K Number
K242669

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-03-26

(202 days)

Product Code
Regulation Number
878.4550
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SnapshotGLO is a handheld imaging tool that allows clinicians diagnosing and treating skin wounds, at the point of care, to (i) View and digitally record images of a wound, (ii) Measure and digitally record the size of a wound, and (iii) View and digitally record images of fluorescence emitted from a wound when exposed to an excitation light. The fluorescence image, when used in combination with clinical signs and symptoms, has been shown to increase the likelihood that clinicians can identify wounds containing bacterial loads >104 CFU per gram as compared to examination of clinical signs and symptoms alone. The SnapshotGLO device should not be used to rule-out the presence of bacteria in a wound. The SnapshotGLO does not diagnose or treat skin wounds.

Device Description

SnapshotGLO is a medical device that operates like a camera. It is a point-of-care, wound imaging device. This device is a non-contact imaging device wherein the wound images are captured from a height of ~12 cm using 395 nm LEDs and a white LED to produce a resultant fluorescence image that aids in visualising the bacteria on the wound and a colour-based "RGB" image or clinical image. Resultant images are viewed on the 7-inch touchscreen display. SnapshotGLO is based on autofluorescence imaging technology and uses native fluorescence of bacteria to determine the presence of bacterial bioburden and displays a two-dimensional, colour-coded highlight of bioburden presence on the wounds. SnapshotGLO is a handheld imaging tool that allows clinicians diagnosing, monitoring and treating skin wounds at the point of care with the help of the following features: . View and digitally record images of a wound, . Measure and digitally record the size of a wound, and View and digitally record images of fluorescence emitted from a wound when exposed to an ● excitation light SnapshotGLO consists of: - SnapshotGLO device - Medical grade adapter - User Manual - Quick Start Guide ● SnapshotGLO is intended for wound care applications as an adjunct tool that uses autofluorescence to detect tissues or structures. The fluorescence image, when used in combination with clinical signs and symptoms, has been shown to increase the likelihood that clinicians can identify wounds containing high bacteria bioburden compared to clinical symptoms alone. SnapshotGLO should not be used to rule-out the presence of bacteria in a wound. This device is not intended to provide a diagnosis.

AI/ML Overview

The provided text is a 510(k) summary for the SnapshotGLO (KB100) device, aiming to demonstrate its substantial equivalence to the MolecuLightDX as a predicate device. While it details the device's function and provides some study information, it does not explicitly state "acceptance criteria" in a tabulated format alongside "reported device performance." Instead, it discusses the outcomes of non-clinical and clinical studies that support the device's equivalence and performance.

Based on the information provided, here's an attempt to structure the response according to your request, extracting the closest equivalents to "acceptance criteria" and "reported performance" from the study descriptions.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly define quantitative acceptance criteria for the clinical study. However, the non-clinical tests imply an acceptance criterion of "comparable performance" or "substantially equivalent performance" to the predicate device. For the clinical study, the acceptance was based on showing "improved accuracy" compared to the predicate device when used with clinical signs and symptoms.

Criterion TypeAcceptance Criterion (Implicit/Derived)Reported Device Performance (SnapshotGLO)
Non-clinical - Bacterial Fluorescence DetectionSubstantially equivalent detection of red fluorescence from porphyrin-producing bacteria (mono- and bi-microbial, biofilms) compared to predicate device.Provided robust evidence that SnapshotGLO is substantially equivalent to MolecuLightDX in detecting bacterial fluorescence. Confirmed effectiveness for identifying porphyrin-producing bacteria and biofilms.
Non-clinical - Wound Dimensions MeasurementComparable performance in manual wound detection modes to the predicate device, demonstrating agreement in measurement accuracy and repeatability.Performs comparably to MolecuLightDX in manual wound detection modes, with strong agreement in measurement accuracy and repeatability. Supports claim of substantial equivalence.
Clinical - Bacterial Load IdentificationWhen used with clinical signs and symptoms (CSS), demonstrated improved accuracy in identifying wounds with bacterial loads >10^4 CFU per gram compared to predicate device alone.When used in conjunction with CSS, showed over 88% positive percent agreement and provided improved accuracy (75-82.5%) compared to MolecuLightDX (52.5-65%) when validated against culture results for identifying wounds with bacterial loads >10^4 CFU per gram.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Clinical Study: 40 patients.
    • Non-clinical Studies: Not explicitly stated, but conducted on "culture plates" and "wound dimensions."
  • Data Provenance: The document does not specify the country of origin for the clinical study. The clinical study was described as retrospective.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Two clinical evaluators were involved in the retrospective clinical study.
  • Qualifications: Not explicitly stated in the document.

4. Adjudication Method for the Test Set

The document states "This blinded assessment" for the clinical study, indicating that the evaluators were blinded to some information, but it does not describe a specific adjudication method (e.g., 2+1, 3+1 consensus).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? A clinical study comparing the SnapshotGLO and the predicate device was done, involving two clinical evaluators. While it's a comparative study with multiple readers, the format described does not fully align with a typical MRMC study designed to assess reader improvement with AI assistance. It rather compares the device's performance (with CSS) against the predicate device's performance (with CSS) validated against culture.
  • Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: The study compared SnapshotGLO + CSS versus MolecuLightDX + CSS versus CSS alone (implicit, as the basis for comparison), all validated against culture results. It demonstrated:
    • SnapshotGLO + CSS accuracy: 75-82.5%
    • MolecuLightDX + CSS accuracy: 52.5-65%
    • The phrase "increase the likelihood that clinicians can identify wounds containing bacterial loads >10^4 CFU per gram as compared to examination of clinical signs and symptoms alone" (from the Indications for Use) suggests that the device, when combined with CSS, improves performance over CSS alone. The specific "effect size" of improvement of human readers with AI vs. without AI assistance (meaning AI as an added tool for human readers) is not directly quantified as a comparative value in terms of reader gain. The comparison shown is between two different devices (both of which are imaging tools that provide additional information to clinicians) when used with CSS, against culture results.

6. Standalone (Algorithm Only) Performance

The document does not report on standalone (algorithm only without human-in-the-loop performance). The indications for use consistently state that the fluorescence image is to be used "in combination with clinical signs and symptoms."

7. Type of Ground Truth Used

  • Clinical Study: The ground truth for identifying wounds with bacterial loads >10^4 CFU per gram was established using culture results.
  • Non-clinical Studies: The ground truth for bacterial fluorescence detection was based on bacterial presence in in vitro culture plates. For wound dimensions, it was likely based on known or carefully measured dimensions.

8. Sample Size for the Training Set

The document is a 510(k) summary for a medical device and does not provide information regarding the training set sample size as it primarily focuses on the device's performance for regulatory submission. This device description points to an "autofluorescence imaging technology" for directly visualizing bacterial compounds, rather than a machine learning algorithm that requires a training set. If there is an AI component for image processing or interpretation not explicitly detailed, the training set information is not included in this document.

9. How Ground Truth for Training Set Was Established

As no training set is discussed concerning an AI/ML algorithm, no information is provided on how its ground truth was established. The device utilizes physical principles of autofluorescence.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".

March 26, 2025

Kent Imaging Inc. Hitalo Arume Regulatory Affairs and Quality Assurance Manager Suite 300, 1210 8 Street SW Calgary, AB T2R 1L3 Canada

Re: K242669

Trade/Device Name: SnapshotGLO (KB100) Regulation Number: 21 CFR 878.4550 Regulation Name: Autofluorescence Detection Device For General Surgery And Dermatological Use Regulatory Class: Class II Product Code: QJF, FXN Dated: March 17, 2025 Received: March 17, 2025

Dear Hitalo Arume:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}------------------------------------------------

See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by TANISHA - Digitally signed by Signed by L. HITHE - C Date: 2025.03.26
16:27:24 -04'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242669

Device Name SnapshotGLO (KB100)

Indications for Use (Describe)

The SnapshotGLO is a handheld imaging tool that allows clinicians diagnosing and treating skin wounds, at the point of care, to

(i) View and digitally record images of a wound,

(ii) Measure and digitally record the size of a wound, and

(iii) View and digitally record images of fluorescence emitted from a wound when exposed to an excitation light.

The fluorescence image, when used in combination with clinical signs and symptoms, has been shown to increase the likelihood that clinicians can identify wounds containing bacterial loads >104 CFU per gram as compared to examination of clinical signs and symptoms alone. The SnapshotGLO device should not be used to rule-out the presence of bacteria in a wound.

The SnapshotGLO does not diagnose or treat skin wounds.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for Kent Imaging. The logo consists of a red square with a white stylized "K" inside, followed by the text "KENT" in large, bold, blue letters. Below the word "KENT" is the word "IMAGING" in smaller, gray letters.

510(k) Summary

March 25, 2025

510(k) Number: K242669

Device Name: SnapshotGLO (KB100)

Submittal Information:

Post-approval contact: Hitalo Arume

Kent Imaging Inc. Suite 300, 1210 8 Street SW Calgary, AB CANADA T2R 1L3

Phone: 403-455-7610 Fax: 877-664-5450

Name of the Device

SnapshotGLO

Device Classification and Product Code

Autofluorescence detection device, 21 CFR 878.4550, Class II, QJF

Tape, Camera, Surgical, 21 CFR 878.4160, Class I, FXN

Legally Marketed Predicate Device

Device Name: MolecuLightDX

510(k) Number: K211901

SnapshotGLO (KB100) information

Indications for Use

The SnapshotGLO is a handheld imaging tool that allows clinicians diagnosing and treating skin wounds, at the point of care, to

  • i. View and digitally record images of a wound,
  • ii. Measure and digitally record the size of a wound, and
  • iii. View and digitally record images of fluorescence emitted from a wound when exposed to an excitation light.

The fluorescence image, when used in combination with clinical signs and symptoms, has been shown to increase the likelihood that clinicians can identify wounds containing bacterial loads >104 CFU per gram as compared to examination of clinical signs and symptoms alone. The SnapshotGLO device should not be used to rule-out the presence of bacteria in a wound.

The SnapshotGLO does not diagnose or treat skin wounds.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Kent Imaging. The logo consists of a red square with a white "K" inside, followed by the word "KENT" in blue, block letters. Below the word "KENT" is the word "IMAGING" in smaller, gray letters.

Device Description

SnapshotGLO is a medical device that operates like a camera. It is a point-of-care, wound imaging device. This device is a non-contact imaging device wherein the wound images are captured from a height of ~12 cm using 395 nm LEDs and a white LED to produce a resultant fluorescence image that aids in visualising the bacteria on the wound and a colour-based "RGB" image or clinical image. Resultant images are viewed on the 7-inch touchscreen display.

SnapshotGLO is based on autofluorescence imaging technology and uses native fluorescence of bacteria to determine the presence of bacterial bioburden and displays a two-dimensional, colour-coded highlight of bioburden presence on the wounds.

SnapshotGLO is a handheld imaging tool that allows clinicians diagnosing, monitoring and treating skin wounds at the point of care with the help of the following features:

  • . View and digitally record images of a wound,
  • . Measure and digitally record the size of a wound, and
  • View and digitally record images of fluorescence emitted from a wound when exposed to an ● excitation light

SnapshotGLO consists of:

  • SnapshotGLO device
  • Medical grade adapter
  • User Manual
  • Quick Start Guide ●

SnapshotGLO is intended for wound care applications as an adjunct tool that uses autofluorescence to detect tissues or structures. The fluorescence image, when used in combination with clinical signs and symptoms, has been shown to increase the likelihood that clinicians can identify wounds containing high bacteria bioburden compared to clinical symptoms alone. SnapshotGLO should not be used to rule-out the presence of bacteria in a wound. This device is not intended to provide a diagnosis.

Fundamental Principles

SnapshotGLO operates using autofluorescence imaging technology. When exposed to 395 nm excitation light, naturally occurring bacterial compounds emit fluorescence in the visible spectrum (400-700 nm). This emitted light is captured by the device to generate fluorescence images that highlight areas of potential bacterial presence. When interpreted alongside clinical signs and symptoms, these images can assist clinicians in identifying wounds with high bacterial bioburden. The device also captures white light images and enables wound measurement, with results displayed on an integrated touchscreen for clinical review and documentation.

Non-clinical study

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for Kent Imaging. The logo consists of a red square with a white "K" inside, followed by the text "KENT IMAGING" in blue. The word "KENT" is in a larger, bolder font than the word "IMAGING", which is located directly below it.

Culture Plate Bench Testing:

This nonclinical test aimed to evaluate the efficacy of the SnapshotGLO device in detecting bacterial fluorescence after ultraviolet excitation and compare it with the predicate device, MoleculightDX. The primary endpoints focused on detecting red fluorescence from porphyrin-producing bacteria, both in mono- and bi-microbial settings, as well as in bacterial biofilms. The secondary endpoints examined the time-series analysis of red fluorescence intensity.

The nonclinical test provided robust evidence that the SnapshotGLO device is substantially equivalent to the MoleculightDX in detecting bacterial fluorescence. The results support the device's effectiveness in clinical settings, confirming its potential for identifying porphyrin-producing bacteria and biofilms in wound care and infection management.

Wound dimensions Testing:

This nonclinical testing discussed a comprehensive comparison of the SnapshotGLO device against the predicate device, MoleculightDX, to assess their equivalence in measuring wound dimensions. The purpose of these tests is to determine whether SnapshotGLO performs similarly to MoleculightDX, supporting a claim of substantial equivalence in the 510(k) submission.

This nonclinical test demonstrates that SnapshotGLO performs comparably to the predicate device, MoleculightDX, in manual wound detection modes. The strong agreement in measurement accuracy and repeatability between the two devices supports the claim of substantial equivalence.

Clinical data

A retrospective clinical study involving 40 patients and two clinical evaluators was conducted to compare the performance of SnapshotGLO and MolecuLightDX, alongside clinical signs and symptoms (CSS). This blinded assessment demonstrated that SnapshotGLO, when used in conjunction with CSS, showed over 88% positive percent agreement and provided improved accuracy (75-82.5%) compared to MolecuLightDX (52.5-65%) when validated against culture results. These findings support the device's substantial equivalence and improved clinical performance.

Substantial Equivalence

The SnapshotGLO (KB100) has the same regulatory information as the predicate device MolecuLightDX (K211901).

DescriptionMoleculightDX (Predicate)SnapshotGLO (Subject)Notes
Regulation ClassClass IIClass IISame
RegulationNumberQJF, FXNQJF, FXNSame
DescriptionMoleculightDX (Predicate)SnapshotGLO (Subject)Notes
ProductClassification21 CFR 878.455021 CFR 878.416021 CFR 878.455021 CFR 878.4160Same
ClassificationNameAutofluorescence detectiondevice for general surgery anddermatological useAutofluorescence detectiondevice for general surgery anddermatological useSame
AccessoriesDark drape/adapter/power cableDrape/adapter/power cableSame
Intended useIntended for general surgery anddermatological use as an adjuncttool that uses autofluorescence todetect tissues or structures.This device is not intended toprovide a diagnosis.Intended for general surgery anddermatological use as an adjuncttool that uses autofluorescence todetect tissues or structures. Thisdevice is not intended to providea diagnosis.Same
Indications for useThe MolecuLightDX is ahandheld imaging tool thatallows clinicians diagnosing andtreating skin wounds, at the pointof care, toi)View and digitallyrecord images of awound,ii)Measure and digitallyrecord the size of awound, andiii)View and digitallyrecord images offluorescence emittedfrom a wound whenexposed to an excitationlight.The fluorescence image, whenused in combination with clinicalsigns and symptoms, has beenshown to increase the likelihoodthat clinicians can identifywounds containing bacterialloads >104 CFU per gram ascompared to examination ofclinical signs and symptomsalone. The MolecuLightDXdevice should not be used toThe SnapshotGLO is a handheldimaging tool that allowsclinicians diagnosing andtreating skin wounds, at the pointof care, toi)View and digitallyrecord images of awound,ii)Measure and digitallyrecord the size of awound, andiii)View and digitallyrecord images offluorescence emittedfrom a wound whenexposed to an excitationlight.The fluorescence image, whenused in combination with clinicalsigns and symptoms, has beenshown to increase the likelihoodthat clinicians can identifywounds containing bacterialloads >104 CFU per gram ascompared to examination ofclinical signs and symptomsalone. The SnapshotGLO deviceshould not be used to rule-outSame
DescriptionMoleculightDX (Predicate)SnapshotGLO (Subject)Notes
rule-out the presence of bacteria in a wound.the presence of bacteria in a wound.
The MolecuLightDX does not diagnose or treat skin wounds.The SnapshotGLO does not diagnose or treat skin wounds.
Target organWoundsWoundsSame
Patient populationAdult patientsAdult patientsSame
Operating modesStandard and fluorescence imaging, video and image captureAutofluorescence and white light imagingSimilar:SnapshotGLO operates in white light and autofluorescence image capture mode only.
Excitation light405 nm light emitted from light emitting diodes (LED)s395 nm light emitted from light emitting diodes (LED)sSimilar. Small difference in excitation wavelength.
White LEDMoleculightDX has a white LED which is referred to as a torch.White LED that covers the visible spectrum (~400-700 nm)Similar, both devices use white light when capturing RGB images
Emission wavelength500-545 nm and 600-665 nm400-700 nmSimilar
Contrast agentNot required - autofluorescence-based targetNot required - autofluorescence-based targetSame
Infrared Laser Distance Finder Wavelength940 nm940 nmSame
Infrared Laser Distance Finder Peak Power Output17 mW17 mWSame
Working distance8 - 12 cm (Fluorescence and Standard Imaging)8 - 20 cm (Measurement Mode)11- 13 cm from the region of interest for both fluorescence and wound measurement modesSimilar. The slight difference in the imaging distance does not affect the performance of the device.
Optical head3 camera system1 RGB cameraDifferent.Only 1 camera required for autofluorescence and wound dimension measurement.
DescriptionMoleculightDX (Predicate)SnapshotGLO (Subject)Notes
Resolution (focalplane)8 megapixelsMaximum 12 MP for RGBcameraSimilar.
Maximum FrameRate30 images/sec30 images/secSame
Camera Bit Depth8 bits8 bits for RGB cameraSame
Image size (pixels)3264 x 2448 pixels1200x1200 pixelsSimilar
Image formatJPEGJPEGSame
Software operatingsystem (OS)compatibilityAndroid 9.1A customized OS provided bythe module manufacturer(Quectel) based on the AOSP ofAndroid 10Q.R8.Similar. Both devices use anAndroid-based OS.
MeasurementFunctionalityStickerless WoundMeasurement: Wound length,width, and area measurementsStickerless WoundMeasurement: Length, Width,and Area measurementsSame
Power SupplyBattery and WallBattery and WallSame
Display5.5" AMOLED display7" LCD displaySimilar.
Patient ContactingMaterialsNon-patient contacting deviceNon-patient contacting deviceSame
SterilityDevice non-sterileDevice non-sterileSame
Electrical SafetyCompliance to IEC 60601-1Compliance to IEC 60601-1Same
Mechanical SafetyCompliance to IEC 60601-1Compliance to IEC 60601-1Same
Chemical SafetyNo chemical delivered or used aspart of the systemNo chemical delivered or used aspart of the systemSame
Standards withwhich the DevicecompliesIEC 60601-1IEC 60601-1-2IEC 60601-2-57IEC 62471IEC 60601-1IEC 60601-1-2IEC 60601-2-57IEC 62471IEC 60601-2-22Similar. The laser module inthe ranging sensor is IEC60601-2-22 certified.
ConnectivityCan be connected to wirelessnetworksCan be connected to wirelessnetworksSame

Detailed Comparative Features between Devices.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Kent Imaging. The logo consists of a red square with a white "K" inside, followed by the words "KENT" in blue, and "IMAGING" in gray underneath. The "K" in the square is stylized with a curved line extending from the top left corner to the bottom right corner.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for Kent Imaging. The logo consists of a red square with a white "K" inside of it. To the right of the square is the word "KENT" in large, blue, sans-serif letters. Below the word "KENT" is the word "IMAGING" in smaller, gray, sans-serif letters.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Kent Imaging. The logo consists of a red square with a white stylized "K" inside, followed by the word "KENT" in large, bold, blue letters. Below the word "KENT" is the word "IMAGING" in smaller, black letters.

Conclusion

The SnapshotGLO (KB100) has the same intended use, indications for use, and fundamental technological characteristics as the predicate device, MolecuLightDX (K211901). Minor differences in design, such as

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Kent Imaging. The logo consists of a red square with a white stylized "K" inside, followed by the word "KENT" in large, bold, blue letters. Below the word "KENT" is the word "IMAGING" in smaller, gray letters. The logo is simple and modern, and the colors are eye-catching.

excitation wavelength and imaging hardware, do not raise new questions of safety or effectiveness. Nonclinical and clinical performance data support that SnapshotGLO is substantially equivalent to the predicate device for its intended purpose in wound imaging and bacterial fluorescence detection.

§ 878.4550 Autofluorescence detection device for general surgery and dermatological use.

(a)
Identification. An autofluorescence detection device for general surgery and dermatological use is an adjunct tool that uses autofluorescence to detect tissues or structures. This device is not intended to provide a diagnosis.(b)
Classification. Class II (special controls). The special controls for this device are:(1) In vivo testing under anticipated conditions of use must characterize the ability of the device to detect autofluorescent signals from tissues or structures consistent with the indications for use.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance testing must demonstrate the electromagnetic compatibility and electrical, mechanical, and thermal safety of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Performance testing must demonstrate the sterility of patient-contacting components of the device.
(6) Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(7) Performance testing must demonstrate laser and light safety for eye, tissue, and skin.
(8) Labeling must include the following:
(i) Instructions for use;
(ii) The detection performance characteristics of the device when used as intended; and
(iii) A shelf life for any sterile components.