DEN180008 · MolecuLight, Inc. · QCR · Jul 31, 2018 · General, Plastic Surgery
Device Facts
Record ID
DEN180008
Device Name
MolecuLight i:X
Applicant
MolecuLight, Inc.
Product Code
QCR · General, Plastic Surgery
Decision Date
Jul 31, 2018
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 878.4165
Device Class
Class 1
Indications for Use
The MolecuLight i:X is a handheld imaging tool that allows clinicians diagnosing and treating skin wounds, at the point of care, to view and digitally record images of a wound, and (i) view and digitally record images of fluorescence emitted from a wound (ii) when exposed to an excitation light. The MolecuLight i:X is for prescription use only.
Device Story
Handheld, battery-operated imaging tool; captures standard white light and fluorescence images of skin wounds. Operates via 450nm excitation light; detects fluorescence in 500-545nm and 600-665nm ranges. Used at point-of-care by clinicians; maintains 8-12cm distance via range finder. Provides visual documentation of wound appearance and fluorescence; does not provide diagnostic or quantitative information. Clinicians use images to monitor wound changes over time. No patient-contacting components; requires intermediate-level disinfection between uses.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing (imaging uniformity, detection depth, linearity, SNR) and a review of standard and fluorescent wound photographs (diabetic foot ulcers, venous leg ulcers, pressure ulcers, surgical wounds, burns) to assess image quality and consistency.
Technological Characteristics
Handheld, battery-operated imaging device. Features 450nm excitation light source; detects fluorescence in 500-545nm and 600-665nm bands. Includes built-in range finder for 8-12cm working distance. Non-sterile; multi-patient reusable. Wireless/Bluetooth connectivity. Software operates on Apple iOS. No patient-contacting materials.
Indications for Use
Indicated for clinicians diagnosing and treating skin wounds to view and record wound images and wound autofluorescence at the point of care.
Regulatory Classification
Identification
A wound autofluorescence imaging device is a tool to view autofluorescence images from skin wounds that are exposed to an excitation light. The device is not intended to provide quantitative or diagnostic information.
Related Devices
K211901 — MolecuLightDX · MolecuLight, Inc. · Jul 21, 2021
K191371 — MolecuLight i:X · MolecuLight, Inc. · Dec 4, 2019
K210882 — MolecuLight I:X · MolecuLight, Inc. · Jun 22, 2021
K242669 — SnapshotGLO (KB100) · Kent Imaging, Inc. · Mar 26, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR MOLECULIGHT I:X
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Wound autofluorescence imaging device. A wound autofluorescence imaging device is a tool to view autofluorescence images from skin wounds that are exposed to an excitation light. The device is not intended to provide quantitative or diagnostic information.
#### NEW REGULATION NUMBER: 21 CFR 878.4165
CLASSIFICATION: Class I (Exempt from premarket notification, subject to the limitations in 21 CFR 878.9)
### PRODUCT CODE: QCR
### BACKGROUND
DEVICE NAME: MolecuLight i:X
SUBMISSION NUMBER: DEN180008
DATE OF DE NOVO: February 16, 2018
- MolecuLight Inc CONTACT: 101 College Street, Suite 200 Toronto. ON M5G 1L7 Canada
#### INDICATIONS FOR USE
The MolecuLight i:X is a handheld imaging tool that allows clinicians diagnosing and treating skin wounds, at the point of care, to
- view and digitally record images of a wound, and (i)
- view and digitally record images of fluorescence emitted from a wound (ii) when exposed to an excitation light.
The MolecuLight i:X is for prescription use only.
### LIMITATIONS
The sale, distribution, and use of MolecuLight are restricted to prescription use in accordance with 21 CFR 801.109.
This device can detect fluorescence at maximum depth of 0.8mm in a wound.
{1}------------------------------------------------
# PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS, AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
The MolecuLight i:X is a handheld, portable, and battery operated white light and fluorescence imaging tool. The MolecuLight i:X can (i) view and digitally record standard (ST) digital images of a wound.
Standard digital images/video capture the appearance of the wound under illumination of broad band white light, like regular photography. This Standard Imaging Mode does not require contact with the patient and the distance between the device and wound is kept consistent (8 - 12 cm) between imaging sessions through the use of the built in range finder
The MolecuLight i:X can also capture fluorescence features of wounds in real-time when used in Fluorescence Imaging Mode and it enables the user to document this fluorescence information. When the level of ambient light in the examination room is acceptable, an indicator light informs the user that fluorescence images (or videos) can be acquired. The distance between the device to the patient is maintained at 8 - 12 cm as guided by the range finder indicator light. The device has 450nm excitation light and is capable to detect fluorescence signals in 500-545 and 600-665nm wavelengths. The Moleculight i:X is not intended to quantify the fluorescence emitted from a wound.
Image /page/1/Picture/5 description: The image shows a blue handheld medical device with a screen displaying an image of what appears to be a wound or tissue sample. The device has a silver-colored top section with indicator lights and a screen in the middle. There is a silver button on the side and a triangular button at the bottom of the device.
Figure 1: MolecuLight i:X
# SUMMARY OF NONCLINICAL/BENCH STUDIES
### BIOCOMPATIBILITY/MATERIALS
There are no patient contacting components.
### SHELF LIFE/STERILITY
No component of the MolecuLight i:X is provided sterile.
{2}------------------------------------------------
The life-time of the MolecuLight i:X has been evaluated to be at least (b) (4)
Shelf-life of the system is greater than 2 years and is determined by the shelf-life of the lithium batteries installed in the product that may slowly discharge into a nonrecoverable state. Other components, such as plastics and electronic printed circuit boards and other circuitry, are not anticipated to degrade in this time, and the device contains no electrolytic capacitors.
# REPROCESSING
The subject device is multi-patient, reusable and is provided non-sterile to the end user. There are no patient contacting components. To mitigate the risk of cross-contamination through indirect patient transmission mechanisms, the subject device is intended to be cleaned and intermediate level disinfected in between uses. The cleaning instructions were validated using an artificial test soil to represent the worse-case constituents (i.e. bloods and protein) of patient material that may contact the device during use. After soiling, the sample was allowed to dry for one hour undisturbed at room temperature. Then the device was cleaned using worst case implementation of the instructions in the user manual. The acceptance criteria of < 6.4 µg/cm² for residual protein and < 2.2 ug/cm2 for residual hemoglobin on the device were met.
The disinfection instructions were validated by inoculating test devices with each of five test microorganisms in suspension with 5% Fetal Bovine Serum, and then disinfecting the device using worst case implementation of the instructions in the user manual. The results demonstrated a minimum 6-log reduction of Staphylococcus aureus, Pseudomonas aeruginosa, Escherichia coli. and Klebsiella pneumoniae, as well as a minimum 3-log reduction of Mycobacterium terrae, thus meeting the acceptance criteria. The testing and results are adequate to support the label cleaning and disinfection instructions for reuse of the MolecuLight i:X.
# ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL/MECHANICAL/THERMAL SAFETY
The following Electrical/Mechanical/Thermal Safety and electromagnetic compatibility (EMC) testing has been performed:
- IEC 60601-1: 2005 +CORR.1:2006+CORR.2: 2007 + AM1:2012 or IEC 60601-. 1:2012 with US deviations. Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2: 2007 (Edition 3), Medical electrical equipment Part 1-2 General . requirements for basic safety and essential performance - Electromagnetic compatibility.
- . IEC 60601-1-6: Collateral Standard: Medical electrical equipment Part 1-6 -General requirements for basic safety and essential performance - Usability.
The MolecuLight i:X device and labeling passed all relevant portions of the testing.
{3}------------------------------------------------
The De Novo request submitted information regarding wireless and Bluetooth connectivity in accordance with the recommendations in the FDA guidance document "Radio Frequency Wireless Technology in Medical Devices."
## SOFTWARE
The device software is a mobile app, operating on Apple iOS operating system. It captures and reviews the standard (white light) and fluorescence images and videos. The software does not save any personal information of the patient. The device software will be upgraded via the Apple App Store.
All of the elements of software and cybersecurity information as outlined in FDA's guidance documents "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005) and "Content of Premarket Submissions for management of Cybersecurity in Medical Devices" were provided.
Overall, the software and cybersecurity documentation included in the De Novo request is in sufficient detail to provide reasonable assurance that the software will operate in a manner described in the specifications.
### PERFORMANCE TESTING - BENCH
Bench testing was conducted to characterize the device performance.
### Imaging field uniformity:
The spatial uniformity of the imaging field was tested with green (fluorescein dye (VWR, CAAAAL13251-22) and red (PpIX / Sigma Aldrich, P8293-1G) fluorescent agents integrated into an optical tissue phantom (made by intralipid solution and diluted hemoglobin in PBS). 2D intensity mapping was used to demonstrate the spatial nonuniformity of the illumination source and fluorescent imager.
{4}------------------------------------------------
Image /page/4/Figure/0 description: The image contains five different diagrams labeled A through E. Diagram A is a grayscale image uniformity map with a scale from 50 to 100%. Diagram B is a green image with two white spots. Diagram C is a green image uniformity map with a scale from 40 to 100%. Diagram D is a red image with two white spots, and diagram E is a red image uniformity map with a scale from 20 to 100%.
### Detection depth:
A red fluorescent phantom (PpIX / Sigma Aldrich, P8293-1G - 1.6mm dia. x 2.5mm depth) was embedded at different depths inside a green fluorescent phantom (VWR. CAAAAL13251-22) both with the same optical tissue properties. The results show the maximum detection depth of the device as 800um.
### Detection limits, linearity, Signal-to-Noise Ratio (SNR):
Different dilutions of green quantum dots (ThermoFisher Scientific, Q10143MP, emission 525 nm) and red quantum dots (ThermoFisher Scientific, Q22063, emission 625 nm) in PBS were used to determine the minimum and maximum fluorescence detection limit of the device at the lowest and highest intensity illuminated area of the imaging field. The same data were used to determine the device linearity and SNR of the device for each spectral bandwidth (green and red).
For each color, the background (B) was considered as mean of the detected fluorescence intensity across all pixels within a blank well (no fluorescent contrast agent). Signal (S) and noise (N) for each fluorescence concentration were considered as the mean and standard deviation of the detected fluorescence intensity across all pixels within the well, respectively. Based on these values, SNR was calculated as the signal with background subtracted, divided by the noise ([S-B]/N). The SNR for the minimum and maximum
{5}------------------------------------------------
fluorescence detection limit in the high and low illumination intensity section of the imaging field are provided in the tables below:
| Item | Specification |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Fluorescence Detection Depth | 0.8 mm |
| Linear range, green, maximum<br>illumination intensity | 1-88% of Full Scale Image Sensor Range 0-125 nM Quantum Dots (ThermoFisher<br>Scientific, Q10143MP, emission 525 nm |
| Linear range, green, minimum<br>illumination intensity | 0-94% of Full Scale Image Sensor Range 0-250 nM Quantum Dots; ThermoFisher<br>Scientific, Q10143MP, emission 525 nm |
| Linear range, red, maximum<br>illumination intensity | 3-90% of Full Scale Image Sensor Range 0.5-5.6 nM Quantum Dots (ThermoFisher<br>Scientific, Q22063, emission 625 nm) |
| Linear range, red, minimum<br>illumination intensity | 2-95% of Full Scale Image Sensor Range 0.5-7.7 nM Quantum Dots (ThermoFisher<br>Scientific, Q22063, emission 625 nm) |
| Limits of detection, green, maximum<br>illumination intensity | 20-111 nM Quantum Dots (ThermoFisher Scientific,<br>Q10143MP, emission 525 nm) |
| Limits of detection, green, minimum<br>illumination intensity | 20-182 nM Quantum Dots (ThermoFisher Scientific,<br>Q10143MP, emission 525 nm) |
| Limits of detection, red, maximum<br>illumination intensity | 0.7-4.6 nM Quantum Dots (Thermo Fisher Scientific,<br>Q22063, emission 625 nm) |
| Limits of detection, red, minimum<br>illumination intensity | 0.5-5.6 nM Quantum Dots (Thermo Fisher Scientific,<br>Q22063, emission 625 nm) |
| Excitation Source Intensity Variation | 100% to 52% at corners (See Figure 34A) |
| Fluorescence Imaging Signal-to-Noise Ratio | | | | |
|--------------------------------------------|---------------------------|------------------------------------|------------------------------------|--|
| Color | Illumination<br>Intensity | SNR<br>(At min limit of detection) | SNR<br>(At max limit of detection) | |
| Green | Maximum | 2.2 | 8.5 | |
| Green | Minimum | 2.4 | 8.3 | |
| Red | Maximum | 1.2 | 7.8 | |
| Red | Minimum | 1.8 | 6.3 | |
Image /page/5/Figure/3 description: The image contains two graphs that show the relationship between quantum dot concentration and normalized intensity. The graph on the left shows the data for Green-HII and Green-LII, with equations y=0.0075x-0.0261 (R^2=0.994) and y=0.0042x-0.0194 (R^2=0.9913), respectively. The graph on the right shows the data for Red-HII and Red-LII, with equations y=0.180x-0.073 (R^2=0.993) and y=0.141x-0.075 (R^2=0.992), respectively. The quantum dot concentration is measured in nM.
Figure 25: Linearity data for green and red fluorophores in areas of high and low illumination intensity. Signal normalized to the maximum output of the camera sensor. HII: high illumination intensity; LII: low illumination intensity; black dash lines: HII linear regression; black long-dash-dot lines: LII linear regression.
### SUMMARY OF CLINICAL INFORMATION
{6}------------------------------------------------
Photographs of standard white light and fluorescent images from different wound types, sizes, and locations were taken with the device in human patients and reviewed to support the safety and effectiveness of MolecuLight i:X.
The standard white light and fluorescent images included diabetic foot ulcers, venous leg ulcers, pressure ulcers, surgical site wounds, and burns. Wound sizes ranged from small to large and the locations included were on the trunk, upper extremities, and lower extremities. The images were evaluated for quality and consistency.
### Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric population. The device is not labeled for use in the pediatric population.
## LABELING
Labeling has been included which consists of a user manual and quick start guide. The user manual and instructions for use include a description of the device, technical parameters, and principles of operation. These documents summarize the main steps for using the device as well as the necessary measures to clean the reusable components of the device.
### RISKS TO HEALTH
The device is a non-contact autofluorescence imaging tool and uses visible excitation source. It does not use any exogenous contrast agent and does not provide any interpretation for the fluorescence signal. The risks associated with MolecuLight i:X include electrical/mechanical/thermal, EMC and optical safety of the device, and the error in fluorescence detection from the wound; these risks can be mitigated with general controls.
#### BENEFIT/RISK DETERMINATION
The probable benefits/risks of the device are based on nonclinical laboratory studies as well as photographs of wound images. The risk includes the error in fluorescence detection from the wound. However, this is a tool type device for capturing and recording white light and fluorescent images of a wound performing its function without any clinical claims. A benefit is seen with capturing and recording wound images to follow changes in wounds over time. No direct clinical benefit has been demonstrated from recording fluorescence, but a tool that records fluorescence images may provide additional information to a clinician. As only photographs were evaluated to see that the tool performs a function, there is no magnitude, probability, duration, or patient perspective of benefit. The probable benefits outweigh the probable risks.
#### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
### Benefit/Risk Conclusion
{7}------------------------------------------------
In conclusion, given the available information above, the data support that for the previously stated indications for use, the probable benefits outweigh the probable risks for the MolecuLight i:X. The device provides a benefit and the risks can be mitigated using general controls.
## CONCLUSION
The De Novo request for the MolecuLight i:X is granted and the device is classified as follows:
Product Code: QCR Device Type: Wound autofluorescence imaging device Class: I (Exempt from premarket notification, subject to the limitations in 878.9) Regulation Number: 21 CFR 878.4165
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.