sensiLift

K170499 · El Global Trade, Ltd. · PAY · Jun 15, 2017 · General, Plastic Surgery

Device Facts

Record IDK170499
Device NamesensiLift
ApplicantEl Global Trade, Ltd.
Product CodePAY · General, Plastic Surgery
Decision DateJun 15, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4420
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The sensiLift is an over-the-counter home use device intended for non-invasive treatment of mild to moderate facial wrinkles for adult women who have Fitzpatrick Skin Types I-IV.

Device Story

Hand-held, home-use device; generates 1 MHz bipolar radiofrequency (RF) energy pulses; heats skin tissue to improve appearance of wrinkles/rhytides. User selects heating level (high/medium/low) and electrode movement speed; device includes RF power module, power supply, and electrodes. Two redundant thermistors monitor skin temperature; automatically deactivates RF if temperature exceeds 40°C. Used by lay users at home. Output is thermal energy; healthcare providers do not see output; user monitors skin appearance. Benefits include non-invasive wrinkle reduction.

Clinical Evidence

Prospective, open-label, single-arm clinical study; 40 female participants (ages 40-64); 35 completed 8 treatments with 3-month follow-up. Primary endpoint: improvement in Fitzpatrick Wrinkle Severity Scale. Results: 89-95% improvement at end of treatment; 97-100% improvement at 3-month follow-up. No adverse events reported.

Technological Characteristics

Hand-held RF device; ABS plastic shell, chrome-coated electrodes. 1 MHz bipolar RF energy source. Max power 9±1W. Two redundant thermistors for temperature monitoring (40°C cutoff). Non-sterile, reusable. Software-controlled via embedded PLC. Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-2.

Indications for Use

Indicated for non-invasive treatment of mild to moderate facial wrinkles in adult women with Fitzpatrick Skin Types I-IV.

Regulatory Classification

Identification

An electrosurgical device for over-the-counter aesthetic use is a device using radiofrequency energy to produce localized heating within tissues for non-invasive aesthetic use.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested: Over-heating, power accuracy radiofrequency, pulse cycle, waveform, pulse duration, and device characterization parameters. (2) Label comprehension and self-selection performance evaluation must demonstrate that the intended over-the-counter users can understand the package labeling and correctly choose the device for the indicated aesthetic use. (3) Usability performance evaluation must demonstrate that the over-the-counter user can correctly use the device, based solely on reading the directions for use, to treat the indicated aesthetic use. (4) Clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use to achieve the intended aesthetic results. (5) The patient-contacting components of the device must be demonstrated to be biocompatible. (6) Instructions for cleaning the device must be validated. (7) Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety, including the mechanical integrity, of the device. (8) Software verification, validation, and hazard analysis must be performed. (9) Labeling must include: (i) Warnings, precautions, and contraindications to ensure the safe use of the device for the over-the-counter users. (ii) A statement that the safety and effectiveness of the device's use for uses other than the indicated aesthetic use are not known. (iii) A summary of the clinical information used to establish effectiveness for each indicated aesthetic usage and observed adverse events.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 15, 2017 El Global Trade Ltd Dr. Yael Liebes-Peer RA/ QA Manager 8 Tzoran St. P.O. Box 8242 Netanya, 4250608 Israel Re: K170499 Trade/Device Name: sensiLift Regulation Number: 21 CFR 878.4420 Regulation Name: Electrosurgical Device for Over-The-Counter Aesthetic Use. Regulatory Class: Class II Product Code: PAY Dated: May 7, 2017 Received: May 12, 2017 Dear Dr. Yael Liebes-Peer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Image /page/1/Picture/9 description: The image shows the text "Jennifer R. Stevenson -S3". The text is in a large, sans-serif font and is black. The background is white. The letters "FDA" are in the background in a light blue color. For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K170499 Device Name sensiLift Indications for Use (Describe) The sensiLift is an over-the-counter home use device intended for non-invasive treatment of mild to moderate facial wrinkles for adult women who have Fitzpatrick Skin Types I-IV. | Type of Use (Select one or both, as applicable) | <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | |-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff [PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov) *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* {3}------------------------------------------------ | EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069 Fax: +972-9-7734831 | | |---------------------------|-------------------------------------------------------------------------------------------------|--| | sensiLift – K170499/ S001 | RD- 11118 B2 | | DATE PREPARED: APRIL 24TH, 2016 ### 510(K) OWNER NAME 1. EL Global Trade Ltd. Tzoran 8th st, P.O.Box 8242, Netanya 425608, Israel. Phone: +972-9-7889069, Fax: +972-9-7734831. Contact person name: Dr. Yael Liebes-Peer, RA/ QA Manager Phone: +972-9-7889069, Fax: +972-9-7734831, E mail: Yael@sensica.com #### 2. DEVICE NAME Common/ Usual Name: OTC device for treatment of facial wrinkles based on RF Proprietary/Trade name: sensiLift Classification: EL Global Trade Ltd.'s sensiLift device has been classified as Class II device under the following classification names: | Classification Name | Product<br>Code | Regulation<br>Number | Panel | |-----------------------------------------------------------------------|-----------------|----------------------|--------------------------------| | Electrosurgical device<br>for over-the-counter<br>(OTC) aesthetic use | PAY | 878.4420 | General and Plastic<br>Surgery | #### 3. PREDICATE DEVICES EL Global Trade Ltd.'s sensiLift device is substantially equivalent to the following Predicate Device: 3.1 EndyMed Medical Ltd.'s NEWA device, cleared under De-Novo number DEN150005 at Dec 18th, 2016. #### 4. DEVICE DESCRIPTION The sensiLift is an OTC, home use hand held device generating pulses of radiofrequency (RF) energy that are emitted into the skin. RF energy heats the {4}------------------------------------------------ | EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069 | | Fax: +972-9-7734831 | |---------------------------|-----------------------------------------------------------------------------|--------------|---------------------| | sensiLift – K170499/ S001 | | RD- 11118 B2 | | tissue to improve the appearance of wrinkles and rhytides. It is a non-invasive, non-ablative device consisting of: - . User Interface - Programmable logic controller (PLC, microcontroller) embedded in PCBA ● - RF power module - Power Supply ● - RF electrodes ● The user interface allows the selection of heating level: high, medium low or off, conjugated to movement speed level of the electrodes (also adjusted by the user): high or low. The PLC (on the PCBA) is a specially configured software that, combined with hardware circuits, provides the operational and safety function of the system. The RF power module provides RF energy to the active tip electrodes, producing a sinusoidal signal of 1MHz frequency (power limited by hardware). This device is supplied non-sterile. ## Technical specifications: - . Maximal power output: 9±1 Watt. - . Skin level power: 3.5±1, 5±1 and 6.5±1 Watt (energy levels 1-3). - . Frequency: 1 MHz. - Maximal temperature allowed: 40°C. ● ### ട. INTENDED USE/ INDICATIONS FOR USE The sensiLift is an over-the-counter home use device intended for non-invasive treatment of mild to moderate facial wrinkles for adult women who have Fitzpatrick Skin Types I-IV. {5}------------------------------------------------ | EL Global Trade Ltd | | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069 Fax: +972-9-7734831 | | |---------------------------|-------------|-------------------------------------------------------------------------------------------------|--| | sensiLift – K170499/ S001 | RD-11118 B2 | | | #### 6. NON-CLINICAL (BENCH) PERFORMANCE DATA The following performance data (bench tests) were provided in support of the performance and safety and efficacy of the sensiLift device as well as the substantial equivalence determination. ## Safety Bench Tests and Verification & Validation (V&V) Summary Bench testing was conducted per IEC standard 60601family to demonstrate that the sensiLift device performs as expected under anticipated conditions of use. This testing included safety performance verification (electrical and mechanical) according to IEC 60601-1, essential requirements according to IEC 60601-1-2 and particular requirements for home use medical device per IEC60601-1-11. In addition, the following bench testing was conducted (among others) to demonstrate the device performance characteristics as part of the V&V: - . Over-heating safety. The two redundant thermistors embedded in the device constantly measure the skin temperature and deactivate RF delivery when temperature exceeds 40℃. This temperature was determined to be acceptable based on published information from the National Institute for Standard and and Technology (http://www.nist.gov/fire/fire_behavior.cfm) which state that human skin begins to feel pain at 44°C and may start to develop skin burns at 48°C. - . Power accuracy. The device was validated for the different power levels on a 200 Ω load, which is appropriate as the reference of the average load of the user. The measured total power was within the error margin, indicating that the device met the acceptance criteria. {6}------------------------------------------------ | EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069 | | Fax: +972-9-7734831 | |---------------------------|-----------------------------------------------------------------------------|--|---------------------| | sensiLift – K170499/ S001 | RD- 11118 B2 | | | - Parameter validation. The device was tested for its varied parameters, . including radio frequency, pulse cycle, wave form and pulse duration. All results were within the acceptance criteria. ### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the sensiLift device. The device complies with: - IEC 60601-1:2005/EN 60601-1:2006, General safety standard: safety . requirements for medical electrical systems. - . IEC 60601-1-2:2007, IEC 60601-1-2:2009, Medical electrical equipment Part 1-2 - General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests. - IEC 60601-1-11:2010. Medical electrical equipment Part 1-11 -. Requirements for the medical electrical equipment and medical electrical systems used in the home healthcare environment. - . IEC 60601-2-2: 2009, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories. ### Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern because inadvertent software errors could result in skin burns to the user. {7}------------------------------------------------ | K170499 | Page 5 of 1 | |---------------------------|----------------------------------------------------------------------------------------------------| | EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069<br>Fax: +972-9-7734831 | | sensiLift – K170499/ S001 | RD- 11118 B2 | ## Human Factors Validation Testing A human Factors Validation Testing was done to evaluate the HFE/ UE concerns, including label comprehension/ self-selection and usability/ user interface studies (one arm, simulated use and summative evaluation). In both studies, participants had received the labeling materials of the device, and given enough time to read it according to their own wish. Labeling Comprehension/ Self Selection Study was aimed at demonstrating potential user comprehension of the device labeling (mainly the outer package) in order to decide if device use is appropriate for her. It had included three main questions regarding the intended use of the device, treatment areas and possibility to usse the device according to the contraindication, with correct answers: ,,Yes" or "No", explenation and further validation by the moderator. A total of 43 participants, 2 men and 41 women, from age 18 to 65 years old (45 ± 13 years old) were enrolled to this study. Participants had different educational levels, as obtained by REALM testing with 20% below 9th grade level. All participants had met the inclusion/ exclusion criteria and signed ICF. 100% of the participants had reported correct answers, as further validated by the moderator («=1). Usability/ User Interface Study was aimed at demonstrating the OTC user can correctly use the device, based on the labeling material for the intended aesthetic puspose. The study was focused on four use scenarios including 22 relevant tasks: "preparing and knowing the device" (7 tasks), "self-preparation" (4 tasks), "device usage" (7 tasks) and "after usage" (3 tasks). None of the participants had required assistance from the moderator while performing the tasks. Altogether, all tasks were completed by 100% of the participants, and the success rate (pass criteria) was 100% per each task scenario. In addition, the vast majority of the participants (14 out of 20) had additionally commented that the device usage is easy and clear and they are curios and would {8}------------------------------------------------ | EL Global Trade Ltd | | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069     Fax: +972-9-7734831 | | |---------------------------|--|-----------------------------------------------------------------------------------------------------|--| | sensiLift – K170499/ S001 | | RD- 11118 B2 | | be interested in using the device. Some participants (7 out of 20) had additionally commented (though succeeded in operation) regarding the comfort of using the buttons. Only few (4 out of 20 participants) were concerned about other elements of the device. Furthermore, when participants were asked regarding their user experience and satisfaction after the usability and user interface study, high satisfaction was obtained, with an average result was 1.66 ± 0.25, using a 1-5 discrete scale (when 1 represent best and 5 worst). These two demonstrates that the intended OTC users can understand the package labeling, correctly choose the device and use it for the indicated aesthetic use, based solely on reading the lableling materials. #### 7. PERFORMANCE TESTING - ANIMAL No animal testing was performed with the subject device #### CLEANING, STERILIZATION, SHELF LIFE AND BIOCOMPATIBILITY 8. The sensiLift is a non-sterile, reusable device, intended for a single user. The device cleaning instructions are based on the cleaning instructions of the predicate device due to the fact that both devices are made from the same materials and used similarly. The shelf life expectancy of the device is 5 years, similarly to the predicate device. The biocompatibility evaluation for the sensiLift device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 {9}------------------------------------------------ | EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069 Fax: +972-9-7734831 | |---------------------------|-------------------------------------------------------------------------------------------------| | sensiLift – K170499/ S001 | RD- 11118 B2 | "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. EL Global has categorized its sensiLift device as: "Surface device, Skin Contact for limited contact duration", with contact duration of less than 24h ("A", up to 24 hours). Therefore, the battery of testing included the following tests: cytotoxicity, sensitization and irritation tests. The body contact materials are biocompatible per: - ISO 10993-5:2009, Biological evaluation of medical devices Part 5: . Tests for in vitro cytotoxicity. - ISO 10993-10:2010, Biological evaluation of medical devices Part 10: . Tests for irritation and skin sensitization. The sensiLift device is made of similar materials that come into contact with user's body as already cleared for the predicate device (DEN150005) and for the auxiliary sensiLight mini device (K140527 and K161089). #### 9. CLINICAL PERFORMANCE DATA The sensiLift safety and efficacy, when used at a home setting, for non-invasive treatment of mild to moderate facial wrinkles of adult women with Fitzpatrick Skin Types I-IV was examined in an open label, prospective, interventional single arm study clinical study. 40 females between the ages 40-64, fulfilling the inclusion/ exclusion criteria, were enrolled to the study by the Principle Investigator (PI) and treated 2-3 treatment regions (perioral, periorbital and chin regions), according to their preference and the physician approval. 35 participants (total of 87 treatment areas) had completed 8 treatments according to the instructions for use and appeared to the 3 months after treatments last follow-up (FU2). The wrinkles appearance was assessed using the Fitzpatrick Wrinkles Severity Scale and improvement was defined as decrease of at least one score in the Scale. This assessment was conducted by 3 independent physicians (beside the PI, 3 board certified {10}------------------------------------------------ | EL Global Trade Ltd<br>Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069 Fax: +972-9-7734831 | | |------------------------------------------------------------------------------------------------------------------------|--------------| | sensiLift – K170499/ S001 | RD- 11118 B2 | dermatologists or plastic surgeons) at the end of the active treatment regimen (T8), 1 and 3 months after the last treatment (follow-up regimen FU1 and FU2, respectively). No adverse events occurred during and after the study, meaning that the success criteria for primary safety endpoint have been met. As for effectiveness, after the completion of the active treatment regimen, 89%, 95% and 95% of the study participants (as assessed by 3 reviewers, respectively) have shown improvement in overall facial wrinkles appearance. Similarly, improvement was observed at the end of the Follow-up regimen, 3 months after the last treatment, for 97%, 100% and 100% of the participant by the three evaluators. In addition, all the participants were highly satisfied with the device at the end of the study. 89% reported on improvement in their skin's appearance. These results demonstrate that the device performs as intended under anticipated conditions of use (home environment) to achieve the intended aesthetic results. #### 10. SUBSTANTIAL EQUIVALENCE The indications for use and technological characteristics of the sensiLift device are substantially equivalent to the indications for use and technological characteristics of the predicate device. A detailed comparison between the sensiLift by EL Global LTD. proposed device and the predicate device is presented at Table 1 below, including the similarities and differences and the related conclusion. {11}------------------------------------------------ Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel Tel: +972-9-7889069 Fax: +972-9-7734831 sensiLift – K170499/ S001 RD- 11118 B2 ### Table 1. Substantial Equivalence of EL Global's sensiLift with Predicate Device. | Feature | EL Global<br>sensiLift<br><i>New Device</i> | EndyMed<br>NewaTM<br>[DEN150005]<br><i>Predicate Device</i> | Characteristics<br>comparison (similarity) | |-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------| | Regulatory information | | | | | Device Class | Class II | Class II | Identical | | Classification<br>Panel | General and Plastic<br>Surgery | General and Plastic<br>Surgery | Identical | | Product code | PAY | PAY | Identical | | Regulation<br>number | 21 CFR 878.4420 | 21 CFR 878.4420 | Identical | | Device | Over-The-Counter<br>Radiofrequency<br>Coagulation Device For<br>Wrinkle Reduction | Over-The-Counter<br>Radiofrequency<br>Coagulation Device For<br>Wrinkle Reduction | Identical | | Regulation<br>description | Electrosurgical device for<br>over-the-counter aesthetic<br>use | Electrosurgical device for<br>over-the-counter aesthetic<br>use | Identical | | Intended use | The sensiLiftTM<br>is an over-the-counter<br>home use device intended<br>for non-invasive treatment<br>of mild to moderate facial<br>wrinkles for adult women<br>who have Fitzpatrick Skin<br>Types I-IV. | The EndyMed NewaTM is<br>an over-the-counter home<br>use device intended for<br>non-invasive treatment of<br>mild to moderate facial<br>wrinkles for adult women<br>who have Fitzpatrick Skin<br>Types I-IV. | Identical | | Device characteristics | | | | | Mode of<br>Operation | The device a home-use<br>hand-held device<br>generating pulses RF<br>energy that emitted into the<br>skin. RF energy heats the<br>tissue to improve the<br>appearance of wrinkles and<br>rhytides. It is non-invasive<br>and non-ablating device. | The device is a home-use<br>hand-held device<br>generating pulses of RF<br>energy that emitted into the<br>skin. RF energy heats the<br>tissue to improve the<br>appearance of wrinkles and<br>rhytides. It is non-invasive<br>and non-ablating device. | Identical | | Energy<br>Source | 1 MHz RF (Bi-polar) | 1 MHz RF (Bi-polar) | Identical | | Number of<br>electrodes | 2 electrodes (1 pair of Bi-<br>polar electrodes) | 6 electrodes (3 pairs of Bi-<br>polar electrodes) | Similar to Predicate<br>Device. See note No. 1 | | | EL Global<br><i>sensiLift</i> | EndyMed<br><i>Newa</i> TM<br>[DEN150005] | Characteristics<br>comparison (similarity) | | Feature | <i>New Device</i> | <i>Predicate Device</i> | | | Treatment<br>areas | Face: lower cheek<br>(perioral), upper cheek<br>(periorbital) and under the<br>chin. | Face: lower cheek, upper<br>cheek and under the chin. | Identical | | Treatment<br>regimen | 5 minute each area, 1 time<br>per week, for 8 consecutive<br>weeks (treatment area<br>about 4X4 cm2). | 4 minute each area, 5 times<br>per week, for 4 consecutive<br>weeks (treatment area<br>about 4X4 cm2). | Different. Treatment<br>regimen was validated via<br>clinical Trial.<br>See note No. 2. | | Intended<br>population | Adult healthy women<br>(Fitzpatrick skin type I-IV)<br>who desire to reduce facial<br>wrinkles. | Adult healthy women<br>(Fitzpatrick skin type I-IV)<br>who desire to reduce facial<br>wrinkles. | Identical | | Use<br>Environment | Home Use, self-operation<br>by an unprofessional user. | Home Use, self-operation<br>by an unprofessional user. | Identical | | Device Features | | | | | Hand piece<br>Dimensions,<br>(LXHXW)<br>[mm³] | 173 X 56.8 X 53 | 73 X 37 X 120 | Similar.<br>See note No. 3. | | Weight [gr] | 122 | 70 | Similar.<br>See note No. 3. | | Electrodes<br>treatment<br>area [mm²] | 176 | 178 | Identical | | Body contact<br>materials | ABS plastic shell.<br>Chrome coated electrodes. | ABS plastic shell.<br>Chrome coated electrodes. | Identical | | User<br>interface | 1 LED: device status<br>indicator (yellow-<br>green)1 LED panel (energy<br>level, green)1 LED panel (speed<br>level, green) | 1 LED, blue or green,<br>reflecting the device<br>statusKnob for energy level<br>selection | Similar.<br>See note No. 4. | | Maximal<br>power by<br>hardware<br>[Watt] | $9 \pm 1$ | $10 \pm 2$ | Identical | | User | 3 energy levels deliver skin | 2 energy levels, deliver | Similar. See note No. 1. | | | EL Global | EndyMed | | | Feature | sensiLift | Newa™ | Characteristics | | | | [DEN150005] | comparison (similarity) | | | New Device | Predicate Device | | | energy levels<br>(time<br>modulation,<br>pulse) [%] | Level 3 - 72% Level 2 - 55% Level 1 - 38% | Level 2 - 60% Level 1 - 40% | | | Safety<br>mechanism | Two redundant thermistors continuously monitoring skin temperature and automatically de-activate RF energy delivery when temperature of 40°C/ 104°F is detected, until temperature lowers enough. Moving electrodes to prevent local over-heating. A flickering or steady green LED light when the device is active. RF is delivered only when both electrodes are with body contact. | One Thermal sensor for skin temperature measurements. Automatically stop delivering RF when temperature of 42°C is detected, until temperature lowers enough. Built in moving sensor stop delivering RF when the device is stationary A steady LED when the device is active Vibratory signal alerts when treatment period completed. | Similar.<br>Usefulness of all safety mechanisms was validated in clinical trial and usability study.<br>See note No. 4. | | | Performance Characteristics | | | | Electrical<br>requirements | 100-240 V,<br>50-60 Hz<br>9 V, 1.5 A | 100-240 V<br>50-60 Hz<br>9 V, 2 A | Identical | | Energy<br>consumption | Up to 10W | Up to 10W | Identical | | Frequency<br>[MHz] | 1 | 1 | Identical | | Safety standards | | | | | Safety/EMC | IEC 60601-1:2005/EN<br>60601-1:2006, General<br>safety standard: safety | IEC 60601-1:2005/EN<br>60601-1:2006, General<br>safety standard: safety | Identical.<br>sensiLift has the same<br>safety/ EMC certifications | | Feature | EL Global<br>sensiLift<br>New Device | EndyMed<br>NewaTM<br>[DEN150005]<br>Predicate Device | Characteristics<br>comparison (similarity) | | | requirements for medical<br>electrical systems. | requirements for medical<br>electrical systems. | as the predicate device and<br>is certified for additional<br>standard. | | | IEC 60601-1-2:2007, IEC<br>60601-1-2:2009, Medical<br>electrical equipment Part<br>1-2 - General requirements<br>for basic safety and<br>essential performance -<br>Collateral Standard:<br>Electromagnetic<br>Compatibility -<br>Requirements And Tests. | IEC 60601-1-2:2007,<br>Medical electrical<br>equipment Part 1-2 -<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>Standard: Electromagnetic<br>Compatibility -<br>Requirements And Tests. | | | | IEC 60601-1-11:2010,<br>Medical electrical<br>equipment Part 1-11 -<br>Requirements for the<br>medical electrical<br>equipment and medical<br>electrical systems used in<br>the home healthcare<br>environment. | IEC 60601-1-11:2010,<br>Medical electrical<br>equipment Part 1-11 -<br>Requirements for the<br>medical electrical<br>equipment and medical<br>electrical systems used in<br>the home healthcare<br>environment. | | | | IEC 60601-2-2: 2009,<br>Medical electrical<br>equipment - Part 2-2:<br>Particular requirements for<br>the basic safety and<br>essential performance of<br>high frequency surgical<br>equipment and high<br>frequency surgical<br>accessories. | | | | Biocompatibility | All parts that are in contact<br>with user comply with the<br>requirements for ISO-<br>10993-1: 2009, Biological | All parts that are in contact<br>with user comply with the<br>requirements for ISO-<br>10993-1: 2009, Biological | Identical<br>See note No. 5 | | | | | | | | EL Global | EndyMed | | | | <i>sensiLift</i> | <i>Newa™</i> | Characteristics | | Feature | <i>New Device</i> | [DEN150005]<br><i>Predicate Device</i> | comparison (similarity) | | | devices - Part 1:<br>Evaluation and testing<br>within a risk management<br>process. | devices - Part 1:<br>Evaluation and testing<br>within a risk management<br>process. | | | | ISO 10993-5:2009,<br>Biological evaluation of<br>medical devices - Part 5:<br>Tests for in vitro<br>cytotoxicity. | ISO 10993-5:2009,<br>Biological evaluation of<br>medical devices - Part 5:<br>Tests for in vitro<br>cytotoxicity. | | | | ISO 10993-10:2010,<br>Biological evaluation of<br>medical devices - Part 10:<br>Tests for irritation and skin<br>sensitization. | ISO 10993-10:2010,<br>Biological evaluation of<br>medical devices - Part 10:<br>Tests for irritation and skin<br>sensitization. | | | Software | The software was verified<br>and validated according to<br>the FDA guidance.<br>Moderate Level of<br>Concern | The software was verified<br>and validated according to<br>the FDA guidance.<br>Moderate Level of<br>Concern | Identical | {12}------------------------------------------------ Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel Tel: +972-9-7889069 Fax: +972-9-7734831 # sensiLift – K170499/ S001 RD- 11118 B2 {13}------------------------------------------------ Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel Tel: +972-9-7889069 Fax: +972-9-7734831 ## sensiLift – K170499/ S001 RD- 11118 B2 {14}------------------------------------------------ Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel Tel: +972-9-7889069 Fax: +972-9-7734831 # sensiLift – K170499/ S001 RD- 11118 B2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . {15}------------------------------------------------ Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel Fax: +972-9-7734831 Tel: +972-9-7889069 sensiLift — K170499/ S001 RD- 11118 B2 ### Note No. 1. The energy delivered and electrodes architecture. The method of RF delivery is the crucial factor, since it will dictate the penetration depth of the RF energy. Both sensiLift and predicate device are 1 MHz Bi-polar RF devices. The power energy output is an additional important factor for safe and effective treatment since it dictates the time required to achieve the desired heating. Both devices have similar maximal power consumption of up to 10 W and the effective delivered energy levels are set on time pulse basis (time modulation), with similar values for both devices. The configuration of the electrodes affects the penetration depth of the RF energy as well as the treatment area and the current density. The differences in both devices configuration does not affect the safety and efficacy of the devices, as s {16}------------------------------------------------ | EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069 Fax: +972-9-7734831 | | | |---------------------------|-------------------------------------------------------------------------------------------------|--------------|--| | sensiLift – K170499/ S001 | | RD- 11118 B2 | | was further demonstrated using varies EMC tests, in-vitro comparisons (Performance Testing - Bench) and a clinical trial (Performance Testing -Clinical). ### Note No. 2. The treatment Regimen. The treatment regimen of the sensiLift was especially planned to achieve its intended use in a safe and effective manner. Both sensiLift predicate devices are designed for the exact same anatomical site. Since the total exposure of the sensiLift device lower than of the predicate device, no safety issues are raised. Both the safety and efficacy of the sensiLift device for home-use environment for the intended use was validated in a clinical trial (Performance Testing - Clinical). ### Note No. 3. Device size and weight. The size and weight of the hand-held device are basic unit characteristics with no effect over the intended use nor the safety nor the efficacy of the device. The sensiLift device was designed to allow ergonomic and fatigue-free grip during several minutes of treatments by the naïve user. ### Note No. 4. User interface and safety mechanism. Both user interface elements and the safety mechanisms the sensiLift device were designed to meet the needs of the naïve user at home and to protect its user from any hazard (including several safety mechanisms). The added feature of the sensiLift electrodes to move along their axis does not alter the intended use nor affect neither the safety profile nor introducing new or additional risk factors. This movement ability of the electrodes is similar to devices classified as class 1 devices, exempt of 510K (Massager, Therapeutic, Electric; Product Code ISA; Regulation Number 890.5660). The suitability of sensiLift device for home-use environment and ability of the naïve lay user to use the device in a safe and effective manner was verified according to safety standards and further validated {17}------------------------------------------------ | EL Global Trade Ltd | Tzoran 8th St, P.O.Box 8242, Netanya 4250608, Israel<br>Tel: +972-9-7889069     Fax: +972-9-7734831 | | | |---------------------------|-----------------------------------------------------------------------------------------------------|--------------|--| | sensiLift – K170499/ S001 | | RD- 11118 B2 | | in a Human Factors Test Validation (including self-screening) study and further validated in a clinical trial data (Performance Testing - Bench and Performance Testing - Clinical, respectively). ### Note No. 5. Biocompatibility. The sensiLift device is made of similar materials that come into contact with user's body as already cleared for the predicate device (DEN150005) and for the auxiliary device (K140527 and K161089). #### 11. CONCLUSIONS Due to these identical clauses and high similarities, the sensiLift device is at least as safe and as effective as the predicate device, since they share the same intended use, technological characteristics, features, specifications and materials and anatomical site for use. The differences between the sensiLift device and the predicate device do not raise new safety or effectiveness issues or questions, as detailed above. Based on the performance testing and comparison to predicate devices, the sensiLift device is substantially equivalent to the previously cleared NEWA predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...