The following studies were conducted to validate the performance of the ORAcollect-Dx device:

  • Reproducibility .
  • . Method Comparison
  • Interfering Substances .

Summaries of the study results are provided. The formal study reports are provided in K152464.

Reproducibility

Reproducibility of the performance of the ORAcollect-Dx device was evaluated to establish multi-center (site-to-site), lot-to-lot, sample-to-sample, day-to-day extraction and operator-tooperator reproducibility.

Sample-to-Sample, Lot-to-Lot, Day-to-Day and Operator-to-Operator Reproducibility

Three samples (collected using three lots of ORAcollect-Dx format OCD-100) from each of ten donors, were processed by three different operators on multiple days. Each operator extracted DNA from each sample using the same Qiagen QIAamp DNA mini kit, followed by determination of DNA concentration and A260/A280 ratio for all samples. Three operators tested the extracted DNA samples on the eSensor Warfarin Saliva Sensitivity Test. Genotyping data was evaluated after first-pass results and all samples were concordant to bi-directional sequencing.

{8}------------------------------------------------

SNPSamplesTestedCorrectCallsIncorrectCallsNo-Calls%Agreement
Operator12C9*2202000100%
2C9*3202000100%
VKOR202000100%
Operator22C9*2202000100%
2C9*3202000100%
VKOR202000100%
Operator32C9*2202000100%
2C9*3202000100%
VKOR202000100%
Combined2C9*2606000100%
2C9*3606000100%
VKOR606000100%

Summary of Results Stratified by Operator

Summary of Results by Sample and Genotype

Genotype by sequencingNumber ofSamples Testedby eSensorNumber ofCorrect Calls%Agreement
Donor ID2C9*22C9*3VKOR
REP01WTWTHET1212100%
REP02HETWTWT1212100%
REP03HETWTMUT1212100%
REP04HETWTWT1212100%
REP05HETHETHET1212100%
REP06WTMUTMUT1212100%
REP07MUTWTWT1212100%
REP08HETHETHET1212100%
REP09WTWTHET1212100%
REP10MUTWTWT1212100%
Total120120100%

Multi-center reproducibility

Thirty (30) donors collected multiple saliva samples each from 3 sites; 2 of the 3 sites were in a professional setting and had supervised collections compared to unsupervised collections at the third site. The 30 donors were selected to encompass a diverse genotype distribution for CYP2C9 and VKORC1 genotype distribution. After sample collection, one sample from each donor was transported at ambient temperatures to three (3) independent sites. Each site had one operator for a study total of 3 operators. Following sample extraction using the QIAamp DNA

{9}------------------------------------------------

mini kit, all purified genomic DNA samples were tested for DNA concentration, yield and A260/A280 at the sites where they were extracted. All purified genomic DNA samples were transported to Site 1 for testing on the eSensor Warfarin Sensitivity Saliva Test where one extracted DNA aliquot from each sample from each site was tested on the Warfarin assay, excluding a single sample that did not meet assay input criteria. When data from all three sites are combined, only one (1) sample did not meet the eSensor Warfarin Sensitivity Saliva Test input requirements. After final pass there was 100% agreement (89/89) with bidirectional sequencing.

Summary of Results by Site (Final Pass)

eSensor Warfarin Sensitivity Saliva Test***
Site of sampleextraction*Number ofSamples Tested**CorrectCallsIncorrectCallsNo-calls% Agreement with Bi-directional Sequencing
Site 1303000100%
Site 2303000100%
Site 3****292900100%

※ DNA extraction using QIAGEN QIAMP DNA Mini kit.

** Donor genotype representation: CYP2C9 (*1, *2, *3); VKOR

*** All eSensor Warfarin Sensitivity Saliva testing was conducted at Site 1 (by Operator 1). Only sample aliquots meeting the WST input criteria were tested.

**** One sample was excluded from WST testing

Method Comparison

In a method comparison study, a total of 156 saliva samples were genotyped using the eSensor® Warfarin Sensitivity Saliva Test and DNA sequencing. The genotyping calls by the eSensor® Warfarin Sensitivity Saliva Test method were 99.4% concordant with genotypes determined by DNA sequencing for all polymorphisms, with a 98.1% first-pass call rate and 99.4% final pass call rate. The following tables summarize the results of the method comparison study after retests.

{10}------------------------------------------------

eSensor® Warfarin Sensitivity Saliva Test Method Comparison After Retest
DNA Sequencing Result2C9 wt/wt2C9 wt/*22C9*2/*2
Correct Calls104465
No-Calls*100
Miscalls000
%Agreement99.0%100.0%100.0%
95% LCB95.6%93.7%54.9%
DNA Sequencing Result2C9 wt/wt2C9 wt/*32C9 *3/*3
Correct Calls135191
No-Calls*010
Miscalls000
%Agreement100.0%95.0%100.0%
95% LCB97.8%78.4%5.0%
DNA Sequencing ResultVKORC1G/GVKORC1G/AVKORC1A/A
Correct Calls637121
No-Calls*100
Miscalls000
%Agreement98.4%100.0%100.0%
95% LCB92.8%95.9%86.7%
On First pass there were 2 miscalls due to operator error (sample mix-up) occurring during the first XT8 testing.
  • After Final pass, one no call remained unresolved, likely due to an operator error at the purification step of the protocol.

Interfering Substances

Interfering substances including salivary α-amylase, hemoglobin, immunoglobulin A (IgA) and total protein were spiked into saliva samples at the highest amounts found in literature. 14 donors provided four saliva samples each which were each spiked with one of the four interfering substances. There was 100% agreement between the eSensor® Warfarin Sensitivity Saliva Test results and bidirectional DNA sequencing for all test substances after re-test, demonstrating no effect of any interfering substances on genotyping.

SubstanceSamplesTestedCorrectCallsIncorrectCallsNo-Calls%Agreement
Amylase141400100%
Hemoglobin141400100%
IgA141400100%
Total Protein141400100%

{11}------------------------------------------------

Effect of Exogenous Interfering Substances:

Potentially interfering exogenous substances introduced into saliva samples through various activities (eating, drinking, chewing gum, using mouthwash, smoking and brushing teeth) were tested. Each activity group was composed of five to nine donors who each provided samples immediately after activity and 30 minutes post-activity. There was 100% agreement between the eSensor® Warfarin Sensitivity Saliva Test results and bidirectional DNA sequencing for all activities tested in first pass, demonstrating no effect of any interfering substances on genotyping.

ActivityTime-pointSamplesTestedCorrectCallIncorrectCallNo-call% Agreement
EatingImmediate8800100%
30 minutes9900100%
DrinkingImmediate8800100%
30 minutes9900100%
Chewing GumImmediate7700100%
30 minutes7700100%
SmokingImmediate5500100%
30 minutes5500100%
MouthwashImmediate5500100%
30 minutes5500100%
Brushing TeethImmediate9900100%
30 minutes9900100%

Conclusion

The above test results support the safety and effectiveness of the eSensor® Warfarin Sensitivity Saliva Test on the eSensor® XT-8 System, and demonstrate substantial equivalence to the predicate device.

§ 862.3360 Drug metabolizing enzyme genotyping system.

(a)
Identification. A drug metabolizing enzyme genotyping system is a device intended for use in testing deoxyribonucleic acid (DNA) extracted from clinical samples to identify the presence or absence of human genotypic markers encoding a drug metabolizing enzyme. This device is used as an aid in determining treatment choice and individualizing treatment dose for therapeutics that are metabolized primarily by the specific enzyme about which the system provides genotypic information.(b)
Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping Test System.” See § 862.1(d) for the availability of this guidance document.