K180623 · Ulthera, Inc. · OHV · May 4, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180623
Device Name
Ulthera System
Applicant
Ulthera, Inc.
Product Code
OHV · General, Plastic Surgery
Decision Date
May 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ulthera® System is indicated for use as a non-invasive dermatological aesthetic treatment to: lift the eyebrow lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions improve lines and wrinkles of the décolleté The Ulthera® System in conjunction with the Ulthera® DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The inaging is to visualize the dermal and subdermal layers of tissue to: ensure proper coupling of the transducer to the skin confirm appropriate depth of treatment such as to avoid bone
Device Story
Ulthera System uses High Intensity Focused Ultrasound (HIFU) for non-invasive dermatological aesthetic treatment; directs micro-focused acoustic waves to create controlled thermal coagulation points beneath epidermis; utilizes DeepSEE transducer for real-time ultrasonic visualization of dermal/subdermal layers up to 8 mm; used in clinical/doctor's office settings by trained operators; imaging output allows operators to ensure proper transducer coupling and confirm treatment depth to avoid bone; thermal energy (< 2 J) lifts tissue and reduces lines/wrinkles; benefits patients via non-invasive skin tightening and rejuvenation without requiring surface cooling.
Clinical Evidence
Open-label clinical trial evaluated submental and neck tissue lifting. 70 patients enrolled; 51/70 (72.9%) achieved ≥20 mm² improvement in lift; 84.3% of these identified as improved by masked evaluators. Adverse events were mild, short-lived, and resolved without incident. No serious adverse events or unanticipated adverse device effects reported.
Technological Characteristics
High Intensity Focused Ultrasound (HIFU) system; thermal energy < 2 J; non-sterile; reusable control unit/handpiece; single-use transducers; biocompatible patient-contact materials; software-controlled; compliant with IEC 60601-1-3.1 ed. and relevant EMC standards.
Indications for Use
Indicated for non-invasive dermatological aesthetic treatment to lift eyebrows, lift lax submental and neck tissue, and improve lines/wrinkles of the décolleté. Includes ultrasonic visualization of dermal/subdermal layers up to 8 mm depth for transducer coupling and depth confirmation.
Regulatory Classification
Identification
A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
K134032 — ULTHERA SYSTEM · Ulthera, Inc. · Jun 20, 2014
K243035 — Ulthera® System · Ulthera, Inc. · Feb 24, 2025
K233996 — Ulthera System (UC-1 Control Unit PRIME) · Ulthera, Inc. · Feb 22, 2024
K250418 — Ulthera System (UC-1 Control Unit PRIME Model 2.1) · Ulthera, Inc. · May 13, 2025
K132028 — ULTHERA SYSTEM · Ulthera, Inc. · Dec 11, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 4, 2018
Ulthera, Inc Mr. Steven J. Kachelmeyer Executive Director, Regulatory Affairs 1840 South Stapley Drive Suite 200 Mesa, Arizona 85204
Re: K180623
Trade/Device Name: Ulthera System Regulation Number: 21 CFR 878.4590 Regulation Name: Focused Ultrasound Stimulator System for Aesthetic Use Regulatory Class: Class II Product Code: OHV, IYO Dated: March 7, 2018 Received: March 9, 2018
Dear Mr. Kachelmeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180623
Device Name Ulthera System
Indications for Use (Describe)
The Ulthera® System is indicated for use as a non-invasive dermatological aesthetic treatment to:
· lift the eyebrow
• Ift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions
· improve lines and wrinkles of the décolleté
The Ulthera® System in conjunction with the Ulthera® DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The inaging is to visualize the dermal and subdermal layers of tissue to:
· ensure proper coupling of the transducer to the skin
· confirm appropriate depth of treatment such as to avoid bone
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
| Applicant: | Ulthera®, Inc. |
|------------|----------------------------------------------------------------|
| Address: | 1840 South Stapley Drive, Suite 200 |
| | Mesa, AZ 85204 |
| Contact: | Steven J. Kachelmeyer - Executive Director, Regulatory Affairs |
| Telephone: | (480) 619-4069 |
| Fax: | (480) 619-4071 |
Date Prepared: May 2, 2018 Device Trade Name: Ulthera® System
Common Name: Focused Ultrasound for Tissue Heat or Mechanical Cellular Disruption System, Imaging, Pulsed Echo, Ultrasonic Classification Name: 21 C.F.R. § 878.4590, Focused Ultrasound Stimulator Use System for Aesthetic Use Regulatory Class: Class II Product Codes: OHV, IYO Predicate Device: Ulthera®, Inc. Ulthera® System (K134032)
Applicable Guidance: Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.
The device also conforms to Guidance for Industry and FDA Staff: Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers and Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Device Description: The Ulthera® System consists of the Ulthera® Control Unit (with system software), a handpiece with cable, and interchangeable transducers. The device produces controlled tissue coagulation below the skin surface (epidermis) within the first few millimeters of tissue (dermis) using highly focused, low-energy ultrasound deposition. The Ulthera System directs micro-focused acoustic waves to the treatment area at desired depths without affecting or requiring a secondary action to protect the skin surface. The operator may also use the device's supplemental imaging capability to visualize the treatment area and aid in assuring full/proper skin contact of the Ulthera® System transducer to the skin in the target area.
Indications for Use: The Ulthera® System is indicated for use as a non-invasive dermatological aesthetic treatment to:
- lift the eyebrow
- lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions
- improve lines and wrinkles of the décolleté ●
{4}------------------------------------------------
The Ulthera® System in conjunction with the Ulthera® DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:
- . ensure proper coupling of the transducer to the skin
- confirm appropriate depth of treatment such as to avoid bone ●
Summary of Technological Characteristics: High Intensity Focused Ultrasound (HIFU) is the underlying technological principle for both the subject and predicate devices. The Ulthera System functions by directing ultrasonic energy beneath the outer dermis in localized points at a specified distance and depth between points. The design and performance characteristics of the subject and predicate devices are identical. The only technological differences between the devices are minor and do not raise different questions of safety or effectiveness for the subject device as compared to the predicate.
| | Predicate Device: Ulthera® System<br>(K134032) | Subject Device: Ulthera® System,<br>Expanded Indications for Use |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 21 C.F.R. § 878.4590 | 21 C.F.R. § 878.4590 |
| Product Code | OHV, IYO | OHV, IYO |
| Intended Use/<br>Indications for<br>Use | Non-invasive dermatological aesthetic<br>treatment to:<br>• lift the eyebrow<br>• lift lax submental (beneath the chin) and<br>neck tissue<br>• improve lines and wrinkles of the<br>décolleté<br>The Ulthera® System in conjunction with<br>the Ulthera® DeepSEE transducer allows<br>for ultrasonic visualization of depths up to<br>8 mm below the surface of the skin. The<br>indicated use of the imaging is to visualize<br>the dermal and subdermal layers of tissue<br>to:<br>• ensure proper coupling of the<br>transducer<br>to the skin<br>• confirm appropriate depth of<br>treatment such as to avoid bone | Non-invasive dermatological aesthetic<br>treatment to:<br>• lift the eyebrow<br>• lift lax submental (beneath the chin)<br>and neck tissue which can also affect the<br>appearance of lax tissue in the<br>submental and neck regions<br>• improve lines and wrinkles of the<br>décolleté<br>The Ulthera® System in conjunction with<br>the Ulthera® DeepSEE transducer allows<br>for ultrasonic visualization of depths up to<br>8 mm below the surface of the skin. The<br>indicated use of the imaging is to<br>visualize the dermal and subdermal<br>layers of tissue to:<br>•ensure proper coupling of the transducer<br>to the skin<br>•confirm appropriate depth of treatment<br>such as to avoid bone |
| Where Used | Clinic/doctor's office | Clinic/doctor's office |
| Anatomical Site | Skin | Skin |
| Type of Energy | Thermal, < 2 J | Thermal, < 2 J |
| Biological Effect | Lifting of tissue and reduction of lines and<br>wrinkles via High Intensity Focused<br>Ultrasound (HIFU) directed beneath the<br>outer dermis in localized points at a<br>specified depth and distance between | Lifting of tissue and reduction of lines and<br>wrinkles via High Intensity Focused<br>Ultrasound (HIFU) directed beneath the<br>outer dermis in localized points at a<br>specified depth and distance between |
| Patient Contact<br>Material | Biocompatible | Biocompatible |
A table comparing the key features of the subject and predicate devices is provided below.
{5}------------------------------------------------
| Medical<br>Electrical<br>Equipment<br>Safety and EMC | Compliant with relevant standards | Compliant with relevant standards |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Thermal<br>Coagulation<br>Point | Shallow (< 5 mm) and confined – no<br>thermal coagulation below focal zone. | Shallow (< 5 mm) and confined – no<br>thermal coagulation below focal zone. |
| Epidermal<br>Impact | Non-invasive; no cooling required | Non-invasive; no cooling required |
| Pigmentation<br>Effect | Chromophore insensitive | Chromophore insensitive |
| System<br>Software | Version 1382 | Version 1700 – minor software updates to<br>have GUI match cleared labeling in<br>K134032, add improved security for<br>transducer detection, and correct minor<br>anomalies. |
| Device Labeling | Instructions for Use(IFU) and Technical<br>Information Manual (TIM) shipped as two<br>separate documents | Integrated Instructions for Use(IFU) and<br>Technical Information Manual (TIM) into a<br>single document to simplify translations.<br>Added UDI labeling<br>Updated symbols for IEC 60601-1-3.1 ed. |
| Packaging | Control Unit and Handpiece: Corrugated<br>carton<br>Transducer: Nylon / Aluminum foil<br>laminate bag with very low water vapor<br>transmission rate, enclosed within a | Control Unit and Handpiece: Corrugated<br>carton<br>Transducer: Nylon / Aluminum foil<br>laminate bag with very low water vapor<br>transmission rate, enclosed within a |
| Sterilization | Non-sterile | Non-sterile |
| Single-use/<br>Reusable | Controller and handpiece: Reusable<br>Transducers: Single-use | Controller and handpiece: Reusable<br>Transducers: Single-use |
| Shelf Life | 12 months (transducer) | 12 months (transducer) |
Performance Data: To support the expanded indication, the Ulthera® System was evaluated in an open-label clinical trial investigating the clinical response following treatment with the Ulthera System to achieve lifting of lax submental (beneath the chin) and neck tissue. Improvement was evaluated through quantitative assessment, qualitative assessment and patient satisfaction questionnaires. There were 51/70 patients that had an improvement of ≥20 mm² in lift, of these patients 84.3% were identified as showing improvement by masked evaluators. The adverse events resulting from treatment with the Ulthera® System during this study were mild, short-lived in duration, and resolved without incident. There were no serious adverse events (SAEs) or unanticipated adverse device effects (UADEs) related to treatment with the Ulthera® System.
Conclusion: The Ulthera® System is as safe and effective as the predicate Ulthera® System cleared by FDA in K134032. The subject device has the same intended use and very similar technological characteristics and principles of operation as its predicate device. The minor differences in indications do not alter the intended therapeutic use of the device, and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the subject Ulthera® System and its predicate raise no new types of safety or effectiveness questions. Thus, the Ulthera® System with modified indications for use is substantially equivalent to the legally marketed predicate device (K134032).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.