WONDFO MULTI-DRUG URINE TEST CUP WONDFO MULTI-DRUG URINE TEST PANEL
Device Facts
| Record ID | K121166 |
|---|---|
| Device Name | WONDFO MULTI-DRUG URINE TEST CUP WONDFO MULTI-DRUG URINE TEST PANEL |
| Applicant | Guangzhou Wondfo Biotech Co., Ltd. |
| Product Code | DIO · Clinical Toxicology |
| Decision Date | May 8, 2012 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.3250 |
| Device Class | Class 2 |
Indications for Use
Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Morphine 2000, Phencyclidine, Notriptyline in human urine at the cutoff concentrations of: Cocaine /COC 300 ng/mL; Methamphetamine (MET) 1000 ng/mL; Methylenedioxymethamphetamine (MDMA) 500 ng/mL; Morphine (MOP) 300 ng/mL; Methadone (MTD) 300 ng/mL; Morphine 2000 (OPI) 2000 ng/mL; Phencyclidine (PCP) 25 ng/mL; Notriptyline (TCA) 1000 ng/mL. Configuration of the Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel can consist of any combination of the above listed drug analytes. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
Device Story
Lateral flow immunochromatographic assay for qualitative detection of drugs of abuse in human urine; competitive binding principle. Input: human urine sample. Output: visual colored lines on test strip; absence of line in test region indicates positive result (drug concentration above cutoff); presence of line indicates negative result or concentration below cutoff; control line confirms test validity. Used in clinical or home settings; operated by healthcare professionals or lay users. Provides preliminary results requiring confirmation via GC/MS. Assists in clinical decision-making regarding drug screening.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding format. Multi-drug test cup or panel configuration (2-8 analytes). Visual readout via colored lines. Standalone device; no energy source required. No software or electronic components.
Indications for Use
Indicated for the qualitative detection of drugs of abuse in human urine samples.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- Wondfo Cocaine Urine Test (k112071)
- Wondfo Methamphetamine Test (k112071)
Related Devices
- K150356 — Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card · Healgen Scientific,, LLC · May 13, 2015
- K151642 — Heagen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card · Healgen Scientific,, LLC · Aug 13, 2015
- K130665 — WONDOFO MULTI-DRUG URINE TEST CUP / PANEL · Guangzhou Wondfo Biotech Co., Ltd. · Apr 9, 2013
- K180255 — CLUNGENE Multi-Drug Test Dip Card, CLUNGENE Multi-Drug Test Easy Cup · Hangzhou Clongene Biotech Co., Ltd. · Feb 28, 2018