InstaStrip Fentanyl Rapid Test (Urine); InstaStrip Fentanyl Dipstick Test (Urine)
K240295 · Instanosis, Inc. · NGL · Mar 28, 2024 · Clinical Toxicology
Device Facts
Record ID
K240295
Device Name
InstaStrip Fentanyl Rapid Test (Urine); InstaStrip Fentanyl Dipstick Test (Urine)
Applicant
Instanosis, Inc.
Product Code
NGL · Clinical Toxicology
Decision Date
Mar 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3650
Device Class
Class 2
Indications for Use
InstaStrip Fentanyl Rapid Test (Urine) is intended for the qualitative detection of fentanyl in human urine at the cutoff value of 1.0 ng/mL. The test is intended for over-the-counter (OTC) use. The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive. The InstaStrip Fentanyl Dipstick Test (Urine) is intended for the qualitative detection of fentanyl in human urine at the cutoff value of 1.0 ng/mL. The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
Device Story
InstaStrip Fentanyl Test is a lateral flow immunoassay for qualitative detection of fentanyl in human urine; cutoff 1.0 ng/mL. Device components include test strip, test tube, and dropper. Principle: competitive binding; urine sample mixed with rabbit monoclonal antibody-gold nanoparticle conjugates; mixture applied to strip; pre-immobilized fentanyl-BSA on test line competes with sample fentanyl for antibody binding. Result: positive (no test line) if fentanyl >1 ng/mL; negative (visible test line) if fentanyl <1 ng/mL. Control line confirms test validity. Intended for OTC use; performed by lay users. Provides preliminary results; requires professional clinical judgment and confirmatory testing via chromatography/mass spectrometry for clinical decisions. Benefits: rapid, accessible screening for fentanyl presence.
Clinical Evidence
No clinical studies performed. Evidence consists of bench testing: precision studies (n=60 per concentration per lot), interference testing with various physiological/pathological substances, specificity/cross-reactivity analysis, and method comparison against LC/MS using 85 clinical samples. Lay-user study (n=140) confirmed usability and performance across diverse educational backgrounds. Results demonstrate high concordance with LC/MS and successful interpretation by lay users.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding principle. Components: test strip, test tube, dropper. Reagents: rabbit monoclonal antibody-gold nanoparticle conjugates, pre-immobilized fentanyl-BSA. Storage: 15-30°C. Standalone device; no connectivity or software. Qualitative visual readout.
Indications for Use
Indicated for qualitative detection of fentanyl in human urine at 1.0 ng/mL cutoff. Intended for over-the-counter (OTC) use by lay users (age 18+). Provides preliminary results only; requires confirmation by GC/MS or LC/MS.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
K242428 — Dochek® Fentanyl Urine Test Strip, Dochek® Fentanyl Urine Test Strip Plus, Dochek® Fentanyl Urine Test Card, Dochek® Fentanyl Urine Test Card Plus, Dochek® Fentanyl Urine Test Cup and Dochek® Fentanyl Urine Test Cup Plus · Guangzhou Decheng Biotechnology Co., Ltd. · Feb 6, 2025
K240949 — Healgen® Accurate Fentanyl Rapid Test Cassette (Urine); Healgen® Accurate Rapid Fentanyl Test Cassette (Urine); Healgen® Accurate Fentanyl Rapid Test Dip Card (Urine); Healgen® Accurate Rapid Fentanyl Test Dip Card (Urine); Healgen® Accurate Fentanyl Rapid Test Strip (Urine); Healgen® Accurate Rapid Fentanyl Test Strip (Urine) · Healgen Scientific,, LLC · May 13, 2024
K233417 — AllTest Fentanyl Urine Test Cassette · Hangzhou AllTest Biotech Co., Ltd. · Oct 26, 2023
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K240295
B Applicant
Instanosis Inc
C Proprietary and Established Names
InstaStrip Fentanyl Rapid Test (Urine); InstaStrip Fentanyl Dipstick Test (Urine)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NGL | Class II | 21 CFR 862.3650 - Opiate Test System | TX - Clinical Toxicology |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Fentanyl
C Type of Test:
Qualitative
## III Intended Use/Indications for Use:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K240295 - Page 2 of 10
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
InstaStrip Fentanyl Rapid Test (Urine) is intended for the qualitative detection of fentanyl in human urine at the cutoff value of 1.0 ng/mL. The test is intended for over-the-counter (OTC) use.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
The InstaStrip Fentanyl Dipstick Test (Urine) is intended for the qualitative detection of fentanyl in human urine at the cutoff value of 1.0 ng/mL.
The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.
C Special Conditions for Use Statement(s):
OTC - Over The Counter
D Special Instrument Requirements:
Not applicable.
IV Device/System Characteristics:
A Device Description:
InstaStrip Fentanyl Rapid Test (Urine) and InstaStrip Fentanyl Dipstick Test (Urine) are an immunoassay technique that is based on competitive lateral flow immunoassay intended for the qualitative detection of fentanyl in human urine. The candidate devices consist of a urine cup, test tube, dropper, test strip, tube stand, and package insert. Each test strip is sealed with sachets of desiccant in an aluminum pouch.
B Principle of Operation:
The urine sample is added to a provided test tube containing dried rabbit monoclonal antibody-gold nanoparticle (Ab-AuNP) conjugates. The test tube is shaken until the Ab-AuNP pellet is dissolved and the sample turns pink. The test strip is then inserted into the test tube and the user caps the test tube. The pre-immobilized fentanyl-BSA on the test line competes with fentanyl in the urine sample for binding to the Ab-AuNPs. The device is designed so that when the fentanyl concentration in the urine sample exceeds 1 ng/mL, the test line is no longer visible. The test line
{2}
will be visible and the result is negative when the fentanyl concentration in the urine sample is less than 1 ng/mL. No matter whether the sample contains the corresponding analyte or not, the quality control area (C) will develop a colored line, which is the criteria for judging whether the chromatography process is normal or not.
V Substantial Equivalence Information:
A Predicate Device Name(s):
AllTest Fentanyl Urine Test Cassette
B Predicate 510(k) Number(s):
K233417
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K240295 | K233417 |
| --- | --- | --- |
| Device Trade Name | InstaStrip Fentanyl Rapid Test (Urine) and InstaStrip Fentanyl Dipstick Test (Urine) | AllTest Fentanyl Rapid Test (Urine) |
| General Device Characteristic Similarities | | |
| Indications For Use | For the qualitative detection of fentanyl in human urine. | Same |
| Calibrator and Cut-off Values | Fentanyl (FTY) 1 ng/mL | Same |
| Methodology | Competitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry. | Same |
| Test Type | Qualitative | Same |
| Intended Use | For Over-The-Counter Use | Same |
| General Device Characteristic Differences | | |
| Configurations | Strip | Cassette |
| Storage | 15-30°C | 4-30°C |
VI Standards/Guidance Documents Referenced:
None referenced.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
K240295 - Page 3 of 10
{3}
K240295 - Page 4 of 10
1. Precision/Reproducibility:
Precision studies for InstaStrip Fentanyl Rapid Test (Urine) and InstaStrip Fentanyl Dipstick Test (Urine) were carried out at one site. The randomized and blinded study was performed for samples with concentrations of -100% cut-off, -75% cut-off, -50% cut-off, -25% cut-off, cut-off, +25% cut-off, +50% cut-off, +75% cut-off and +100% cut-off where the fentanyl cut-off concentration was 1 ng/mL. Samples with concentration of -100% cut-off were drug-free urine samples. Other samples were prepared by spiking target drug fentanyl in drug-free urine samples. Each drug concentration was confirmed by LC/MS. The samples were tested by three operators using three different device lots over a 10-day period, each operator tested a single lot. Six tests per day per concentration per lot were tested for a total 54 tests per day per lot.
| Lot Number | -100% cut-off | -75% cut-off | -50% cut-off | -25% cut-off | cut-off | 25% cut-off | 50% cut-off | 75% cut-off | 100% cut-off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Lot 1 | 60-/0+ | 60-/0+ | 60-/0+ | 54-/6+ | 32+/28- | 60+/0- | 60+/0- | 60+/0- | 60+/0- |
| Lot 2 | 60-/0+ | 60-/0+ | 60-/0+ | 55-/5+ | 30+/30- | 60+/0- | 60+/0- | 60+/0- | 60+/0- |
| Lot 3 | 60-/0+ | 60-/0+ | 60-/0+ | 55-/5+ | 34+/26- | 60+/0- | 60+/0- | 60+/0- | 60+/0- |
2. Linearity:
Not applicable, this device is intended for qualitative use only.
3. Analytical Specificity/Interference:
Cross-Reactivity:
To test specificity, similarly structured drug metabolites and other components that are likely to interfere in urine samples were tested using three lots of device by three different operators. Results are expressed as a minimum concentration of metabolite or compound required to produce a response approximately equivalent to the cutoff concentration of the assay. The percent cross reactivity of those compounds is listed below (if no cross reactivity was observed, the highest concentration tested is shown as <0.01%):
| Compounds for cross reactivity | Concentration Approximately Equivalent to the Cut-off (ng/mL) | Percent (%) cross reactivity |
| --- | --- | --- |
| Acetyl fentanyl | 1 | 100 |
| Acetyl norfentanyl | >10,000 | <0.01 |
| Acrylfentanyl | 1.8 | 56 |
| Alfentanil | >10,000 | <0.01 |
| Benzodioxole fentanyl | 4.8 | 21 |
| Butyryl fentanyl | 1 | 100 |
| Carfentanil | >10,000 | <0.01 |
| Crotonyl fentanyl | 1 | 100 |
{4}
The following structurally unrelated opioid compounds were tested at a concentration of 100 $\mu \mathrm{g} / \mathrm{mL}$ . Negative results were obtained for all these compounds.
| 6-Acetyl morphine | Morphine |
| --- | --- |
| Amphetamine | Morphine-3-glucuronide |
| Buprenorphine | Naloxone |
| Buprenorphine glucuronide | Naltrexone |
| Ceftriaxone | Norbuprenorphine |
| M-Chlorophenylpiperazine | Norcodeine |
| Ciprofloxacin | Norketamine |
| Cocaine | Normeperidine |
| Codeine | Normorphine |
| Dextromethorphan | Noroxycodone |
| Dihydrocodeine | Ofloxacin |
| Diphehydramine | Oxycodone |
K240295 - Page 5 of 10
{5}
K240295 - Page 6 of 10
| Duloxetine | Oxymorphone |
| --- | --- |
| EDDP | Pentazocine (Talwin) |
| EMDP | Pipamperone |
| Fluoxetine | 1-(3-chlorophenyl)
Piperazine (hydrochloride) |
| Haloperidol | Quinidine |
| Heroin | Risperidone |
| Hydrocodone | Tapentadol |
| Hydromorphone | Thioridazine |
| Ketamine | Tilidine |
| Levorphanol | Tramadol |
| MDMA | Tramadol-O- Desmethyl |
| Meperidine | Tramadol-N- Desmethyl |
| Methadone | Trazodone |
| Methamphetamine | |
Interference:
Potential endogenous and exogenous interfering substances commonly found in human urine, were added to drug-free urine and target drug fentanyl urines with concentrations at 50% below and 50% above cut-off levels. Tests were performed for compounds at a concentration of 100 µg/mL or specified concentrations using three lots of the device. No compounds showed any interference. Refer to the tables below for the compounds and levels tested.
Compounds tested at a concentration of 100 µg/mL are shown below.
| Acetophenetidin | DL-Tyrosine | Octopamine |
| --- | --- | --- |
| Acetylsalicylic acid | Doxepin | O-Hydroxyhippuric acid |
| Albuterol | Ecgonine methyl ester | Oxazepam |
| Aminopyrine | Ephedrine | Oxolinic acid |
| Amitriptyline | Erythromycin | Oxymetazoline |
| Amobarbital | Fenoprofen | Papaverine |
| Amoxicillin | Fluphenazine | Penicillin G |
| Ampicillin | Furosemide | Perphenazine |
| Apomorphine | Gentisic acid | Phencyclidine |
| Aspartame | Hydralazine | Phenelzine |
| Atropine | Hydrochlorothiazide | Phenobarbital |
| Benzilic acid | Hydrocortisone | Prednisone |
| Benzoic acid | Hydroxytyramine | Propoxyphene |
| Benzoylecgonine | Imipramine | Propranolol |
| Bilirubin | Isoproterenol | Pseudoephedrine |
| Bupropion | Ketamine | Ranitidine |
| Caffeine | Ketoprofen | Salicylic acid |
| Carbamazepine | Lidocaine | Secobarbital |
{6}
K240295 - Page 7 of 10
| Chloral hydrate | Loperamide | Serotonin (5-hydroxytyramine) |
| --- | --- | --- |
| Chloramphenicol | Maprotiline | Sulfamethazine |
| Chlorothiazide | Meperidine | Sulindac |
| Chlorpromazine | Meprobamate | Tetrahydrocortisone 3-(β-Dglucuronide) |
| Cholesterol | Methapyrilene | Tetrahydrocortisone 3-acetate |
| Clomipramine | Methaqualone | Tetrahydrozoline |
| Clonidine | Methoxyphenamine | Thiamine |
| Cortisone | N-Acetylprocainamide | Thioridazine |
| Cotinine | Nalidixic acid | Triamterene |
| Cyclobenzaprine | Naloxone | Trifluoperazine |
| Deoxycorticosterone | Naltrexone | Trimethoprim |
| Desipramine | Naproxen | Tyramine |
| Dextromethorphan | Niacinamide | Uric acid |
| Diclofenac | Nicotine | Venlafaxine |
| Diflunisal | Nifedipine | Zomepirac |
| Digoxin | Norethindrone | β-Estradiol |
| Diphenhydramine | Nortriptyline | |
| DL-Tryptophan | Noscapine | |
Compounds tested at specified concentrations are shown below.
| Acetaminophen (500 μg/mL) |
| --- |
| Acetone (1000 mg/dL) |
| Albumin (500 mg/dL) |
| Ascorbic Acid (560 mg/dL) |
| Boric acid (1% w/v) |
| Creatinine (500 mg/dL) |
| Ethanol (1%) |
| Galactose (10 mg/dL) |
| Gamma globulin (500 mg/dL) |
| Glucose (3000 mg/dL) |
| Hemoglobin (500 mg/dL) |
| Ibuprofen (500 μg/mL) |
| Isoxsuprine (20 μg/mL) |
| Labetalol (15 μg/mL) |
| Metronidazole (300 μg/mL) |
| NaCl (4000 mg/dL) |
| Oxalic acid (100 mg/dL) |
{7}
K240295 - Page 8 of 10
Quinine (15 µg/mL)
Riboflavin (10 mg/dL)
Urea (2000 mg/dL)
Valproic acid (250 µg/mL)
Verapamil (20 µg/mL)
Effect of Urinary Specific Gravity and pH:
To investigate the effect of urine specific gravity and urine pH, urine samples with 1.000, 1.002, 1.008, 1.014, 1.019, 1.022, 1.030, and 1.035 specific gravity and urine samples with pH of 4.0, 5.0, 6.0, 7.0, 8.0, and 9.0 were spiked with target fentanyl at 50% below and 50% above cut-off levels. These samples were tested using three lots of the device. Results were all positive for samples at and above +50% cut-off and all negative for samples at and below -50% cut-off.
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The drug calibrator for this device is traceable to a commercially available standard.
6. Detection Limit:
Characterization of how the device performs at low concentrations appears in the precision section, VII.A.1, above.
7. Assay Cut-Off:
Characterization of how the device performs at low concentrations appears in the precision section, VII.A.1, above.
B Comparison Studies:
1. Method Comparison:
Method comparison studies were performed by three operators using three device lots. Unaltered clinical samples included a total of 85 urine samples (42 Negative and 43 positive). The samples were blind labeled and compared to LC/MS results. The results are summarized in the tables below.
| | | Negative | Low Negative by LC/MS (less than -50%) | Near Cut-off Negative by LC/MS (Between -50% and cut-off) | Near Cut-off Positive by LC/MS (Between the cut-off and +50%) | High Positive by LC/MS (greater than +50%) |
| --- | --- | --- | --- | --- | --- | --- |
| | Positive | 0 | 0 | 1 | 7 | 36 |
{8}
| Operator 1 | Negative | 11 | 21 | 9 | 0 | 0 |
| --- | --- | --- | --- | --- | --- | --- |
| Operator 2 | Positive | 0 | 0 | 0 | 7 | 36 |
| | Negative | 11 | 21 | 10 | 0 | 0 |
| Operator 3 | Positive | 0 | 0 | 0 | 6 | 36 |
| | Negative | 11 | 21 | 10 | 1 | 0 |
Discordant Results
| Operator | Sample Number | LC-MS/MS Result | InstaStrip Result |
| --- | --- | --- | --- |
| Operator 1 | 310-13724 | 0.9 | Positive |
| Operator 3 | 280-06705 | 1 | Negative |
2. Matrix Comparison:
Not applicable. These devices are for use with urine samples only.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Lay User Study:
A lay user was performed at three intended user sites with 140 lay persons. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to $>50$ years. Urine samples were prepared at the following concentrations: $-100\%$ , $+/-75\%$ , $+/-50\%$ , $+/-25\%$ of the cut-off by spiking fentanyl into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers, blind-labeled and randomized. Each participant was provided with the package insert, 1 blind labeled sample and a device. A summary of the results is shown in the table below.
| % of Cut-off by LC/MS | Number of samples | No. of Positive | No. of Negative | The percentage of correct results (%) |
| --- | --- | --- | --- | --- |
| -100% Cut-off | 20 | 0 | 20 | 100% |
| -75% Cut-off | 20 | 0 | 20 | 100% |
| -50% Cut-off | 20 | 0 | 20 | 100% |
| -25% Cut-off | 20 | 1 | 19 | 95% |
| +25% Cut-off | 20 | 20 | 0 | 100% |
K240295 - Page 9 of 10
{9}
| +50% Cut-off | 20 | 20 | 0 | 100% |
| --- | --- | --- | --- | --- |
| +75% Cut-off | 20 | 20 | 0 | 100% |
Lay-users were also given surveys on the ease of understanding the package insert instructions and device use, and the results were found to be acceptable. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.
Read Time Flex Study:
The sponsor provided data to support the recommendation for read time. The sponsor recommends that the results should be read at 5 minutes and not after 20 minutes.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K240295 - Page 10 of 10
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.