K193334 · GE Healthcare · MUE · May 15, 2020 · Radiology
Device Facts
Record ID
K193334
Device Name
Pristina Serena Bright
Applicant
GE Healthcare
Product Code
MUE · Radiology
Decision Date
May 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1715
Device Class
Class 2
Indications for Use
Pristina Serena Bright is an optional accessory of Senographe Pristina intended to provide accurate location of lesions in the breast in three dimensions.
Device Story
Pristina Serena Bright is a software-based biopsy guidance option for the Senographe Pristina mammography platform. It utilizes Contrast Enhanced Spectral Mammography (CESM) to identify suspicious lesions not clearly visible on standard mammography or ultrasound. The system acquires stereotactic pairs of 2D X-ray images under breast compression to calculate 3D (X, Y, Z) coordinates of target lesions. The device uses a motorized biopsy holder, compatible with vertical or horizontal needle approaches, to guide biopsy needles or localization wires to the target. Operated by radiologists in clinical settings, the system provides visual guidance for needle placement. By enabling contrast-enhanced targeting, it allows for earlier cancer detection and treatment initiation compared to standard imaging, while offering a less invasive, faster, and more comfortable alternative to MRI-guided biopsy for patients with claustrophobia or metallic implants.
Clinical Evidence
Bench testing only. No clinical data. Verification included image quality and dose testing demonstrating performance equivalent to SenoBright HD. Biopsy accuracy testing verified geometric alignment between target lesions on X-ray stereo pairs and actual needle tip/notch position. Compatibility testing confirmed the biopsy application functions within the timeframe of contrast agent visibility.
Technological Characteristics
Software-based option for Senographe Pristina platform. Uses CESM imaging chain. Motorized biopsy device holder supports vertical and horizontal needle approaches. Connectivity: Integrated into existing Senographe Pristina system. No hardware modifications required. Developed under ISO 13485 and 21 CFR 820.
Indications for Use
Indicated for patients with suspicious lesions only seen with certainty when imaged with a contrast agent or that do not have a definite correlate on mammography or ultrasound. Provides 3D location of target lesions using stereotactic pairs of 2D X-ray images acquired with Contrast Enhanced Spectral Mammography (CESM) under breast compression. Guides minimally invasive/interventional breast procedures: vacuum-assisted biopsy, core biopsy, pre-surgical localization (e.g., hookwire), and fine needle aspirations (FNA).
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
K182951 — Pristina Serena 3D · GE Healthcare · Jan 18, 2019
K173576 — Pristina Serena · GE Healthcare · May 14, 2018
K260087 — Senographe Pristina; SenoBright HD; Pristina Serena; Pristina Serena Bright · GE Medical Systems SCS · Mar 24, 2026
K211215 — SenoBright HD · GE Healthcare · Jun 22, 2021
K172404 — SenoBright HD · GE Healthcare · Oct 30, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
May 15, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Healthcare % Barthelemy Arman Regulatory Affairs Leader 283 rue de la Miniere 78530 Buc FRANCE
Re: K193334
Trade/Device Name: Pristina Serena Bright Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: Class II Product Code: MUE Dated: April 10, 2020 Received: April 17, 2020
Dear Barthelemy Arman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting
{1}------------------------------------------------
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K193334
Device Name Pristina Serena Bright
#### Indications for Use (Describe)
The Pristina Serena Bright option provides the three-dimensional location of target lesions, using information obtained from stereotactic pairs of two-dimensional X-ray images acquired with Contrast Enhanced Spectral Mammography (CESM) under the same breast compression. This information provides guidance for a variety of minimally invasive or interventional procedures in the breast such as: vacuum assisted biopsy, pre-surgical localization (e.g. hookwire), and fine needle aspirations (FNA).
CESM-Biopsy application is indicated for patients with suspicious lesions only seen with certainty when imaged with a contrast agent or that do not have a definite correlate on mammography or ultrasound.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Premarket Notification Submission
Image /page/3/Picture/2 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by three curved lines on the top and bottom, resembling water droplets or stylized waves. The logo is simple, recognizable, and represents the brand identity of General Electric.
Section 5: 510(k) Summary
PRISTINA SERENA BRIGHT
{4}------------------------------------------------
# GE Healthcare
### 510(k) Premarket Notification Submission
Image /page/4/Picture/2 description: The image shows the GE logo, which is a blue circle with the letters "GE" in white inside. The letters are stylized and connected with swirling lines. Below the logo is the text "K193334" in black.
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | May 13th, 2020 |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare<br>GE Medical Systems SCS<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE |
| Primary Contact Person: | Barthélémy Arman<br>Regulatory Affairs Leader<br>283 rue de la Minière<br>78530 Buc - FRANCE<br>Phone : +33 1 30 70 40 40<br>Email : Barthelemy.ARMAN1@ge.com |
| Secondary Contact Person: | Gregory Pessato,<br>Regulatory Affairs Program Manager,<br>GE Medical Systems SCS<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE<br>Phone : + 33 1 30 70 93 16<br>Email : gregorypessato@ge.com |
| Device Trade Name: | Pristina Serena Bright |
| Common/Usual Name: | Stereotaxy biopsy guidance application using contrast imaging |
| Classification Names:<br>Product Code: | 21 CFR 892.1715, Class II<br>MUE |
| Predicate / Reference<br>Device(s): | Pristina Serena (K173576) - predicate / SenoBright HD (K172404) -<br>reference |
| Device Description: | Pristina Serena Bright is a Biopsy System for Senographe Pristina. It is<br>an additional software option that builds upon the Pristina Serena<br>device (GE Healthcare Stereotaxy biopsy option for Senographe Pristina<br>platform). Pristina Serena was cleared on May 14, 2018 (K173576). |
| | |
| | Pristina Serena Bright enables biopsy medical application to be done<br>using Contrast Enhanced Spectral Mammography images.<br>The Pristina Serena Bright add-on includes the following items:<br>Software: A new software version for the Senographe Pristina platform which includes software to manage the Pristina Serena Bright option. Labeling for the CESM Biopsy Medical application. |
| | Pristina Serena Bright option is compatible with previously installed<br>Senographe Pristina systems. Pristina Serena Bright does not require<br>any hardware modification on the Senographe Pristina platform. The<br>hardware that was cleared on Pristina Serena (K173576) was also not<br>modified. |
| Intended Use: | Pristina Serena Bright is an optional accessory of Senographe Pristina<br>intended to provide accurate location of lesions in the breast in three<br>dimensions. |
| Indications for Use | The Pristina Serena Bright option provides the three-dimensional<br>location of target lesions, using information obtained from stereotactic<br>pairs of two-dimensional X-ray images acquired with Contrast Enhanced<br>Spectral Mammography (CESM) under the same breast compression.<br>This information provides guidance for a variety of minimally invasive or<br>interventional procedures in the breast such as: vacuum assisted<br>biopsy, core biopsy, pre-surgical localization (e.g. hookwire), and fine<br>needle aspirations (FNA).<br>CESM-Biopsy application is indicated for patients with suspicious lesions<br>only seen with certainty when imaged with a contrast agent or that do<br>not have a definite correlate on mammography or ultrasound. |
| Technology: | Pristina Serena Bright reuses the Stereotaxy principle of the cleared<br>device Pristina Serena with Contrast Enhanced Spectral Mammography<br>(CESM) images instead of 2D images to determine the three-<br>dimensional (3D) location (X, Y and Z coordinates) of an object of<br>interest in the breast (such as a suspicious lesion).<br><br>With Pristina Serena Bright, the stereotaxy biopsy process uses a stereo<br> |
| | |
| | breast at the exact target position (biopsy of sample tissue or<br>placement of a hook wire for guidance of surgical interventions). With<br><b>Pristina Serena Bright</b> reuses the positioning of the biopsy device<br>holder from <b>Pristina Serena</b> which is motorized and takes into account<br>the geometry of the biopsy needles, so when the biopsy device holder is<br>in place, the user introduces the needle in the breast until reaching the<br>mechanical stop of the biopsy device holder. As the biopsy needle is<br>fixed on and guided by the biopsy device holder (not handed), the<br>needle tip (or notch) will then be at the target lesion 3D coordinates. |
| | As it was for <b>Pristina Serena</b> option, the <b>Pristina Serena Bright</b> allows<br>two different needle approaches for Biopsy: vertical and horizontal (left<br>or right). |
| | In vertical approach, the needle is introduced from the "top" of the compressed breast. In horizontal approach, the needle is introduced from the side of the compressed breast. Depending on patient morphology and location of the lesion to be<br>biopsied in the breast, a radiologist might choose an approach or the<br>other. Usually an effort is made to use an approach that would limit the<br>distance crossed by the needle in the breast to reach the target<br>location. |
| | <b>Pristina Serena Bright</b> option uses the Image chain of the Senographe<br><b>Pristina</b> platform with <b>SenoBright HD</b> (CESM application) thus the image<br>quality of images during the biopsy procedure is equivalent to that of<br><b>SenoBright HD</b> . |
| Substantial Equivalence /<br>Predicate and Reference<br>Devices | The <b>Pristina Serena Bright</b> has identical intended use and substantially<br>equivalent indications for use to its legally marketed predicate device,<br>the <b>Pristina Serena</b> .<br>The functionalities, specifications and technological characteristics of<br><b>Pristina Serena Bright</b> option are identical or equivalent to <b>Pristina</b><br><b>Serena</b> .<br>The principles of operation, image quality and dose characteristics and<br>performances of the <b>Pristina Serena Bright</b> are equivalent to those of<br>the reference device <b>SenoBright HD</b> .<br>The <b>Pristina Serena Bright</b> was designed to provide an image quality<br>performance equivalent to <b>SenoBright HD</b> and the same accuracy as<br><b>Pristina Serena</b> as demonstrated in Image Quality and Accuracy<br>Performance Testing. |
| Benefit-Risk Analysis: | The proposed Pristina Serena Bright device offers the same benefits<br>and risk as the predicate device Pristina Serena except for the following:<br>Improved benefit by allowing to perform breast biopsy for<br>patients with suspicious findings only clearly seen when imaged<br>with a contrast agent or that do not have a definite correlate on<br>mammography or ultrasound, meaning potential earlier<br>detection of breast cancer and earlier initiation of treatment. Increased risk due to the use of contrast agent injection. The<br>same contrast injection protocols are used in CESM and CESM-<br>guided Biopsy, so this risk is identical to the Reference device<br>SenoBright HD for CESM diagnostic imaging. As compared to the alternative standard of care MRI-guided Biopsy,<br>Pristina Serena Bright might offer the following benefits:<br>From a workflow and access perspective: potential benefit with<br>regards to reduced cost and faster access. From a patient perspective: potentially improved patient<br>comfort for patients with claustrophobia and allows to handle<br>patients that cannot be imaged with MRI due to the presence<br>of metallic objects in their body and is a shorter procedure. Regarding the other alternative to CESM-guided Biopsy, i.e. surgical<br>biopsy, it presents a higher risk profile as it is a more invasive<br>procedure, it has a longer recovery time and it has a higher risk of<br>infection and bruising. Besides it might remain difficult for the surgeon<br>to localize the appropriate area to be biopsied as no biopsy hook/wire<br>would have been able to be placed in the breast prior to surgery. |
| | Overall GE Healthcare believes that as the increased risk is accompanied<br>by an increase in benefit, the proposed device has a comparable<br>benefit-risk profile to the predicate device. |
| Determination of Substantial<br>Equivalence: | Pristina Serena Bright has successfully completed required design<br>control testing per GE Healthcare's quality management system. No<br>unexpected test results were obtained. This new device was designed<br>and will be manufactured under the Quality System Regulations of<br>21CFR 820 and ISO 13485. The following quality assurance measures<br>were applied to the development of the system:<br>Risk Analysis Design Reviews Software Development Lifecycle Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) |
| | - Safety testing (Verification)<br>- Simulated use testing (Validation) |
| | The safety and performance of the <b>Pristina Serena Bright</b> option was<br>demonstrated through full verification testing and additional<br>engineering bench performance testing such as:<br>- Non-Clinical Data – Image Quality and Dose test that<br>demonstrates that images acquired with <b>Pristina Serena <br/>Bright</b> are of same quality as images acquired with<br>Senographe Pristina with SenoBright HD at similar dose<br>levels.<br>- Non-Clinical Data – Biopsy accuracy testing: verification of<br>the geometrical accuracy between the target lesion<br>identified on the X-ray Stereo pair images and the actual<br>position of the biopsy needle tip (or needle notch).<br>- Non-clinical Data – Study that demonstrates the<br>compatibility of the biopsy application on the Pristina<br>Serena platform with the timeframe of the contrast agent<br>visibility in the breast.<br>These tests were performed to provide the requisite data to<br>substantiate performance and substantial equivalence. The testing<br>demonstrated that <b>Pristina Serena Bright</b> performs according to<br>specifications and functions as intended. |
| Conclusion: | Based on: conformance to standards; development under GE<br>Healthcare's quality management system and design controls; benefit-<br>risk analysis, successful verification/validation testing and additional<br>bench performance testing and clinical testing, GE Healthcare believes<br>that the <b>Pristina Serena Bright</b> option is substantially equivalent to its<br>predicate device Pristina Serena (K173576) and reference device<br>SenoBright HD (K162268). Therefore, GE concludes that <b>Pristina Serena <br/>Bright</b> is as safe and effective for its intended use as its predicate<br>device. |
{5}------------------------------------------------
# GE Healthcare
______________________________________________________________________________________________________________________________________________________________________________
Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has decorative swirls around its perimeter, giving it a dynamic and recognizable appearance.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are water droplet-like shapes around the edge of the circle. The logo is simple and recognizable.
{7}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
Image /page/7/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular frame. The frame has decorative flourishes resembling water droplets or stylized leaves. The logo is presented in a solid blue color, providing a clean and recognizable brand representation.
{8}------------------------------------------------
# GE Healthcare
______________________________________________________________________________________________________________________________________________________________________________
Image /page/8/Picture/1 description: The image shows the logo of General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has a white border and four white stylized water droplets around the letters.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.