K Number
K233996

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-02-22

(66 days)

Product Code
Regulation Number
878.4590
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ulthera System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin.

The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to:

· Lift the eyebrow

• Lift lax submental (beneath the chin) and neck tissue. which can also affect the appearance of lax tissue in the submental and neck regions

· Improve lines and wrinkles of the décolleté

The Ulthera® System, in conjunction with the Ulthera® DeepSEE® transducer, allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:

· Ensure proper coupling of the transducer to the skin

  • · Confirm appropriate depth of treatment such as to avoid bone
Device Description

The Ulthera® System consists of the Ulthera® Control Unit (with system software), a handpiece with cable, and interchangeable transducers. The device produces controlled tissue coagulation below the skin surface (epidermis) within the first few millimeters of tissue (dermis) using highly focused, low-energy ultrasound deposition. The Ulthera® System directs micro-focused acoustic waves to the treatment area at desired depths without affecting or requiring a secondary action to protect the skin surface. The operator may also use the device's supplemental imaging capability to visualize the treatment area and aid in assuring full/proper skin contact of the Ulthera® System transducer to the skin in the target area.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Ulthera System (UC-1 Control Unit PRIME). The submission aims to demonstrate substantial equivalence to a predicate device (Ulthera System, K180623), primarily focusing on modifications to the console.

Based on the provided text, the device is an aesthetic device (Focused Ultrasound Stimulator System for Aesthetic Use) and NOT a diagnostic device or an AI/ML-driven device that involves human readers or complex image interpretation.

Therefore, most of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of diagnostic AI/ML devices (e.g., sample size, expert ground truth, MRMC studies, standalone performance) is not applicable or not detailed in this submission.

The "study" described is a series of non-clinical performance tests validating the modified console's functionality and safety, rather than a clinical study evaluating diagnostic accuracy or reader improvement.

Here's an attempt to address the points based on the available information, noting where information is not present due to the nature of the device and submission:


Acceptance Criteria and Device Performance for Ulthera System (UC-1 Control Unit PRIME)

The 510(k) submission primarily focuses on demonstrating that modifications to the Ulthera System's console do not significantly affect its clinical functionality or performance compared to the predicate device. The acceptance criteria are essentially that the modified device functions "as intended" and remains "as safe and effective" as the predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

Since this is not a diagnostic device with specific performance metrics like sensitivity/specificity, the "acceptance criteria" relate to engineering, software, and physical performance.

Acceptance Criteria CategoryReported Device Performance/Verification
BiocompatibilityUser-contacting components verified in accordance with ISO 10993-1. (Performance: Verified)
Electrical Safety and EMCVerified in accordance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, and IEC 60601-2-62. (Performance: Compliant)
Software Verification & ValidationConducted as recommended by FDA Guidance, "Content of Premarket Submissions Device Software Functions." A basic documentation level was used as a failure would not present a hazardous situation with a probable risk of death or serious injury. (Performance: Conducted, Passed; Functions as intended.)
UsabilityConducted in accordance with IEC 62366-1 and FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices." Specific checks included console's portability, GUI visibility and layout, and display responsiveness. (Performance: Clinicians able to use device in representative environment; no new risks identified.)
Mechanical TestingDisplay hinge of the console verified via mechanical cyclic testing. (Performance: Verified)
Packaging, Transit, & EnvironmentalPackaging and transit testing per ASTM D4169 and ASTM D4332. Environmental conditioning testing conducted to verify device functionality at defined operating and storage conditions. (Performance: Verified)
Ultrasound Output TestingAcoustic reliability verified using simulated treatment over the device's service life. Power output and imaging were verified to be within specification. (Performance: Within specification, operates as intended)
Overall Comparison (Subject vs. Predicate)The submission states: "These non-clinical performance tests of the Ulthera System demonstrated the modified console continues to operate as intended and that it is as safe and effective as its predicate for the proposed indications for use." (Performance: Demonstrates substantial equivalence, safety, and effectiveness compared to predicate)

2. Sample Size Used for the Test Set and Data Provenance

This is not a clinical study involving a "test set" of patient data in the typical sense for AI/ML. The "testing" involves engineering and software validation on the device itself. Therefore, concepts like "sample size" for patient data and "data provenance" (country, retrospective/prospective) are not applicable here.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. The ground truth for this device's performance is established by engineering specifications, regulatory standards compliance, and direct verification of the device's physical and software functions. There are no "experts" establishing ground truth in the context of interpreting medical images or clinical outcomes for a test set.

4. Adjudication Method for the Test Set

Not applicable, as there is no "test set" of cases requiring adjudication by multiple readers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. An MRMC study is relevant for diagnostic devices where human reader performance (with or without AI assistance) is being evaluated relative to a reference standard. This device is an aesthetic treatment system, and the submission is for console modifications, not a new diagnostic algorithm.

6. If Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No. This is not an AI/ML diagnostic algorithm. The device has an "imaging mode" for visualization and user guidance (proper coupling, depth avoidance) but is fundamentally an energy delivery system for aesthetic treatment. Its performance is tied to its physical and software functions (power output, imaging capabilities, safety features), which were tested directly.

7. The Type of Ground Truth Used

The "ground truth" for the various tests were:

  • Engineering specifications and design requirements: For mechanical, electrical, and acoustic performance.
  • International standards (ISO, IEC, ASTM): For biocompatibility, electrical safety, EMC, usability, packaging, and transit.
  • FDA Guidance: For software verification and validation, and usability/human factors.
  • Predicate device's established safety and effectiveness: The ultimate "ground truth" for substantial equivalence is demonstrating that the modified device performs comparably to the legally marketed predicate.

8. The Sample Size for the Training Set

Not applicable. This device does not employ machine learning that would require a "training set" of data.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for machine learning.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out as "U.S. Food & Drug Administration".

February 22, 2024

Ulthera, Inc. Kim Kelly Sr. Director, Regulatory Affairs 1840 S. Stapley Dr. Suite 200 Mesa, Arizona 85204

Re: K233996

Trade/Device Name: Ulthera System (UC-1 Control Unit PRIME) Regulation Number: 21 CFR 878.4590 Regulation Name: Focused Ultrasound Stimulator System For Aesthetic Use Regulatory Class: Class II Product Code: OHV Dated: January 23, 2024 Received: January 23, 2024

Dear Kim Kelly:

We have reviewed vour section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.02.22 15:01:25 -05'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233996

Device Name

Ulthera System (UC-1 Control Unit PRIME)

Indications for Use (Describe)

The Ulthera System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin.

The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to:

· Lift the eyebrow

• Lift lax submental (beneath the chin) and neck tissue. which can also affect the appearance of lax tissue in the submental and neck regions

· Improve lines and wrinkles of the décolleté

The Ulthera® System, in conjunction with the Ulthera® DeepSEE® transducer, allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:

· Ensure proper coupling of the transducer to the skin

  • · Confirm appropriate depth of treatment such as to avoid bone
    Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Description 3.0 (Part 2: Support & Support 2)

I he-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

APPLICANT

Company's Name:Ulthera, Inc.
Company's Address:1820 South Stapley Drive. Suite 200Mesa, AZ 85204
Telephone:(984) 286-9190
Contact Person:Kim Kelly, Senior Director, Regulatory Affairs
Date Prepared:February 22, 2024

DEVICE

Device Name:Ulthera System (UC-1 Control Unit PRIME)
Classification Name:Focused Ultrasound Stimulator System for Aesthetic Use
Classification Name:21 C.F.R § 878.4590, Focused Ultrasound Stimulator Use
Regulatory Class:Class II
Product Codes:OHV, IYO
Applicable Guidances:Focused Ultrasound Stimulator System for Aesthetic Use
Applicable Guidances:Marketing Clearance of Diagnostic Ultrasound Systems andTransducers
Applicable Guidances:Content of Premarket Submissions for Software Containedin Medical Devices.

PREDICATE DEVICE

Ulthera System, Ulthera, Inc., K180623

{5}------------------------------------------------

DEVICE DESCRIPTION

The Ulthera® System consists of the Ulthera® Control Unit (with system software), a handpiece with cable, and interchangeable transducers. The device produces controlled tissue coagulation below the skin surface (epidermis) within the first few millimeters of tissue (dermis) using highly focused, low-energy ultrasound deposition. The Ulthera® System directs micro-focused acoustic waves to the treatment area at desired depths without affecting or requiring a secondary action to protect the skin surface. The operator may also use the device's supplemental imaging capability to visualize the treatment area and aid in assuring full/proper skin contact of the Ulthera® System transducer to the skin in the target area.

INTENDED USE / INDICATIONS FOR USE

The Ulthera System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin.

The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to:

  • Lift the eyebrow .
  • . Lift lax submental (beneath the chin) and neck tissue, which can also affect the appearance of lax tissue in the submental and neck regions
  • Improve lines and wrinkles of the décolleté ●

The Ulthera® System, in conjunction with the Ulthera® DeepSEE® transducer, allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:

  • Ensure proper coupling of the transducer to the skin
  • . Confirm appropriate depth of treatment such as to avoid bone

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS

The purpose of this 510(k) notification relates to modifications to the console of the Ulthera System. Low-intensity, highly focused ultrasound is the main technological principle of both the subject and the predicate device. In both devices, focused ultrasound energy is delivered below the skin and produces discrete points of thermal coagulation that results in contraction of the skin, which produces a lifting or tightening effect and improves appearance of lax tissue, lines, and wrinkles. Both the subject and predicate device share the following similar technological elements:

  • · Identical ultrasound signal energy and ultrasound treatment guidelines

{6}------------------------------------------------

  • · Same main components, including a console with integrated touchscreen, connected handpiece, and interchangeable transducers
  • · An imaging mode of operation to visualize the treatment area and aid in assuring full/proper skin contact of the transducer to the skin

The following differences between the subject and predicate device do not significantly affect clinical functionality or performance specifications of the device and have been verified and tested:

  • · Updates to the console's industrial design and modernization of the graphics within the graphical user interface (GUI)
  • Hardware updates to comply with the latest electromedical standards and address end-of-life components

Comparison between the subject and predicate device are summarized in Table 1 below:

Ulthera Modified ConsoleUlthera Console K180623
Subject Device(Predicate)
Intended UseThe Ulthera System is intended to applyfocused ultrasound energy to the body toachieve temporary changes in thephysical appearance of the skin.The Ulthera System is intended to applyfocused ultrasound energy to the body toachieve temporary changes in thephysical appearance of the skin.
Indications forUseThe Ulthera® System is indicated for useas a non-invasive dermatologicalaesthetic treatment to:The Ulthera® System is indicated foruse as a non-invasive dermatologicalaesthetic treatment to:
● lift the eyebrow● lift the eyebrow
● lift lax submental (beneath thechin) and neck tissue; which canalso affect the appearance of laxtissue in the submental and neckregions● lift lax submental (beneath thechin) and neck tissue; whichcan also affect the appearanceof lax tissue in the submentaland neck regions
● improve lines and wrinkles ofthe décolleté.● improve lines and wrinkles ofthe décolleté.
The Ulthera® System in conjunctionwith the Ulthera® DeepSEE transducerallows for ultrasonic visualization ofdepths up to 8 mm below the surface ofthe skin. The indicated use of theimaging is to visualize the dermal andsubdermal layers of tissue to:The Ulthera® System in conjunctionwith the Ulthera® DeepSEE transducerallows for ultrasonic visualization ofdepths up to 8 mm below the surface ofthe skin. The indicated use of theimaging is to visualize the dermal andsubdermal layers of tissue to:
● ensure proper coupling of thetransducer to the skin● ensure proper coupling of thetransducer to the skin
● confirm appropriate depth oftreatment such as to avoid bone● confirm appropriate depth oftreatment such as to avoid bone
Table 1: Substantial Equivalence Table Comparing Subject & Predicate Devices

{7}------------------------------------------------

User PopulationTreatment of adult patient population by trained medical professionalTreatment of adult patient population by trained medical professional
Main System Components• Control console• Handpiece• Transducers• Control console• Handpiece• Transducers
Additional Components Required for Operation• ACLF Power Cord with pigtail adapter• USB Access Key• Power cord• USB Access Key
Dimensions(l x w x h)Height: <16.69" (424 mm)Width: 19.4" (493.6 mm)Depth: 13.1" (333 mm)Height: 15.3" (389 mm)Width: 16.5" (419 mm)Depth: 13.0" (330 mm)
WeightWeight: ≤27 lbs (12.2 kg)Weight: 22 lbs (10 kg)
Display18.5" screen; aspect ratio of 16:9 with1920 x 1080 resolution15" screen; aspect ratio of 4:3 with 1024x 768 resolution
Power Source100-240 VAC, 50/60 Hz, 3A maxFuse: (2) 5x20mm, 6.3A fast acting,250V100-240 VAC, 50/60 Hz, 3A maxFuse: (2) 5x20mm, 6.3A fast acting,250V
BiocompatibilityBiocompatible user contacting surfacesBiocompatible user contacting surfaces
Softwareversion 2.1.2030version 1700
OperatingSystemWindows 10Windows XP
ElectricalSafety & EMCstandardsCompliant with relevant IEC standardsfor console and ACLF power cordCompliant with relevant IEC standardsfor console

PERFORMANCE DATA

The following non-clinical data were provided to support the substantial equivalence of the subject device to the predicate device. In all instances, the subject device functioned as intended.

Biocompatibility

Biocompatibility of the user-contacting components of the device was verified in accordance with ISO 10993-1.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC of the subject device were verified in accordance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, and IEC 60601-2-62.

Software Verification and Validation Testing

{8}------------------------------------------------

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA Guidance, "Content of Premarket Submissions Device Software Functions." A basic documentation level was used, as a failure or flaw of any device software function(s) would not present a hazardous situation with a probable risk of death or serious injury prior to the implementation of risk control measures.

Usability

Usability testing was conducted in accordance with IEC 62366-1 and FDA Guidance, "Applving Human Factors and Usability Engineering to Medical Devices." The console's portability, GUI visibility and layout, and display responsiveness were also verified. Testing demonstrated that clinicians were able to use the device in a representative environment and use conditions. No new risks were identified during the stimulated use study.

Mechanical Testing

The display hinge of the console was verified via mechanical cyclic testing.

Packaging, Transit, and Environmental Testing

Packaging and transit testing were conducted per ASTM D4169 and ASTM D4332. Environmental conditioning testing was conducted to verify device functionality at defined operating and storage conditions.

Ultrasound Output Testing

Acoustic reliability was verified using simulated treatment over the device's service life. Power output and imaging were verified to be within specification.

CONCLUSIONS

The modified console has undergone robust performance testing, including software, mechanical, electrical safety, electromagnetic compatibility, packaging, environmental, and usability testing. Imaging and acoustical testing were also conducted to verify the ultrasound signal. These non-clinical performance tests of the Ulthera System demonstrated the modified console continues to operate as intended and that it is as safe and effective as its predicate for the proposed indications for use.

§ 878.4590 Focused ultrasound stimulator system for aesthetic use.

(a)
Identification. A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.