SAFECARE Multi-Drug Urine Test Cup, SAFECARE Multi-Drug Urine Test Dip-Card
Device Facts
| Record ID | K202453 |
|---|---|
| Device Name | SAFECARE Multi-Drug Urine Test Cup, SAFECARE Multi-Drug Urine Test Dip-Card |
| Applicant | Safecare Biotech (Hangzhou) Co., Ltd. |
| Product Code | NFT · Clinical Toxicology |
| Decision Date | Mar 24, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Indications for Use
SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of: [Table of cut-offs]. Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes. The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use. SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine. Oxazepam. Marijuana. Methamphetamine, Morphine. Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of: [Table of cut-offs]. Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes. The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.
Device Story
Single-use lateral flow immunochromatographic assays for qualitative drug detection in human urine. Input: urine sample. Principle: competitive binding; target drugs below cutoff allow antibody-coated particles to bind immobilized drug-conjugate, forming visible test line; drug presence above cutoff saturates binding sites, preventing line formation. Output: visual qualitative result (line/no line). Used in home/OTC settings by lay users. Results are preliminary; require GC/MS or LC/MS confirmation. Benefits: rapid, simultaneous screening for multiple substances.
Clinical Evidence
Bench testing only. Performance validated via precision studies (25 days, 2 runs/day), interference testing, specificity/cross-reactivity analysis, and method comparison against LC/MS (80 samples per drug). Lay-user study conducted with 310 participants across diverse demographics, demonstrating high accuracy (90-100%) across various drug concentrations relative to cut-offs. Flesch-Kincaid analysis confirmed 7th-grade reading level for instructions.
Technological Characteristics
Lateral flow immunochromatographic assay. Single-use dip card or cup format. Reagent pads contain monoclonal mouse antibodies. Qualitative visual readout. Stable at 4-30°C for 24 months. No electronic components or software.
Indications for Use
Indicated for qualitative, simultaneous detection of drugs of abuse (Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, MDMA, Phencyclidine, Methadone, Nortriptyline, Propoxyphene, EDDP) in human urine. Intended for over-the-counter use by lay users. Not intended to distinguish between prescription use and abuse.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
- K233019 — AllTest Multi-Drug Rapid Test Cup; AllTest Multi-Drug Rapid Test Panel; AllTest Multi-Drug Rapid Test Cup Rx; AllTest Multi-Drug Rapid Test Panel Rx · Hangzhou AllTest Biotech Co., Ltd. · Dec 13, 2023
- K171695 — First Sign Multi-Drug Dip Card Test, First Sign Multi-Drug Cup Test, First Sign Drug of Abuse Dip Card Test Marijuana, First Sign Drug of Abuse Cup Test Marijuana · W.H.P.M., Inc. · Feb 5, 2018
- K180255 — CLUNGENE Multi-Drug Test Dip Card, CLUNGENE Multi-Drug Test Easy Cup · Hangzhou Clongene Biotech Co., Ltd. · Feb 28, 2018
- K233062 — BioSieveTM Multi-Drug Urine Test Panel; BioSieveTM Multi-Drug Urine Test Panel Rx · Vivachek Biotech (Hangzhou) Co., Ltd. · Nov 2, 2023