K Number
K211215

Validate with FDA (Live)

Device Name
SenoBright HD
Manufacturer
Date Cleared
2021-06-22

(60 days)

Product Code
Regulation Number
892.1715
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SenoBright HD is an extension of the existing indication for diagnostic mammography with Senographe Pristina. The SenoBright HD application shall enable contrast enhanced breast imaging a dual energy technique. This imaging technique can be used as an adjunct following mammography and ultrasound exams to help localize a known or suspected lesion.

Device Description

This submission is proposing a software update to SenoBright HD consisting of an improvement of the existing "recombination" algorithm with the New Image Recombination Algorithm (NIRA) by adding a local estimation of breast thickness in the images recombination to account for the non-uniformity of the breast thickness, and by compensating for potential patient movement between the 2 CESM acquisitions (Low Energy and High Energy).

SenoBright HD (K172404) is the name of Senographe Pristina FFDM system allowing to perform Contrast Enhanced Spectral Mammography (CESM) application.

The CESM acquisition technique consists in acquiring two images (one High Energy and one Low Energy) in sequence and under the same breast compression after patient injection with an iodinated contrast media. The two images are then recombined through a post-processing algorithm.

This design change is a software and labeling only option, compatible with SenoBright HD installed base and does not require any hardware modification on the Senographe Pristina platform.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the SenoBright HD device with the New Image Recombination Algorithm (NIRA), based on the provided document:

1. Table of Acceptance Criteria & Reported Device Performance

The acceptance criteria are not explicitly stated in a quantitative manner (e.g., "must achieve X% sensitivity"). Instead, the document focuses on demonstrating non-inferiority or improvement compared to the predicate device (SenoBright HD K172404) and showing clinical acceptability. The performance is reported in terms of assessments made by radiologists.

Acceptance Criterion (Implicit)Reported Device Performance (SenoBright HD with NIRA)
Equivalence/Improvement in Contrast Uptake VisibilityAssessed equivalent or better in 97% of the cases.
Reduction/Equivalence in Artifacts VisibilityAssessed equivalent or lower in 99% of the cases.
Improvement in Overall Clinical Image QualityAssessed superior in more than 98% of the cases.
Clinical Image Acceptability (for NIRA images)Illustrated through objective criteria defined and evaluated by radiologists. (Specific quantitative results for this are not provided, only that it was "illustrated").
Non-clinical Performance (Phantoms)Demonstrated that SenoBright HD with NIRA performs at least as well as the cleared device (K172404) and brings Image Quality improvements. Demonstrated reduction of artifacts in case of patient motion or breast thickness non-uniformity to increased lesion visibility. (Specific quantitative results for phantom studies are not provided in this summary).
Safety and EffectivenessDemonstrated through full verification testing, additional performance testing, and clinical image evaluations. Concluded that NIRA for SenoBright HD is substantially equivalent to the predicate device and raises no new questions of safety and effectiveness.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Clinical Image Evaluation (Initial Acceptability): 10 images
    • Clinical Image Evaluation (Comparative): 50 clinical images
  • Data Provenance: The document does not explicitly state the country of origin for the clinical data used in the studies. Given GE Healthcare's presence in France (as per the submitter's address), it is plausible the data could originate from there or other international sites. The data is described as "clinical images," implying retrospective data from a clinical setting. It is not stated whether it was prospective or retrospective, but the description "clinical images" often implies retrospective collection for such comparative studies.

3. Number of Experts and Qualifications

  • Number of Experts: 3 independent MQSA-qualified radiologists used for both clinical image evaluations.
  • Qualifications: "MQSA-qualified radiologists." MQSA (Mammography Quality Standards Act) qualification indicates that these radiologists meet specific federal standards for interpreting mammograms in the United States, including training, experience, and continuing education requirements. The specific years of experience for each expert are not provided.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method (like 2+1, 3+1). It states that the evaluations were "performed by 3 independent MQSA-qualified radiologists." This suggests that each radiologist independently assessed the images, and the reported percentages (97%, 99%, 98%) likely represent the proportion of cases where at least two out of three (or possibly all three) agreed on the assessment, or an aggregatetion of individual assessments, but the specific consensus/adjudication rule is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size

  • A comparative effectiveness study was performed as a form of reader study, but it wasn't a "human readers improve with AI vs without AI assistance" MRMC study in the sense of AI assisting the human reader.
  • Instead, this study compared the quality of images generated by the new algorithm (NIRA) against the quality of images generated by the previous algorithm (Predicate SenoBright HD), both assessed by human readers.
  • Therefore, there's no direct "effect size of how much human readers improve with AI vs without AI assistance" as the AI (NIRA) is the image generation method being evaluated, not an assistance tool for the human reader's diagnostic performance. The human readers are evaluating the output of the AI.

6. Standalone Performance (Algorithm Only)

The document primarily focuses on the output quality of the algorithm as perceived by human readers, rather than a quantifiable "standalone" diagnostic performance (e.g., sensitivity/specificity for detecting lesions). The phantom testing is a form of standalone performance evaluation for image quality metrics, but not for diagnostic accuracy in a clinical context.

7. Type of Ground Truth Used

The ground truth for the clinical image evaluations was effectively the expert consensus/assessment of the image quality metrics (contrast uptake visibility, artifact visibility, overall image quality) by the 3 MQSA-qualified radiologists. There is no mention of pathology or outcomes data being used as the clinical ground truth for lesion presence/absence for diagnostic performance evaluation, as the study was focused on image quality assessment.

8. Sample Size for the Training Set

The document does not provide any information regarding the sample size used for training the New Image Recombination Algorithm (NIRA). This is typical for premarket notifications where the focus is on verification and validation of changes rather than the internal development details of the algorithm itself.

9. How Ground Truth for Training Set was Established

The document does not provide any information on how the ground truth for the training set (if any was used for supervised learning of NIRA) was established. As NIRA is described as an "evolution" of an existing recombination algorithm, it might involve engineering improvements rather than a machine learning model that requires a labeled training set in the typical sense. It states NIRA accounts for "non-uniformity of the breast thickness" and compensates for "potential patient movement," suggesting algorithmic improvements based on physical principles and image characteristics rather than purely data-driven supervised learning.

{0}------------------------------------------------

June 22, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

GE Healthcare % Florian Akpakpa Regulatory Affairs Program Manager 283 Rue de la Minière Buc, 78530 FRANCE

Re: K211215

Trade/Device Name: SenoBright HD Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: Class II Product Code: MUE Dated: April 22, 2021 Received: April 23, 2021

Dear Florian Akpakpa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211215

Device Name

SenoBright HD

Indications for Use (Describe)

SenoBright HD is an extension of the existing indication for diagnostic mammography with Senographe Pristina. The SenoBright HD application shall enable contrast enhanced breast imaging a dual energy technique. This imaging technique can be used as an adjunct following mammography and ultrasound exams to help localize a known or suspected lesion.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Premarket Notification Submission

Image /page/4/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has a white border and four white teardrop shapes evenly spaced around the perimeter, giving the impression of motion or energy.

Section 5: 510(k) Summary

NIRA for SENOBRIGHT HD

510(k) Number: K211215

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the text "510(k) Premarket Notification Submission". The text is written in a clear, sans-serif font and is horizontally aligned. The text appears to be the title or heading of a document or presentation related to medical device regulation.

Image /page/5/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has small, droplet-like shapes around it, giving it a dynamic and fluid appearance. The logo is presented in a blue color scheme.

510(k) Summary

K211215

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:June 17, 2021
Submitter:GE HealthcareGE Medical Systems SCS283 RUE DE LA MINIERE78530 BUC - FRANCE
Primary Contact Person:Florian AkpakpaRegulatory Affairs Program Manager283 rue de la Minière78530 Buc - FRANCEPhone : +33 1 30 70 40 40Email : florian.akpakpa@ge.com
Secondary Contact Person:Gregory Pessato,Regulatory Affairs Senior ManagerGE Medical Systems SCS283 RUE DE LA MINIERE78530 BUC – FRANCEPhone : +33 1 30 70 40 40Email : gregorypessato@ge.com
Device Trade Name:SenoBright HD
Common/Usual Name:Contrast Enhanced Spectral Mammography
Classification Names:Product Code:21 CFR 892.1715, Class IIMUE
Predicate Device:Predicate device: SenoBright HD (K172404)
Device Description:This submission is proposing a software update to SenoBright HDconsisting of an improvement of the existing "recombination" algorithmwith the New Image Recombination Algorithm (NIRA) by adding a localestimation of breast thickness in the images recombination to accountfor the non-uniformity of the breast thickness, and by compensating forpotential patient movement between the 2 CESM acquisitions (LowEnergy and High Energy).
SenoBright HD (K172404) is the name of Senographe Pristina FFDMsystem allowing to perform Contrast Enhanced Spectral Mammography(CESM) application.
The CESM acquisition technique consists in acquiring two images (oneHigh Energy and one Low Energy) in sequence and under the samebreast compression after patient injection with an iodinated contrastmedia. The two images are then recombined through a post-processingalgorithm.
This design change is a software and labeling only option, compatiblewith SenoBright HD installed base and does not require any hardwaremodification on the Senographe Pristina platform.
Intended Use:SenoBright HD is an extension of the existing indication for diagnosticmammography with Senographe Pristina. The SenoBright HDapplication shall enable contrast enhanced breast imaging using a dualenergy technique. This imaging technique can be used as an adjunctfollowing mammography and ultrasound exams to help localize aknown or suspected lesion.Note: The intended use of SenoBright HD cleared in K172404 is notchanged.
Indications for UseRefer to Intended use.
Technology:The update of the existing recombination algorithm to New ImageRecombination Algorithm (NIRA) for SenoBright HD is completelyindependent from all other parts of the digital mammographicequipment Senographe Pristina. It therefore does not affect the dual-energy image acquisition.
The changes introduced with NIRA are specific to the recombinationalgorithm and can affect exclusively the recombined image. Thus, it alsodoes not affect the acquisition, processing and display of the low-energy image when used as a regular FFDM image.
NIRA is an evolution of the SenoBright HD Recombination algorithmbringing the following improvements:- Use of a local estimation of breast thickness in the imagerecombination to account for the non-uniformity of the breastthickness.- Compensation for potential patient movement between the 2CESM acquisitions (Low Energy and High Energy).
Substantial Equivalence /Predicate DeviceThe update of the existing recombination algorithm to the New ImageRecombination Algorithm (NIRA) for SenoBright HD does not impactthe intended use/indications for use of the legally marketed deviceSenoBright HD
The fundamental principles of operation, functionalities, specificationsand technological characteristics of SenoBright HD with NIRA are
identical to those of the Predicate device SenoBright HD.
The image processing of SenoBright HD is improved with the New
Image Recombination Algorithm by adding a local estimation of breast
thickness in the images recombination to account for the non-
uniformity of the breast thickness, and by compensating for potential
patient movement between the 2 CESM acquisitions (LE and HE).
Image Quality Performance Testing results demonstrated that the
image quality of SenoBright HD with NIRA performs similarly as its
legally marketed predicate device, SenoBright HD (K172404).
Determination of SubstantialEquivalence:NIRA for SenoBright HD successfully completed required design controltesting per GE Healthcare's quality management system. No
unexpected test results were obtained. The device is designed and will
be manufactured in compliance with the Quality System Regulations of
21CFR 820 and ISO 13485. The following quality assurance measures
were applied to the development of the system:
- Risk Analysis
- Design Reviews
- Software Development Lifecycle
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Performance testing (Verification)
- Safety testing (Verification)
- Simulated use testing (Validation)
The safety and performance of NIRA for SenoBright HD was
demonstrated through full verification testing and additional
performance testing such as:
1. A Non-clinical Performance Testing on phantoms, including
Image quality testing to demonstrate that SenoBright HD
with NIRA performs at least as well as the cleared device
SenoBright HD (K172404) and brings Image Quality
improvements. The performance testing also demonstrates
reduction of artifacts in case of either patient motion or
breast thickness non-uniformity to increased lesion visibility.
2. A clinical image evaluation performed by 3 independent
MQSA-qualified radiologists on 10 images with objective
criteria defined. It illustrated the clinical image acceptability
of images generated by SenoBright HD with NIRA.
3. A clinical image evaluation comparing SenoBright HD with NIRA to the predicate device SenoBright HD, performed by 3 independent MQSA-qualified radiologists on 50 clinical images, showing results including:The contrast uptake visibility assessed equivalent or better in 97% of the cases. The visibility of artifacts assessed equivalent or lower in 99% of the cases. The overall clinical image quality assessed superior in more than 98% of the cases. The results of verification and performance testing demonstrate the safe and effective use of SenoBright HD with NIRA .SenoBright HD with NIRA is substantially equivalent to the predicate device SenoBright HD (K172404) and there is no new question of safety and effectiveness that was raised.
Conclusion:The fundamental principles of operation, functionalities, specifications and technological characteristics of SenoBright HD with NIRA are identical to those of the Predicate device SenoBright HD.Furthermore, results of successful verification activities and additional bench performance testing and clinical image evaluations do not raise any new issue regarding the safety and effectiveness of the device.Based on this information, GE Healthcare believe that SenoBright HD with NIRA is substantially equivalent to its predicate device SenoBright HD (K172404).

{6}------------------------------------------------

GE Healthcare


Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font inside. There are decorative swirls around the letters, giving the logo a classic and recognizable appearance. The blue color is consistent throughout the logo.

510(k) Premarket Notification Submission

{7}------------------------------------------------

GE Healthcare

Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a circle. There are decorative flourishes around the circle, resembling water droplets or stylized leaves. The logo is presented in a blue color.

510(k) Premarket Notification Submission

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The letters and the frame are in a light blue color. There are also some white curved lines around the letters.

510(k) Premarket Notification Submission

§ 892.1715 Full-field digital mammography system.

(a)
Identification. A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”See § 892.1(e) for the availability of this guidance document.