K Number
K211773

Validate with FDA (Live)

Date Cleared
2021-12-02

(177 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Single Use Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. This is a single use device provided non-sterile.

Device Description

The Single-Use Surgical Mask is a 3-ply, disposable surgical face mask. The outer and inner layers of the mask are made of spunbond polypropylene, while the middle layer consists of a meltblown polypropylene filter fabric. The device is secured on the user's face via elastic earloops and are labeled non-sterile, single-use, and disposable. This device is a surface device intended to come in contact with intact skin for a duration of less than 24 hours.

AI/ML Overview

The provided document describes the acceptance criteria and the study that proves the performance of a Single-Use Surgical Mask (K211773). This is a non-clinical study, as stated in section 8. Clinical Testing.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

Item/Test MethodPurposeAcceptance CriteriaReported Device Performance (Result)
Fluid Resistance (ASTM 1862)Demonstrate resistance to fluid sprayPass at 80 mmHg (level 1 fluid resistance)Pass
Bacterial Filtration Eff. (ASTM F2101-19)Demonstrate resistance to bacterial penetrationPass at ≥ 99%Pass
Particulate Filtration Eff. (ASTM F2299)Demonstrate resistance to penetration of particulate matterPass at ≥ 95%Pass
Differential Pressure (EN 14683)Demonstrate breathabilityPass at < 5.0 mm H2O/cm²Pass
Flammability (16 CFR 1610)Demonstrate low flammability Class 1(Implicitly Class 1, based on predicate)Pass
Biocompatibility: Cytotoxicity (ISO 10993-5)Demonstrate low cytotoxic potentialUnder the conditions of the study, non-cytotoxicPass
Biocompatibility: Skin Sensitization (ISO 10993-10)Demonstrate low skin sensitization potentialUnder the conditions of the study, not a sensitizerPass
Biocompatibility: Skin Irritation (ISO 10993-10)Demonstrate low skin irritation potentialUnder the conditions of the study, not an irritantPass
ASTM F2100 Level (ASTM F2100)Demonstrates face mask performanceMeet Level 1 requirementsPass

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample sizes used for each individual test. It mentions "non-clinical testing" and the results are presented as "Pass" for each criterion. The data provenance is also not specified in terms of country of origin or whether it's retrospective or prospective, beyond being non-clinical (laboratory-based) tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable. The device is a surgical mask, and its performance criteria are objectively measured via standardized laboratory tests (e.g., ASTM, ISO, EN standards) rather than requiring expert interpretation for ground truth.

4. Adjudication method for the test set

This information is not applicable. The tests are laboratory-based and produce quantitative results compared against predefined thresholds, not subjective assessments requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study. This document describes the performance of a physical device (surgical mask) through non-clinical testing, not an AI or diagnostic imaging device that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The subject is a physical surgical mask, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the performance of the surgical mask is established by the specified industry and regulatory standards and test methods (e.g., ASTM F1862 for fluid resistance, ASTM F2101-19 for Bacterial Filtration Efficiency, ISO 10993 for biocompatibility). These standards define objective criteria and measurement procedures.

8. The sample size for the training set

This information is not applicable. This is not a study involving a training set for an algorithm or a similar machine learning context. The tests are for a physical product.

9. How the ground truth for the training set was established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 2, 2021

Taiwan Seafood and Fish Corporation Eric Ly Business Development Assistant 733 Gladys Avenue Los Angeles, California 90021

Re: K211773

Trade/Device Name: Single-use Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 26, 2021 Received: September 1, 2021

Dear Eric Ly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211773

Device Name Single-Use Surgical Mask

Indications for Use (Describe)

The Single Use Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. This is a single use device provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. Submitter Information

Submitted by:

Date Prepared:

Contact Person:

2. Subject Device

Proprietary Name: Common Name: Classification Name and Reference: Device Product Code, Device Panel:

3. Predicate Device

  • Predicate Device 510(k) Number: Submitter: Proprietary Name: Device Product Code: Regulation:

4. Device Description

The Single-Use Surgical Mask is a 3-ply, disposable surgical face mask. The outer and inner layers of the mask are made of spunbond polypropylene, while the middle layer consists of a meltblown polypropylene filter fabric. The device is secured on the user's face via elastic earloops and are labeled non-sterile, single-use, and disposable. This device is a surface device intended to come in contact with intact skin for a duration of less than 24 hours.

5. Indications for Use

The Single Use Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. This is a single use device provided non-sterile.

The Indications for Use statement for the Single-Use Surgical Mask is not identical to the predicate device; however, the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for the protection of both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material.

6. Comparison of Technological Characteristics

Taiwan Seafood and Fish Corporation 733 Gladys Avenue Los Angeles, CA 90021 213-624-2927

11/18/2021

Eric Ly Business Development Assistant erickltsf@gmail.com

Single-Use Surgical Mask Surgical Face Mask 21 CFR 878.4040, Class II FXX, General Hospital

K202824 Blackbriar Regulatory Services, LLC BRS Procedure Face Mask FXX 21 CFR 878.4040

{4}------------------------------------------------

The Single-Use Surgical Mask is made of an outer layer consisting of spunbond polypropylene fabric, a middle layer of meltblown polypropylene filter, an inner layer of spunbond polypropylene, 2 earloops made out of nylon and spandex, and a polyethylene nose wire. The technological characteristics of this device are similar to the predicate device because the materials are similar or identical.

ItemSubject DevicePredicate DeviceComparison
510(k) NumberK211773K202824-
Trade NameSingle-Use Surgical MaskBRS Procedure Face Mask-
Indications ForUseThe Single Use Surgical Mask isintended to be worn to protectboth the patient andhealthcare personnel from thetransfer of microorganisms,body fluids, and particulatematerial. This is a single usedevice provided non-sterile.BRS Procedure Face Mask is a non-sterile, single use, disposable face maskintended to be worn to protect bothhealthcare personnel and patients fromthe transfer of microorganisms, bloodand body fluids, and particulate material.This equipment is intended for the use ininfection control practices to reduce thepotential exposure to blood and fluids.BRS Level 1 Procedure Face Mask(Model# 00850021617004)Similar
MaterialsOuter Layer: SpunbondpolypropyleneMiddle Layer: Meltblownpolypropylene filter InnerLayer: SpunbondPolypropyleneEarloop: Spandex and NylonNose piece: Polyethylene wireOuter Layer: Spunbond polypropyleneMiddle Layer: Meltblown polypropylenefilter Inner Layer: SpunbondpolypropyleneEarloop: Spandex and NylonNosepiece: iron wire covered in plasticSimilar
Mask Style3-ply, disposable, non-sterile3-ply, disposable, nonsterileSame
Mask Dimensions175mm x 95mm175mm x 95mmSame
ASTM F2100 LevelLevel 1Level 1 variantSimilar
Biocompatibility(ISO 10993-5, 10)Under the conditions of study,non-cytotoxic, non-irritating,and non-sensitizingUnder the conditions of study, non-cytotoxic, non-irritating, and non-sensitizingSame

The subject device meets the same ASTM F2100 requirements as the ASTM F2100 Level 1 variant of the predicate device.

7. Summary of Nonclinical Testing

Item/Test MethodPurposeCriteriaResult
Fluid Resistance(mmHg)(ASTM 1862)Demonstrate resistance to fluid sprayPass at 80 mmHg (level 1 fluid resistance)Pass
Bacterial Filtration Efficiency (%)(ASTM F2101-19)Demonstrate resistance to bacterial penetrationPass at $≥$ 99%Pass

{5}------------------------------------------------

Particulate FiltrationEfficiency (%)(ASTM F2299)Demonstrate resistance topenetration of particulatematterPass at ≥ 95%Pass
Differential Pressure(EN 14683)Demonstrate breathabilityPass at < 5.0 mm H2O/cm²Pass
Flammability(16 CFR 1610)Demonstrate low flammability Class 1Pass
Biocompatibility:Cytotoxicity(ISO 10993-5)Demonstrate low cytotoxicpotentialUnder the conditions of the study, non-cytotoxicPass
Biocompatibility:Skin sensitization(ISO 10993-10)Demonstrate low skinsensitization potentialUnder the conditions of the study, not asensitizerPass
Biocompatibility:Skin Irritation(ISO 10993-10)Demonstrate low skinirritation potentialUnder the conditions of the study, notan irritantPass
ASTM F2100 Level(ASTM F2100)Demonstrates face maskperformanceMeet Level 1 requirementsPass

8. Clinical Testing

No clinical testing is included in this submission.

9. Conclusion

The conclusions from the non-clinical testing demonstrate that the Single Use Surgical Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device, BRS Level 1 Procedure Face Mask (K202824).

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.