K251798 · Ray Co., Ltd. · OAS · Nov 25, 2025 · Radiology
Device Facts
Record ID
K251798
Device Name
RCT700
Applicant
Ray Co., Ltd.
Product Code
OAS · Radiology
Decision Date
Nov 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
RCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals
Device Story
RCT700 is a dental CBCT, panoramic, and cephalometric X-ray system. It uses a rotating C-arm with an X-ray tube and detector to capture 3D volumetric data of dental/maxillofacial structures and 2D panoramic/cephalometric images. Operated by dentists or qualified professionals in clinical settings, the device transforms X-ray projections into diagnostic images via software. Output is viewed by clinicians to support diagnosis and orthodontic treatment planning. The system includes an optional model scan and carpus imaging for growth assessment. It benefits patients by providing high-resolution diagnostic imaging for dental and ENT conditions.
Clinical Evidence
Clinical evidence consisted of a review of sample clinical images collected from the new detectors by a licensed practitioner. Images were gathered from diverse patients (varying age, gender, size) using standard protocols. The practitioner deemed the image quality acceptable for the intended diagnostic use. No formal clinical trial with primary endpoints was required; bench testing per FDA guidance for solid-state X-ray imaging devices provided the primary performance validation.
Technological Characteristics
System includes C-arm, X-ray generator, and digital detectors (e.g., FXDD-0606CA, Jupi0606X1, Pluto0900X). Operates at 60-100kVp, 1-17mA. Connectivity via Giga-Ethernet. DICOM 3.0 compatible. Standards: IEC 60601-1, IEC 60601-1-3, IEC 60601-2-63, IEC 60601-1-2. Software version 2.5.0.0.
Indications for Use
Indicated for radiographic examination of dento-maxillofacial, sinus, TMJ, airway, and ENT structures in adult and pediatric patients (recommended >5 years old). Includes carpus imaging for orthodontic growth assessment. No restrictions on gender, ethnicity, or health condition.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 25, 2025
RAY Co., Ltd.
% Suyeon Baek
RA Team
1F~3F, 4F(Part), 5F, 265, Daeji-Ro, Suji-gu
Yongin-si, Gyeonggi-do 16882
REPUBLIC OF KOREA
Re: K251798
Trade/Device Name: RCT700
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: OAS
Dated: July 1, 2025
Received: October 30, 2025
Dear Suyeon Baek:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251798 - Suyeon Baek
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251798 - Suyeon Baek
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251798 | |
| Device Name RCT700 | |
| Indications for Use (Describe) RCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Submission- RCT700
# 510(k) Summary
K251798
# 1. 510(k) Summary
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
# 2. Date
July 4, 2025
3. Administrative Information
| Applicant | | Ray Co., Ltd. |
| --- | --- | --- |
| Address | | 1F~3F, 4F(Part), 5F, 265, Daeji-ro, Suji-gu, Yongin-si, 16882, Korea |
| Manufacturer | Name | Ray Co., Ltd. |
| | Address | 1F~3F, 4F(Part), 5F, 265, Daeji-ro, Suji-gu, Yongin-si, 16882, Korea |
| | Tel | +82-31-605-1000 |
| | Fax | +82-2-6280-5534 |
| Contact Person | Name | Suyeon Baek |
| | Email | suyeon.baek@raymedical.com |
4. Device Information
| Trade/Proprietary Name | | RCT700 |
| --- | --- | --- |
| Common Name | | Dental Panoramic/Tomography and Cephalometric X-ray System |
| Classification Name | Device | Computed tomography x-ray system |
| | Regulation Number | 21 CFR 892.1750 |
| | Class | 2 |
| | Product Code | OAS |
| | Review Panel | Radiology |
{5}
510(k) Submission- RCT700
# 5. Predicate device
| Parameter | Predicate Device |
| --- | --- |
| Device Name | RCT700 |
| Manufacturer | Ray Co., Ltd |
| 510(K) Number | K213226 |
| Classification name | Computed tomography x-ray system |
| Regulation number | 892.1750 |
| Primary product code | OAS |
# 6. Device Description
RCT700 provides 3D computed tomography for scanning hard tissues such as bone and teeth. By rotating the C-arm, which houses a high-voltage generator, an X-ray tube and a detector on each end, CBCT images of dental maxillofacial structures are obtained by recombining data scanned from the same level at different angles. Functionalities include panoramic image scanning for obtaining images of whole teeth, and a cephalometric option for obtaining cephalometric images.
# 7. Indications for Use
RCT700 is CBCT and panoramic x-ray imaging system with cephalometric.
Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.
The device is to be operated and used by dentists or other legally qualified heath care professionals
# 8. Patient population
The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure.
There is no restriction for ethnic group, Gender, age, weight, health, or condition.
We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.
{6}
510(k) Submission- RCT700
# 9. Comparison with predicate device
The following table provides a summary of the technological characteristics of RCT700 compared to the predicate device.
| Parameter | | Subject Device | Predicate Device |
| --- | --- | --- | --- |
| Manufacturer | | RAY Co., Ltd. | RAY Co., Ltd. |
| Device name | | RCT700 | RCT700 |
| 510(K) Number | | Traditional 510K | K213226 (Traditional 510K) |
| Common Name | | Dental panoramic/tomography and cephalometric x-ray system | Dental panoramic/tomography and cephalometric x-ray system |
| Indications for use | | Same as predicate device | RCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals |
| Mode of Operation | | Same as predicate device | Continuous operation with intermittent, stated permissible loading |
| 3D technology | | Same as predicate device | CBCT Cone beam Computed Tomography |
| Performance Specification | | Same as predicate device | 1) CBCT Computed tomography - Patient 2) Panoramic 3) Cephalometric(optional) - One shot type - Scan type |
| Functional Option | | Same as predicate device | Base CT+PANO Option(CEPH) CT + PANO + SCAN CEPH CT + PANO + One shot(One shot, Standard Type) CT + PANO + One shot(One shot, Large Type). |
| Detector Type | CT | Same as predicate device | FXDD-0606CA |
| | | Same as predicate device | Jupi0606X1 |
{7}
510(k) Submission- RCT700
| | PANO | Same as predicate device | FXDD-0606CA |
| --- | --- | --- | --- |
| | | Same as predicate device | Jupi0606X1 |
| | Ceph (Scan) | Same as predicate device | XID-C24DC |
| | | Pluto0900X | N/A |
| | Ceph (One shot) | Same as predicate device | FXDD-1012CA |
| | | Same as predicate device | FXRD-1717VA |
| Exposure switch Type | | Same as predicate device | “Deadman” Button type |
| Main Components | Same as predicate device | Ceph Apparatus | |
| | | Same as predicate device | Vertical Carriage |
| | | Same as predicate device | Rotator |
| | | Same as predicate device | X-RAY Generator |
| | | Same as predicate device | X-ray tube |
| | | Same as predicate device | High Frequency Generator |
| | | Same as predicate device | Column |
| | | Control panel | Touch monitor (panel) |
| | | Detector | |
| | | - CT | Detector |
| | | FXDD-0606CA | - CT |
| | | Jupi0606X1 | FXDD-0606CA |
| | | | Jupi0606X1 |
| | | - PANO | - PANO |
| | | FXDD-0606CA | FXDD-0606CA |
| | | Jupi0606X1 | Jupi0606X1 |
| | | - Ceph | - Ceph |
| | | Pluto0900X(Scan) | XID-C24DC(Scan) |
| | | XID-C24DC(Scan) | FXDD-1012CA(One shot, Standard Size) |
| FXDD-1717VA(One shot, Large Size) | FXRD-1717VA(One shot, Large Size) | | |
| Same as predicate device | | Chinrest | |
| Same as predicate device | | Head rest | |
| Same as predicate device | | Automatic Collimator | |
| Same as predicate device | | Exposure switch | |
| Same as predicate device | | Emergency stop switch | |
| Same as predicate device | | Console PC set | |
| Automatic Collimator | | Same as predicate device | CT examsPanoramic examsCephalometric exams |
{8}
510(k) Submission- RCT700
| Class | | Same as predicate device | Class I with type B applied parts according to IEC 60601-1 |
| --- | --- | --- | --- |
| Focal size | | Same as predicate device | 0.5 |
| Field of View(CT) | | Same as predicate device | Max.160x100 mm |
| X-ray Voltage | | Same as predicate device | 60~100kVp |
| X-ray Current | | Same as predicate device | 1~17mA |
| Total Filtration | | Same as predicate device | Min. 2.8 mm Al equivalent |
| Detector Pixel size | CT | Same as predicate device | FXDD-0606CA: 119μm |
| | | Same as predicate device | Jupi0606X1: 100μm |
| | PANO | Same as predicate device | FXDD-0606CA: 119μm |
| | | Same as predicate device | Jupi0606X1: 100μm |
| | Ceph (Scan) | Same as predicate device | XID-C24DC: 100μm |
| | | Pluto0900X: 100μm | N/A |
| | Ceph(One shot) | Same as predicate device | FXDD-1012CA: 124μm |
| | | Same as predicate device | FXRD-1717VA: 140μm |
| Format compatible | | Same as predicate device | DICOM 3.0 Format compatible |
| Image Viewing Software | | RayScan(version 2.5.0.0) | RayScan(version 2.4.0.0) |
| Image acquisition | | Same as predicate device | Giga-Ethernet Network |
| Total Height | | Same as predicate device | Max 2,296mm |
| Weight | | Same as predicate device | 1) Computed Tomography(CT) + Panoramic(PANO)=185kg(407.9lb) ± 10%2) Computed Tomography(CT) + Panoramic(PANO) + Ceph (Scan type)= 212.5kg (468.5lb) ± 10%3) Computed Tomography(CT) + Panoramic(PANO) + Ceph (One shot type, installed in Standard size)= 211kg (465.2lb) ± 10%4) Computed Tomography(CT) + Panoramic(PANO) + Ceph (One shot type, installed in Large size) 211kg (465.2lb) ± 10% |
| Type of installation | | Same as predicate device | Wall or floor mount |
| Patient position | | Same as predicate device | Standing / Wheelchair |
| Applicable Standards | | Same as predicate device | IEC 60601-1IEC 60601-1-3IEC 60601-2-63IEC 60601-1-2 |
{9}
510(k) Submission- RCT700
The product is principally just the same as in the previous 510(k) #K213226.
The table 1 provides the summary of the technological characteristics of RCT700 compared to the predicate device.
The complete of differences of the subject device to the predicate device is as follows
- Detector (Added a detector of scan ceph type)
- Software version upgrade
The software of RCT700 saves the patient and image data and offers an inquiry function, in addition, supports the image generate function intended to obtain images using the RCT700 equipment and various sensors for diagnosis. And the structure and function of the software is the same as that of the predicate device(K213226).
The 510(k) for the existing detector used in our equipment is provided below.
| Modality | Manufacturer | Detector Model | Cleared | 510(k) No. |
| --- | --- | --- | --- | --- |
| CT | Iray | Jupi0606X1 | No PMA | K213226 |
| CT | Viewworks | FXDD-0606CA | No PMA | K181452 |
| Pano | Viewworks | FXDD-0606CA | No PMA | K181452 |
| Pano | Iray | Jupi0606X1 | No PMA | K213226 |
| Scan Ceph | i3system | XID-C24DC | No PMA | K181452 |
| Scan Ceph | Iray | Pluto0900X | No PMA | K211159 |
| One shot Ceph | Viewworks | FXRD-1717VA | No PMA | K181003 |
| One shot Ceph | Viewworks | FXDD-1012CA | No PMA | K182805 |
{10}
510(k) Submission- RCT700
# 10. Safety and Effectiveness Information
RCT700 system described in this 510(k) is similar to the predicate device in terms of indications for use, materials, safety characteristics, and X-ray source.
The following information further substantiates the substantial equivalence between the subject device and predicate device. The fundamental technological characteristics of the subject and predicate device are similar. The imaging modes are similar; PANO, CEPH (Optional), CBCT. The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the performance of the subject and the predicate device for comparison. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.
The complete of differences of the subject device to the predicate device is as follows
- Detector (Added a detector of scan ceph type)
- Software version upgrade
Electrical, mechanical and environmental safety testing according to the standard of IEC 60601-1:2005/AMD1:2020 (3.2 Edition), IEC 60601-1-3:2008/AMD1:2013/AMD2:2021 (2.2 Edition), IEC 60601-1-6:2010/AMD1:2013/AMD2:2020 (3.2 Edition), and IEC 60601-2-63:2012/AMD1:2017/AMD2:2021 (1.2 Edition) were performed.
EMC testing was conducted in accordance with the standard IEC 60601-1-2:2014+AMD1:2020 (Edition 4.1).
The software of RCT700 saves patient and image data and offers an inquiry function. In addition, it supports the image generate function intended to obtain images using the RCT700 equipment and various sensors for diagnosis. That has been validated according to the FDA Guidance for the "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" to assure substantial equivalence. The software for this device was considered a "Basic Documentation Level" of concern since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.
As a result, we identified the level of concern associated with a new device and provided documentation consistent with that level. Based on our risk analysis of software, the difference does not affect its safety and effectiveness.
Bench testing was conducted according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices". Bench testing is used to assess whether the parameters required to describe functionalities related to imaging properties of the dental X-ray device and patient dosage satisfy the designated tolerance.
Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4 and IEC 61223-3-7. All test results were satisfactory.
Non-clinical considerations were conducted in accordance with FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices". Because the subject device used the same detector as the predicate device, there is no significant difference between the two devices as a result of non-clinical testing.
Clinical considerations were conducted according to the FDA Guidance "Electronic Submission Template for Medical Device 510(k) Submissions: Guidance for Industry and Food and Drug Administration Staff," Section "Performance Testing - Clinical" Clinical
{11}
510(k) Submission- RCT700
images were provided, and they provide further evidence, in addition to the laboratory performance data, to show that the complete system works as intended.
The features of RCT700 were clinically tested and approved by one licensed practitioners/clinicians. Clinical imaging samples were collected from new detectors on the proposed device at the two offices where the predicate device was installed for the clinical test images. These images were gathered from all detectors installed with RCT700 using protocols with random patient age, gender, and size. A licensed practitioner reviewed the sample clinical images and deemed them to be of acceptable quality for the intended use.
## 11. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Ray Co., Ltd. concludes that the new RCT700 is safe, effective and substantially equivalent to the predicate device as described herein.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.