METHAMPHETAMINE ORAL FLUID HOMOGENEOUS ENZYME IMMUNOASSAY; CALIBRATORS AND CONTROLS
K062242 · Lin-Zhi International, Inc. · LAF · Sep 29, 2006 · Clinical Toxicology
Device Facts
Record ID
K062242
Device Name
METHAMPHETAMINE ORAL FLUID HOMOGENEOUS ENZYME IMMUNOASSAY; CALIBRATORS AND CONTROLS
Applicant
Lin-Zhi International, Inc.
Product Code
LAF · Clinical Toxicology
Decision Date
Sep 29, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3610
Device Class
Class 2
Indications for Use
The Methamphetamine Enzyme Immunoassays for Drugs of Abuse in Oral Fluid is a homogeneous enzyme immunoassay system to detect methamphetamine in human saliva with a cutoff of 45 ng/mL when testing oral fluid specimens collected with a Salivette collector (manufactured by Sarstedt) and diluted with 1 mL of buffer. The calibrators and controls of the analyte (d-methamphetamine) are prepared with oral fluid buffer so that it can be used to verify and validate the assay. The assay is intended for use in the qualitative determination for methamphetamine. The assay is designed for professional use with a number of automated clinical chemistry analyzers. The Oral Fluid Methamphetamine Enzyme Immunoassay is a homogeneous enzyme immunoassay system provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
Device Story
Ready-to-use liquid reagent homogeneous enzyme immunoassay; detects methamphetamine in human saliva. Principle: competition between drug-labeled G6PDH enzyme and free drug in saliva sample for fixed amount of specific antibody. In absence of free drug, antibody binds labeled enzyme, decreasing activity. Enzyme activity measured spectrophotometrically at 340 nm via conversion of NAD to NADH; drug concentration proportional to enzyme activity. Used in clinical laboratories by professionals on automated clinical chemistry analyzers. Provides preliminary analytical results; requires confirmation by GC/MS. Assists clinicians in identifying methamphetamine presence in oral fluid specimens.
Clinical Evidence
Bench testing only. Method comparison study of 112 clinical oral fluid specimens against GC/MS reference method. Results: 71 true positives, 39 true negatives, 2 false positives (near cutoff). Overall agreement 98.2%. Precision studies (n=6, twice daily for 10 days) showed within-run %CV 0.39-0.64% and total precision %CV 0.58-0.80%. Analytical specificity evaluated against various cross-reactants and interferents.
Technological Characteristics
Homogeneous enzyme immunoassay; liquid reagents (mouse monoclonal antibodies, G6PDH conjugate). Sensing principle: spectrophotometric measurement of NADH at 340 nm. Designed for automated clinical chemistry analyzers. Storage: 2-8 °C. Stability: 18 months shelf life; 22 days for collected samples (refrigerated or frozen).
Indications for Use
Indicated for qualitative detection of methamphetamine in human saliva for professional use in clinical chemistry laboratories. Not for point-of-care settings.
Regulatory Classification
Identification
A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.
Special Controls
*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
K063024 — LZI AMPHETAMINE-SPECIFIC ORAL FLUID HOMOGENEOUS ENZTME IMMUNOASSAY: LZI AMPHETAMINE ORAL FLUID CALIBRATORS AND CONTROLS · Lin-Zhi International, Inc. · Oct 30, 2006
K051058 — METHADONE ORAL FLUID HOMOGENEOUS ENZYME IMMUNOASSAY, CALIBRATORS AND CONTROLS · Lin-Zhi International, Inc. · Mar 29, 2006
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k062242
B. Purpose for Submission:
New device
C. Measurand:
Methamphetamine
D. Type of Test:
Qualitative Enzyme Immunoassay
E. Applicant:
Lin-Zhi International, Inc.
F. Proprietary and Established Names:
LZI Oral Fluid Methamphetamine Enzyme Immunoassay
LZI Oral Fluid Methamphetamine Calibrators
LZI Oral Fluid Methamphetamine Controls
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.3610, Methamphetamine test system
21 CFR § 862.3200, Clinical toxicology calibrator
21 CFR § 862.3280, Clinical toxicology control material
2. Classification:
Class II
3. Product code:
LAF
DLJ
LAS
{1}
4. Panel:
Toxicology (91)
H. Intended Use:
1. Intended use(s):
Refer to Indications for Use
2. Indication(s) for use:
The Methamphetamine Enzyme Immunoassays for Drugs of Abuse in Oral Fluid is a homogeneous enzyme immunoassay system to detect methamphetamine in human saliva with a cutoff of 45 ng/mL when testing oral fluid specimens collected with a Salivette collector (manufactured by Sarstedt) and diluted with 1 mL of buffer. The calibrators and controls of the analyte (d-methamphetamine) are prepared with oral fluid buffer so that it can be used to verify and validate the assay. The assay is intended for use in the qualitative determination for methamphetamine. The assay is designed for professional use with a number of automated clinical chemistry analyzers.
The Oral Fluid Methamphetamine Enzyme Immunoassay is a homogeneous enzyme immunoassay system provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
3. Special conditions for use statement(s):
For prescription, professional use only in clinical chemistry laboratories.
The assay is not designated for use in point-of-care settings.
4. Special instrument requirements:
The device is designed for use on automated clinical chemistry analyzers.
Performance data for this submission was collected using a Hitachi 717 analyzer.
I. Device Description:
The assay consists of ready-to-use liquid reagents. Reagent 1 contains mouse monoclonal antibodies to methamphetamine, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide as preservative. Reagent 2 contains methamphetamine-labeled glucose-6-phosphate dehydrogenase (G6PDH) in buffer with sodium azide as a preservative.
Calibrators for this assay are sold separately and consist of a negative, cutoff, and high calibrator in an oral fluid buffer, ranging in concentration from 0 to 90 ng/mL.
Assayed negative and positive controls are sold separately and are run with unknown
2
{2}
samples to monitor the performance of the assay. They consist of oral fluid to which methamphetamine has been added.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
- OraSure Methamphetamine Intercept® Micro-plate EIA
- LZI Oral Fluid Cocaine Enzyme Immunoassay
- LZI Oral Fluid Opiate Enzyme Immunoassay
- LZI Oral Fluid Methadone Enzyme Immunoassay
2. Predicate 510(k) number(s):
- k993208
- k050945
- k050988
- k051058
3. Comparison with predicates:
| Similarities | | |
| --- | --- | --- |
| Item | Device | OraSure Methamphetamine Intercept® Micro-plate EIA |
| Intended Use | Qualitative detection of methamphetamine | Same |
| Assay Type | Enzyme immunoassay | Same |
| Matrix | Oral Fluid | Same |
| Controls | 2 levels | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Cutoff | 45 ng/mL | 40 ng/mL |
| Calibrators | 3 levels | 2 levels |
The LZI predicate devices are similar to the new device in terms of matrix, calibrators, controls, and assay type. They differ in that they measure different analytes than the new device.
## K. Standard/Guidance Document Referenced (if applicable):
- CLSI EP5-A, Evaluation of Precision Performance of Clinical Chemistry
- CLSI EP7-A, Interference Testing in Clinical Chemistry
- CLSI EP9-A, Method Comparison and Bias Estimation Using Patient Samples
- CLSI ILA21-A, Clinical Evaluation of Immunoassays
{3}
L. Test Principle:
The assay is based on competition for anti-methamphetamine antibody binding sites between methamphetamine in the sample and methamphetamine conjugated to glucose-6-phosphate dehydrogenase (G6PDH). In the absence of free drug in the sample, the antibody binds the conjugated methamphetamine thus decreasing the enzymatic activity of the G6PDH. The G6PDH activity is measured spectrophotometrically at 340 nm due to conversion of NAD to NADH.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
Performance data for this submission was collected using a Hitachi 717 analyzer.
a. Precision/Reproducibility:
Precision studies were performed by using the negative, cutoff and high calibrators and two levels of controls. Testing was performed in replicates of 6, twice a day for 10 days for all concentrations. The within-run % CV ranged from 0.39 to 0.64 % and the Total precision %CV ranged from 0.58 to 0.80%.
The recovery study showed that the assay correctly identified spiked samples above the cutoff as positive and below the cutoff as negative.
b. Linearity/assay reportable range:
Not applicable. This is a qualitative assay.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
A commercially available standard solution is made into a secondary (lower concentration) stock solution. The secondary stock solution is then spiked into the calibrators and controls to the desired concentration. The concentrations are confirmed by GC/MS.
Stability Studies: Real time and accelerated studies have been conducted. Protocols and acceptance criteria were described and found to be acceptable. The manufacturer claims the following expiration date:
When stored at 2-8 °C product is good until expiration date which is 18 months.
The stability of methamphetamine in the Salivette collection device was determined by taking a pool of negative oral fluid samples spiked at three different concentrations and split into three hundred and thirty samples. On day one 10 samples were run to determine the baseline that subsequent runs were compared to. Of the remaining one hundred samples for each concentration, half were stored at 2-8 °C and the other half were stored frozen
4
{4}
-20 °C. The study was conducted over 22 days and samples were run on days 1, 2, 8, 12, 15 and 22. The sample is stable for 22 days when stored refrigerated at 2-8 °C or frozen at -20 °C.
d. Detection limit:
See the Precision/Reproducibility section above for performance around the stated cutoff concentration.
e. Analytical specificity:
Various potentially cross-reacting or interfering substances were evaluated to determine whether they interfere with assay results. Test compounds were spiked into the drug-free calibrator to various concentrations and evaluated against the cutoff calibrator. Results were as follows:
| Potential Cross-reactant | Conc. (ng/mL) | X-reactivity |
| --- | --- | --- |
| d-Methamphetamine | 30 | Positive |
| l-Methamphetamine | 500 | Negative |
| d-Amphetamine | 1000 | Negative |
| l-Amphetamine | 10000 | Negative |
| Potential Interferent | Conc. (μg/mL) | X-reactivity |
| --- | --- | --- |
| Acetaminophen | 100 | Negative |
| Acetylsalicylic acid | 250 | Negative |
| Amobarbital | 100 | Negative |
| Benzoylecgonine | 100 | Negative |
| Benzphetamine | 50 | Negative |
| Brompheniramine | 200 | Negative |
| Bupropion | 250 | Negative |
| Buspirone | 250 | Negative |
| Caffeine | 300 | Negative |
| Chlorpheniramine | 250 | Negative |
| Chlorpromazine | 300 | Negative |
| Codeine | 250 | Negative |
| Dextromethorphan | 300 | Negative |
| Doxepine | 50 | Negative |
| d-Ephedrine | 100 | Negative |
| d,l-Ephedrine | 45 | Negative |
| l-Ephedrine | 7 | Negative |
| Fenfluramine | 1 | Negative |
| 3-OH Tyramine | 125 | Negative |
| Isoxsuprine | 250 | Negative |
| Meperidine | 200 | Negative |
{5}
| Potential Interferent | Conc. (μg/mL) | X-reactivity |
| --- | --- | --- |
| Mephentermine | 3 | Negative |
| Methadone | 200 | Negative |
| Morphine | 200 | Negative |
| Oxazepam | 200 | Negative |
| Phencyclidine | 25 | Negative |
| Phenethylamine | 5 | Negative |
| Phenmetrazine | 15 | Negative |
| Phenobarbital | 200 | Negative |
| Phentermine | 5 | Negative |
| d-Phenylpropanolamine | 50 | Negative |
| d,l-Phenylpropanolamine | 12.5 | Negative |
| l-Phenylpropanolamine | 6.25 | Negative |
| Procainamide | 125 | Negative |
| Promethazine | 300 | Negative |
| Propoxyphene | 200 | Negative |
| Propranolol | 300 | Negative |
| d-Pseudoephedrine | 5 | Negative |
| l-Pseudoephedrine | 100 | Negative |
| Ranitidine | 10 | Negative |
| Secobarbital | 200 | Negative |
| Sertraline | 125 | Negative |
| Trazodone | 100 | Negative |
| Trifluoperazine | 250 | Negative |
| Tyramine | 50 | Negative |
| Valproic Acid | 300 | Negative |
f. Assay cut-off:
The stated cutoff of this assay, which includes the dilution of the sample with the Salivette collection device, is 45 ng/mL. Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision/reproducibility section above.
2. Comparison studies:
a. Method comparison with predicate device:
One hundred and twelve clinical oral fluid specimens were collected. The oral fluid specimens were tested with LZI Oral Fluid Methamphetamine Enzyme Immunoassay and compared to a Gas Chromatography/ Mass Spectrophotometer (GC/MS).
Results from the study are presented below. The table describes the agreement between the device and the GC/MS.
{6}
| | GC-MS | | |
| --- | --- | --- | --- |
| | | Positive | Negative |
| LZI EIA | Positive | 71 | 2* |
| | Negative | 0 | 39 |
*The two discrepant samples contained 41 and 44 ng/mL methamphetamine
% Agreement = 98.2%
b. Matrix comparison:
Not applicable. This device is intended for use with oral fluid only.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.