K Number
K233062

Validate with FDA (Live)

Date Cleared
2023-11-02

(37 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BioSieve™ Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, 2- ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP 300/OPI 2000)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

BioSieve™ Multi-Drug Urine Test Panel offers any combinations from 1 to 15 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use only. It is intended for OTC use.

The tests may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene, and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

BioSieve™ Multi-Drug Urine Test Panel Rx tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1.5-dimenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Cannabinoids in human urine at the cutoff concentrations of:

DrugCalibratorCut-off (ng/mL)
Amphetamine (AMP)D-Amphetamine1,000 or 500
Barbiturates (BAR)Secobarbital300
Buprenorphine (BUP)Buprenorphine10
Oxazepam (BZO)Oxazepam300
Cocaine (COC)Benzoylecgonine300 or 150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
Ecstasy (MDMA)D,L- Methylenedioxy-methamphetamine500
Methamphetamine (MET)D-Methamphetamine1,000 or 500
Morphine (MOP/OPI)Morphine2,000 or 300
Methadone (MTD)Methadone300
Oxycodone (OXY)Oxycodone100
Phencyclidine (PCP)Phencyclidine25
Propoxyphene (PPX)Propoxyphene300
Nortriptyline (TCA)Nortriptyline1,000
Marijuana (THC)11-nor-Δ9-THC-9 COOH50

BioSieve™ Multi-Drug Urine Test Panel Rx offers any combinations from 1 to 15 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use only. It is intended for prescription use.

The tests may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene, and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

Device Description

The BioSieve™ Multi-Drug Urine Test Panel and BioSieve™ Multi-Drug Urine Test Panel Rx are rapid, single-use in vitro diagnostic devices. Each test kit contains a test device in one pouch. One pouch contains a test BioSieve™ Panel and two desiccants, and a package insert. The BioSieve™ Multi-Drug Urine Test Panel is intended for over-the-counter use and the BioSieve™ Multi-Drug Urine Test Panel Rx is intended for prescription use.

AI/ML Overview

Acceptance Criteria and Device Performance for BioSieve™ Multi-Drug Urine Test Panel

This document describes the acceptance criteria and the studies performed to demonstrate that the BioSieve™ Multi-Drug Urine Test Panel meets these criteria.

1. Acceptance Criteria and Reported Device Performance

The core acceptance criterion for the BioSieve™ Multi-Drug Urine Test Panel, as derived from the provided precision studies, is the accurate qualitative detection of drugs at various concentrations relative to the defined cutoff level. Specifically, the device is expected to:

  • Consistently report Negative (no drug detected) for samples with drug concentrations at or below -25% of the cutoff level (e.g., -100% cutoff, -75% cutoff, -50% cutoff, -25% cutoff).
  • Consistently report Positive (drug detected) for samples with drug concentrations at or above +25% of the cutoff level (e.g., +25% cutoff, +50% cutoff, +75% cutoff, +100% cutoff).
  • Demonstrate a balanced distribution of Positive and Negative results around the cutoff level (e.g., approximately 50% positive and 50% negative).

The reported device performance, based on the precision studies for each drug, generally aligns with these criteria. For concentrations at or below -25% of the cutoff, almost all results were negative (e.g., 50-/0+ indicating 50 negative and 0 positive). For concentrations at or above +25% of the cutoff, almost all results were positive (e.g., 0-/50+ indicating 0 negative and 50 positive). At the cutoff concentration, the results showed a mix of positive and negative, reflecting the expected performance of a qualitative assay near its detection threshold (e.g., 23-/27+, 26-/24+, etc.).

Table of Acceptance Criteria and Reported Device Performance (Example for BUP 10, PCP 25, THC 50, OXY 100)

Concentration by LC/MS (relative to cutoff)Acceptance Criteria for Qualitative ResultReported Performance (Example: BUP 10)Reported Performance (Example: PCP 25)Reported Performance (Example: THC 50)Reported Performance (Example: OXY 100)
-100% cutoffNegative50-/0+50-/0+50-/0+50-/0+
-75% cutoffNegative50-/0+50-/0+50-/0+50-/0+
-50% cutoffNegative50-/0+50-/0+50-/0+50-/0+
-25% cutoffNegative50-/0+50-/0+50-/0+50-/0+
CutoffMixed (approx. 50% Pos / 50% Neg)26-/24+ (Avg. across lots)24-/26+ (Avg. across lots)26-/24+ (Avg. across lots)25-/25+ (Avg. across lots)
+25% cutoffPositive0-/50+0-/50+0-/50+0-/50+
+50% cutoffPositive0-/50+0-/50+0-/50+0-/50+
+75% cutoffPositive0-/50+0-/50+0-/50+0-/50+
+100% cutoffPositive0-/50+0-/50+0-/50+0-/50+

Note: The table above provides a representative sample. The full document lists extensive data for each drug, demonstrating similar performance.

2. Sample Sizes Used for the Test Set and Data Provenance

The primary analytical performance test (Precision Study) used a substantial sample size. For each drug, tests were performed over 25 days, with two runs per day, using three different lots of test panels. This means for each concentration level tested per drug (e.g., -100% cutoff, -50% cutoff, cutoff, +50% cutoff, +100% cutoff), there were 25 days * 2 runs * 3 lots = 150 individual tests.

The method comparison studies utilized 80 unaltered urine samples for each drug, divided into 40 negative and 40 positive. These samples were run by three different operators.

The lay-user study included a total of 280 participants.

  • Configuration 1: 64 males and 76 females (140 participants)
  • Configuration 2: 67 males and 73 females (140 participants)
    For each drug and each concentration level in the lay-user study, there were 20 test results. Given there are 7 concentration levels per drug, this amounts to 140 results per drug per configuration.

Data Provenance: The document does not explicitly state the country of origin for the data. However, the studies were performed by VivaChek Biotech (Hangzhou) Co., Ltd. located in Hangzhou, China, suggesting the data was collected there. The studies appear to be retrospective as they involve prepared samples or collected unaltered urine samples that were then tested.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for the analytical performance and method comparison studies was established using LC/MS (Liquid Chromatography/Mass Spectrometry) or LC/MS/MS. LC/MS is a highly accurate and widely accepted gold-standard analytical method for confirming drug concentrations in urine.

The document does not specify the "number of experts" or their "qualifications" in the context of establishing LC/MS ground truth, as LC/MS is an instrumental method that provides objective quantitative results. The operation and interpretation of LC/MS data typically rely on trained laboratory personnel with expertise in analytical chemistry, but their specific "qualifications" (e.g., years of experience) are not detailed in this submission.

For the lay-user study, the ground truth was also established by spiking known concentrations of drugs into drug-free urine, confirming these concentrations via LC/MS or LC/MS/MS.

4. Adjudication Method for the Test Set

For the analytical performance (precision) studies, the raw results for each concentration level are presented (e.g., 26-/24+). This implies no specific "adjudication method" beyond the direct comparison of the device's qualitative result (positive/negative) against the quantitative LC/MS ground truth. The acceptance criteria for the cutoff concentration already imply a mixed outcome is expected.

For the method comparison study, discordant results are explicitly listed with the LC/MS result compared against the device's result by each operator. This indicates a direct comparison to the LC/MS ground truth rather than an adjudication process between device results.

For the lay-user study, the "agreement (%)" is calculated based on how often the lay user's interpretation of the device result matched the expected result based on the spiked concentration (and confirmed by LC/MS). Again, this is a direct comparison to established ground truth, not an adjudication of human interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done in the traditional sense of comparing human readers with and without AI assistance, as this device (BioSieve™ Multi-Drug Urine Test Panel) is a lateral flow immunochromatographic assay, not an AI-powered diagnostic tool for image or data interpretation.

The studies involve:

  • Analytical performance: Device-to-LC/MS comparison.
  • Method comparison: Three operators reading the device and comparing to LC/MS. This is a multi-reader study, but comparing the device's output to a gold standard, not assessing the improvement of human readers with AI assistance.
  • Lay-user study: Multiple lay users interpreting the device results based on provided instructions. This evaluates the ease of use and interpretability for the intended over-the-counter audience, not the effectiveness of AI assistance.

Therefore, an "effect size of how much human readers improve with AI vs without AI assistance" is not applicable to this device or the studies presented.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Given that the BioSieve™ Multi-Drug Urine Test Panel is a physical, lateral flow immunochromatographic assay, it inherently requires a human-in-the-loop to visually interpret the test lines. It is not an "algorithm only" device. Therefore, a standalone (algorithm only) performance study was not done and is not applicable. The precision, comparison, and lay-user studies all involve human observation and interpretation of the visual reaction.

7. The Type of Ground Truth Used

The primary type of ground truth used across all analytical studies (precision, specificity, interference, method comparison, and lay-user studies) is analytical confirmation by LC/MS (Liquid Chromatography/Mass Spectrometry) or LC/MS/MS. This a quantitative chemical method considered a gold standard for drug detection and concentration measurement.

For the precision studies, various concentrations of each drug were "spiking target drug in drug-free urine samples" and "confirmed by LC/MS."
For the method comparison studies, "unaltered urine samples were blind labeled and compared to LC/MS results."
For the lay-user study, samples were "spiked with drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS or LC/MS."

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or algorithm development. This device is a biochemical immunoassay, not an AI/ML-based diagnostic. Therefore, there is no separate "training set" as understood in a computational context.

The studies described are for verification and validation of the device's analytical performance.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" for an AI/ML algorithm for this device, this question is not applicable. The chemical and biological principles of the immunoassay define its detection capabilities, rather than a learned model from a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

VivaChek Biotech (Hangzhou) Co., Ltd. % Joe Shia LSI International 504 E Diamond Ave., Suite I Gaithersburg, Maryland 20877

Re: K233062

Trade/Device Name: BioSieve™ Multi-Drug Urine Test Panel: BioSieve™ Multi-Drug Urine Test Panel Rx Regulation Number: 21 CFR 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: NFT, NFY, NGL, PTH, NFV, PTG, NGG, LCM, QBF, QAW, NFW Dated: September 21, 2023 Received: September 26, 2023

Dear Joe Shia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Joseph A. Digitally signed by Joseph
Kotarek -S
Date: 2023.11.02 14:50:57 Kotarek -S Joseph Kotarek, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233062

Device Name

BioSieve™ Multi-Drug Urine Test Panel

Indications for Use (Describe)

BioSieve™ Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, 2- ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP 300/OPI 2000)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

BioSieve™ Multi-Drug Urine Test Panel offers any combinations from 1 to 15 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use only. It is intended for OTC use.

The tests may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene, and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K233062

Device Name

BioSieve™ Multi-Drug Urine Test Panel Rx

Indications for Use (Describe)

BioSieve™ Multi-Drug Urine Test Panel Rx tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1.5-dimenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Cannabinoids in human urine at the cutoff concentrations of:

DrugCalibratorCut-off (ng/mL)
Amphetamine (AMP)D-Amphetamine1,000 or 500
Barbiturates (BAR)Secobarbital300
Buprenorphine (BUP)Buprenorphine10
Oxazepam (BZO)Oxazepam300
Cocaine (COC)Benzoylecgonine300 or 150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
Ecstasy (MDMA)D,L- Methylenedioxy-methamphetamine500
Methamphetamine (MET)D-Methamphetamine1,000 or 500
Morphine (MOP/OPI)Morphine2,000 or 300
Methadone (MTD)Methadone300
Oxycodone (OXY)Oxycodone100
Phencyclidine (PCP)Phencyclidine25
Propoxyphene (PPX)Propoxyphene300
Nortriptyline (TCA)Nortriptyline1,000
Marijuana (THC)11-nor-Δ9-THC-9 COOH50

BioSieve™ Multi-Drug Urine Test Panel Rx offers any combinations from 1 to 15 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use only. It is intended for prescription use.

The tests may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene, and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

{5}------------------------------------------------

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{6}------------------------------------------------

510(k) SUMMARY

K233062

1DateOctober 25, 2023
2SubmitterVivaChek Biotech (Hangzhou) Co., Ltd.Level 2, Block 2, 146 East Chaofeng Rd.Hangzhou, China
3Contact PersonJoe ShiaLSI International Inc.504 East Diamond Ave., Suite HGaithersburg, MD 20877Telephone: 240-505-7880Fax: 301-916-6213Email: shiajl@yahoo.com
  • BioSieve™ Multi-Drug Urine Test Panel 4 Device Name BioSieve™ Multi-Drug Urine Test Panel Rx
  • 5 Classification Class II
Class II
Product CodeTarget DrugRegulation SectionPanel
NFTAmphetamine (AMP)862.3100, Amphetamine Test SystemToxicology
NGLBuprenorphine (BUP)862.3650, Opiate Test SystemToxicology
PTHSecobarbital (BAR)862.3150, Barbiturate Test SystemToxicology
NFVOxazepam (BZO)862.3170,Benzodiazepine Test SystemToxicology
NFYCocaine (COC)862.3250, Cocaine Test SystemToxicology
PTG2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine(EDDP)862.3620, Methadone Test SystemToxicology
NGGMethamphetamine(MET)862.3610,Methamphetamine Test SystemToxicology
NGGMethylenedioxymethamphetamine (MDMA)862.3610,Methamphetamine Test SystemToxicology
NGL862.3650, Opiate Test SystemToxicology

{7}------------------------------------------------

Morphine (MOP/OPI)
PTGMethadone (MTD)862.3620, Methadone Test SystemToxicology
NGLOxycodone (OXY)862.3650, Opiate Test SystemToxicology
LCMPhencyclidine (PCP)UnclassifiedToxicology
QBFPropoxyphene (PPX)862.3700 Propoxyphene testsystem.Toxicology
QAWNortriptyline (TCA)862.3910 Tricyclic antidepressantdrugs test systemToxicology
NFWCannabinoids (THC 50)862.3870, Cannabinoids TestSystemToxicology

6. Predicate Device K202567

Wondfo T-Dip® Multi-Drug Urine Test Panel

7. Intended Use

BioSieve™ Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2- ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP 300/OPI 2000)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL

{8}------------------------------------------------

Marijuana (THC)50 ng/mL
---------------------------

BioSieve™ Multi-Drug Urine Test Panel offers any combinations from 1 to 15 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use only. It is intended for OTC use.

The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Nortriptyline, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternative chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

BioSieve™ Multi-Drug Urine Test Panel Rx tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2- ethylidene-1.5-dimethyl-3.3diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Cannabinoids in human urine at the cutoff concentrations of:

DrugCalibratorCut-off (ng/mL)
Amphetamine (AMP)D-Amphetamine1,000 or 500
Secobarbital (BAR)Secobarbital300
Buprenorphine (BUP)Buprenorphine10
Oxazepam (BZO)Oxazepam300
Cocaine (COC )Benzoylecgonine300 or 150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300
Ecstasy (MDMA)D,L- Methylenedioxy-methamphetamine500
Methamphetamine (MET)D-Methamphetamine1,000 or 500
Morphine (MOP/OPI )Morphine2,000 or 300
Methadone (MTD)Methadone300
Oxycodone (OXY )Oxycodone100
Phencyclidine (PCP)Phencyclidine25
Propoxyphene (PPX)Propoxyphene300
Nortriptyline (TCA)Nortriptyline1,000
Marijuana (THC )11-nor-Δ9-THC-9 COOH50

{9}------------------------------------------------

BioSieve™ Multi-Drug Urine Test Panel Rx offers any combinations from 1 to 15 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use only. It is intended for prescription use.

The tests may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene, and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.

Device Description 8.

The BioSieve™ Multi-Drug Urine Test Panel and BioSieve™ Multi-Drug Urine Test Panel Rx are rapid, single-use in vitro diagnostic devices. Each test kit contains a test device in one pouch. One pouch contains a test BioSieve™ Panel and two desiccants, and a package insert. The BioSieve™ Multi-Drug Urine Test Panel is intended for over-the-counter use and the BioSieve™ Multi-Drug Urine Test Panel Rx is intended for prescription use.

ItemProposed DevicePredicate(K202567)
Indication(s) foruseFor the qualitative determination of Amphetamine,Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine,Methamphetamine, Methylenedioxymethamphetamine,Morphine, Methadone, Oxycodone, Phencyclidine,Propoxyphene, Nortriptyline and Cannabinoids in human urine.Same
MethodologyCompetitive binding, lateral flow immunochromatographicassay based on antigen-antibody reactionSame
Type of TestQualitativeSame
Specimen TypeHuman urineSame
Target Drugand Cut OffValuesTarget DrugCutoff (ng/mL)
Amphetamine (AMP)1000 or 500
Buprenorphine (BUP)10
Secobarbital (BAR)300
Oxazepam (BZO)300
Cocaine (COC)300 or 150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300
Methamphetamine (MET)1000 or 500

Substantial Equivalence Information 9.

{10}------------------------------------------------

Methylenedioxymethamphetamine (MDMA)500
Morphine (MOP 300/OPI 2000)2000 or 300
Methadone (MTD)300
Oxycodone (OXY)100
Phencyclidine (PCP)25
Propoxyphene (PPX)300
Nortriptyline (TCA)1000
Cannabinoids (THC 50)50
ConfigurationsTest PanelPanel
Intended UsePrescription Use and over-the-counter usePrescription Use and over-the-counter use

10. Test Principle

BioSieve™ Multi-Drug Urine Test Panel and BioSieve™M Multi-Drug Urine Test Panel Rx are rapid tests for the qualitative detection of Amphetamine, Buprenorphine, Secobarbital. Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Cannabinoids in urine samples. They are lateral flow chromatographic immunoassay. When urine sample is added to the Panel device, urine is absorbed into the test strip and migrates upwards by capillary action. If the concentration of target drug presented in the urine sample is below the cutoff level, the target drug will not saturate the binding sites of its specific monoclonal antibody-coated particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored band will be formed on the test line region. If the concentration of target is beyond the cutoff level, the target drug will saturate the binding sites of its specific monoclonal antibody-particles, thus the antibody-coated particles will not be captured by immobilized drug-conjugate hence no colored band will be formed on the test line region.

A band should be formed on the control line region regardless of the presence of target drug or metabolite in the sample to indicate that the tests have been performed properly.

11. Performance Characteristics

    1. Analytical Performance
    • Precision a.

Precision studies were carried out for samples with concentrations of -100% cut off, - 50% cut off, -25% cut off, cutoff, +25% cut off, +75% cut off, +75% cut off and +100% cut off. Samples with concentration of -100% cutoff were drug-free urines samples. Other samples were prepared by spiking target drug in drug-free urine samples. Each drug concentration was confirmed by LC/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test Panels. The results obtained are summarized in the following tables:

{11}------------------------------------------------

BioSieve™ Multi-Drug Urine Test Panel BUP 10

Concentration byLC/MS(ng/mL)+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
20.017.113.611.810.26.95.42.70
Lot Number
Lot I0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel PCP 25

Concentration byLC/MS(ng/mL)+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
Lot Number52.143.137.329.425.217.712.26.50
Lot I0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel THC 50

Concentration by+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
LC/MScutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
(ng/mL)96.584.375.760.152.535.924.112.10
Lot Number
Lot I0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel OXY 100

Concentration byLC/MS(ng/mL)Lot Number+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
200.3177.1158.9131.7108.578.051.627.60
Lot I0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+

{12}------------------------------------------------

BioSieve™ Multi-Drug Urine Test Panel BAR 300

Concentration byLC/MS(ng/mL)+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
588.4525.8457.6383.8301.6228.3157.180.20
Lot Number
Lot I0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel BZO 300

Concentration by+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
LC/MScutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
(ng/mL)Lot Number596.1536.5470.4370.3290.4219.8157.378.50
Lot I0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel EDDP 300

Concentration by+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
LC/MS(ng/mL)cutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
646.2545.3455.1371.0290.7229.5148.877.00
Lot Number
Lot I0-/50+0-/50+0-/50+0-/50+28-/22+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel MTD 300

Concentration byLC/MS(ng/mL)Lot Number+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
620.5547.0469.9380.9328.6240.2143.971.40
Lot I0-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+

{13}------------------------------------------------

BioSieve™ Multi-Drug Urine Test Panel MOP 300

Concentration by+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
LC/MS(ng/mL)cutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
Lot Number622.4530.2468.8381.8322.8220.8159.075.40
Lot I0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel PPX 300

Concentration by+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
LC/MScutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
(ng/mL)622.3562.7451.3383.3297.7218.3152.575.80
Lot Number
Lot I0-/50+0-/50+0-/50+0-/50+28-/22+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel COC 150

Concentration byLC/MS(ng/mL)Lot Number+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
298.2246.1237.0193.6157.7106.576.236.00
Lot I0-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel MDMA 500

Concentration byLC/MS(ng/mL)+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
Lot Number
1048.5861.5740.8614.9522.8342.0250.6128.40
Lot I0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+

{14}------------------------------------------------

BioSieve™ Multi-Drug Urine Test Panel TCA 1000

Concentration byLC/MS(ng/mL)Lot Number+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
2175.21841.21597.51261.61081.5708.2493.1251.50
Lot I0-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel AMP 500

Concentration+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
by LC/MScutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
(ng/mL)Lot Number1011.8846.4772.7646.8544.3357.6225.0120.90
Lot I0-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel MET 500

Concentration byLC/MS(ng/mL)+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
1072.7873.9731.7633.1477.8386.1249.2122.50
Lot Number
Lot I0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel OPI 2000

Concentration byLC/MS(ng/mL)Lot Number+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
4208.23672.93119.02590.52050.01460.41007.5493.00
Lot I0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+

{15}------------------------------------------------

BioSieve™ Multi-Drug Urine Test Panel COC 300

Concentration by+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
LC/MScutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
(ng/mL)Lot Number610.6558.5461.6373.4329.9235.6156.774.50
Lot I0-/50+0-/50+0-/50+0-/50+24-/26+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel AMP 1000

Concentration by+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
LC/MScutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
(ng/mL)Lot Number1933.31805.21562.71262.01051.1812.0540.9271.90
Lot I0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+25-/25+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+

BioSieve™ Multi-Drug Urine Test Panel MET 1000

Concentration by+100%+75%+50%+25%Cutoff-25%-50%-75%-100%
LC/MScutoffcutoffcutoffcutoffcutoffcutoffcutoffcut-off
(ng/mL)Lot Number1954.61824.21593.11304.91003.4736.7464.8251.10
Lot I0-/50+0-/50+0-/50+0-/50+27-/23+50-/0+50-/0+50-/0+50-/0+
Lot II0-/50+0-/50+0-/50+0-/50+23-/27+50-/0+50-/0+50-/0+50-/0+
Lot III0-/50+0-/50+0-/50+0-/50+26-/24+50-/0+50-/0+50-/0+50-/0+

The following cutoff values are verified:

Target DrugCut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL

{16}------------------------------------------------

Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (MOP 300/OPI 2000)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Cannabinoids (THC 50)50 ng/mL

b. Linearity

Not applicable

c. Stability

The devices are stable at 2-30°C for 24 months based on real time stability studies at 2°C and 30°C.

d. Interference

Potential interfering substances were added to drug-free urine sample and samples with target drugs of -25% cutoff and +25% cutoff level.

Compounds that show no interference at a concentration of 100µg/mL are summarized in the following table.

AcetaminophenEffexorNikethamide
AcetophenetidinEnalapril MaleateNimodipine
Acetylsalicylic AcidEpinephrine HydrochlorideNitroglycerin
AcyclovirErythromycinNorethindrone
AfrinEsomeprazole MagnesiumO-Hydroxyhippuric Acid
Albumin (100mg/dL)ß-EstradiolOlanzapine
AminophyllineEthanol (1%)Omeprazole
AminopyrineFenofibrateOndansetran
Amiodarone HydrochlorideFenoprofenOxalic Acid
Amlodipine MesylateFentanyl CitrateOxolinic Acid
AmoxicillinFluoxetine HydrochlorideOxymetazoline
AmpicillinFluvoxaminePaliperidone
ApomorphineFurosemidePantoprazole
AripiprazoleGabapentinPapaverine
AspartameGentisic AcidParoxetine Hydrochloride
AtomoxetineGlibenclamidePenfluridol
Atorvastatin CalciumGliclazidePenicillin-G
AtropineGlipizidePenicillinV Potassium
Benzilic AcidGlucosePhenelzine
Benzoic AcidHaloperidolPioglitazone Hydrochloride
BilirubinHemoglobinPiracetam
BupropionHydrochlorothiazidePravastatin Sodium
CaptoprilHydrocortisonePrednisone
Carbamazepine3-HydroxytyraminePromethazine
CefradineIbuprofenPropylthiouracil
CephalexinIsosorbide DinitrateQuetiapine Fumarate
Chloral HydrateIsoxsuprineQuinine
ChloramphenicolKetamineRanitidine
ChlorothiazideKetoconazoleRifampicin
chlorpheniramineKetoprofenRisperidone
CholesterolKratom powderSalicylic Acid
Ciprofloxacin HydrochlorideLabetalolSerotonin
CitalopramLamotrigineSertraline Hydrochloride
ClarithromycinLevofloxacin HydrochlorideSildenafil Citrate
ClonidineLevonorgestrelSimvastatin
Clopidogrel Hydrogen SulphateLevothyroxine SodiumSodium Valproate
ClozapineLidocaine HydrochlorideSpironolactone
Conjugated EstrogensLisinoprilSulfamethazine
CortisoneLithium CarbonateSulindac
(-) CotinineLiveriteTetracycline
CreatinineLoperamideTetrahydrocortisone 3- (β-D glucuronide)
D-PseudoephedrineLoratadineTetrahydrocortsone 3 -acetate
D,L-OctopamineMagnesiumTetrahydrozoline
D,L-PropranololMeperidineThiamine
D,L-TyrosineMeprobamateThioridazine
Deoxy- corticosteroneMetoprolol TartrateTopiramate
DextromethorphanMifepristoneTramadol Hydrochloride
DiclofenacMinocyclineTrazodone Hydrochloride
DicyclomineMirtazapineTriamterene
DiflunisalMontelukast SodiumTrifluoperazine
DigoxinMosapride CitrateTrimethoprim
DiphenhydramineN-Acetylprocain-amideUric Acid
DirithromycinNalidixic AcidValproate
DomperidoneNaproxenVerapamil
DoxylamineNiacinamideVitamin B2
DuloxetineNifedipineVitamin C
Ecgonine Methyl Ester

{17}------------------------------------------------

{18}------------------------------------------------

  • Specificity e.
    To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine. These urine samples were tested using three lots of each device.

Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100; compounds that did not yield a positive result at the highest concentration tested have relative cross reactivity results represented by a dash in the table below:

BUP 10 (Buprenorphine,Cutoff=10 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Buprenorphine -3-D-Glucuronide1566.67%
Norbuprenorphine2050%
Norbuprenorphine-3-D-Glucuronide2005%
Morphine>100000-
Oxymorphone>100000-
Hydromorphone>100000-
PCP (Phencyclidine)(Phencyclidine,Cutoff=25 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
4-Hydroxyphencyclidine125000.2%
THC 50(11-nor-Δ9-THC-9-COOH,Cutoff=50 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
(-)-11-nor-9-carboxy-Δ9-THC50100%
11-nor-Δ8-THC-9-COOH50100%
11-nor-Δ9-THC-carboxy glucuronide10050%
Cannabidiol100,000--
Cannabinol100,000--
Δ8- Tetrahydrocannabinol15,0000.3%
Δ9- Tetrahydrocannabinol15,0000.3%
11-hydroxy-Δ9-Tetrahydrocannabinol5,0001%
OXY 100(Oxycodone, Cutoff=100 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Dihydrocodeine20,0000.5%
Hydrocodone80125%

{19}------------------------------------------------

Oxymorphone1,00010%
Codeine100,000--
Hydromorphone36,0000.28%
Morphine100,000--
Acetylmorphine100,000--
Buprenorphine100,000--
Ethylmorphine100,000--
Thebaine100,000--
COC 150(Benzoylecgonine, Cutoff=150 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Cocaine37540%
Cocaethylene6,2502.4%
Ecgonine16,0000.9%
Ecgonine methyl ester100,000--
Norcocaine100,000--
BAR 300(Secobarbital, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Amobarbital300100%
Alphenol60050%
Aprobarbital200150%
Butabarbital100300%
Butethal200150%
Butalbital2,00015%
Cyclopentobarbital40075%
Pentobarbital200150%
Phenobarbital200150%
BZO 300(Oxazepam, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Alprazolam190157.9%
a-Hydroxyalprazolam300100%
Bromazepam50060%
Chlordiazepoxide1,50020%
Clobazam110272.7%
Clonazepam100,000--
Clorazepate dipotassium300100%
Delorazepam100,000--

{20}------------------------------------------------

Desalkylflurazepam200150%
Diazepam190157.9%
Estazolam5,0006%
Flunitrazepam40075%
Midazolam2,20013.6%
Nitrazepam200150%
Norchlordiazepoxide80037.5%
Nordiazepam150200%
Temazepam100300%
Triazolam6,0005%
Demoxepam2,00015%
Flurazepam100,000--
D,L-Lorazepam75,0000.4%
EDDP 300(2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Cutoff = 300 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Methadone>100000-
EMDP>100000-
Doxylamine>100000-
Disopyramide>100000-
LAAM (Levo-alpha-acetylmethadol) HCl>100000-
Alpha Methadol>100000-
MTD 300(Methadone, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Doxylamine>100000-
EDDP>100000-
EMDP>100000-
LAAM>100000-
Alpha Methadol>100000-
MOP 300(Morphine, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Normorphine300100%
Codeine300100%
s-Monoacetylmorphine300100%
Ethyl Morphine200150%
Heroin300100%
Hydrocodone70042.9%

{21}------------------------------------------------

Hydromorphone200150%
Morphinie-3-β-d-glucuronide1,00030%
Oxycodone100,000--
Oxymorphone100,000--
Thebaine20,0001.5%
Levorphanol10,0003%
6-Monoacetylmorphine (6-MAM)300100%
Norcodeine6,2504.8%
Procaine100,000--
PPX 300(d-Propoxyphene, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
d-Norpropoxyphene300100%
MDMA 500(3,4-Methylenedioxymethamphetamine HCl,Cutoff=500ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
3,4-Methylenedioxyamphetamine HCl (MDA)4,00012.5%
3,4-Methylenedioxyethylamphetamine (MDE)400125%
d-methamphetamine>100000-
d-amphetamine>100000-
l-methamphetamine>100000-
l-amphetamine>100000-
AMP (Amphetamine) (Amphetamine,Cutoff=500ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
l-Amphetamine>100000-
dl- Amphetamine1,50033.3%
(+/-) 3,4-Methylenedioxyamphetamine (MDA)500100%
Phentermine6,0008.3%
Hydroxyamphetamine>100000-
d-Methamphetamine>100000-
l-Methamphetamine>100000-
(+/-) 3,4-Methylenedioxyethylamphetamine (MDE)>100000-
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)>100000-
β-Phenylethylamine>100000-
Tyramine>100000-
p-Hydroxynorephedrine>100000-

{22}------------------------------------------------

Phenylpropanolamine>100000-
(+)Phenylpropanolamine>100000-
p-Hydroxyamphetamine>100000-
d/l-Norephedrine>100000-
Benzphetamine>100000-
l-Ephedrine>100000-
l-Epinephrine>100000-
d/l-Epinephrine>100000-
Ephedrine>100000-
MET 500(D(+)-Methamphetamine, Cutoff=500ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDE)12,5004%
D/L-Methamphetamine500100%
p-Hydroxymethamphetamine15,0003.3%
D-Amphetamine>100000-
L-Amphetamine>100000-
Chloroquine50,0001%
(+/-)-Ephedrine100,000--
(-)-Methamphetamine65,0000.8%
(+/-)3,4-Methylenedioxyamphetamine (MDA)>100000-
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)4,00012.5%
β-Phenylethylamine25,0002%
Trimethobenzamide10,0005%
d,l-Amphetamine>100000-
Mephentermine25,0002%
(1R,2S)-(-)-Ephedrine>100000-
1-phenylephrine>100000-
L-Methamphetamine65,0000.8%
TCA 1000(Nortriptyline, Cutoff=1000ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Nordoxepine1,000100%
Trimipramine3,00033.3%
Amitriptyline450222.2%
Promazine1,50066.7%
Desipramine200500%

{23}------------------------------------------------

Imipramine801250%
Clomipramine1,20083.3%
Doxepin2,00050%
Maprotiline2,00050%
Promethazine>100,000--
Cyclobenzaprine800125%
Norclomipramine12,5008%
COC 300(Benzoylecgonine, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Cocaine78038.5%
Cocaethylene12,5002.4%
Ecgonine32,0000.9%
Ecgonine methyl ester>100000-
Norcocaine>100000-
AMP 1000(d-Amphetamine, Cutoff=1000ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
l-Amphetamine>100000-
dl- Amphetamine3,00033.3%
(+/-)3,4-Methylenedioxyamphetamine (MDA)1,000100%,
Phentermine6,00016.7%
Hydroxyamphetamine>100000-
d-Methamphetamine>100000-
l-Methamphetamine>100000-
(+/-)3,4-Methylenedioxyethylamphetamine(MDE)>100000-
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)>100000-
β-Phenylethylamine>100000-
Tyramine>100000-
p-Hydroxynorephedrine>100000-
Phenylpropanolamine>100000-
(±)Phenylpropanolamine>100000-
p-Hydroxyamphetamine>100000-
d/l-Norephedrine>100000-
Benzphetamine>100000-
1-Ephedrine>100000-
l-Epinephrine>100000-
d/l-Epinephrine>100000-

{24}------------------------------------------------

Company of Children Company of Children Company Company Comments of Children Comments of Children Comments of Children Comments of Children Comments of Children Comments of CEphedrine>100000
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
MET 1000(D(+)-Methamphetamine, Cutoff=1000ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDE)25,0004%
D/L-Methamphetamine1,000100%
p-Hydroxymethamphetamine30,0003.3%
D-Amphetamine>100000-
L-Amphetamine>100000-
Chloroquine50,0002%
(+/-)-Ephedrine>100000-
(-)-Methamphetamine>100000-
(+/-)3,4-Methylenedioxyamphetamine (MDA)>100000-
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)8,00012.5%
β-Phenylethylamine50,0002%
Trimethobenzamide20,0005%
d,l-Amphetamine>100000-
Mephetermine50,0002%
(1R,2S)-(-)-Ephedrine>100000-
L-phenylephrine>100000-
L-Methamphetamine>100000-
OPI 2000(Morphine, Cutoff=2000ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Normorphine50,0004%
Codeine2,000100%
s-Monoacetylmorphine2,000100%
Ethyl Morphine1,500133.3%
Heroin2,000100%
Hydrocodone12,50016%
Hydromorphone3,50057.1%
Morphinie-3-β-d-glucuronide2,000100%
Oxycodone25,0008%
Oxymorphone25,0008%
Thebaine50,0004%
Levorphanol75,0002.7%
6-Monoacetylmorphine (6-MAM)2,000100%

{25}------------------------------------------------

Norcodeine12,50016%
Procaine>100,000--

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity, urine samples with specific gravity from 1.000 to 1.035 were spiked with target drugs at +25% cutoff and -25% cutoff levels. Three Operators tested each sample using test devices from three different lots. The results were all positive for samples at +25% cutoff and all negative for samples at -25% cutoff, indicating that urine specific gravity between 1.000 and 1.035 has no effect on the accuracy and precision of the test device.

To investigate the effect of urine pH, urine samples with pH value from 4 to 9 were spiked with target drugs at +25% cutoff and -25% cutoff levels. Three Operators tested each sample using test devices from three different lots. The results were all positive for samples at +25% cutoff and all negative for samples at -25% cutoff, indicating that urine pH value between 4.0 and 9.0 has no effect on the accuracy and precision of the test device.

  • g. Reading Time Study
    Reading time studies were performed for drug free urine samples and urine samples spiked with drug concentrations of -50% cutoff, -25% cutoff, +25% cutoff and +50% cutoff. It demonstrated that test results can be read from 5 to 10 minutes.

  • Comparison Studies 2.
    The method comparison studies for BioSieve™ Multi-Drug Urine Test Panel were performed inhouse with three operators.

Operators ran 80 (40 negative and 40 positive) unaltered urine samples were blind labeled and compared to LC/MS results. The results are presented in the table below:

For BioSieve™ Multi-Drug Urine Test Panel:

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0022020
Negative10121600
Operator BPositive0002020
Negative10121800
Operator CPositive0012020

AMP 500

{26}------------------------------------------------

Negative10121700
------------------------------

Discordant Results for AMP 500:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator AAMP132385.9+
Operator AAMP136499.0+
Operator CAMP028477.4+

BUP 10

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0011820
Negative10151420
Operator BPositive0021920
Negative10151310
Operator CPositive0021920
Negative10151310

Discordant Results for BUP 10:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator ABUP01011.3-
Operator ABUP02910.2-
Operator BBUP02910.2-
Operator CBUP02910.2-
Operator ABUP0709.9+
Operator BBUP0529.5+
Operator BBUP0709.9+
Operator CBUP0589.8+
Operator CBUP0709.9+

BAR 300

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0011720
Negative10161330
Operator BPositive0001920

{27}------------------------------------------------

Negative10161410
Operator CPositive0011820
Negative10161320

Discordant Results for BAR 300:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator ABAR017300.9-
Operator ABAR011303.8-
Operator ABAR033312.2-
Operator BBAR017300.9-
Operator CBAR017300.9-
Operator CBAR011303.8-
Operator ABAR054285.0+
Operator CBAR054285.0+

BZO 300

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0001723
Negative10151500
Operator BPositive0001523
Negative10151520
Operator CPositive0001523
Negative10151520

Discordant Results for BZO 300:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator BBZO018303.6-
Operator BBZO058307.2-
Operator CBZO018303.6-
Operator CBZO058307.2-

COC 150

BioSieve™PanelDrug-FreeLowNegative byLC/MS(less than -50%)Near CutoffNegative byLC/MSNear CutoffPositive byLC/MSHighPositive byLC/MS(greaterthan +50%)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{28}------------------------------------------------

(Between -50% and theCutoff)(Between thecutoff and+50%)
Operator APositive0011822
Negative10161300
Operator BPositive0011722
Negative10161310
Operator CPositive0021722
Negative10161210

Discordant Results for COC 150:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator ACOC046144.8+
Operator BCOC146148.9+
Operator CCOC046144.8+
Operator CCOC146148.9+
Operator BCOC128162.8-
Operator CCOC128162.8-

EDDP 300

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50% and theCutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0011721
Negative10151420
Operator BPositive0021921
Negative10151300
Operator CPositive0011821
Negative10151410

Discordant Results for EDDP 300:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator AEDDP069288.4+
Operator BEDDP069288.4+
Operator BEDDP075290.6+
Operator CEDDP075290.6+
Operator AEDDP010318.6-
Operator AEDDP061318.5-
Operator CEDDP010318.6-

{29}------------------------------------------------

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0001820
Negative10151520
Operator BPositive0001920
Negative10151510
Operator CPositive0001920
Negative10151510

Discordant Results for MET 500:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator AMET062519.5-
Operator AMET102521.1-
Operator BMET102521.1-
Operator CMET062519.5-

MDMA 500

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0022020
Negative10171100
Operator BPositive0012020
Negative10171200
Operator CPositive0022020
Negative10171100

Discordant Results for MDMA 500:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator AMDMA026488.3+
Operator AMDMA060492.0+
Operator BMDMA060492.0+
Operator CMDMA026488.3+
Operator CMDMA060492.0+

MOP 300

{30}------------------------------------------------

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive
Negative10141510
Operator BPositive0021822
Negative10141400
Operator CPositive0021822
Negative10141400

Discordant Results for MOP 300:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator AMOP057293.2+
Operator BMOP150282.8+
Operator BMOP057293.2+
Operator CMOP057293.2+
Operator CMOP150282.8+
Operator AMOP114315.6-

MTD 300

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021820
Negative10161220
Operator BPositive0011920
Negative10161310
Operator CPositive0012020
Negative10161300

Discordant Results for MTD 300:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator AMTD022298.2+
Operator AMTD049289.2+
Operator BMTD022298.2+
Operator CMTD049289.2+
Operator AMTD003309.1-

{31}------------------------------------------------

Operator AMTD045301.7-
Operator BMTD045301.7-

OXY 100

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021921
Negative10141400
Operator BPositive0011821
Negative10141510
Operator CPositive0011721
Negative10141520

Discordant Results for OXY 100:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator AOXY01296.3+
Operator AOXY07195.2+
Operator BOXY07195.2+
Operator COXY01296.3+
Operator BOXY002101.4-
Operator COXY002101.4-
Operator COXY006111.0-

РСР 25

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0
Negative10181020
Operator BPositive0012118
Negative10181110
Operator CPositive0022118
Negative10181010

Discordant Results for PCP 25:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator APCP00322.4+

{32}------------------------------------------------

Operator APCP03420.2+
Operator BPCP06022.7+
Operator CPCP00322.4+
Operator CPCP03420.2+
Operator APCP01229.2-
Operator APCP02325.5-
Operator BPCP02325.5-
Operator CPCP01229.2-

PPX 300

BioSieveTMPanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive00217
Negative10161220
Operator BPositive0011721
Negative10161320
Operator CPositive0021821
Negative10161210

Discordant Results for PPX 300:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator APPX024292.5+
Operator APPX029291.4+
Operator BPPX029291.4+
Operator CPPX024292.5+
Operator CPPX029291.4+
Operator APPX043300.7-
Operator APPX053300.8-
Operator BPPX043300.7-
Operator BPPX053300.8-
Operator CPPX043300.7-

TCA 1000

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0011622

{33}------------------------------------------------

Negative10151420
Operator BPositive0011722
Negative10151410
Operator CPositive0021722
Negative10151310

Discordant Results for TCA 1000:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator ATCA043969.0+
Operator BTCA005991.3+
Operator CTCA005991.3+
Operator CTCA043969.0+
Operator ATCA0101015.1-
Operator ATCA0521015.9-
Operator BTCA0521015.9-
Operator CTCA0101015.1-

THC 50

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021622
Negative10161220
Operator BPositive0011622
Negative10161320
Operator CPositive0011622
Negative10161320

Discordant Results for THC 50:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator ATHC01447.8+
Operator ATHC05446.8+
Operator BTHC05446.8+
Operator CTHC01447.8+
Operator ATHC06250.9-
Operator ATHC06953.5-
Operator BTHC06953.5-
Operator BTHC07653.9-
Operator CTHC03650.5-

{34}------------------------------------------------

Operator CTHC06250.9-
AMP 1000
BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0001721
Negative10151520
Operator BPositive0001821
Negative10151510
Operator CPositive0011821
Negative10151410

Discordant Results for AMP 1000:

OperatorSample NumberLC/MS Result (ng/mL)Result
Operator CAMP116998.8+
Operator AAMP0951035.1-
Operator AAMP1021048.4-
Operator BAMP1021048.4-
Operator CAMP1021048.4-

COC 300

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021623
Negative10141410
Operator BPositive0011523
Negative10141520
Operator CPositive0021723
Negative10141400

Discordant Results for COC 300:

OperatorSample NumberLC/MS ResultResult
Operator ACOC028296.4+
Operator ACOC143283.8+
Operator BCOC028296.4+

{35}------------------------------------------------

Operator CCOC028296.4+
Operator CCOC143283.8+
Operator ACOC138318.7-
Operator BCOC033317.7-
Operator BCOC138318.7-

MET 1000

BioSieveTMPanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0002019
Negative10131710
Operator BPositive0001919
Negative10131720
Operator CPositive0002019
Negative10131710

Discordant Results for MET 1000:

OperatorSample NumberLC/MS ResultResult
Operator AMET1231049.5-
Operator BMET1231049.5-
Operator BMET1381068.2-
Operator CMET1381068.2-

OPI 2000

BioSieve™PanelDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and +50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0001622
Negative10161420
Operator BPositive0011822
Negative10161300
Operator CPositive0001722
Negative10161410

Discordant Results for OPI 2000:

OperatorSample NumberLC/MS ResultResult
Operator BMOP0761943.3+

{36}------------------------------------------------

Operator AMOP0892070.0-
Operator AMOP1392105.7-
Operator CMOP1252156.7-

Lay-user study:

A lay user study was performed using urine samples prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS or LC/MS. Each sample was aliquoted into individual containers and blind-labeled. A total of 280 participants with diverse educational and professional backgrounds aged 20 years and older were recruited from three sites. Sixty-four males and 76 females tested BioSieve™ Multi-Drug Urine Test Panel Configuration 1 (including AMP 500, MET 500, MOP 300, COC 150); 67 male and 73 females tested BioSieve™ Multi-Drug Urine Test Panel Configuration 2 (including AMP 1000, MET 1000, MOP 2000 (OPI), COC 300). Each participant was provided one package insert, one blind labeled test solution, and one test device. The results are summarized below:

Lay-User Study Results for BioSieve™ Multi-Drug Urine Test Panel Configuration 1 (including AMP 500, MET 500, MOP 300, COC 150):

Concentration
AssayResults-100%cutoff-75%cutoff-50%cutoff-25%cutoff+25%cutoff+50%cutoff+75%cutoff
Negative20202020100
Positive0000192020
AMPTotal20202020202020
Agreement (%)10010010010095100100
Negative20202019100
Positive0001192020
BUPTotal20202020202020
Agreement (%)1001001009595100100
Negative20202019000
Positive0001202020
BARTotal20202020202020
Agreement (%)10010010095100100100
Negative20202020100
Positive0000192020
BZOTotal20202020202020
Agreement (%)10010010010095100100
Negative20202018100
COCPositive0002192020
Total20202020202020
Agreement (%)1001001009095100100
EDDPNegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
MDMANegative20202020000
Positive0000202020
Total20202020202020
Agreement (%)100100100100100100100
METNegative20202019100
Positive0001192020
Total20202020202020
Agreement (%)1001001009595100100
MOPNegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
MTDNegative20202020100
Positive0000192020
Total20202020202020
Agreement (%)10010010010095100100
OXYNegative20202019100
Positive0001192020
Total20202020202020
Agreement (%)1001001009595100100
PCPNegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
PPXNegative20202020000
Positive0000202020
Total20202020202020
Agreement (%)100100100100100100100
TCANegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
THCNegative20202019200
Positive0001182020
Total20202020202020
Agreement (%)1001001009590100100
AssayResultsConcentration
-100% cutoff-75% cutoff-50% cutoff-25% cutoff+25% cutoff+50% cutoff+75% cutoff
AMPNegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
BUPNegative20202020100
Positive0000192020
Total20202020202020
Agreement (%)10010010010095100100
BARNegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
BZONegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
COCNegative20202019100
Positive0001192020
Total20202020202020
Agreement (%)1001001009595100100
EDDPNegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
MDMANegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
METNegative20202020000
Positive0000202020
Total20202020202020
Agreement (%)100100100100100100100
OPINegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
Agreement (%)10010010095100100100
MTDNegative20202020100
Positive0000192020
Total20202020202020
Agreement (%)10010010010095100100
OXYNegative20202019100
Positive0001192020
Total20202020202020
Agreement (%)10010010010095100100
PCPNegative20202020100
Positive0000192020
Total20202020202020
Agreement (%)10010010010095100100
PPXNegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010010095100100
TCANegative20202019000
Positive0001202020
Total20202020202020
Agreement (%)10010010095100100100
THCNegative20202018000
Positive0002202020
Total20202020202020
Agreement (%)10010010090100100100

{37}------------------------------------------------

{38}------------------------------------------------

Lay-User Study Results for BioSieve™ Multi-Drug Urine Test Panel Configuration 2 (AMP 1000, MET 1000, MOP 2000 (OPI), COC 300):

{39}------------------------------------------------

Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

12. Conclusion

Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that BioSieve™ Multi-Drug Urine Test Panel and BioSieve™ Multi-Drug Urine Test Panel Rx are substantially equivalent to the predicate devices.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).