K Number
K182805

Validate with FDA (Live)

Device Name
RCT800
Manufacturer
Date Cleared
2018-12-14

(73 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RCT800 is CBCT and panoramic x-ray imaging system with cephalometric.

Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option.

Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified heath care professionals.

Device Description

RCT800 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle.

Panoramic image scanning function for attaining image of whole teeth, cephalometric scanning option for attaining cephalic image, and Model Scan option for attaining dental model CBCT image are included.

AI/ML Overview

The provided text does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria in terms of diagnostic accuracy or effect size with AI assistance. The document is a 510(k) summary for the RCT800, a dental CBCT and panoramic x-ray system, comparing it to a predicate device.

The document discusses safety and effectiveness information and performance (imaging performance) testing according to certain standards, but these tests are primarily related to the physical and technical specifications of the imaging system and its components (like X-ray generation, detectors, electrical safety, EMC, software validation, etc.), rather than an evaluation of its diagnostic capabilities or comparative effectiveness with human readers.

Here's an analysis of the provided sections based on your request, highlighting the missing information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document lists various parameters of the proposed device and compares them to the predicate device. However, these are technical specifications (e.g., detector type, FOV, X-ray voltage, pixel size, scan time) and not diagnostic acceptance criteria (e.g., sensitivity, specificity for a particular condition).

ParameterAcceptance Criteria (Not Explicitly Stated)Reported Device Performance
Indications for UseSame as predicate device #1RCT800 is CBCT and panoramic x-ray imaging system with cephalometric. Intended for radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. Model scan option. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. Operated by dentists or other legally qualified healthcare professionals. (Same as predicate)
Performance (Imaging Performance)Satisfies designated tolerance for parameters required for imaging properties and patient dosage."All test results were satisfactory." (General statement, no specific diagnostic performance metrics)
FXDD-1012CHA Detector PerformanceSimilar to FXDD-0606CA (predicate device's detector)"Based on Non-Clinical Test results of FXDD-1012CHA for the subject device, is similar to that of the FXDD-0606CA for the Predicate device. All test results were satisfactory."
Electrical, Mechanical, Environmental SafetyCompliance with IEC 60601 series and IEC 60601-1-2 (EMC).Tests performed, results satisfactory.
Software ValidationCompliance with FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".Documentation provided, level of concern "moderate" identified, difference does not affect safety/effectiveness.
Bench TestingSatisfy designated tolerance for parameters describing functionalities related to imaging properties and patient dosage."All test results were satisfactory."
Non-clinical ConsiderationsCompliance with FDA Guidance "submissions of 510(k)'s for Solid State X-ray Imaging Devices".Conducted, results satisfactory.

2. Sample Size Used for the Test Set and Data Provenance:

The document mentions "Clinical imaging samples are collected from the new detector on propose device at the 2 offices where the predicate device is installed". It also states these images were "gathered from the new detector installed with RCT800 on any protocols with random patient age, gender, and size."

  • Sample Size: The exact number of clinical imaging samples or "patients" is not specified.
  • Data Provenance: The images were collected at "2 offices," implying real-world clinical settings. The location (country) is not specified, but the applicant and manufacturer are from Korea. The data collection seems to be prospective in the sense that the images were taken with the new detector for evaluation, but it's not a formal prospective clinical trial for diagnostic performance. It's described as "Clinical considerations were conducted according to FDA Guidance 'Format for Traditional and Abbreviated 510(k)s, section 20'".

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • Number of Experts: "two licensed practitioners/clinicians observed and verified that dental X ray system from RCT800."
  • Qualifications: "licensed practitioners/clinicians." No specific details on their specialty (e.g., radiologist, dentist, orthodontist) or years of experience are provided.

4. Adjudication Method for the Test Set:

Not applicable or not described in the context of diagnostic performance. The two practitioners "observed and verified" the images, but there's no mention of a formal adjudication process (e.g., 2+1, 3+1 consensus) for establishing ground truth for diagnostic accuracy.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done:

No, an MRMC comparative effectiveness study was not done. The document does not describe any study comparing human readers with AI assistance versus without AI assistance. The RCT800 is an imaging system, and the filing is for its substantial equivalence to a predicate imaging system, not for an AI diagnostic aid.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

Not applicable. The RCT800 is an imaging device, not an AI algorithm. Its performance is as an imaging system.

7. The Type of Ground Truth Used:

For the "clinical imaging samples," the ground truth seems to be implicitly expert diagnosis/interpretation. The document states, "As licensed practitioners or clinician diagnoses of the images, it might be proved that the clinical diagnosis and structures are acceptable in the region of interests." This suggests that the practitioners visually assessed the images for diagnostic acceptability and structural clarity, rather than comparing them to a definitive pathological diagnosis or outcomes data.

8. The Sample Size for the Training Set:

Not applicable for this type of device and submission. The RCT800 is a hardware imaging system, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the right side features the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 14, 2018

RAY Co., Ltd Changhwan Lee RA Manager 332-7, Samsung 1-ro Hwaseong-si, Gyeonggi-do 18380 KOREA

Re: K182805

Trade/Device Name: RCT800 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: September 27, 2018 Received: October 2, 2018

Dear Changhwan Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hsl 2. Nils

for Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

4. Indications for Use Statement

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K182805

Device Name RCT800

Indications for Use (Describe)

RCT800 is CBCT and panoramic x-ray imaging system with cephalometric.

Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option.

Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified heath care professionals.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

See PRA Statement below.

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

5. 510(K) Summary

{5}------------------------------------------------

510(k) Summary

- K182805 1. 510(k) Summary

The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

    1. Date: September 27, 2018

3. Administrative Information

APPLICANTRAY Co.,Ltd
ADDRESS#332-7, Samsung 1-ro, Hwaseong-si, Gyeonggi-do, 18380, Korea
ManufacturerRAY Co.,Ltd
332-7, Samsung 1-ro, Hwaseong-si, Gyeonggi-do, 18380, Korea
TEL : +82-31-605-1000FAX : +82-2-6280-5534
Contact Persone-mail : ch0406.lee@raymedical.co.kr

4. Device Information

Device Name

Trade/Proprietary Name: RCT800 Common Name: Dental panoramic/tomography and cephalometric x-ray system

Classification

Classification Name: Computed tomography x-ray system Regulation Number : 21 CFR 892.1750 Class : Il Product code : OAS Panel : Radiology

5. Predicate device

ParameterPredicate Device-1
Device NameRCT800
ManufacturerRAY Co., Ltd

{6}------------------------------------------------

510(K) NumberK181452Traditional 510k
Classification nameComputed tomography x-ray system
Regulation number892.1750
Primary product codeOAS

7. Device Description

System purpose RCT800 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle.

Panoramic image scanning function for attaining image of whole teeth, cephalometric scanning option for attaining cephalic image, and Model Scan option for attaining dental model CBCT image are included.

8. Indication for use

RCT800 is CBCT and panoramic x-ray imaging system with cephalometric.

Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified heath care professionals.

9. Patient population

The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure.

There is no restriction for ethnic group, Gender, weight, health, or condition.

We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.

10. Operating principle

The purpose of this unit is diagnose clinical structures of a tooth and head areas mainly by using the characteristics of permeability from X-ray. The principle of formations are as following. The machine is made of X-ray generator and arms in which transfers X-ray signals to a sensor in 2D. Also, an object that has a magnification is required in a distance. Moreover, the unit has to be adjustable depending on height of a patient and PC system to reconstruct an image.

The arm parts are controlled for rotating and linear moving to synchronize between the sensor and X-

{7}------------------------------------------------

ray generator to get the image of interests. The purpose of this mechanism is to provide the images in 2D or/and 3D as preferred to diagnose in a monitor. CBCT provides in 3D images as reconstructed and Panorama is to diagnose the structures in a panoramic view. Cephalometric allows for orthodontic treatment. These 3 functions could be in 1 system, Panorama with Cephalometric, or Panorama only system depending on the needs. To provide the features as mentioned above, digital transferring from permeated X-ray to absorbing to the sensor is essential and all the process are proceed in Detector. Detector transfers X-ray to light depending on the structure materials. Detector is separated into indirect method that the light is changed to digital signals on photodiode and direct method in which the light is directly transferred to digital signal. This unit is using both direct and indirect method depending on the interior structure materials.

{8}------------------------------------------------

11. Comparison with predicate device

The product is principally just the same as in the previous 510(k) #K181452. We added a CT/Pano detector to the predicate Device #K181452. And, CT protocol and FOV size have changed slightly due to the added detector.

The compared technical features for imaging technology, FOV, imaging parameters, resolution, and other basic characteristics are matching very closely, and the differences are so small that they do not have any effect on performance in practice. Both devices conform to given international performance standards.

ParameterProposed DevicePredicate Device
ManufacturerRAY Co., Ltd.RAY Co., Ltd.
Device nameRCT800RCT800
510(K) Number(Special 510K)K181452(Traditional 510K)
Common NameDental panoramic/tomography andcephalometric x-ray systemDental panoramic/tomography andcephalometric x-ray system
Indicationsfor useSame as predicate device #1RCT800 is CBCT and panoramic x-rayimaging system with cephalometric.Which is intended to radiographicexamination of the dento-maxillofacial,sinus, TMJ, Airway for diagnosticsupport for adult and pediatric patients.And a model scan is included as anoption.Cephalometric image is also includeswrist to obtain carpus images for growthand maturity assessment for orthodontictreatment.The device is to be operated and usedby dentists or other legally qualifiedheath care professionals.
Mode ofOperationSame as predicate device #1Continuous operation withintermittent, stated permissibleloading
3D technologySame as predicate device #1CBCT Cone beamComputed Tomography
PerformanceSpecificationSame as predicate device #11) CBCT Computed tomography- Patient- Dental Model Scan(Optional)2) Panoramic3) Cephalometric(optional)- One shot type- Scan type
FunctionalOptionSame as predicate device #1BaseCT+PANOOption(CEPH)CT + PANO + SCAN CEPHCT + PANO + One shot(One shot,Standard Type)CT + PANO + One shot(One shot,Large Type)
DetectorTypeCTSame as predicate device #1FXDD-0606CA
FXDD-1012CHAN/A
PANOSame as predicate device #1FXDD-0606CA
FXDD-1012CHAN/A
Ceph (Scan)Same as predicate device #1XID-C24DC
Ceph (One shot)Same as predicate device #11717SCC
Same as predicate device #1PaxScan 2530C
Exposure switchTypeSame as predicate device #1"Deadman" Button type
Same as predicate device #1Ceph Apparatus
Same as predicate device #1Vertical Carriage
Same as predicate device #1Rotator
Same as predicate device #1X-RAY Generator
Same as predicate device #1X-ray tube
Same as predicate device #1High Frequency Generator
MainComponentsSame as predicate device #1Column
Same as predicate device #1Touch monitor (panel)
Detector- CTFXDD-0606CAFXDD-1012CHADetector- CTFXDD-0606CA
- PANOFXDD-0606CAFXDD-1012CHA- PANOFXDD-0606CA
- CephXID-C24DC(Scan)1717SCC(One shot, Large Size)PaxScan 2530C(One shot, StandardSize)- CephXID-C24DC(Scan)1717SCC(One shot, Large Size)PaxScan 2530C(One shot, StandardSize)
Same as predicate device #1Chinrest
Same as predicate device #1Head rest
Same as predicate device #1Automatic Collimator
Same as predicate device #1Exposure switch
Same as predicate device #1Emergency stop switch
Same as predicate device #1Console PC set
AutomaticCollimatorSame as predicate device #1CT examsPanoramic examsCephalometric exams
Display TypeSame as predicate device #1TFT LCD type(Normally black)*1280x800 pixel
ClassSame as predicate device #1Class I with type B applied partsaccording to IEC 60601-1
Focal sizeSame as predicate device #1Patient 0.5Model scan 0.04 (Optional)
Field ofView(CT)FXDD-0606CA : Max.160x100 mmFXDD-1012CHA : Max. 200x200 mmFXDD-0606CA : Max.160x100 mm
X-ray Voltage(Patient)Same as predicate device #160~90kVp
X-ray Current(Patient)Same as predicate device #14~17mA
X-ray Voltage(Model Scan,Optional)Same as predicate device #150~80kVp
X-ray Current(Model Scan,Optional)Same as predicate device #10.4~0.7mA
Total FiltrationSame as predicate device #1Min. 2.8 mm Al equivalent
DetectorPixel sizeCTSame as predicate device #1FXDD-0606CA: 119µm
FXDD-1012CHA: 124µmN/A
PANOSame as predicate device #1FXDD-0606CA: 119µm
FXDD-1012CHA: 124µmN/A
Ceph (Scan)Same as predicate device #1XID-C24DC: 100µm
Ceph(One shot)Same as predicate device #11717SCC: 127µm
Same as predicate device #1PaxScan 2530C: 139µm
MagnificationCTSame as predicate device #1FXDD-0606CA: 1.44(Patient)1.91(Model Scan)
FXDD-1012CHA: 1.44(Patient)1.91(Model Scan)N/A
PANOSame as predicate device #1FXDD-0606CA: 1.3
FXDD-1012CHA: 1.3N/A
Ceph (Scan)Same as predicate device #1XID-C24DC: 1.11
Ceph(One shot)Same as predicate device #11717SCC: 1.13
Same as predicate device #1PaxScan 2530C: 1.12
Scan timeSame as predicate device #1CT : below 14sec(Patient)CT : below 180sec(Model Scan)
Same as predicate device #1Pano : below 14sec
Same as predicate device #1Ceph[Scan type] : below 19sec
Same as predicate device #1Ceph[One shot type]:below 2sec
Format compatibleSame as predicate device #1DICOM 3.0 Format compatible
Image Viewing SoftwareSame as predicate device #1RayScan(Cleared under K181452)
Image acquisitionSame as predicate device #1Giga-Ethernet Network
Total HeightSame as predicate device #1Max 2.296mm
WeightSame as predicate device #11) Computed Tomography(CT) +Panoramic(PANO)=189kg(416.6lb) ±10%2) Computed Tomography(CT) +Panoramic(PANO) + Ceph(Scan type)= 219kg (482.8lb) ± 10%3) Computed Tomography(CT) +Panoramic(PANO) + Ceph(One shot type, installed in Standardsize)= 217kg (478.4lb) ± 10%4) Computed Tomography(CT) +Panoramic(PANO) + Ceph(One shot type, installed in Large size)212kg (467.3lb) ± 10%
Type ofinstallationSame as predicate device #1Wall or floor mount
Patient positionSame as predicate device #1Standing / Wheelchair
ApplicableStandardsSame as predicate device #1IEC 60601-1IEC 60601-1-3IEC 60601-2-63IEC 60601-1-2

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

12. Safety and Effectiveness Information

RCT800 system described in this 510(k) is similar to the predicate device in terms of indications for use, materials, safety characteristics, and X-ray source.

The following information further substantial equivalence between the subject device and predicate device.

The fundamental technological characteristics of the subject and predicate device are similar.

The imaging modes are similar; PANO, CEPH (Optional), CBCT, Model Scan All viewing software programs have been cleared with previous 510k submissions; RAYSCAN(K181452).

The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the performance of the subject and the predicate device for comparison. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.

The differences are as follows.

  • The subject device is equipped with new CT/PANO detectors, FXDD-1012CHA.

Electrical, mechanical and environmental safety testing according to standard of IEC 60601-1: 2005/AMD1:2012(3.1 Edition), IEC 60601-1-3: 2008/AMD1:2013(Second Edition), IEC 60601-1-6:2010(Third Edition) and IEC 60601-2-63: 2012(first Edition) were performed. EMC testing was conducted in accordance with the standard IEC 60601-1-2: 2014(Edition 4.0).

The software of RCT800 saves the patient and image data and offers an inquiry function, in addition, supports the image generate function intended to obtain images using the RCT800 equipment and various sensors for diagnosis. And that has been validated according to FDA "Guidance for the Content d Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" to assure substantial equivalence. The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator. As a result, we identified the level of concern associated with new device and provided documentation consistent with that level. Based on our risk analysis of software, the difference does not affect its safety and effectiveness.

Bench testing was conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, section 18, Performance Testing - Bench"

Bench testing is used to assess whether or not the parameter measured required for describing functionalities related to imaging properties of the dental X-ray device and patient dosage satisfies the designated tolerance.

Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4 and IEC 61223-3-5.

All test results were satisfactory.

Non-clinical considerations were conducted in accordance with FDA Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices".

RCT800 is equipped with FXDD-1012CHA. FXDD-1012CHA is a new SSXI detector, which is used to capture an image in panoramic, CBCT and Model Scan mode.

Based on Non-Clinical Test results of FXDD-1012CHA for the subject device, is similar to that of the FXDD-0606CA for the Predicate device.

All test results were satisfactory.

Clinical considerations were conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, section 20".

{13}------------------------------------------------

Clinical images were provided these images were not necessary to establish substantial equivalence based on the modifications to the device but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended.

For clinical testing, two licensed practitioners/clinicians observed and verified that dental X ray system from RCT800.

The clinical imaging samples are collected from the new detector on propose device at the 2 offices where the predicate device is installed on clinical consideration report for the clinical test images. These images were gathered from the new detector installed with RCT800 on any protocols with random patient age, gender, and size. As licensed practitioners or clinician diagnoses of the images, it might be proved that the clinical diagnosis and structures are acceptable in the region of interests.

12. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. RAY Co., Ltd. concludes that the newly RCT800 is safe and effective and substantially equivalent to predicate device as described herein.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.