K243342 · Kp Trading Co., Ltd. · FXX · Dec 22, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243342
Device Name
KP Protective Face Mask
Applicant
Kp Trading Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
Dec 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The KP Protective Face Mask is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and operating room personnel from transfer of microorganisms, body fluids, and particulate material.
Device Story
KP Protective Face Mask is a non-sterile, single-use surgical mask; provides physical barrier to fluids, microorganisms, and particulate matter. Worn by operating room personnel during surgical procedures. Constructed of three nonwoven polypropylene layers: outer spunbond, melt-blown filter, and inner comfort layer; includes polypropylene-clad adjustable nose wire and elastic nylon/polyurethane earloops. Device functions as a passive barrier; no active components or software. Benefits include reduction of cross-contamination risk between surgical staff and patients.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM F2101 (BFE >99%), ASTM F3502-25 (PFE ≥85%), ASTM F1862 (120 mmHg fluid resistance), and EN 14683/MIL-M-36954 (breathability <6.0 mmH₂O/cm²). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation, PII=0.0), and ISO 10993-10 (sensitization, 0% response).
Technological Characteristics
Three-layer nonwoven polypropylene construction (spunbond/melt-blown/spunbond). Features polypropylene-clad adjustable nose wire and nylon/polyurethane earloops. Flat-fold style. Non-sterile. Complies with ASTM F2100-19 Level 2, ASTM F2101, ASTM F3502, ASTM F1862, and 16 CFR 1610 (Class I flammability). Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for operating room personnel during surgical procedures to protect the patient and personnel from transfer of microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
K203161 — Surgical Face Mask · Wuhan Raytex Protection Co., Ltd. · Jan 20, 2022
K212299 — Medical Surgical Mask · Wepon Medical Technology Co., Ltd. · Jan 4, 2022
K211773 — Single-use Surgical Mask · Taiwan Seafood and Fish Corporation · Dec 2, 2021
K210840 — Disposable Surgical Mask · Safeway Industry Healthcare. · Jun 22, 2021
K202746 — Surgical mask · Ningbo Jingeao Electronics, Inc. · Aug 27, 2021
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
KP Trading Co., Ltd.
Ji Kwon
Director
5,6 Dong 572054, Chorok-ro, Yanggam-myeon
Hwaseong-si, Gyeonggido 18627
Korea, South
Re: K243342/S001
Trade/Device Name: KP Protective Face Mask
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical Apparel
Regulatory Class: Class II
Product Code: FXX
Dated: November 20, 2025
Received: November 20, 2025
Dear Ji Kwon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243342 - Ji Kwon
Page 2
cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K243342 - Ji Kwon
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# BIFENG QIAN -S
Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243342 | |
| Device Name KP Protective Face Mask | |
| Indications for Use (Describe) The KP Protective Face Mask is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and operating room personnel from transfer of microorganisms, body fluids, and particulate material. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) SUMMARY K243342
# I. SUBMITTER
Company Name: KP TRADING CO., LTD.
Company Address: 5,6 Dong 372-54 Chorok-ro, Yanggam myeon, Hwaseong-si, Gyeonggi do, Republic of Korea
Contact Person: JI HYE KWON
+63 945 750 1811
+82 10 5358 2538
jihyekwon@kptrading.co.kr
Date of summary: December 18, 2025
# II. SUBJECT DEVICE
Proprietary Name: KP Protective Face Mask
Common Name: Surgical Mask
Classification Name: Surgical Apparel
Regulation Number: 21 CFR 878.4040
Product Code: FXX
Device Class: Class II
# III. PREDICATE DEVICE
Predicate Device: Techno Web Surgical Mask
Predicate 510(k) Number: K172500
Product Code: FXX
# IV. DEVICE DESCRIPTION
The KP Protective Face Mask is a non-sterile, single-use surgical mask intended to cover the user’s nose and mouth, providing a physical barrier to fluids, microorganisms, and particulate matter.
The mask consists of three nonwoven polypropylene layers, including:
- Outer layer: Spunbond non-woven
- Filter layer: Melt-blown polypropylene
- Inner comfort layer: Spunbond non-woven
- Nose wire: Polypropylene-clad adjustable wire
- Attachment: Elastic earlops (A mixture of nylon and polyurethane)
The mask is white, flat-fold style, supplied non-sterile, and packaged in standard pouches.
1 of 5
{5}
2 of 5
V. INDICATIONS FOR USE
The KP Protective Face Mask is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and operating room personnel from transfer of microorganisms, body fluids, and particulate material.
VI. COMPARISON TABLE
A comparison between the proposed device and predicate device are shown in the table below for the purpose of presenting the equivalence in relation to effectiveness and safety.
| Item | Primary Predicate Device (K172500) | Subject Device (K243342) | Comparison |
| --- | --- | --- | --- |
| Regulatory Class | Class II | Class II | Same |
| Product Code | FXX | FXX | Same |
| Intended Use | Techno Web Surgical Mask is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and operating room personnel from transfer of microorganisms, body fluids, and particulate material. | KP Protective Face Mask is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and operating room personnel from transfer of microorganisms, body fluids, and particulate material. | Same |
| Materials | | | |
| --- | --- | --- | --- |
| Outer Cover Web | Felt (Nonwoven filtration web) | Polypropylene Spunbond Non-woven fabric woven fabric (40gsm, SSS) | Similar |
| Stiffener Web | N/A | N/A | Same |
{6}
| Filter Web | Felt (Nonwoven filtration web) | Polypropylene Melt blown non-woven fabric (40g sm. FFP2, Water Agent process | Similar |
| --- | --- | --- | --- |
| Inner Web | Felt (Nonwoven filtration web) | Polypropylene Spunbond Non-woven fabric (20gsm, SSS) | Similar |
| Nose-Clip | Aluminum | Polypropylene clad wire | Similar |
| Staple | N/A | N/A | Same |
| Ear loop | Ear loop: Nylon | Ear loops: A mixture of Nylon and Polyurethane strap | Similar |
| Nose Foam | Plastic coating | N/A | Different |
| Surgical face mask style | | | |
| Fold | Flat fold | Flat fold | Same |
| Layers | Multi (three-layers) | Multi (three layers) | Same |
| Sterility | | | |
| --- | --- | --- | --- |
| | Non-Sterile Single Use | Non-Sterile Single Use | Same |
| Specifications and Dimensions | | | |
| Color | White | White | Same |
| Length | 177.8mm | 205 ± 10mm | Similar |
| Width | 74mm | 85.1 ± 6.4mm | Similar |
| Ear loops Length | 138mm | 165 ± 16mm | Similar |
| Performance Specifications | | | |
{7}
| Particulate Filtration Efficiency (PFE) | Pass ASTM F2299 At least 99.71% | Pass 3 non-consecutive ASTM F3502 (75±20 nm NaCl): 92.15-95.96% | Similar |
| --- | --- | --- | --- |
| Differential Pressure Delta P | Pass — EN 14683:2019 < 6.0 mmH₂O/cm² (Level 2) | Pass — ASTM F2100 (Level 2) Tested on 3 non-consecutive production lots Differential pressure range: approx. 4.3–5.5mmH₂O/cm² | Similar |
| Fluid Resistance | Pass 10/10 at 120mmHg | Pass 3 non-consecutive lots 32/32 at 120mmHg ASTM F1862/F1862M-17 | Similar |
| Bacterial Filtration Efficiency | Pass ≥ 99 (%) | Pass >99% ASTM F2101-2019 | Same |
| Flammability (CFR 16 1610) | Pass Class I | Pass 16 CFR 1610 (Class I) | Same |
| Certification | | | |
| Surgical face mask | Classified as Surgical Mask | Classified as Surgical Mask | Same |
| Biocompatibility | | | |
| Cytotoxicity (ISO 10993-5) | Biocompatibility assessment performed on patient-contact materials (per publicly available 510(k) Summary) | Sample showed a non-cytotoxic effect | Similar |
| Irritation (ISO 10993-10:2010) | | Primary Irritation Index (PII) = 0; negligible irritation | Similar |
| Cytotoxicity (ISO 10993-10: 2010) | | Sensitization rate = 0%; no skin sensitization observed | Similar |
4 of 5
{8}
5 of 5
# VII. PERFORMANCE TESTING SUMMARY
The KP Protective Face Mask was evaluated in accordance with ASTM F2100-19 Level 2 requirements. Testing included BFE, PFE, Synthetic Blood Penetration, Differential Pressure, Flammability, and Biocompatibility (ISO 10993). All results met acceptance criteria.
| Test Method | Purpose | Acceptance | Results |
| --- | --- | --- | --- |
| ASTM F2101 | Measure filtration efficiency against S. aureus aerosol | ≥98% (Level 2) | PASS ≥99% |
| ASTM F3502-25 | Measure submicron particle filtration (0.1μm) | ≥85 (Level 2) | PASS ≥85% |
| ASTM F1862 | Evaluate resistance to synthetic blood at pressure | 120 mmHg - No penetration | PASS - no penetration |
| EN 14683 / MIL-M-36954 | Measure breathability (ΔP) | <6.0 mmH₂O/cm² (Level 2) | PASS |
| 16 CFR 1610 | Evaluate flammability of mask materials | Class 1 (Normal flammability) | PASS - Class 1 |
# Biocompatibility Testing
The relevant standards for biocompatibility testing of the subject device are presented as follows.
| Test Method | Purpose | Acceptance | Results |
| --- | --- | --- | --- |
| ISO 10993-10:2010 | Irritation | Primary irritation index (PII) ≤ 0.4 = negligible irritation | PII = 0.0, No irritation observed |
| ISO 10993-10:2010 | Sensitization | 0% sensitization response = Pass | 0% sensitization response, no sensitization observed |
| ISO 10993-5:2019 | In vitro Cytotoxicity | ≥70% cell viability = non-cytotoxic | Non-Cytotoxic |
# Clinical Testing
Clinical testing is not applicable to the subject device.
# VIII. CONCLUSION
The conclusions drawn from the nonclinical tests demonstrate that the KP Protective Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate devices, the Techno Web Surgical Mask (K172500).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.