K Number
K161089

Validate with FDA (Live)

Device Name
sensiLight Mini
Date Cleared
2016-07-08

(81 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The sensiLight Mini is an over the counter device intended for the removal of unwanted hair. The sensiLight Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of treatment regimen.

Device Description

The sensiLight Mini device is a pulsed light hair removal device. Light-based hair removal is based on the theory of selective photothermolysis in which optical energy is used to disable hair growth. The sensiLight Mini device is composed of a hand held applicator and an external power supply. The spot size (treatment area) in the sensiLight Mini device is 3 cm2.

The device contains a lamp, a skin proximity sensor and a skin pigmentation sensor to detect appropriate skin tones. If the sensiLight Mini is not properly applied (in full contact with the skin) or user skin tone is too dark/tanned, the sensiLight Mini will not trigger a pulse.

AI/ML Overview

This document is a 510(k) premarket notification for the sensiLight Mini, which is a pulsed light hair removal device. The device described in this submission (K161089) is stated to be identical to a previously cleared device (K140527). Therefore, no new performance data, non-clinical data, or clinical data are reported in this specific submission. The substantiation relies on the data from the predicate devices.

However, based on the information provided, we can infer some details about the clinical performance for the original device that K161089 references (K140527), particularly regarding its "permanent reduction in hair regrowth" claim.

Here's an attempt to answer your questions based on the available information and reasonable inferences:

1. A table of acceptance criteria and the reported device performance

Since this submission explicitly states "No new clinical performance data is reported in this submission" and "The sensiLight Mini device is identical in all details to the sensiLight Mini device cleared under K140527," the acceptance criteria and performance would be associated with the original K140527 submission. The document defines "permanent reduction in hair regrowth" as:

Acceptance Criteria (Implied for K140527)Reported Device Performance (Implied for K140527)
Long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of treatment regimen.(Details not provided in this specific K161089 document, but the device was cleared based on meeting this definition.)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the K161089 document as it refers back to a previous submission (K140527). To obtain these details, one would need to consult the K140527 submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the K161089 document. Clinical studies for hair removal devices often involve trained technicians or physicians for evaluation, but specifics are missing here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the K161089 document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Hair removal devices like the sensiLight Mini are typically consumer-use devices and do not involve "readers" in the context of diagnostic imaging. Therefore, an MRMC study and AI assistance as described are not applicable to this type of device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This device is a physical light-based hair removal device for consumer use, not an algorithm. Therefore, "standalone (algorithm only)" performance is not applicable. The device's performance is inherently tied to human use and application according to instructions.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For hair removal studies of this nature, the "ground truth" for efficacy (hair reduction) is typically established by outcomes data in the form of objective measurements of hair count or hair density in treated vs. untreated areas, often performed by trained personnel or experts following a standardized protocol. This would be observed at specified time points (6, 9, and 12 months as indicated by the definition).

8. The sample size for the training set

The device described is a physical medical device, not an AI/ML algorithm that requires a "training set" in the conventional sense of machine learning. Therefore, a "training set" as it applies to AI/ML is not applicable. Its development would involve engineering, safety, and clinical testing, not algorithmic training.

9. How the ground truth for the training set was established

As explained in point 8, the concept of a "training set" with ground truth establishment for an AI/ML algorithm is not applicable to this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three curved lines.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 8, 2016

El Global Trade Ltd Yael Liebes-Peer RA/ QA Manager 8 Tzoran St, P.O. Box 8242 Netanya, 4250604 IL

Re: K161089

Trade/Device Name: Sensilight Mini Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: June 1, 2016 Received: June 6, 2016

Dear Yael Liebes-Peer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K161089

Device Name sensiLight Mini

Indications for Use (Describe)

The sensiLight Mini is an over the counter device intended for the removal of unwanted hair. The sensiLight Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of treatment regimen.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

PSC Publishing Services (301) 443-6740

EF

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

sensiLight Mini – K161089 amendment

RD-17063 A0

510(K) SUMMARY FOR EL GLOBAL TRADE LTD.'S SENSILIGHT MINI

DATE PREPARED: MAY 19™, 2016

510(K) OWNER NAME 1.

EL Global Trade Ltd.

Tzoran 8th st, P.O.Box 8242, Netanya 425068, Israel.

Phone: +972-9-7889069, Fax: +972-9-7734831.

Contact person name: Dr. Yael Liebes-Peer, RA/ QA Manager

Phone: +972-9-7889069, Fax: +972-9-7734831, E mail: Yael@sensica.com

2. DEVICE NAME

Common/Usual Name: Light based hair removal deices

Proprietarv/Trade name: sensiLight Mini

Classification: EL Global Trade Ltd.'s sensiLight Mini device has been classified as Class II device under the following classification names:

Classification NameProductCodeRegulationNumberPanel
Light Based Over-The-Counter Hair RemovalOHT878.4810General and PlasticSurgery

3. PREDICATE DEVICES

EL Global Trade Ltd.'s sensiLight Mini device is substantially equivalent to the following Predicate Devices:

  • 3.1 EL Global Trade Ltd.'s sensiLight Mini device,
    cleared under 510(k) number K140527 at June 19th, 2014.

  • 3.2 Home Skinovations Ltd.'s Silk'n Glide device,
    cleared under 510(k) number K141242 at May 8th, 2014.

{4}------------------------------------------------

EL Global Trade LtdTzoran 8th St, P.O.Box 8242, Netanya 4250608, IsraelTel: +972-9-7889069 Fax: +972-9-7734831
sensilight Mini – K161089 amendmentRD-17063 A0

4. DEVICE DESCRIPTION

The sensiLight Mini device is a pulsed light hair removal device. Light-based hair removal is based on the theory of selective photothermolysis in which optical energy is used to disable hair growth. The sensiLight Mini device is composed of a hand held applicator and an external power supply. The spot size (treatment area) in the sensiLight Mini device is 3 cm2.

The device contains a lamp, a skin proximity sensor and a skin pigmentation sensor to detect appropriate skin tones. If the sensiLight Mini is not properly applied (in full contact with the skin) or user skin tone is too dark/tanned, the sensiLight Mini will not trigger a pulse.

INTENDED USE/ INDICATIONS FOR USE 5.

The sensiLight Mini is an over the counter device intended for the removal of unwanted hair. The sensiLight Mini is also intended for permanent reduction in hair regrowth, defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of treatment regimen.

6. PERFORMANCE DATA

The sensiLight Mini device is identical in all details to the sensiLight Mini device cleared under K140527. No new performance data is reported in this submission.

7. NON-CLINICAL (BENCH) PERFORMANCE DATA

The sensiLight Mini device is identical in all details to the sensiLight Mini device cleared under K140527. No new non-clinical performance data is reported in this submission.

{5}------------------------------------------------

EL Global Trade LtdTzoran 8th St, P.O.Box 8242, Netanya 4250608, IsraelTel: +972-9-7889069Fax: +972-9-7734831
sensilight Mini – K161089 amendmentRD-17063 A0

CLINICAL PERFORMANCE DATA 8.

No new clinical performance data is reported in this submission.

9. SUBSTANTIAL EQUIVALENCE

EL Global Trade Ltd.'s sensiLight Mini is identical to the sensiLight Mini device previously cleared under K140527. The indications for use and technological characteristics of the sensiLight Mini device are substantially equivalent to the indications for use and technological characteristics of the previously cleared sensiLight Mini device (K140527) and Glide device (K141242).

Consequently, it can be concluded that the sensiLight Mini device is substantially equivalent to the Predicate sensiLight Mini and Glide devices, cleared under 510(k) K140527 and K141242, and therefore, may be legally marketed in the USA.

10. CONCLUSIONS

Based on the performance testing and comparison to predicate devices, the sensiLight Mini device is substantially equivalent to the previously cleared sensiLight Mini and Glide predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.